Comparison of the blood gas analyser Edan i15 Vet with the blood gas analyser Gem Premier 3000 for the analysis of blood gases, electrolytes, hemoglobin and hematocrit in lactating Holstein
{"title":"Comparison of the blood gas analyser Edan i15 Vet with the blood gas analyser Gem Premier 3000 for the analysis of blood gases, electrolytes, hemoglobin and hematocrit in lactating Holstein","authors":"M. Metin, K. Aksoy, Aytaç Pekmezci, A. Deniz","doi":"10.21521/mw.6728","DOIUrl":null,"url":null,"abstract":"A point-of-care blood gas analyser, Edan i15 Vet (EDAN), was compared with a benchtop blood gas analyser, GEM Premier 3000 (GEM). GEM and EDAN were used to analyse whole blood from 123 lactating Holsteins within one month of calving for blood gases, electrolytes, hematocrit and hemoglobin. EDAN and GEM showed significant linear correlations for blood gases, electrolytes, Hct and Hb. The 95% confidence intervals (CI) of intercept in Passing-Bablok regressions included zero in pCO2, bicarbonate, TCO2, BE ecf, BE B, hematocrit, and pH, but not for Na+, K+, pO2 and sO2. The CI of the slope included 1.0 for Na+, K+, pCO2, bicarbonate, TCO2, BE ecf, BE B, hematocrit, and pH, but not for pO2, sO2, and hemoglobin. The Bland-Altman plots between EDAN and GEM showed a bias of 1.4% for Na+, 2.4% for K+, –1.6% for pCO2, 3.0% for pO2, –5.3% for bicarbonate, 2.8% for SO2, –7.3% for TCO2, 10.4% for Hct, 21.2% for Hb, –25.1% for BE B and –38.5% for BE ecf. The biases in the analysis of certain estimated parameters were much higher (> 5%) than for measured parameters except for Hct. Parity did not correlate with blood gas parameters but blood pH correlated negatively with K+, pCO2 and positively with pO2, TCO2, sO2, bicarbonate, BE ecf and BE B. The postpartum time correlated positively with pCO2, TCO2, BE and bicarbonate and negatively with Hct and Hb values. Reference values (2.5-97.5% quartiles) were determined for each parameter. Conclusively, EDAN can be used interchangeably with GEM for the analysis of blood pH, K+, pCO2 and HCO3 – as they show acceptable moderate agreement. However, they did not agree for other parameters such as TCO2, BE ecf, BE B, Hb, Hct, sO2, Na+, and pO2, therefore, reference values for each parameter were set for GEM and EDAN.","PeriodicalId":49017,"journal":{"name":"Medycyna Weterynaryjna-Veterinary Medicine-Science and Practice","volume":null,"pages":null},"PeriodicalIF":0.4000,"publicationDate":"2023-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Medycyna Weterynaryjna-Veterinary Medicine-Science and Practice","FirstCategoryId":"97","ListUrlMain":"https://doi.org/10.21521/mw.6728","RegionNum":4,"RegionCategory":"农林科学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"VETERINARY SCIENCES","Score":null,"Total":0}
引用次数: 0
Abstract
A point-of-care blood gas analyser, Edan i15 Vet (EDAN), was compared with a benchtop blood gas analyser, GEM Premier 3000 (GEM). GEM and EDAN were used to analyse whole blood from 123 lactating Holsteins within one month of calving for blood gases, electrolytes, hematocrit and hemoglobin. EDAN and GEM showed significant linear correlations for blood gases, electrolytes, Hct and Hb. The 95% confidence intervals (CI) of intercept in Passing-Bablok regressions included zero in pCO2, bicarbonate, TCO2, BE ecf, BE B, hematocrit, and pH, but not for Na+, K+, pO2 and sO2. The CI of the slope included 1.0 for Na+, K+, pCO2, bicarbonate, TCO2, BE ecf, BE B, hematocrit, and pH, but not for pO2, sO2, and hemoglobin. The Bland-Altman plots between EDAN and GEM showed a bias of 1.4% for Na+, 2.4% for K+, –1.6% for pCO2, 3.0% for pO2, –5.3% for bicarbonate, 2.8% for SO2, –7.3% for TCO2, 10.4% for Hct, 21.2% for Hb, –25.1% for BE B and –38.5% for BE ecf. The biases in the analysis of certain estimated parameters were much higher (> 5%) than for measured parameters except for Hct. Parity did not correlate with blood gas parameters but blood pH correlated negatively with K+, pCO2 and positively with pO2, TCO2, sO2, bicarbonate, BE ecf and BE B. The postpartum time correlated positively with pCO2, TCO2, BE and bicarbonate and negatively with Hct and Hb values. Reference values (2.5-97.5% quartiles) were determined for each parameter. Conclusively, EDAN can be used interchangeably with GEM for the analysis of blood pH, K+, pCO2 and HCO3 – as they show acceptable moderate agreement. However, they did not agree for other parameters such as TCO2, BE ecf, BE B, Hb, Hct, sO2, Na+, and pO2, therefore, reference values for each parameter were set for GEM and EDAN.
期刊介绍:
"Medycyna Weterynaryjna" publishes various types of articles which are grouped in the following editorial categories: reviews, original studies, scientific and professional problems, the history of veterinary medicine, posthumous memoirs, as well as chronicles that briefly relate scientific advances and developments in the veterinary profession and medicine. The most important are the first two categories, which are published with short summaries in English. Moreover, from 2001 the editors of "Medycyna Weterynaryjna", bearing in mind market demands, has also started publishing entire works in English. Since 2008 the periodical has appeared in an electronic version. The following are available in this version: summaries of studies published from 1999 to 2005, full versions of all the studies published in the years 2006-2011 (in pdf files), and full versions of the English studies published in the current year (pdf). Only summaries of the remaining studies from the current year are available. In accordance with the principles accepted by the editors, the full versions of these texts will not be made available until next year.
All articles are evaluated twice by leading Polish scientists and professionals before they are considered for publication. For years now "Medycyna Weterynaryjna" has maintained a high standard thanks to this system. The review articles are actually succinct monographs dealing with specific scientific and professional problems that are based on the most recent findings. Original works have a particular value, since they present research carried out in Polish and international scientific centers.