微创手术中腹腔镜抓取器与大肠之间的摩擦行为

IF 1.6 Q4 ENGINEERING, BIOMEDICAL Biosurface and Biotribology Pub Date : 2021-11-16 DOI:10.1049/bsb2.12028
Yan Guo, Chengmo Cai, Wei Li
{"title":"微创手术中腹腔镜抓取器与大肠之间的摩擦行为","authors":"Yan Guo,&nbsp;Chengmo Cai,&nbsp;Wei Li","doi":"10.1049/bsb2.12028","DOIUrl":null,"url":null,"abstract":"<p>Slippage is a common phenomenon between laparoscopic graspers and tissues during minimally invasive surgery, which may lead to inefficient surgical operations, prolonged operation time, and increased patient suffering. The stability factors related to the friction behaviour between laparoscopic graspers and the large intestine, including bio-surface liquids, pulling angle, and surface profile of graspers, were studied. The friction behaviour at the large intestine–grasper interface was tested using a UMT-II tribometer under the conditions of clamping force of 1–4 N, sliding displacement of 15 mm, and sliding velocity of 2 mm/s to simulate the grasping and pulling operations of soft tissue. The results showed that the bio-surface liquid (serum) of the large intestine significantly decreased the friction coefficient, thus reducing the grasping efficiency. A pulling angle of 15° could generate the peak frictional force and enhance the grasping stability. The frictional force increased with the ratio of the profile surface area of the grasper. These results demonstrate that the grasping stability can be improved by changing either the bio-surface liquid condition or the pulling angle. In addition, a grasper with a larger profile surface area can also prevent slippage due to its significant influence on the pressure distribution and actual contact area for tissue retention.</p>","PeriodicalId":52235,"journal":{"name":"Biosurface and Biotribology","volume":"8 1","pages":"58-68"},"PeriodicalIF":1.6000,"publicationDate":"2021-11-16","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://ietresearch.onlinelibrary.wiley.com/doi/epdf/10.1049/bsb2.12028","citationCount":"1","resultStr":"{\"title\":\"Friction behaviour between a laparoscopic grasper and the large intestine during minimally invasive surgery\",\"authors\":\"Yan Guo,&nbsp;Chengmo Cai,&nbsp;Wei Li\",\"doi\":\"10.1049/bsb2.12028\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p>Slippage is a common phenomenon between laparoscopic graspers and tissues during minimally invasive surgery, which may lead to inefficient surgical operations, prolonged operation time, and increased patient suffering. The stability factors related to the friction behaviour between laparoscopic graspers and the large intestine, including bio-surface liquids, pulling angle, and surface profile of graspers, were studied. The friction behaviour at the large intestine–grasper interface was tested using a UMT-II tribometer under the conditions of clamping force of 1–4 N, sliding displacement of 15 mm, and sliding velocity of 2 mm/s to simulate the grasping and pulling operations of soft tissue. The results showed that the bio-surface liquid (serum) of the large intestine significantly decreased the friction coefficient, thus reducing the grasping efficiency. A pulling angle of 15° could generate the peak frictional force and enhance the grasping stability. The frictional force increased with the ratio of the profile surface area of the grasper. These results demonstrate that the grasping stability can be improved by changing either the bio-surface liquid condition or the pulling angle. In addition, a grasper with a larger profile surface area can also prevent slippage due to its significant influence on the pressure distribution and actual contact area for tissue retention.</p>\",\"PeriodicalId\":52235,\"journal\":{\"name\":\"Biosurface and Biotribology\",\"volume\":\"8 1\",\"pages\":\"58-68\"},\"PeriodicalIF\":1.6000,\"publicationDate\":\"2021-11-16\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://ietresearch.onlinelibrary.wiley.com/doi/epdf/10.1049/bsb2.12028\",\"citationCount\":\"1\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Biosurface and Biotribology\",\"FirstCategoryId\":\"1087\",\"ListUrlMain\":\"https://onlinelibrary.wiley.com/doi/10.1049/bsb2.12028\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"ENGINEERING, BIOMEDICAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Biosurface and Biotribology","FirstCategoryId":"1087","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1049/bsb2.12028","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 1

摘要

在微创手术中,腹腔镜握持器与组织之间普遍存在滑脱现象,可能导致手术效率低下,延长手术时间,增加患者痛苦。研究了影响腹腔镜抓握器与大肠摩擦行为的因素,包括生物表面液体、抓握器的拉伸角度和抓握器的表面轮廓。采用UMT-II型摩擦计,在夹紧力为1 ~ 4 N、滑动位移为15 mm、滑动速度为2 mm/s的条件下,测试了大肠-抓取器界面的摩擦行为,模拟了软组织的抓取和抓取操作。结果表明,大肠生物表面液(血清)明显降低了摩擦系数,从而降低了抓取效率。当拉力角为15°时,可产生最大摩擦力,提高抓握稳定性。摩擦力随抓握器轮廓表面积的比值增大而增大。实验结果表明,通过改变生物表面液体条件和抓取角度可以提高抓取的稳定性。此外,由于对压力分布和组织保留的实际接触面积有重大影响,具有较大轮廓表面积的抓握器也可以防止滑移。
本文章由计算机程序翻译,如有差异,请以英文原文为准。

摘要图片

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Friction behaviour between a laparoscopic grasper and the large intestine during minimally invasive surgery

Slippage is a common phenomenon between laparoscopic graspers and tissues during minimally invasive surgery, which may lead to inefficient surgical operations, prolonged operation time, and increased patient suffering. The stability factors related to the friction behaviour between laparoscopic graspers and the large intestine, including bio-surface liquids, pulling angle, and surface profile of graspers, were studied. The friction behaviour at the large intestine–grasper interface was tested using a UMT-II tribometer under the conditions of clamping force of 1–4 N, sliding displacement of 15 mm, and sliding velocity of 2 mm/s to simulate the grasping and pulling operations of soft tissue. The results showed that the bio-surface liquid (serum) of the large intestine significantly decreased the friction coefficient, thus reducing the grasping efficiency. A pulling angle of 15° could generate the peak frictional force and enhance the grasping stability. The frictional force increased with the ratio of the profile surface area of the grasper. These results demonstrate that the grasping stability can be improved by changing either the bio-surface liquid condition or the pulling angle. In addition, a grasper with a larger profile surface area can also prevent slippage due to its significant influence on the pressure distribution and actual contact area for tissue retention.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Biosurface and Biotribology
Biosurface and Biotribology Engineering-Mechanical Engineering
CiteScore
1.70
自引率
0.00%
发文量
27
审稿时长
11 weeks
期刊最新文献
Protein hydrogels for biomedical applications Flow field characteristics and drag reduction performance of high–low velocity stripes on the biomimetic imbricated fish scale surfaces Advancements and challenges in bionic joint lubrication biomaterials for sports medicine Biofunctionalisation strategies of material surface and the inspired biological effects for bone repair Enhancing the biological functionality of poly (lactic-co-glycolic acid) cage-like structures through surface modification with micro- and nano-sized hydroxyapatite particles
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1