灭菌和消毒方法对数字化设计的外科植入物导板精确度的影响:体外研究。

IF 3.7 2区 医学 Q1 DENTISTRY, ORAL SURGERY & MEDICINE Clinical Implant Dentistry and Related Research Pub Date : 2024-05-29 DOI:10.1111/cid.13350
Ruikun Li, Jie Zhu, Shuo Wang, Xueyuan Li, Songhang Li
{"title":"灭菌和消毒方法对数字化设计的外科植入物导板精确度的影响:体外研究。","authors":"Ruikun Li,&nbsp;Jie Zhu,&nbsp;Shuo Wang,&nbsp;Xueyuan Li,&nbsp;Songhang Li","doi":"10.1111/cid.13350","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Introduction</h3>\n \n <p>Surgical guides are commonly used to assist with dental implant placement. This study investigated the effects of five sterilization and disinfection methods on the accuracy of implant guides.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>Thirty surgical guides (five in each group) were designed and printed (with digital light processing technology) using different sterilization or disinfection methods categorized into six groups: hydrogen peroxide sterilization (group one); glutaraldehyde sterilization (group two); autoclaving (group three); plasma sterilization (group four); iodophor disinfection (group five); and blank group (group six). Verification was determined using three methods: distance and angle between the cross-shaped marks, deformation after superimposing the guides, and displacement and axial changes in the virtual implant.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>After disinfection and sterilization, the guides in the autoclaving and iodophor groups showed a more pronounced color change and the guide in the autoclaving group had visible cracks. More significant changes were observed in the H<sub>2</sub>O<sub>2</sub>, glutaraldehyde, autoclaving, and iodophor groups regarding deformation after superimposing the guides and the distance and angle between the cross-shaped marks. The average labial deformation values (mm) of the first through fifth groups of guides were 0.283, 0.172, 0.289, 0.153, and 0.188, respectively. All groups were statistically different from the blank group for displacement and axial changes of the virtual implant (<i>p</i> &lt; 0.05).</p>\n </section>\n \n <section>\n \n <h3> Conclusion</h3>\n \n <p>The sizes of almost all surgical guides changed after sterilization and disinfection treatments, with between-group differences. Plasma sterilization was more suitable for surgical guide sterilization because of the smaller deformations after treatment.</p>\n </section>\n </div>","PeriodicalId":50679,"journal":{"name":"Clinical Implant Dentistry and Related Research","volume":"26 5","pages":"889-898"},"PeriodicalIF":3.7000,"publicationDate":"2024-05-29","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Effects of sterilization and disinfection methods on digitally designed surgical implant guide accuracy: An in vitro study\",\"authors\":\"Ruikun Li,&nbsp;Jie Zhu,&nbsp;Shuo Wang,&nbsp;Xueyuan Li,&nbsp;Songhang Li\",\"doi\":\"10.1111/cid.13350\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div>\\n \\n \\n <section>\\n \\n <h3> Introduction</h3>\\n \\n <p>Surgical guides are commonly used to assist with dental implant placement. This study investigated the effects of five sterilization and disinfection methods on the accuracy of implant guides.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Methods</h3>\\n \\n <p>Thirty surgical guides (five in each group) were designed and printed (with digital light processing technology) using different sterilization or disinfection methods categorized into six groups: hydrogen peroxide sterilization (group one); glutaraldehyde sterilization (group two); autoclaving (group three); plasma sterilization (group four); iodophor disinfection (group five); and blank group (group six). Verification was determined using three methods: distance and angle between the cross-shaped marks, deformation after superimposing the guides, and displacement and axial changes in the virtual implant.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Results</h3>\\n \\n <p>After disinfection and sterilization, the guides in the autoclaving and iodophor groups showed a more pronounced color change and the guide in the autoclaving group had visible cracks. More significant changes were observed in the H<sub>2</sub>O<sub>2</sub>, glutaraldehyde, autoclaving, and iodophor groups regarding deformation after superimposing the guides and the distance and angle between the cross-shaped marks. The average labial deformation values (mm) of the first through fifth groups of guides were 0.283, 0.172, 0.289, 0.153, and 0.188, respectively. All groups were statistically different from the blank group for displacement and axial changes of the virtual implant (<i>p</i> &lt; 0.05).</p>\\n </section>\\n \\n <section>\\n \\n <h3> Conclusion</h3>\\n \\n <p>The sizes of almost all surgical guides changed after sterilization and disinfection treatments, with between-group differences. Plasma sterilization was more suitable for surgical guide sterilization because of the smaller deformations after treatment.</p>\\n </section>\\n </div>\",\"PeriodicalId\":50679,\"journal\":{\"name\":\"Clinical Implant Dentistry and Related Research\",\"volume\":\"26 5\",\"pages\":\"889-898\"},\"PeriodicalIF\":3.7000,\"publicationDate\":\"2024-05-29\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Clinical Implant Dentistry and Related Research\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://onlinelibrary.wiley.com/doi/10.1111/cid.13350\",\"RegionNum\":2,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"DENTISTRY, ORAL SURGERY & MEDICINE\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical Implant Dentistry and Related Research","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1111/cid.13350","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

摘要

导言:手术导板通常用于辅助牙科种植体的植入。本研究调查了五种消毒和灭菌方法对种植导板准确性的影响:采用不同的消毒或灭菌方法设计和打印了 30 个手术导板(每组 5 个),并将其分为六组:过氧化氢消毒组(第一组)、戊二醛消毒组(第二组)、高压灭菌组(第三组)、等离子消毒组(第四组)、碘伏消毒组(第五组)和空白组(第六组)。验证采用三种方法:十字形标记之间的距离和角度、叠加导板后的变形以及虚拟种植体的位移和轴向变化:消毒灭菌后,高压灭菌组和碘伏组的导板颜色变化更明显,高压灭菌组的导板有明显裂纹。H2O2组、戊二醛组、高压灭菌组和碘伏组的导板叠加后的变形以及十字形标记之间的距离和角度都有更明显的变化。第一至第五组导板的平均唇侧变形值(毫米)分别为 0.283、0.172、0.289、0.153 和 0.188。在虚拟种植体的位移和轴向变化方面,所有组与空白组都有统计学差异(P 结论:所有手术导板的位移和轴向变化都与空白组不同:几乎所有手术导板的尺寸在消毒和灭菌处理后都发生了变化,且存在组间差异。等离子消毒更适合用于手术导板消毒,因为处理后的变形较小。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Effects of sterilization and disinfection methods on digitally designed surgical implant guide accuracy: An in vitro study

Introduction

Surgical guides are commonly used to assist with dental implant placement. This study investigated the effects of five sterilization and disinfection methods on the accuracy of implant guides.

Methods

Thirty surgical guides (five in each group) were designed and printed (with digital light processing technology) using different sterilization or disinfection methods categorized into six groups: hydrogen peroxide sterilization (group one); glutaraldehyde sterilization (group two); autoclaving (group three); plasma sterilization (group four); iodophor disinfection (group five); and blank group (group six). Verification was determined using three methods: distance and angle between the cross-shaped marks, deformation after superimposing the guides, and displacement and axial changes in the virtual implant.

Results

After disinfection and sterilization, the guides in the autoclaving and iodophor groups showed a more pronounced color change and the guide in the autoclaving group had visible cracks. More significant changes were observed in the H2O2, glutaraldehyde, autoclaving, and iodophor groups regarding deformation after superimposing the guides and the distance and angle between the cross-shaped marks. The average labial deformation values (mm) of the first through fifth groups of guides were 0.283, 0.172, 0.289, 0.153, and 0.188, respectively. All groups were statistically different from the blank group for displacement and axial changes of the virtual implant (p < 0.05).

Conclusion

The sizes of almost all surgical guides changed after sterilization and disinfection treatments, with between-group differences. Plasma sterilization was more suitable for surgical guide sterilization because of the smaller deformations after treatment.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
6.00
自引率
13.90%
发文量
103
审稿时长
4-8 weeks
期刊介绍: The goal of Clinical Implant Dentistry and Related Research is to advance the scientific and technical aspects relating to dental implants and related scientific subjects. Dissemination of new and evolving information related to dental implants and the related science is the primary goal of our journal. The range of topics covered by the journals will include but be not limited to: New scientific developments relating to bone Implant surfaces and their relationship to the surrounding tissues Computer aided implant designs Computer aided prosthetic designs Immediate implant loading Immediate implant placement Materials relating to bone induction and conduction New surgical methods relating to implant placement New materials and methods relating to implant restorations Methods for determining implant stability A primary focus of the journal is publication of evidenced based articles evaluating to new dental implants, techniques and multicenter studies evaluating these treatments. In addition basic science research relating to wound healing and osseointegration will be an important focus for the journal.
期刊最新文献
Issue Information Featured Cover A transcrestal sinus floor elevation strategy based on a haptic robot system: An in vitro study Influence of repeated implant‐abutment manipulation on the prevalence of peri‐implant diseases in complete arch restorations. A retrospective analysis Biocompatibility and dimensional stability through the use of 3D‐printed scaffolds made by polycaprolactone and bioglass‐7: An in vitro and in vivo study
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1