北京地区外束流放疗输出剂量调查与分析

Feng Zechen, Ma Yongzhong, W. Hongfang, Su Yaru, Zhang Zhibin, Lou Yun
{"title":"北京地区外束流放疗输出剂量调查与分析","authors":"Feng Zechen, Ma Yongzhong, W. Hongfang, Su Yaru, Zhang Zhibin, Lou Yun","doi":"10.13491/J.ISSN.1004-714X.2021.02.001","DOIUrl":null,"url":null,"abstract":"Objective To study the output dose of external beam radiotherapy in Beijing and analyze the\n cause of output dose deviation. Guidance is given on the selection of measurement\n methods, calculation formulas and parameters.\n Methods During 2020, the output doses of medical electron accelerators, tomotherapy, cyberknife,\n and γ-knife were measured based on the water absorbed dose calibration factor combined\n with the requirements in the IAEA TRS398 and TRS483 reports, and low-energy X-rays\n were measured based on the air kerma calibration factor combined with the requirements\n in the AAPM protocal. The output dose was measured and compared with the preset irradiation\n value.\n Results A total of 59 external beam radiation therapy equipments’s output dose were investigated.\n 2 of the 49 accelerators’s absolute value of relative deviation between the system\n indication value and the measurement exceeded 3%; the deviation of other external\n beam radiotherapy equipments was less than 3%.\n Conclusion A unified method and multi-party quality control can ensure the accuracy of the output\n dose. In the routine operation of the equipment, the hospital should strictly implement\n periodic inspections to ensure accurate and reliable radiotherapy.\n 摘要: 目的 调查北京地方所属外照射放射治疗设备输出剂量的情况, 分析输出剂量相对偏差原因, 明确测量方法、 计算公式及参数的选取。\n 方法 2020 年期间, 采用水中吸收剂量校准因子结合国际原子能机构 (IAEA) TRS398 和 TRS483 号报告中的要求对医用电子加速器、螺旋断层放射治疗装置、机械臂放射治疗装置、\n γ 射线立体定向放射治疗 系统的输出剂量进行测量, 采用空气比释动能校准因子结合 AAPM 方案中的要求对低能 X 射线治疗机的输出剂量进 行测量, 并与预置照射值进行比较分析输出剂量相对偏差。\n 结果共调查了 59 台外照射放射治疗设备的输出剂量, 49 台医用电子加速器中 2 台输出剂量相对偏差绝对值超过 3%; 其他外照射放射治疗设备的输出剂量相对偏差绝对值\n 均小于 3%。\n 结论 统一明确的测量方法和多方的质量控制检测可确保放射治疗设备输出剂量的准确, 在设备常规运 行中医院应严格执行周期性检验保证实施精确可靠的放射治疗。","PeriodicalId":58844,"journal":{"name":"中国辐射卫生","volume":" ","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2021-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Output dose investigation and analysis of external beam radiotherapy in Beijing\",\"authors\":\"Feng Zechen, Ma Yongzhong, W. Hongfang, Su Yaru, Zhang Zhibin, Lou Yun\",\"doi\":\"10.13491/J.ISSN.1004-714X.2021.02.001\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Objective To study the output dose of external beam radiotherapy in Beijing and analyze the\\n cause of output dose deviation. Guidance is given on the selection of measurement\\n methods, calculation formulas and parameters.\\n Methods During 2020, the output doses of medical electron accelerators, tomotherapy, cyberknife,\\n and γ-knife were measured based on the water absorbed dose calibration factor combined\\n with the requirements in the IAEA TRS398 and TRS483 reports, and low-energy X-rays\\n were measured based on the air kerma calibration factor combined with the requirements\\n in the AAPM protocal. The output dose was measured and compared with the preset irradiation\\n value.\\n Results A total of 59 external beam radiation therapy equipments’s output dose were investigated.\\n 2 of the 49 accelerators’s absolute value of relative deviation between the system\\n indication value and the measurement exceeded 3%; the deviation of other external\\n beam radiotherapy equipments was less than 3%.\\n Conclusion A unified method and multi-party quality control can ensure the accuracy of the output\\n dose. In the routine operation of the equipment, the hospital should strictly implement\\n periodic inspections to ensure accurate and reliable radiotherapy.\\n 摘要: 目的 调查北京地方所属外照射放射治疗设备输出剂量的情况, 分析输出剂量相对偏差原因, 明确测量方法、 计算公式及参数的选取。\\n 方法 2020 年期间, 采用水中吸收剂量校准因子结合国际原子能机构 (IAEA) TRS398 和 TRS483 号报告中的要求对医用电子加速器、螺旋断层放射治疗装置、机械臂放射治疗装置、\\n γ 射线立体定向放射治疗 系统的输出剂量进行测量, 采用空气比释动能校准因子结合 AAPM 方案中的要求对低能 X 射线治疗机的输出剂量进 行测量, 并与预置照射值进行比较分析输出剂量相对偏差。\\n 结果共调查了 59 台外照射放射治疗设备的输出剂量, 49 台医用电子加速器中 2 台输出剂量相对偏差绝对值超过 3%; 其他外照射放射治疗设备的输出剂量相对偏差绝对值\\n 均小于 3%。\\n 结论 统一明确的测量方法和多方的质量控制检测可确保放射治疗设备输出剂量的准确, 在设备常规运 行中医院应严格执行周期性检验保证实施精确可靠的放射治疗。\",\"PeriodicalId\":58844,\"journal\":{\"name\":\"中国辐射卫生\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2021-04-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"中国辐射卫生\",\"FirstCategoryId\":\"1087\",\"ListUrlMain\":\"https://doi.org/10.13491/J.ISSN.1004-714X.2021.02.001\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"中国辐射卫生","FirstCategoryId":"1087","ListUrlMain":"https://doi.org/10.13491/J.ISSN.1004-714X.2021.02.001","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

Objective To study the output dose of external beam radiation in Beijing and analyze the cause of output dose deviation Guidance is given on the selection of measurement methods, calculation formulas and parameters Methods During 2020, the output doses of medical electron accelerators, tomotherapy, cyberknife, and γ- Know were measured based on the water absorbed dose calibration factor combined with the requirements in the IAEA TRS398 and TRS483 reports, and low energy X-rays were measured based on the air kernel calibration factor combined with the requirements in the AAPM protocol The output did was measured and compared with the preset radiation value Results A total of 59 external beam radiation therapy equipment's output did not have been invested 2 of the 49 accelerators' absolute value of relative deviation between the system indication value and the measurement exceeded 3%; The deviation of other external beam radiation equipment was less than 3% Conclusion A unified method and multi party quality control can ensure the accuracy of the output dose In the routing operation of the equipment, the hospital should strictly implement periodic inspections to ensure accuracy and related radiation Abstract: Objective: To investigate the output dose of external radiation therapy equipment in Beijing, analyze the reasons for the relative deviation of output dose, and clarify the selection of measurement methods, calculation formulas, and parameters. During the period of 2020, the calibration factor for absorbed dose in water was used in conjunction with the requirements of the International Atomic Energy Agency (IAEA) reports TRS398 and TRS483 for medical electron accelerators, spiral tomography radiotherapy devices, robotic arm radiotherapy devices γ The output dose of the X-ray stereotactic radiotherapy system is measured, using an air specific release energy calibration factor combined with the requirements of the AAPM scheme to measure the output dose of the low-energy X-ray therapy machine, and comparing and analyzing the relative deviation of the output dose with the preset exposure value. The results showed that a total of 59 external radiation therapy equipment were investigated for their output doses, and 2 out of 49 medical electron accelerators had relative deviation absolute values of output doses exceeding 3%; The absolute deviation of the output dose of other external radiation therapy equipment is less than 3%. Conclusion: A unified and clear measurement method and multi-party quality control testing can ensure the accuracy of the output dose of radiation therapy equipment. Traditional Chinese medicine hospitals should strictly implement periodic testing during the routine operation of the equipment to ensure the implementation of accurate and reliable radiation therapy.
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Output dose investigation and analysis of external beam radiotherapy in Beijing
Objective To study the output dose of external beam radiotherapy in Beijing and analyze the cause of output dose deviation. Guidance is given on the selection of measurement methods, calculation formulas and parameters. Methods During 2020, the output doses of medical electron accelerators, tomotherapy, cyberknife, and γ-knife were measured based on the water absorbed dose calibration factor combined with the requirements in the IAEA TRS398 and TRS483 reports, and low-energy X-rays were measured based on the air kerma calibration factor combined with the requirements in the AAPM protocal. The output dose was measured and compared with the preset irradiation value. Results A total of 59 external beam radiation therapy equipments’s output dose were investigated. 2 of the 49 accelerators’s absolute value of relative deviation between the system indication value and the measurement exceeded 3%; the deviation of other external beam radiotherapy equipments was less than 3%. Conclusion A unified method and multi-party quality control can ensure the accuracy of the output dose. In the routine operation of the equipment, the hospital should strictly implement periodic inspections to ensure accurate and reliable radiotherapy. 摘要: 目的 调查北京地方所属外照射放射治疗设备输出剂量的情况, 分析输出剂量相对偏差原因, 明确测量方法、 计算公式及参数的选取。 方法 2020 年期间, 采用水中吸收剂量校准因子结合国际原子能机构 (IAEA) TRS398 和 TRS483 号报告中的要求对医用电子加速器、螺旋断层放射治疗装置、机械臂放射治疗装置、 γ 射线立体定向放射治疗 系统的输出剂量进行测量, 采用空气比释动能校准因子结合 AAPM 方案中的要求对低能 X 射线治疗机的输出剂量进 行测量, 并与预置照射值进行比较分析输出剂量相对偏差。 结果共调查了 59 台外照射放射治疗设备的输出剂量, 49 台医用电子加速器中 2 台输出剂量相对偏差绝对值超过 3%; 其他外照射放射治疗设备的输出剂量相对偏差绝对值 均小于 3%。 结论 统一明确的测量方法和多方的质量控制检测可确保放射治疗设备输出剂量的准确, 在设备常规运 行中医院应严格执行周期性检验保证实施精确可靠的放射治疗。
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
0.80
自引率
0.00%
发文量
7142
期刊介绍: Chinese Journal of Radiological Health is one of the Source Journals for Chinese Scientific and Technical Papers and Citations and belongs to the series published by Chinese Preventive Medicine Association (CPMA). It is a national academic journal supervised by National Health Commission of the People’s Republic of China and co-sponsored by Institute of Radiation Medicine, Shandong Academy of Medical Sciences and CPMA, and is a professional academic journal publishing research findings and management experience in the field of radiological health, issued to the public in China and abroad. Under the guidance of the Communist Party of China and the national press and publication policies, the Journal actively publicizes the guidelines and policies of the Party and the state on health work, promotes the implementation of relevant laws, regulations and standards, and timely reports new achievements, new information, new methods and new products in the specialty, with the aim of organizing and promoting the academic communication of radiological health in China and improving the academic level of the specialty, and for the purpose of protecting the health of radiation workers and the public while promoting the extensive use of radioisotopes and radiation devices in the national economy. The main columns include Original Articles, Expert Comments, Experience Exchange, Standards and Guidelines, and Review Articles.
期刊最新文献
Analysis of gross radioactivity in drinking water around Tianwan Nuclear Power Plant from 2016 to 2018 Current research status of ionizing radiation bleeding syndrome The predictive value of MSCT imaging features on the pathological risk of gastrointestinal stromal tumors Analysis of quality control test results of medical electron linear accelerators in Guangxi Province 2017—2019 Analysis on the distribution status and concentration degree of radiological diagnosis and treatment resources in Beijing
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1