{"title":"[The preparation procedure of TiAdapt abutment effect on the implant to abutment joint stability].","authors":"Feimin Zhang, Yunfeng Zhao, Huarong Wang, Michael E Razzoog, Brien R Lang","doi":"","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>This study was designed to determine the difference in the implant to abutment joint stability resulting from the preparation procedure in adjusting the TiAdapt abutment.</p><p><strong>Methods: </strong>Thirty Branemark implants (10RP, 10NP, 10WP) and 10 TiAdapt RP, TiAdapt NP and TiAdapt WP abutments formed the experimental populations of 3 groups. Samples in each groups were further divided into subgroups A and B. Subgroup A was operated with preparation procedure whereas the other served as control. The torque required and the gap between abutment and implant before and after preparation were measured.</p><p><strong>Results: </strong>There were no significant differences in group RP and group WP; there were significant differences in group NP at the significance level of 0.05.</p><p><strong>Conclusion: </strong>These results indicate that the implant to abutment joint stability has been strengthened by the abutment preparation procedure.</p>","PeriodicalId":13173,"journal":{"name":"Hua xi yi ke da xue xue bao = Journal of West China University of Medical Sciences = Huaxi yike daxue xuebao","volume":"33 2","pages":"247-9"},"PeriodicalIF":0.0000,"publicationDate":"2002-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Hua xi yi ke da xue xue bao = Journal of West China University of Medical Sciences = Huaxi yike daxue xuebao","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
Objective: This study was designed to determine the difference in the implant to abutment joint stability resulting from the preparation procedure in adjusting the TiAdapt abutment.
Methods: Thirty Branemark implants (10RP, 10NP, 10WP) and 10 TiAdapt RP, TiAdapt NP and TiAdapt WP abutments formed the experimental populations of 3 groups. Samples in each groups were further divided into subgroups A and B. Subgroup A was operated with preparation procedure whereas the other served as control. The torque required and the gap between abutment and implant before and after preparation were measured.
Results: There were no significant differences in group RP and group WP; there were significant differences in group NP at the significance level of 0.05.
Conclusion: These results indicate that the implant to abutment joint stability has been strengthened by the abutment preparation procedure.