F A Alnaimat, H A Owida, Jamal I Al-Nabulsi, B Al-Naami, A Al-Ghraibah, M Al-Ayyad, Muneera Altayeb, Ashraf Al Sharah
{"title":"硅胶假指关节的设计修改与评价。","authors":"F A Alnaimat, H A Owida, Jamal I Al-Nabulsi, B Al-Naami, A Al-Ghraibah, M Al-Ayyad, Muneera Altayeb, Ashraf Al Sharah","doi":"10.3233/BME-222527","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>There are many reasons that could lead to finger joint arthroplasty, and the most familiar reason is osteoarthritis. Silicone finger joint are the most commonly used implants. However, these implants might fracture with time and cause wear which will lead to chronic inflammation and synovitis for the patient and then implant failure.</p><p><strong>Objective: </strong>The aim of this study is to improve the design of the silicone finger joint and simulate the different designs using finite element analysis (FEA) simulation.</p><p><strong>Method: </strong>Three different designs were drawn and FEA has been used in this study using Solidworks software. The first design is the silicone finger joint design without any modification, the second one is modified design with added ribs to the junction of distal stem and hinge and the third design was added filler material inside the body of the artificial joint. An axial force with 625 N that was applied on the upper part of the distal stem which is nearly represents the maximum value of the grip strength for normal males.</p><p><strong>Results: </strong>The results showed improvement on the design in which the concentrated stress at the junction of the distal stem and hinge of the design was distributed. In addition, the Von Mises stress was stable for the modified design with added ribs and the added filler material designs after 15°.</p><p><strong>Conclusion: </strong>The design modification could improve the stress distribution and stability of the artificial finger joint and increase the lifetime expectancy of these implants.</p>","PeriodicalId":9109,"journal":{"name":"Bio-medical materials and engineering","volume":" ","pages":"481-490"},"PeriodicalIF":1.0000,"publicationDate":"2023-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Design modifications and evaluation of the silicone artificial finger joints.\",\"authors\":\"F A Alnaimat, H A Owida, Jamal I Al-Nabulsi, B Al-Naami, A Al-Ghraibah, M Al-Ayyad, Muneera Altayeb, Ashraf Al Sharah\",\"doi\":\"10.3233/BME-222527\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>There are many reasons that could lead to finger joint arthroplasty, and the most familiar reason is osteoarthritis. Silicone finger joint are the most commonly used implants. However, these implants might fracture with time and cause wear which will lead to chronic inflammation and synovitis for the patient and then implant failure.</p><p><strong>Objective: </strong>The aim of this study is to improve the design of the silicone finger joint and simulate the different designs using finite element analysis (FEA) simulation.</p><p><strong>Method: </strong>Three different designs were drawn and FEA has been used in this study using Solidworks software. The first design is the silicone finger joint design without any modification, the second one is modified design with added ribs to the junction of distal stem and hinge and the third design was added filler material inside the body of the artificial joint. An axial force with 625 N that was applied on the upper part of the distal stem which is nearly represents the maximum value of the grip strength for normal males.</p><p><strong>Results: </strong>The results showed improvement on the design in which the concentrated stress at the junction of the distal stem and hinge of the design was distributed. In addition, the Von Mises stress was stable for the modified design with added ribs and the added filler material designs after 15°.</p><p><strong>Conclusion: </strong>The design modification could improve the stress distribution and stability of the artificial finger joint and increase the lifetime expectancy of these implants.</p>\",\"PeriodicalId\":9109,\"journal\":{\"name\":\"Bio-medical materials and engineering\",\"volume\":\" \",\"pages\":\"481-490\"},\"PeriodicalIF\":1.0000,\"publicationDate\":\"2023-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Bio-medical materials and engineering\",\"FirstCategoryId\":\"5\",\"ListUrlMain\":\"https://doi.org/10.3233/BME-222527\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"ENGINEERING, BIOMEDICAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Bio-medical materials and engineering","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.3233/BME-222527","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
Design modifications and evaluation of the silicone artificial finger joints.
Background: There are many reasons that could lead to finger joint arthroplasty, and the most familiar reason is osteoarthritis. Silicone finger joint are the most commonly used implants. However, these implants might fracture with time and cause wear which will lead to chronic inflammation and synovitis for the patient and then implant failure.
Objective: The aim of this study is to improve the design of the silicone finger joint and simulate the different designs using finite element analysis (FEA) simulation.
Method: Three different designs were drawn and FEA has been used in this study using Solidworks software. The first design is the silicone finger joint design without any modification, the second one is modified design with added ribs to the junction of distal stem and hinge and the third design was added filler material inside the body of the artificial joint. An axial force with 625 N that was applied on the upper part of the distal stem which is nearly represents the maximum value of the grip strength for normal males.
Results: The results showed improvement on the design in which the concentrated stress at the junction of the distal stem and hinge of the design was distributed. In addition, the Von Mises stress was stable for the modified design with added ribs and the added filler material designs after 15°.
Conclusion: The design modification could improve the stress distribution and stability of the artificial finger joint and increase the lifetime expectancy of these implants.
期刊介绍:
The aim of Bio-Medical Materials and Engineering is to promote the welfare of humans and to help them keep healthy. This international journal is an interdisciplinary journal that publishes original research papers, review articles and brief notes on materials and engineering for biological and medical systems. Articles in this peer-reviewed journal cover a wide range of topics, including, but not limited to: Engineering as applied to improving diagnosis, therapy, and prevention of disease and injury, and better substitutes for damaged or disabled human organs; Studies of biomaterial interactions with the human body, bio-compatibility, interfacial and interaction problems; Biomechanical behavior under biological and/or medical conditions; Mechanical and biological properties of membrane biomaterials; Cellular and tissue engineering, physiological, biophysical, biochemical bioengineering aspects; Implant failure fields and degradation of implants. Biomimetics engineering and materials including system analysis as supporter for aged people and as rehabilitation; Bioengineering and materials technology as applied to the decontamination against environmental problems; Biosensors, bioreactors, bioprocess instrumentation and control system; Application to food engineering; Standardization problems on biomaterials and related products; Assessment of reliability and safety of biomedical materials and man-machine systems; and Product liability of biomaterials and related products.