Broken Adjustable Moore Pins Made From Cobalt-Chromium Alloy 由钴铬合金制成的破碎可调摩尔引脚
Pub Date : 2019-06-01 DOI: 10.31399/asm.fach.med.c0048399
Two of four adjustable Moore pins, which had been used to stabilize a proximal femur fracture, were found to be broken and deformed at their threads. The pins were made from a cobalt-chromium alloy and were not in the same condition. Brittle precipitates in the grains and grain boundaries were seen in one of the pins and hence the fracture was revealed to have occurred along the grain boundaries. The other pin made from cold-worked cobalt-chromium alloy was observed to have randomly lines of primary inclusions. Intermingled dimples and fatigue striations were exhibited on the fracture surface of this pin. Thus, the effect of different conditions of cobalt-chromium alloys on failure behavior was demonstrated as a result of this study.
用于稳定股骨近端骨折的四个可调节摩尔钉中的两个被发现在其螺纹处断裂和变形。这些别针是由钴铬合金制成的,情况不同。在其中一个销中发现了晶粒和晶界中的脆性沉淀,因此断裂是沿着晶界发生的。另一种由冷加工钴铬合金制成的销被观察到有随机的初级夹杂物线。断口表面出现了混杂的韧窝和疲劳条纹。因此,本研究的结果证明了不同条件对钴铬合金失效行为的影响。
{"title":"Broken Adjustable Moore Pins Made From Cobalt-Chromium Alloy","authors":"","doi":"10.31399/asm.fach.med.c0048399","DOIUrl":"https://doi.org/10.31399/asm.fach.med.c0048399","url":null,"abstract":"\u0000 Two of four adjustable Moore pins, which had been used to stabilize a proximal femur fracture, were found to be broken and deformed at their threads. The pins were made from a cobalt-chromium alloy and were not in the same condition. Brittle precipitates in the grains and grain boundaries were seen in one of the pins and hence the fracture was revealed to have occurred along the grain boundaries. The other pin made from cold-worked cobalt-chromium alloy was observed to have randomly lines of primary inclusions. Intermingled dimples and fatigue striations were exhibited on the fracture surface of this pin. Thus, the effect of different conditions of cobalt-chromium alloys on failure behavior was demonstrated as a result of this study.","PeriodicalId":125471,"journal":{"name":"ASM Failure Analysis Case Histories: Medical and Biomedical Devices","volume":"216 1","pages":"0"},"PeriodicalIF":0.0,"publicationDate":"2019-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"114593837","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Titanium Screw Head With Fretting Structure at Contact Area With Plate Hole 钛螺丝头与板孔接触处有微动结构
Pub Date : 2019-06-01 DOI: 10.31399/asm.fach.med.c0048421
Wear on a titanium screw head with a lip of material that that was transported by fretting at a plate-hole edge was studied. A flat fretting zone was visible on the screw surface over the material lip. A cellular wear structure containing wear debris was found. No morphological signs of corrosion were observed in connection with fretting structures.
研究了在板孔边缘通过微动输送材料的钛螺钉头的磨损。在料唇上方的螺杆表面可见平坦的微动区。发现了含有磨损碎片的细胞状磨损结构。在微动结构中没有观察到腐蚀的形态学迹象。
{"title":"Titanium Screw Head With Fretting Structure at Contact Area With Plate Hole","authors":"","doi":"10.31399/asm.fach.med.c0048421","DOIUrl":"https://doi.org/10.31399/asm.fach.med.c0048421","url":null,"abstract":"\u0000 Wear on a titanium screw head with a lip of material that that was transported by fretting at a plate-hole edge was studied. A flat fretting zone was visible on the screw surface over the material lip. A cellular wear structure containing wear debris was found. No morphological signs of corrosion were observed in connection with fretting structures.","PeriodicalId":125471,"journal":{"name":"ASM Failure Analysis Case Histories: Medical and Biomedical Devices","volume":"17 1","pages":"0"},"PeriodicalIF":0.0,"publicationDate":"2019-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"123996564","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Severe General Corrosion on Historic Lane Plate Made From Chromium Steel 铬钢历史巷板普遍腐蚀严重
Pub Date : 2019-06-01 DOI: 10.31399/asm.fach.med.c0048397
A large portion of the four-hole Lane plate disintegrated and consisted mainly of corrosion products after remaining in the body for 26 years. Transformation structures and carbides were exhibited by the plate which was made from chromium steel. Minimal corrosion was exhibited by the soft austenitic 304 stainless steel used to make the screws. The corrosion products of the plate were revealed by microprobe analysis to impregnate the surrounding tissues. Improper material selection was concluded to be the reason for the general corrosion behavior.
四孔车道板在体内放置26年后,大部分已经解体,主要由腐蚀产物组成。由铬钢制成的板材显示出相变组织和碳化物。软奥氏体304不锈钢用于制造螺丝,腐蚀最小。通过微探针分析发现,腐蚀产物浸渍了周围组织。材料选择不当是造成一般腐蚀行为的原因。
{"title":"Severe General Corrosion on Historic Lane Plate Made From Chromium Steel","authors":"","doi":"10.31399/asm.fach.med.c0048397","DOIUrl":"https://doi.org/10.31399/asm.fach.med.c0048397","url":null,"abstract":"\u0000 A large portion of the four-hole Lane plate disintegrated and consisted mainly of corrosion products after remaining in the body for 26 years. Transformation structures and carbides were exhibited by the plate which was made from chromium steel. Minimal corrosion was exhibited by the soft austenitic 304 stainless steel used to make the screws. The corrosion products of the plate were revealed by microprobe analysis to impregnate the surrounding tissues. Improper material selection was concluded to be the reason for the general corrosion behavior.","PeriodicalId":125471,"journal":{"name":"ASM Failure Analysis Case Histories: Medical and Biomedical Devices","volume":"187 1","pages":"0"},"PeriodicalIF":0.0,"publicationDate":"2019-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"121209392","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Fatigue Fracture of a Type 316L Stainless Steel Angled Plate 316L型不锈钢斜板的疲劳断裂
Pub Date : 2019-06-01 DOI: 10.31399/asm.fach.med.c0048413
A type 316L stainless steel angled plate failed. The fatigue fracture was found to have occurred at a plate hole. Symmetric cyclic bending forces were revealed by the fatigue damage at the fracture edge at the top surface of the plate. Fatigue striations and slip bands produced on the surface during cyclic loading were observed. The material was showed by the deformation structure to be in the cold-worked condition and was termed to not be the cause of the implant failure.
316L型不锈钢斜板失效。发现疲劳断裂发生在板孔处。板顶表面断口处的疲劳损伤表现出对称的循环弯曲力。在循环加载过程中,试样表面产生了疲劳条纹和滑移带。材料的变形结构表明材料处于冷加工状态,并不是导致植入体失效的原因。
{"title":"Fatigue Fracture of a Type 316L Stainless Steel Angled Plate","authors":"","doi":"10.31399/asm.fach.med.c0048413","DOIUrl":"https://doi.org/10.31399/asm.fach.med.c0048413","url":null,"abstract":"\u0000 A type 316L stainless steel angled plate failed. The fatigue fracture was found to have occurred at a plate hole. Symmetric cyclic bending forces were revealed by the fatigue damage at the fracture edge at the top surface of the plate. Fatigue striations and slip bands produced on the surface during cyclic loading were observed. The material was showed by the deformation structure to be in the cold-worked condition and was termed to not be the cause of the implant failure.","PeriodicalId":125471,"journal":{"name":"ASM Failure Analysis Case Histories: Medical and Biomedical Devices","volume":"145 1","pages":"0"},"PeriodicalIF":0.0,"publicationDate":"2019-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"128177306","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Shearing Fracture of a Type 316LR Stainless Steel Screw 316LR型不锈钢螺钉的剪切断裂
Pub Date : 2019-06-01 DOI: 10.31399/asm.fach.med.c0048403
During the internal fixation, the type 316LR stainless steel cortical bone screw failed. Extensive spiral deformation was revealed by the fracture surface. Dimple structure characteristic of a ductile failure mode was observed with dimples oriented uniformly in the deformation direction. A zone of heavily deformed grains at the fracture edge was revealed by longitudinal metallographic examination. The shearing fractures of a commercially pure titanium screw and a cast cobalt-chromium-molybdenum alloy were discussed for purpose of comparison.
内固定过程中,316LR型不锈钢皮质骨螺钉失效。断口处出现大面积的螺旋形变形。观察到具有延性破坏模式特征的韧窝结构,韧窝在变形方向上取向均匀。纵向金相检查发现,断口边缘有严重变形的晶粒带。对工业纯钛螺杆和铸造钴铬钼合金的剪切断口进行了比较。
{"title":"Shearing Fracture of a Type 316LR Stainless Steel Screw","authors":"","doi":"10.31399/asm.fach.med.c0048403","DOIUrl":"https://doi.org/10.31399/asm.fach.med.c0048403","url":null,"abstract":"\u0000 During the internal fixation, the type 316LR stainless steel cortical bone screw failed. Extensive spiral deformation was revealed by the fracture surface. Dimple structure characteristic of a ductile failure mode was observed with dimples oriented uniformly in the deformation direction. A zone of heavily deformed grains at the fracture edge was revealed by longitudinal metallographic examination. The shearing fractures of a commercially pure titanium screw and a cast cobalt-chromium-molybdenum alloy were discussed for purpose of comparison.","PeriodicalId":125471,"journal":{"name":"ASM Failure Analysis Case Histories: Medical and Biomedical Devices","volume":"2 1","pages":"0"},"PeriodicalIF":0.0,"publicationDate":"2019-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"121378379","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Characteristic Observations on Type 316LR Stainless Steel Screws That Failed by Fatigue 316LR型不锈钢螺钉疲劳失效特性观察
Pub Date : 2019-06-01 DOI: 10.31399/asm.fach.med.c0048405
Type 316LR stainless steel screws that failed by fatigue were studied. It was found that fatigue fracture can occur on different thread levels, depending on the loading situation. The initiation of secondary fatigue cracks was occasionally found parallel to the fracture plane. The screws were used with a relatively rigid plate to treat a fracture complication in the upper end of the femur. The fatigue failures were explained by signs of unstable fixation revealed by radiographs.
对316LR型不锈钢螺钉疲劳失效进行了研究。结果表明,不同载荷情况下,不同螺纹水平均可发生疲劳断裂。偶有平行于断裂面产生二次疲劳裂纹的现象。螺钉与相对刚性钢板一起用于治疗股骨上端骨折并发症。x线片显示的不稳定固定迹象解释了疲劳失效。
{"title":"Characteristic Observations on Type 316LR Stainless Steel Screws That Failed by Fatigue","authors":"","doi":"10.31399/asm.fach.med.c0048405","DOIUrl":"https://doi.org/10.31399/asm.fach.med.c0048405","url":null,"abstract":"\u0000 Type 316LR stainless steel screws that failed by fatigue were studied. It was found that fatigue fracture can occur on different thread levels, depending on the loading situation. The initiation of secondary fatigue cracks was occasionally found parallel to the fracture plane. The screws were used with a relatively rigid plate to treat a fracture complication in the upper end of the femur. The fatigue failures were explained by signs of unstable fixation revealed by radiographs.","PeriodicalId":125471,"journal":{"name":"ASM Failure Analysis Case Histories: Medical and Biomedical Devices","volume":"30 1","pages":"0"},"PeriodicalIF":0.0,"publicationDate":"2019-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"134318291","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Screw Hole With Fretting and Fretting Corrosion of a Type 316LR Stainless Steel Plate 316LR型不锈钢板的微动和微动腐蚀螺钉孔
Pub Date : 2019-06-01 DOI: 10.31399/asm.fach.med.c0048420
Fretting and fretting corrosion at the contact area between the screw hole of a type 316LR stainless steel bone plate and the corresponding screw head was studied. The attack on the 316LR stainless steel was only shallow. Mechanical grinding and polishing structures were exhibited by a large portion of the contact area. Fine corrosion pits in the periphery were observed and intense mechanical material transfer that can take place during fretting was revealed. Smearing of material layers over each other during wear was observed and attack by pitting corrosion was interpreted to be possible.
研究了316LR型不锈钢骨板螺钉孔与螺钉头接触区域的微动及微动腐蚀。对316LR不锈钢的攻击很浅。机械磨削和抛光结构表现为很大一部分的接触面积。在外围观察到细小的腐蚀坑,并揭示了微动过程中可能发生的强烈的机械材料转移。在磨损过程中观察到材料层之间的涂抹,并且解释了点蚀的攻击是可能的。
{"title":"Screw Hole With Fretting and Fretting Corrosion of a Type 316LR Stainless Steel Plate","authors":"","doi":"10.31399/asm.fach.med.c0048420","DOIUrl":"https://doi.org/10.31399/asm.fach.med.c0048420","url":null,"abstract":"\u0000 Fretting and fretting corrosion at the contact area between the screw hole of a type 316LR stainless steel bone plate and the corresponding screw head was studied. The attack on the 316LR stainless steel was only shallow. Mechanical grinding and polishing structures were exhibited by a large portion of the contact area. Fine corrosion pits in the periphery were observed and intense mechanical material transfer that can take place during fretting was revealed. Smearing of material layers over each other during wear was observed and attack by pitting corrosion was interpreted to be possible.","PeriodicalId":125471,"journal":{"name":"ASM Failure Analysis Case Histories: Medical and Biomedical Devices","volume":"2 1","pages":"0"},"PeriodicalIF":0.0,"publicationDate":"2019-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"117083801","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Prongs on Forceps That Split and Fractured Due to Forging Seams 钳子上的尖头由于锻造接缝而裂开和断裂
Pub Date : 2019-06-01 DOI: 10.31399/asm.fach.med.c0047158
The pointed ends of several stainless steel forceps split or completely fractured where split portions broke off. All the forceps were delivered in the same lot. The pointed ends of the forceps are used for probing and gripping very small objects and must be true, sound, and sharp. Analysis supported the conclusion that the failures to be the result of seams in the steel that were not joined during hot working. Recommendations included that closer inspection of the product take place at all stages of manufacturing. Inspection at the mill will minimize discrepancies at the source, and the inspection of the finished product will help detect obscure seams.
几个不锈钢钳的尖端裂开或完全断裂,裂开的部分断裂。所有的产钳都是同一批次的。镊子的尖端用于探测和夹住非常小的物体,必须是真实的、声音的和锋利的。分析结果支持以下结论,即失效是由于热加工过程中未连接钢中的接缝造成的。建议包括在生产的所有阶段对产品进行更仔细的检查。在工厂的检查将最大限度地减少源头上的差异,对成品的检查将有助于发现不明显的接缝。
{"title":"Prongs on Forceps That Split and Fractured Due to Forging Seams","authors":"","doi":"10.31399/asm.fach.med.c0047158","DOIUrl":"https://doi.org/10.31399/asm.fach.med.c0047158","url":null,"abstract":"\u0000 The pointed ends of several stainless steel forceps split or completely fractured where split portions broke off. All the forceps were delivered in the same lot. The pointed ends of the forceps are used for probing and gripping very small objects and must be true, sound, and sharp. Analysis supported the conclusion that the failures to be the result of seams in the steel that were not joined during hot working. Recommendations included that closer inspection of the product take place at all stages of manufacturing. Inspection at the mill will minimize discrepancies at the source, and the inspection of the finished product will help detect obscure seams.","PeriodicalId":125471,"journal":{"name":"ASM Failure Analysis Case Histories: Medical and Biomedical Devices","volume":"7 1","pages":"0"},"PeriodicalIF":0.0,"publicationDate":"2019-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"124476798","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Fatigue Crack on a Type 316 Stainless Steel Bone Plate and Corresponding Broken Screw 316型不锈钢接骨板的疲劳裂纹及相应的断裂螺钉
Pub Date : 2019-06-01 DOI: 10.31399/asm.fach.med.c0048410
A narrow bone plate made of type 316 stainless steel and used to stabilize an open midshaft femur fracture failed. A crack at a plate hole next to the fracture site had been revealed by a radiograph taken 13 weeks after the operation. The plate was revealed to be slightly bent in the horizontal plane, and the fracture gap was considerably open. The screws and plates supplied by different manufacturers were revealed to be different with respect to microcleanliness (primary inclusion content) of the materials and only one of them was found to be according to specifications. The local crack formation was influenced by the presence of larger inclusions. The screw failed was revealed to have failed through a fatigue mechanism by the presence of striations in the scanning electron micrograph. The crack in the plate was revealed to have originated at the upper, outer corner of the plate by the beach marks which indicated the action of asymmetric bending and rotational forces.
用316型不锈钢制成的窄骨板固定开放式中轴股骨骨折失败。术后13周的x线片显示骨折部位附近的钢板孔处有裂缝。显示钢板在水平面上有轻微弯曲,骨折间隙相当开放。不同制造商提供的螺钉和板在材料的微清洁度(主要夹杂物含量)方面存在差异,其中只有一种符合规格。局部裂纹的形成受较大夹杂物的影响。通过扫描电子显微照片上的条纹显示,螺旋失效是通过疲劳机制失效的。在不对称弯曲力和旋转力的作用下,滩痕显示板的裂缝起源于板的上外角。
{"title":"Fatigue Crack on a Type 316 Stainless Steel Bone Plate and Corresponding Broken Screw","authors":"","doi":"10.31399/asm.fach.med.c0048410","DOIUrl":"https://doi.org/10.31399/asm.fach.med.c0048410","url":null,"abstract":"\u0000 A narrow bone plate made of type 316 stainless steel and used to stabilize an open midshaft femur fracture failed. A crack at a plate hole next to the fracture site had been revealed by a radiograph taken 13 weeks after the operation. The plate was revealed to be slightly bent in the horizontal plane, and the fracture gap was considerably open. The screws and plates supplied by different manufacturers were revealed to be different with respect to microcleanliness (primary inclusion content) of the materials and only one of them was found to be according to specifications. The local crack formation was influenced by the presence of larger inclusions. The screw failed was revealed to have failed through a fatigue mechanism by the presence of striations in the scanning electron micrograph. The crack in the plate was revealed to have originated at the upper, outer corner of the plate by the beach marks which indicated the action of asymmetric bending and rotational forces.","PeriodicalId":125471,"journal":{"name":"ASM Failure Analysis Case Histories: Medical and Biomedical Devices","volume":"193 1","pages":"0"},"PeriodicalIF":0.0,"publicationDate":"2019-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"133017445","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Heavy Pitting Corrosion on a Type 304 Stainless Steel Screw 304型不锈钢螺丝严重点蚀
Pub Date : 2019-06-01 DOI: 10.31399/asm.fach.med.c0048419
Heavy pitting corrosion on type 304 stainless steel bone screw was studied. A screw head that exhibited heavy pitting corrosion attack was observed. Deep tunnels that penetrated the screw head and followed the inclusion lines were revealed. The screw was inserted in a plate made of type 316LR stainless steel and some mechanical fretting and very few corrosion pits were revealed. Type 304 stainless steel was deemed not to be satisfactory as an implant material.
对304型不锈钢骨螺钉的重度点蚀进行了研究。观察到螺钉头表现出严重的点蚀攻击。钻穿螺旋头并沿夹杂物线延伸的深隧道显露出来。将螺钉插入316LR型不锈钢板中,发现有一些机械微动和很少的腐蚀坑。304型不锈钢被认为是不令人满意的植入材料。
{"title":"Heavy Pitting Corrosion on a Type 304 Stainless Steel Screw","authors":"","doi":"10.31399/asm.fach.med.c0048419","DOIUrl":"https://doi.org/10.31399/asm.fach.med.c0048419","url":null,"abstract":"\u0000 Heavy pitting corrosion on type 304 stainless steel bone screw was studied. A screw head that exhibited heavy pitting corrosion attack was observed. Deep tunnels that penetrated the screw head and followed the inclusion lines were revealed. The screw was inserted in a plate made of type 316LR stainless steel and some mechanical fretting and very few corrosion pits were revealed. Type 304 stainless steel was deemed not to be satisfactory as an implant material.","PeriodicalId":125471,"journal":{"name":"ASM Failure Analysis Case Histories: Medical and Biomedical Devices","volume":"36 1","pages":"0"},"PeriodicalIF":0.0,"publicationDate":"2019-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"134233776","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
期刊
ASM Failure Analysis Case Histories: Medical and Biomedical Devices
全部 Acc. Chem. Res. ACS Applied Bio Materials ACS Appl. Electron. Mater. ACS Appl. Energy Mater. ACS Appl. Mater. Interfaces ACS Appl. Nano Mater. ACS Appl. Polym. Mater. ACS BIOMATER-SCI ENG ACS Catal. ACS Cent. Sci. ACS Chem. Biol. ACS Chemical Health & Safety ACS Chem. Neurosci. ACS Comb. Sci. ACS Earth Space Chem. ACS Energy Lett. ACS Infect. Dis. ACS Macro Lett. ACS Mater. Lett. ACS Med. Chem. Lett. ACS Nano ACS Omega ACS Photonics ACS Sens. ACS Sustainable Chem. Eng. ACS Synth. Biol. Anal. Chem. BIOCHEMISTRY-US Bioconjugate Chem. BIOMACROMOLECULES Chem. Res. Toxicol. Chem. Rev. Chem. Mater. CRYST GROWTH DES ENERG FUEL Environ. Sci. Technol. Environ. Sci. Technol. Lett. Eur. J. Inorg. Chem. IND ENG CHEM RES Inorg. Chem. J. Agric. Food. Chem. J. Chem. Eng. Data J. Chem. Educ. J. Chem. Inf. Model. J. Chem. Theory Comput. J. Med. Chem. J. Nat. Prod. J PROTEOME RES J. Am. Chem. Soc. LANGMUIR MACROMOLECULES Mol. Pharmaceutics Nano Lett. Org. Lett. ORG PROCESS RES DEV ORGANOMETALLICS J. Org. Chem. J. Phys. Chem. J. Phys. Chem. A J. Phys. Chem. B J. Phys. Chem. C J. Phys. Chem. Lett. Analyst Anal. Methods Biomater. Sci. Catal. Sci. Technol. Chem. Commun. Chem. Soc. Rev. CHEM EDUC RES PRACT CRYSTENGCOMM Dalton Trans. Energy Environ. Sci. ENVIRON SCI-NANO ENVIRON SCI-PROC IMP ENVIRON SCI-WAT RES Faraday Discuss. Food Funct. Green Chem. Inorg. Chem. Front. Integr. Biol. J. Anal. At. Spectrom. J. Mater. Chem. A J. Mater. Chem. B J. Mater. Chem. C Lab Chip Mater. Chem. Front. Mater. Horiz. MEDCHEMCOMM Metallomics Mol. Biosyst. Mol. Syst. Des. Eng. Nanoscale Nanoscale Horiz. Nat. Prod. Rep. New J. Chem. Org. Biomol. Chem. Org. Chem. Front. PHOTOCH PHOTOBIO SCI PCCP Polym. Chem.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1