首页 > 最新文献

Journal of Applied Clinical Medical Physics最新文献

英文 中文
Exposure index in digital radiography and its dependence on acquisition parameters, anatomy, and manufacturer 数字放射照相的曝光指数及其对采集参数、解剖结构和制造商的依赖。
IF 2.2 4区 医学 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2026-01-14 DOI: 10.1002/acm2.70331
Ioannis A. Tsalafoutas, Shady AlKhazzam, Mohammed Hassan Kharita

PURPOSE

The exposure index (EI), the target exposure index (EIT), and the deviation index (DI) have been defined in the IEC Standard 62494-1 Ed.1 2008-08. This study investigates the impact of certain acquisition parameters, the imaged anatomy, and the manufacturer's specificities on the EI of radiological images and how these may affect EIT setting procedure.

METHODS

Images were acquired using two digital radiography (DR) systems of two different manufacturers, using aluminum attenuators and an anthropomorphic phantom. Acquisition parameters like the tube potential (kVp), the tube loading (mAs), the exposure time, the automatic exposure control (AEC) system settings (sensor and dose level selection), the grid (with or without), the additional filtration, the field size, and the imaged anatomy were varied and their effect on the EI was quantified separately for each system.

RESULTS

EI is linearly related to the incident air kerma (IAK) on the detector as expected (by definition). For constant IAK, EI increases with increasing kVp. While EI in general is reduced in the presence of scatter, this may not always be the case. Under AEC operation, even the exposure time can make a difference. EI is strongly affected by the imaged anatomy in combination with the AEC sensor and field size selections, the examination protocol, and the manufacturer.

CONCLUSIONS

Many parameters affect the EI calculation apart from IAK. Among them, the most important are the imaged anatomy and the manufacturer. Since the EI calculation is a complex procedure, setting of the EIT values should be done with caution on a per-examination and manufacturer basis, since the values that apply for one digital system are not always applicable to another. Furthermore, when EI is used as an image quality tool, a DI variation of at least ±2 should be allowed before a possibly meaningful red flag is activated.

目的:在IEC标准62494-1 Ed.1 2008-08中定义了暴露指数(EI)、目标暴露指数(EIT)和偏差指数(DI)。本研究探讨了某些采集参数、成像解剖结构和制造商对放射图像EI的影响,以及这些因素如何影响EIT设置过程。方法:使用两个不同制造商的数字x线摄影(DR)系统获取图像,使用铝制衰减器和拟人化幻影。采集参数如管电位(kVp)、管负载(mAs)、曝光时间、自动曝光控制(AEC)系统设置(传感器和剂量水平选择)、网格(带或不带)、额外过滤、场大小和成像解剖等都是不同的,并且它们对每个系统的EI的影响分别被量化。结果:EI与探测器上的入射空气kerma (IAK)呈线性相关(根据定义)。当IAK恒定时,EI随kVp的增加而增加。虽然EI通常在存在散射时降低,但情况并非总是如此。在AEC操作下,即使是曝光时间也会产生影响。EI受成像解剖、AEC传感器和场大小选择、检查方案和制造商的影响很大。结论:除IAK外,影响EI计算的参数还有很多。其中,最重要的是成像解剖和制造商。由于EI计算是一个复杂的过程,因此应谨慎设置每个检查和制造商的EIT值,因为适用于一个数字系统的值并不总是适用于另一个数字系统。此外,当EI用作图像质量工具时,在激活可能有意义的红旗之前,应该允许DI至少±2的变化。
{"title":"Exposure index in digital radiography and its dependence on acquisition parameters, anatomy, and manufacturer","authors":"Ioannis A. Tsalafoutas,&nbsp;Shady AlKhazzam,&nbsp;Mohammed Hassan Kharita","doi":"10.1002/acm2.70331","DOIUrl":"10.1002/acm2.70331","url":null,"abstract":"<div>\u0000 \u0000 \u0000 <section>\u0000 \u0000 <h3> PURPOSE</h3>\u0000 \u0000 <p>The exposure index (EI), the target exposure index (EI<sub>T</sub>), and the deviation index (DI) have been defined in the IEC Standard 62494-1 Ed.1 2008-08. This study investigates the impact of certain acquisition parameters, the imaged anatomy, and the manufacturer's specificities on the EI of radiological images and how these may affect EI<sub>T</sub> setting procedure.</p>\u0000 </section>\u0000 \u0000 <section>\u0000 \u0000 <h3> METHODS</h3>\u0000 \u0000 <p>Images were acquired using two digital radiography (DR) systems of two different manufacturers, using aluminum attenuators and an anthropomorphic phantom. Acquisition parameters like the tube potential (kVp), the tube loading (mAs), the exposure time, the automatic exposure control (AEC) system settings (sensor and dose level selection), the grid (with or without), the additional filtration, the field size, and the imaged anatomy were varied and their effect on the EI was quantified separately for each system.</p>\u0000 </section>\u0000 \u0000 <section>\u0000 \u0000 <h3> RESULTS</h3>\u0000 \u0000 <p>EI is linearly related to the incident air kerma (IAK) on the detector as expected (by definition). For constant IAK, EI increases with increasing kVp. While EI in general is reduced in the presence of scatter, this may not always be the case. Under AEC operation, even the exposure time can make a difference. EI is strongly affected by the imaged anatomy in combination with the AEC sensor and field size selections, the examination protocol, and the manufacturer.</p>\u0000 </section>\u0000 \u0000 <section>\u0000 \u0000 <h3> CONCLUSIONS</h3>\u0000 \u0000 <p>Many parameters affect the EI calculation apart from IAK. Among them, the most important are the imaged anatomy and the manufacturer. Since the EI calculation is a complex procedure, setting of the EI<sub>T</sub> values should be done with caution on a per-examination and manufacturer basis, since the values that apply for one digital system are not always applicable to another. Furthermore, when EI is used as an image quality tool, a DI variation of at least ±2 should be allowed before a possibly meaningful red flag is activated.</p>\u0000 </section>\u0000 </div>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":"27 1","pages":""},"PeriodicalIF":2.2,"publicationDate":"2026-01-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12802571/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145965985","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
RapidArc dynamic versus RapidArc: A multi anatomical-site dosimetric evaluation RapidArc动态与RapidArc:多解剖部位剂量学评价。
IF 2.2 4区 医学 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2026-01-14 DOI: 10.1002/acm2.70404
Aram Rostami, Carole Naim, Renilmon Pottanplackal Sukumaran, Mohammad Usman, Abbass Yousef Mkanna, Alla Fuad Al-Sabahi, Ahamed Basith, Shelton Chuck, Mojtaba Barzegar, Bevan Orville Peltier, Bassim Aklan, Samaneh Baradaran, Satheesh Prasad Paloor, Rabih Wafiq Hammoud

Background

RapidArc (RA) has advanced VMAT delivery; however, challenges remain in achieving optimal organ-at-risk (OAR) sparing in complex cases. RapidArc Dynamic (RAD), a next-generation Volumetric Modulation Arc Therapy (VMAT) technique, incorporates dynamic collimator rotation, embedded static fields, and a faster optimizer to enhance dose modulation. This study compares dosimetric outcomes between conventional RA and RAD across four anatomical sites.

Methods

A retrospective analysis was conducted on 40 patients (10 each for breast, lung SBRT, prostate, and head and neck cancers). Each case was initially treated with RA and subsequently replanned using RAD in Eclipse v18.1. Plans were evaluated using dose–volume histograms (DVH), and paired t-tests assessed differences in target coverage and OAR sparing (p < 0.05).

Results

RAD maintained equivalent target coverage compared to RA across all sites. Significant OAR sparing was observed with RAD, including reduced mean heart and contralateral lung doses in breast cases, lower rectal and bladder doses in prostate cases, improved conformity and reduced spinal cord and lung doses in lung SBRT, and superior sparing of parotids, cochleae, and mucosal structures in head and neck cases.

Conclusion

RAD offers superior OAR sparing and improved dose conformity without compromising target coverage compared to conventional RA. These results support RAD's clinical adoption, particularly for anatomically complex or high-precision treatments. Further prospective studies are needed to assess the long-term clinical benefits.

背景:RapidArc (RA)拥有先进的VMAT交付技术;然而,在复杂病例中实现最佳的风险器官(OAR)保留仍然存在挑战。RapidArc Dynamic (RAD)是下一代体积调制电弧治疗(VMAT)技术,结合了动态准直器旋转、嵌入式静态场和更快的优化器来增强剂量调制。本研究比较了常规RA和RAD在四个解剖部位的剂量学结果。方法:对40例患者(乳腺癌、肺癌、前列腺癌、头颈癌各10例)进行回顾性分析。每种情况最初都使用RA进行治疗,随后在Eclipse v18.1中使用RAD进行重新规划。使用剂量-体积直方图(DVH)评估计划,配对t检验评估目标覆盖率和OAR保留的差异(p)。结果:与RA相比,RAD在所有地点保持相同的目标覆盖率。RAD可显著节省OAR,包括乳腺病例降低心脏和对侧肺的平均剂量,前列腺病例降低直肠和膀胱下段剂量,肺SBRT改善一致性,减少脊髓和肺剂量,头颈部病例保留腮腺、耳蜗和粘膜结构。结论:与传统RA相比,RAD提供了更好的OAR保留和改善的剂量一致性,而不影响靶覆盖。这些结果支持RAD的临床应用,特别是在解剖复杂或高精度的治疗中。需要进一步的前瞻性研究来评估长期临床益处。
{"title":"RapidArc dynamic versus RapidArc: A multi anatomical-site dosimetric evaluation","authors":"Aram Rostami,&nbsp;Carole Naim,&nbsp;Renilmon Pottanplackal Sukumaran,&nbsp;Mohammad Usman,&nbsp;Abbass Yousef Mkanna,&nbsp;Alla Fuad Al-Sabahi,&nbsp;Ahamed Basith,&nbsp;Shelton Chuck,&nbsp;Mojtaba Barzegar,&nbsp;Bevan Orville Peltier,&nbsp;Bassim Aklan,&nbsp;Samaneh Baradaran,&nbsp;Satheesh Prasad Paloor,&nbsp;Rabih Wafiq Hammoud","doi":"10.1002/acm2.70404","DOIUrl":"10.1002/acm2.70404","url":null,"abstract":"<div>\u0000 \u0000 \u0000 <section>\u0000 \u0000 <h3> Background</h3>\u0000 \u0000 <p>RapidArc (RA) has advanced VMAT delivery; however, challenges remain in achieving optimal organ-at-risk (OAR) sparing in complex cases. RapidArc Dynamic (RAD), a next-generation Volumetric Modulation Arc Therapy (VMAT) technique, incorporates dynamic collimator rotation, embedded static fields, and a faster optimizer to enhance dose modulation. This study compares dosimetric outcomes between conventional RA and RAD across four anatomical sites.</p>\u0000 </section>\u0000 \u0000 <section>\u0000 \u0000 <h3> Methods</h3>\u0000 \u0000 <p>A retrospective analysis was conducted on 40 patients (10 each for breast, lung SBRT, prostate, and head and neck cancers). Each case was initially treated with RA and subsequently replanned using RAD in Eclipse v18.1. Plans were evaluated using dose–volume histograms (DVH), and paired t-tests assessed differences in target coverage and OAR sparing (<i>p</i> &lt; 0.05).</p>\u0000 </section>\u0000 \u0000 <section>\u0000 \u0000 <h3> Results</h3>\u0000 \u0000 <p>RAD maintained equivalent target coverage compared to RA across all sites. Significant OAR sparing was observed with RAD, including reduced mean heart and contralateral lung doses in breast cases, lower rectal and bladder doses in prostate cases, improved conformity and reduced spinal cord and lung doses in lung SBRT, and superior sparing of parotids, cochleae, and mucosal structures in head and neck cases.</p>\u0000 </section>\u0000 \u0000 <section>\u0000 \u0000 <h3> Conclusion</h3>\u0000 \u0000 <p>RAD offers superior OAR sparing and improved dose conformity without compromising target coverage compared to conventional RA. These results support RAD's clinical adoption, particularly for anatomically complex or high-precision treatments. Further prospective studies are needed to assess the long-term clinical benefits.</p>\u0000 </section>\u0000 </div>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":"27 1","pages":""},"PeriodicalIF":2.2,"publicationDate":"2026-01-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12800903/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145965967","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Influence of scan mode, tilt, and radiation dose on CT radiomic metrics 扫描方式、倾斜和辐射剂量对CT放射学指标的影响。
IF 2.2 4区 医学 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2026-01-14 DOI: 10.1002/acm2.70462
Neha Yadav, Xiaomeng Lei, Steven Y. Cen, Joshua Levy, Kristin Jensen, Bino A. Varghese
<div> <section> <h3> Background</h3> <p>Radiomic features derived from computed tomography (CT) are highly susceptible to variations in acquisition parameters, which can introduce confounding effects in multicenter research and reduce diagnostic accuracy. While the effects of parameters such as scanning mode and dose have been studied, the impact of gantry tilt—despite its routine clinical use—remains underexplored in radiomics literature.</p> </section> <section> <h3> Purpose</h3> <p>To systematically evaluate how scan mode (axial vs. helical), gantry tilt (0° vs. 5°), and radiation dose affect CT-based radiomic metrics using an anthropomorphic liver phantom containing six 3D-printed texture inserts, with special emphasis on the novel inclusion of tilt.</p> </section> <section> <h3> Methods</h3> <p>Twelve unique image acquisition configurations were scanned on a GE Revolution Apex CT scanner, with each configuration repeated once. Manual segmentation of volumes of interest (VOIs) was performed, and 93 radiomic features spanning six texture families were extracted using PyRadiomics. First-order dispersion metrics (standard deviation, interquartile range, and coefficient of variation) were analyzed alongside higher-order features via regression with heatmap visualization, and repeatable, robust, and calibratable features were identified.</p> </section> <section> <h3> Results</h3> <p>Helical scans without tilt generally exhibited lower first-order dispersion than axial scans. Introducing a 5° tilt reduced dispersion in axial scans but had inconsistent effects in helical scans, with no coherent trend observed. Radiation dose demonstrated an expected inverse relationship with dispersion metrics. Intraclass correlation coefficient (ICC) analysis revealed that 34% of radiomic metrics exhibited good or excellent repeatability across all trials (ICC ≥ 0.6), but only 13% demonstrated good or excellent robustness, highlighting the sensitivity of radiomic metrics to scanning conditions. Regression analysis yielded 31 metrics (33%) that can be calibrated using their significant linear relationships with the parameters varied in this study, thereby allowing researchers to correct for variations in acquisition settings.</p> </section> <section> <h3> Conclusions</h3> <p>These findings underscore the importance of accounting for acquisition variability—including less frequently examined parameters such as tilt—when designing radiomic studies, selecting robust feat
背景:来自计算机断层扫描(CT)的放射学特征非常容易受到采集参数变化的影响,这可能在多中心研究中引入混淆效应,降低诊断准确性。虽然扫描模式和剂量等参数的影响已经被研究过,但放射组学文献中对龙门倾斜的影响仍未充分探讨,尽管它是常规的临床应用。目的:系统地评估扫描模式(轴向与螺旋)、支架倾斜(0°与5°)和辐射剂量如何影响基于ct的放射学指标,使用包含六个3d打印纹理插入的拟人化肝脏模型,特别强调倾斜的新包含。方法:在GE Revolution Apex CT扫描仪上扫描12种不同的图像采集配置,每种配置重复一次。对感兴趣体积(VOIs)进行人工分割,并使用PyRadiomics提取了6个纹理族的93个放射学特征。通过热图可视化回归分析一阶离散度指标(标准差、四分位间距和变异系数)和高阶特征,确定了可重复、鲁棒和可校准的特征。结果:无倾斜的螺旋扫描通常比轴向扫描显示更低的一级弥散度。在轴向扫描中引入5°倾斜可以减少色散,但在螺旋扫描中效果不一致,没有观察到一致的趋势。辐射剂量与色散指标表现出预期的反比关系。类内相关系数(ICC)分析显示,34%的放射学指标在所有试验中表现出良好或极好的重复性(ICC≥0.6),但只有13%表现出良好或极好的鲁棒性,突出了放射学指标对扫描条件的敏感性。回归分析产生了31个指标(33%),可以使用它们与本研究中变化的参数之间的显著线性关系进行校准,从而允许研究人员纠正获取设置的变化。结论:这些发现强调了在设计放射学研究、选择可靠特征以及解释临床和多中心研究结果时,考虑获取变异性(包括较少检查的参数,如倾斜)的重要性。这种方法有助于从成像伪影中区分有意义的生物变异,从而提高个性化医学放射组学分析的可靠性。
{"title":"Influence of scan mode, tilt, and radiation dose on CT radiomic metrics","authors":"Neha Yadav,&nbsp;Xiaomeng Lei,&nbsp;Steven Y. Cen,&nbsp;Joshua Levy,&nbsp;Kristin Jensen,&nbsp;Bino A. Varghese","doi":"10.1002/acm2.70462","DOIUrl":"10.1002/acm2.70462","url":null,"abstract":"&lt;div&gt;\u0000 \u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Background&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;Radiomic features derived from computed tomography (CT) are highly susceptible to variations in acquisition parameters, which can introduce confounding effects in multicenter research and reduce diagnostic accuracy. While the effects of parameters such as scanning mode and dose have been studied, the impact of gantry tilt—despite its routine clinical use—remains underexplored in radiomics literature.&lt;/p&gt;\u0000 &lt;/section&gt;\u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Purpose&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;To systematically evaluate how scan mode (axial vs. helical), gantry tilt (0° vs. 5°), and radiation dose affect CT-based radiomic metrics using an anthropomorphic liver phantom containing six 3D-printed texture inserts, with special emphasis on the novel inclusion of tilt.&lt;/p&gt;\u0000 &lt;/section&gt;\u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Methods&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;Twelve unique image acquisition configurations were scanned on a GE Revolution Apex CT scanner, with each configuration repeated once. Manual segmentation of volumes of interest (VOIs) was performed, and 93 radiomic features spanning six texture families were extracted using PyRadiomics. First-order dispersion metrics (standard deviation, interquartile range, and coefficient of variation) were analyzed alongside higher-order features via regression with heatmap visualization, and repeatable, robust, and calibratable features were identified.&lt;/p&gt;\u0000 &lt;/section&gt;\u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Results&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;Helical scans without tilt generally exhibited lower first-order dispersion than axial scans. Introducing a 5° tilt reduced dispersion in axial scans but had inconsistent effects in helical scans, with no coherent trend observed. Radiation dose demonstrated an expected inverse relationship with dispersion metrics. Intraclass correlation coefficient (ICC) analysis revealed that 34% of radiomic metrics exhibited good or excellent repeatability across all trials (ICC ≥ 0.6), but only 13% demonstrated good or excellent robustness, highlighting the sensitivity of radiomic metrics to scanning conditions. Regression analysis yielded 31 metrics (33%) that can be calibrated using their significant linear relationships with the parameters varied in this study, thereby allowing researchers to correct for variations in acquisition settings.&lt;/p&gt;\u0000 &lt;/section&gt;\u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Conclusions&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;These findings underscore the importance of accounting for acquisition variability—including less frequently examined parameters such as tilt—when designing radiomic studies, selecting robust feat","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":"27 1","pages":""},"PeriodicalIF":2.2,"publicationDate":"2026-01-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12800920/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145965969","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Evaluation of hybrid DIR performance using controlling structures and points of interest in MR-guided adaptive radiotherapy for prostate cancer patients 利用控制结构和兴趣点评价磁共振引导下前列腺癌患者适应放疗的混合DIR性能。
IF 2.2 4区 医学 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2026-01-14 DOI: 10.1002/acm2.70437
Victor Malkov, Iymad R. Mansour, Vickie Kong, Winnie Li, Jennifer Dang, Parisa Sadeghi, Inmaculada Navarro, Jerusha Padayachee, Peter Chung, Jeff D. Winter, Tony Tadic
<div> <section> <h3> Background</h3> <p>MR-guided adaptive radiotherapy (ART) allows for daily plan optimization based on patient-specific anatomy. Accumulated doses, driven by deformable image registration (DIR), of daily fractions can provide cumulative dose metrics and insights into toxicity and tumor control. In prostate ART, inter- and intra-factional deformations, particularly due to bladder and rectum, pose a challenge to accurate DIR generation.</p> </section> <section> <h3> Purpose</h3> <p>To quantify geometric and dosimetric accuracy of a proposed prostate MR-to-MR DIR approach to support MR-guided ART dose accumulation.</p> </section> <section> <h3> Methods</h3> <p>We evaluated DIR accuracy in 25 patients treated with 30 Gy in five fractions on a 1.5 T MR-linac using an adaptive workflow. For all patients, a reference MR was used for planning, with three images collected at each fraction: adapt MR for adaptive planning, verify MR for pretreatment position verification and beam-on for capturing anatomy during radiation delivery. We assessed three DIR approaches: intensity-based, intensity-based with controlling structures (CS), and intensity-based with controlling structures and points of interest (CS + P). DIRs were performed between the reference and fraction images and within fractions (adapt-to-verify and adapt-to-beam-on). For the evaluation, we propagated CTV, bladder, and rectum contours using the DIRs and compared each to manually delineated contours using Dice similarity coefficient, mean distance to agreement, and dose–volume metrics.</p> </section> <section> <h3> Results</h3> <p>CS and CS + P improved geometric agreement between manual and propagated contours over intensity-only DIR. For example, mean distance to agreement (DTA<sub>mean</sub>) for reference-to-beam-on intensity-only DIR was 0.131 ± 0.009 cm (CTV), 0.46 ± 0.08 cm (bladder), and 0.154 ± 0.013 cm (rectum). For the CS, the DTA<sub>mean</sub> values were 0.018 ± 0.002, 0.388 ± 0.14, and 0.036 ± 0.013 cm. Finally, for CS + P, these values were 0.015 ± 0.001, 0.025 ± 0.004, and 0.021 ± 0.002 cm. Dosimetrically, comparing CS and CS + P for reference to beam-on DIRs resulted in a change of CTV D98% from [−29 cGy, 19 cGy] to [−18 cGy, 26 cGy], bladder D5cc from [−51 cGy, 544 cGy] to [−79 cGy, 36 cGy], and rectum D1cc from [−106 cGy, 72 cGy] to [−52 cGy, 74 cGy].</p> </section> <section> <h3> Conclusion</h3> <p>CS improved geometric and dosimetric accuracy over intensity-
背景:磁共振引导的适应性放疗(ART)允许基于患者特定解剖结构的日常计划优化。由可变形图像配准(DIR)驱动的每日组分累积剂量可以提供累积剂量指标和对毒性和肿瘤控制的见解。在前列腺ART中,派别间和派别内的变形,特别是由于膀胱和直肠的变形,对准确的DIR生成提出了挑战。目的:量化拟议的前列腺MR-to-MR DIR方法的几何和剂量学准确性,以支持mr引导的ART剂量积累。方法:我们使用自适应工作流程评估25例患者在1.5 T MR-linac上接受5次30 Gy治疗的DIR准确性。对于所有患者,使用参考MR进行规划,每个部分收集三张图像:适应MR进行自适应规划,验证MR进行预处理位置验证,以及在放射传递期间捕获解剖结构。我们评估了三种DIR方法:基于强度、基于控制结构的强度(CS)和基于控制结构和兴趣点的强度(CS + P)。在参考图像和分数图像之间以及分数内(适应-验证和适应-光束)进行dir。为了评估,我们使用DIRs传播CTV,膀胱和直肠轮廓,并使用Dice相似系数,平均一致距离和剂量-体积指标将每个轮廓与手动描绘的轮廓进行比较。结果:CS和CS + P比仅强度DIR改善了手动和传播轮廓之间的几何一致性。例如,参考到光束强度仅DIR的平均一致距离(DTAmean)为0.131±0.009 cm (CTV), 0.46±0.08 cm(膀胱)和0.154±0.013 cm(直肠)。CS的DTAmean值分别为0.018±0.002、0.388±0.14和0.036±0.013 cm。最后,对于CS + P,这些值分别为0.015±0.001、0.025±0.004和0.021±0.002 cm。剂量学上,比较CS和CS + P以参考束上DIRs导致CTV D98%从[-29 cGy, 19 cGy]变化到[-18 cGy, 26 cGy],膀胱D5cc从[-51 cGy, 544 cGy]变化到[-79 cGy, 36 cGy],直肠D1cc从[-106 cGy, 72 cGy]变化到[-52 cGy, 74 cGy]。结论:CS比仅强度的DIR提高了几何和剂量学准确性,CS + P提供了进一步的性能改善,特别是膀胱。然而,会话图像分割仍然是一个挑战,这可以通过自动轮廓来解决。
{"title":"Evaluation of hybrid DIR performance using controlling structures and points of interest in MR-guided adaptive radiotherapy for prostate cancer patients","authors":"Victor Malkov,&nbsp;Iymad R. Mansour,&nbsp;Vickie Kong,&nbsp;Winnie Li,&nbsp;Jennifer Dang,&nbsp;Parisa Sadeghi,&nbsp;Inmaculada Navarro,&nbsp;Jerusha Padayachee,&nbsp;Peter Chung,&nbsp;Jeff D. Winter,&nbsp;Tony Tadic","doi":"10.1002/acm2.70437","DOIUrl":"10.1002/acm2.70437","url":null,"abstract":"&lt;div&gt;\u0000 \u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Background&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;MR-guided adaptive radiotherapy (ART) allows for daily plan optimization based on patient-specific anatomy. Accumulated doses, driven by deformable image registration (DIR), of daily fractions can provide cumulative dose metrics and insights into toxicity and tumor control. In prostate ART, inter- and intra-factional deformations, particularly due to bladder and rectum, pose a challenge to accurate DIR generation.&lt;/p&gt;\u0000 &lt;/section&gt;\u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Purpose&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;To quantify geometric and dosimetric accuracy of a proposed prostate MR-to-MR DIR approach to support MR-guided ART dose accumulation.&lt;/p&gt;\u0000 &lt;/section&gt;\u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Methods&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;We evaluated DIR accuracy in 25 patients treated with 30 Gy in five fractions on a 1.5 T MR-linac using an adaptive workflow. For all patients, a reference MR was used for planning, with three images collected at each fraction: adapt MR for adaptive planning, verify MR for pretreatment position verification and beam-on for capturing anatomy during radiation delivery. We assessed three DIR approaches: intensity-based, intensity-based with controlling structures (CS), and intensity-based with controlling structures and points of interest (CS + P). DIRs were performed between the reference and fraction images and within fractions (adapt-to-verify and adapt-to-beam-on). For the evaluation, we propagated CTV, bladder, and rectum contours using the DIRs and compared each to manually delineated contours using Dice similarity coefficient, mean distance to agreement, and dose–volume metrics.&lt;/p&gt;\u0000 &lt;/section&gt;\u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Results&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;CS and CS + P improved geometric agreement between manual and propagated contours over intensity-only DIR. For example, mean distance to agreement (DTA&lt;sub&gt;mean&lt;/sub&gt;) for reference-to-beam-on intensity-only DIR was 0.131 ± 0.009 cm (CTV), 0.46 ± 0.08 cm (bladder), and 0.154 ± 0.013 cm (rectum). For the CS, the DTA&lt;sub&gt;mean&lt;/sub&gt; values were 0.018 ± 0.002, 0.388 ± 0.14, and 0.036 ± 0.013 cm. Finally, for CS + P, these values were 0.015 ± 0.001, 0.025 ± 0.004, and 0.021 ± 0.002 cm. Dosimetrically, comparing CS and CS + P for reference to beam-on DIRs resulted in a change of CTV D98% from [−29 cGy, 19 cGy] to [−18 cGy, 26 cGy], bladder D5cc from [−51 cGy, 544 cGy] to [−79 cGy, 36 cGy], and rectum D1cc from [−106 cGy, 72 cGy] to [−52 cGy, 74 cGy].&lt;/p&gt;\u0000 &lt;/section&gt;\u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Conclusion&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;CS improved geometric and dosimetric accuracy over intensity-","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":"27 1","pages":""},"PeriodicalIF":2.2,"publicationDate":"2026-01-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12802555/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145965947","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Evaluation of the stray radiation distribution around a mobile cone beam computed tomography system in a simulated operating room environment 模拟手术室环境下移动锥束计算机断层扫描系统杂散辐射分布的评估。
IF 2.2 4区 医学 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2026-01-12 DOI: 10.1002/acm2.70434
A. Kyle Jones
<div> <section> <h3> Background</h3> <p>Radiation protection in the operating room (OR) environment is a subject of much discussion in both the surgery and medical physics communities. Radiation exposure is often infrequent during image-guided procedures, especially when only 3D imaging for navigation is used. This is accompanied by unique personnel considerations, including staff that rotate in and out of the OR and staff that are scrubbed in and do not have the opportunity to easily don and doff radioprotective garments. These communities seek clear guidance about the magnitude of stray radiation dose in the OR environment. However, prior studies have reported conflicting data on the topic and have used different methods and instruments.</p> </section> <section> <h3> Purpose</h3> <p>To systematically measure the magnitude of stray radiation doses in a simulated OR environment in locations relevant to the placement of personnel during image-guided spine surgery and to make recommendations for radiation protection based on these data when using a mobile cone beam computed tomography (CBCT) system and a realistic anthropomorphic phantom on an actual spine surgery table.</p> </section> <section> <h3> Methods</h3> <p>Measurements of stray radiation dose were performed in a grid pattern in a simulated OR environment using pressurized ionization chamber survey meters and two configurations of a tissue-mimicking anthropomorphic phantom, large (L) and extra-large (XL). The phantom was imaged using “navigation” mode (i.e., CBCT) with standard and high definition (HD) protocols. Stray radiation dose was measured at heights corresponding to chest level (125 cm) and eye level (175 cm) of a typical operator and additional heights of 100 and 150 cm.</p> </section> <section> <h3> Results</h3> <p>The absolute per scan whole body dose in the shadow of the gantry console 2 m from isocenter at a height of 125 cm (chest level) was 1.26 µSv for an equal mix of L and XL patients, and at a height of 175 cm (eye lens level) the air kerma was 9.22 µGy. The dose at 2 m from isocenter on the head side of the patient at a height of 125 cm was 7.93 µSv and air kerma at a height of 175 cm was 14.7 µGy. Doses at the same distance from isocenter and same heights on the side of the gantry opposite the console were 13.5 µSv and 15.0 µGy.</p> </section> <section> <h3> Conclusions</h3> <p>Stray radiation doses were lowest in the shadow of the gantry console and were higher at a heig
背景:手术室(OR)环境中的辐射防护是外科和医学物理界讨论较多的主题。在图像引导过程中,辐射暴露通常不常见,特别是当仅使用3D成像进行导航时。这伴随着独特的人员考虑,包括进出手术室的工作人员和被擦洗进去的工作人员,他们没有机会轻易地穿上和脱下辐射防护服。这些社区寻求关于手术室环境中杂散辐射剂量大小的明确指导。然而,先前的研究报告了关于该主题的相互矛盾的数据,并使用了不同的方法和工具。目的:系统地测量图像引导脊柱手术中与人员位置相关的模拟手术室环境中杂散辐射剂量的大小,并在使用移动锥束计算机断层扫描(CBCT)系统和实际脊柱手术台上的逼真拟人化幻影时,根据这些数据提出辐射防护建议。方法:在模拟手术室环境中,使用加压电离室测量仪和大(L)和超大(XL)两种仿组织拟人模型在网格模式下进行杂散辐射剂量测量。采用标准和高清协议的“导航”模式(即CBCT)对幻体进行成像。测量的高度对应于典型操作人员的胸部水平(125厘米)和眼水平(175厘米)以及额外高度为100和150厘米。结果:L和XL患者在距等中心2 m的龙门台阴影下,在125 cm(胸部水平)高度处,每次扫描的全身绝对剂量为1.26µSv,在175 cm(眼球晶状体水平)高度处,空气kerma为9.22µGy。患者头侧距等中心2 m处125 cm高度处的剂量为7.93µSv, 175 cm高度处的风量为14.7µGy。距离等中心相同距离和与控制台相对的龙台一侧相同高度处的剂量分别为13.5µSv和15.0µGy。结论:杂散辐射剂量在龙门台阴影处最低,175 cm处高于125 cm处。根据距离等中心2 m处测量的杂散辐射剂量,可以采用多种辐射防护策略,使手术室人员的职业辐射剂量保持在合理可行的最低水平。
{"title":"Evaluation of the stray radiation distribution around a mobile cone beam computed tomography system in a simulated operating room environment","authors":"A. Kyle Jones","doi":"10.1002/acm2.70434","DOIUrl":"10.1002/acm2.70434","url":null,"abstract":"&lt;div&gt;\u0000 \u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Background&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;Radiation protection in the operating room (OR) environment is a subject of much discussion in both the surgery and medical physics communities. Radiation exposure is often infrequent during image-guided procedures, especially when only 3D imaging for navigation is used. This is accompanied by unique personnel considerations, including staff that rotate in and out of the OR and staff that are scrubbed in and do not have the opportunity to easily don and doff radioprotective garments. These communities seek clear guidance about the magnitude of stray radiation dose in the OR environment. However, prior studies have reported conflicting data on the topic and have used different methods and instruments.&lt;/p&gt;\u0000 &lt;/section&gt;\u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Purpose&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;To systematically measure the magnitude of stray radiation doses in a simulated OR environment in locations relevant to the placement of personnel during image-guided spine surgery and to make recommendations for radiation protection based on these data when using a mobile cone beam computed tomography (CBCT) system and a realistic anthropomorphic phantom on an actual spine surgery table.&lt;/p&gt;\u0000 &lt;/section&gt;\u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Methods&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;Measurements of stray radiation dose were performed in a grid pattern in a simulated OR environment using pressurized ionization chamber survey meters and two configurations of a tissue-mimicking anthropomorphic phantom, large (L) and extra-large (XL). The phantom was imaged using “navigation” mode (i.e., CBCT) with standard and high definition (HD) protocols. Stray radiation dose was measured at heights corresponding to chest level (125 cm) and eye level (175 cm) of a typical operator and additional heights of 100 and 150 cm.&lt;/p&gt;\u0000 &lt;/section&gt;\u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Results&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;The absolute per scan whole body dose in the shadow of the gantry console 2 m from isocenter at a height of 125 cm (chest level) was 1.26 µSv for an equal mix of L and XL patients, and at a height of 175 cm (eye lens level) the air kerma was 9.22 µGy. The dose at 2 m from isocenter on the head side of the patient at a height of 125 cm was 7.93 µSv and air kerma at a height of 175 cm was 14.7 µGy. Doses at the same distance from isocenter and same heights on the side of the gantry opposite the console were 13.5 µSv and 15.0 µGy.&lt;/p&gt;\u0000 &lt;/section&gt;\u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Conclusions&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;Stray radiation doses were lowest in the shadow of the gantry console and were higher at a heig","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":"27 1","pages":""},"PeriodicalIF":2.2,"publicationDate":"2026-01-12","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12796718/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145959490","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Focal spot motion in digital breast tomosynthesis and its effect on spatial resolution 数字乳房断层合成中的焦点运动及其对空间分辨率的影响。
IF 2.2 4区 医学 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2026-01-09 DOI: 10.1002/acm2.70443
Colin J. Schaeffer, Katie W. Hulme, Ashley E. Rubinstein

Background

Digital breast tomosynthesis (DBT) has become standard practice; however, the acquisition method of DBT between vendors is far from standardized. Currently, there are three commercially available DBT tube motion techniques: (1) continuous motion, (2) step-and-shoot, and (3) continuous motion with flying focal spot. Each of these methods represents a trade-off between total acquisition time and focal spot blur.

Purpose

The aim of the study was to characterize the increase in effective focal spot size in DBT relative to standard 2D projections and assess the influence of this increase on spatial resolution using the modulation transfer function (MTF).

Methods

Focal spot size was measured for both a 2D acquisition and the 0° DBT projection using a 10 µm slit phantom. Imaging techniques were set to those used for a 2, 4, and 8 cm thick breast of 50/50 adipose/fat composition. MTF curves were measured using a copper edge phantom both at the breast support plane and 4 cm above the breast support.

Results

The effective focal spot size increase from 2D to DBT increased with breast thickness for all systems. The continuous motion systems showed the greatest increase in effective focal spot size with percent increases of 101% to 462% depending on unit and breast thickness. The flying focal spot system showed the smallest increase in effective focal spot size in DBT acquisitions, being 3%, 21%, and 25% for a 2, 4, and 8 cm thick breast, respectively. The step-and-shoot and flying focal spot systems showed no degradation in MTF curves due to increasing effective focal spot size in DBT acquisitions, while the continuous motion systems showed a reduction in the frequency at which the MTF curve reached 50% of 26%–45%.

Conclusion

Both step-and-shoot and flying focal spot systems minimized effective focal spot size increase in DBT acquisitions compared to continuous tube motion systems.

背景:数字乳腺断层合成(DBT)已成为标准做法;然而,厂商之间DBT的获取方式还远远没有标准化。目前,商业上可获得的DBT管运动技术有三种:(1)连续运动,(2)步进射击,(3)带飞行焦斑的连续运动。这些方法中的每一种都代表了总采集时间和焦斑模糊之间的权衡。目的:本研究的目的是描述DBT相对于标准二维投影的有效焦斑大小的增加,并利用调制传递函数(MTF)评估这种增加对空间分辨率的影响。方法:使用10µm狭缝模体测量二维采集和0°DBT投影的焦点光斑大小。成像技术设置为脂肪/脂肪组成各占50/50的2、4和8 cm厚乳房所使用的成像技术。在乳房支撑平面和乳房支撑上方4cm处使用铜边模体测量MTF曲线。结果:各系统的有效焦斑大小随乳房厚度的增加而增加。连续运动系统的有效焦斑尺寸增加最大,根据单位和乳房厚度的不同,有效焦斑尺寸增加了101%至462%。飞行焦斑系统在DBT采集中显示的有效焦斑大小增加最小,分别为2、4和8 cm厚乳房的3%、21%和25%。步进射击和飞行焦斑系统在DBT采集过程中,由于有效焦斑尺寸的增加,MTF曲线没有下降,而连续运动系统的MTF曲线达到26%-45%的50%的频率下降。结论:与连续管运动系统相比,步进射击和飞行焦斑系统都能最大限度地减少DBT捕获的有效焦斑尺寸增加。
{"title":"Focal spot motion in digital breast tomosynthesis and its effect on spatial resolution","authors":"Colin J. Schaeffer,&nbsp;Katie W. Hulme,&nbsp;Ashley E. Rubinstein","doi":"10.1002/acm2.70443","DOIUrl":"10.1002/acm2.70443","url":null,"abstract":"<div>\u0000 \u0000 \u0000 <section>\u0000 \u0000 <h3> Background</h3>\u0000 \u0000 <p>Digital breast tomosynthesis (DBT) has become standard practice; however, the acquisition method of DBT between vendors is far from standardized. Currently, there are three commercially available DBT tube motion techniques: (1) continuous motion, (2) step-and-shoot, and (3) continuous motion with flying focal spot. Each of these methods represents a trade-off between total acquisition time and focal spot blur.</p>\u0000 </section>\u0000 \u0000 <section>\u0000 \u0000 <h3> Purpose</h3>\u0000 \u0000 <p>The aim of the study was to characterize the increase in effective focal spot size in DBT relative to standard 2D projections and assess the influence of this increase on spatial resolution using the modulation transfer function (MTF).</p>\u0000 </section>\u0000 \u0000 <section>\u0000 \u0000 <h3> Methods</h3>\u0000 \u0000 <p>Focal spot size was measured for both a 2D acquisition and the 0° DBT projection using a 10 µm slit phantom. Imaging techniques were set to those used for a 2, 4, and 8 cm thick breast of 50/50 adipose/fat composition. MTF curves were measured using a copper edge phantom both at the breast support plane and 4 cm above the breast support.</p>\u0000 </section>\u0000 \u0000 <section>\u0000 \u0000 <h3> Results</h3>\u0000 \u0000 <p>The effective focal spot size increase from 2D to DBT increased with breast thickness for all systems. The continuous motion systems showed the greatest increase in effective focal spot size with percent increases of 101% to 462% depending on unit and breast thickness. The flying focal spot system showed the smallest increase in effective focal spot size in DBT acquisitions, being 3%, 21%, and 25% for a 2, 4, and 8 cm thick breast, respectively. The step-and-shoot and flying focal spot systems showed no degradation in MTF curves due to increasing effective focal spot size in DBT acquisitions, while the continuous motion systems showed a reduction in the frequency at which the MTF curve reached 50% of 26%–45%.</p>\u0000 </section>\u0000 \u0000 <section>\u0000 \u0000 <h3> Conclusion</h3>\u0000 \u0000 <p>Both step-and-shoot and flying focal spot systems minimized effective focal spot size increase in DBT acquisitions compared to continuous tube motion systems.</p>\u0000 </section>\u0000 </div>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":"27 1","pages":""},"PeriodicalIF":2.2,"publicationDate":"2026-01-09","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12788966/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145944034","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Ultra-fast dosimetric data collection with a commercial plastic scintillation detector in an MR-linac 超快速剂量学数据收集与商用塑料闪烁探测器在磁共振直线加速器。
IF 2.2 4区 医学 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2026-01-07 DOI: 10.1002/acm2.70460
Carlos Ferrer, Concepción Huertas, Marcos Feijoo, Alessandro Cardin, Moisés Sáez

Background

Plastic scintillation detectors (PSD) are widely used for detecting and measuring ionizing radiation. These detectors are versatile, with high efficiency, fast response and the ability to provide real-time measurements.

Purpose

Evaluate the suitability of Blue Physics PSD (BP-PSD) for performing ultra-fast dosimetric commissioning measurements with high accuracy and precision in a very short time.

Methods

Ultra-fast measurements were performed in water using a BP-PSD on an Elekta Unity MR-linac. Percentage depth doses (PDD) and profiles at different depths were measured at two movement velocities, 10 mm/s and 20 mm/s, for field sizes ranging from 10 × 10 cm2 to 1 × 1 cm2. Gamma analysis was conducted to compare these measurements with those obtained during machine commissioning using a PTW Semiflex 3D ionization chamber (for PDD) and a PTW micro-Diamond detector (for PDD and profiles). Gamma criteria of 2%/2 mm and 1%/1 mm dose difference/distance to agreement were studied, alongside field size, penumbra, and measurement time.

Results

All PDD and profile gamma passing rates were 100% at 2%/2 mm. At the stricter 1%/1 mm criteria, all PDD showed a passing rate above 96.97% for both velocities, with most of the profiles exceeding 95% at 10 mm/s and 90% at 20 mm/s. Gamma analysis results were superior for smaller fields (1 × 1 cm2 and 2 × 2 cm2) and generally better at 10 mm/s. On average, the penumbra measurements obtained with the PSD were greater than those achieved with the micro-Diamond detector. Measurement times were found to be between 7 and 14 times shorter for PDD, and between 5 and 9 times shorter for profiles at speeds of 10 mm/s and 20 mm/s, respectively.

Conclusions

Ultra-fast measurements using the Blue Physics PSD are suitable for acquiring dosimetric commissioning data with high accuracy and precision, and can be performed in a much shorter timeframe than with commonly used detectors.

背景:塑料闪烁探测器(PSD)广泛应用于电离辐射的探测和测量。这些探测器是多功能的,具有高效率,快速响应和提供实时测量的能力。目的:评估Blue Physics PSD (BP-PSD)在极短时间内进行高精度、超快速剂量学调试测量的适用性。方法:在Elekta Unity MR-linac上使用BP-PSD在水中进行超快速测量。在10 × 10 cm2至1 × 1 cm2范围内,以10 mm/s和20 mm/s两种移动速度测量了不同深度的百分比深度剂量(PDD)和剖面。伽马分析将这些测量结果与机器调试期间使用PTW Semiflex 3D电离室(用于PDD)和PTW微型金刚石探测器(用于PDD和轮廓)获得的测量结果进行比较。研究了2%/ 2mm和1%/ 1mm剂量差/一致距离的Gamma标准,以及视场大小、半影和测量时间。结果:所有PDD和剖面图γ在2%/ 2mm时通过率均为100%。在更严格的1%/1 mm标准下,所有PDD对两种速度的通过率均超过96.97%,其中大多数剖面在10 mm/s速度下超过95%,在20 mm/s速度下超过90%。伽玛分析结果在较小的区域(1 × 1 cm2和2 × 2 cm2)较好,通常在10 mm/s时较好。平均而言,使用PSD获得的半影测量值大于使用微金刚石探测器获得的测量值。在速度为10 mm/s和20 mm/s时,PDD的测量时间分别缩短了7到14倍,型材的测量时间缩短了5到9倍。结论:使用Blue Physics PSD的超快速测量适合于获得高精度和精密度的剂量学调试数据,并且可以在比常用检测器更短的时间内完成。
{"title":"Ultra-fast dosimetric data collection with a commercial plastic scintillation detector in an MR-linac","authors":"Carlos Ferrer,&nbsp;Concepción Huertas,&nbsp;Marcos Feijoo,&nbsp;Alessandro Cardin,&nbsp;Moisés Sáez","doi":"10.1002/acm2.70460","DOIUrl":"10.1002/acm2.70460","url":null,"abstract":"<div>\u0000 \u0000 \u0000 <section>\u0000 \u0000 <h3> Background</h3>\u0000 \u0000 <p>Plastic scintillation detectors (PSD) are widely used for detecting and measuring ionizing radiation. These detectors are versatile, with high efficiency, fast response and the ability to provide real-time measurements.</p>\u0000 </section>\u0000 \u0000 <section>\u0000 \u0000 <h3> Purpose</h3>\u0000 \u0000 <p>Evaluate the suitability of Blue Physics PSD (BP-PSD) for performing ultra-fast dosimetric commissioning measurements with high accuracy and precision in a very short time.</p>\u0000 </section>\u0000 \u0000 <section>\u0000 \u0000 <h3> Methods</h3>\u0000 \u0000 <p>Ultra-fast measurements were performed in water using a BP-PSD on an Elekta Unity MR-linac. Percentage depth doses (PDD) and profiles at different depths were measured at two movement velocities, 10 mm/s and 20 mm/s, for field sizes ranging from 10 × 10 cm<sup>2</sup> to 1 × 1 cm<sup>2</sup>. Gamma analysis was conducted to compare these measurements with those obtained during machine commissioning using a PTW Semiflex 3D ionization chamber (for PDD) and a PTW micro-Diamond detector (for PDD and profiles). Gamma criteria of 2%/2 mm and 1%/1 mm dose difference/distance to agreement were studied, alongside field size, penumbra, and measurement time.</p>\u0000 </section>\u0000 \u0000 <section>\u0000 \u0000 <h3> Results</h3>\u0000 \u0000 <p>All PDD and profile gamma passing rates were 100% at 2%/2 mm. At the stricter 1%/1 mm criteria, all PDD showed a passing rate above 96.97% for both velocities, with most of the profiles exceeding 95% at 10 mm/s and 90% at 20 mm/s. Gamma analysis results were superior for smaller fields (1 × 1 cm<sup>2</sup> and 2 × 2 cm<sup>2</sup>) and generally better at 10 mm/s. On average, the penumbra measurements obtained with the PSD were greater than those achieved with the micro-Diamond detector. Measurement times were found to be between 7 and 14 times shorter for PDD, and between 5 and 9 times shorter for profiles at speeds of 10 mm/s and 20 mm/s, respectively.</p>\u0000 </section>\u0000 \u0000 <section>\u0000 \u0000 <h3> Conclusions</h3>\u0000 \u0000 <p>Ultra-fast measurements using the Blue Physics PSD are suitable for acquiring dosimetric commissioning data with high accuracy and precision, and can be performed in a much shorter timeframe than with commonly used detectors.</p>\u0000 </section>\u0000 </div>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":"27 1","pages":""},"PeriodicalIF":2.2,"publicationDate":"2026-01-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12779937/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145917694","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Initial experience with remote MRI scanning support in an oncology focused practice: Opportunities for expanded access to radiology care 在肿瘤学重点实践中获得远程MRI扫描支持的初步经验:扩大获得放射学护理的机会。
IF 2.2 4区 医学 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2026-01-07 DOI: 10.1002/acm2.70461
Christopher M. Walker, Maria G. Maldonado, Megan C. Jacobsen, Suprateek Kundu, Michelle L. Underwood, Joshua P. Yung, Brandy J. Reed, David Jaffray, R. Jason Stafford, Marshall E. Hicks, Caroline Chung, Aradhana M. Venkatesan

Background

As medical imaging demand grows, there is increasing stress on the currently available workforce to deliver consistent, high-quality imaging studies while ensuring rapid study turnaround times and round-the-clock radiology coverage. Advances in remote access technology facilitating remote scan assistance and control are now commercially available to address these pressing clinical needs.

Methods

This work evaluated an early clinical application of a virtual scanner operations system (syngo Virtual Cockpit (VA13A, Siemens Healthineers, Erlangen, Germany) for remote magnetic resonance imaging (MRI) monitoring and scan control at three geographically distant outpatient sites associated with our primary institution.

Results

The system facilitated execution of technically complex oncologic MRI exams at these geographically distant clinics with no measurable impact on acquisition time compared to MR imaging performed at our primary hospital location. Additional operational improvements were realized with the use of the system, including remote staff training, technical assistance, and scanning during staff shortages. This early iteration of remote scanning had some limitations including limited utility for additional assistance in the scanning of those protocols that require complex physical setup. Moreover, connectivity issues were noted to be a limiting factor that contributed to operational delays. It was still necessary to have an onsite MRI technologist at the scanner console to interface with the patient and ensure safe operation.

Conclusion

Despite these limitations, our initial experience demonstrates that the use of remote MRI scanning support facilitates staffing flexibility while providing expanded patient access to oncology MRI services.

背景:随着医学成像需求的增长,在确保快速的研究周转时间和24小时放射学覆盖的同时,现有的工作人员提供一致的、高质量的成像研究的压力越来越大。远程访问技术的进步促进了远程扫描协助和控制,现在可以商业化地解决这些迫切的临床需求。方法:本研究评估了虚拟扫描仪操作系统(syngo虚拟座舱(VA13A, Siemens Healthineers, Erlangen, Germany)在与我们的主要机构相关的三个地理位置遥远的门诊点进行远程磁共振成像(MRI)监测和扫描控制的早期临床应用。结果:该系统促进了在这些地理位置遥远的诊所执行技术复杂的肿瘤MRI检查,与在我们的主要医院位置进行的MR成像相比,对采集时间没有可测量的影响。通过使用该系统,进一步改进了业务,包括远程工作人员培训、技术援助和工作人员短缺期间的扫描。这种远程扫描的早期迭代有一些局限性,包括在扫描那些需要复杂物理设置的协议时提供额外帮助的效用有限。此外,连接问题被认为是导致操作延迟的限制因素。仍然需要有一名现场MRI技术人员在扫描仪控制台与患者沟通并确保操作安全。结论:尽管存在这些限制,我们的初步经验表明,使用远程MRI扫描支持促进了人员配置的灵活性,同时为肿瘤MRI服务提供了更多的患者机会。
{"title":"Initial experience with remote MRI scanning support in an oncology focused practice: Opportunities for expanded access to radiology care","authors":"Christopher M. Walker,&nbsp;Maria G. Maldonado,&nbsp;Megan C. Jacobsen,&nbsp;Suprateek Kundu,&nbsp;Michelle L. Underwood,&nbsp;Joshua P. Yung,&nbsp;Brandy J. Reed,&nbsp;David Jaffray,&nbsp;R. Jason Stafford,&nbsp;Marshall E. Hicks,&nbsp;Caroline Chung,&nbsp;Aradhana M. Venkatesan","doi":"10.1002/acm2.70461","DOIUrl":"10.1002/acm2.70461","url":null,"abstract":"<div>\u0000 \u0000 \u0000 <section>\u0000 \u0000 <h3> Background</h3>\u0000 \u0000 <p>As medical imaging demand grows, there is increasing stress on the currently available workforce to deliver consistent, high-quality imaging studies while ensuring rapid study turnaround times and round-the-clock radiology coverage. Advances in remote access technology facilitating remote scan assistance and control are now commercially available to address these pressing clinical needs.</p>\u0000 </section>\u0000 \u0000 <section>\u0000 \u0000 <h3> Methods</h3>\u0000 \u0000 <p>This work evaluated an early clinical application of a virtual scanner operations system (syngo Virtual Cockpit (VA13A, Siemens Healthineers, Erlangen, Germany) for remote magnetic resonance imaging (MRI) monitoring and scan control at three geographically distant outpatient sites associated with our primary institution.</p>\u0000 </section>\u0000 \u0000 <section>\u0000 \u0000 <h3> Results</h3>\u0000 \u0000 <p>The system facilitated execution of technically complex oncologic MRI exams at these geographically distant clinics with no measurable impact on acquisition time compared to MR imaging performed at our primary hospital location. Additional operational improvements were realized with the use of the system, including remote staff training, technical assistance, and scanning during staff shortages. This early iteration of remote scanning had some limitations including limited utility for additional assistance in the scanning of those protocols that require complex physical setup. Moreover, connectivity issues were noted to be a limiting factor that contributed to operational delays. It was still necessary to have an onsite MRI technologist at the scanner console to interface with the patient and ensure safe operation.</p>\u0000 </section>\u0000 \u0000 <section>\u0000 \u0000 <h3> Conclusion</h3>\u0000 \u0000 <p>Despite these limitations, our initial experience demonstrates that the use of remote MRI scanning support facilitates staffing flexibility while providing expanded patient access to oncology MRI services.</p>\u0000 </section>\u0000 </div>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":"27 1","pages":""},"PeriodicalIF":2.2,"publicationDate":"2026-01-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12779931/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145917682","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Margin reduction and optimal prescription isodose model for liver stereotactic radiotherapy with respiratory motion 肝立体定向放射治疗呼吸运动的切缘缩小及最佳处方等剂量模型。
IF 2.2 4区 医学 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2026-01-07 DOI: 10.1002/acm2.70455
Daisuke Kawahara, Hirokazu Masuda, Takuya Wada, Misato Kishi, Tsuyoshi Katsuta, Yuji Murakami

Purpose

This study aimed to establish a framework for optimizing planning target volume (PTV) margins in liver SBRT using virtual 4DCT (v4DCT), by introducing the concept of the Dosimetric Coverage Amplitude (DCA) to quantify motion tolerance and deriving an Optimal Margin (OM) that balances tumor coverage with normal-tissue sparing.

Methods

VMAT plans were developed using a whole-body phantom with a virtual tumor, ensuring that the prescription dose corresponded to D95% of the PTV. The 60%–80% isodose levels were defined relative to the maximum dose to represent alternative prescription surfaces. v4DCT simulated free-breathing conditions across 10 respiratory phases to generate virtual four-dimensional radiotherapy (v4DRT) dose distributions. The DCA was defined as the maximum respiratory amplitude at which GTV dose coverage (D100% or D99%) was maintained. Based on DCA analysis, the OM was determined as the clinically applicable margin derived from the isodose plan that satisfied the DCA condition and minimized normal tissue dose.

Results

DCA analysis revealed that GTV dose coverage was maintained for respiratory motion amplitudes up to 1.9–2.2 times larger than the conventional PTV margin of 10 mm. Taking into account GTV dose coverage, average GTV dose relative to the 80%-static reference, and normal-tissue dose, the 60% isodose plan was identified as the optimal prescription level. From this plan, the OM was calculated as 5.6 mm, representing a 44% reduction compared with the conventional 10-mm PTV, while maintaining GTV coverage and minimizing liver dose. This margin reduction was attributed to the steeper dose gradient associated with prescribing to a lower isodose level. The v4DCT approach enabled the generation of multiple respiratory phase images, allowing quantitative evaluation of motion-induced dose distribution and optimization of margins based on respiratory variability.

Conclusion

The proposed v4DCT-based framework demonstrated that the 60% isodose plan provided the optimal balance between tumor coverage and normal-tissue sparing, yielding an OM of 5.6 mm (44% reduction compared with the conventional 10-mm PTV). This approach offers a clinically applicable strategy for margin reduction in liver SBRT while maintaining robust dosimetric coverage.

目的:本研究旨在通过引入剂量覆盖幅度(DCA)的概念来量化运动耐受性,并得出平衡肿瘤覆盖与正常组织保留的最佳边界(OM),建立一个使用虚拟4DCT (v4DCT)优化肝脏SBRT规划靶体积(PTV)边界的框架。方法:采用带虚拟肿瘤的全身假体制定VMAT计划,确保处方剂量符合PTV的D95%。60%-80%等剂量水平被定义为相对于最大剂量来代表可选择的处方表面。v4DCT模拟10个呼吸期的自由呼吸条件,生成虚拟四维放疗(v4DRT)剂量分布。DCA定义为维持GTV剂量覆盖(D100%或D99%)时的最大呼吸振幅。根据DCA分析,确定OM为满足DCA条件和最小正常组织剂量的等剂量方案得出的临床适用边界。结果:DCA分析显示,GTV对呼吸运动幅度的剂量覆盖范围比常规PTV的10 mm大1.9-2.2倍。综合考虑GTV剂量覆盖率、相对于80%静态参考的平均GTV剂量和正常组织剂量,确定60%等剂量计划为最佳处方水平。根据该方案,OM计算为5.6 mm,与传统的10 mm PTV相比减少了44%,同时保持GTV覆盖范围并使肝脏剂量最小化。这种边际减少归因于与较低等剂量水平处方相关的更陡峭的剂量梯度。v4DCT方法能够生成多个呼吸期图像,允许定量评估运动引起的剂量分布和基于呼吸变异性的边缘优化。结论:提出的基于v4dct的框架表明,60%等剂量计划提供了肿瘤覆盖和正常组织保留之间的最佳平衡,产生5.6 mm的OM(与传统的10 mm PTV相比减少44%)。该方法提供了一种临床适用的策略,可以在保持稳健剂量覆盖的同时减少肝脏SBRT的边缘。
{"title":"Margin reduction and optimal prescription isodose model for liver stereotactic radiotherapy with respiratory motion","authors":"Daisuke Kawahara,&nbsp;Hirokazu Masuda,&nbsp;Takuya Wada,&nbsp;Misato Kishi,&nbsp;Tsuyoshi Katsuta,&nbsp;Yuji Murakami","doi":"10.1002/acm2.70455","DOIUrl":"10.1002/acm2.70455","url":null,"abstract":"<div>\u0000 \u0000 \u0000 <section>\u0000 \u0000 <h3> Purpose</h3>\u0000 \u0000 <p>This study aimed to establish a framework for optimizing planning target volume (PTV) margins in liver SBRT using virtual 4DCT (v4DCT), by introducing the concept of the Dosimetric Coverage Amplitude (DCA) to quantify motion tolerance and deriving an Optimal Margin (OM) that balances tumor coverage with normal-tissue sparing.</p>\u0000 </section>\u0000 \u0000 <section>\u0000 \u0000 <h3> Methods</h3>\u0000 \u0000 <p>VMAT plans were developed using a whole-body phantom with a virtual tumor, ensuring that the prescription dose corresponded to D<sub>95%</sub> of the PTV. The 60%–80% isodose levels were defined relative to the maximum dose to represent alternative prescription surfaces. v4DCT simulated free-breathing conditions across 10 respiratory phases to generate virtual four-dimensional radiotherapy (v4DRT) dose distributions. The DCA was defined as the maximum respiratory amplitude at which GTV dose coverage (D100% or D99%) was maintained. Based on DCA analysis, the OM was determined as the clinically applicable margin derived from the isodose plan that satisfied the DCA condition and minimized normal tissue dose.</p>\u0000 </section>\u0000 \u0000 <section>\u0000 \u0000 <h3> Results</h3>\u0000 \u0000 <p>DCA analysis revealed that GTV dose coverage was maintained for respiratory motion amplitudes up to 1.9–2.2 times larger than the conventional PTV margin of 10 mm. Taking into account GTV dose coverage, average GTV dose relative to the 80%-static reference, and normal-tissue dose, the 60% isodose plan was identified as the optimal prescription level. From this plan, the OM was calculated as 5.6 mm, representing a 44% reduction compared with the conventional 10-mm PTV, while maintaining GTV coverage and minimizing liver dose. This margin reduction was attributed to the steeper dose gradient associated with prescribing to a lower isodose level. The v4DCT approach enabled the generation of multiple respiratory phase images, allowing quantitative evaluation of motion-induced dose distribution and optimization of margins based on respiratory variability.</p>\u0000 </section>\u0000 \u0000 <section>\u0000 \u0000 <h3> Conclusion</h3>\u0000 \u0000 <p>The proposed v4DCT-based framework demonstrated that the 60% isodose plan provided the optimal balance between tumor coverage and normal-tissue sparing, yielding an OM of 5.6 mm (44% reduction compared with the conventional 10-mm PTV). This approach offers a clinically applicable strategy for margin reduction in liver SBRT while maintaining robust dosimetric coverage.</p>\u0000 </section>\u0000 </div>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":"27 1","pages":""},"PeriodicalIF":2.2,"publicationDate":"2026-01-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12779934/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145917653","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Clinical evaluation of the golden beam data in the monaco treatment planning system for the harmony pro and infinity linear accelerators harmony pro和无限线性加速器摩纳哥治疗计划系统中金束数据的临床评价。
IF 2.2 4区 医学 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2026-01-05 DOI: 10.1002/acm2.70457
Wenchao Gao, Min Zhang, Shuchao Zhang, Jie Qi, Xin Wang, Liang Zhao, Ping Liu, Zhijun Yang, HongXiang Cheng
<div> <section> <h3> Background</h3> <p>Accurate dose calculations in radiotherapy depend on high-quality beam models in the Monaco treatment planning system (TPS). The accelerated go live (AGL) workflow, using a golden beam data (GBD) model, has improved beam modeling accuracy for various linear accelerators (linacs). However, similar studies specifically for the Harmony Pro, recently introduced for online adaptive radiotherapy, have not yet been reported internationally. Moreover, studies on dosimetric differences between TPS beam models with GBD and those with measured beam data (MBD) are limited, and no such studies have been published specifically for Elekta linacs.</p> </section> <section> <h3> Purpose</h3> <p>This study aimed to assess the clinical performance of GBD in the Monaco TPS for the harmony pro and infinity linacs.</p> </section> <section> <h3> Methods</h3> <p>Beam tuning and data collection were performed on the harmony pro and infinity linacs based on GBD. Subsequently, percentage depth doses (PDDs), off-axis dose profiles, output factors (OFs), absolute doses, and test fields were measured to evaluate the GBD model. Additionally, 31 clinical plans from multiple anatomical sites, including 17 conventional fractionated radiotherapy (CFRT) plans and 14 stereotactic body radiotherapy (SBRT) plans, were designed using the Monaco TPS (GBD model) and practically tested. An Infinity linac with MBD was introduced as a control.</p> </section> <section> <h3> Results</h3> <p>PDDs and profiles on both GBD linacs showed 100% passing rate (2%/2 mm). OFs and absolute doses on both GBD linacs agreed within ±1% and ±1.5%, respectively. Additionally, verification of the test fields yielded passing rates above 98% (2%/2 mm) for both GBD linacs. Furthermore, for CFRT plans, measurements on three linacs achieved a passing rate above 95% (3%/2 mm). The absolute dose deviations were within 3%, whereas one MBD linac case exceeded 3% (-3.73%). For SBRT plans, the gamma passing rates were 98.18 ± 1.58%, 98.76 ± 1.54%, and 94.72 ± 0.04% (3%/2 mm) and 96.69 ± 1.96, 96.29 ± 2.26, and 89.51 ± 0.06% (2%/2 mm), for the two GDB linacs and the MDB linac, respectively. The absolute dose deviations were within 3%, whereas two MBD linac cases exceeded 3% (−3.51%, −4.50%).</p> </section> <section> <h3> Conclusions</h3> <p>The harmony pro-GBD and infinity-GBD linac results demonstrated strong agreement with GBD. Clinical plans designed with Monaco TPS (GBD model) were
背景:精确的放射剂量计算依赖于摩纳哥治疗计划系统(TPS)中高质量的束流模型。采用金束数据(GBD)模型的加速上线(AGL)工作流,提高了各种直线加速器(linacs)的光束建模精度。然而,专门针对Harmony Pro的类似研究,最近用于在线适应性放疗,尚未在国际上报道。此外,关于具有GBD的TPS光束模型与具有测量光束数据(MBD)的TPS光束模型之间剂量学差异的研究是有限的,并且没有专门针对Elekta linacs的此类研究发表。目的:本研究旨在评估摩纳哥TPS对和谐线和无限线的GBD的临床表现。方法:对基于GBD的和谐直线机和无限直线机进行光束调谐和数据采集。随后,测量了百分比深度剂量(pdd)、离轴剂量曲线、输出因子(OFs)、绝对剂量和试验场,以评估GBD模型。此外,采用Monaco TPS (GBD模型)设计了31个多解剖部位的临床方案,包括17个常规分割放疗(CFRT)方案和14个立体定向体放疗(SBRT)方案,并进行了实际检验。引入了一个带MBD的无限直线加速器作为控制。结果:两种GBD直线的pdd和剖面图通过率均为100% (2%/ 2mm)。两种GBD直线的OFs和绝对剂量分别在±1%和±1.5%范围内一致。此外,测试现场的验证表明,两种GBD直线的合格率均超过98% (2%/ 2mm)。此外,对于CFRT方案,在三个直线机上的测量达到了95%以上的通过率(3%/ 2mm)。绝对剂量偏差在3%以内,而1例MBD直线病例超过3%(-3.73%)。对于SBRT方案,两种GDB直线器和MDB直线器的伽玛通良率分别为98.18±1.58%、98.76±1.54%和94.72±0.04% (3%/2 mm)和96.69±1.96、96.29±2.26和89.51±0.06% (2%/2 mm)。绝对剂量偏差在3%以内,2例MBD直线患者剂量偏差超过3%(-3.51%,-4.50%)。结论:和谐的亲GBD和无限的GBD线性结果与GBD有很强的一致性。采用摩纳哥TPS (GBD模型)设计的临床方案在两种GBD直线上均可接受。尽管Infinity-MBD直线上提供的GBD模型方案的测试结果与两种GBD直线上的测试结果存在一定差异,但大多数方案仍然可以接受。这表明摩纳哥TPS中基于gbd的建模在不同直线类型中提供了可靠的临床表现。
{"title":"Clinical evaluation of the golden beam data in the monaco treatment planning system for the harmony pro and infinity linear accelerators","authors":"Wenchao Gao,&nbsp;Min Zhang,&nbsp;Shuchao Zhang,&nbsp;Jie Qi,&nbsp;Xin Wang,&nbsp;Liang Zhao,&nbsp;Ping Liu,&nbsp;Zhijun Yang,&nbsp;HongXiang Cheng","doi":"10.1002/acm2.70457","DOIUrl":"10.1002/acm2.70457","url":null,"abstract":"&lt;div&gt;\u0000 \u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Background&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;Accurate dose calculations in radiotherapy depend on high-quality beam models in the Monaco treatment planning system (TPS). The accelerated go live (AGL) workflow, using a golden beam data (GBD) model, has improved beam modeling accuracy for various linear accelerators (linacs). However, similar studies specifically for the Harmony Pro, recently introduced for online adaptive radiotherapy, have not yet been reported internationally. Moreover, studies on dosimetric differences between TPS beam models with GBD and those with measured beam data (MBD) are limited, and no such studies have been published specifically for Elekta linacs.&lt;/p&gt;\u0000 &lt;/section&gt;\u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Purpose&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;This study aimed to assess the clinical performance of GBD in the Monaco TPS for the harmony pro and infinity linacs.&lt;/p&gt;\u0000 &lt;/section&gt;\u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Methods&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;Beam tuning and data collection were performed on the harmony pro and infinity linacs based on GBD. Subsequently, percentage depth doses (PDDs), off-axis dose profiles, output factors (OFs), absolute doses, and test fields were measured to evaluate the GBD model. Additionally, 31 clinical plans from multiple anatomical sites, including 17 conventional fractionated radiotherapy (CFRT) plans and 14 stereotactic body radiotherapy (SBRT) plans, were designed using the Monaco TPS (GBD model) and practically tested. An Infinity linac with MBD was introduced as a control.&lt;/p&gt;\u0000 &lt;/section&gt;\u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Results&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;PDDs and profiles on both GBD linacs showed 100% passing rate (2%/2 mm). OFs and absolute doses on both GBD linacs agreed within ±1% and ±1.5%, respectively. Additionally, verification of the test fields yielded passing rates above 98% (2%/2 mm) for both GBD linacs. Furthermore, for CFRT plans, measurements on three linacs achieved a passing rate above 95% (3%/2 mm). The absolute dose deviations were within 3%, whereas one MBD linac case exceeded 3% (-3.73%). For SBRT plans, the gamma passing rates were 98.18 ± 1.58%, 98.76 ± 1.54%, and 94.72 ± 0.04% (3%/2 mm) and 96.69 ± 1.96, 96.29 ± 2.26, and 89.51 ± 0.06% (2%/2 mm), for the two GDB linacs and the MDB linac, respectively. The absolute dose deviations were within 3%, whereas two MBD linac cases exceeded 3% (−3.51%, −4.50%).&lt;/p&gt;\u0000 &lt;/section&gt;\u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Conclusions&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;The harmony pro-GBD and infinity-GBD linac results demonstrated strong agreement with GBD. Clinical plans designed with Monaco TPS (GBD model) were","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":"27 1","pages":""},"PeriodicalIF":2.2,"publicationDate":"2026-01-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12766699/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145900539","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
期刊
Journal of Applied Clinical Medical Physics
全部 Acc. Chem. Res. ACS Applied Bio Materials ACS Appl. Electron. Mater. ACS Appl. Energy Mater. ACS Appl. Mater. Interfaces ACS Appl. Nano Mater. ACS Appl. Polym. Mater. ACS BIOMATER-SCI ENG ACS Catal. ACS Cent. Sci. ACS Chem. Biol. ACS Chemical Health & Safety ACS Chem. Neurosci. ACS Comb. Sci. ACS Earth Space Chem. ACS Energy Lett. ACS Infect. Dis. ACS Macro Lett. ACS Mater. Lett. ACS Med. Chem. Lett. ACS Nano ACS Omega ACS Photonics ACS Sens. ACS Sustainable Chem. Eng. ACS Synth. Biol. Anal. Chem. BIOCHEMISTRY-US Bioconjugate Chem. BIOMACROMOLECULES Chem. Res. Toxicol. Chem. Rev. Chem. Mater. CRYST GROWTH DES ENERG FUEL Environ. Sci. Technol. Environ. Sci. Technol. Lett. Eur. J. Inorg. Chem. IND ENG CHEM RES Inorg. Chem. J. Agric. Food. Chem. J. Chem. Eng. Data J. Chem. Educ. J. Chem. Inf. Model. J. Chem. Theory Comput. J. Med. Chem. J. Nat. Prod. J PROTEOME RES J. Am. Chem. Soc. LANGMUIR MACROMOLECULES Mol. Pharmaceutics Nano Lett. Org. Lett. ORG PROCESS RES DEV ORGANOMETALLICS J. Org. Chem. J. Phys. Chem. J. Phys. Chem. A J. Phys. Chem. B J. Phys. Chem. C J. Phys. Chem. Lett. Analyst Anal. Methods Biomater. Sci. Catal. Sci. Technol. Chem. Commun. Chem. Soc. Rev. CHEM EDUC RES PRACT CRYSTENGCOMM Dalton Trans. Energy Environ. Sci. ENVIRON SCI-NANO ENVIRON SCI-PROC IMP ENVIRON SCI-WAT RES Faraday Discuss. Food Funct. Green Chem. Inorg. Chem. Front. Integr. Biol. J. Anal. At. Spectrom. J. Mater. Chem. A J. Mater. Chem. B J. Mater. Chem. C Lab Chip Mater. Chem. Front. Mater. Horiz. MEDCHEMCOMM Metallomics Mol. Biosyst. Mol. Syst. Des. Eng. Nanoscale Nanoscale Horiz. Nat. Prod. Rep. New J. Chem. Org. Biomol. Chem. Org. Chem. Front. PHOTOCH PHOTOBIO SCI PCCP Polym. Chem.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1