{"title":"血色素ACT- lr和ACT+试验皿监测中低肝素浓度的性能:一项体外研究。","authors":"Janne Moilanen MD , Marika Pada MD , Pasi Ohtonen MSc , Timo Kaakinen MD, PhD , Matti Pokela MD, PhD , Eeva-Riitta Savolainen MD, PhD , Tiina Erkinaro MD, PhD","doi":"10.1053/j.jvca.2024.12.001","DOIUrl":null,"url":null,"abstract":"<div><h3>Objectives</h3><div>According to the manufacturer, the Hemochron ACT-LR cuvette is designated for heparin concentrations of 0 to 2.5 IU/mL, while the optimal concentration range for the ACT+ cartridge is 1 to 6 IU/mL. We hypothesized that at low to moderate heparin concentrations, the ACT-LR is more reliable than the ACT+.</div></div><div><h3>Design</h3><div>An in vitro study.</div></div><div><h3>Setting</h3><div>A tertiary care university hospital.</div></div><div><h3>Participants</h3><div>Twenty healthy adult volunteers.</div></div><div><h3>Interventions</h3><div>Donor blood samples were spiked with unfractionated heparin to concentrations of 0, 0.5, 1.0, 1.5, 2.0, and 2.5 IU/mL. From each concentration, 3 measurements were made with each test cassette using 3 different Hemochron Signature Elite devices, for a total of 720 ACT tests. The Bland-Altman method and intraclass correlation coefficients were used to analyze the data.</div></div><div><h3>Measurements and Main Results</h3><div>All 360 ACT+ tests yielded a result. Of 360, 38 (11%) ACT-LR tests ran above the upper measurement range; of these, 36 were from the 120 measurements at heparin concentrations ≥2.0 IU/mL. The bias between all ACT-LR and ACT+ measurements was 52 (95% CI, 48-56) celite seconds, and the limits of agreement were wide. Based on the intraclass correlation coefficients, the reliability of ACT+ was better than or equal to that of ACT-LR.</div></div><div><h3>Conclusions</h3><div>The performance of ACT+ is equal to that of ACT-LR up to a concentration of 0.5 IU/mL, above which ACT+ is superior to ACT-LR, questioning the rationale for a specific low-range test. However, there is a significant bias between ACT-LR and ACT+ measurements that needs to be considered if switching from one test to the other.</div></div>","PeriodicalId":15176,"journal":{"name":"Journal of cardiothoracic and vascular anesthesia","volume":"39 2","pages":"Pages 447-452"},"PeriodicalIF":2.3000,"publicationDate":"2025-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Performance of Hemochron ACT-LR and ACT+ Test Cuvettes in Monitoring Low to Moderate Heparin Concentrations: An In Vitro Study\",\"authors\":\"Janne Moilanen MD , Marika Pada MD , Pasi Ohtonen MSc , Timo Kaakinen MD, PhD , Matti Pokela MD, PhD , Eeva-Riitta Savolainen MD, PhD , Tiina Erkinaro MD, PhD\",\"doi\":\"10.1053/j.jvca.2024.12.001\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Objectives</h3><div>According to the manufacturer, the Hemochron ACT-LR cuvette is designated for heparin concentrations of 0 to 2.5 IU/mL, while the optimal concentration range for the ACT+ cartridge is 1 to 6 IU/mL. We hypothesized that at low to moderate heparin concentrations, the ACT-LR is more reliable than the ACT+.</div></div><div><h3>Design</h3><div>An in vitro study.</div></div><div><h3>Setting</h3><div>A tertiary care university hospital.</div></div><div><h3>Participants</h3><div>Twenty healthy adult volunteers.</div></div><div><h3>Interventions</h3><div>Donor blood samples were spiked with unfractionated heparin to concentrations of 0, 0.5, 1.0, 1.5, 2.0, and 2.5 IU/mL. From each concentration, 3 measurements were made with each test cassette using 3 different Hemochron Signature Elite devices, for a total of 720 ACT tests. The Bland-Altman method and intraclass correlation coefficients were used to analyze the data.</div></div><div><h3>Measurements and Main Results</h3><div>All 360 ACT+ tests yielded a result. Of 360, 38 (11%) ACT-LR tests ran above the upper measurement range; of these, 36 were from the 120 measurements at heparin concentrations ≥2.0 IU/mL. The bias between all ACT-LR and ACT+ measurements was 52 (95% CI, 48-56) celite seconds, and the limits of agreement were wide. Based on the intraclass correlation coefficients, the reliability of ACT+ was better than or equal to that of ACT-LR.</div></div><div><h3>Conclusions</h3><div>The performance of ACT+ is equal to that of ACT-LR up to a concentration of 0.5 IU/mL, above which ACT+ is superior to ACT-LR, questioning the rationale for a specific low-range test. However, there is a significant bias between ACT-LR and ACT+ measurements that needs to be considered if switching from one test to the other.</div></div>\",\"PeriodicalId\":15176,\"journal\":{\"name\":\"Journal of cardiothoracic and vascular anesthesia\",\"volume\":\"39 2\",\"pages\":\"Pages 447-452\"},\"PeriodicalIF\":2.3000,\"publicationDate\":\"2025-02-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of cardiothoracic and vascular anesthesia\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S105307702400939X\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"ANESTHESIOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of cardiothoracic and vascular anesthesia","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S105307702400939X","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ANESTHESIOLOGY","Score":null,"Total":0}
Performance of Hemochron ACT-LR and ACT+ Test Cuvettes in Monitoring Low to Moderate Heparin Concentrations: An In Vitro Study
Objectives
According to the manufacturer, the Hemochron ACT-LR cuvette is designated for heparin concentrations of 0 to 2.5 IU/mL, while the optimal concentration range for the ACT+ cartridge is 1 to 6 IU/mL. We hypothesized that at low to moderate heparin concentrations, the ACT-LR is more reliable than the ACT+.
Design
An in vitro study.
Setting
A tertiary care university hospital.
Participants
Twenty healthy adult volunteers.
Interventions
Donor blood samples were spiked with unfractionated heparin to concentrations of 0, 0.5, 1.0, 1.5, 2.0, and 2.5 IU/mL. From each concentration, 3 measurements were made with each test cassette using 3 different Hemochron Signature Elite devices, for a total of 720 ACT tests. The Bland-Altman method and intraclass correlation coefficients were used to analyze the data.
Measurements and Main Results
All 360 ACT+ tests yielded a result. Of 360, 38 (11%) ACT-LR tests ran above the upper measurement range; of these, 36 were from the 120 measurements at heparin concentrations ≥2.0 IU/mL. The bias between all ACT-LR and ACT+ measurements was 52 (95% CI, 48-56) celite seconds, and the limits of agreement were wide. Based on the intraclass correlation coefficients, the reliability of ACT+ was better than or equal to that of ACT-LR.
Conclusions
The performance of ACT+ is equal to that of ACT-LR up to a concentration of 0.5 IU/mL, above which ACT+ is superior to ACT-LR, questioning the rationale for a specific low-range test. However, there is a significant bias between ACT-LR and ACT+ measurements that needs to be considered if switching from one test to the other.
期刊介绍:
The Journal of Cardiothoracic and Vascular Anesthesia is primarily aimed at anesthesiologists who deal with patients undergoing cardiac, thoracic or vascular surgical procedures. JCVA features a multidisciplinary approach, with contributions from cardiac, vascular and thoracic surgeons, cardiologists, and other related specialists. Emphasis is placed on rapid publication of clinically relevant material.