{"title":"Comparative study of commercial pedicle screw based on maximum pull out force","authors":"Radhi Maladzi, S. Supriadi","doi":"10.1063/1.5139337","DOIUrl":null,"url":null,"abstract":"Pedicle screw has a role in stabilizing vertebral bone motion and resisting all types of load such as axially load. If pedicle screw cannot maintain axially load that happened, pull out failure can be occurred. Experimental pedicle screw pull out testing is referring to American Society for Testing and Materials (ASTM) F543. Each pedicle screw displaced at 5 mm/min until pedicle screw pulled from the test block material. Note the displacement graph between load and displacement, and the highest peak means maximum pull out force of the specimen. The purpose of this study to determine the maximum axially force of pedicle screws can bear. Commercial 1 pedicle screw shows higher pull out maximum force with 1497 N than commercial two pedicle screw with 469 N.Pedicle screw has a role in stabilizing vertebral bone motion and resisting all types of load such as axially load. If pedicle screw cannot maintain axially load that happened, pull out failure can be occurred. Experimental pedicle screw pull out testing is referring to American Society for Testing and Materials (ASTM) F543. Each pedicle screw displaced at 5 mm/min until pedicle screw pulled from the test block material. Note the displacement graph between load and displacement, and the highest peak means maximum pull out force of the specimen. The purpose of this study to determine the maximum axially force of pedicle screws can bear. Commercial 1 pedicle screw shows higher pull out maximum force with 1497 N than commercial two pedicle screw with 469 N.","PeriodicalId":22239,"journal":{"name":"THE 4TH BIOMEDICAL ENGINEERING’S RECENT PROGRESS IN BIOMATERIALS, DRUGS DEVELOPMENT, HEALTH, AND MEDICAL DEVICES: Proceedings of the International Symposium of Biomedical Engineering (ISBE) 2019","volume":"5 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2019-12-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"THE 4TH BIOMEDICAL ENGINEERING’S RECENT PROGRESS IN BIOMATERIALS, DRUGS DEVELOPMENT, HEALTH, AND MEDICAL DEVICES: Proceedings of the International Symposium of Biomedical Engineering (ISBE) 2019","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1063/1.5139337","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1

Abstract

Pedicle screw has a role in stabilizing vertebral bone motion and resisting all types of load such as axially load. If pedicle screw cannot maintain axially load that happened, pull out failure can be occurred. Experimental pedicle screw pull out testing is referring to American Society for Testing and Materials (ASTM) F543. Each pedicle screw displaced at 5 mm/min until pedicle screw pulled from the test block material. Note the displacement graph between load and displacement, and the highest peak means maximum pull out force of the specimen. The purpose of this study to determine the maximum axially force of pedicle screws can bear. Commercial 1 pedicle screw shows higher pull out maximum force with 1497 N than commercial two pedicle screw with 469 N.Pedicle screw has a role in stabilizing vertebral bone motion and resisting all types of load such as axially load. If pedicle screw cannot maintain axially load that happened, pull out failure can be occurred. Experimental pedicle screw pull out testing is referring to American Society for Testing and Materials (ASTM) F543. Each pedicle screw displaced at 5 mm/min until pedicle screw pulled from the test block material. Note the displacement graph between load and displacement, and the highest peak means maximum pull out force of the specimen. The purpose of this study to determine the maximum axially force of pedicle screws can bear. Commercial 1 pedicle screw shows higher pull out maximum force with 1497 N than commercial two pedicle screw with 469 N.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
基于最大拔出力的商用椎弓根螺钉的比较研究
椎弓根螺钉具有稳定椎体骨运动和抵抗各种载荷(如轴向载荷)的作用。如果椎弓根螺钉不能维持已发生的轴向载荷,则可能发生拔出失败。实验性椎弓根螺钉拔出试验参照美国材料试验学会(ASTM) F543。每个椎弓根螺钉以5mm /min的速度移位,直到椎弓根螺钉从试验块材料中拔出。注意荷载与位移之间的位移曲线图,峰值表示试件的最大拉拔力。本研究的目的是确定椎弓根螺钉所能承受的最大轴向力。商用1号椎弓根螺钉在1497 N时的最大拔出力高于商用2号椎弓根螺钉在469 N时的最大拔出力。椎弓根螺钉具有稳定椎体骨运动和抵抗轴向载荷等各种载荷的作用。如果椎弓根螺钉不能维持已发生的轴向载荷,则可能发生拔出失败。实验性椎弓根螺钉拔出试验参照美国材料试验学会(ASTM) F543。每个椎弓根螺钉以5mm /min的速度移位,直到椎弓根螺钉从试验块材料中拔出。注意荷载与位移之间的位移曲线图,峰值表示试件的最大拉拔力。本研究的目的是确定椎弓根螺钉所能承受的最大轴向力。商用1号椎弓根螺钉1497 N的最大拔出力高于商用2号椎弓根螺钉469 N的最大拔出力。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Development of tools utilization monitoring system on labor-intensive manufacturing industries Topology optimization on geometry of 3D printed “Impulse RC Alien 4 Inch” racing quadcopter frame with polylactic acid material Design and analysis of regenerative shock absorber using ball screw mechanism for vehicle suspension Effect of Sundanese music on daytime sleep quality based on EEG signal Static analysis of an energy storage and return (ESAR) prosthetic foot
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1