小鼠模型骨肉瘤定量 T2 和 T2* 映像的再现性和可重复性。

IF 3.7 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING European Radiology Experimental Pub Date : 2024-06-14 DOI:10.1186/s41747-024-00467-9
Raheleh Roudi, Laura J Pisani, Fabrizio Pisani, Tie Liang, Heike E Daldrup-Link
{"title":"小鼠模型骨肉瘤定量 T2 和 T2* 映像的再现性和可重复性。","authors":"Raheleh Roudi, Laura J Pisani, Fabrizio Pisani, Tie Liang, Heike E Daldrup-Link","doi":"10.1186/s41747-024-00467-9","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>New immunotherapies activate tumor-associated macrophages (TAMs) in the osteosarcoma microenvironment. Iron oxide nanoparticles (IONPs) are phagocytosed by TAMs and, therefore, enable TAM detection on T2*- and T2-weighted magnetic resonance images. We assessed the repeatability and reproducibility of T2*- and T2-mapping of osteosarcomas in a mouse model.</p><p><strong>Methods: </strong>Fifteen BALB/c mice bearing-murine osteosarcomas underwent magnetic resonance imaging (MRI) on 3-T and 7-T scanners before and after intravenous IONP infusion, using T2*-weighted multi-gradient-echo, T2-weighted fast spin-echo, and T2-weighted multi-echo sequences. Each sequence was repeated twice. Tumor T2 and T2* relaxation times were measured twice by two independent investigators. Repeatability and reproducibility of measurements were assessed.</p><p><strong>Results: </strong>We found excellent agreement between duplicate acquisitions for both T2* and T2 measurements at either magnetic field strength, by the same individual (repeatability), and between individuals (reproducibility). The repeatability concordance correlation coefficient (CCC) for T2* values were 0.99 (coefficients of variation (CoV) 4.43%) for reader 1 and 0.98 (CoV 5.82%) for reader 2. The reproducibility of T2* values between the two readers was 0.99 (CoV 3.32%) for the first acquisitions and 0.99 (CoV 6.30%) for the second acquisitions. Regarding T2 values, the repeatability of CCC was similar for both readers, 0.98 (CoV 3.64% for reader 1 and 4.45% for reader 2). The CCC of the reproducibility of T2 was 0.99 (CoV 3.1%) for the first acquisition and 0.98 (CoV 4.38%) for the second acquisition.</p><p><strong>Conclusions: </strong>Our results demonstrated high repeatability and reproducibility of quantitative T2* and T2 mapping for monitoring the presence of TAMs in osteosarcomas.</p><p><strong>Relevance statement: </strong>T2* and T2 measurements of osteosarcomas on IONP-enhanced MRI could allow identifying patients who may benefit from TAM-modulating immunotherapies and for monitoring treatment response. The technique described here could be also applied across a wide range of other solid tumors.</p><p><strong>Key points: </strong>• Optimal integration of TAM-modulating immunotherapies with conventional chemotherapy remains poorly elucidated. • We found high repeatability of T2* and T2 measurements of osteosarcomas in a mouse model, both with and without IONPs contrast, at 3-T and 7-T MRI field strengths. • T2 and T2* mapping may be used to determine response to macrophage-modulating cancer immunotherapies.</p>","PeriodicalId":36926,"journal":{"name":"European Radiology Experimental","volume":"8 1","pages":"74"},"PeriodicalIF":3.7000,"publicationDate":"2024-06-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11176138/pdf/","citationCount":"0","resultStr":"{\"title\":\"Reproducibility and repeatability of quantitative T2 and T2* mapping of osteosarcomas in a mouse model.\",\"authors\":\"Raheleh Roudi, Laura J Pisani, Fabrizio Pisani, Tie Liang, Heike E Daldrup-Link\",\"doi\":\"10.1186/s41747-024-00467-9\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>New immunotherapies activate tumor-associated macrophages (TAMs) in the osteosarcoma microenvironment. Iron oxide nanoparticles (IONPs) are phagocytosed by TAMs and, therefore, enable TAM detection on T2*- and T2-weighted magnetic resonance images. We assessed the repeatability and reproducibility of T2*- and T2-mapping of osteosarcomas in a mouse model.</p><p><strong>Methods: </strong>Fifteen BALB/c mice bearing-murine osteosarcomas underwent magnetic resonance imaging (MRI) on 3-T and 7-T scanners before and after intravenous IONP infusion, using T2*-weighted multi-gradient-echo, T2-weighted fast spin-echo, and T2-weighted multi-echo sequences. Each sequence was repeated twice. Tumor T2 and T2* relaxation times were measured twice by two independent investigators. Repeatability and reproducibility of measurements were assessed.</p><p><strong>Results: </strong>We found excellent agreement between duplicate acquisitions for both T2* and T2 measurements at either magnetic field strength, by the same individual (repeatability), and between individuals (reproducibility). The repeatability concordance correlation coefficient (CCC) for T2* values were 0.99 (coefficients of variation (CoV) 4.43%) for reader 1 and 0.98 (CoV 5.82%) for reader 2. The reproducibility of T2* values between the two readers was 0.99 (CoV 3.32%) for the first acquisitions and 0.99 (CoV 6.30%) for the second acquisitions. Regarding T2 values, the repeatability of CCC was similar for both readers, 0.98 (CoV 3.64% for reader 1 and 4.45% for reader 2). The CCC of the reproducibility of T2 was 0.99 (CoV 3.1%) for the first acquisition and 0.98 (CoV 4.38%) for the second acquisition.</p><p><strong>Conclusions: </strong>Our results demonstrated high repeatability and reproducibility of quantitative T2* and T2 mapping for monitoring the presence of TAMs in osteosarcomas.</p><p><strong>Relevance statement: </strong>T2* and T2 measurements of osteosarcomas on IONP-enhanced MRI could allow identifying patients who may benefit from TAM-modulating immunotherapies and for monitoring treatment response. The technique described here could be also applied across a wide range of other solid tumors.</p><p><strong>Key points: </strong>• Optimal integration of TAM-modulating immunotherapies with conventional chemotherapy remains poorly elucidated. • We found high repeatability of T2* and T2 measurements of osteosarcomas in a mouse model, both with and without IONPs contrast, at 3-T and 7-T MRI field strengths. • T2 and T2* mapping may be used to determine response to macrophage-modulating cancer immunotherapies.</p>\",\"PeriodicalId\":36926,\"journal\":{\"name\":\"European Radiology Experimental\",\"volume\":\"8 1\",\"pages\":\"74\"},\"PeriodicalIF\":3.7000,\"publicationDate\":\"2024-06-14\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11176138/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"European Radiology Experimental\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1186/s41747-024-00467-9\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"European Radiology Experimental","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1186/s41747-024-00467-9","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

摘要

背景:新的免疫疗法可激活骨肉瘤微环境中的肿瘤相关巨噬细胞(TAMs)。氧化铁纳米颗粒(IONPs)会被TAMs吞噬,因此能在T2*和T2加权磁共振图像上检测到TAM。我们在小鼠模型中评估了骨肉瘤 T2* 和 T2 映射的可重复性和再现性:在静脉注射 IONP 之前和之后,15 只患有鼠骨肉瘤的 BALB/c 小鼠在 3-T 和 7-T 扫描仪上接受了磁共振成像 (MRI),使用的是 T2* 加权多梯度回波、T2 加权快速自旋回波和 T2 加权多回波序列。每个序列重复两次。肿瘤 T2 和 T2* 驰豫时间由两名独立研究人员测量两次。对测量的重复性和再现性进行了评估:我们发现,在任何一种磁场强度下,同一人(重复性)和不同人(再现性)重复采集的 T2* 和 T2 测量结果都非常一致。阅读器 1 的 T2* 值重复性一致性相关系数 (CCC) 为 0.99(变异系数 (CoV) 为 4.43%),阅读器 2 的重复性一致性相关系数 (CCC) 为 0.98(变异系数 (CoV) 为 5.82%)。两个阅读器之间 T2* 值的可重复性为:第一次采集为 0.99(变异系数为 3.32%),第二次采集为 0.99(变异系数为 6.30%)。在 T2 值方面,两个读数器的 CCC 重复性相似,均为 0.98(读数器 1 的 CoV 为 3.64%,读数器 2 为 4.45%)。第一次采集的 T2 可重复性 CCC 为 0.99(CoV 为 3.1%),第二次采集为 0.98(CoV 为 4.38%):我们的结果表明,定量 T2* 和 T2 图谱在监测骨肉瘤中 TAM 的存在方面具有很高的重复性和再现性:IONP增强磁共振成像对骨肉瘤的T2*和T2测量可用于识别可能从TAM调节免疫疗法中获益的患者并监测治疗反应。本文所述技术还可广泛应用于其他实体瘤:- 要点:TAM调节免疫疗法与传统化疗的最佳结合仍未得到充分阐明。- 我们发现,在3T和7T磁共振成像场强下,对小鼠模型中的骨肉瘤进行T2*和T2测量时,无论是否使用IONPs对比剂,重复性都很高。- T2和T2*图谱可用于确定对巨噬细胞调节癌症免疫疗法的反应。
本文章由计算机程序翻译,如有差异,请以英文原文为准。

摘要图片

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Reproducibility and repeatability of quantitative T2 and T2* mapping of osteosarcomas in a mouse model.

Background: New immunotherapies activate tumor-associated macrophages (TAMs) in the osteosarcoma microenvironment. Iron oxide nanoparticles (IONPs) are phagocytosed by TAMs and, therefore, enable TAM detection on T2*- and T2-weighted magnetic resonance images. We assessed the repeatability and reproducibility of T2*- and T2-mapping of osteosarcomas in a mouse model.

Methods: Fifteen BALB/c mice bearing-murine osteosarcomas underwent magnetic resonance imaging (MRI) on 3-T and 7-T scanners before and after intravenous IONP infusion, using T2*-weighted multi-gradient-echo, T2-weighted fast spin-echo, and T2-weighted multi-echo sequences. Each sequence was repeated twice. Tumor T2 and T2* relaxation times were measured twice by two independent investigators. Repeatability and reproducibility of measurements were assessed.

Results: We found excellent agreement between duplicate acquisitions for both T2* and T2 measurements at either magnetic field strength, by the same individual (repeatability), and between individuals (reproducibility). The repeatability concordance correlation coefficient (CCC) for T2* values were 0.99 (coefficients of variation (CoV) 4.43%) for reader 1 and 0.98 (CoV 5.82%) for reader 2. The reproducibility of T2* values between the two readers was 0.99 (CoV 3.32%) for the first acquisitions and 0.99 (CoV 6.30%) for the second acquisitions. Regarding T2 values, the repeatability of CCC was similar for both readers, 0.98 (CoV 3.64% for reader 1 and 4.45% for reader 2). The CCC of the reproducibility of T2 was 0.99 (CoV 3.1%) for the first acquisition and 0.98 (CoV 4.38%) for the second acquisition.

Conclusions: Our results demonstrated high repeatability and reproducibility of quantitative T2* and T2 mapping for monitoring the presence of TAMs in osteosarcomas.

Relevance statement: T2* and T2 measurements of osteosarcomas on IONP-enhanced MRI could allow identifying patients who may benefit from TAM-modulating immunotherapies and for monitoring treatment response. The technique described here could be also applied across a wide range of other solid tumors.

Key points: • Optimal integration of TAM-modulating immunotherapies with conventional chemotherapy remains poorly elucidated. • We found high repeatability of T2* and T2 measurements of osteosarcomas in a mouse model, both with and without IONPs contrast, at 3-T and 7-T MRI field strengths. • T2 and T2* mapping may be used to determine response to macrophage-modulating cancer immunotherapies.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
European Radiology Experimental
European Radiology Experimental Medicine-Radiology, Nuclear Medicine and Imaging
CiteScore
6.70
自引率
2.60%
发文量
56
审稿时长
18 weeks
期刊最新文献
An AI deep learning algorithm for detecting pulmonary nodules on ultra-low-dose CT in an emergency setting: a reader study. Evaluation of pulmonary artery pressure, blood indices, and myocardial microcirculation in rats returning from high altitude to moderate altitude. Image biomarkers and explainable AI: handcrafted features versus deep learned features. Technical feasibility of automated blur detection in digital mammography using convolutional neural network. Quantification of breast biopsy clip marker artifact on routine breast MRI sequences: a phantom study.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1