{"title":"手持式结构光 3D 扫描仪在下肢假体多部位设置中的准确性、可重复性和可再现性","authors":"A. Cutti, Maria Grazia Santi, Andrew H. Hansen, Stefania Fatone","doi":"10.3390/s24072350","DOIUrl":null,"url":null,"abstract":"The aim of this work was to assess the accuracy, repeatability, and reproducibility of a hand-held, structured-light 3D scanner (EINScan Pro 2X Plus with High Definition Prime Pack, SHINING 3D Tech. Co., Ltd., Hangzhou, China), to support its potential use in multi-site settings on lower limb prosthetics. Four limb models with different shapes were fabricated and scanned with a metrological 3D scanner (EINScan Laser FreeScan 5X, SHINING 3D Tech. Co., Ltd., Hangzhou, China) by a professional operator (OP0). Limb models were then mailed to three sites where two operators (OP1, OP2) scanned them using their own structured-light 3D scanner (same model). OP1 scanned limb models twice (OP1-A, OP1-B). OP0, OP1-A, and OP2 scans were compared for accuracy, OP1-A and OP1-B for repeatability, and OP1-A and OP2 for reproducibility. Among all comparisons, the mean radial error was <0.25 mm, mean angular error was <4°, and root mean square error of the radial distance was <1 mm. Moreover, limits of agreement were <3.5% for perimeters and volumes. By comparing these results with respect to clinically-relevant thresholds and to the literature available on other 3D scanners, we conclude that the EINScan Pro 2X Plus 3D Scanner with High Definition Prime Pack has good accuracy, repeatability, and reproducibility, supporting its use in multi-site settings.","PeriodicalId":221960,"journal":{"name":"Sensors (Basel, Switzerland)","volume":"9 23","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2024-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Accuracy, Repeatability, and Reproducibility of a Hand-Held Structured-Light 3D Scanner across Multi-Site Settings in Lower Limb Prosthetics\",\"authors\":\"A. Cutti, Maria Grazia Santi, Andrew H. Hansen, Stefania Fatone\",\"doi\":\"10.3390/s24072350\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"The aim of this work was to assess the accuracy, repeatability, and reproducibility of a hand-held, structured-light 3D scanner (EINScan Pro 2X Plus with High Definition Prime Pack, SHINING 3D Tech. Co., Ltd., Hangzhou, China), to support its potential use in multi-site settings on lower limb prosthetics. Four limb models with different shapes were fabricated and scanned with a metrological 3D scanner (EINScan Laser FreeScan 5X, SHINING 3D Tech. Co., Ltd., Hangzhou, China) by a professional operator (OP0). Limb models were then mailed to three sites where two operators (OP1, OP2) scanned them using their own structured-light 3D scanner (same model). OP1 scanned limb models twice (OP1-A, OP1-B). OP0, OP1-A, and OP2 scans were compared for accuracy, OP1-A and OP1-B for repeatability, and OP1-A and OP2 for reproducibility. Among all comparisons, the mean radial error was <0.25 mm, mean angular error was <4°, and root mean square error of the radial distance was <1 mm. Moreover, limits of agreement were <3.5% for perimeters and volumes. By comparing these results with respect to clinically-relevant thresholds and to the literature available on other 3D scanners, we conclude that the EINScan Pro 2X Plus 3D Scanner with High Definition Prime Pack has good accuracy, repeatability, and reproducibility, supporting its use in multi-site settings.\",\"PeriodicalId\":221960,\"journal\":{\"name\":\"Sensors (Basel, Switzerland)\",\"volume\":\"9 23\",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2024-04-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Sensors (Basel, Switzerland)\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.3390/s24072350\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Sensors (Basel, Switzerland)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.3390/s24072350","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

这项工作的目的是评估手持式结构光三维扫描仪(EINScan Pro 2X Plus,带高清原色包,信宁三维科技有限公司,中国杭州)的准确性、可重复性和再现性。中国杭州,信宁三维科技有限公司),以支持其在下肢假肢多部位设置中的潜在应用。专业操作员(OP0)制作了四个不同形状的肢体模型,并使用计量三维扫描仪(EINScan Laser FreeScan 5X,信宁三维科技有限公司,中国杭州)进行扫描。然后,肢体模型被邮寄到三个地点,由两名操作员(OP1、OP2)使用自己的结构光三维扫描仪(相同型号)进行扫描。OP1 扫描肢体模型两次(OP1-A、OP1-B)。对 OP0、OP1-A 和 OP2 扫描的准确性进行比较,对 OP1-A 和 OP1-B 扫描的重复性进行比较,对 OP1-A 和 OP2 扫描的再现性进行比较。在所有比较中,平均径向误差小于 0.25 毫米,平均角度误差小于 4°,径向距离的均方根误差小于 1 毫米。此外,周长和体积的一致性均小于 3.5%。通过将这些结果与临床相关阈值和其他三维扫描仪的文献进行比较,我们得出结论:EINScan Pro 2X Plus 三维扫描仪(带高清晰度主件包)具有良好的准确性、可重复性和再现性,支持在多站点环境中使用。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Accuracy, Repeatability, and Reproducibility of a Hand-Held Structured-Light 3D Scanner across Multi-Site Settings in Lower Limb Prosthetics
The aim of this work was to assess the accuracy, repeatability, and reproducibility of a hand-held, structured-light 3D scanner (EINScan Pro 2X Plus with High Definition Prime Pack, SHINING 3D Tech. Co., Ltd., Hangzhou, China), to support its potential use in multi-site settings on lower limb prosthetics. Four limb models with different shapes were fabricated and scanned with a metrological 3D scanner (EINScan Laser FreeScan 5X, SHINING 3D Tech. Co., Ltd., Hangzhou, China) by a professional operator (OP0). Limb models were then mailed to three sites where two operators (OP1, OP2) scanned them using their own structured-light 3D scanner (same model). OP1 scanned limb models twice (OP1-A, OP1-B). OP0, OP1-A, and OP2 scans were compared for accuracy, OP1-A and OP1-B for repeatability, and OP1-A and OP2 for reproducibility. Among all comparisons, the mean radial error was <0.25 mm, mean angular error was <4°, and root mean square error of the radial distance was <1 mm. Moreover, limits of agreement were <3.5% for perimeters and volumes. By comparing these results with respect to clinically-relevant thresholds and to the literature available on other 3D scanners, we conclude that the EINScan Pro 2X Plus 3D Scanner with High Definition Prime Pack has good accuracy, repeatability, and reproducibility, supporting its use in multi-site settings.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
A Flexible Ammonia Gas Sensor Based on a Grafted Polyaniline Grown on a Polyethylene Terephthalate Film Investigation of Appropriate Scaling of Networks and Images for Convolutional Neural Network-Based Nerve Detection in Ultrasound-Guided Nerve Blocks EMR-HRNet: A Multi-Scale Feature Fusion Network for Landslide Segmentation from Remote Sensing Images Measuring DNI with a New Radiometer Based on an Optical Fiber and Photodiode Highly Sensitive Balloon-like Fiber Interferometer Based on Ethanol Coated for Temperature Measurement
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1