{"title":"Nuss整形术治疗漏斗胸的模拟","authors":"Yan-bin Wei, Yi-sheng Shi, Huan Wang, Ya Gao","doi":"10.1109/BMEI.2009.5305262","DOIUrl":null,"url":null,"abstract":"Nuss Orthopedic of Pectus Excavatum(PE) is a novel minimal invasive surgical technique for the treatment of PE, however, the surgery will fail as the Nuss bar is slipped or rotated. A 3D model for thorax is built according to the helical CT images of PE, the process of Nuss orthopedic was simulated by finite element analysis. The results show as the depressed sternum was jacking to normal position gradually by Nuss bar, the sternum, the ribs and the rib cartilages will all move upward which make the Nuss bar support the sternum at the location lower about 1~6 mm than the end of depressed sternum end. However, original plan is that the bar should be supported just at the end of depressed sternum. After the bar is fixed, there is a torque on the bar for the pressure imposed by sternum and the force produced by the fixed place of bar forms a pair of torque. This torque may be the reason for the failure of surgery. The improvement is that the fix place of bar should be move upward 1~6mm or a distance of one rib. Keywords-pectus excavatum; Nuss orthopedic; 3D modeling; finite element; simulation","PeriodicalId":6389,"journal":{"name":"2009 2nd International Conference on Biomedical Engineering and Informatics","volume":"2 1","pages":"1-4"},"PeriodicalIF":0.0000,"publicationDate":"2009-10-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"4","resultStr":"{\"title\":\"Simulation of Nuss Orthopedic for Pectus Excavatum\",\"authors\":\"Yan-bin Wei, Yi-sheng Shi, Huan Wang, Ya Gao\",\"doi\":\"10.1109/BMEI.2009.5305262\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Nuss Orthopedic of Pectus Excavatum(PE) is a novel minimal invasive surgical technique for the treatment of PE, however, the surgery will fail as the Nuss bar is slipped or rotated. A 3D model for thorax is built according to the helical CT images of PE, the process of Nuss orthopedic was simulated by finite element analysis. The results show as the depressed sternum was jacking to normal position gradually by Nuss bar, the sternum, the ribs and the rib cartilages will all move upward which make the Nuss bar support the sternum at the location lower about 1~6 mm than the end of depressed sternum end. However, original plan is that the bar should be supported just at the end of depressed sternum. After the bar is fixed, there is a torque on the bar for the pressure imposed by sternum and the force produced by the fixed place of bar forms a pair of torque. This torque may be the reason for the failure of surgery. The improvement is that the fix place of bar should be move upward 1~6mm or a distance of one rib. Keywords-pectus excavatum; Nuss orthopedic; 3D modeling; finite element; simulation\",\"PeriodicalId\":6389,\"journal\":{\"name\":\"2009 2nd International Conference on Biomedical Engineering and Informatics\",\"volume\":\"2 1\",\"pages\":\"1-4\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2009-10-30\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"4\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"2009 2nd International Conference on Biomedical Engineering and Informatics\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1109/BMEI.2009.5305262\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"2009 2nd International Conference on Biomedical Engineering and Informatics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/BMEI.2009.5305262","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 4

摘要

Nuss骨科的漏斗胸(PE)是一种新型的微创手术技术,用于PE的治疗,然而,由于Nuss棒滑动或旋转,手术会失败。根据PE的螺旋CT图像建立胸腔三维模型,采用有限元方法模拟Nuss矫形过程。结果表明,当下行胸骨被Nuss棒逐渐顶至正常位置时,胸骨、肋骨和肋软骨均向上移动,使Nuss棒在低于下行胸骨末端约1~ 6mm的位置支撑胸骨。然而,最初的计划是在胸骨凹陷的末端支撑杆。钢条固定后,由于胸骨施加的压力与钢条固定处产生的力在钢条上形成扭矩对。这种扭力可能是手术失败的原因。改进之处是将钢筋的固定位置向上移动1~6mm或一肋的距离。Keywords-pectus举;主犯整形;三维建模;有限元;模拟
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Simulation of Nuss Orthopedic for Pectus Excavatum
Nuss Orthopedic of Pectus Excavatum(PE) is a novel minimal invasive surgical technique for the treatment of PE, however, the surgery will fail as the Nuss bar is slipped or rotated. A 3D model for thorax is built according to the helical CT images of PE, the process of Nuss orthopedic was simulated by finite element analysis. The results show as the depressed sternum was jacking to normal position gradually by Nuss bar, the sternum, the ribs and the rib cartilages will all move upward which make the Nuss bar support the sternum at the location lower about 1~6 mm than the end of depressed sternum end. However, original plan is that the bar should be supported just at the end of depressed sternum. After the bar is fixed, there is a torque on the bar for the pressure imposed by sternum and the force produced by the fixed place of bar forms a pair of torque. This torque may be the reason for the failure of surgery. The improvement is that the fix place of bar should be move upward 1~6mm or a distance of one rib. Keywords-pectus excavatum; Nuss orthopedic; 3D modeling; finite element; simulation
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
A Novel Approach for Blood Vessel Edge Detection in Retinal Images Skin Response During Irradiation by Intense Pulsed Light Based on Optical Imaging Technology and Histology Physical Properties of LYSO Scintillator for NN-PET Detectors A High Security Framework for SMS An Efficient Antenna Selection Algorithm for MIMO Systems
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1