{"title":"Study on the Instrument for Testing Radiofrequency Ablation Catheter","authors":"Lei Zhang, Wenbo Lu, Jianjun Sun, Chongchong Ai, Zechen Lin, Naishui Han, Longfei Li, Yufei Jia","doi":"10.32629/jcmr.v4i2.1234","DOIUrl":null,"url":null,"abstract":"Objective — To design a detection instrument for detecting the quality of radiofrequency ablation electrode, which can quickly detect the physical properties of radiofrequency ablation electrode. Methods — The DC constant current source is introduced to measure the small resistance, test whether the thermocouple is well connected, and measure the resistance between the positive and negative wires of the thermocouple at the plug (1-core and 2-core) to determine whether the thermocouple is well connected. The resistance of the good product is about 40Ω. Detect whether the inner tube and outer tube of RF ablation are in good physical contact through resistance, and judge whether there is physical contact between the inner tube and outer tube by detecting the resistance between the positive conductor of thermocouple (2 cores) and RF connecting wire (3 cores). When in good contact, the resistance is about 0.5 Ω; Measure the short circuit between 4~8 cores of the plug and judge whether the code is correct. The measurement method of pipeline absolute tightness is adopted to detect the change of pipeline pressure. Results — The developed radiofrequency ablation electrode detector can accurately measure the resistance, circuit and pipeline tightness of radiofrequency ablation, and can meet the requirements of detection accuracy. Conclusion — The quality of the prototype is stable and can meet the needs of users.","PeriodicalId":15431,"journal":{"name":"Journal of Clinical Medicine Research","volume":"43 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2023-07-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Clinical Medicine Research","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.32629/jcmr.v4i2.1234","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
Objective — To design a detection instrument for detecting the quality of radiofrequency ablation electrode, which can quickly detect the physical properties of radiofrequency ablation electrode. Methods — The DC constant current source is introduced to measure the small resistance, test whether the thermocouple is well connected, and measure the resistance between the positive and negative wires of the thermocouple at the plug (1-core and 2-core) to determine whether the thermocouple is well connected. The resistance of the good product is about 40Ω. Detect whether the inner tube and outer tube of RF ablation are in good physical contact through resistance, and judge whether there is physical contact between the inner tube and outer tube by detecting the resistance between the positive conductor of thermocouple (2 cores) and RF connecting wire (3 cores). When in good contact, the resistance is about 0.5 Ω; Measure the short circuit between 4~8 cores of the plug and judge whether the code is correct. The measurement method of pipeline absolute tightness is adopted to detect the change of pipeline pressure. Results — The developed radiofrequency ablation electrode detector can accurately measure the resistance, circuit and pipeline tightness of radiofrequency ablation, and can meet the requirements of detection accuracy. Conclusion — The quality of the prototype is stable and can meet the needs of users.