眼底相机与智能手机相机附件:图像质量分析

Nicole Barritt, Mohana Kuppuswamy Parthasarathy, Ibrahim Faruq, J. Zelek, V. Lakshminarayanan
{"title":"眼底相机与智能手机相机附件:图像质量分析","authors":"Nicole Barritt, Mohana Kuppuswamy Parthasarathy, Ibrahim Faruq, J. Zelek, V. Lakshminarayanan","doi":"10.1117/12.2528965","DOIUrl":null,"url":null,"abstract":"Fundus cameras are the current clinical standard for capturing retinal images, which are used to diagnose a variety of sight-threatening conditions. Traditional fundus cameras are not easily transported, making them unsuitable for field use. In addition, traditional fundus cameras are expensive. Due to this, a variety of technologies have been developed such as the D-EYE Digital Ophthalmoscope (D-EYE Srl, Padova, Italy) which is compatible with various cellphone cameras. This paper reports on the comparison of the image quality of the Nidek RS-330 OCT Retina Scan Duo (Nidek, Tokyo, Japan) and the D-EYE paired with an iPhone 6 (Apple, Cupertino, USA). Twenty-one participants were enrolled in the study of whom 14 underwent nonmydriatic and mydriatic imaging with the D-EYE and the Nidek. Seven participants underwent nonmydriatic imaging with the D-EYE and the Nidek. The images were co-registered and cropped so that the region of interest was equal in both the D-EYE and Nidek images, as the D-EYE had a smaller field of view. Using the Nidek image as the reference, objective full-reference image quality analysis was performed. Metrics such as structural similarity index and peak signal noise ratio were obtained. It was found that the image quality of the D-EYE is limited by the attached iPhone camera, and is lower when compared to the Nidek. Quantification of the differences between the D-EYE and Nidek allows for targeted development of smartphone camera attachments that can help to bridge the current gap in image quality.","PeriodicalId":10843,"journal":{"name":"Current Developments in Lens Design and Optical Engineering XX","volume":"41 4 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2019-08-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"4","resultStr":"{\"title\":\"Fundus camera versus smartphone camera attachment: image quality analysis\",\"authors\":\"Nicole Barritt, Mohana Kuppuswamy Parthasarathy, Ibrahim Faruq, J. Zelek, V. Lakshminarayanan\",\"doi\":\"10.1117/12.2528965\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Fundus cameras are the current clinical standard for capturing retinal images, which are used to diagnose a variety of sight-threatening conditions. Traditional fundus cameras are not easily transported, making them unsuitable for field use. In addition, traditional fundus cameras are expensive. Due to this, a variety of technologies have been developed such as the D-EYE Digital Ophthalmoscope (D-EYE Srl, Padova, Italy) which is compatible with various cellphone cameras. This paper reports on the comparison of the image quality of the Nidek RS-330 OCT Retina Scan Duo (Nidek, Tokyo, Japan) and the D-EYE paired with an iPhone 6 (Apple, Cupertino, USA). Twenty-one participants were enrolled in the study of whom 14 underwent nonmydriatic and mydriatic imaging with the D-EYE and the Nidek. Seven participants underwent nonmydriatic imaging with the D-EYE and the Nidek. The images were co-registered and cropped so that the region of interest was equal in both the D-EYE and Nidek images, as the D-EYE had a smaller field of view. Using the Nidek image as the reference, objective full-reference image quality analysis was performed. Metrics such as structural similarity index and peak signal noise ratio were obtained. It was found that the image quality of the D-EYE is limited by the attached iPhone camera, and is lower when compared to the Nidek. Quantification of the differences between the D-EYE and Nidek allows for targeted development of smartphone camera attachments that can help to bridge the current gap in image quality.\",\"PeriodicalId\":10843,\"journal\":{\"name\":\"Current Developments in Lens Design and Optical Engineering XX\",\"volume\":\"41 4 1\",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2019-08-30\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"4\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Current Developments in Lens Design and Optical Engineering XX\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1117/12.2528965\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Current Developments in Lens Design and Optical Engineering XX","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1117/12.2528965","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 4

摘要

眼底相机是目前用于捕捉视网膜图像的临床标准,用于诊断各种视力威胁状况。传统的眼底相机不容易运输,不适合野外使用。此外,传统眼底相机价格昂贵。因此,开发了多种技术,如D-EYE数字检眼镜(D-EYE Srl,帕多瓦,意大利),与各种手机摄像头兼容。本文报道了Nidek RS-330 OCT视网膜扫描Duo (Nidek,东京,日本)和D-EYE与iPhone 6(苹果,库比蒂诺,美国)配对的图像质量的比较。21名参与者参加了这项研究,其中14名接受了D-EYE和Nidek的无肌痛和有肌痛成像。7名参与者接受了D-EYE和Nidek的非散体成像。由于D-EYE具有较小的视野,因此这些图像被共同配准和裁剪,因此D-EYE和Nidek图像中的感兴趣区域是相等的。以Nidek图像为参照物,进行客观的全参照物图像质量分析。得到了结构相似指数、峰值信噪比等指标。结果发现,D-EYE的图像质量受到iPhone相机的限制,比Nidek低。对D-EYE和Nidek之间的差异进行量化,可以有针对性地开发智能手机相机附件,帮助弥合目前在图像质量方面的差距。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Fundus camera versus smartphone camera attachment: image quality analysis
Fundus cameras are the current clinical standard for capturing retinal images, which are used to diagnose a variety of sight-threatening conditions. Traditional fundus cameras are not easily transported, making them unsuitable for field use. In addition, traditional fundus cameras are expensive. Due to this, a variety of technologies have been developed such as the D-EYE Digital Ophthalmoscope (D-EYE Srl, Padova, Italy) which is compatible with various cellphone cameras. This paper reports on the comparison of the image quality of the Nidek RS-330 OCT Retina Scan Duo (Nidek, Tokyo, Japan) and the D-EYE paired with an iPhone 6 (Apple, Cupertino, USA). Twenty-one participants were enrolled in the study of whom 14 underwent nonmydriatic and mydriatic imaging with the D-EYE and the Nidek. Seven participants underwent nonmydriatic imaging with the D-EYE and the Nidek. The images were co-registered and cropped so that the region of interest was equal in both the D-EYE and Nidek images, as the D-EYE had a smaller field of view. Using the Nidek image as the reference, objective full-reference image quality analysis was performed. Metrics such as structural similarity index and peak signal noise ratio were obtained. It was found that the image quality of the D-EYE is limited by the attached iPhone camera, and is lower when compared to the Nidek. Quantification of the differences between the D-EYE and Nidek allows for targeted development of smartphone camera attachments that can help to bridge the current gap in image quality.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
A laser pumping double-light-source module with photon-recycling Optical systems for large-aperture phased laser array including diffractive optics Deployment of combined higher order aberrations to extend the depth of focus of lenses Exposure of Restore-L camera optical elements to a simulated orbital radiation environment Application of GPUs in optical design software
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1