Elliot Greenwald, Cheng Chen, N. Thakor, C. Maier, G. Cauwenberghs
{"title":"带有片上DAC校准和电荷平衡的CMOS神经刺激器","authors":"Elliot Greenwald, Cheng Chen, N. Thakor, C. Maier, G. Cauwenberghs","doi":"10.1109/BioCAS.2013.6679646","DOIUrl":null,"url":null,"abstract":"A multi-channel biphasic neural stimulator with on-chip DAC calibration and current matching capabilities is presented. Each channel consists of two sub-binary radix DACs, for the anodic and cathodic stimulation phases, and a wideswing high output impedance current source and sink. A single integrator is shared among channels and serves to calibrate DAC coefficients and to closely match the anodic and cathodic stimulation phases. After calibration, the differential non-linearity is bounded between +/- 0.5 LSBs at 8-bit resolution, and the two stimulation phases can be matched to better than 50 nA. We demonstrate operation with stimulation through a tungsten microelectrode in saline, and show stimulator induced modulation of neural activity in the cortex of a rat. Our novel architecture allows for blind self-calibration, amenable to implantable neural interfaces.","PeriodicalId":344317,"journal":{"name":"2013 IEEE Biomedical Circuits and Systems Conference (BioCAS)","volume":"4 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2013-12-12","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"11","resultStr":"{\"title\":\"A CMOS neurostimulator with on-chip DAC calibration and charge balancing\",\"authors\":\"Elliot Greenwald, Cheng Chen, N. Thakor, C. Maier, G. Cauwenberghs\",\"doi\":\"10.1109/BioCAS.2013.6679646\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"A multi-channel biphasic neural stimulator with on-chip DAC calibration and current matching capabilities is presented. Each channel consists of two sub-binary radix DACs, for the anodic and cathodic stimulation phases, and a wideswing high output impedance current source and sink. A single integrator is shared among channels and serves to calibrate DAC coefficients and to closely match the anodic and cathodic stimulation phases. After calibration, the differential non-linearity is bounded between +/- 0.5 LSBs at 8-bit resolution, and the two stimulation phases can be matched to better than 50 nA. We demonstrate operation with stimulation through a tungsten microelectrode in saline, and show stimulator induced modulation of neural activity in the cortex of a rat. Our novel architecture allows for blind self-calibration, amenable to implantable neural interfaces.\",\"PeriodicalId\":344317,\"journal\":{\"name\":\"2013 IEEE Biomedical Circuits and Systems Conference (BioCAS)\",\"volume\":\"4 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2013-12-12\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"11\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"2013 IEEE Biomedical Circuits and Systems Conference (BioCAS)\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1109/BioCAS.2013.6679646\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"2013 IEEE Biomedical Circuits and Systems Conference (BioCAS)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/BioCAS.2013.6679646","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
A CMOS neurostimulator with on-chip DAC calibration and charge balancing
A multi-channel biphasic neural stimulator with on-chip DAC calibration and current matching capabilities is presented. Each channel consists of two sub-binary radix DACs, for the anodic and cathodic stimulation phases, and a wideswing high output impedance current source and sink. A single integrator is shared among channels and serves to calibrate DAC coefficients and to closely match the anodic and cathodic stimulation phases. After calibration, the differential non-linearity is bounded between +/- 0.5 LSBs at 8-bit resolution, and the two stimulation phases can be matched to better than 50 nA. We demonstrate operation with stimulation through a tungsten microelectrode in saline, and show stimulator induced modulation of neural activity in the cortex of a rat. Our novel architecture allows for blind self-calibration, amenable to implantable neural interfaces.