{"title":"胸骨闭合系统生理测试方法的开发与验证","authors":"Annette Kienle, Nadine Renner, Constanze Bischoff","doi":"10.1115/1.4053687","DOIUrl":null,"url":null,"abstract":"\n After a median sternotomy, the sternal bone has to be refixated using implants such as wires, plates or clamps. However, there is still a lack of specifically applicable test methods to investigate the mechanical safety and effectiveness of such implant systems. The aim of this study was to develop a new test method that replicates the in vivo loads acting on the sternum and that is applicable to all types of sternal closure systems.\n Based on the scientific literature a setup was developed that incorporates the physiological loads acting on the sternum during breathing (91 N) and coughing (328 N). These loads are applied to a sternum replicate at 5 Hz for approximately 1.8 million cycles. This cycle number is assumed to represent the healing period. For validation, the new method was applied to two different sternal closure systems: a PEEK clamp system and wires.\n The new test method proved to be easily applicable. The validation tests with the two sternal closure systems showed reasonable and reproducible results regarding all outcome parameters. The pre-tension exerted by the implants significantly differed between the two implant groups and decreased after the first coughing cycles. The fracture gap separation during breathing also significantly different between the two test groups but it was similar during coughing. No implant failed.\n Using this new test method it is possible to compare sternal closure systems under reproducible conditions and interpret their mechanical characteristics regarding their in vivo safety and effectiveness.","PeriodicalId":49305,"journal":{"name":"Journal of Medical Devices-Transactions of the Asme","volume":null,"pages":null},"PeriodicalIF":0.8000,"publicationDate":"2022-01-26","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Development and Validation of a Physiological Testing Method for Sternal Closure Systems\",\"authors\":\"Annette Kienle, Nadine Renner, Constanze Bischoff\",\"doi\":\"10.1115/1.4053687\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"\\n After a median sternotomy, the sternal bone has to be refixated using implants such as wires, plates or clamps. However, there is still a lack of specifically applicable test methods to investigate the mechanical safety and effectiveness of such implant systems. The aim of this study was to develop a new test method that replicates the in vivo loads acting on the sternum and that is applicable to all types of sternal closure systems.\\n Based on the scientific literature a setup was developed that incorporates the physiological loads acting on the sternum during breathing (91 N) and coughing (328 N). These loads are applied to a sternum replicate at 5 Hz for approximately 1.8 million cycles. This cycle number is assumed to represent the healing period. For validation, the new method was applied to two different sternal closure systems: a PEEK clamp system and wires.\\n The new test method proved to be easily applicable. The validation tests with the two sternal closure systems showed reasonable and reproducible results regarding all outcome parameters. The pre-tension exerted by the implants significantly differed between the two implant groups and decreased after the first coughing cycles. The fracture gap separation during breathing also significantly different between the two test groups but it was similar during coughing. No implant failed.\\n Using this new test method it is possible to compare sternal closure systems under reproducible conditions and interpret their mechanical characteristics regarding their in vivo safety and effectiveness.\",\"PeriodicalId\":49305,\"journal\":{\"name\":\"Journal of Medical Devices-Transactions of the Asme\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.8000,\"publicationDate\":\"2022-01-26\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Medical Devices-Transactions of the Asme\",\"FirstCategoryId\":\"5\",\"ListUrlMain\":\"https://doi.org/10.1115/1.4053687\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"ENGINEERING, BIOMEDICAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Medical Devices-Transactions of the Asme","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1115/1.4053687","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0

摘要

胸骨正中切开术后,胸骨必须用金属丝、钢板或夹子等植入物进行再固定。然而,仍然缺乏专门适用的测试方法来研究这种植入系统的机械安全性和有效性。本研究的目的是开发一种新的测试方法,该方法可以复制作用在胸骨上的体内负荷,并适用于所有类型的胸骨闭合系统。在科学文献的基础上,开发了一种装置,该装置结合了呼吸(91牛)和咳嗽(328牛)时作用于胸骨的生理负荷。这些负荷以5赫兹的频率在胸骨上重复约180万次。这个周期数被假定为表示治疗周期。为了验证,新方法应用于两种不同的胸骨闭合系统:PEEK夹紧系统和钢丝。新的测试方法被证明是易于应用的。两种胸骨闭合系统的验证试验在所有结局参数方面显示出合理和可重复的结果。两组患者的预张力差异显著,且在第一次咳嗽后降低。呼吸时骨折间隙分离在两组之间也有显著差异,但在咳嗽时相似。没有植入失败。使用这种新的测试方法,可以在可重复的条件下比较胸骨闭合系统,并解释其体内安全性和有效性的机械特性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Development and Validation of a Physiological Testing Method for Sternal Closure Systems
After a median sternotomy, the sternal bone has to be refixated using implants such as wires, plates or clamps. However, there is still a lack of specifically applicable test methods to investigate the mechanical safety and effectiveness of such implant systems. The aim of this study was to develop a new test method that replicates the in vivo loads acting on the sternum and that is applicable to all types of sternal closure systems. Based on the scientific literature a setup was developed that incorporates the physiological loads acting on the sternum during breathing (91 N) and coughing (328 N). These loads are applied to a sternum replicate at 5 Hz for approximately 1.8 million cycles. This cycle number is assumed to represent the healing period. For validation, the new method was applied to two different sternal closure systems: a PEEK clamp system and wires. The new test method proved to be easily applicable. The validation tests with the two sternal closure systems showed reasonable and reproducible results regarding all outcome parameters. The pre-tension exerted by the implants significantly differed between the two implant groups and decreased after the first coughing cycles. The fracture gap separation during breathing also significantly different between the two test groups but it was similar during coughing. No implant failed. Using this new test method it is possible to compare sternal closure systems under reproducible conditions and interpret their mechanical characteristics regarding their in vivo safety and effectiveness.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
1.80
自引率
11.10%
发文量
56
审稿时长
6-12 weeks
期刊介绍: The Journal of Medical Devices presents papers on medical devices that improve diagnostic, interventional and therapeutic treatments focusing on applied research and the development of new medical devices or instrumentation. It provides special coverage of novel devices that allow new surgical strategies, new methods of drug delivery, or possible reductions in the complexity, cost, or adverse results of health care. The Design Innovation category features papers focusing on novel devices, including papers with limited clinical or engineering results. The Medical Device News section provides coverage of advances, trends, and events.
期刊最新文献
A Novel Design Method for the Knee Joint of the Exoskeleton Based On the Modular Wearable Sensor Experimental Investigation of the Calcified Plaque Material Removal Rate in Coronary Rotational Atherectomy Assessment of a Novel Application of the Capture-Trap-Terminate Approach for Treating Aerosol Products During Dental Procedures Development And Mechanical Testing Of Implant For Cranial Reconstruction After Burr Hole Trepanation In Vitro Thrombogenicity Testing of Biomaterials in a Dynamic Flow Loop: Effects of Length and Quantity of Test Samples.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1