Characterization of nonlinear stress relaxation of the femoral and tibial trabecular bone for computational modeling

IF 1.7 4区 医学 Q3 ENGINEERING, BIOMEDICAL Medical Engineering & Physics Pub Date : 2025-03-08 DOI:10.1016/j.medengphy.2025.104324
Thomas Gersie , Thom Bitter , David Wolfson , Robert Freeman , Nico Verdonschot , Dennis Janssen
{"title":"Characterization of nonlinear stress relaxation of the femoral and tibial trabecular bone for computational modeling","authors":"Thomas Gersie ,&nbsp;Thom Bitter ,&nbsp;David Wolfson ,&nbsp;Robert Freeman ,&nbsp;Nico Verdonschot ,&nbsp;Dennis Janssen","doi":"10.1016/j.medengphy.2025.104324","DOIUrl":null,"url":null,"abstract":"<div><div>Computational models of orthopedic reconstructions are reliant on bone material properties, but viscoelastic behavior of trabecular bone is often ignored in numerical simulations. The inclusion of stress relaxation could be of importance for the accuracy of models simulating the primary stability of cementless implants. In this study, a material model to describe the nonlinear viscoelastic behavior of human trabecular bone was constructed based on uniaxial stress relaxation experiments. The relationship of bone mineral density (BMD) and stress relaxation was explored, and the material model was implemented in sample-specific finite element (FE) simulations.</div><div>Cylindrical trabecular human bone specimens, from the distal femur and proximal tibia, were subjected to stress relaxation tests, undergoing compression with strains from 0.2 % to 0.8 % for 30 min on four consecutive days. The experimental data were extrapolated to 24 h. Similar levels of stress relaxation were found for femoral and tibial specimens, with an average 54.4 % stress relaxation and a maximum level of 81.6 %. Using a modified superposition model, the specimen-specific nonlinear stress relaxation behavior was captured. However, when the samples were considered collectively, no correlation was found between applied strain, BMD and the viscoelastic response. Therefore, the average level of stress relaxation in combination with existing BMD-stiffness relationships were implemented in FE simulations for each individual specimen. While the FE models, on average, overestimated the overall stiffness by 64 %, they were able to adequately capture the stress relaxation response.</div></div>","PeriodicalId":49836,"journal":{"name":"Medical Engineering & Physics","volume":"138 ","pages":"Article 104324"},"PeriodicalIF":1.7000,"publicationDate":"2025-03-08","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Medical Engineering & Physics","FirstCategoryId":"5","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1350453325000438","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0

Abstract

Computational models of orthopedic reconstructions are reliant on bone material properties, but viscoelastic behavior of trabecular bone is often ignored in numerical simulations. The inclusion of stress relaxation could be of importance for the accuracy of models simulating the primary stability of cementless implants. In this study, a material model to describe the nonlinear viscoelastic behavior of human trabecular bone was constructed based on uniaxial stress relaxation experiments. The relationship of bone mineral density (BMD) and stress relaxation was explored, and the material model was implemented in sample-specific finite element (FE) simulations.
Cylindrical trabecular human bone specimens, from the distal femur and proximal tibia, were subjected to stress relaxation tests, undergoing compression with strains from 0.2 % to 0.8 % for 30 min on four consecutive days. The experimental data were extrapolated to 24 h. Similar levels of stress relaxation were found for femoral and tibial specimens, with an average 54.4 % stress relaxation and a maximum level of 81.6 %. Using a modified superposition model, the specimen-specific nonlinear stress relaxation behavior was captured. However, when the samples were considered collectively, no correlation was found between applied strain, BMD and the viscoelastic response. Therefore, the average level of stress relaxation in combination with existing BMD-stiffness relationships were implemented in FE simulations for each individual specimen. While the FE models, on average, overestimated the overall stiffness by 64 %, they were able to adequately capture the stress relaxation response.

Abstract Image

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
求助全文
约1分钟内获得全文 去求助
来源期刊
Medical Engineering & Physics
Medical Engineering & Physics 工程技术-工程:生物医学
CiteScore
4.30
自引率
4.50%
发文量
172
审稿时长
3.0 months
期刊介绍: Medical Engineering & Physics provides a forum for the publication of the latest developments in biomedical engineering, and reflects the essential multidisciplinary nature of the subject. The journal publishes in-depth critical reviews, scientific papers and technical notes. Our focus encompasses the application of the basic principles of physics and engineering to the development of medical devices and technology, with the ultimate aim of producing improvements in the quality of health care.Topics covered include biomechanics, biomaterials, mechanobiology, rehabilitation engineering, biomedical signal processing and medical device development. Medical Engineering & Physics aims to keep both engineers and clinicians abreast of the latest applications of technology to health care.
期刊最新文献
Measurement of heart rate from long-distance videos via projection of rotated orthogonal bases in POS Surgical technique and implant design affect abduction kinematics and functional outcomes after reverse shoulder arthroplasty The effect of bone plasticity models on simulations of primary fixation in total knee arthroplasty Blinded prediction of custom-made pelvic implant failure using patient-specific finite element modeling Characterization of nonlinear stress relaxation of the femoral and tibial trabecular bone for computational modeling
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1