Fadillah Ahmad, Afdhal Muttaqin, D. Fitriyani, Ridwan Ridwan
{"title":"基于多叶准直变化的食相治疗计划系统(TPS)中各向异性分析算法(AAA)的剂量学验证","authors":"Fadillah Ahmad, Afdhal Muttaqin, D. Fitriyani, Ridwan Ridwan","doi":"10.25077/jif.15.1.48-55.2023","DOIUrl":null,"url":null,"abstract":"Dosimetry verification of anisotropic analytical (AAA) algorithms based on variations of the multileaf collimator has been investigated using the Eclipse Treatment Planning System. This study used a 0.6cc ionization chamber farmer detector and Linac Clinical CX variant equipped with TPS Eclipse with AAA. This study used 6 MV energy and 2 Gy dose. The multileaf collimator was varied into six groups with the size (A, B, C, D, E, F) of the irradiation field used 20 cm x 20 cm. The measurement results were a dose deviation value or a dose ratio presentation in each irradiation area. The dose deviation of the multileaf collimator variation was A 0.86% for each group. Group B had 6.8%, Group C had -0.43%, Group D had 0.73%, Group E had 1.11%, and Group F had 0.84%. The mean dose deviation value for all multileaf collimator forms was 1.67%, where this value is within the tolerance value recommended by ICRU, namely 3-5%. The p-value in the Analysis of variance (ANOVA) for the entire group was 0.00. This data shows that there is no effect of variation in the multileaf collimator on the dose given.","PeriodicalId":52720,"journal":{"name":"JIF Jurnal Ilmu Fisika","volume":" ","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2022-12-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Dosimetry Verification of Anisotropic Analytical Algorithm (AAA) In Eclipse Treatment Planning System (TPS) Based on Multileaf Collimator Variations\",\"authors\":\"Fadillah Ahmad, Afdhal Muttaqin, D. Fitriyani, Ridwan Ridwan\",\"doi\":\"10.25077/jif.15.1.48-55.2023\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Dosimetry verification of anisotropic analytical (AAA) algorithms based on variations of the multileaf collimator has been investigated using the Eclipse Treatment Planning System. This study used a 0.6cc ionization chamber farmer detector and Linac Clinical CX variant equipped with TPS Eclipse with AAA. This study used 6 MV energy and 2 Gy dose. The multileaf collimator was varied into six groups with the size (A, B, C, D, E, F) of the irradiation field used 20 cm x 20 cm. The measurement results were a dose deviation value or a dose ratio presentation in each irradiation area. The dose deviation of the multileaf collimator variation was A 0.86% for each group. Group B had 6.8%, Group C had -0.43%, Group D had 0.73%, Group E had 1.11%, and Group F had 0.84%. The mean dose deviation value for all multileaf collimator forms was 1.67%, where this value is within the tolerance value recommended by ICRU, namely 3-5%. The p-value in the Analysis of variance (ANOVA) for the entire group was 0.00. This data shows that there is no effect of variation in the multileaf collimator on the dose given.\",\"PeriodicalId\":52720,\"journal\":{\"name\":\"JIF Jurnal Ilmu Fisika\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2022-12-30\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"JIF Jurnal Ilmu Fisika\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.25077/jif.15.1.48-55.2023\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"JIF Jurnal Ilmu Fisika","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.25077/jif.15.1.48-55.2023","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
摘要
使用Eclipse治疗计划系统研究了基于多叶准直器变化的各向异性分析(AAA)算法的剂量测定验证。本研究使用了0.6cc电离室农民探测器和配备了TPS Eclipse和AAA的Linac Clinical CX变体。本研究使用6MV能量和2Gy剂量。将多叶准直器分为六组,照射场大小(A、B、C、D、E、F)为20 cm x 20 cm。测量结果为每个照射区域的剂量偏差值或剂量比。各组多叶准直器变化的剂量偏差为A 0.86%。B组为6.8%,C组为-0.43%,D组为0.73%,E组为1.11%,F组为0.84%。所有多叶准直器形式的平均剂量偏差值为1.67%,在ICRU推荐的容许值范围内,即3-5%。整个组的方差分析(ANOVA)中的p值为0.00。该数据表明,多叶准直器的变化对所给剂量没有影响。
Dosimetry Verification of Anisotropic Analytical Algorithm (AAA) In Eclipse Treatment Planning System (TPS) Based on Multileaf Collimator Variations
Dosimetry verification of anisotropic analytical (AAA) algorithms based on variations of the multileaf collimator has been investigated using the Eclipse Treatment Planning System. This study used a 0.6cc ionization chamber farmer detector and Linac Clinical CX variant equipped with TPS Eclipse with AAA. This study used 6 MV energy and 2 Gy dose. The multileaf collimator was varied into six groups with the size (A, B, C, D, E, F) of the irradiation field used 20 cm x 20 cm. The measurement results were a dose deviation value or a dose ratio presentation in each irradiation area. The dose deviation of the multileaf collimator variation was A 0.86% for each group. Group B had 6.8%, Group C had -0.43%, Group D had 0.73%, Group E had 1.11%, and Group F had 0.84%. The mean dose deviation value for all multileaf collimator forms was 1.67%, where this value is within the tolerance value recommended by ICRU, namely 3-5%. The p-value in the Analysis of variance (ANOVA) for the entire group was 0.00. This data shows that there is no effect of variation in the multileaf collimator on the dose given.