{"title":"Lung ultrasound: A comparison of image interpretation accuracy between curvilinear and phased array transducers","authors":"Michael H. Walsh, Leo M. Smyth, Janeve R. Desy, Ernest A. Fischer, Alberto Goffi, Na Li, Matthew Lee, Joëlle St-Pierre, Irene W. Y. Ma","doi":"10.1002/ajum.12347","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Introduction</h3>\n \n <p>Both curvilinear and phased array transducers are commonly used to perform lung ultrasound (LUS). This study seeks to compare LUS interpretation accuracy of images obtained using a curvilinear transducer with those obtained using a phased array transducer.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>We invited 166 internists and trainees to interpret 16 LUS images/cineloops of eight patients in an online survey: eight curvilinear and eight phased array, performed on the same lung location. Images depicted normal lung, pneumothorax, pleural irregularities, consolidation/hepatisation, pleural effusions and B-lines. Primary outcome for each participant is the difference in image interpretation accuracy scores between the two transducers.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>A total of 112 (67%) participants completed the survey. The mean paired accuracy score difference between the curvilinear and phased array images was 3.0% (95% CI: 0.6 to 5.4%, P = 0.015). For novices, scores were higher on curvilinear images (mean difference: 5.4%, 95% CI: 0.9 to 9.9%, P = 0.020). For non-novices, there were no differences between the two transducers (mean difference: 1.4%, 95% CI: −1.1 to 3.9%, P = 0.263). For pleural-based findings, the mean of the paired differences between transducers was higher in the novice group (estimated mean difference-in-differences: 9.5%, 95% CI: 0.6 to 18.4%; P = 0.036). No difference in mean accuracies was noted between novices and non-novices for non-pleural-based pathologies (estimated mean difference-in-differences: 0.6%, 95% CI to 5.4–6.6%; P = 0.837).</p>\n </section>\n \n <section>\n \n <h3> Conclusions</h3>\n \n <p>Lung ultrasound images obtained using the curvilinear transducer are associated with higher interpretation accuracy than the phased array transducer. This is especially true for novices interpreting pleural-based pathologies.</p>\n </section>\n </div>","PeriodicalId":36517,"journal":{"name":"Australasian Journal of Ultrasound in Medicine","volume":"26 3","pages":"150-156"},"PeriodicalIF":0.0000,"publicationDate":"2023-05-27","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1002/ajum.12347","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Australasian Journal of Ultrasound in Medicine","FirstCategoryId":"1085","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/ajum.12347","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 1
Abstract
Introduction
Both curvilinear and phased array transducers are commonly used to perform lung ultrasound (LUS). This study seeks to compare LUS interpretation accuracy of images obtained using a curvilinear transducer with those obtained using a phased array transducer.
Methods
We invited 166 internists and trainees to interpret 16 LUS images/cineloops of eight patients in an online survey: eight curvilinear and eight phased array, performed on the same lung location. Images depicted normal lung, pneumothorax, pleural irregularities, consolidation/hepatisation, pleural effusions and B-lines. Primary outcome for each participant is the difference in image interpretation accuracy scores between the two transducers.
Results
A total of 112 (67%) participants completed the survey. The mean paired accuracy score difference between the curvilinear and phased array images was 3.0% (95% CI: 0.6 to 5.4%, P = 0.015). For novices, scores were higher on curvilinear images (mean difference: 5.4%, 95% CI: 0.9 to 9.9%, P = 0.020). For non-novices, there were no differences between the two transducers (mean difference: 1.4%, 95% CI: −1.1 to 3.9%, P = 0.263). For pleural-based findings, the mean of the paired differences between transducers was higher in the novice group (estimated mean difference-in-differences: 9.5%, 95% CI: 0.6 to 18.4%; P = 0.036). No difference in mean accuracies was noted between novices and non-novices for non-pleural-based pathologies (estimated mean difference-in-differences: 0.6%, 95% CI to 5.4–6.6%; P = 0.837).
Conclusions
Lung ultrasound images obtained using the curvilinear transducer are associated with higher interpretation accuracy than the phased array transducer. This is especially true for novices interpreting pleural-based pathologies.