首页 > 最新文献

Biochemia Medica最新文献

英文 中文
Evaluation and management of leukolysis-mediated pseudohyperkalemia in paediatric leukemic samples. 评估和处理白血病患儿白血病溶解介导的假性高钾血症。
IF 3.3 3区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2022-02-15 DOI: 10.11613/BM.2022.010904
Lourens Jan Peter Nonkes, Valérie de Haas, Hans Kemperman, Albert Huisman, Ruben Eduardus Antonius Musson, Wouter Marcel Tiel Groenestege

Introduction: Leukolysis-related pseudohyperkalemia due to preanalytical procedures may lead to erroneous (or absence of) treatment based on an invalid lab test result. We aimed to obtain a leukocyte threshold above which leukolysis-related pseudohyperkalemia becomes clinical relevant. Secondly, temporal dynamics of treatment-induced leukocyte decrease were studied to allow tailored implementation of laboratory information system (LIS) decision rules based on the leukocyte threshold to avoid leukolysis-related pseudohyperkalemia.

Materials and methods: Potassium results of AU5811 routine chemistry (Beckman Coulter, Brea, California, USA) and iStat point of care (POC) (Abbott Diagnostics, Chicago, Illinois, USA) analysers were compared, the latter method being insensitive to leukolysis caused by pre-analytical procedures. Potassium results were combined with leukocyte counts obtained using a Cell-Dyn Sapphire haematology analyser (Abbott Diagnostics, Santa Clara, California, USA), resulting in 132 unique data triplets. Regression analysis was performed to establish a leukocyte threshold. The Reference Change Value (√2 x Z x √(CVa 2 + CVi 2)) was used to calculate maximum allowable difference between routine analyser and POC potassium results (deltamax + 0.58 mmol/L). Temporal analysis on the treatment-induced leukocyte decrease was performed by plotting leukocyte counts in time for all patients above the threshold leukocyte count (N = 41).

Results: Established leukocyte threshold was 63 x109/L. Temporal analysis showed leukocyte counts below the threshold within 8 days of treatment for all patients.

Conclusions: Based on performed analyses we were able to implement LIS decision rules to reduce pseudohyperkalemia due to preanalytical procedures. This implementation can contribute to a reduction in erroneous (or absence of) treatments in the clinic.

导读:分析前程序导致的与白血病相关的假性高钾血症可能导致基于无效实验室检测结果的错误(或缺乏)治疗。我们的目的是获得一个白细胞阈值,超过这个阈值,白细胞溶解相关的假性高钾血症就具有临床相关性。其次,研究了治疗诱导的白细胞减少的时间动态,以便根据白细胞阈值定制实验室信息系统(LIS)决策规则,以避免白细胞溶解相关的假性高钾血症。材料和方法:比较AU5811常规化学(Beckman Coulter, Brea, California, USA)和iStat护理点(POC) (Abbott Diagnostics, Chicago, Illinois, USA)分析仪的钾检测结果,后者对前分析程序引起的白细胞溶解不敏感。钾结果与使用Cell-Dyn Sapphire血液学分析仪(Abbott Diagnostics, Santa Clara, California, USA)获得的白细胞计数相结合,产生132个独特的三胞胎数据。采用回归分析建立白细胞阈值。参考变化值(√2 × Z ×√(CVa 2 + CVi 2))用于计算常规分析仪与POC钾结果(deltamax + 0.58 mmol/L)之间的最大允许差值。通过及时绘制白细胞计数高于阈值的所有患者(N = 41)的白细胞计数,对治疗诱导的白细胞减少进行时间分析。结果:建立的白细胞阈值为63 × 109/L。时间分析显示,所有患者在治疗8天内白细胞计数均低于阈值。结论:根据所进行的分析,我们能够实施LIS决策规则,以减少由于分析前程序引起的假性高钾血症。这种实现有助于减少临床错误(或缺乏)治疗。
{"title":"Evaluation and management of leukolysis-mediated pseudohyperkalemia in paediatric leukemic samples.","authors":"Lourens Jan Peter Nonkes,&nbsp;Valérie de Haas,&nbsp;Hans Kemperman,&nbsp;Albert Huisman,&nbsp;Ruben Eduardus Antonius Musson,&nbsp;Wouter Marcel Tiel Groenestege","doi":"10.11613/BM.2022.010904","DOIUrl":"https://doi.org/10.11613/BM.2022.010904","url":null,"abstract":"<p><strong>Introduction: </strong>Leukolysis-related pseudohyperkalemia due to preanalytical procedures may lead to erroneous (or absence of) treatment based on an invalid lab test result. We aimed to obtain a leukocyte threshold above which leukolysis-related pseudohyperkalemia becomes clinical relevant. Secondly, temporal dynamics of treatment-induced leukocyte decrease were studied to allow tailored implementation of laboratory information system (LIS) decision rules based on the leukocyte threshold to avoid leukolysis-related pseudohyperkalemia.</p><p><strong>Materials and methods: </strong>Potassium results of AU5811 routine chemistry (Beckman Coulter, Brea, California, USA) and iStat point of care (POC) (Abbott Diagnostics, Chicago, Illinois, USA) analysers were compared, the latter method being insensitive to leukolysis caused by pre-analytical procedures. Potassium results were combined with leukocyte counts obtained using a Cell-Dyn Sapphire haematology analyser (Abbott Diagnostics, Santa Clara, California, USA), resulting in 132 unique data triplets. Regression analysis was performed to establish a leukocyte threshold. The Reference Change Value (√2 x Z x √(CV<sub>a</sub> <sup>2</sup> + CV<sub>i</sub> <sup>2</sup>)) was used to calculate maximum allowable difference between routine analyser and POC potassium results (delta<sub>max</sub> + 0.58 mmol/L). Temporal analysis on the treatment-induced leukocyte decrease was performed by plotting leukocyte counts in time for all patients above the threshold leukocyte count (N = 41).</p><p><strong>Results: </strong>Established leukocyte threshold was 63 x10<sup>9</sup>/L. Temporal analysis showed leukocyte counts below the threshold within 8 days of treatment for all patients.</p><p><strong>Conclusions: </strong>Based on performed analyses we were able to implement LIS decision rules to reduce pseudohyperkalemia due to preanalytical procedures. This implementation can contribute to a reduction in erroneous (or absence of) treatments in the clinic.</p>","PeriodicalId":9021,"journal":{"name":"Biochemia Medica","volume":"32 1","pages":"010904"},"PeriodicalIF":3.3,"publicationDate":"2022-02-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8833246/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"39834174","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 1
Interpretative comments - need for harmonization? Results of the Croatian survey by the Working Group for Post-analytics. 解释性评论——需要统一吗?后分析工作小组克罗地亚调查的结果。
IF 3.3 3区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2022-02-15 Epub Date: 2021-12-15 DOI: 10.11613/BM.2022.010901
Vladimira Rimac, Sonja Podolar, Anja Jokic, Jelena Vlasic Tanaskovic, Lorena Honovic, Jasna Lenicek Krleza

Introduction: Interpretation of laboratory test results is a complex post-analytical activity that requires not only understanding of the clinical significance of laboratory results but also the analytical phase of laboratory work. The aims of this study were to determine: 1) the general opinion of Croatian medical biochemistry laboratories (MBLs) about the importance of interpretative comments on laboratory test reports, and 2) to find out whether harmonization of interpretative comments is needed.

Materials and methods: This retrospective study was designed as a survey by the Working Group for Post-analytics as part of national External Quality Assessment (EQA) program. All 195 MBLs participating in the national EQA scheme, were invited to participate in the survey. Results are reported as percentages of the total number of survey participants.

Results: Out of 195 MBLs, 162 participated in the survey (83%). Among them 59% MBLs implemented test result comments in routine according to national recommendations. The majority of laboratories (92%) state that interpretative comments added value to the laboratory reports, and a substantial part (72%) does not have feedback from physicians on their significance. Although physicians and patients ask for expert opinion, participants stated that the lack of interest of physicians (64%) as well as the inability to access patient's medical record (62%) affects the quality of expert opinion.

Conclusion: Although most participants state that they use interpretative comments and provide expert opinions regarding test results, results of the present study indicate that harmonization for interpretative comments is needed.

实验室检测结果的解释是一项复杂的分析后活动,不仅需要了解实验室结果的临床意义,还需要了解实验室工作的分析阶段。本研究的目的是确定:1)克罗地亚医学生物化学实验室(MBLs)对实验室检测报告解释性评论的重要性的一般意见,以及2)查明是否需要统一解释性评论。材料和方法:本回顾性研究是由后分析工作小组设计的一项调查,作为国家外部质量评估(EQA)计划的一部分。所有参加国家EQA计划的195个mbl都被邀请参加了调查。结果以调查参与者总数的百分比报告。结果:在195名MBLs中,162名(83%)参与了调查。其中59%的卫生院按照国家建议在日常工作中实施检测结果评议。大多数实验室(92%)表示解释性评论增加了实验室报告的价值,而且相当一部分(72%)没有医生对其重要性的反馈。虽然医生和病人都要求专家意见,但与会者表示,医生缺乏兴趣(64%)以及无法获得病人的医疗记录(62%)影响了专家意见的质量。结论:尽管大多数参与者表示他们使用解释性评论并就测试结果提供专家意见,但本研究的结果表明,解释性评论的协调是必要的。
{"title":"Interpretative comments - need for harmonization? Results of the Croatian survey by the Working Group for Post-analytics.","authors":"Vladimira Rimac,&nbsp;Sonja Podolar,&nbsp;Anja Jokic,&nbsp;Jelena Vlasic Tanaskovic,&nbsp;Lorena Honovic,&nbsp;Jasna Lenicek Krleza","doi":"10.11613/BM.2022.010901","DOIUrl":"https://doi.org/10.11613/BM.2022.010901","url":null,"abstract":"<p><strong>Introduction: </strong>Interpretation of laboratory test results is a complex post-analytical activity that requires not only understanding of the clinical significance of laboratory results but also the analytical phase of laboratory work. The aims of this study were to determine: 1) the general opinion of Croatian medical biochemistry laboratories (MBLs) about the importance of interpretative comments on laboratory test reports, and 2) to find out whether harmonization of interpretative comments is needed.</p><p><strong>Materials and methods: </strong>This retrospective study was designed as a survey by the Working Group for Post-analytics as part of national External Quality Assessment (EQA) program. All 195 MBLs participating in the national EQA scheme, were invited to participate in the survey. Results are reported as percentages of the total number of survey participants.</p><p><strong>Results: </strong>Out of 195 MBLs, 162 participated in the survey (83%). Among them 59% MBLs implemented test result comments in routine according to national recommendations. The majority of laboratories (92%) state that interpretative comments added value to the laboratory reports, and a substantial part (72%) does not have feedback from physicians on their significance. Although physicians and patients ask for expert opinion, participants stated that the lack of interest of physicians (64%) as well as the inability to access patient's medical record (62%) affects the quality of expert opinion.</p><p><strong>Conclusion: </strong>Although most participants state that they use interpretative comments and provide expert opinions regarding test results, results of the present study indicate that harmonization for interpretative comments is needed.</p>","PeriodicalId":9021,"journal":{"name":"Biochemia Medica","volume":"32 1","pages":"010901"},"PeriodicalIF":3.3,"publicationDate":"2022-02-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8672392/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"39851748","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Evaluation of the clinical chemistry tests analytical performance with Sigma Metric by using different quality specifications - Comparison of analyser actual performance with manufacturer data. 使用不同的质量规格,用西格玛标准评估临床化学检验的分析性能 - 比较分析仪的实际性能与制造商的数据。
IF 3.3 3区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2022-02-15 Epub Date: 2021-12-15 DOI: 10.11613/BM.2022.010703
Murat Keleş

Introduction: The interest in quality management tools/methodologies is gradually increasing to ensure quality and accurate results in line with international standards in clinical laboratories. Six Sigma stands apart from other methodologies with its total quality management system approach. However, the lack of standardization in tolerance limits restricts the advantages for the process. Our study aimed both to evaluate the applicability of analytical quality goals with Roche Cobas c 702 analyser and to determine achievable goals specific to the analyser used.

Materials and methods: The study examined under two main headings as Sigmalaboratory and Sigmaanalyser. Sigmalaboratory was calculated using internal and external quality control data by using Roche Cobas c 702 analyser for 21 routine biochemistry parameters and, Sigmaanalyser calculation was based on the manufacturer data presented in the package inserts of the reagents used in our laboratory during the study. Sigma values were calculated with the six sigma formula.

Results: Considering the total number of targets achieved, Sigmaanalyser performed best by meeting all CLIA goals, while Sigmalaboratory showed the lowest performance relative to biological variation (BV) desirable goals.

Conclusions: The balance between the applicability and analytical assurance of "goal-setting models" should be well established. Even if the package insert data provided by the manufacturer were used in our study, it was observed that almost a quarter of the evaluated analytes failed to achieve even "acceptable" level performance according to BV-based goals. Therefore, "state-of-the-art" goals for the Six Sigma methodology are considered to be more reasonable, achievable, and compatible with today's technologies.

导言:为了确保临床实验室的质量和结果的准确性符合国际标准,人们对质量管理工具/方法的兴趣逐渐增加。六西格玛以其全面质量管理系统的方法从其他方法中脱颖而出。然而,公差限值缺乏标准化限制了该方法的优势。我们的研究旨在评估罗氏 Cobas c 702 分析仪分析质量目标的适用性,并确定分析仪可实现的具体目标:研究主要从两个方面进行考察,即 Sigmalaboratory 和 Sigmaanalyser。Sigmalaboratory 是通过使用罗氏 Cobas c 702 分析仪对 21 项常规生化参数进行内部和外部质控数据计算得出的,而 Sigmaanalyser 的计算则基于本实验室在研究期间使用的试剂包装插页中提供的制造商数据。西格玛值用六西格玛公式计算:结果:考虑到达到的目标总数,Sigmaanalyser 表现最好,达到了 CLIA 的所有目标,而 Sigmalaboratory 在生物变异 (BV) 理想目标方面表现最低:结论:应在 "目标设定模型 "的适用性和分析保证之间取得平衡。在我们的研究中,即使使用了制造商提供的包装说明书数据,也发现几乎四分之一的受评估分析物甚至达不到基于生物变异的目标的 "可接受 "水平。因此,"六西格玛 "方法的 "最新 "目标被认为是更合理、更可实现、更符合当今技术的。
{"title":"Evaluation of the clinical chemistry tests analytical performance with Sigma Metric by using different quality specifications - Comparison of analyser actual performance with manufacturer data.","authors":"Murat Keleş","doi":"10.11613/BM.2022.010703","DOIUrl":"10.11613/BM.2022.010703","url":null,"abstract":"<p><strong>Introduction: </strong>The interest in quality management tools/methodologies is gradually increasing to ensure quality and accurate results in line with international standards in clinical laboratories. Six Sigma stands apart from other methodologies with its total quality management system approach. However, the lack of standardization in tolerance limits restricts the advantages for the process. Our study aimed both to evaluate the applicability of analytical quality goals with Roche Cobas c 702 analyser and to determine achievable goals specific to the analyser used.</p><p><strong>Materials and methods: </strong>The study examined under two main headings as Sigma<sub>laboratory</sub> and Sigma<sub>analyser</sub>. Sigma<sub>laboratory</sub> was calculated using internal and external quality control data by using Roche Cobas c 702 analyser for 21 routine biochemistry parameters and, Sigma<sub>analyser</sub> calculation was based on the manufacturer data presented in the package inserts of the reagents used in our laboratory during the study. Sigma values were calculated with the six sigma formula.</p><p><strong>Results: </strong>Considering the total number of targets achieved, Sigma<sub>analyser</sub> performed best by meeting all CLIA goals, while Sigma<sub>laboratory</sub> showed the lowest performance relative to biological variation (BV) desirable goals.</p><p><strong>Conclusions: </strong>The balance between the applicability and analytical assurance of \"goal-setting models\" should be well established. Even if the package insert data provided by the manufacturer were used in our study, it was observed that almost a quarter of the evaluated analytes failed to achieve even \"acceptable\" level performance according to BV-based goals. Therefore, \"state-of-the-art\" goals for the Six Sigma methodology are considered to be more reasonable, achievable, and compatible with today's technologies.</p>","PeriodicalId":9021,"journal":{"name":"Biochemia Medica","volume":"32 1","pages":"010703"},"PeriodicalIF":3.3,"publicationDate":"2022-02-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8672391/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"39763731","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Reevaluation of von Willebrand disease diagnosis in a Croatian paediatric cohort combining bleeding scores, phenotypic laboratory assays and next generation sequencing: a pilot study. 结合出血评分、表型实验室分析和下一代测序对克罗地亚儿科队列血管性血友病诊断的重新评估:一项试点研究
IF 3.3 3区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2022-02-15 DOI: 10.11613/BM.2022.010707
Ivana Lapić, Margareta Radić Antolic, Sara Dejanović Bekić, Désirée Coen-Herak, Ernest Bilić, Dunja Rogić, Renata Zadro

Introduction: This study reevaluated von Willebrand disease (vWD) diagnosis in a Croatian paediatric cohort by combining bleeding scores (BS), phenotypic laboratory testing, and next-generation sequencing (NGS).

Materials and methods: A total of 25 children (11 males and 14 females, median age 10 years, from 2 to 17) previously diagnosed with vWD were included. BS were calculated using an online bleeding assessment tool. Phenotypic laboratory analyses included platelet count, platelet function analyser closure times, prothrombin time, activated partial thromboplastin time, von Willebrand factor antigen (vWF:Ag), vWF gain-of-function mutant glycoprotein Ib binding activity (vWF:GPIbM), vWF collagen binding activity (vWF:CBA), factor VIII activity (FVIII:C) and multimeric analysis. Next-generation sequencing covered regions of both vWF and FVIII genes and was performed on MiSeq (Illumina, San Diego, USA).

Results: Disease-associated variants identified in 15 patients comprised 11 distinct heterozygous vWF gene variants in 13 patients and one novel FVIII gene variant (p.Glu2085Lys) in two male siblings. Four vWF variants were novel (p.Gln499Pro, p.Asp1277Tyr, p.Asp1277His, p.Lys1491Glu). Three patients without distinctive variants had vWF:GPIbM between 30 and 50%. Patients with identified vWF gene variants had statistically significant lower values of vWF:GPIbM (P = 0.002), vWF:Ag (P = 0.007), vWF:CBA (P < 0.001) and FVIII:C (P = 0.002), compared to those without. Correlations between BS and phenotypic laboratory test results were not statistically significant for either of the tests.

Conclusion: The applied diagnostic approach confirmed the diagnosis of vWD in 13 patients and mild haemophilia A in two. Limited utility of BS in the paediatric population was evidenced.

本研究通过结合出血评分(BS)、表型实验室检测和下一代测序(NGS),重新评估克罗地亚儿科队列中的血管性血友病(vWD)诊断。材料和方法:纳入既往诊断为vWD的儿童25例(男11例,女14例,中位年龄10岁,2 ~ 17岁)。使用在线出血评估工具计算BS。表型实验室分析包括血小板计数、血小板功能分析仪关闭时间、凝血酶原时间、活化的部分凝血酶时间、血管性血友病因子抗原(vWF:Ag)、vWF功能获得型突变糖蛋白Ib结合活性(vWF:GPIbM)、vWF胶原结合活性(vWF:CBA)、因子VIII活性(FVIII:C)和多聚体分析。下一代测序覆盖了vWF和FVIII基因的区域,并在MiSeq (Illumina, San Diego, USA)上进行。结果:在15例患者中发现的疾病相关变异体包括13例患者中11种不同的杂合vWF基因变异体和2例男性兄弟姐妹中1种新的FVIII基因变异体(p.Glu2085Lys)。4个vWF变异是新发现的(p.Gln499Pro, p.Asp1277Tyr, p.Asp1277His, p.Lys1491Glu)。三名无明显变异的患者的vWF:GPIbM在30 - 50%之间。发现vWF基因变异的患者vWF:GPIbM (P = 0.002)、vWF:Ag (P = 0.007)、vWF:CBA (P < 0.001)和FVIII:C (P = 0.002)值均低于未发现vWF基因变异的患者。BS与表型实验室测试结果之间的相关性在两项测试中均无统计学意义。结论:应用诊断方法确诊vWD 13例,轻度A型血友病2例。证明了BS在儿科人群中的有限效用。
{"title":"Reevaluation of von Willebrand disease diagnosis in a Croatian paediatric cohort combining bleeding scores, phenotypic laboratory assays and next generation sequencing: a pilot study.","authors":"Ivana Lapić,&nbsp;Margareta Radić Antolic,&nbsp;Sara Dejanović Bekić,&nbsp;Désirée Coen-Herak,&nbsp;Ernest Bilić,&nbsp;Dunja Rogić,&nbsp;Renata Zadro","doi":"10.11613/BM.2022.010707","DOIUrl":"https://doi.org/10.11613/BM.2022.010707","url":null,"abstract":"<p><strong>Introduction: </strong>This study reevaluated von Willebrand disease (vWD) diagnosis in a Croatian paediatric cohort by combining bleeding scores (BS), phenotypic laboratory testing, and next-generation sequencing (NGS).</p><p><strong>Materials and methods: </strong>A total of 25 children (11 males and 14 females, median age 10 years, from 2 to 17) previously diagnosed with vWD were included. BS were calculated using an online bleeding assessment tool. Phenotypic laboratory analyses included platelet count, platelet function analyser closure times, prothrombin time, activated partial thromboplastin time, von Willebrand factor antigen (vWF:Ag), vWF gain-of-function mutant glycoprotein Ib binding activity (vWF:GPIbM), vWF collagen binding activity (vWF:CBA), factor VIII activity (FVIII:C) and multimeric analysis. Next-generation sequencing covered regions of both vWF and FVIII genes and was performed on MiSeq (Illumina, San Diego, USA).</p><p><strong>Results: </strong>Disease-associated variants identified in 15 patients comprised 11 distinct heterozygous vWF gene variants in 13 patients and one novel FVIII gene variant (p.Glu2085Lys) in two male siblings. Four vWF variants were novel (p.Gln499Pro, p.Asp1277Tyr, p.Asp1277His, p.Lys1491Glu). Three patients without distinctive variants had vWF:GPIbM between 30 and 50%. Patients with identified vWF gene variants had statistically significant lower values of vWF:GPIbM (P = 0.002), vWF:Ag (P = 0.007), vWF:CBA (P < 0.001) and FVIII:C (P = 0.002), compared to those without. Correlations between BS and phenotypic laboratory test results were not statistically significant for either of the tests.</p><p><strong>Conclusion: </strong>The applied diagnostic approach confirmed the diagnosis of vWD in 13 patients and mild haemophilia A in two. Limited utility of BS in the paediatric population was evidenced.</p>","PeriodicalId":9021,"journal":{"name":"Biochemia Medica","volume":"32 1","pages":"010707"},"PeriodicalIF":3.3,"publicationDate":"2022-02-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8833252/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"39834172","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 1
The importance of regulation (EU) 2017/746 for quality control in medical laboratories. 法规(EU) 2017/746对医学实验室质量控制的重要性
IF 3.3 3区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2022-02-15 Epub Date: 2021-12-15 DOI: 10.11613/BM.2022.010301
Cristiano Ialongo, Maria Sapio, Leda Elisabetta Antetomaso, Antonio Angeloni
The guideline C24 (now in its 4th edition) issued by the Clinical and Laboratory Standards Institute (CLSI) recommends that for purchased quality control (QC), the laboratory should never use the manufacturer’s declared target (T) value but that identified by testing the product at least ten times with each new batch (1). Of course, verifying T is quite different from estimating the variability of the analytical process, albeit both are used for the statistical process control (SPC) since T should be the production target of the QC material on which the manufacturer has complete control.
{"title":"The importance of regulation (EU) 2017/746 for quality control in medical laboratories.","authors":"Cristiano Ialongo,&nbsp;Maria Sapio,&nbsp;Leda Elisabetta Antetomaso,&nbsp;Antonio Angeloni","doi":"10.11613/BM.2022.010301","DOIUrl":"https://doi.org/10.11613/BM.2022.010301","url":null,"abstract":"The guideline C24 (now in its 4th edition) issued by the Clinical and Laboratory Standards Institute (CLSI) recommends that for purchased quality control (QC), the laboratory should never use the manufacturer’s declared target (T) value but that identified by testing the product at least ten times with each new batch (1). Of course, verifying T is quite different from estimating the variability of the analytical process, albeit both are used for the statistical process control (SPC) since T should be the production target of the QC material on which the manufacturer has complete control.","PeriodicalId":9021,"journal":{"name":"Biochemia Medica","volume":"32 1","pages":"010301"},"PeriodicalIF":3.3,"publicationDate":"2022-02-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8672393/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"39763728","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 1
Verification of Atellica 1500 and comparison with Iris urine analyser and urine culture. Atellica 1500的验证及与Iris尿液分析仪和尿液培养的比较。
IF 3.3 3区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2022-02-15 Epub Date: 2021-12-15 DOI: 10.11613/BM.2022.010701
Ana Nikler, Helena Čičak, Danijela Bejuk, Vanja Radišić Biljak, Ana-Maria Šimundić

Introduction: The aims of study were to assess: 1) performance specifications of Atellica 1500, 2) comparability of Atellica 1500 and Iris, 3) the accuracy of both analysers in their ability to detect bacteria.

Materials and methods: Carryover, linearity, precision, reproducibility, and limit of blank (LoB) verification were evaluated for erythrocyte and leukocyte counts. ICSH 2014 protocol was used for estimation of carryover, CLSI EP15-A3 for precision, and CLSI EP17 for LoB verification. Comparison for quantitative parameters was evaluated by Bland-Altman plot and Passing-Bablok regression. Qualitative parameters were evaluated by Weighted kappa analysis. Sixty-five urine samples were randomly selected and sent for urine culture which was used as reference method to determine the accuracy of bacteria detection by analysers.

Results: Analytical specifications of Atellica 1500 were successfully verified. Total of 393 samples were used for qualitative comparison, while 269 for sediment urinalysis. Bland-Altman analysis showed statistically significant proportional bias for erythrocytes and leukocytes. Passing-Bablok analysis for leukocytes pointed to significant constant and minor proportional difference, while it was not performed for erythrocytes due to significant data deviation from linearity. Kappa analysis resulted in the strongest agreements for pH, ketones, glucose concentrations and leukocytes, while the poorest agreement for bacteria. The sensitivity and specificity of bacteria detection were: 91 (59-100)% and 76 (66-87)% for Atellica 1500 and 46 (17-77)% and 96 (87-100)% for Iris.

Conclusion: There are large differences between Atellica 1500 and Iris analysers, due to which they are not comparable and can not be used interchangeably. While there was no difference in specificity of bacteria detection, Iris analyser had greater sensitivity.

本研究的目的是评估:1)Atellica 1500的性能指标,2)Atellica 1500和Iris的可比性,3)两种分析仪检测细菌能力的准确性。材料和方法:对红细胞和白细胞计数的结转、线性、精密度、重现性和空白(LoB)验证限进行了评价。结转估计采用ICSH 2014协议,精度采用CLSI EP15-A3协议,LoB验证采用CLSI EP17协议。定量参数比较采用Bland-Altman图和Passing-Bablok回归。采用加权kappa分析法对定性参数进行评价。随机抽取65份尿液样本进行尿液培养,作为参考方法,以确定分析仪检测细菌的准确性。结果:成功验证了Atellica 1500的分析规范。共有393份样本用于定性比较,269份样本用于沉积物分析。Bland-Altman分析显示红细胞和白细胞的比例偏倚具有统计学意义。白细胞的pass - bablok分析显示有显著的常数和较小的比例差异,而红细胞由于数据明显偏离线性而未进行pass - bablok分析。Kappa分析结果表明,pH值、酮类、葡萄糖浓度和白细胞的一致性最强,而细菌的一致性最差。细菌检测的灵敏度和特异度分别为:Atellica 1500的91(59 ~ 100)%和76 (66 ~ 87)%,Iris的46(17 ~ 77)%和96(87 ~ 100)%。结论:Atellica 1500型分析仪与Iris型分析仪存在较大差异,不具有可比性,不能互换使用。细菌检测的特异性没有差异,但虹膜分析仪的灵敏度更高。
{"title":"Verification of Atellica 1500 and comparison with Iris urine analyser and urine culture.","authors":"Ana Nikler,&nbsp;Helena Čičak,&nbsp;Danijela Bejuk,&nbsp;Vanja Radišić Biljak,&nbsp;Ana-Maria Šimundić","doi":"10.11613/BM.2022.010701","DOIUrl":"https://doi.org/10.11613/BM.2022.010701","url":null,"abstract":"<p><strong>Introduction: </strong>The aims of study were to assess: 1) performance specifications of Atellica 1500, 2) comparability of Atellica 1500 and Iris, 3) the accuracy of both analysers in their ability to detect bacteria.</p><p><strong>Materials and methods: </strong>Carryover, linearity, precision, reproducibility, and limit of blank (LoB) verification were evaluated for erythrocyte and leukocyte counts. ICSH 2014 protocol was used for estimation of carryover, CLSI EP15-A3 for precision, and CLSI EP17 for LoB verification. Comparison for quantitative parameters was evaluated by Bland-Altman plot and Passing-Bablok regression. Qualitative parameters were evaluated by Weighted kappa analysis. Sixty-five urine samples were randomly selected and sent for urine culture which was used as reference method to determine the accuracy of bacteria detection by analysers.</p><p><strong>Results: </strong>Analytical specifications of Atellica 1500 were successfully verified. Total of 393 samples were used for qualitative comparison, while 269 for sediment urinalysis. Bland-Altman analysis showed statistically significant proportional bias for erythrocytes and leukocytes. Passing-Bablok analysis for leukocytes pointed to significant constant and minor proportional difference, while it was not performed for erythrocytes due to significant data deviation from linearity. Kappa analysis resulted in the strongest agreements for pH, ketones, glucose concentrations and leukocytes, while the poorest agreement for bacteria. The sensitivity and specificity of bacteria detection were: 91 (59-100)% and 76 (66-87)% for Atellica 1500 and 46 (17-77)% and 96 (87-100)% for Iris.</p><p><strong>Conclusion: </strong>There are large differences between Atellica 1500 and Iris analysers, due to which they are not comparable and can not be used interchangeably. While there was no difference in specificity of bacteria detection, Iris analyser had greater sensitivity.</p>","PeriodicalId":9021,"journal":{"name":"Biochemia Medica","volume":"32 1","pages":"010701"},"PeriodicalIF":3.3,"publicationDate":"2022-02-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8672386/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"39763729","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 1
Deviating glucose results in an international dual-center study. A root cause investigation. 一项国际双中心研究结果显示葡萄糖偏离。根本原因调查。
IF 3.3 3区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2022-02-15 Epub Date: 2021-12-15 DOI: 10.11613/BM.2022.011001
Janne Cadamuro, Peter Bergsten, Katharina Mörwald, Anders Forslund, Marie Dahlbom, Jonas Bergquist, Iris Ciba, Susanne M Brunner, Jeanne Jabbour, Daniel Weghuber

During a dual-center study on obese and normal weight children and adolescents, focusing on glucose metabolism, we observed a marked difference in glucose results (N = 16,840) between the two sites, Salzburg, Austria and Uppsala, Sweden (P < 0.001). After excluding differences in patient characteristics between the two populations as cause of this finding, we investigated other preanalytic influences. Finally, only the tubes used for blood collection at the two sites were left to evaluate. While the Vacuette FC-Mix tube (Greiner Bio-One, Kremsmünster, Austria) was used in Uppsala, in Salzburg blood collections were performed with a lithium heparin tube (LH-Monovette, Sarstedt, Germany). To prove our hypothesis, we collected two blood samples in either of these tubes from 51 children (Salzburg N = 27, Uppsala N = 24) and compared the measured glucose results. Indeed, we found the suspected bias and calculated a correction formula, which significantly diminished the differences of glucose results between the two sites (P = 0.023). Our finding is in line with those of other studies and although this issue should be widely known, we feel that it is widely neglected, especially when comparing glucose concentrations across Europe, using large databases without any information on preanalytic sample handling.

在一项针对肥胖和正常体重儿童和青少年的双中心研究中,我们观察到两个地点(奥地利萨尔茨堡和瑞典乌普萨拉)的葡萄糖结果有显著差异(N = 16,840) (P < 0.001)。在排除了两个人群之间患者特征的差异作为这一发现的原因后,我们调查了其他分析前影响。最后,只留下两个地点用于采血的管子进行评估。乌普萨拉使用Vacuette FC-Mix管(Greiner Bio-One, kremsmnster,奥地利),在萨尔茨堡采集血液时使用锂肝素管(hl - monovette,德国萨斯泰特)。为了证明我们的假设,我们从51名儿童(Salzburg N = 27, Uppsala N = 24)中采集了两份血样,并比较了测量的葡萄糖结果。事实上,我们发现了可疑的偏倚,并计算了一个校正公式,显著减小了两个地点之间葡萄糖结果的差异(P = 0.023)。我们的发现与其他研究一致,尽管这个问题应该广为人知,但我们觉得它被广泛忽视了,特别是在比较欧洲各地的葡萄糖浓度时,使用大型数据库而没有任何分析前样品处理的信息。
{"title":"Deviating glucose results in an international dual-center study. A root cause investigation.","authors":"Janne Cadamuro,&nbsp;Peter Bergsten,&nbsp;Katharina Mörwald,&nbsp;Anders Forslund,&nbsp;Marie Dahlbom,&nbsp;Jonas Bergquist,&nbsp;Iris Ciba,&nbsp;Susanne M Brunner,&nbsp;Jeanne Jabbour,&nbsp;Daniel Weghuber","doi":"10.11613/BM.2022.011001","DOIUrl":"https://doi.org/10.11613/BM.2022.011001","url":null,"abstract":"<p><p>During a dual-center study on obese and normal weight children and adolescents, focusing on glucose metabolism, we observed a marked difference in glucose results (N = 16,840) between the two sites, Salzburg, Austria and Uppsala, Sweden (P < 0.001). After excluding differences in patient characteristics between the two populations as cause of this finding, we investigated other preanalytic influences. Finally, only the tubes used for blood collection at the two sites were left to evaluate. While the Vacuette FC-Mix tube (Greiner Bio-One, Kremsmünster, Austria) was used in Uppsala, in Salzburg blood collections were performed with a lithium heparin tube (LH-Monovette, Sarstedt, Germany). To prove our hypothesis, we collected two blood samples in either of these tubes from 51 children (Salzburg N = 27, Uppsala N = 24) and compared the measured glucose results. Indeed, we found the suspected bias and calculated a correction formula, which significantly diminished the differences of glucose results between the two sites (<i>P</i> = 0.023). Our finding is in line with those of other studies and although this issue should be widely known, we feel that it is widely neglected, especially when comparing glucose concentrations across Europe, using large databases without any information on preanalytic sample handling.</p>","PeriodicalId":9021,"journal":{"name":"Biochemia Medica","volume":"32 1","pages":"011001"},"PeriodicalIF":3.3,"publicationDate":"2022-02-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8672384/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"39851751","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
A sudden creatinine increase: A case report. 肌酸酐突然升高1例。
IF 3.3 3区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2022-02-15 DOI: 10.11613/BM.2022.011002
Bernhard Strasser, Sebastian Strasser, Josef Tomasits

Creatinine and estimated glomerular filtration rate (eGFR) are first-line laboratory parameters in the diagnosis of various renal diseases. In recent decades, cystatin C (cysC) has furthered the laboratory repertoire regarding renal status assessment and has been implemented in many clinical guidelines. Accordingly, with the establishment of cysC as a renal routine biomarker, further opportunities for assessing eGFR have been attained. Nevertheless, various limitations are still associated with cysC and creatinine analysis. Preanalytical errors could cause false results in both biomarkers. In our case, we were confronted with implausibly elevated creatinine levels due to preanalytical errors.

肌酐和估计肾小球滤过率(eGFR)是诊断各种肾脏疾病的一线实验室参数。近几十年来,胱抑素C (cystatin C, cysC)在肾脏状态评估方面得到了进一步的实验室应用,并在许多临床指南中得到了应用。因此,随着cysC作为肾脏常规生物标志物的确立,进一步获得了评估eGFR的机会。然而,各种限制仍然与cysC和肌酐分析有关。分析前错误可能导致两种生物标志物的错误结果。在我们的病例中,由于分析前的错误,我们面临着令人难以置信的肌酐水平升高。
{"title":"A sudden creatinine increase: A case report.","authors":"Bernhard Strasser,&nbsp;Sebastian Strasser,&nbsp;Josef Tomasits","doi":"10.11613/BM.2022.011002","DOIUrl":"https://doi.org/10.11613/BM.2022.011002","url":null,"abstract":"<p><p>Creatinine and estimated glomerular filtration rate (eGFR) are first-line laboratory parameters in the diagnosis of various renal diseases. In recent decades, cystatin C (cysC) has furthered the laboratory repertoire regarding renal status assessment and has been implemented in many clinical guidelines. Accordingly, with the establishment of cysC as a renal routine biomarker, further opportunities for assessing eGFR have been attained. Nevertheless, various limitations are still associated with cysC and creatinine analysis. Preanalytical errors could cause false results in both biomarkers. In our case, we were confronted with implausibly elevated creatinine levels due to preanalytical errors.</p>","PeriodicalId":9021,"journal":{"name":"Biochemia Medica","volume":"32 1","pages":"011002"},"PeriodicalIF":3.3,"publicationDate":"2022-02-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8833249/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"39670809","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Moving average procedures as an additional tool for real-time analytical quality control: challenges and opportunities of implementation in small-volume medical laboratories. 移动平均程序作为实时分析质量控制的额外工具:在小批量医学实验室实施的挑战和机遇。
IF 3.3 3区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2022-02-15 Epub Date: 2021-12-15 DOI: 10.11613/BM.2022.010705
Vera Lukić, Svetlana Ignjatović

Introduction: Moving average (MA) is one possible way to use patient results for analytical quality control in medical laboratories. The aims of this study were to: (1) implement previously optimized MA procedures for 10 clinical chemistry analytes into the laboratory information system (LIS); (2) monitor their performance as a real-time quality control tool, and (3) define an algorithm for MA alarm management in a small-volume laboratory to suit the specific laboratory.

Materials and methods: Moving average alarms were monitored and analysed over a period of 6 months on all patient results (total of 73,059) obtained for 10 clinical chemistry parameters. The optimal MA procedures were selected previously using an already described technique called the bias detection simulation method, considering the ability of bias detection the size of total allowable error as the key parameter for optimization.

Results: During 6 months, 17 MA alarms were registered, which is 0.023% of the total number of generated MA values. In 65% of cases, their cause was of pre-analytical origin, in 12% of analytical origin, and in 23% the cause was not found. The highest alarm rate was determined on sodium (0.10%), and the lowest on calcium and chloride.

Conclusions: This paper showed that even in a small-volume laboratory, previously optimized MA procedures could be successfully implemented in the LIS and used for continuous quality control. Review of patient results, re-analysis of samples from the stable period, analysis of internal quality control samples and assessment of the analyser malfunctions and maintenance log have been proposed for the algorithm for managing MA alarms.

简介:移动平均(MA)是医学实验室中使用患者结果进行分析质量控制的一种可能方法。本研究的目的是:(1)在实验室信息系统(LIS)中实施先前优化的10种临床化学分析物的MA程序;(2)监测其性能,作为实时质量控制工具;(3)定义适合特定实验室的小容量实验室MA报警管理算法。材料和方法:对10项临床化学参数的所有患者结果(共73,059例)进行为期6个月的移动平均警报监测和分析。考虑到偏差检测能力和总允许误差的大小作为优化的关键参数,之前使用已经描述的技术称为偏差检测模拟方法来选择最优的MA程序。结果:6个月共登记MA报警17例,占MA总生成数的0.023%。在65%的病例中,他们的原因是分析前的起源,在12%的分析起源,在23%的原因未被发现。钠的报警率最高(0.10%),钙和氯的报警率最低。结论:本文表明,即使在小容量实验室中,先前优化的MA程序也可以在LIS中成功实施并用于持续质量控制。针对MA报警管理算法,提出了对患者结果的回顾、对稳定期样本的重新分析、对内部质量控制样本的分析以及对分析仪故障和维护日志的评估。
{"title":"Moving average procedures as an additional tool for real-time analytical quality control: challenges and opportunities of implementation in small-volume medical laboratories.","authors":"Vera Lukić,&nbsp;Svetlana Ignjatović","doi":"10.11613/BM.2022.010705","DOIUrl":"https://doi.org/10.11613/BM.2022.010705","url":null,"abstract":"<p><strong>Introduction: </strong>Moving average (MA) is one possible way to use patient results for analytical quality control in medical laboratories. The aims of this study were to: (1) implement previously optimized MA procedures for 10 clinical chemistry analytes into the laboratory information system (LIS); (2) monitor their performance as a real-time quality control tool, and (3) define an algorithm for MA alarm management in a small-volume laboratory to suit the specific laboratory.</p><p><strong>Materials and methods: </strong>Moving average alarms were monitored and analysed over a period of 6 months on all patient results (total of 73,059) obtained for 10 clinical chemistry parameters. The optimal MA procedures were selected previously using an already described technique called the bias detection simulation method, considering the ability of bias detection the size of total allowable error as the key parameter for optimization.</p><p><strong>Results: </strong>During 6 months, 17 MA alarms were registered, which is 0.023% of the total number of generated MA values. In 65% of cases, their cause was of pre-analytical origin, in 12% of analytical origin, and in 23% the cause was not found. The highest alarm rate was determined on sodium (0.10%), and the lowest on calcium and chloride.</p><p><strong>Conclusions: </strong>This paper showed that even in a small-volume laboratory, previously optimized MA procedures could be successfully implemented in the LIS and used for continuous quality control. Review of patient results, re-analysis of samples from the stable period, analysis of internal quality control samples and assessment of the analyser malfunctions and maintenance log have been proposed for the algorithm for managing MA alarms.</p>","PeriodicalId":9021,"journal":{"name":"Biochemia Medica","volume":"32 1","pages":"010705"},"PeriodicalIF":3.3,"publicationDate":"2022-02-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8672389/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"39851747","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 2
Poor correlation between alcohol concentration in oral fluid and breath in subjects consuming beverages immediately before testing. 在测试前立即饮用饮料的受试者口服液中的酒精浓度与呼吸之间的相关性较差。
IF 3.3 3区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2022-02-15 Epub Date: 2021-12-15 DOI: 10.11613/BM.2022.010902
Hallvard Gjerde, Anne Line Bretteville-Jensen, Håvard Furuhaugen

Introduction: In previous research projects and clinical settings, alcohol analysis in oral fluid (saliva) has been used as an alternative to breath or blood alcohol testing. In this study we examined whether it is possible to obtain clinically relevant data regarding alcohol consumption in individuals who recently consumed alcohol by analysing oral fluid samples when the recommended rinsing of the mouth is impossible before sample collection.

Materials and methods: We conducted a study of 89 nightclub patrons in Norway. Before collecting oral fluid samples and performing breath alcohol testing, participants were required to drink a glass of water to remove residual alcohol from the mouth. Oral fluid samples were collected with the Quantisal oral fluid collection device and analysed using an enzymatic method for alcohol. The alcohol concentration in the neat (undiluted) oral fluid was then calculated. Breath alcohol testing was performed using Lion Alcolmeter 500 instruments.

Results: No false-negative or false-positive results for alcohol were detected in the oral fluid when compared with those in the breath. The Intraclass Correlation Coefficient of 0.40 indicated a poor correlation between alcohol concentrations in the two sample types.

Conclusions: The procedure for collecting oral fluid was suitable for the qualitative determination of alcohol intake but not for quantitative assessment. We recommend that oral fluid samples should not be used for estimating blood or breath alcohol concentrations in people who have recently consumed alcohol or non-alcoholic beverages, as recommended in the instructions for use.

在以前的研究项目和临床环境中,口服液(唾液)中的酒精分析已被用作呼气或血液酒精测试的替代方法。在这项研究中,我们研究了是否有可能通过分析最近饮酒的个体的口腔液样本来获得有关酒精消耗的临床相关数据,因为在收集样本之前不可能建议漱口。材料和方法:我们对挪威89名夜总会顾客进行了一项研究。在收集口服液样本和进行呼气酒精测试之前,参与者被要求喝一杯水以清除口腔中的残留酒精。用定量口服液收集装置收集口服液样品,用酶法分析酒精。然后计算纯(未稀释)口服液中的酒精浓度。呼气酒精测试使用Lion Alcolmeter 500仪器进行。结果:与呼气中的酒精相比,口腔液中没有检测到假阴性或假阳性的结果。类内相关系数为0.40,表明两种样品中酒精浓度相关性较差。结论:口服液采集法适用于酒精摄入量的定性测定,但不适用于定量评价。我们建议,不应按照使用说明中的建议,将口服液样本用于估计最近饮用过酒精或非酒精饮料的人的血液或呼吸中的酒精浓度。
{"title":"Poor correlation between alcohol concentration in oral fluid and breath in subjects consuming beverages immediately before testing.","authors":"Hallvard Gjerde,&nbsp;Anne Line Bretteville-Jensen,&nbsp;Håvard Furuhaugen","doi":"10.11613/BM.2022.010902","DOIUrl":"https://doi.org/10.11613/BM.2022.010902","url":null,"abstract":"<p><strong>Introduction: </strong>In previous research projects and clinical settings, alcohol analysis in oral fluid (saliva) has been used as an alternative to breath or blood alcohol testing. In this study we examined whether it is possible to obtain clinically relevant data regarding alcohol consumption in individuals who recently consumed alcohol by analysing oral fluid samples when the recommended rinsing of the mouth is impossible before sample collection.</p><p><strong>Materials and methods: </strong>We conducted a study of 89 nightclub patrons in Norway. Before collecting oral fluid samples and performing breath alcohol testing, participants were required to drink a glass of water to remove residual alcohol from the mouth. Oral fluid samples were collected with the Quantisal oral fluid collection device and analysed using an enzymatic method for alcohol. The alcohol concentration in the neat (undiluted) oral fluid was then calculated. Breath alcohol testing was performed using Lion Alcolmeter 500 instruments.</p><p><strong>Results: </strong>No false-negative or false-positive results for alcohol were detected in the oral fluid when compared with those in the breath. The Intraclass Correlation Coefficient of 0.40 indicated a poor correlation between alcohol concentrations in the two sample types.</p><p><strong>Conclusions: </strong>The procedure for collecting oral fluid was suitable for the qualitative determination of alcohol intake but not for quantitative assessment. We recommend that oral fluid samples should not be used for estimating blood or breath alcohol concentrations in people who have recently consumed alcohol or non-alcoholic beverages, as recommended in the instructions for use.</p>","PeriodicalId":9021,"journal":{"name":"Biochemia Medica","volume":"32 1","pages":"010902"},"PeriodicalIF":3.3,"publicationDate":"2022-02-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8672390/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"39851749","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
期刊
Biochemia Medica
全部 Acc. Chem. Res. ACS Applied Bio Materials ACS Appl. Electron. Mater. ACS Appl. Energy Mater. ACS Appl. Mater. Interfaces ACS Appl. Nano Mater. ACS Appl. Polym. Mater. ACS BIOMATER-SCI ENG ACS Catal. ACS Cent. Sci. ACS Chem. Biol. ACS Chemical Health & Safety ACS Chem. Neurosci. ACS Comb. Sci. ACS Earth Space Chem. ACS Energy Lett. ACS Infect. Dis. ACS Macro Lett. ACS Mater. Lett. ACS Med. Chem. Lett. ACS Nano ACS Omega ACS Photonics ACS Sens. ACS Sustainable Chem. Eng. ACS Synth. Biol. Anal. Chem. BIOCHEMISTRY-US Bioconjugate Chem. BIOMACROMOLECULES Chem. Res. Toxicol. Chem. Rev. Chem. Mater. CRYST GROWTH DES ENERG FUEL Environ. Sci. Technol. Environ. Sci. Technol. Lett. Eur. J. Inorg. Chem. IND ENG CHEM RES Inorg. Chem. J. Agric. Food. Chem. J. Chem. Eng. Data J. Chem. Educ. J. Chem. Inf. Model. J. Chem. Theory Comput. J. Med. Chem. J. Nat. Prod. J PROTEOME RES J. Am. Chem. Soc. LANGMUIR MACROMOLECULES Mol. Pharmaceutics Nano Lett. Org. Lett. ORG PROCESS RES DEV ORGANOMETALLICS J. Org. Chem. J. Phys. Chem. J. Phys. Chem. A J. Phys. Chem. B J. Phys. Chem. C J. Phys. Chem. Lett. Analyst Anal. Methods Biomater. Sci. Catal. Sci. Technol. Chem. Commun. Chem. Soc. Rev. CHEM EDUC RES PRACT CRYSTENGCOMM Dalton Trans. Energy Environ. Sci. ENVIRON SCI-NANO ENVIRON SCI-PROC IMP ENVIRON SCI-WAT RES Faraday Discuss. Food Funct. Green Chem. Inorg. Chem. Front. Integr. Biol. J. Anal. At. Spectrom. J. Mater. Chem. A J. Mater. Chem. B J. Mater. Chem. C Lab Chip Mater. Chem. Front. Mater. Horiz. MEDCHEMCOMM Metallomics Mol. Biosyst. Mol. Syst. Des. Eng. Nanoscale Nanoscale Horiz. Nat. Prod. Rep. New J. Chem. Org. Biomol. Chem. Org. Chem. Front. PHOTOCH PHOTOBIO SCI PCCP Polym. Chem.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1