{"title":"从水的三相点到铟的凝固点的 ITS-90 子范围内的新插值公式","authors":"Kang Lan, Lide Fang, Ying Wang, Zhiru Kang, Suli Sang","doi":"10.1088/1681-7575/ad1bfb","DOIUrl":null,"url":null,"abstract":"\n For all the subranges above 0.01 ℃, other than the interpolation between the triple point of water (TPW) to the melting point of gallium, temperatures on the ITS-90 are specified without the melting point of gallium. This study suggests a new interpolation equation in the subrange from the triple point of water to the freezing point of indium which is a polynomial of fractional third order with two coefficients to be determined by the measuring values of standard platinum resistance thermometers (SPRTs)’ resistance ratios at the melting point of gallium (WGa) and at the freezing point of indium (WIn). A sample set of 30 SPRTs from various countries and laboratories is used to quantify the comparisons between this new interpolation and the ITS-90. Analyses show that the reproducibility is significantly improved compared with the ITS-90 in the same subrange. The peak value of overlapping subrange inconsistencies (SRI) from TPW to the melting point of gallium is reduced from 0.91 mK to 0.21 mK. Both the mean and standard deviation of the SRI peak values for the new equation decrease by approximately a factor of four compared to those of the ITS-90 (mean decreasing form 0.20 mK to 0.05 mK and standard deviation decreasing form 0.32 mK to 0.07 mK). Type 3 non-uniqueness (NU3) determinations of the same subrange have also been looked into with a data set of four SPRTs compared in thermostatic baths using a copper block and the results show that NU3 scales down from a range -0.15 mK to 0.25 mK of the ITS-90 to -0.10 mK to 0.15 mK of the new interpolation equation. Propagation of Uncertainties investigation shows that the new equation inflates merely 7% the measurement uncertainties of fixed points in a relatively short interval from 30 ℃ to 60 ℃.","PeriodicalId":18444,"journal":{"name":"Metrologia","volume":"36 22","pages":""},"PeriodicalIF":2.1000,"publicationDate":"2024-01-08","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"A New Interpolation Equation in the ITS-90 Subrange from the Triple Point of Water to the Freezing Point of Indium\",\"authors\":\"Kang Lan, Lide Fang, Ying Wang, Zhiru Kang, Suli Sang\",\"doi\":\"10.1088/1681-7575/ad1bfb\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"\\n For all the subranges above 0.01 ℃, other than the interpolation between the triple point of water (TPW) to the melting point of gallium, temperatures on the ITS-90 are specified without the melting point of gallium. This study suggests a new interpolation equation in the subrange from the triple point of water to the freezing point of indium which is a polynomial of fractional third order with two coefficients to be determined by the measuring values of standard platinum resistance thermometers (SPRTs)’ resistance ratios at the melting point of gallium (WGa) and at the freezing point of indium (WIn). A sample set of 30 SPRTs from various countries and laboratories is used to quantify the comparisons between this new interpolation and the ITS-90. Analyses show that the reproducibility is significantly improved compared with the ITS-90 in the same subrange. The peak value of overlapping subrange inconsistencies (SRI) from TPW to the melting point of gallium is reduced from 0.91 mK to 0.21 mK. Both the mean and standard deviation of the SRI peak values for the new equation decrease by approximately a factor of four compared to those of the ITS-90 (mean decreasing form 0.20 mK to 0.05 mK and standard deviation decreasing form 0.32 mK to 0.07 mK). Type 3 non-uniqueness (NU3) determinations of the same subrange have also been looked into with a data set of four SPRTs compared in thermostatic baths using a copper block and the results show that NU3 scales down from a range -0.15 mK to 0.25 mK of the ITS-90 to -0.10 mK to 0.15 mK of the new interpolation equation. Propagation of Uncertainties investigation shows that the new equation inflates merely 7% the measurement uncertainties of fixed points in a relatively short interval from 30 ℃ to 60 ℃.\",\"PeriodicalId\":18444,\"journal\":{\"name\":\"Metrologia\",\"volume\":\"36 22\",\"pages\":\"\"},\"PeriodicalIF\":2.1000,\"publicationDate\":\"2024-01-08\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Metrologia\",\"FirstCategoryId\":\"5\",\"ListUrlMain\":\"https://doi.org/10.1088/1681-7575/ad1bfb\",\"RegionNum\":3,\"RegionCategory\":\"工程技术\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"INSTRUMENTS & INSTRUMENTATION\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Metrologia","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1088/1681-7575/ad1bfb","RegionNum":3,"RegionCategory":"工程技术","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"INSTRUMENTS & INSTRUMENTATION","Score":null,"Total":0}
引用次数: 0

摘要

对于 0.01 ℃ 以上的所有子范围,除了水的三相点(TPW)到镓的熔点之间的插值以外,ITS-90 上的温度都是在没有镓的熔点的情况下指定的。本研究提出了一个从水的三重点到铟的凝固点的新插值方程,它是一个分数三阶多项式,有两个系数,由标准铂电阻温度计(SPRT)在镓的熔点(WGa)和铟的凝固点(WIn)的电阻比测量值决定。一组来自不同国家和实验室的 30 个 SPRT 样本用于量化这种新插值法与 ITS-90 之间的比较。分析表明,在相同的子范围内,与 ITS-90 相比,重现性明显提高。从 TPW 到镓熔点的重叠子范围不一致性 (SRI) 峰值从 0.91 mK 降至 0.21 mK。与 ITS-90 相比,新方程 SRI 峰值的平均值和标准偏差都降低了约 4 倍(平均值从 0.20 mK 降至 0.05 mK,标准偏差从 0.32 mK 降至 0.07 mK)。此外,还使用铜块在恒温水浴中比较了四个 SPRT 的数据集,对同一子范围的第 3 类非唯一性(NU3)进行了测定,结果表明,NU3 的范围从 ITS-90 的 -0.15 mK 至 0.25 mK 缩小到新插值方程的 -0.10 mK 至 0.15 mK。不确定度的传播调查表明,新方程在 30 ℃ 至 60 ℃ 的较短区间内仅使固定点的测量不确定度增加了 7%。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
A New Interpolation Equation in the ITS-90 Subrange from the Triple Point of Water to the Freezing Point of Indium
For all the subranges above 0.01 ℃, other than the interpolation between the triple point of water (TPW) to the melting point of gallium, temperatures on the ITS-90 are specified without the melting point of gallium. This study suggests a new interpolation equation in the subrange from the triple point of water to the freezing point of indium which is a polynomial of fractional third order with two coefficients to be determined by the measuring values of standard platinum resistance thermometers (SPRTs)’ resistance ratios at the melting point of gallium (WGa) and at the freezing point of indium (WIn). A sample set of 30 SPRTs from various countries and laboratories is used to quantify the comparisons between this new interpolation and the ITS-90. Analyses show that the reproducibility is significantly improved compared with the ITS-90 in the same subrange. The peak value of overlapping subrange inconsistencies (SRI) from TPW to the melting point of gallium is reduced from 0.91 mK to 0.21 mK. Both the mean and standard deviation of the SRI peak values for the new equation decrease by approximately a factor of four compared to those of the ITS-90 (mean decreasing form 0.20 mK to 0.05 mK and standard deviation decreasing form 0.32 mK to 0.07 mK). Type 3 non-uniqueness (NU3) determinations of the same subrange have also been looked into with a data set of four SPRTs compared in thermostatic baths using a copper block and the results show that NU3 scales down from a range -0.15 mK to 0.25 mK of the ITS-90 to -0.10 mK to 0.15 mK of the new interpolation equation. Propagation of Uncertainties investigation shows that the new equation inflates merely 7% the measurement uncertainties of fixed points in a relatively short interval from 30 ℃ to 60 ℃.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Metrologia
Metrologia 工程技术-物理:应用
CiteScore
2.80
自引率
25.00%
发文量
137
审稿时长
12 months
期刊介绍: Published 6 times per year, Metrologia covers the fundamentals of measurements, particularly those dealing with the seven base units of the International System of Units (metre, kilogram, second, ampere, kelvin, candela, mole) or proposals to replace them. The journal also publishes papers that contribute to the solution of difficult measurement problems and improve the accuracy of derived units and constants that are of fundamental importance to physics. In addition to regular papers, the journal publishes review articles, issues devoted to single topics of timely interest and occasional conference proceedings. Letters to the Editor and Short Communications (generally three pages or less) are also considered.
期刊最新文献
Conversion of a piston-cylinder dimensional dataset to the effective area of a mechanical pressure generator. Evaluating the structural and predictive validity of a derivational morphology task with struggling adult readers. A robust time scale for space applications using the student’s t-distribution A digital four-arm bridge for the comparison of resistance with capacitance Traceable characterisation of fibre-coupled single-photon detectors
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1