Bryan Zamorano, Yolla Sakr, Frédérique Galopin-Dubois
{"title":"[评估用于血红蛋白床旁检测的新型便携式分析仪 HemoCue® Hb801]。","authors":"Bryan Zamorano, Yolla Sakr, Frédérique Galopin-Dubois","doi":"10.1684/abc.2024.1868","DOIUrl":null,"url":null,"abstract":"<p><p>The use of portable hemoglobin measuring devices is widespread. In this context, the company HemoCue® has put on the market a new device, the Hb801. It uses a whole blood absorbance measurement method and not the azidmethemoglobin measurement method used by HemoCue's older devices. We evaluated this new equipment on EDTA venous blood. Hb801 is lightweight, compact, requires a volume of 10 μL of blood and renders its result in less than a second. The repeatability and intermediate precision are close to the values expected according to Ricos, with coefficients of variation respectively for a low level of hemoglobin: 2.1% and 1.9%, for an average level: 0.8% and 1.5% and for a high level: 1.5% and 1.3%. Comparison to our laboratory reference method (XN-10 Sysmex®) and HemoCue® Hb201+ was performed on 96 samples. Bias (SD) found were: XN-10: +0.42 g/dL (0.17), HemoCue® Hb201+: +0.17 g/dL (0.41). Clinically acceptable performance (within ± 1 g/dL of reference hemoglobin) was high: 93.8%. In the end, this device seems to us to be suitable for hemoglobin point-of-care testing.</p>","PeriodicalId":93870,"journal":{"name":"Annales de biologie clinique","volume":"82 1","pages":"103-111"},"PeriodicalIF":0.0000,"publicationDate":"2024-04-19","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"[Evaluation of a new portable analyzer, HemoCue<sup>®</sup> Hb801, for hemoglobin point-of-care testing].\",\"authors\":\"Bryan Zamorano, Yolla Sakr, Frédérique Galopin-Dubois\",\"doi\":\"10.1684/abc.2024.1868\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>The use of portable hemoglobin measuring devices is widespread. In this context, the company HemoCue® has put on the market a new device, the Hb801. It uses a whole blood absorbance measurement method and not the azidmethemoglobin measurement method used by HemoCue's older devices. We evaluated this new equipment on EDTA venous blood. Hb801 is lightweight, compact, requires a volume of 10 μL of blood and renders its result in less than a second. The repeatability and intermediate precision are close to the values expected according to Ricos, with coefficients of variation respectively for a low level of hemoglobin: 2.1% and 1.9%, for an average level: 0.8% and 1.5% and for a high level: 1.5% and 1.3%. Comparison to our laboratory reference method (XN-10 Sysmex®) and HemoCue® Hb201+ was performed on 96 samples. Bias (SD) found were: XN-10: +0.42 g/dL (0.17), HemoCue® Hb201+: +0.17 g/dL (0.41). Clinically acceptable performance (within ± 1 g/dL of reference hemoglobin) was high: 93.8%. In the end, this device seems to us to be suitable for hemoglobin point-of-care testing.</p>\",\"PeriodicalId\":93870,\"journal\":{\"name\":\"Annales de biologie clinique\",\"volume\":\"82 1\",\"pages\":\"103-111\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2024-04-19\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Annales de biologie clinique\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1684/abc.2024.1868\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annales de biologie clinique","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1684/abc.2024.1868","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
[Evaluation of a new portable analyzer, HemoCue® Hb801, for hemoglobin point-of-care testing].
The use of portable hemoglobin measuring devices is widespread. In this context, the company HemoCue® has put on the market a new device, the Hb801. It uses a whole blood absorbance measurement method and not the azidmethemoglobin measurement method used by HemoCue's older devices. We evaluated this new equipment on EDTA venous blood. Hb801 is lightweight, compact, requires a volume of 10 μL of blood and renders its result in less than a second. The repeatability and intermediate precision are close to the values expected according to Ricos, with coefficients of variation respectively for a low level of hemoglobin: 2.1% and 1.9%, for an average level: 0.8% and 1.5% and for a high level: 1.5% and 1.3%. Comparison to our laboratory reference method (XN-10 Sysmex®) and HemoCue® Hb201+ was performed on 96 samples. Bias (SD) found were: XN-10: +0.42 g/dL (0.17), HemoCue® Hb201+: +0.17 g/dL (0.41). Clinically acceptable performance (within ± 1 g/dL of reference hemoglobin) was high: 93.8%. In the end, this device seems to us to be suitable for hemoglobin point-of-care testing.