S.-K. Lee, Matthew R. Tarasek, Keith Park, Desmond T.-B. Yeo, Thomas K.-F. Foo
{"title":"用于降低高性能人体头部磁共振成像系统声压级的可插入式双密度介质屏障。","authors":"S.-K. Lee, Matthew R. Tarasek, Keith Park, Desmond T.-B. Yeo, Thomas K.-F. Foo","doi":"10.1016/j.mri.2024.110224","DOIUrl":null,"url":null,"abstract":"<div><p>We report use of a dual-density dielectric barrier surrounding a detachable high-pass radiofrequency (RF) birdcage coil to achieve an order-of-magnitude reduction of acoustic noise in a high-performance head gradient system. The barrier consisted of a 4.5 mm-thick mass-loaded vinyl and a 6 mm-thick polyurethane foam. It was inserted into the radial gap between the birdcage coil and the RF shield in a prototype head-only gradient system at 3 T. More than 9 dB<sub>A</sub> reduction of sound pressure level was achieved on the average with representative, high acoustic-noise imaging sequences. Increased acoustic damping was apparent from acoustic impulse response functions. High dielectric constant of the mass-loaded vinyl effectively added distributed capacitance to the birdcage coil, lowering the resonance frequency, but not seriously degrading the RF transmission performance. The barrier occupied the radial space normally used for air cooling of the RF coil and the RF shield. The resulting omission of air cooling was found to be acceptable with efficient gradient thermal management and use of a high-resistivity RF shield for eddy current reduction. The proposed method can improve patient experience while preserving image quality in a high-power head-only gradient system.</p></div>","PeriodicalId":18165,"journal":{"name":"Magnetic resonance imaging","volume":"114 ","pages":"Article 110224"},"PeriodicalIF":2.1000,"publicationDate":"2024-09-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Insertable, dual-density dielectric barrier for acoustic pressure level reduction in a high-performance human head-only MRI system\",\"authors\":\"S.-K. Lee, Matthew R. Tarasek, Keith Park, Desmond T.-B. Yeo, Thomas K.-F. Foo\",\"doi\":\"10.1016/j.mri.2024.110224\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><p>We report use of a dual-density dielectric barrier surrounding a detachable high-pass radiofrequency (RF) birdcage coil to achieve an order-of-magnitude reduction of acoustic noise in a high-performance head gradient system. The barrier consisted of a 4.5 mm-thick mass-loaded vinyl and a 6 mm-thick polyurethane foam. It was inserted into the radial gap between the birdcage coil and the RF shield in a prototype head-only gradient system at 3 T. More than 9 dB<sub>A</sub> reduction of sound pressure level was achieved on the average with representative, high acoustic-noise imaging sequences. Increased acoustic damping was apparent from acoustic impulse response functions. High dielectric constant of the mass-loaded vinyl effectively added distributed capacitance to the birdcage coil, lowering the resonance frequency, but not seriously degrading the RF transmission performance. The barrier occupied the radial space normally used for air cooling of the RF coil and the RF shield. The resulting omission of air cooling was found to be acceptable with efficient gradient thermal management and use of a high-resistivity RF shield for eddy current reduction. The proposed method can improve patient experience while preserving image quality in a high-power head-only gradient system.</p></div>\",\"PeriodicalId\":18165,\"journal\":{\"name\":\"Magnetic resonance imaging\",\"volume\":\"114 \",\"pages\":\"Article 110224\"},\"PeriodicalIF\":2.1000,\"publicationDate\":\"2024-09-02\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Magnetic resonance imaging\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S0730725X24002054\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Magnetic resonance imaging","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0730725X24002054","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
摘要
我们报告了在可拆卸的高通射频(RF)鸟笼线圈周围使用双密度介质屏障,从而在高性能头部梯度系统中实现了数量级的声噪降低。屏障由 4.5 毫米厚的大质量乙烯基和 6 毫米厚的聚氨酯泡沫组成。在 3 T 的原型头部梯度系统中,将其插入鸟笼线圈和射频屏蔽之间的径向间隙。在具有代表性的高声噪成像序列中,声压级平均降低了 9 分贝以上。从声学脉冲响应函数中可以明显看到声学阻尼的增加。质量负载乙烯基的高介电常数有效地增加了鸟笼线圈的分布电容,降低了共振频率,但并没有严重降低射频传输性能。阻挡层占据了通常用于对射频线圈和射频屏蔽进行空气冷却的径向空间。通过有效的梯度热管理和使用高电阻率射频屏蔽以减少涡流,发现省去空气冷却是可以接受的。所提出的方法可以改善患者的就医体验,同时保持大功率纯头部梯度系统的图像质量。
Insertable, dual-density dielectric barrier for acoustic pressure level reduction in a high-performance human head-only MRI system
We report use of a dual-density dielectric barrier surrounding a detachable high-pass radiofrequency (RF) birdcage coil to achieve an order-of-magnitude reduction of acoustic noise in a high-performance head gradient system. The barrier consisted of a 4.5 mm-thick mass-loaded vinyl and a 6 mm-thick polyurethane foam. It was inserted into the radial gap between the birdcage coil and the RF shield in a prototype head-only gradient system at 3 T. More than 9 dBA reduction of sound pressure level was achieved on the average with representative, high acoustic-noise imaging sequences. Increased acoustic damping was apparent from acoustic impulse response functions. High dielectric constant of the mass-loaded vinyl effectively added distributed capacitance to the birdcage coil, lowering the resonance frequency, but not seriously degrading the RF transmission performance. The barrier occupied the radial space normally used for air cooling of the RF coil and the RF shield. The resulting omission of air cooling was found to be acceptable with efficient gradient thermal management and use of a high-resistivity RF shield for eddy current reduction. The proposed method can improve patient experience while preserving image quality in a high-power head-only gradient system.
期刊介绍:
Magnetic Resonance Imaging (MRI) is the first international multidisciplinary journal encompassing physical, life, and clinical science investigations as they relate to the development and use of magnetic resonance imaging. MRI is dedicated to both basic research, technological innovation and applications, providing a single forum for communication among radiologists, physicists, chemists, biochemists, biologists, engineers, internists, pathologists, physiologists, computer scientists, and mathematicians.