Changfu Wu, Chad Green, Salvador Marquez, Paolo Monelli, Craig Weinberg, Matthew Weston, Patricia Lawford, Duke Cameron, Ajit Yoganathan, Ulrich Steinseifer
{"title":"有效孔径:手术瓣膜假体的新尺寸参数,为瓣膜选择提供参考。","authors":"Changfu Wu, Chad Green, Salvador Marquez, Paolo Monelli, Craig Weinberg, Matthew Weston, Patricia Lawford, Duke Cameron, Ajit Yoganathan, Ulrich Steinseifer","doi":"10.21037/acs-2024-aae-0057","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The labeled sizes of surgical valve prostheses and their discordance with the physical internal valve orifice sizes has long been a controversy in the cardiac surgery community, leading many to believe it to be a contributing factor in prosthesis-patient mismatch following valvular replacement surgery. In an attempt to address this issue, the International Organization for Standardization (ISO) 5840-2:2021 standard for surgical valve prostheses recommends that a new sizing parameter, namely, the effective orifice diameter, be provided in labeling by all manufacturers as an indicator of the true flow-passing capacity of a prosthetic valve.</p><p><strong>Methods: </strong>The ISO Cardiac Valves Working Group conducted a multi-laboratory round-robin study to investigate whether the effective orifice diameter of a prosthetic surgical valve could be derived repeatably and reproducibly through steady forward-flow testing. A total of seven valve models, each with multiple sizes, were tested, including a mechanical heart valve and multiple biological heart valves.</p><p><strong>Results: </strong>The round-robin study confirmed that the steady forward-flow test had good intra-laboratory repeatability and inter-laboratory reproducibility in deriving the effective orifice diameters of surgical valve prostheses. On average, among the participating laboratories, the experimentally derived effective orifice diameter of a prosthetic heart valve was 3-12 mm smaller than its labeled size.</p><p><strong>Conclusions: </strong>The effective orifice diameter provides better characterization of the hydrodynamic characteristics of a surgical valve prosthesis and can be derived using a validated steady forward-flow test method. This new sizing parameter will soon be adopted by surgical valve manufacturers and provided in device labeling to inform valve selection by surgeons.</p>","PeriodicalId":8067,"journal":{"name":"Annals of cardiothoracic surgery","volume":"13 3","pages":"283-290"},"PeriodicalIF":3.3000,"publicationDate":"2024-05-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11148754/pdf/","citationCount":"0","resultStr":"{\"title\":\"Effective orifice diameter: a new sizing parameter of surgical valve prostheses to inform valve selection.\",\"authors\":\"Changfu Wu, Chad Green, Salvador Marquez, Paolo Monelli, Craig Weinberg, Matthew Weston, Patricia Lawford, Duke Cameron, Ajit Yoganathan, Ulrich Steinseifer\",\"doi\":\"10.21037/acs-2024-aae-0057\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>The labeled sizes of surgical valve prostheses and their discordance with the physical internal valve orifice sizes has long been a controversy in the cardiac surgery community, leading many to believe it to be a contributing factor in prosthesis-patient mismatch following valvular replacement surgery. In an attempt to address this issue, the International Organization for Standardization (ISO) 5840-2:2021 standard for surgical valve prostheses recommends that a new sizing parameter, namely, the effective orifice diameter, be provided in labeling by all manufacturers as an indicator of the true flow-passing capacity of a prosthetic valve.</p><p><strong>Methods: </strong>The ISO Cardiac Valves Working Group conducted a multi-laboratory round-robin study to investigate whether the effective orifice diameter of a prosthetic surgical valve could be derived repeatably and reproducibly through steady forward-flow testing. A total of seven valve models, each with multiple sizes, were tested, including a mechanical heart valve and multiple biological heart valves.</p><p><strong>Results: </strong>The round-robin study confirmed that the steady forward-flow test had good intra-laboratory repeatability and inter-laboratory reproducibility in deriving the effective orifice diameters of surgical valve prostheses. On average, among the participating laboratories, the experimentally derived effective orifice diameter of a prosthetic heart valve was 3-12 mm smaller than its labeled size.</p><p><strong>Conclusions: </strong>The effective orifice diameter provides better characterization of the hydrodynamic characteristics of a surgical valve prosthesis and can be derived using a validated steady forward-flow test method. This new sizing parameter will soon be adopted by surgical valve manufacturers and provided in device labeling to inform valve selection by surgeons.</p>\",\"PeriodicalId\":8067,\"journal\":{\"name\":\"Annals of cardiothoracic surgery\",\"volume\":\"13 3\",\"pages\":\"283-290\"},\"PeriodicalIF\":3.3000,\"publicationDate\":\"2024-05-31\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11148754/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Annals of cardiothoracic surgery\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.21037/acs-2024-aae-0057\",\"RegionNum\":2,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/5/16 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q2\",\"JCRName\":\"CARDIAC & CARDIOVASCULAR SYSTEMS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annals of cardiothoracic surgery","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.21037/acs-2024-aae-0057","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/5/16 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
Effective orifice diameter: a new sizing parameter of surgical valve prostheses to inform valve selection.
Background: The labeled sizes of surgical valve prostheses and their discordance with the physical internal valve orifice sizes has long been a controversy in the cardiac surgery community, leading many to believe it to be a contributing factor in prosthesis-patient mismatch following valvular replacement surgery. In an attempt to address this issue, the International Organization for Standardization (ISO) 5840-2:2021 standard for surgical valve prostheses recommends that a new sizing parameter, namely, the effective orifice diameter, be provided in labeling by all manufacturers as an indicator of the true flow-passing capacity of a prosthetic valve.
Methods: The ISO Cardiac Valves Working Group conducted a multi-laboratory round-robin study to investigate whether the effective orifice diameter of a prosthetic surgical valve could be derived repeatably and reproducibly through steady forward-flow testing. A total of seven valve models, each with multiple sizes, were tested, including a mechanical heart valve and multiple biological heart valves.
Results: The round-robin study confirmed that the steady forward-flow test had good intra-laboratory repeatability and inter-laboratory reproducibility in deriving the effective orifice diameters of surgical valve prostheses. On average, among the participating laboratories, the experimentally derived effective orifice diameter of a prosthetic heart valve was 3-12 mm smaller than its labeled size.
Conclusions: The effective orifice diameter provides better characterization of the hydrodynamic characteristics of a surgical valve prosthesis and can be derived using a validated steady forward-flow test method. This new sizing parameter will soon be adopted by surgical valve manufacturers and provided in device labeling to inform valve selection by surgeons.