基于梯形复合椭球谐振光声单元和偏最小二乘法的高灵敏度和超低浓度范围光声光谱技术

IF 7.1 1区 医学 Q1 ENGINEERING, BIOMEDICAL Photoacoustics Pub Date : 2023-12-30 DOI:10.1016/j.pacs.2023.100583
Qiaoyun Wang , Shunyuan Xu , Ziheng Zhu , Jilong Wang , Xin Zou , Chu Zhang , Qiang Liu
{"title":"基于梯形复合椭球谐振光声单元和偏最小二乘法的高灵敏度和超低浓度范围光声光谱技术","authors":"Qiaoyun Wang ,&nbsp;Shunyuan Xu ,&nbsp;Ziheng Zhu ,&nbsp;Jilong Wang ,&nbsp;Xin Zou ,&nbsp;Chu Zhang ,&nbsp;Qiang Liu","doi":"10.1016/j.pacs.2023.100583","DOIUrl":null,"url":null,"abstract":"<div><p>A high sensitivity and ultra-low concentration range photoacoustic spectroscopy (PAS) gas detection system, which was based on a novel trapezoid compound ellipsoid resonant photoacoustic cell (TCER-PAC) and partial least square (PLS), was proposed to detect acetylene (C<sub>2</sub>H<sub>2</sub>) gas. In the concentration range of 0.5 ppm ∼ 10.0 ppm, the limit of detection (LOD) values of TCER-PAC-based PAS system without data processing was 66.4 ppb, which was lower than that of the traditional trapezoid compound cylindrical resonant photoacoustic cell (TCCR-PAC). The experimental results indicated that the TCER-PAC had higher sensitivity than of TCCR-PAC. Within the concentration range of 12.5 ppb ∼ 125.0 ppb, the LOD and limit of quantification (LOQ) of TCER-PAC-based PAS system combined with PLS regression algorithm were 1.1 ppb and 3.7 ppb, respectively. The results showed that higher detection sensitivity and lower LOD were obtained by PAS system with TCER-PAC and PLS than that of TCCR-PAC-based PAS system.</p></div>","PeriodicalId":56025,"journal":{"name":"Photoacoustics","volume":"35 ","pages":"Article 100583"},"PeriodicalIF":7.1000,"publicationDate":"2023-12-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S2213597923001362/pdfft?md5=43741e7d877ecd3cb000992aa09cafd8&pid=1-s2.0-S2213597923001362-main.pdf","citationCount":"0","resultStr":"{\"title\":\"High sensitivity and ultra-low concentration range photoacoustic spectroscopy based on trapezoid compound ellipsoid resonant photoacoustic cell and partial least square\",\"authors\":\"Qiaoyun Wang ,&nbsp;Shunyuan Xu ,&nbsp;Ziheng Zhu ,&nbsp;Jilong Wang ,&nbsp;Xin Zou ,&nbsp;Chu Zhang ,&nbsp;Qiang Liu\",\"doi\":\"10.1016/j.pacs.2023.100583\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><p>A high sensitivity and ultra-low concentration range photoacoustic spectroscopy (PAS) gas detection system, which was based on a novel trapezoid compound ellipsoid resonant photoacoustic cell (TCER-PAC) and partial least square (PLS), was proposed to detect acetylene (C<sub>2</sub>H<sub>2</sub>) gas. In the concentration range of 0.5 ppm ∼ 10.0 ppm, the limit of detection (LOD) values of TCER-PAC-based PAS system without data processing was 66.4 ppb, which was lower than that of the traditional trapezoid compound cylindrical resonant photoacoustic cell (TCCR-PAC). The experimental results indicated that the TCER-PAC had higher sensitivity than of TCCR-PAC. Within the concentration range of 12.5 ppb ∼ 125.0 ppb, the LOD and limit of quantification (LOQ) of TCER-PAC-based PAS system combined with PLS regression algorithm were 1.1 ppb and 3.7 ppb, respectively. The results showed that higher detection sensitivity and lower LOD were obtained by PAS system with TCER-PAC and PLS than that of TCCR-PAC-based PAS system.</p></div>\",\"PeriodicalId\":56025,\"journal\":{\"name\":\"Photoacoustics\",\"volume\":\"35 \",\"pages\":\"Article 100583\"},\"PeriodicalIF\":7.1000,\"publicationDate\":\"2023-12-30\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.sciencedirect.com/science/article/pii/S2213597923001362/pdfft?md5=43741e7d877ecd3cb000992aa09cafd8&pid=1-s2.0-S2213597923001362-main.pdf\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Photoacoustics\",\"FirstCategoryId\":\"5\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S2213597923001362\",\"RegionNum\":1,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"ENGINEERING, BIOMEDICAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Photoacoustics","FirstCategoryId":"5","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2213597923001362","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0

摘要

基于新型梯形复合椭球谐振光声池(TCER-PAC)和偏最小二乘法(PLS)的高灵敏度和超低浓度范围光声谱(PAS)气体检测系统被提出来用于检测乙炔(C2H2)气体。在 0.5 ppm ~ 10.0 ppm 的浓度范围内,基于 TCER-PAC 的 PAS 系统在不进行数据处理的情况下的检出限(LOD)值为 66.4 ppb,低于传统的梯形复合圆柱谐振光声电池(TCCR-PAC)。实验结果表明,TCER-PAC 的灵敏度高于 TCCR-PAC。在 12.5 ppb ∼ 125.0 ppb 的浓度范围内,基于 TCER-PAC 的 PAS 系统结合 PLS 回归算法的检测灵敏度(LOD)和定量限(LOQ)分别为 1.1 ppb 和 3.7 ppb。结果表明,与基于 TCCR-PAC 的 PAS 系统相比,基于 TCER-PAC 和 PLS 的 PAS 系统具有更高的检测灵敏度和更低的 LOD。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
High sensitivity and ultra-low concentration range photoacoustic spectroscopy based on trapezoid compound ellipsoid resonant photoacoustic cell and partial least square

A high sensitivity and ultra-low concentration range photoacoustic spectroscopy (PAS) gas detection system, which was based on a novel trapezoid compound ellipsoid resonant photoacoustic cell (TCER-PAC) and partial least square (PLS), was proposed to detect acetylene (C2H2) gas. In the concentration range of 0.5 ppm ∼ 10.0 ppm, the limit of detection (LOD) values of TCER-PAC-based PAS system without data processing was 66.4 ppb, which was lower than that of the traditional trapezoid compound cylindrical resonant photoacoustic cell (TCCR-PAC). The experimental results indicated that the TCER-PAC had higher sensitivity than of TCCR-PAC. Within the concentration range of 12.5 ppb ∼ 125.0 ppb, the LOD and limit of quantification (LOQ) of TCER-PAC-based PAS system combined with PLS regression algorithm were 1.1 ppb and 3.7 ppb, respectively. The results showed that higher detection sensitivity and lower LOD were obtained by PAS system with TCER-PAC and PLS than that of TCCR-PAC-based PAS system.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Photoacoustics
Photoacoustics Physics and Astronomy-Atomic and Molecular Physics, and Optics
CiteScore
11.40
自引率
16.50%
发文量
96
审稿时长
53 days
期刊介绍: The open access Photoacoustics journal (PACS) aims to publish original research and review contributions in the field of photoacoustics-optoacoustics-thermoacoustics. This field utilizes acoustical and ultrasonic phenomena excited by electromagnetic radiation for the detection, visualization, and characterization of various materials and biological tissues, including living organisms. Recent advancements in laser technologies, ultrasound detection approaches, inverse theory, and fast reconstruction algorithms have greatly supported the rapid progress in this field. The unique contrast provided by molecular absorption in photoacoustic-optoacoustic-thermoacoustic methods has allowed for addressing unmet biological and medical needs such as pre-clinical research, clinical imaging of vasculature, tissue and disease physiology, drug efficacy, surgery guidance, and therapy monitoring. Applications of this field encompass a wide range of medical imaging and sensing applications, including cancer, vascular diseases, brain neurophysiology, ophthalmology, and diabetes. Moreover, photoacoustics-optoacoustics-thermoacoustics is a multidisciplinary field, with contributions from chemistry and nanotechnology, where novel materials such as biodegradable nanoparticles, organic dyes, targeted agents, theranostic probes, and genetically expressed markers are being actively developed. These advanced materials have significantly improved the signal-to-noise ratio and tissue contrast in photoacoustic methods.
期刊最新文献
Quantitative volumetric photoacoustic assessment of vasoconstriction by topical corticosteroid application in mice skin Spiral volumetric optoacoustic and ultrasound (SVOPUS) tomography of mice Miniature optical fiber photoacoustic spectroscopy gas sensor based on a 3D micro-printed planar-spiral spring optomechanical resonator Calibration-free infrared absorption spectroscopy using cantilever-enhanced photoacoustic detection of the optical power Application of multispectral optoacoustic tomography for lower limb musculoskeletal sports injuries in adults
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1