{"title":"用于骨科触觉仿真的机器人单轴力矩装置的设计综合","authors":"T. Cotter, R. Mongrain, Mark Driscoll","doi":"10.1115/1.4054344","DOIUrl":null,"url":null,"abstract":"\n Robotic devices are commonly used in surgical simulators to provide tactile, or haptic, feedback. They can provide customized feedback that can be rapidly modified with minimal hardware changes in comparison to non-robotic systems. This work describes the design, development, and evaluation of one such tool: a novel uniaxial torque haptic device for a surgical training simulator. The objective of the work was to design a single connection haptic device that could augment an existing six degree of freedom haptic device to mimic a Concorde Clear vacuum curette. Design and evaluations focused on the tool's ability to deliver adequate torque, imitate a surgical tool, and be integrated into the haptic device. Twenty-nine surgeons tested the tool in the simulator and evaluated it via a questionnaire. The device was found to deliver the 800 N·mm of torque necessary to mimic an orthopedic procedure. Surgeons found it accurately imitated surgical tool physical appearance and maneuverability, scoring them 3.9±1.0 and 3.3±1.2, respectively, on a 1-5 Likert scale. By virtue of the functionality necessary for testing and evaluation, the device could be connected to the haptic device for mechanical and electrical engagement. This device is a step forward in the field of augmentable haptic devices for surgical simulation. By changing the number of robotically-controlled degrees of freedom of a haptic device, existing devices can be tuned to meet the demands of a particular simulator, which has the potential to improve surgeon training standards.","PeriodicalId":49305,"journal":{"name":"Journal of Medical Devices-Transactions of the Asme","volume":null,"pages":null},"PeriodicalIF":0.8000,"publicationDate":"2022-04-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":"{\"title\":\"Design Synthesis of a Robotic Uniaxial Torque Device for Orthopedic Haptic Simulation\",\"authors\":\"T. Cotter, R. Mongrain, Mark Driscoll\",\"doi\":\"10.1115/1.4054344\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"\\n Robotic devices are commonly used in surgical simulators to provide tactile, or haptic, feedback. They can provide customized feedback that can be rapidly modified with minimal hardware changes in comparison to non-robotic systems. This work describes the design, development, and evaluation of one such tool: a novel uniaxial torque haptic device for a surgical training simulator. The objective of the work was to design a single connection haptic device that could augment an existing six degree of freedom haptic device to mimic a Concorde Clear vacuum curette. Design and evaluations focused on the tool's ability to deliver adequate torque, imitate a surgical tool, and be integrated into the haptic device. Twenty-nine surgeons tested the tool in the simulator and evaluated it via a questionnaire. The device was found to deliver the 800 N·mm of torque necessary to mimic an orthopedic procedure. Surgeons found it accurately imitated surgical tool physical appearance and maneuverability, scoring them 3.9±1.0 and 3.3±1.2, respectively, on a 1-5 Likert scale. By virtue of the functionality necessary for testing and evaluation, the device could be connected to the haptic device for mechanical and electrical engagement. This device is a step forward in the field of augmentable haptic devices for surgical simulation. By changing the number of robotically-controlled degrees of freedom of a haptic device, existing devices can be tuned to meet the demands of a particular simulator, which has the potential to improve surgeon training standards.\",\"PeriodicalId\":49305,\"journal\":{\"name\":\"Journal of Medical Devices-Transactions of the Asme\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.8000,\"publicationDate\":\"2022-04-14\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"1\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Medical Devices-Transactions of the Asme\",\"FirstCategoryId\":\"5\",\"ListUrlMain\":\"https://doi.org/10.1115/1.4054344\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"ENGINEERING, BIOMEDICAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Medical Devices-Transactions of the Asme","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1115/1.4054344","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 1

摘要

机器人设备通常用于外科模拟器,以提供触觉或触觉反馈。与非机器人系统相比,它们可以提供定制的反馈,与最小的硬件更改相比,可以快速修改。这项工作描述了一种这样的工具的设计、开发和评估:一种用于外科训练模拟器的新型单轴扭矩触觉装置。这项工作的目的是设计一个单连接触觉装置,可以增加现有的六自由度触觉装置,以模仿协和式真空刮管。设计和评估的重点是工具提供足够扭矩的能力,模仿手术工具,并集成到触觉设备中。29名外科医生在模拟器中测试了该工具,并通过问卷对其进行了评估。研究发现,该装置可提供模拟骨科手术所需的800 N·mm扭矩。外科医生发现它准确地模仿了手术工具的物理外观和可操作性,在1-5的李克特量表上分别获得3.9±1.0和3.3±1.2分。由于测试和评估所需的功能,该设备可以连接到触觉设备进行机械和电气接触。该设备是外科模拟增强触觉设备领域的一个进步。通过改变触觉设备的机器人控制自由度的数量,现有的设备可以调整以满足特定模拟器的要求,这有可能提高外科医生的培训标准。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Design Synthesis of a Robotic Uniaxial Torque Device for Orthopedic Haptic Simulation
Robotic devices are commonly used in surgical simulators to provide tactile, or haptic, feedback. They can provide customized feedback that can be rapidly modified with minimal hardware changes in comparison to non-robotic systems. This work describes the design, development, and evaluation of one such tool: a novel uniaxial torque haptic device for a surgical training simulator. The objective of the work was to design a single connection haptic device that could augment an existing six degree of freedom haptic device to mimic a Concorde Clear vacuum curette. Design and evaluations focused on the tool's ability to deliver adequate torque, imitate a surgical tool, and be integrated into the haptic device. Twenty-nine surgeons tested the tool in the simulator and evaluated it via a questionnaire. The device was found to deliver the 800 N·mm of torque necessary to mimic an orthopedic procedure. Surgeons found it accurately imitated surgical tool physical appearance and maneuverability, scoring them 3.9±1.0 and 3.3±1.2, respectively, on a 1-5 Likert scale. By virtue of the functionality necessary for testing and evaluation, the device could be connected to the haptic device for mechanical and electrical engagement. This device is a step forward in the field of augmentable haptic devices for surgical simulation. By changing the number of robotically-controlled degrees of freedom of a haptic device, existing devices can be tuned to meet the demands of a particular simulator, which has the potential to improve surgeon training standards.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
1.80
自引率
11.10%
发文量
56
审稿时长
6-12 weeks
期刊介绍: The Journal of Medical Devices presents papers on medical devices that improve diagnostic, interventional and therapeutic treatments focusing on applied research and the development of new medical devices or instrumentation. It provides special coverage of novel devices that allow new surgical strategies, new methods of drug delivery, or possible reductions in the complexity, cost, or adverse results of health care. The Design Innovation category features papers focusing on novel devices, including papers with limited clinical or engineering results. The Medical Device News section provides coverage of advances, trends, and events.
期刊最新文献
Mechanical Viability and Functionality Assessment of a New Sutureless Endoluminal Microvascular Device: A Preliminary In Vivo Rabbit Study. Design and Implementation of a Computer-Controlled Hybrid Oscillatory Ventilator. Controlled Ice Nucleation With a Sand-PDMS Film Device Enhances Cryopreservation of Mouse Preantral Ovarian Follicles. A Novel Design Method for the Knee Joint of the Exoskeleton Based On the Modular Wearable Sensor Experimental Investigation of the Calcified Plaque Material Removal Rate in Coronary Rotational Atherectomy
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1