Evaluation of large volume SrI2(Eu) scintillator detectors

B. Sturm, N. Cherepy, O. Drury, P. Thelin, S. Fisher, A. Magyar, S. Payne, A. Burger, L. Boatner, J. O. Ramey, K. Shah, R. Hawrami
{"title":"Evaluation of large volume SrI2(Eu) scintillator detectors","authors":"B. Sturm, N. Cherepy, O. Drury, P. Thelin, S. Fisher, A. Magyar, S. Payne, A. Burger, L. Boatner, J. O. Ramey, K. Shah, R. Hawrami","doi":"10.1109/NSSMIC.2010.5874047","DOIUrl":null,"url":null,"abstract":"There is an ever increasing demand for gamma-ray detectors which can achieve good energy resolution, high detection efficiency, and room-temperature operation. We are working to address each of these requirements through the development of large volume SrI2(Eu) scintillator detectors. In this work, we have evaluated a variety of SrI2 crystals with volumes >10 cm3. The goal of this research was to examine the causes of energy resolution degradation for larger detectors and to determine what can be done to mitigate these effects. Testing both packaged and unpackaged detectors, we have consistently achieved better resolution with the packaged detectors. Using a collimated gamma-ray source, it was determined that better energy resolution for the packaged detectors is correlated with better light collection uniformity. A number of packaged detectors were fabricated and tested and the best spectroscopic performance was achieved for a 3% Eu doped crystal with an energy resolution of 2.93% FWHM at 662keV. Simulations of SrI2(Eu) crystals were also performed to better understand the light transport physics in scintillators and are reported. This study has important implications for the development of SrI2(Eu) detectors for national security purposes.","PeriodicalId":13048,"journal":{"name":"IEEE Nuclear Science Symposuim & Medical Imaging Conference","volume":"152 1","pages":"1607-1611"},"PeriodicalIF":0.0000,"publicationDate":"2010-11-18","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"21","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"IEEE Nuclear Science Symposuim & Medical Imaging Conference","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/NSSMIC.2010.5874047","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 21

Abstract

There is an ever increasing demand for gamma-ray detectors which can achieve good energy resolution, high detection efficiency, and room-temperature operation. We are working to address each of these requirements through the development of large volume SrI2(Eu) scintillator detectors. In this work, we have evaluated a variety of SrI2 crystals with volumes >10 cm3. The goal of this research was to examine the causes of energy resolution degradation for larger detectors and to determine what can be done to mitigate these effects. Testing both packaged and unpackaged detectors, we have consistently achieved better resolution with the packaged detectors. Using a collimated gamma-ray source, it was determined that better energy resolution for the packaged detectors is correlated with better light collection uniformity. A number of packaged detectors were fabricated and tested and the best spectroscopic performance was achieved for a 3% Eu doped crystal with an energy resolution of 2.93% FWHM at 662keV. Simulations of SrI2(Eu) crystals were also performed to better understand the light transport physics in scintillators and are reported. This study has important implications for the development of SrI2(Eu) detectors for national security purposes.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
大体积SrI2(Eu)闪烁体探测器的评价
对能量分辨率高、探测效率高、能在室温下工作的伽玛射线探测器的需求日益增长。我们正在努力通过开发大体积SrI2(Eu)闪烁体探测器来满足这些要求。在这项工作中,我们评估了各种体积为bbb10 cm3的SrI2晶体。这项研究的目的是检查大型探测器能量分辨率下降的原因,并确定可以做些什么来减轻这些影响。测试了封装和未封装的检测器,我们一致地使用封装检测器获得了更好的分辨率。使用准直伽玛射线源,确定了封装探测器的更好的能量分辨率与更好的光收集均匀性相关。制作并测试了多个封装的探测器,在662keV下,3% Eu掺杂晶体的能量分辨率为2.93% FWHM,获得了最佳的光谱性能。为了更好地理解闪烁体中的光输运物理,还对SrI2(Eu)晶体进行了模拟。该研究对用于国家安全目的的SrI2(Eu)探测器的开发具有重要意义。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
CaF2(Eu): An “Old” scintillator revisited Evaluation of the accuracy and robustness of a motion correction algorithm for PET using a novel phantom approach The performance of GridPix detectors Applying the neutron scatter camera to treaty verification and warhead monitoring Development of an 8-channel time based readout ASIC for PET applications
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1