Comparative analysis of fetal dose sparing between a C-arm linac and an O-ring linac in a SIB-VMAT sarcoma treatment for a pregnant patient: A technical note and case report.
Wesley Rivais, Louis Constine, Matthew Pacella, Neil Joyce, Maimuna Nagey, Matthew Webster, Jihyung Yoon, Hyunuk Jung, Sean Tanny, Olga Maria Dona Lemus, Dandan Zheng
{"title":"Comparative analysis of fetal dose sparing between a C-arm linac and an O-ring linac in a SIB-VMAT sarcoma treatment for a pregnant patient: A technical note and case report.","authors":"Wesley Rivais, Louis Constine, Matthew Pacella, Neil Joyce, Maimuna Nagey, Matthew Webster, Jihyung Yoon, Hyunuk Jung, Sean Tanny, Olga Maria Dona Lemus, Dandan Zheng","doi":"10.1002/acm2.14556","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To compare the effect of two linacs designs on fetal dose sparing on a pregnant patient, including estimation of the fetal dose, and the effect of a lead apron.</p><p><strong>Methods: </strong>A patient with a high-grade sarcoma located in the right knee/lower thigh was prescribed 51 Gy (1.7 Gy/Fx) with a simultaneous-integrated-boost (SIB) of 60 Gy to a smaller volume, starting in the 26th gestational week. Volumetric modulated radiation therapy (VMAT) plans with 6MV-FFF were developed using identical dosimetric constraints on a Varian Truebeam Edge with HD-MLC and a Varian Halcyon with double-stacked MLC. Based on patient dimension measurements, an anthropomorphic phantom was constructed using a Rando phantom and saline bags in the patient's Vac-Lok bag. Phantom measurements were performed using OSLDs and TLDs placed at three different planes, corresponding to the pubis, the umbilicus, and the fundus based on patient measurements and projected gestational age, to estimate the fetal dose. Three experimental scenarios were measured, each with CBCT-based image guidance for an accurate, reproducible setup: Edge, Halcyon, and Halcyon with a tri-folded lead apron (0.5 mm × 3 = 1.5 mm Pb) over the phantom abdomen.</p><p><strong>Results: </strong>Plan quality and total MUs are comparable between the Edge and Halcyon plans. The OSLD-measured whole-course dose to the pubis, umbilicus, and fundus were 18.8, 13.1, and 11.7 cGy, respectively, on Halcyon, on average 27.8% lower than Edge. The repeatability within either dosimeter was good, although TLD showed systematically lower doses. Importantly, both dosimetry systems showed a lower measured fetal dose for the Halcyon plan compared with the Edge plan. Adding a tri-folded lead apron over the abdomen did not meaningfully lower the measured dose.</p><p><strong>Conclusion: </strong>In this case study, Halcyon demonstrated a better sparing of out-of-field fetal dose compared to TrueBeam Edge. It was shown that lead aprons do not provide additional fetal dose sparing.</p>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":" ","pages":"e14556"},"PeriodicalIF":2.0000,"publicationDate":"2024-11-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Applied Clinical Medical Physics","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1002/acm2.14556","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: To compare the effect of two linacs designs on fetal dose sparing on a pregnant patient, including estimation of the fetal dose, and the effect of a lead apron.
Methods: A patient with a high-grade sarcoma located in the right knee/lower thigh was prescribed 51 Gy (1.7 Gy/Fx) with a simultaneous-integrated-boost (SIB) of 60 Gy to a smaller volume, starting in the 26th gestational week. Volumetric modulated radiation therapy (VMAT) plans with 6MV-FFF were developed using identical dosimetric constraints on a Varian Truebeam Edge with HD-MLC and a Varian Halcyon with double-stacked MLC. Based on patient dimension measurements, an anthropomorphic phantom was constructed using a Rando phantom and saline bags in the patient's Vac-Lok bag. Phantom measurements were performed using OSLDs and TLDs placed at three different planes, corresponding to the pubis, the umbilicus, and the fundus based on patient measurements and projected gestational age, to estimate the fetal dose. Three experimental scenarios were measured, each with CBCT-based image guidance for an accurate, reproducible setup: Edge, Halcyon, and Halcyon with a tri-folded lead apron (0.5 mm × 3 = 1.5 mm Pb) over the phantom abdomen.
Results: Plan quality and total MUs are comparable between the Edge and Halcyon plans. The OSLD-measured whole-course dose to the pubis, umbilicus, and fundus were 18.8, 13.1, and 11.7 cGy, respectively, on Halcyon, on average 27.8% lower than Edge. The repeatability within either dosimeter was good, although TLD showed systematically lower doses. Importantly, both dosimetry systems showed a lower measured fetal dose for the Halcyon plan compared with the Edge plan. Adding a tri-folded lead apron over the abdomen did not meaningfully lower the measured dose.
Conclusion: In this case study, Halcyon demonstrated a better sparing of out-of-field fetal dose compared to TrueBeam Edge. It was shown that lead aprons do not provide additional fetal dose sparing.
期刊介绍:
Journal of Applied Clinical Medical Physics is an international Open Access publication dedicated to clinical medical physics. JACMP welcomes original contributions dealing with all aspects of medical physics from scientists working in the clinical medical physics around the world. JACMP accepts only online submission.
JACMP will publish:
-Original Contributions: Peer-reviewed, investigations that represent new and significant contributions to the field. Recommended word count: up to 7500.
-Review Articles: Reviews of major areas or sub-areas in the field of clinical medical physics. These articles may be of any length and are peer reviewed.
-Technical Notes: These should be no longer than 3000 words, including key references.
-Letters to the Editor: Comments on papers published in JACMP or on any other matters of interest to clinical medical physics. These should not be more than 1250 (including the literature) and their publication is only based on the decision of the editor, who occasionally asks experts on the merit of the contents.
-Book Reviews: The editorial office solicits Book Reviews.
-Announcements of Forthcoming Meetings: The Editor may provide notice of forthcoming meetings, course offerings, and other events relevant to clinical medical physics.
-Parallel Opposed Editorial: We welcome topics relevant to clinical practice and medical physics profession. The contents can be controversial debate or opposed aspects of an issue. One author argues for the position and the other against. Each side of the debate contains an opening statement up to 800 words, followed by a rebuttal up to 500 words. Readers interested in participating in this series should contact the moderator with a proposed title and a short description of the topic