金超微电极阵列溶出伏安法检测痕量重金属

Manuel Gutiérrez-Capitán, A. Baldi, C. Jiménez-Jorquera, C. Fernández-Sánchez, Raquel Gomez, V. Garcia
{"title":"金超微电极阵列溶出伏安法检测痕量重金属","authors":"Manuel Gutiérrez-Capitán, A. Baldi, C. Jiménez-Jorquera, C. Fernández-Sánchez, Raquel Gomez, V. Garcia","doi":"10.1109/IBERSENSOR.2014.6995557","DOIUrl":null,"url":null,"abstract":"The sensitive and simultaneous detection of zinc (Zn), cadmium (Cd) and lead (Pb) at gold ultramicroelectrode arrays (UMEAs) is reported. The device performance was based on the square-wave anodic stripping voltammetric technique. This consisted of the co-deposition of a layer of bismuth (Bi) together with the heavy metal target analytes onto the UMEA surface by applying a constant negative potential of -1.3 V for 360 s, followed by the stripping of the heavy metal alloy deposited by an anodic potential scan in a suitable potential window. Peaks at potentials ca. -1.0, -0.75 and -0.5 V were recorded and could be ascribed to the Zn, Cd and Pb stripping processes, respectively. Measurements were carried out in standard 0.1 M acetate buffer solutions pH 4.5, containing increasing concentrations of the three analytes. By using the peak areas, the corresponding calibration curves were constructed. The results showed a good sensitivity with a linear range from 10 to 70 ppb and a limit of detection of 7.5 ppb Pb, 5.3 ppb Cd and 9.2 ppb Zn, respectively.","PeriodicalId":296271,"journal":{"name":"2014 IEEE 9th IberoAmerican Congress on Sensors","volume":"128 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2014-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":"{\"title\":\"Stripping voltammetric detection of trace heavy metals using gold ultramicroelectrode arrays\",\"authors\":\"Manuel Gutiérrez-Capitán, A. Baldi, C. Jiménez-Jorquera, C. Fernández-Sánchez, Raquel Gomez, V. Garcia\",\"doi\":\"10.1109/IBERSENSOR.2014.6995557\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"The sensitive and simultaneous detection of zinc (Zn), cadmium (Cd) and lead (Pb) at gold ultramicroelectrode arrays (UMEAs) is reported. The device performance was based on the square-wave anodic stripping voltammetric technique. This consisted of the co-deposition of a layer of bismuth (Bi) together with the heavy metal target analytes onto the UMEA surface by applying a constant negative potential of -1.3 V for 360 s, followed by the stripping of the heavy metal alloy deposited by an anodic potential scan in a suitable potential window. Peaks at potentials ca. -1.0, -0.75 and -0.5 V were recorded and could be ascribed to the Zn, Cd and Pb stripping processes, respectively. Measurements were carried out in standard 0.1 M acetate buffer solutions pH 4.5, containing increasing concentrations of the three analytes. By using the peak areas, the corresponding calibration curves were constructed. The results showed a good sensitivity with a linear range from 10 to 70 ppb and a limit of detection of 7.5 ppb Pb, 5.3 ppb Cd and 9.2 ppb Zn, respectively.\",\"PeriodicalId\":296271,\"journal\":{\"name\":\"2014 IEEE 9th IberoAmerican Congress on Sensors\",\"volume\":\"128 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2014-10-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"1\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"2014 IEEE 9th IberoAmerican Congress on Sensors\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1109/IBERSENSOR.2014.6995557\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"2014 IEEE 9th IberoAmerican Congress on Sensors","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/IBERSENSOR.2014.6995557","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1

摘要

报道了金超微电极阵列(UMEAs)对锌(Zn)、镉(Cd)和铅(Pb)的灵敏同时检测。器件性能基于方波阳极溶出伏安技术。这包括通过施加-1.3 V的恒定负电位360 s,在UMEA表面上沉积一层铋(Bi)和重金属目标分析物,然后在合适的电位窗口内通过阳极电位扫描剥离沉积的重金属合金。在-1.0 V、-0.75 V和-0.5 V电位处记录到峰,分别归因于Zn、Cd和Pb剥离过程。测量在标准的0.1 M醋酸缓冲溶液pH 4.5中进行,其中含有三种分析物的浓度增加。利用峰面积,构建相应的标定曲线。结果表明,该方法具有良好的灵敏度,线性范围为10 ~ 70 ppb,检出限分别为7.5 ppb Pb、5.3 ppb Cd和9.2 ppb Zn。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Stripping voltammetric detection of trace heavy metals using gold ultramicroelectrode arrays
The sensitive and simultaneous detection of zinc (Zn), cadmium (Cd) and lead (Pb) at gold ultramicroelectrode arrays (UMEAs) is reported. The device performance was based on the square-wave anodic stripping voltammetric technique. This consisted of the co-deposition of a layer of bismuth (Bi) together with the heavy metal target analytes onto the UMEA surface by applying a constant negative potential of -1.3 V for 360 s, followed by the stripping of the heavy metal alloy deposited by an anodic potential scan in a suitable potential window. Peaks at potentials ca. -1.0, -0.75 and -0.5 V were recorded and could be ascribed to the Zn, Cd and Pb stripping processes, respectively. Measurements were carried out in standard 0.1 M acetate buffer solutions pH 4.5, containing increasing concentrations of the three analytes. By using the peak areas, the corresponding calibration curves were constructed. The results showed a good sensitivity with a linear range from 10 to 70 ppb and a limit of detection of 7.5 ppb Pb, 5.3 ppb Cd and 9.2 ppb Zn, respectively.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Design and implementation of a planar capacitive pressure sensor Flow-injection amperometric sensor for quantification and speciation of iron Development of a low-cost packaging for MEMS pressure sensors Detection of pathogenic bacteria using immunofiltration and lightscribe technology TEOS plasma polymerized films: Impact of clustered material in humidity measurement
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1