首页 > 最新文献

Clinical chemistry and laboratory medicine最新文献

英文 中文
Measurement of lipoprotein(a) via a novel point of care approach with comparison to established laboratory assays. 测量脂蛋白(a)通过一个新的点护理方法与建立实验室分析比较。
IF 3.7 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2026-01-06 DOI: 10.1515/cclm-2025-1605
Caroline A E Bachmeier, Greg J Ward, Andrew J Kassianos, Alex Dechavez, Chantelle Ebersohn, Andrew Liu, Karam M Kostner

Objectives: Lipoprotein(a) is an atherogenic particle causative of atherosclerotic cardiovascular disease. Novel treatments have been developed that lower lipoprotein(a) to unprecedented levels with cardiovascular outcomes trials ongoing. Many guidelines recommend testing once in the lifetime of everyone, but testing rates remain low. In this study we compare a lipoprotein(a) point of care testing device to laboratory analysers and assess its performance.

Methods: Lipoprotein(a) concentrations on residual patient samples measured on the Randox and Roche assays were compared to a novel point of care device, iProtin. Furthermore, assessment of iProtin performance characteristics were performed, including intra- and inter-assay coefficient of variation and dilutional studies.

Results: Lipoprotein(a) concentrations measured on the Randox and Roche assays showed strong correlation with iProtin. Regression analysis using Passing-Bablock showed the best fits for iProtin based on 58 serum samples were: 1.15 × Randox + 7.28 nmol/L and 1.02 × Roche + 17.54 nmol/L. The R2 values for Randox/iProtin and Roche/iProtin were 0.906 and 0.912 respectively. Correlation between Roche and Randox showed Roche=1.15 × Randox - 13.33 nmol/L with an R2 value of 0.973. Inter-assay coefficient of variation of the iProtin device showed a day-to-day imprecision over 5 days of 15.5 % (low concentration quality control) and 6.2 % (high concentration quality control). Within day imprecision was 13.2 % (lower concentration patient sample) and 14.3 % (higher concentration patient sample).

Conclusions: Point of care testing could be a complimentary option to laboratory testing of lipoprotein(a), especially in remote areas. It may help (re-)stratify cardiovascular risk and help tailor treatment decisions.

目的:脂蛋白(a)是导致动脉粥样硬化性心血管疾病的致动脉粥样硬化颗粒。新的治疗方法已经开发出来,将脂蛋白(a)降低到前所未有的水平,心血管结局试验正在进行中。许多指南建议每个人一生中检测一次,但检测率仍然很低。在这项研究中,我们比较脂蛋白(a)点护理测试设备实验室分析仪和评估其性能。方法:将Randox和Roche检测的残余患者样本上的脂蛋白(a)浓度与一种新的护理点设备iprotein进行比较。此外,还进行了iprotein性能特征的评估,包括测定内和测定间的变异系数和稀释研究。结果:Randox和Roche测定的脂蛋白(a)浓度与iprotein有很强的相关性。通过pass - bablock回归分析,58份血清样本iprotein的最佳拟合值分别为:1.15 × Randox + 7.28 nmol/L和1.02 × Roche + 17.54 nmol/L。Randox/iProtin和Roche/iProtin的R2分别为0.906和0.912。Roche与Randox的相关性为Roche=1.15 × Randox - 13.33 nmol/L, R2值为0.973。iprotein装置的测定间变异系数在5天内每天的不精确性为15.5 %(低浓度质量控制)和6.2 %(高浓度质量控制)。日内不精确性为13.2 %(低浓度患者样本)和14.3 %(高浓度患者样本)。结论:护理点检测可作为实验室检测脂蛋白(a)的补充选择,特别是在偏远地区。它可能有助于(重新)分层心血管风险,并有助于制定治疗方案。
{"title":"Measurement of lipoprotein(a) via a novel point of care approach with comparison to established laboratory assays.","authors":"Caroline A E Bachmeier, Greg J Ward, Andrew J Kassianos, Alex Dechavez, Chantelle Ebersohn, Andrew Liu, Karam M Kostner","doi":"10.1515/cclm-2025-1605","DOIUrl":"https://doi.org/10.1515/cclm-2025-1605","url":null,"abstract":"<p><strong>Objectives: </strong>Lipoprotein(a) is an atherogenic particle causative of atherosclerotic cardiovascular disease. Novel treatments have been developed that lower lipoprotein(a) to unprecedented levels with cardiovascular outcomes trials ongoing. Many guidelines recommend testing once in the lifetime of everyone, but testing rates remain low. In this study we compare a lipoprotein(a) point of care testing device to laboratory analysers and assess its performance.</p><p><strong>Methods: </strong>Lipoprotein(a) concentrations on residual patient samples measured on the Randox and Roche assays were compared to a novel point of care device, iProtin. Furthermore, assessment of iProtin performance characteristics were performed, including intra- and inter-assay coefficient of variation and dilutional studies.</p><p><strong>Results: </strong>Lipoprotein(a) concentrations measured on the Randox and Roche assays showed strong correlation with iProtin. Regression analysis using Passing-Bablock showed the best fits for iProtin based on 58 serum samples were: 1.15 × Randox + 7.28 nmol/L and 1.02 × Roche + 17.54 nmol/L. The R<sup>2</sup> values for Randox/iProtin and Roche/iProtin were 0.906 and 0.912 respectively. Correlation between Roche and Randox showed Roche=1.15 × Randox - 13.33 nmol/L with an R<sup>2</sup> value of 0.973. Inter-assay coefficient of variation of the iProtin device showed a day-to-day imprecision over 5 days of 15.5 % (low concentration quality control) and 6.2 % (high concentration quality control). Within day imprecision was 13.2 % (lower concentration patient sample) and 14.3 % (higher concentration patient sample).</p><p><strong>Conclusions: </strong>Point of care testing could be a complimentary option to laboratory testing of lipoprotein(a), especially in remote areas. It may help (re-)stratify cardiovascular risk and help tailor treatment decisions.</p>","PeriodicalId":10390,"journal":{"name":"Clinical chemistry and laboratory medicine","volume":" ","pages":""},"PeriodicalIF":3.7,"publicationDate":"2026-01-06","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145899400","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Assessment of drone transport for biological samples: a real-world experience at a tertiary hospital. 生物样本的无人机运输评估:在三级医院的实际经验。
IF 3.7 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2026-01-06 DOI: 10.1515/cclm-2025-1205
Paula Calomarde-Pastor, Álvaro Piedra-Aguilera, Arnau Pulido-Gracia, Carla Fernández-Prendes, Marta Álvarez-Álvarez, Silvia Martínez-Couselo, María Martínez-Bujidos, María Antonia Huertas-Contreras, María Carmen García-Martín, Laura Jiménez-Añón, Jennifer Rodríguez-Domínguez, Vanessa García-Bayarri, Berta Llebot-Casajuana, Núria Abdón-Giménez, Núria Sánchez-Mercader, Sara Iglesias-Lorente, Yolanda García-González, Ignacio Blanco-Guillermo, Cristian Morales-Indiano, Susana Malumbres-Serrano

Objectives: Our aim is to determine whether drone transport introduces relevant variability in analytical results and to assess its suitability for integration into routine hospital laboratory operations.

Methods: A two-phase study was conducted. The stability phase involved three sets of samples from 31 healthy volunteers transported by drone, by road, and processed without transport. The pilot phase included paired samples from 104 primary care patients transported by drone and by road. A broad panel of 53 biochemical, hematological and urinary analytes was assessed. Analytical performance was evaluated using mean bias, comparison with reference change values (RCVs) and hemolysis grading.

Results: Most analytes showed minimal differences between transport methods. In the stability study, glucose and MCH in road-transported samples, and lactate dehydrogenase (LDH) and mean corpuscular hemoglobin (MCH) in drone-transported samples, exceeded their respective RCVs, with LDH likely affected by mild hemolysis. In the pilot study, statistically significant differences were observed between transport methods for potassium, exhibiting a bias of -1.7 % (-2.6 to -0.8 %). Hemolysis was slightly more frequent in drone samples during the stability study but comparable in the pilot study.

Conclusions: Drone transport preserved the analytical integrity of a comprehensive test panel and has the potential to reduce logistical constraints. These findings support its implementation as a reliable and sustainable alternative to conventional transport in clinical laboratory settings.

目的:我们的目的是确定无人机运输是否会在分析结果中引入相关的可变性,并评估其整合到常规医院实验室操作中的适用性。方法:采用两期研究。稳定阶段包括来自31名健康志愿者的三组样本,分别通过无人机、公路运输和无运输处理。试点阶段包括通过无人机和公路运输的104名初级保健患者的配对样本。对53项生化、血液学和尿液分析进行了评估。采用平均偏倚、参考变化值(rcv)比较和溶血分级来评估分析效果。结果:大多数分析显示运输方法之间的差异很小。在稳定性研究中,公路运输样品的葡萄糖和MCH以及无人机运输样品的乳酸脱氢酶(LDH)和平均红细胞血红蛋白(MCH)均超过了各自的rcv, LDH可能受到轻度溶血的影响。在初步研究中,观察到钾转运方式之间的统计学显著差异,偏差为-1.7 %(-2.6至- 0.8%)。在稳定性研究期间,溶血在无人机样本中稍微更频繁,但在试点研究中是相似的。结论:无人机运输保留了综合测试小组的分析完整性,并有可能减少后勤限制。这些发现支持其作为一种可靠和可持续的替代传统运输在临床实验室设置的实施。
{"title":"Assessment of drone transport for biological samples: a real-world experience at a tertiary hospital.","authors":"Paula Calomarde-Pastor, Álvaro Piedra-Aguilera, Arnau Pulido-Gracia, Carla Fernández-Prendes, Marta Álvarez-Álvarez, Silvia Martínez-Couselo, María Martínez-Bujidos, María Antonia Huertas-Contreras, María Carmen García-Martín, Laura Jiménez-Añón, Jennifer Rodríguez-Domínguez, Vanessa García-Bayarri, Berta Llebot-Casajuana, Núria Abdón-Giménez, Núria Sánchez-Mercader, Sara Iglesias-Lorente, Yolanda García-González, Ignacio Blanco-Guillermo, Cristian Morales-Indiano, Susana Malumbres-Serrano","doi":"10.1515/cclm-2025-1205","DOIUrl":"10.1515/cclm-2025-1205","url":null,"abstract":"<p><strong>Objectives: </strong>Our aim is to determine whether drone transport introduces relevant variability in analytical results and to assess its suitability for integration into routine hospital laboratory operations.</p><p><strong>Methods: </strong>A two-phase study was conducted. The stability phase involved three sets of samples from 31 healthy volunteers transported by drone, by road, and processed without transport. The pilot phase included paired samples from 104 primary care patients transported by drone and by road. A broad panel of 53 biochemical, hematological and urinary analytes was assessed. Analytical performance was evaluated using mean bias, comparison with reference change values (RCVs) and hemolysis grading.</p><p><strong>Results: </strong>Most analytes showed minimal differences between transport methods. In the stability study, glucose and MCH in road-transported samples, and lactate dehydrogenase (LDH) and mean corpuscular hemoglobin (MCH) in drone-transported samples, exceeded their respective RCVs, with LDH likely affected by mild hemolysis. In the pilot study, statistically significant differences were observed between transport methods for potassium, exhibiting a bias of -1.7 % (-2.6 to -0.8 %). Hemolysis was slightly more frequent in drone samples during the stability study but comparable in the pilot study.</p><p><strong>Conclusions: </strong>Drone transport preserved the analytical integrity of a comprehensive test panel and has the potential to reduce logistical constraints. These findings support its implementation as a reliable and sustainable alternative to conventional transport in clinical laboratory settings.</p>","PeriodicalId":10390,"journal":{"name":"Clinical chemistry and laboratory medicine","volume":" ","pages":""},"PeriodicalIF":3.7,"publicationDate":"2026-01-06","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145899392","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
From venipuncture to self-sampling dried blood spots: a shift in monitoring testosterone levels. 从静脉穿刺到自我取样干血点:监测睾酮水平的转变。
IF 3.7 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2026-01-05 DOI: 10.1515/cclm-2025-1287
Anouk Olthof, Jacquelien J Hillebrand, Lysette H A van Deursen-Nagel, Anita Boelen, Andreas Meißner, Annemieke C Heijboer

Objectives: Dried blood spot (DBS) sampling enables convenient at-home blood collection, suitable for patients needing regular check-ups, such as those receiving testosterone replacement therapy (TRT). This study assessed the feasibility and reliability of DBS sampling for monitoring testosterone concentrations in patients undergoing TRT, aiming to provide a convenient at-home alternative to venipuncture.

Methods: Twenty healthy volunteers (12F/8M) and 56 male patients (n=59 samples) receiving TRT were included. Blood was collected parallel by venipuncture (reference) and patient-performed finger prick (DBS) following instruction materials and additional verbal guidance from the andrology consultant. Feasibility and patients' experience were evaluated through questionnaires. Testosterone concentrations in both serum and DBS were quantified using liquid chromatography-tandem mass spectrometry (LC-MS/MS).

Results: In healthy volunteers, serum and DBS testosterone concentrations were highly correlated (y=-0.021 + 0.66x; r=0.99). For patients receiving intramuscular testosterone undecanoate, this correlation was strong (y=1.3 + 0.54x; r=0.87). However, DBS samples from patients using testosterone gel showed extremely elevated testosterone levels compared to serum, likely due to local contamination. Most healthy volunteers and in general all patients preferred at-home DBS-sampling over venipuncture at the hospital when regular check-ups are necessary, citing convenience and reduced travel time.

Conclusions: At-home DBS sampling is a feasible and reliable method for monitoring testosterone in patients treated with intramuscular testosterone undecanoate, improving patient comfort and convenience. However, DBS sampling at-home is currently unsuitable for patients using testosterone gel, likely due to contamination issues. Comprehensive patient education on collection of DBS samples is crucial to ensure high-quality DBS and accurate testosterone quantification.

目的:干血斑(DBS)取样可以方便地在家中采集血液,适用于需要定期检查的患者,例如接受睾酮替代疗法(TRT)的患者。本研究评估了DBS取样监测TRT患者睾酮浓度的可行性和可靠性,旨在提供一种方便的在家替代静脉穿刺的方法。方法:选取健康志愿者20例(12F/8M)和接受TRT治疗的男性患者56例(n=59例)。通过静脉穿刺(参考)和患者手指穿刺(DBS)平行采集血液,并遵循指导材料和男科顾问的额外口头指导。通过问卷调查的方式评价其可行性及患者体验。采用液相色谱-串联质谱法(LC-MS/MS)定量测定血清和DBS中的睾酮浓度。结果:在健康志愿者中,血清和DBS睾酮浓度高度相关(y=-0.021 + 0.66x; r=0.99)。对于肌肉注射十一酸睾酮的患者,这种相关性很强(y=1.3 + 0.54x; r=0.87)。然而,使用睾酮凝胶的患者的DBS样本显示,与血清相比,睾酮水平异常升高,可能是由于局部污染。当需要定期检查时,大多数健康志愿者和所有患者都更喜欢在家进行dbs采样,而不是在医院进行静脉穿刺,理由是方便和减少了旅行时间。结论:在家DBS取样是肌注十一酸睾酮患者睾酮监测的一种可行、可靠的方法,提高了患者的舒适度和便捷性。然而,可能由于污染问题,目前在家进行DBS采样不适合使用睾酮凝胶的患者。全面的患者教育收集DBS样本是至关重要的,以确保高质量的DBS和准确的睾酮定量。
{"title":"From venipuncture to self-sampling dried blood spots: a shift in monitoring testosterone levels.","authors":"Anouk Olthof, Jacquelien J Hillebrand, Lysette H A van Deursen-Nagel, Anita Boelen, Andreas Meißner, Annemieke C Heijboer","doi":"10.1515/cclm-2025-1287","DOIUrl":"https://doi.org/10.1515/cclm-2025-1287","url":null,"abstract":"<p><strong>Objectives: </strong>Dried blood spot (DBS) sampling enables convenient at-home blood collection, suitable for patients needing regular check-ups, such as those receiving testosterone replacement therapy (TRT). This study assessed the feasibility and reliability of DBS sampling for monitoring testosterone concentrations in patients undergoing TRT, aiming to provide a convenient at-home alternative to venipuncture.</p><p><strong>Methods: </strong>Twenty healthy volunteers (12F/8M) and 56 male patients (n=59 samples) receiving TRT were included. Blood was collected parallel by venipuncture (reference) and patient-performed finger prick (DBS) following instruction materials and additional verbal guidance from the andrology consultant. Feasibility and patients' experience were evaluated through questionnaires. Testosterone concentrations in both serum and DBS were quantified using liquid chromatography-tandem mass spectrometry (LC-MS/MS).</p><p><strong>Results: </strong>In healthy volunteers, serum and DBS testosterone concentrations were highly correlated (y=-0.021 + 0.66x; r=0.99). For patients receiving intramuscular testosterone undecanoate, this correlation was strong (y=1.3 + 0.54x; r=0.87). However, DBS samples from patients using testosterone gel showed extremely elevated testosterone levels compared to serum, likely due to local contamination. Most healthy volunteers and in general all patients preferred at-home DBS-sampling over venipuncture at the hospital when regular check-ups are necessary, citing convenience and reduced travel time.</p><p><strong>Conclusions: </strong>At-home DBS sampling is a feasible and reliable method for monitoring testosterone in patients treated with intramuscular testosterone undecanoate, improving patient comfort and convenience. However, DBS sampling at-home is currently unsuitable for patients using testosterone gel, likely due to contamination issues. Comprehensive patient education on collection of DBS samples is crucial to ensure high-quality DBS and accurate testosterone quantification.</p>","PeriodicalId":10390,"journal":{"name":"Clinical chemistry and laboratory medicine","volume":" ","pages":""},"PeriodicalIF":3.7,"publicationDate":"2026-01-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145892299","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Laboratory findings in patients treated with complement factor C3 inhibitor pegcetacoplan. 补体因子C3抑制剂pegcetacoplan治疗患者的实验室结果。
IF 3.7 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2026-01-05 DOI: 10.1515/cclm-2025-1270
Saskia M C Langemeijer, Jeroen D Langereis, Mendy Ter Avest, Caroline Duineveld, Anne-Els van de Logt, Corrie M De Kat Angelino, Amber van Deelen, Marloes A H M Michels, Jolein Gloerich, Evelien G G Sprenkeler, Renate G van der Molen, Nicole C A J van de Kar, Joannes F M Jacobs

Objectives: Complement inhibitor pegcetacoplan binds to C3 and its activation product C3b. Pegcetacoplan has been approved for the treatment of paroxysmal nocturnal hemoglobinuria. Because pegcetacoplan exerts broad inhibition of the complement cascade its efficacy is also investigated in numerous other diseases caused by complement dysregulation, such as C3 glomerulopathy. Pegcetacoplan causes a number of counterintuitive changes in laboratory results.

Methods: In-depth complement analysis in two patients with PNH and three patients with C3 glomerulopathy, all treated with pegcetacoplan.

Results: C3 levels increase up to 300 % above reference levels. In vitro testing showed that this is not a turbidimetric artifact in the C3 immunoassay due to pegcetacoplan-C3 complex formation but appears to be caused by increased half-life of C3 bound to pegcetacoplan. Unbiased mass spectrometric plasma proteome analysis confirmed the dramatic pegcetacoplan-induced increase in circulating C3. Surprisingly, also a three-fold increase of properdin was observed during pegcetacoplan treatment. Serum protein electrophoresis showed an additional band in all patients after pegcetacoplan exposure. This C3-band does not migrate at its expected position because of changes in the mass and charge of C3 bound to pegcetacoplan and should therefore not be misinterpreted as an M-protein. Both in vitro experiments and real clinical practice laboratory results demonstrated that pegcetacoplan completely blocked the alternative complement pathway while the classical pathway is affected but remains largely intact.

Conclusions: Because of the increasing use of pegcetacoplan in routine clinical practice, it is important that both clinicians and laboratory specialists are aware of these unexpected therapy-induced laboratory findings.

目的:补体抑制剂pegcetacoplan与C3及其激活产物C3b结合。Pegcetacoplan已被批准用于治疗阵发性夜间血红蛋白尿。由于pegcetacoplan对补体级联具有广泛的抑制作用,其在补体失调引起的许多其他疾病(如C3肾小球病)中的疗效也被研究。Pegcetacoplan在实验室结果中引起了一些违反直觉的变化。方法:对2例PNH患者和3例C3型肾小球病变患者进行深入补体分析。结果:C3水平比参考水平高出300 %。体外测试表明,这不是C3免疫测定中由于pegcetacoplan-C3复合物形成的浊度伪影,而是由于C3与pegcetacoplan结合的半衰期增加引起的。无偏质谱血浆蛋白质组分析证实了pegcetacoplan引起的循环C3的急剧增加。令人惊讶的是,在pegcetacoplan治疗期间,也观察到properdin增加了三倍。血清蛋白电泳显示,所有患者暴露于pegcetacoplan后都有一个额外的条带。由于与pegcetacoplan结合的C3的质量和电荷的变化,该C3带不会迁移到预期的位置,因此不应被误解为m蛋白。体外实验和实际临床实验结果均表明,pegcetacoplan完全阻断了替代补体途径,而经典途径受到影响,但基本保持完整。结论:由于在常规临床实践中越来越多地使用pegcetacoplan,临床医生和实验室专家都必须意识到这些意想不到的治疗引起的实验室结果。
{"title":"Laboratory findings in patients treated with complement factor C3 inhibitor pegcetacoplan.","authors":"Saskia M C Langemeijer, Jeroen D Langereis, Mendy Ter Avest, Caroline Duineveld, Anne-Els van de Logt, Corrie M De Kat Angelino, Amber van Deelen, Marloes A H M Michels, Jolein Gloerich, Evelien G G Sprenkeler, Renate G van der Molen, Nicole C A J van de Kar, Joannes F M Jacobs","doi":"10.1515/cclm-2025-1270","DOIUrl":"10.1515/cclm-2025-1270","url":null,"abstract":"<p><strong>Objectives: </strong>Complement inhibitor pegcetacoplan binds to C3 and its activation product C3b. Pegcetacoplan has been approved for the treatment of paroxysmal nocturnal hemoglobinuria. Because pegcetacoplan exerts broad inhibition of the complement cascade its efficacy is also investigated in numerous other diseases caused by complement dysregulation, such as C3 glomerulopathy. Pegcetacoplan causes a number of counterintuitive changes in laboratory results.</p><p><strong>Methods: </strong>In-depth complement analysis in two patients with PNH and three patients with C3 glomerulopathy, all treated with pegcetacoplan.</p><p><strong>Results: </strong>C3 levels increase up to 300 % above reference levels. <i>In vitro</i> testing showed that this is not a turbidimetric artifact in the C3 immunoassay due to pegcetacoplan-C3 complex formation but appears to be caused by increased half-life of C3 bound to pegcetacoplan. Unbiased mass spectrometric plasma proteome analysis confirmed the dramatic pegcetacoplan-induced increase in circulating C3. Surprisingly, also a three-fold increase of properdin was observed during pegcetacoplan treatment. Serum protein electrophoresis showed an additional band in all patients after pegcetacoplan exposure. This C3-band does not migrate at its expected position because of changes in the mass and charge of C3 bound to pegcetacoplan and should therefore not be misinterpreted as an M-protein. Both <i>in vitro</i> experiments and real clinical practice laboratory results demonstrated that pegcetacoplan completely blocked the alternative complement pathway while the classical pathway is affected but remains largely intact.</p><p><strong>Conclusions: </strong>Because of the increasing use of pegcetacoplan in routine clinical practice, it is important that both clinicians and laboratory specialists are aware of these unexpected therapy-induced laboratory findings.</p>","PeriodicalId":10390,"journal":{"name":"Clinical chemistry and laboratory medicine","volume":" ","pages":""},"PeriodicalIF":3.7,"publicationDate":"2026-01-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145888794","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Validation study of the Helena V8 UltraCE capillary elecrophoresis analyser. Helena V8 UltraCE毛细管电泳分析仪的验证研究。
IF 3.7 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2026-01-05 DOI: 10.1515/cclm-2025-1523
Joris R Delanghe, Maaike Godefroid, Thomas Maenhout

Objectives: Serum protein electrophoresis is one of the keystone investigations for screening for monoclonal proteins and assessing the serum protein profile. The newly released Helena V8ultra capillary electrophoresis system has been evaluated.

Methods: In total, 164 serum samples were assessed on the Helena V8 UltraCE system and compared with the Sebia Capillarys™ instrument. Abnormalities suggestive of monoclonal proteins were confirmed by immunosubtraction. Imprecision studies intervals were determined. Special attention was paid to complement C3 polymorphism.

Results: The imprecision of the Helena V8ultra was inferior or equal from 0.73 % (albumin) to 4.32 % (alpha2 globulin). The mean bias of Helena V8ultraCE vs. Sebia Capillarys was about -0.225 % for albumin; -0.791 % for alpha1 globulins; 1.353 % for alpha2 globulins; -2.317 % for beta globulins; 0.066 % for gamma globulins. Among the 6 samples with monoclonal proteins confirmed by immunofixation, all were seen on both methods, with only 1 discordant result. The theoretical plate number (albumin fraction) was 2,371 ± 718 (Helena V8UltraCE) vs. 445 ± 115 (Sebia Capillarys 2). The high resolution also allows to distinguish Complement C3 phenotypes. CV values for the various fractions were low, ranging from 0.47 % (albumin) to 0.89 % (gamma globulins). IgG, IgA and IgM M-proteins could be detected in the electropherogram with an excellent sensitivity (<0.1 g/L).

Conclusions: Our evaluation confirms the good analytical performance of the Helena V8 analyzer as a suitable alternative to the Sebia Capillarys instrument. The high resolution allows detailed analysis of individual protein fractions which is an excellent basis for studying microheterogeneity.

目的:血清蛋白电泳是筛选单克隆蛋白和评价血清蛋白谱的重要手段之一。对新推出的Helena v8超毛细管电泳系统进行了评价。方法:164份血清样本在Helena V8 UltraCE系统上进行评估,并与Sebia Capillarys™仪器进行比较。单克隆蛋白提示异常经免疫减影证实。确定了不精确研究间隔。特别注意补体C3多态性。结果:Helena V8ultra在0.73 %(白蛋白)和4.32 % (α - 2球蛋白)之间的不精确性较差或相同。Helena V8ultraCE与Sebia capillys的白蛋白平均偏差约为-0.225 %;α 1球蛋白-0.791 %;α 2球蛋白1.353 %;-2.317 %的β球蛋白;0.066 %为γ球蛋白。免疫固定法证实单克隆蛋白的6份样品,两种方法均可观察到,仅有1份结果不一致。理论板数(白蛋白分数)分别为2371±718 (Helena V8UltraCE)和445±115 (Sebia capillys 2)。高分辨率也允许区分补体C3表型。各组分的CV值都很低,范围从0.47 %(白蛋白)到0.89 % (γ球蛋白)。IgG、IgA和IgM m蛋白均可在电泳图中检测到,灵敏度极好(结论:我们的评估证实了Helena V8分析仪具有良好的分析性能,是Sebia capillys仪器的合适替代品。高分辨率允许对单个蛋白质组分进行详细分析,这是研究微观异质性的良好基础。
{"title":"Validation study of the Helena V8 UltraCE capillary elecrophoresis analyser.","authors":"Joris R Delanghe, Maaike Godefroid, Thomas Maenhout","doi":"10.1515/cclm-2025-1523","DOIUrl":"10.1515/cclm-2025-1523","url":null,"abstract":"<p><strong>Objectives: </strong>Serum protein electrophoresis is one of the keystone investigations for screening for monoclonal proteins and assessing the serum protein profile. The newly released Helena V8ultra capillary electrophoresis system has been evaluated.</p><p><strong>Methods: </strong>In total, 164 serum samples were assessed on the Helena V8 UltraCE system and compared with the Sebia Capillarys™ instrument. Abnormalities suggestive of monoclonal proteins were confirmed by immunosubtraction. Imprecision studies intervals were determined. Special attention was paid to complement C3 polymorphism.</p><p><strong>Results: </strong>The imprecision of the Helena V8ultra was inferior or equal from 0.73 % (albumin) to 4.32 % (alpha<sub>2</sub> globulin). The mean bias of Helena V8ultraCE vs. Sebia Capillarys was about -0.225 % for albumin; -0.791 % for alpha<sub>1</sub> globulins; 1.353 % for alpha<sub>2</sub> globulins; -2.317 % for beta globulins; 0.066 % for gamma globulins. Among the 6 samples with monoclonal proteins confirmed by immunofixation, all were seen on both methods, with only 1 discordant result. The theoretical plate number (albumin fraction) was 2,371 ± 718 (Helena V8UltraCE) vs. 445 ± 115 (Sebia Capillarys 2). The high resolution also allows to distinguish Complement C3 phenotypes. CV values for the various fractions were low, ranging from 0.47 % (albumin) to 0.89 % (gamma globulins). IgG, IgA and IgM M-proteins could be detected in the electropherogram with an excellent sensitivity (<0.1 g/L).</p><p><strong>Conclusions: </strong>Our evaluation confirms the good analytical performance of the Helena V8 analyzer as a suitable alternative to the Sebia Capillarys instrument. The high resolution allows detailed analysis of individual protein fractions which is an excellent basis for studying microheterogeneity.</p>","PeriodicalId":10390,"journal":{"name":"Clinical chemistry and laboratory medicine","volume":" ","pages":""},"PeriodicalIF":3.7,"publicationDate":"2026-01-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145892302","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Analytical and diagnostic evaluation of the Anvajo fluidlab 2 analyzer: a novel urine particle analyzer for clinical application using digital holographic microscopy? Anvajo fluidlab 2分析仪的分析和诊断评价:一种用于临床应用的新型数字全息显微镜尿液颗粒分析仪?
IF 3.7 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2026-01-02 DOI: 10.1515/cclm-2025-1202
Sigrid Deprez, Marijn Speeckaert, Joris Delanghe, Matthijs Oyaert

Objectives: The performance of a novel urine particle analyzer, fluidlab 2 (Anvajo GmbH, Dresden, Germany), was evaluated against phase-contrast visual microscopy according to the most recent EFLM European Urinalysis Guideline.

Methods: The fluidlab 2 device combines digital holographic microscopy with neural network-based object detection for particle classification. Its compact benchtop design is suitable for bedside use, reducing turnaround times. The analytical performance (imprecision, linearity, LoQ) was evaluated according to the 2023 EFLM Urinalysis Guideline. Method comparison involved the analysis of 450 urine samples, assessing RBC, WBC, and SEC counts against visual microscopy using Passing-Bablok regression and Spearman's correlation. Bland-Altman plots were used to evaluate the agreement with clinical performance standards, while weighted Cohen's kappa was used to measure diagnostic agreement on an ordinal scale.

Results: By applying Dahlberg's procedure, a desirable relative coefficient of variation R(CV) ≤2.0 was obtained for RBC and WBC. Linearity of up to 7 × 106/L and 6 × 106/L was achieved. The estimated LoQ at CV=30 % reached 20 × 106/L for RBC and 5 × 106/L for WBC. Spearman's correlation coefficient against visual microscopy was 0.86, 0.92 and 0.94 for RBC, WBC and SEC, respectively. Agreement with visual microscopy (Cohen's weighted kappa) was 0.92 for RBC, 0.93 for WBC, 0.96 for SEC, 0.86 for casts, 0.82 for non-SEC, 0.33 for crystals and 0.51 for bacterial counts.

Conclusions: Fluidlab 2 provides desirable imprecision for RBC and WBC, and meets the criteria for linearity and LoQ. Cohen's weighted kappa coefficients show an optimal comparison to visual microscopy for RBC, WBC and SEC and a minimum comparison for casts and non-SEC. This evaluation demonstrated promising results for the use of the fluidlab 2 analyzer in a clinical setting to detect kidney-related diseases based on urine particle analysis.

目的:根据最新的EFLM欧洲尿液分析指南,对新型尿液颗粒分析仪fluidlab 2 (Anvajo GmbH,德累斯顿,德国)的性能进行相对比视觉显微镜评估。方法:fluidlab 2装置将数字全息显微镜与基于神经网络的物体检测相结合,进行颗粒分类。其紧凑的台式设计适合床边使用,减少周转时间。根据2023 EFLM尿液分析指南评估分析性能(不精密度、线性度、定量限)。方法比较包括对450份尿液样本进行分析,使用Passing-Bablok回归和Spearman相关对视觉显微镜评估RBC、WBC和SEC计数。Bland-Altman图用于评估与临床表现标准的一致性,而加权Cohen’s kappa用于在顺序量表上衡量诊断一致性。结果:应用Dahlberg方法,RBC和WBC的相对变异系数R(CV)≤2.0。线性度可达7 × 106/L和6 × 106/L。在CV=30 %时,RBC的LoQ为20 × 106/L, WBC为5 × 106/L。RBC、WBC和SEC的Spearman相关系数分别为0.86、0.92和0.94。与目视显微镜(Cohen’s加权kappa)的一致性RBC为0.92,WBC为0.93,SEC为0.96,铸型为0.86,非SEC为0.82,晶体为0.33,细菌计数为0.51。结论:Fluidlab 2提供了理想的红细胞和白细胞不精密度,并满足线性和定量限标准。Cohen的加权kappa系数显示,与肉眼显微镜相比,RBC、WBC和SEC的比较效果最佳,而与cast和非SEC的比较效果最低。该评估显示了在临床环境中使用fluidlab 2分析仪检测基于尿液颗粒分析的肾脏相关疾病的有希望的结果。
{"title":"Analytical and diagnostic evaluation of the Anvajo fluidlab 2 analyzer: a novel urine particle analyzer for clinical application using digital holographic microscopy?","authors":"Sigrid Deprez, Marijn Speeckaert, Joris Delanghe, Matthijs Oyaert","doi":"10.1515/cclm-2025-1202","DOIUrl":"https://doi.org/10.1515/cclm-2025-1202","url":null,"abstract":"<p><strong>Objectives: </strong>The performance of a novel urine particle analyzer, fluidlab 2 (Anvajo GmbH, Dresden, Germany), was evaluated against phase-contrast visual microscopy according to the most recent EFLM European Urinalysis Guideline.</p><p><strong>Methods: </strong>The fluidlab 2 device combines digital holographic microscopy with neural network-based object detection for particle classification. Its compact benchtop design is suitable for bedside use, reducing turnaround times. The analytical performance (imprecision, linearity, LoQ) was evaluated according to the 2023 EFLM Urinalysis Guideline. Method comparison involved the analysis of 450 urine samples, assessing RBC, WBC, and SEC counts against visual microscopy using Passing-Bablok regression and Spearman's correlation. Bland-Altman plots were used to evaluate the agreement with clinical performance standards, while weighted Cohen's kappa was used to measure diagnostic agreement on an ordinal scale.</p><p><strong>Results: </strong>By applying Dahlberg's procedure, a desirable relative coefficient of variation R(CV) ≤2.0 was obtained for RBC and WBC. Linearity of up to 7 × 10<sup>6</sup>/L and 6 × 10<sup>6</sup>/L was achieved. The estimated LoQ at CV=30 % reached 20 × 10<sup>6</sup>/L for RBC and 5 × 10<sup>6</sup>/L for WBC. Spearman's correlation coefficient against visual microscopy was 0.86, 0.92 and 0.94 for RBC, WBC and SEC, respectively. Agreement with visual microscopy (Cohen's weighted kappa) was 0.92 for RBC, 0.93 for WBC, 0.96 for SEC, 0.86 for casts, 0.82 for non-SEC, 0.33 for crystals and 0.51 for bacterial counts.</p><p><strong>Conclusions: </strong>Fluidlab 2 provides desirable imprecision for RBC and WBC, and meets the criteria for linearity and LoQ. Cohen's weighted kappa coefficients show an optimal comparison to visual microscopy for RBC, WBC and SEC and a minimum comparison for casts and non-SEC. This evaluation demonstrated promising results for the use of the fluidlab 2 analyzer in a clinical setting to detect kidney-related diseases based on urine particle analysis.</p>","PeriodicalId":10390,"journal":{"name":"Clinical chemistry and laboratory medicine","volume":" ","pages":""},"PeriodicalIF":3.7,"publicationDate":"2026-01-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145888781","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
From automation to agentic artificial intelligence in laboratory medicine: an opinion of the IFCC Division on Emerging Technologies. 从自动化到实验室医学中的人工智能:IFCC新兴技术司的意见。
IF 3.7 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2025-12-29 DOI: 10.1515/cclm-2025-1314
Damien Gruson, Bernard Gouget, Woochang Lee, Ronda Greaves, Yan Liu, Sven Ebert, He Sarina Yang, Swarup Shah

Agentic artificial intelligence (AI) systems are distinguished by their ability to invoke multiple tools, compose command chains, and combine chain-of-thought reasoning with deep research to execute complex tasks and take actions. This represents a major evolution beyond machine learning and large language models (LLM), toward systems capable of planning, executing, and coordinating complex workflows. In contrast to traditional LLMs, which primarily focus on generating and classifying information, agentic AI introduces elements of autonomy, reasoning, and orchestration, while digital twins extend this concept to dynamic virtual representations of patients and laboratory processes, capable of continuous learning, simulation and adaptation. This transition has profound implications for laboratory medicine, a field characterized by high data complexity, multi-omics integration, and stringent operational demands. At the same time, laboratories face growing expectations regarding efficiency, resource stewardship, and value-based healthcare delivery. This article explores both the opportunities and limitations of agentic AI in laboratory medicine, highlighting its potential to move beyond static automation toward autonomous, outcome-driven innovation. It also examines the ethical, interpretability, and governance considerations that must accompany its implementation.

人工智能(AI)系统的特点是能够调用多个工具,组成命令链,并将思维链推理与深度研究相结合,以执行复杂的任务并采取行动。这代表了超越机器学习和大型语言模型(LLM),向能够规划、执行和协调复杂工作流的系统的主要演变。传统的法学硕士主要专注于生成和分类信息,与之相反,人工智能引入了自主、推理和编排的元素,而数字孪生则将这一概念扩展到患者和实验室过程的动态虚拟表示,能够持续学习、模拟和适应。这种转变对以高数据复杂性、多组学集成和严格操作要求为特征的检验医学领域具有深远的影响。与此同时,实验室在效率、资源管理和基于价值的医疗保健服务方面面临着日益增长的期望。本文探讨了人工智能在实验室医学中的机遇和局限性,强调了其超越静态自动化走向自主、结果驱动创新的潜力。它还检查了必须伴随其实现的道德、可解释性和治理方面的考虑。
{"title":"From automation to agentic artificial intelligence in laboratory medicine: an opinion of the IFCC Division on Emerging Technologies.","authors":"Damien Gruson, Bernard Gouget, Woochang Lee, Ronda Greaves, Yan Liu, Sven Ebert, He Sarina Yang, Swarup Shah","doi":"10.1515/cclm-2025-1314","DOIUrl":"https://doi.org/10.1515/cclm-2025-1314","url":null,"abstract":"<p><p>Agentic artificial intelligence (AI) systems are distinguished by their ability to invoke multiple tools, compose command chains, and combine chain-of-thought reasoning with deep research to execute complex tasks and take actions. This represents a major evolution beyond machine learning and large language models (LLM), toward systems capable of planning, executing, and coordinating complex workflows. In contrast to traditional LLMs, which primarily focus on generating and classifying information, agentic AI introduces elements of autonomy, reasoning, and orchestration, while digital twins extend this concept to dynamic virtual representations of patients and laboratory processes, capable of continuous learning, simulation and adaptation. This transition has profound implications for laboratory medicine, a field characterized by high data complexity, multi-omics integration, and stringent operational demands. At the same time, laboratories face growing expectations regarding efficiency, resource stewardship, and value-based healthcare delivery. This article explores both the opportunities and limitations of agentic AI in laboratory medicine, highlighting its potential to move beyond static automation toward autonomous, outcome-driven innovation. It also examines the ethical, interpretability, and governance considerations that must accompany its implementation.</p>","PeriodicalId":10390,"journal":{"name":"Clinical chemistry and laboratory medicine","volume":" ","pages":""},"PeriodicalIF":3.7,"publicationDate":"2025-12-29","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145846436","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Kappa Index predicts disease activity and transition to high-efficacy therapies in multiple sclerosis. Kappa指数预测多发性硬化症的疾病活动性和向高效疗法的过渡。
IF 3.7 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2025-12-29 DOI: 10.1515/cclm-2025-1339
Simone Marcheselli, Federica Galota, Patrizia Natali, Francesco Corrado, Alessia Fiore, Francesca Vitetta, Krzysztof Smolik, Giulia De Napoli, Martina Cardi, Diana Ferraro

Objectives: The Kappa Index has proven its diagnostic value for multiple sclerosis (MS), while its prognostic potential remains to be fully explored. The objective of this study is thus to investigate the value of the Kappa Index at disease onset in predicting disease activity and high-efficacy therapy (HET) initiation.

Methods: We enrolled MS patients with available Kappa Index values at disease onset and a follow-up of at least two years. Primary outcome was the time to loss of NEDA3 (no evidence of disease activity-3) defined as the absence of relapses, MRI activity, and disability progression. Secondary outcome was the time to HET initiation.

Results: Of 120 enrolled patients (36 M, 84 F, mean age: 35 ± 11 years), NEDA3 loss occurred in 89 (74 %) by the end of the follow-up period. A total of 98 (82 %) initiated a moderate efficacy therapy (MET); of these, 34 (28 %) transitioned to a HET during follow-up. Kappa Index values above the maximally selected log-rank statistic-derived cut-off of 38 were independent risk factors for NEDA3 loss (HR 1.75, 95 % CI: 1.09-2.80, p=0.021) and HET initiation (3.25, 95 % CI: 1.54-6.87, p=0.002) and also independently predicted HET following MET failure (2.54, 95 % CI: 1.17-5.51, p=0.018).

Conclusions: Elevated Kappa Index values at diagnosis predict disease activity, MET failure and HET initiation and may be a valuable adjunctive tool in identifying patients in need of prompt HET initiation.

目的:Kappa指数对多发性硬化症(MS)的诊断价值已被证实,但其预后潜力仍有待充分挖掘。因此,本研究的目的是探讨Kappa指数在疾病发病时预测疾病活动性和高效治疗(HET)开始的价值。方法:我们招募了发病时Kappa指数可用的MS患者,随访至少2年。主要终点是NEDA3消失的时间(无疾病活动证据-3),定义为无复发、MRI活动和残疾进展。次要终点是到HET开始的时间。结果:在120例入组患者(36例男性,84例女性,平均年龄:35±11岁)中,89例(74 %)在随访结束时发生NEDA3丢失。共有98人(82% %)开始了中等疗效治疗(MET);其中34例(28% %)在随访期间转变为HET。Kappa指数值高于最大选择的log-rank统计推导的截止值38,是NEDA3丢失(HR 1.75, 95 % CI: 1.09-2.80, p=0.021)和HET起始(3.25,95 % CI: 1.54-6.87, p=0.002)的独立危险因素,也是MET失败后HET的独立预测因素(2.54,95 % CI: 1.17-5.51, p=0.018)。结论:Kappa指数在诊断时的升高可预测疾病活动性、MET失败和HET启动,可能是识别需要及时启动HET的患者的有价值的辅助工具。
{"title":"Kappa Index predicts disease activity and transition to high-efficacy therapies in multiple sclerosis.","authors":"Simone Marcheselli, Federica Galota, Patrizia Natali, Francesco Corrado, Alessia Fiore, Francesca Vitetta, Krzysztof Smolik, Giulia De Napoli, Martina Cardi, Diana Ferraro","doi":"10.1515/cclm-2025-1339","DOIUrl":"https://doi.org/10.1515/cclm-2025-1339","url":null,"abstract":"<p><strong>Objectives: </strong>The Kappa Index has proven its diagnostic value for multiple sclerosis (MS), while its prognostic potential remains to be fully explored. The objective of this study is thus to investigate the value of the Kappa Index at disease onset in predicting disease activity and high-efficacy therapy (HET) initiation.</p><p><strong>Methods: </strong>We enrolled MS patients with available Kappa Index values at disease onset and a follow-up of at least two years. Primary outcome was the time to loss of NEDA3 (no evidence of disease activity-3) defined as the absence of relapses, MRI activity, and disability progression. Secondary outcome was the time to HET initiation.</p><p><strong>Results: </strong>Of 120 enrolled patients (36 M, 84 F, mean age: 35 ± 11 years), NEDA3 loss occurred in 89 (74 %) by the end of the follow-up period. A total of 98 (82 %) initiated a moderate efficacy therapy (MET); of these, 34 (28 %) transitioned to a HET during follow-up. Kappa Index values above the maximally selected log-rank statistic-derived cut-off of 38 were independent risk factors for NEDA3 loss (HR 1.75, 95 % CI: 1.09-2.80, p=0.021) and HET initiation (3.25, 95 % CI: 1.54-6.87, p=0.002) and also independently predicted HET following MET failure (2.54, 95 % CI: 1.17-5.51, p=0.018).</p><p><strong>Conclusions: </strong>Elevated Kappa Index values at diagnosis predict disease activity, MET failure and HET initiation and may be a valuable adjunctive tool in identifying patients in need of prompt HET initiation.</p>","PeriodicalId":10390,"journal":{"name":"Clinical chemistry and laboratory medicine","volume":" ","pages":""},"PeriodicalIF":3.7,"publicationDate":"2025-12-29","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145846439","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Serum GFAP as a biomarker for progression in multiple sclerosis: assay comparison and a large reference database of healthy controls. 血清GFAP作为多发性硬化症进展的生物标志物:测定比较和健康对照的大型参考数据库
IF 3.7 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2025-12-29 DOI: 10.1515/cclm-2025-1480
Eline A J Willemse, Sofia Sandgren, Pascal Benkert, Sabine Schaedelin, Aleksandra Maleska Maceski, Johanna Oechtering, Nafiye Genc, Klaus Berger, Marco Hermesdorf, Stefanie Müller, Sebastian Finkener, Juan F Vilchez Gomez, Amar Zadic, Giulio Disanto, Marcus D'Souza, Cristina Granziera, Caroline Pot, Chiara Zecca, Patrice H Lalive, Robert Hoepner, Patrick Roth, Marina Herwerth, Claudio Gobbi, David Leppert, Maximilian Einsiedler, Jens Kuhle

Objectives: Compare Elecsys (Roche) and Simoa (Quanterix) immunoassays for serum glial fibrillary acidic protein (GFAP) using our reference database and Z scores, and evaluate their prognostic value for progression independent of relapse activity (PIRA) in multiple sclerosis (MS).

Methods: Platform correlation was assessed in 612 samples from healthy controls (n=188; median [interquartile range, IQR] age 45.1 [36.4-61.7] years) and people with MS (n=424; 45.3 [35.2-53.9] years). Elecsys values were converted to Z scores via Passing-Bablok-derived regression and validated in fingolimod (n=414), and B-cell depleting therapy (BCDT; n=353) cohorts. Z scores and hazard ratios (HRs) for time-to-PIRA were compared using Cox regression.

Results: GFAPSimoa and GFAPElecsys measurements were correlated (r=0.94), with Elecsys values ∼54 % lower (GFAPElecsys, ng/L=2.847 [95 % confidence interval, CI: 1.335 - 4.98] + 0.457 [0.434 - 0.478] * GFAPSimoa, ng/L). In univariable Cox models, GFAPSimoa and GFAPElecsys Z scores were associated with time-to-PIRA in both validation cohorts. In multivariable Cox models, higher GFAPSimoa Z scores were associated with shorter time-to-PIRA in fingolimod cohort (HR: 1.27 [95 % CI 1.08 - 1.50], p=0.0031) and trended toward significance in BCDT (1.18 [0.99 - 1.41, p=0.0693). In contrast, GFAPElecsys Z scores were associated with time-to-PIRA in both cohorts (fingolimod: 1.27 [1.09 - 1.48], p=0.0023; BCDT: (1.19 [1.00 - 1.40], p=0.0487).

Conclusions: Serum GFAP measured by Elecsys shows a comparable association with time-to-PIRA as Simoa, and GFAPSimoa Z scores can be successfully bridged to GFAPElecsys Z scores, supporting Elecsys`s potential for clinical implementation.

目的:比较Elecsys(罗氏)和Simoa (Quanterix)的血清胶质纤维酸性蛋白(GFAP)免疫测定方法,使用我们的参考数据库和Z评分,并评估它们在多发性硬化症(MS)中独立于复发活动(PIRA)的进展的预后价值。方法:对612例健康对照(n=188,中位年龄45.1[36.4-61.7]岁)和MS患者(n=424; 45.3[35.2-53.9]岁)进行平台相关性评估。Elecsys值通过passing - bablok衍生回归转换为Z评分,并在fingolimod (n=414)和b细胞消耗治疗(BCDT, n=353)队列中进行验证。采用Cox回归比较Z评分和风险比(hr)。结果:GFAPSimoa和GFAPElecsys测量值相关(r=0.94), Elecsys值低~ 54 % (GFAPElecsys, ng/L=2.847[95 %置信区间,CI: 1.335 - 4.98] + 0.457 [0.434 - 0.478] * GFAPSimoa, ng/L)。在单变量Cox模型中,在两个验证队列中,GFAPSimoa和GFAPElecsys Z评分与到达pira的时间相关。在多变量Cox模型中,高GFAPSimoa Z评分与fingolimod队列中较短的到pira时间相关(HR: 1.27[95 % CI 1.08 - 1.50], p=0.0031),并且在BCDT中具有显著性(1.18 [0.99 - 1.41,p=0.0693)。相比之下,GFAPElecsys Z评分与两个队列中到达pira的时间相关(fingolimod: 1.27 [1.09 - 1.48], p=0.0023; BCDT: 1.19 [1.00 - 1.40], p=0.0487)。结论:Elecsys测量的血清GFAP显示出与Simoa相当的到pira时间的相关性,并且GFAPSimoa Z评分可以成功地桥接GFAPElecsys Z评分,支持Elecsys在临床应用的潜力。
{"title":"Serum GFAP as a biomarker for progression in multiple sclerosis: assay comparison and a large reference database of healthy controls.","authors":"Eline A J Willemse, Sofia Sandgren, Pascal Benkert, Sabine Schaedelin, Aleksandra Maleska Maceski, Johanna Oechtering, Nafiye Genc, Klaus Berger, Marco Hermesdorf, Stefanie Müller, Sebastian Finkener, Juan F Vilchez Gomez, Amar Zadic, Giulio Disanto, Marcus D'Souza, Cristina Granziera, Caroline Pot, Chiara Zecca, Patrice H Lalive, Robert Hoepner, Patrick Roth, Marina Herwerth, Claudio Gobbi, David Leppert, Maximilian Einsiedler, Jens Kuhle","doi":"10.1515/cclm-2025-1480","DOIUrl":"https://doi.org/10.1515/cclm-2025-1480","url":null,"abstract":"<p><strong>Objectives: </strong>Compare Elecsys (Roche) and Simoa (Quanterix) immunoassays for serum glial fibrillary acidic protein (GFAP) using our reference database and Z scores, and evaluate their prognostic value for progression independent of relapse activity (PIRA) in multiple sclerosis (MS).</p><p><strong>Methods: </strong>Platform correlation was assessed in 612 samples from healthy controls (n=188; median [interquartile range, IQR] age 45.1 [36.4-61.7] years) and people with MS (n=424; 45.3 [35.2-53.9] years). Elecsys values were converted to Z scores via Passing-Bablok-derived regression and validated in fingolimod (n=414), and B-cell depleting therapy (BCDT; n=353) cohorts. Z scores and hazard ratios (HRs) for time-to-PIRA were compared using Cox regression.</p><p><strong>Results: </strong>GFAP<sub>Simoa</sub> and GFAP<sub>Elecsys</sub> measurements were correlated (r=0.94), with Elecsys values ∼54 % lower (GFAP<sub>Elecsys</sub>, ng/L=2.847 [95 % confidence interval, CI: 1.335 - 4.98] + 0.457 [0.434 - 0.478] * GFAP<sub>Simoa</sub>, ng/L). In univariable Cox models, GFAP<sub>Simoa</sub> and GFAP<sub>Elecsys</sub> Z scores were associated with time-to-PIRA in both validation cohorts. In multivariable Cox models, higher GFAP<sub>Simoa</sub> Z scores were associated with shorter time-to-PIRA in fingolimod cohort (HR: 1.27 [95 % CI 1.08 - 1.50], p=0.0031) and trended toward significance in BCDT (1.18 [0.99 - 1.41, p=0.0693). In contrast, GFAP<sub>Elecsys</sub> Z scores were associated with time-to-PIRA in both cohorts (fingolimod: 1.27 [1.09 - 1.48], p=0.0023; BCDT: (1.19 [1.00 - 1.40], p=0.0487).</p><p><strong>Conclusions: </strong>Serum GFAP measured by Elecsys shows a comparable association with time-to-PIRA as Simoa, and GFAP<sub>Simoa</sub> Z scores can be successfully bridged to GFAP<sub>Elecsys</sub> Z scores, supporting Elecsys`s potential for clinical implementation.</p>","PeriodicalId":10390,"journal":{"name":"Clinical chemistry and laboratory medicine","volume":" ","pages":""},"PeriodicalIF":3.7,"publicationDate":"2025-12-29","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145843217","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Analytical verification and comparative assessment of the new Atellica IM high-sensitivity prostate specific antigen assay. 新型Atellica IM高灵敏度前列腺特异性抗原测定的分析验证和比较评价。
IF 3.7 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2025-12-23 DOI: 10.1515/cclm-2025-1431
Xavier Filella, Cristina González-Escribano, María Rodríguez-García, Núria Medina-Esteban, Esther Fernández-Galan

Objectives: Despite standardization efforts, significant inter-assay variability persists among prostate-specific antigen (PSA) tests, impacting prostate cancer (PCa) diagnosis and monitoring. We aimed to evaluate the analytical and clinical performance of the newly developed Atellica IM high-sensitivity PSA (hsPSA Atellica) assay compared with established PSA assays.

Methods: A total of 236 serum samples from healthy individuals and patients with or without PCa were analyzed using the hsPSA Atellica assay and four FDA-approved PSA assays: Hybritech Access, Architect, Atellica, and Cobas. Analytical performance included limit of detection (LOD), limit of quantification (LOQ), linearity, inter-assay precision, and hook effect. Method comparison was performed using Passing-Bablok regression, Bland-Altman analysis, and kappa index concordance.

Results: The hsPSA Atellica assay demonstrated a LOD of 0.01 μg/L and LOQ of 0.028 μg/L (CV: 3.2 %, accuracy: 115 %). Precision was maintained across concentrations, with CVs of 4.2 %, 3.8 %, and 2.4 % at low, medium, and high levels of PSA. Strong agreement was observed with the compared tests, particularly with Cobas and Hybritech PSA assays. Diagnostic sensitivity and specificity at the 4 μg/L clinical decision threshold were 98 % and 35 %, respectively. In 87 samples between 3 and 10 μg/L, concordance between hsPSA Atellica and Hybritech reached 96.6 % (κ=0.82). The assay remained accurate up to PSA concentrations of 13,311 μg/L, showing minimal hook effect.

Conclusions: The hsPSA Atellica assay shows excellent analytical sensitivity and strong agreement with established assays. To our knowledge, this is the first published evaluation of this assay, supporting its clinical utility in both PCa diagnosis and follow-up.

目的:尽管标准化的努力,显着的分析之间的差异仍然存在于前列腺特异性抗原(PSA)测试,影响前列腺癌(PCa)的诊断和监测。我们的目的是评估新开发的Atellica IM高灵敏度PSA (hsPSA Atellica)检测与现有PSA检测的分析和临床性能。方法:使用hsPSA Atellica检测和fda批准的四种PSA检测方法(Hybritech Access、Architect、Atellica和Cobas)对健康个体和患有或不患有PCa的患者的236份血清样本进行分析。分析性能包括检出限(LOD)、定量限(LOQ)、线性、测定间精密度和钩效应。方法采用Passing-Bablok回归、Bland-Altman分析和kappa指数一致性进行比较。结果:hsPSA Atellica法的检出限为0.01 μg/L,检出限为0.028 μg/L (CV: 3.2 %,准确度:115 %)。在不同浓度下保持精度,低、中、高水平PSA的CVs分别为4.2 %、3.8 %和2.4 %。与比较试验,特别是与Cobas和Hybritech PSA测定,观察到强烈的一致性。在4 μg/L的临床决策阈值下,诊断敏感性和特异性分别为98 %和35 %。在3 ~ 10 μg/L的87份样品中,hsPSA Atellica与Hybritech的一致性达到96.6% % (κ=0.82)。当PSA浓度达到13311 μg/L时,检测结果仍然准确,钩效应最小。结论:hsPSA Atellica检测方法具有良好的分析灵敏度,与已有的检测方法具有较强的一致性。据我们所知,这是首次发表对该检测的评价,支持其在前列腺癌诊断和随访中的临床应用。
{"title":"Analytical verification and comparative assessment of the new Atellica IM high-sensitivity prostate specific antigen assay.","authors":"Xavier Filella, Cristina González-Escribano, María Rodríguez-García, Núria Medina-Esteban, Esther Fernández-Galan","doi":"10.1515/cclm-2025-1431","DOIUrl":"https://doi.org/10.1515/cclm-2025-1431","url":null,"abstract":"<p><strong>Objectives: </strong>Despite standardization efforts, significant inter-assay variability persists among prostate-specific antigen (PSA) tests, impacting prostate cancer (PCa) diagnosis and monitoring. We aimed to evaluate the analytical and clinical performance of the newly developed Atellica IM high-sensitivity PSA (hsPSA Atellica) assay compared with established PSA assays.</p><p><strong>Methods: </strong>A total of 236 serum samples from healthy individuals and patients with or without PCa were analyzed using the hsPSA Atellica assay and four FDA-approved PSA assays: Hybritech Access, Architect, Atellica, and Cobas. Analytical performance included limit of detection (LOD), limit of quantification (LOQ), linearity, inter-assay precision, and hook effect. Method comparison was performed using Passing-Bablok regression, Bland-Altman analysis, and kappa index concordance.</p><p><strong>Results: </strong>The hsPSA Atellica assay demonstrated a LOD of 0.01 μg/L and LOQ of 0.028 μg/L (CV: 3.2 %, accuracy: 115 %). Precision was maintained across concentrations, with CVs of 4.2 %, 3.8 %, and 2.4 % at low, medium, and high levels of PSA. Strong agreement was observed with the compared tests, particularly with Cobas and Hybritech PSA assays. Diagnostic sensitivity and specificity at the 4 μg/L clinical decision threshold were 98 % and 35 %, respectively. In 87 samples between 3 and 10 μg/L, concordance between hsPSA Atellica and Hybritech reached 96.6 % (κ=0.82). The assay remained accurate up to PSA concentrations of 13,311 μg/L, showing minimal hook effect.</p><p><strong>Conclusions: </strong>The hsPSA Atellica assay shows excellent analytical sensitivity and strong agreement with established assays. To our knowledge, this is the first published evaluation of this assay, supporting its clinical utility in both PCa diagnosis and follow-up.</p>","PeriodicalId":10390,"journal":{"name":"Clinical chemistry and laboratory medicine","volume":" ","pages":""},"PeriodicalIF":3.7,"publicationDate":"2025-12-23","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145803328","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
期刊
Clinical chemistry and laboratory medicine
全部 Acc. Chem. Res. ACS Applied Bio Materials ACS Appl. Electron. Mater. ACS Appl. Energy Mater. ACS Appl. Mater. Interfaces ACS Appl. Nano Mater. ACS Appl. Polym. Mater. ACS BIOMATER-SCI ENG ACS Catal. ACS Cent. Sci. ACS Chem. Biol. ACS Chemical Health & Safety ACS Chem. Neurosci. ACS Comb. Sci. ACS Earth Space Chem. ACS Energy Lett. ACS Infect. Dis. ACS Macro Lett. ACS Mater. Lett. ACS Med. Chem. Lett. ACS Nano ACS Omega ACS Photonics ACS Sens. ACS Sustainable Chem. Eng. ACS Synth. Biol. Anal. Chem. BIOCHEMISTRY-US Bioconjugate Chem. BIOMACROMOLECULES Chem. Res. Toxicol. Chem. Rev. Chem. Mater. CRYST GROWTH DES ENERG FUEL Environ. Sci. Technol. Environ. Sci. Technol. Lett. Eur. J. Inorg. Chem. IND ENG CHEM RES Inorg. Chem. J. Agric. Food. Chem. J. Chem. Eng. Data J. Chem. Educ. J. Chem. Inf. Model. J. Chem. Theory Comput. J. Med. Chem. J. Nat. Prod. J PROTEOME RES J. Am. Chem. Soc. LANGMUIR MACROMOLECULES Mol. Pharmaceutics Nano Lett. Org. Lett. ORG PROCESS RES DEV ORGANOMETALLICS J. Org. Chem. J. Phys. Chem. J. Phys. Chem. A J. Phys. Chem. B J. Phys. Chem. C J. Phys. Chem. Lett. Analyst Anal. Methods Biomater. Sci. Catal. Sci. Technol. Chem. Commun. Chem. Soc. Rev. CHEM EDUC RES PRACT CRYSTENGCOMM Dalton Trans. Energy Environ. Sci. ENVIRON SCI-NANO ENVIRON SCI-PROC IMP ENVIRON SCI-WAT RES Faraday Discuss. Food Funct. Green Chem. Inorg. Chem. Front. Integr. Biol. J. Anal. At. Spectrom. J. Mater. Chem. A J. Mater. Chem. B J. Mater. Chem. C Lab Chip Mater. Chem. Front. Mater. Horiz. MEDCHEMCOMM Metallomics Mol. Biosyst. Mol. Syst. Des. Eng. Nanoscale Nanoscale Horiz. Nat. Prod. Rep. New J. Chem. Org. Biomol. Chem. Org. Chem. Front. PHOTOCH PHOTOBIO SCI PCCP Polym. Chem.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1