Mojtaba Behzadipour, Tianjun Ma, Rabten K Datsang, Brandon Lee, Dane Pittock, Elisabeth Weiss, William Y Song
{"title":"Automated daily dose accumulation workflow for treatment quality assurance during online adaptive radiotherapy with a 0.35T MR-linac.","authors":"Mojtaba Behzadipour, Tianjun Ma, Rabten K Datsang, Brandon Lee, Dane Pittock, Elisabeth Weiss, William Y Song","doi":"10.1002/acm2.14594","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>This study assesses a novel, automated dose accumulation process during MR-guided online adaptive radiotherapy (MRgART) for prostate cancer, focusing on inter-fractional anatomical changes and discrepancies between delivered and planned doses.</p><p><strong>Methods: </strong>A retrospective analysis was conducted on seven prostate cancer patients treated with a five-fraction stereotactic body radiation therapy (SBRT), using a 0.35T MRIdian MR-LINAC system. Daily plans were adapted when dose thresholds were exceeded. Planning MRI (pMRI) and daily MRIs (dMRIs) were imported into MIM software for automated and manual dose accumulation procedures. Rigid and deformable image registrations were followed by dose accumulation to compare delivered and planned doses. Manual and automated image registrations were compared by calculating the Hausdorff distance (HD), Jaccard, and DICE metrics.</p><p><strong>Results: </strong>Moderate discrepancies in dosimetric parameters for the planning target volume (PTV) were observed between auto-accumulated and planned doses, such as <math> <semantics><msub><mi>D</mi> <mrow><mn>95</mn> <mo>%</mo></mrow> </msub> <annotation>${D}_{95\\% }$</annotation></semantics> </math> and <math> <semantics><msub><mi>D</mi> <mrow><mn>0.03</mn> <mrow><mspace></mspace> <mi>cc</mi></mrow> </mrow> </msub> <annotation>${D}_{0.03{\\mathrm{\\ cc}}}$</annotation></semantics> </math> , with average differences of <math> <semantics><mrow><mo>-</mo> <mn>0.60</mn> <mo>±</mo> <mn>0.61</mn></mrow> <annotation>$ - 0.60 \\pm 0.61$</annotation></semantics> </math> Gy and <math> <semantics><mrow><mo>-</mo> <mn>1.31</mn> <mo>±</mo> <mn>0.42</mn></mrow> <annotation>$ - 1.31 \\pm 0.42$</annotation></semantics> </math> Gy, respectively. Volume differences of <math> <semantics><msub><mi>V</mi> <mrow><mn>34.4</mn> <mspace></mspace> <mi>Gy</mi></mrow> </msub> <annotation>${V}_{34.4\\ {\\mathrm{Gy}}}$</annotation></semantics> </math> and <math> <semantics><msub><mi>V</mi> <mrow><mn>36.25</mn> <mrow><mspace></mspace> <mi>Gy</mi></mrow> </mrow> </msub> <annotation>${V}_{36.25{\\mathrm{\\ Gy}}}$</annotation></semantics> </math> indicated that auto-accumulated doses consistently had lower numbers compared to planned doses, with mean discrepancies of <math> <semantics><mrow><mo>-</mo> <mn>1.80</mn> <mo>%</mo> <mo>±</mo> <mn>1.05</mn> <mo>%</mo></mrow> <annotation>$ - 1.80\\% \\pm 1.05\\% $</annotation></semantics> </math> and <math> <semantics><mrow><mo>-</mo> <mn>2.82</mn> <mo>%</mo> <mo>±</mo> <mn>1.72</mn> <mo>%</mo></mrow> <annotation>$ - 2.82\\% \\pm 1.72\\% $</annotation></semantics> </math> , respectively. Organs at risk (OAR) dosimetric parameters exhibited higher dose volumes in auto-accumulated doses, with moderate differences (planned [cc] vs. auto-accumulated [cc]) observed in parameters such as urethra PRV <math> <semantics><msub><mi>V</mi> <mrow><mn>8.4</mn> <mrow><mspace></mspace> <mi>Gy</mi></mrow> </mrow> </msub> <annotation>${V}_{8.4{\\mathrm{\\ Gy}}}$</annotation></semantics> </math> at <math> <semantics><mrow><mo>(</mo> <mrow><mn>4.33</mn> <mo>±</mo> <mn>1.90</mn> <mspace></mspace> <mi>vs</mi> <mo>.</mo> <mspace></mspace> <mn>4.34</mn> <mo>±</mo> <mn>1.90</mn></mrow> <mo>)</mo></mrow> <annotation>$( {4.33 \\pm 1.90\\ {\\mathrm{vs}}.\\ 4.34 \\pm 1.90} )$</annotation></semantics> </math> , rectum <math> <semantics><msub><mi>V</mi> <mrow><mn>24</mn> <mspace></mspace> <mi>Gy</mi></mrow> </msub> <annotation>${V}_{24\\ {\\mathrm{Gy}}}$</annotation></semantics> </math> at <math> <semantics><mrow><mo>(</mo> <mrow><mn>1.17</mn> <mo>±</mo> <mn>1.53</mn> <mspace></mspace> <mi>vs</mi> <mo>.</mo> <mspace></mspace> <mn>1.74</mn> <mo>±</mo> <mn>1.91</mn></mrow> <mo>)</mo></mrow> <annotation>$( {1.17 \\pm 1.53\\ {\\mathrm{vs}}.\\ 1.74 \\pm 1.91} )$</annotation></semantics> </math> and rectum <math> <semantics><msub><mi>V</mi> <mrow><mn>28.2</mn> <mrow><mspace></mspace> <mi>Gy</mi></mrow> </mrow> </msub> <annotation>${V}_{28.2{\\mathrm{\\ Gy}}}$</annotation></semantics> </math> at <math> <semantics><mrow><mo>(</mo> <mrow><mn>0.38</mn> <mo>±</mo> <mn>0.55</mn> <mspace></mspace> <mi>vs</mi> <mo>.</mo> <mspace></mspace> <mn>0.59</mn> <mo>±</mo> <mn>0.71</mn></mrow> <mo>)</mo></mrow> <annotation>$( {0.38 \\pm 0.55\\ {\\mathrm{vs}}.\\ 0.59 \\pm 0.71} )$</annotation></semantics> </math> . The comparison between manually and auto-accumulated doses revealed negligible variations, as also indicated by strong concordance in geometric indices and t-test p-values above 0.7.</p><p><strong>Conclusion: </strong>The automated workflow, developed in collaboration with MIM Software Inc., demonstrates high accuracy compared to manual accumulation. The moderate differences observed between planned and accumulated doses emphasize the need for accurate dose accumulation for adaptive plans.</p>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":" ","pages":"e14594"},"PeriodicalIF":2.0000,"publicationDate":"2024-12-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Applied Clinical Medical Physics","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1002/acm2.14594","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: This study assesses a novel, automated dose accumulation process during MR-guided online adaptive radiotherapy (MRgART) for prostate cancer, focusing on inter-fractional anatomical changes and discrepancies between delivered and planned doses.
Methods: A retrospective analysis was conducted on seven prostate cancer patients treated with a five-fraction stereotactic body radiation therapy (SBRT), using a 0.35T MRIdian MR-LINAC system. Daily plans were adapted when dose thresholds were exceeded. Planning MRI (pMRI) and daily MRIs (dMRIs) were imported into MIM software for automated and manual dose accumulation procedures. Rigid and deformable image registrations were followed by dose accumulation to compare delivered and planned doses. Manual and automated image registrations were compared by calculating the Hausdorff distance (HD), Jaccard, and DICE metrics.
Results: Moderate discrepancies in dosimetric parameters for the planning target volume (PTV) were observed between auto-accumulated and planned doses, such as and , with average differences of Gy and Gy, respectively. Volume differences of and indicated that auto-accumulated doses consistently had lower numbers compared to planned doses, with mean discrepancies of and , respectively. Organs at risk (OAR) dosimetric parameters exhibited higher dose volumes in auto-accumulated doses, with moderate differences (planned [cc] vs. auto-accumulated [cc]) observed in parameters such as urethra PRV at , rectum at and rectum at . The comparison between manually and auto-accumulated doses revealed negligible variations, as also indicated by strong concordance in geometric indices and t-test p-values above 0.7.
Conclusion: The automated workflow, developed in collaboration with MIM Software Inc., demonstrates high accuracy compared to manual accumulation. The moderate differences observed between planned and accumulated doses emphasize the need for accurate dose accumulation for adaptive plans.
期刊介绍:
Journal of Applied Clinical Medical Physics is an international Open Access publication dedicated to clinical medical physics. JACMP welcomes original contributions dealing with all aspects of medical physics from scientists working in the clinical medical physics around the world. JACMP accepts only online submission.
JACMP will publish:
-Original Contributions: Peer-reviewed, investigations that represent new and significant contributions to the field. Recommended word count: up to 7500.
-Review Articles: Reviews of major areas or sub-areas in the field of clinical medical physics. These articles may be of any length and are peer reviewed.
-Technical Notes: These should be no longer than 3000 words, including key references.
-Letters to the Editor: Comments on papers published in JACMP or on any other matters of interest to clinical medical physics. These should not be more than 1250 (including the literature) and their publication is only based on the decision of the editor, who occasionally asks experts on the merit of the contents.
-Book Reviews: The editorial office solicits Book Reviews.
-Announcements of Forthcoming Meetings: The Editor may provide notice of forthcoming meetings, course offerings, and other events relevant to clinical medical physics.
-Parallel Opposed Editorial: We welcome topics relevant to clinical practice and medical physics profession. The contents can be controversial debate or opposed aspects of an issue. One author argues for the position and the other against. Each side of the debate contains an opening statement up to 800 words, followed by a rebuttal up to 500 words. Readers interested in participating in this series should contact the moderator with a proposed title and a short description of the topic