{"title":"po2引导细针肿瘤活检装置的研制。","authors":"Robert C McDonald","doi":"10.1115/1.4052900","DOIUrl":null,"url":null,"abstract":"<p><p>Tumor biopsies are an important aspect of oncology providing a guide for medical treatment and evaluation of disease progression. Highly heterogenous tumors have complex regions of active cancer cells interdigitated with necrotic tissue and healthy noncancerous tissue. The reliable access to tumor tissue pathology is therefore challenging and usually requires multiple needle insertions with accompanying patient discomfort and risk of infection. Oxygen levels provide a means of detecting and evaluating tumor tissue with levels reduced by 2-fold to 22-fold, depending on the type of organ. However, if the biopsy needle is placed in an area of normal tissue, there is always a chance that no diagnostic cells will be acquired for meaningful pathology and molecular analysis. While not the case in all tumors, there are cases where the in vivo oxygen levels differ with tumor cells having a value of pO<sub>2</sub> lying between the anoxic necrotic tissue and normoxic normal tissue. The level of oxygen in tumor cells can also vary with time as related to complex biochemical pathways. The efficacy of radiation therapy is also sensitive to oxygen levels in tumors. Lower levels of oxygen present greater resistance to treatment. To address these concerns, a pO<sub>2</sub>-guided biopsy needle (OGBN) was developed to determine oxygen levels and fluctuations in highly resolved regions of tumors, in order to aide in determining the optimal region for cell sampling help in determining medical treatment options.</p>","PeriodicalId":49305,"journal":{"name":"Journal of Medical Devices-Transactions of the Asme","volume":null,"pages":null},"PeriodicalIF":0.8000,"publicationDate":"2022-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8822461/pdf/med-21-1061_021003.pdf","citationCount":"2","resultStr":"{\"title\":\"Development of a pO<sub>2</sub>-Guided Fine Needle Tumor Biopsy Device.\",\"authors\":\"Robert C McDonald\",\"doi\":\"10.1115/1.4052900\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Tumor biopsies are an important aspect of oncology providing a guide for medical treatment and evaluation of disease progression. Highly heterogenous tumors have complex regions of active cancer cells interdigitated with necrotic tissue and healthy noncancerous tissue. The reliable access to tumor tissue pathology is therefore challenging and usually requires multiple needle insertions with accompanying patient discomfort and risk of infection. Oxygen levels provide a means of detecting and evaluating tumor tissue with levels reduced by 2-fold to 22-fold, depending on the type of organ. However, if the biopsy needle is placed in an area of normal tissue, there is always a chance that no diagnostic cells will be acquired for meaningful pathology and molecular analysis. While not the case in all tumors, there are cases where the in vivo oxygen levels differ with tumor cells having a value of pO<sub>2</sub> lying between the anoxic necrotic tissue and normoxic normal tissue. The level of oxygen in tumor cells can also vary with time as related to complex biochemical pathways. The efficacy of radiation therapy is also sensitive to oxygen levels in tumors. Lower levels of oxygen present greater resistance to treatment. To address these concerns, a pO<sub>2</sub>-guided biopsy needle (OGBN) was developed to determine oxygen levels and fluctuations in highly resolved regions of tumors, in order to aide in determining the optimal region for cell sampling help in determining medical treatment options.</p>\",\"PeriodicalId\":49305,\"journal\":{\"name\":\"Journal of Medical Devices-Transactions of the Asme\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.8000,\"publicationDate\":\"2022-06-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8822461/pdf/med-21-1061_021003.pdf\",\"citationCount\":\"2\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Medical Devices-Transactions of the Asme\",\"FirstCategoryId\":\"5\",\"ListUrlMain\":\"https://doi.org/10.1115/1.4052900\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"ENGINEERING, BIOMEDICAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Medical Devices-Transactions of the Asme","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1115/1.4052900","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 2

摘要

肿瘤活检是肿瘤学的一个重要方面,为医学治疗和疾病进展评估提供指导。高度异质肿瘤具有活跃癌细胞与坏死组织和健康非癌组织交叉的复杂区域。因此,获得肿瘤组织病理的可靠途径具有挑战性,通常需要多次针头插入,同时伴有患者不适和感染风险。氧水平提供了一种检测和评估肿瘤组织的手段,根据器官的类型,氧水平可降低2至22倍。然而,如果活检针放置在正常组织的区域,总是有可能无法获得诊断细胞进行有意义的病理和分子分析。虽然并非所有肿瘤都是如此,但在某些情况下,体内氧水平不同,肿瘤细胞的pO2值介于缺氧坏死组织和常氧正常组织之间。肿瘤细胞中的氧水平也可以随时间变化,与复杂的生化途径有关。放射治疗的效果对肿瘤中的氧含量也很敏感。较低的氧气水平对治疗有更大的抵抗力。为了解决这些问题,开发了一种po2引导活检针(OGBN),以确定肿瘤高度溶解区域的氧水平和波动,以帮助确定细胞取样的最佳区域,从而帮助确定医疗选择。
本文章由计算机程序翻译,如有差异,请以英文原文为准。

摘要图片

摘要图片

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Development of a pO2-Guided Fine Needle Tumor Biopsy Device.

Tumor biopsies are an important aspect of oncology providing a guide for medical treatment and evaluation of disease progression. Highly heterogenous tumors have complex regions of active cancer cells interdigitated with necrotic tissue and healthy noncancerous tissue. The reliable access to tumor tissue pathology is therefore challenging and usually requires multiple needle insertions with accompanying patient discomfort and risk of infection. Oxygen levels provide a means of detecting and evaluating tumor tissue with levels reduced by 2-fold to 22-fold, depending on the type of organ. However, if the biopsy needle is placed in an area of normal tissue, there is always a chance that no diagnostic cells will be acquired for meaningful pathology and molecular analysis. While not the case in all tumors, there are cases where the in vivo oxygen levels differ with tumor cells having a value of pO2 lying between the anoxic necrotic tissue and normoxic normal tissue. The level of oxygen in tumor cells can also vary with time as related to complex biochemical pathways. The efficacy of radiation therapy is also sensitive to oxygen levels in tumors. Lower levels of oxygen present greater resistance to treatment. To address these concerns, a pO2-guided biopsy needle (OGBN) was developed to determine oxygen levels and fluctuations in highly resolved regions of tumors, in order to aide in determining the optimal region for cell sampling help in determining medical treatment options.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
1.80
自引率
11.10%
发文量
56
审稿时长
6-12 weeks
期刊介绍: The Journal of Medical Devices presents papers on medical devices that improve diagnostic, interventional and therapeutic treatments focusing on applied research and the development of new medical devices or instrumentation. It provides special coverage of novel devices that allow new surgical strategies, new methods of drug delivery, or possible reductions in the complexity, cost, or adverse results of health care. The Design Innovation category features papers focusing on novel devices, including papers with limited clinical or engineering results. The Medical Device News section provides coverage of advances, trends, and events.
期刊最新文献
Mechanical Viability and Functionality Assessment of a New Sutureless Endoluminal Microvascular Device: A Preliminary In Vivo Rabbit Study. Design and Implementation of a Computer-Controlled Hybrid Oscillatory Ventilator. Controlled Ice Nucleation With a Sand-PDMS Film Device Enhances Cryopreservation of Mouse Preantral Ovarian Follicles. A Novel Design Method for the Knee Joint of the Exoskeleton Based On the Modular Wearable Sensor Experimental Investigation of the Calcified Plaque Material Removal Rate in Coronary Rotational Atherectomy
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1