In-vivo damage development in Vena Cava Filter: Study of a retrieved device

Anne Marie Lutz, T. Goswami
{"title":"In-vivo damage development in Vena Cava Filter: Study of a retrieved device","authors":"Anne Marie Lutz, T. Goswami","doi":"10.25082/mer.2022.01.001","DOIUrl":null,"url":null,"abstract":"This article was prepared from a project assigned in a graduate class, BME 7371, Failure Assessment of Medical Devices, taught at Wright State University by the senior author. The device was donated for the study which was successfully retrieved after 93 days in-vivo. Even-though the mechanical integrity of the device held in-tact, the microscopic observations revealed that the damage via corrosion and scratching took place on the surface of the device. The image reconstructed in 3D using image-J software to determine device roughness and depth-of-pits. Results presented in this paper show that damage starts developing in these devices after the implantation that resulted in premature failure in many cases as reported in the media and literature.","PeriodicalId":63081,"journal":{"name":"材料工程研究(英文)","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2022-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"材料工程研究(英文)","FirstCategoryId":"1087","ListUrlMain":"https://doi.org/10.25082/mer.2022.01.001","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

This article was prepared from a project assigned in a graduate class, BME 7371, Failure Assessment of Medical Devices, taught at Wright State University by the senior author. The device was donated for the study which was successfully retrieved after 93 days in-vivo. Even-though the mechanical integrity of the device held in-tact, the microscopic observations revealed that the damage via corrosion and scratching took place on the surface of the device. The image reconstructed in 3D using image-J software to determine device roughness and depth-of-pits. Results presented in this paper show that damage starts developing in these devices after the implantation that resulted in premature failure in many cases as reported in the media and literature.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
腔静脉过滤器的体内损伤发展:一种回收装置的研究
这篇文章是根据研究生班BME 7371,医疗器械失效评估的一个项目编写的,该课程由资深作者在赖特州立大学授课。该装置被捐赠用于该研究,并在体内使用93天后成功回收。尽管设备的机械完整性完好无损,但显微镜观察显示,设备表面发生了腐蚀和刮擦造成的损伤。利用image- j软件对图像进行三维重建,确定器件粗糙度和凹坑深度。本文的结果表明,在植入这些装置后,损伤开始发展,导致媒体和文献中报道的许多情况下过早失效。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
11
期刊最新文献
Revolutionizing energy storage: Overcoming challenges and unleashing the potential of next generation Lithium-ion battery technology Load-induced local phase transformation and modulus of shape memory alloys under spherical indentation by finite element method Polymer electrolyte design strategies for high-performance and safe lithium-ion batteries: Recent developments and future prospects Residual properties of silicone (MED-4719) lead with leads from retrieved devices Synthesis of an eco-friendly composite of palygorskite-gypsum associated ore using corn starch and waste biomass
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1