Study on setup errors for different body carcinoma radiotherapy with image guidance in TOMO-HD

Wang Yinliang, Wang Huitao, Zheng Anping, Sun Xiaodong, Wang Dong, Liu Haojia, Qin Yongpeng
{"title":"Study on setup errors for different body carcinoma radiotherapy with image guidance in TOMO-HD","authors":"Wang Yinliang, Wang Huitao, Zheng Anping, Sun Xiaodong, Wang Dong, Liu Haojia, Qin Yongpeng","doi":"10.13491/J.ISSN.1004-714X.2021.03.015","DOIUrl":null,"url":null,"abstract":"Objective The literature study the setup errors of head and neck, thoracic, abdominal and pelvic\n tumors by megavoltage fan-beam CT based image guidance in TOMO-HD to provide the margin\n enlarging from clinic target volume (CTV) to planning target volume (PTV) in treatment\n planning system of TOMO-HD.\n Methods 103 patients with head and neck (30 patients), thoracic (42 patients), abdominal\n and pelvic (31 patients) carcinoma were enrolled. Megavoltage fan-beam CT based image\n guidance in tomotherapy-HD was used to acquire CT scan before every treatment. The\n left-right (X), superior-inferior (Y), anterior-posterior (Z) and rotation (Fy) setup\n errors of patients can be obtained from the tomography image automatically restructured\n by the system. Calculating the systematic error and the random error in the three\n dimensions and check whether the setup data accord with the normal distribution or\n not, then acquire the data expand in the three directions.\n Results According to 2 593 fan-beam CT scans, the shift errors (µ ± s) in X, Y, Z and Fy\n (rotation) of three study group were [(−0.31 ± 2.16) mm, (1.09 ± 3.56) mm, (2.36 ±\n 2.27) mm, (0.29 ± 0.96)°] (head and neck tumor), [(−0.98 ± 2.95) mm, (0.45 ± 6.86)\n mm, (3.79 ± 2.47) mm, (0.18 ± 0.60)°] (thoracic cancer) and [(−0.86 ± 2.85) mm, (−1.59\n ± 6.91) mm, (5.77 ± 2.40) mm, (0.20 ± 0.68)°] (abdominal and pelvic carcinoma). The\n systematic errors (Σ) and random errors (σ) in X, Y, Z dimensions of patients with\n head and neck, thoracic, abdominal and pelvic tumors were (1.06 mm and 1.84 mm), (1.93\n mm and 3.43 mm), (2.41 mm and 2.71 mm), (1.10 mm and 2.56 mm), (3.79 mm and 5.46 mm),\n (1.38 mm and 1.99 mm) and (1.39 mm and 0.87 mm), (4.98 mm and 5.69 mm), (1.19 mm and\n 2.05 mm), respectively.\n Conclusion It is recommended as a reference for image guidance in TOMO-HD according to the frequency\n distribution of setup errors, for patients with head and neck, chest and abdominal\n and pelvic tumors, the maximum range of motion in three dimensions are (5.00, 5.00,\n 5.00) mm, (6,63, 17.25, 16.00) mm and (6.49, 16.24, 13.60) mm.\n 摘要: 目的 利用 TOMO-HD 螺旋加速器的 MV 级扇形束 CT 研宄头颈部、胸部、腹盆部肿瘤在放疗中的摆位误差, 为 PTV 边界的外扩提供数据支持。\n 方法 采用 TOMO-HD 螺旋断层系统分别对 30 例头颈部、42 例胸部和 31 例腹盆 部肿瘤患者每次放疗前行 MV 级扇形束 CT 扫描, 系统自动重建图像并与治疗计划\n CT 图像进行配准, 获得患者左右 (X 轴)、头脚 (Y 轴)、腹背 (Z 轴)、沿Y轴方向的旋转 (Fy) 方向的摆位误差。计算出X、Y、Z方向的系统误差(Σ)\n 和随 机误差(σ)并对摆位误差进行正态性检验, 计算出 PTV 的各方向外扩边界。\n 结果 共行 2 593 次扫描, 头颈部、胸部、腹盆部患者在 X 轴、Y 轴、Z轴、Fy 方向 (沿 Y 轴方向的旋转) 的摆位误差 (µ ± s) 分别为 [(−0.31\n ± 2.16)mm、(1.09 ± 3.56)mm、(2.36 ± 2.27)mm、(0.29 ± 0.96)°], [(−0.98 ± 2.95)mm、(−0.45\n ± 6.86)mm、(3.79 ± 2.47)mm、(0.18 ± 0.60)°] 和 [(−0.86 ± 2.85)mm、(−1.59 ± 6.91)mm、(5.77\n ± 2.40)mm、(0.20 ± 0.68)°]。头颈部、胸部、腹盆部患者在 X 轴、Y 轴、Z轴方向的系统误差(Σ)和随机误差(σ)分别为 (1.06 mm 和\n 1.84 mm)、(1.93 mm 和 3.43 mm)、(2.41 mm 和 2.71 mm)、(1.10 mm 和 2.56 mm)、(3.79 mm 和 5.46\n mm)、(1.38 mm 和 1.99 mm) 和 (1.39 mm 和 0.87 mm)、(4.98 mm 和 5.69 mm)、(1.19 mm 和 2.05\n mm)。\n 结论 根据不同部位的摆位误差频数分布, 头颈部、胸部、腹盆部不同 部位肿瘤患者的靶区在三维方向上最大运动范围参考值为 (5.00, 5.00, 5.00)mm, (6.63,\n 17.25, 16.00)mm 和 (6.49, 16.24,13.60)mm, 推荐以此作为图像引导参考。","PeriodicalId":58844,"journal":{"name":"中国辐射卫生","volume":" ","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2021-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"中国辐射卫生","FirstCategoryId":"1087","ListUrlMain":"https://doi.org/10.13491/J.ISSN.1004-714X.2021.03.015","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Objective The literature study the setup errors of head and neck, thoracic, abdominal and pelvic tumors by megavoltage fan-beam CT based image guidance in TOMO-HD to provide the margin enlarging from clinic target volume (CTV) to planning target volume (PTV) in treatment planning system of TOMO-HD. Methods 103 patients with head and neck (30 patients), thoracic (42 patients), abdominal and pelvic (31 patients) carcinoma were enrolled. Megavoltage fan-beam CT based image guidance in tomotherapy-HD was used to acquire CT scan before every treatment. The left-right (X), superior-inferior (Y), anterior-posterior (Z) and rotation (Fy) setup errors of patients can be obtained from the tomography image automatically restructured by the system. Calculating the systematic error and the random error in the three dimensions and check whether the setup data accord with the normal distribution or not, then acquire the data expand in the three directions. Results According to 2 593 fan-beam CT scans, the shift errors (µ ± s) in X, Y, Z and Fy (rotation) of three study group were [(−0.31 ± 2.16) mm, (1.09 ± 3.56) mm, (2.36 ± 2.27) mm, (0.29 ± 0.96)°] (head and neck tumor), [(−0.98 ± 2.95) mm, (0.45 ± 6.86) mm, (3.79 ± 2.47) mm, (0.18 ± 0.60)°] (thoracic cancer) and [(−0.86 ± 2.85) mm, (−1.59 ± 6.91) mm, (5.77 ± 2.40) mm, (0.20 ± 0.68)°] (abdominal and pelvic carcinoma). The systematic errors (Σ) and random errors (σ) in X, Y, Z dimensions of patients with head and neck, thoracic, abdominal and pelvic tumors were (1.06 mm and 1.84 mm), (1.93 mm and 3.43 mm), (2.41 mm and 2.71 mm), (1.10 mm and 2.56 mm), (3.79 mm and 5.46 mm), (1.38 mm and 1.99 mm) and (1.39 mm and 0.87 mm), (4.98 mm and 5.69 mm), (1.19 mm and 2.05 mm), respectively. Conclusion It is recommended as a reference for image guidance in TOMO-HD according to the frequency distribution of setup errors, for patients with head and neck, chest and abdominal and pelvic tumors, the maximum range of motion in three dimensions are (5.00, 5.00, 5.00) mm, (6,63, 17.25, 16.00) mm and (6.49, 16.24, 13.60) mm. 摘要: 目的 利用 TOMO-HD 螺旋加速器的 MV 级扇形束 CT 研宄头颈部、胸部、腹盆部肿瘤在放疗中的摆位误差, 为 PTV 边界的外扩提供数据支持。 方法 采用 TOMO-HD 螺旋断层系统分别对 30 例头颈部、42 例胸部和 31 例腹盆 部肿瘤患者每次放疗前行 MV 级扇形束 CT 扫描, 系统自动重建图像并与治疗计划 CT 图像进行配准, 获得患者左右 (X 轴)、头脚 (Y 轴)、腹背 (Z 轴)、沿Y轴方向的旋转 (Fy) 方向的摆位误差。计算出X、Y、Z方向的系统误差(Σ) 和随 机误差(σ)并对摆位误差进行正态性检验, 计算出 PTV 的各方向外扩边界。 结果 共行 2 593 次扫描, 头颈部、胸部、腹盆部患者在 X 轴、Y 轴、Z轴、Fy 方向 (沿 Y 轴方向的旋转) 的摆位误差 (µ ± s) 分别为 [(−0.31 ± 2.16)mm、(1.09 ± 3.56)mm、(2.36 ± 2.27)mm、(0.29 ± 0.96)°], [(−0.98 ± 2.95)mm、(−0.45 ± 6.86)mm、(3.79 ± 2.47)mm、(0.18 ± 0.60)°] 和 [(−0.86 ± 2.85)mm、(−1.59 ± 6.91)mm、(5.77 ± 2.40)mm、(0.20 ± 0.68)°]。头颈部、胸部、腹盆部患者在 X 轴、Y 轴、Z轴方向的系统误差(Σ)和随机误差(σ)分别为 (1.06 mm 和 1.84 mm)、(1.93 mm 和 3.43 mm)、(2.41 mm 和 2.71 mm)、(1.10 mm 和 2.56 mm)、(3.79 mm 和 5.46 mm)、(1.38 mm 和 1.99 mm) 和 (1.39 mm 和 0.87 mm)、(4.98 mm 和 5.69 mm)、(1.19 mm 和 2.05 mm)。 结论 根据不同部位的摆位误差频数分布, 头颈部、胸部、腹盆部不同 部位肿瘤患者的靶区在三维方向上最大运动范围参考值为 (5.00, 5.00, 5.00)mm, (6.63, 17.25, 16.00)mm 和 (6.49, 16.24,13.60)mm, 推荐以此作为图像引导参考。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
TOMO-HD影像引导下不同体癌放疗设置误差的研究
目的研究基于巨压扇束CT的TOMO-HD影像引导对头颈部、胸部、腹部和盆腔肿瘤的设置误差,为TOMO-HD治疗计划系统提供从临床靶体积(CTV)到计划靶体积(PTV)的扩大边界。方法103例头颈部癌(30例)、胸椎癌(42例)、腹盆腔癌(31例)。在每次治疗前,采用基于超高电压扇束CT的图像引导技术获取CT扫描结果。从系统自动重构的断层图像中可获得患者的左右(X)、上下(Y)、前后(Z)和旋转(Fy)设置误差。计算三个维度的系统误差和随机误差,并检查设置的数据是否符合正态分布,然后获得三个方向上展开的数据。根据593扇形波束CT扫描结果,这种转变错误(µ±s)在X, Y, Z和财政年度(旋转)的三个研究小组(毫米(−0.31±2.16),(1.09±3.56)毫米,mm(2.36±2.27),(0.29±0.96)°)(头部和颈部肿瘤),[(−0.98±2.95)毫米,mm(0.45±6.86),(3.79±2.47)毫米,(0.18±0.60)°)(胸癌)和((−0.86±2.85)毫米,(−1.59±6.91)毫米,mm(5.77±2.40),(0.20±0.68)°)(腹部和骨盆癌)。头颈、胸腹、盆腔肿瘤患者X、Y、Z维度的系统误差(Σ)和随机误差(Σ)分别为(1.06 mm和1.84 mm)、(1.93 mm和3.43 mm)、(2.41 mm和2.71 mm)、(1.10 mm和2.56 mm)、(3.79 mm和5.46 mm)、(1.38 mm和1.99 mm)、(1.39 mm和0.87 mm)、(4.98 mm和5.69 mm)、(1.19 mm和2.05 mm)。结论建议作为参考图像指导TOMO-HD根据设置错误的频率分布,患者的头部和颈部,胸部和腹部和盆腔肿瘤,运动在三维空间的最大射程是mm(5.00, 5.00, 5.00),(63、17.25、16.00)毫米,(6.49,16.24,13.60)毫米。摘要:目的利用TOMO-HD螺旋加速器的MV级扇形束CT研宄头颈部,胸部、腹盆部肿瘤在放疗中的摆位误差,为PTV边界的外扩提供数据支持。方法采用TOMO-HD螺旋断层系统分别对30例头颈部,42例胸部和31例腹盆部肿瘤患者每次放疗前行MV级扇形束CT扫描,系统自动重建图像并与治疗计划CT图像进行配准,获得患者左右(X轴),头脚(Y轴),腹背(Z轴),沿Y轴方向的旋转(年度)方向的摆位误差。计算出X, Y, Z的方向的系统误差(Σ)和随机误差(σ)并对摆位误差进行正态性检验,计算出PTV的各方向外扩边界。结果共行593次扫描,头颈部,胸部、腹盆部患者在X轴、Y轴,Z轴,方财政年度向(沿Y轴方向的旋转)的摆位误差(µ±s)分别为[(−0.31±2.16)毫米(1.09±3.56)毫米,mm(2.36±2.27),(0.29±0.96)°],[(−0.98±2.95)毫米,(−0.45±6.86)毫米,mm(3.79±2.47),(0.18±0.60)°]和[(−0.86±2.85)毫米,(−1.59±6.91)毫米,mm(5.77±2.40),(0.20±0.68)°)。头颈部,胸部、腹盆部患者在X轴、Y轴、Z轴方向的系统误差(Σ)和随机误差(σ)分别为(1.06毫米和1.84毫米)(1.93毫米和3.43毫米)(2.41毫米和2.71毫米)(1.10毫米和2.56毫米)(3.79毫米和5.46毫米)(1.38毫米和1.99毫米)和(1.39毫米和0.87毫米)(4.98毫米和5.69毫米)(1.19毫米和2.05毫米)。结论根据不同部位的摆位误差频数分布,头颈部,胸部、腹盆部不同部位肿瘤患者的靶区在三维方向上最大运动范围参考值为(5.00,5.00,5.00)毫米,mm(6.63, 17.25, 16.00)和(6.49,16.24,13.60)毫米,推荐以此作为图像引导参考。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
0.80
自引率
0.00%
发文量
7142
期刊介绍: Chinese Journal of Radiological Health is one of the Source Journals for Chinese Scientific and Technical Papers and Citations and belongs to the series published by Chinese Preventive Medicine Association (CPMA). It is a national academic journal supervised by National Health Commission of the People’s Republic of China and co-sponsored by Institute of Radiation Medicine, Shandong Academy of Medical Sciences and CPMA, and is a professional academic journal publishing research findings and management experience in the field of radiological health, issued to the public in China and abroad. Under the guidance of the Communist Party of China and the national press and publication policies, the Journal actively publicizes the guidelines and policies of the Party and the state on health work, promotes the implementation of relevant laws, regulations and standards, and timely reports new achievements, new information, new methods and new products in the specialty, with the aim of organizing and promoting the academic communication of radiological health in China and improving the academic level of the specialty, and for the purpose of protecting the health of radiation workers and the public while promoting the extensive use of radioisotopes and radiation devices in the national economy. The main columns include Original Articles, Expert Comments, Experience Exchange, Standards and Guidelines, and Review Articles.
期刊最新文献
Analysis of gross radioactivity in drinking water around Tianwan Nuclear Power Plant from 2016 to 2018 Current research status of ionizing radiation bleeding syndrome The predictive value of MSCT imaging features on the pathological risk of gastrointestinal stromal tumors Analysis of quality control test results of medical electron linear accelerators in Guangxi Province 2017—2019 Analysis on the distribution status and concentration degree of radiological diagnosis and treatment resources in Beijing
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1