射频消融导管检测仪的研究

Lei Zhang, Wenbo Lu, Jianjun Sun, Chongchong Ai, Zechen Lin, Naishui Han, Longfei Li, Yufei Jia
{"title":"射频消融导管检测仪的研究","authors":"Lei Zhang, Wenbo Lu, Jianjun Sun, Chongchong Ai, Zechen Lin, Naishui Han, Longfei Li, Yufei Jia","doi":"10.32629/jcmr.v4i2.1234","DOIUrl":null,"url":null,"abstract":"Objective — To design a detection instrument for detecting the quality of radiofrequency ablation electrode, which can quickly detect the physical properties of radiofrequency ablation electrode. Methods — The DC constant current source is introduced to measure the small resistance, test whether the thermocouple is well connected, and measure the resistance between the positive and negative wires of the thermocouple at the plug (1-core and 2-core) to determine whether the thermocouple is well connected. The resistance of the good product is about 40Ω. Detect whether the inner tube and outer tube of RF ablation are in good physical contact through resistance, and judge whether there is physical contact between the inner tube and outer tube by detecting the resistance between the positive conductor of thermocouple (2 cores) and RF connecting wire (3 cores). When in good contact, the resistance is about 0.5 Ω; Measure the short circuit between 4~8 cores of the plug and judge whether the code is correct. The measurement method of pipeline absolute tightness is adopted to detect the change of pipeline pressure. Results — The developed radiofrequency ablation electrode detector can accurately measure the resistance, circuit and pipeline tightness of radiofrequency ablation, and can meet the requirements of detection accuracy. Conclusion — The quality of the prototype is stable and can meet the needs of users.","PeriodicalId":15431,"journal":{"name":"Journal of Clinical Medicine Research","volume":"43 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2023-07-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Study on the Instrument for Testing Radiofrequency Ablation Catheter\",\"authors\":\"Lei Zhang, Wenbo Lu, Jianjun Sun, Chongchong Ai, Zechen Lin, Naishui Han, Longfei Li, Yufei Jia\",\"doi\":\"10.32629/jcmr.v4i2.1234\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Objective — To design a detection instrument for detecting the quality of radiofrequency ablation electrode, which can quickly detect the physical properties of radiofrequency ablation electrode. Methods — The DC constant current source is introduced to measure the small resistance, test whether the thermocouple is well connected, and measure the resistance between the positive and negative wires of the thermocouple at the plug (1-core and 2-core) to determine whether the thermocouple is well connected. The resistance of the good product is about 40Ω. Detect whether the inner tube and outer tube of RF ablation are in good physical contact through resistance, and judge whether there is physical contact between the inner tube and outer tube by detecting the resistance between the positive conductor of thermocouple (2 cores) and RF connecting wire (3 cores). When in good contact, the resistance is about 0.5 Ω; Measure the short circuit between 4~8 cores of the plug and judge whether the code is correct. The measurement method of pipeline absolute tightness is adopted to detect the change of pipeline pressure. Results — The developed radiofrequency ablation electrode detector can accurately measure the resistance, circuit and pipeline tightness of radiofrequency ablation, and can meet the requirements of detection accuracy. Conclusion — The quality of the prototype is stable and can meet the needs of users.\",\"PeriodicalId\":15431,\"journal\":{\"name\":\"Journal of Clinical Medicine Research\",\"volume\":\"43 1\",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2023-07-07\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Clinical Medicine Research\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.32629/jcmr.v4i2.1234\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Clinical Medicine Research","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.32629/jcmr.v4i2.1234","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

目的-设计一种射频消融电极质量检测仪器,能够快速检测射频消融电极的物理性能。方法-采用直流恒流源测量小电阻,测试热电偶连接是否良好,并测量热电偶插头(1芯和2芯)正负极线之间的电阻,以确定热电偶是否连接良好。好产品的电阻约为40Ω。通过电阻检测射频烧蚀内管与外管是否物理接触良好,通过检测热电偶正导体(2芯)与射频连接线(3芯)之间的电阻,判断内管与外管之间是否存在物理接触。接触良好时,电阻约0.5 Ω;测量插头4~8芯之间的短路,判断代码是否正确。采用管道绝对紧密度测量法检测管道压力的变化。结果——研制的射频烧蚀电极检测仪能准确测量射频烧蚀的电阻、电路和管路的紧密性,满足检测精度要求。结论-样机质量稳定,能满足用户需求。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Study on the Instrument for Testing Radiofrequency Ablation Catheter
Objective — To design a detection instrument for detecting the quality of radiofrequency ablation electrode, which can quickly detect the physical properties of radiofrequency ablation electrode. Methods — The DC constant current source is introduced to measure the small resistance, test whether the thermocouple is well connected, and measure the resistance between the positive and negative wires of the thermocouple at the plug (1-core and 2-core) to determine whether the thermocouple is well connected. The resistance of the good product is about 40Ω. Detect whether the inner tube and outer tube of RF ablation are in good physical contact through resistance, and judge whether there is physical contact between the inner tube and outer tube by detecting the resistance between the positive conductor of thermocouple (2 cores) and RF connecting wire (3 cores). When in good contact, the resistance is about 0.5 Ω; Measure the short circuit between 4~8 cores of the plug and judge whether the code is correct. The measurement method of pipeline absolute tightness is adopted to detect the change of pipeline pressure. Results — The developed radiofrequency ablation electrode detector can accurately measure the resistance, circuit and pipeline tightness of radiofrequency ablation, and can meet the requirements of detection accuracy. Conclusion — The quality of the prototype is stable and can meet the needs of users.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
The Effect of Preventive Care and Rehabilitation Nursing on Elderly Hypertension Meta-analysis of In-stent Restenosis Factors after Coronary Intervention Effect of Dietary Carotenoids and Lutein on Eye Health Maintenance in Elderly People One Case of Solar Dermatitis Treated with Umbilical Needle "Four Positions" Combined with Body Acupuncture Primary Presacral Neuroendocrine Tumors Presented by Lumbosacral Pain: a Case Report
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1