首页 > 最新文献

Medical Devices-Evidence and Research最新文献

英文 中文
The Comparison Between Arclight Loupe and Handheld Slit Lamp in Anterior Segment Eye Disease Diagnosis. 弓光镜与手持式裂隙灯在前段眼病诊断中的比较。
IF 1.5 Q4 ENGINEERING, BIOMEDICAL Pub Date : 2025-11-11 eCollection Date: 2025-01-01 DOI: 10.2147/MDER.S550130
Immaculate Atukunda, Rebecca Claire Lusobya, Andrew Weil Semulimi, John Mukisa, Abubakar Kalinaki, Caroline Nalukenge, Pamela Okwir Apio, Denis Erima, Charles Batte, David Mukunya, Juliet Otiti-Sengeri, Obaid Kousha, Andrew Blaikie

Introduction: Early diagnosis of eye diseases improves outcomes and reduces blindness. However, diagnostic capacity in resource-poor settings is limited by a shortage of trained and equipped healthcare professionals. The Arclight device is a user-friendly, cost-effective diagnostic tool for both anterior and posterior segment eye diseases that requires minimal training. Studies have shown it to be effective for assessing the fundal reflex and performing fundoscopy; however, its accuracy and acceptability for anterior segment eye disease are still unknown.

Methods: This was a cross-sectional study carried out at Mulago National Super Specialised Hospital, where 21 ophthalmic clinical officers (OCOs) were recruited from surrounding districts. Each participant evaluated eleven patients with anterior segment pathologies and one normal case using the Arclight loupe and a handheld Slit Lamp. The diagnosis and the proportion of correctly identified conditions were recorded. Focus group discussions were held to assess the acceptability of using the devices. Discussions were transcribed verbatim and analysed using latent and manifest content analysis.

Results: Of the 21 participants, 13 (61.9%) were female, with a median age of 39 (IQR: 33-49), and had practiced for 6 to 10 years, representing 76.2%. The majority of conditions were correctly identified by both the Arclight (71.2%) and handheld slit lamp (72.3%). Pterygium was the most accurately diagnosed condition (100%) with the Arclight. Based on themes from Sekhon's framework and the Technology Acceptance Model, the Arclight device was found to be highly acceptable and easy for OCOs to use.

Conclusion: The performance of the OCOs in diagnosing anterior segment diseases with the Arclight is comparable to that of a handheld Slit lamp. It was also found to be an acceptable device for diagnostic purposes. In conclusion, the Arclight device is a suitable, low-cost alternative to the handheld Slit Lamp, especially in resource-limited settings.

眼部疾病的早期诊断可以改善预后,减少失明。然而,在资源贫乏的环境中,诊断能力受到训练有素和装备精良的医疗保健专业人员短缺的限制。Arclight设备是一种用户友好、成本效益高的前、后段眼病诊断工具,只需最少的培训。研究表明,它是有效的评估眼底反射和进行眼底镜检查;然而,其对前段眼病的准确性和可接受性尚不清楚。方法:本研究是在穆拉戈国家超级专科医院进行的横断面研究,从周边地区招募了21名眼科临床官员(oco)。每位参与者使用Arclight镜和手持式裂隙灯评估了11例前段病变患者和1例正常患者。记录诊断结果和正确诊断的比例。举行了焦点小组讨论,以评估使用这些装置的可接受性。讨论被逐字记录下来,并使用潜在和显内容分析进行分析。结果:在21名参与者中,13名(61.9%)是女性,年龄中位数为39岁(IQR: 33-49),练习了6至10年,占76.2%。Arclight(71.2%)和手持式裂隙灯(72.3%)均能正确识别大多数条件。Arclight对翼状胬肉的诊断准确率最高(100%)。基于Sekhon框架和技术接受模型的主题,Arclight设备被认为是高度可接受的,并且易于oco使用。结论:弓形光对前节病变的诊断效果与手持式裂隙灯相当。它也被发现是一种可接受的诊断设备。总之,Arclight设备是手持式裂隙灯的一种合适的低成本替代品,特别是在资源有限的环境中。
{"title":"The Comparison Between Arclight Loupe and Handheld Slit Lamp in Anterior Segment Eye Disease Diagnosis.","authors":"Immaculate Atukunda, Rebecca Claire Lusobya, Andrew Weil Semulimi, John Mukisa, Abubakar Kalinaki, Caroline Nalukenge, Pamela Okwir Apio, Denis Erima, Charles Batte, David Mukunya, Juliet Otiti-Sengeri, Obaid Kousha, Andrew Blaikie","doi":"10.2147/MDER.S550130","DOIUrl":"10.2147/MDER.S550130","url":null,"abstract":"<p><strong>Introduction: </strong>Early diagnosis of eye diseases improves outcomes and reduces blindness. However, diagnostic capacity in resource-poor settings is limited by a shortage of trained and equipped healthcare professionals. The Arclight device is a user-friendly, cost-effective diagnostic tool for both anterior and posterior segment eye diseases that requires minimal training. Studies have shown it to be effective for assessing the fundal reflex and performing fundoscopy; however, its accuracy and acceptability for anterior segment eye disease are still unknown.</p><p><strong>Methods: </strong>This was a cross-sectional study carried out at Mulago National Super Specialised Hospital, where 21 ophthalmic clinical officers (OCOs) were recruited from surrounding districts. Each participant evaluated eleven patients with anterior segment pathologies and one normal case using the Arclight loupe and a handheld Slit Lamp. The diagnosis and the proportion of correctly identified conditions were recorded. Focus group discussions were held to assess the acceptability of using the devices. Discussions were transcribed verbatim and analysed using latent and manifest content analysis.</p><p><strong>Results: </strong>Of the 21 participants, 13 (61.9%) were female, with a median age of 39 (IQR: 33-49), and had practiced for 6 to 10 years, representing 76.2%. The majority of conditions were correctly identified by both the Arclight (71.2%) and handheld slit lamp (72.3%). Pterygium was the most accurately diagnosed condition (100%) with the Arclight. Based on themes from Sekhon's framework and the Technology Acceptance Model, the Arclight device was found to be highly acceptable and easy for OCOs to use.</p><p><strong>Conclusion: </strong>The performance of the OCOs in diagnosing anterior segment diseases with the Arclight is comparable to that of a handheld Slit lamp. It was also found to be an acceptable device for diagnostic purposes. In conclusion, the Arclight device is a suitable, low-cost alternative to the handheld Slit Lamp, especially in resource-limited settings.</p>","PeriodicalId":47140,"journal":{"name":"Medical Devices-Evidence and Research","volume":"18 ","pages":"553-564"},"PeriodicalIF":1.5,"publicationDate":"2025-11-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12619609/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145543090","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Passive, Active, or Both-What Hemostats Do Surgeons Choose in the Real World, and Should We Rethink It? 被动,主动,还是两者兼而有之——外科医生在现实世界中选择什么样的止血器,我们是否应该重新考虑?
IF 1.5 Q4 ENGINEERING, BIOMEDICAL Pub Date : 2025-11-04 eCollection Date: 2025-01-01 DOI: 10.2147/MDER.S549147
Kristen Knapp, Isabella G Armento, Rhea Parreno, Natalia Peres Martinez, Rory Tippit, John Chovanes

Background: This study evaluates the association between the type of topical haemostatic agents used during surgery-categorized as active, passive, or a combination of both-and their relationship with intraoperative blood loss and bleeding-related postoperative outcomes.

Methods: A retrospective analysis was conducted at Cooper Health System, Camden, New Jersey, US, using patient records to compare the clinical performance of three haemostatic strategies: active agents, passive agents, and a combined approach.

Results: A total of 149 patients were included in the analysis: 51 treated with passive agents, 51 with active agents, and 47 with both. The mean age at admission was 55.9 years (SD ±16.3). General surgery was the most frequently represented specialty (24.8%), followed by spine (22.1%), neurosurgery (16.8%), and solid organ procedures (16.1%). The combination group showed the highest estimated blood loss (EBL) at 521.3 mL (SD ±1456.83; 95% CI: 93.5-949.0), compared to 390.7 mL (SD ±1004.81; CI: 108.1-673.3) in the active group and 301.5 mL (SD ±295.78; CI: 218.3-384.7) in the passive group. However, ANCOVA results showed no statistically significant difference in adjusted mean EBL between groups (p = 0.309). Notably, patients in the combination group experienced longer operative times, extended ICU stays, and higher mortality.

Conclusion: Numerical trends suggest greater bleeding and more complex postoperative courses in patients treated with the combination of active and passive haemostats. The use of standardized, validated measures of intraoperative bleeding, paired with a strategic approach that anticipates and manages bleeding complications, may support improved clinical outcomes.

背景:本研究评估了手术中使用的局部止血药物类型(分为主动、被动或两者联合)与术中出血量和术后出血相关结果的关系。方法:回顾性分析美国新泽西州卡姆登市库珀健康系统的患者记录,比较三种止血策略的临床表现:主动药物、被动药物和联合方法。结果:共有149例患者纳入分析:51例接受被动药物治疗,51例接受主动药物治疗,47例同时接受两种药物治疗。入院时平均年龄55.9岁(SD±16.3)。普通外科是最常见的专科(24.8%),其次是脊柱(22.1%)、神经外科(16.8%)和实体器官手术(16.1%)。联合治疗组的估计失血量(EBL)最高,为521.3 mL (SD±1456.83;95% CI: 93.5-949.0),而主动治疗组为390.7 mL (SD±1004.81;CI: 108.1-673.3),被动治疗组为301.5 mL (SD±295.78;CI: 218.3-384.7)。然而,ANCOVA结果显示,两组校正后的平均EBL无统计学差异(p = 0.309)。值得注意的是,联合组患者手术时间更长,ICU住院时间更长,死亡率更高。结论:数字趋势表明,联合使用主动和被动止血剂的患者出血更大,术后病程更复杂。采用标准化的、经过验证的术中出血措施,结合预测和管理出血并发症的战略方法,可能有助于改善临床结果。
{"title":"Passive, Active, or Both-What Hemostats Do Surgeons Choose in the Real World, and Should We Rethink It?","authors":"Kristen Knapp, Isabella G Armento, Rhea Parreno, Natalia Peres Martinez, Rory Tippit, John Chovanes","doi":"10.2147/MDER.S549147","DOIUrl":"10.2147/MDER.S549147","url":null,"abstract":"<p><strong>Background: </strong>This study evaluates the association between the type of topical haemostatic agents used during surgery-categorized as active, passive, or a combination of both-and their relationship with intraoperative blood loss and bleeding-related postoperative outcomes.</p><p><strong>Methods: </strong>A retrospective analysis was conducted at <i>Cooper Health System, Camden, New Jersey, US</i>, using patient records to compare the clinical performance of three haemostatic strategies: active agents, passive agents, and a combined approach.</p><p><strong>Results: </strong>A total of 149 patients were included in the analysis: 51 treated with passive agents, 51 with active agents, and 47 with both. The mean age at admission was 55.9 years (SD ±16.3). General surgery was the most frequently represented specialty (24.8%), followed by spine (22.1%), neurosurgery (16.8%), and solid organ procedures (16.1%). The combination group showed the highest estimated blood loss (EBL) at 521.3 mL (SD ±1456.83; 95% CI: 93.5-949.0), compared to 390.7 mL (SD ±1004.81; CI: 108.1-673.3) in the active group and 301.5 mL (SD ±295.78; CI: 218.3-384.7) in the passive group. However, ANCOVA results showed no statistically significant difference in adjusted mean EBL between groups (p = 0.309). Notably, patients in the combination group experienced longer operative times, extended ICU stays, and higher mortality.</p><p><strong>Conclusion: </strong>Numerical trends suggest greater bleeding and more complex postoperative courses in patients treated with the combination of active and passive haemostats. The use of standardized, validated measures of intraoperative bleeding, paired with a strategic approach that anticipates and manages bleeding complications, may support improved clinical outcomes.</p>","PeriodicalId":47140,"journal":{"name":"Medical Devices-Evidence and Research","volume":"18 ","pages":"545-551"},"PeriodicalIF":1.5,"publicationDate":"2025-11-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12595961/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145483316","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Comparative Immunogenicity and Safety of Measles-Mumps-Rubella Vaccine When Administered Using Needle Free Injection System to That with Conventional Needle-Syringe in India: A Randomized, Parallel Group Study. 在印度,麻疹-腮腺炎-风疹疫苗使用无针注射系统与传统针筒注射系统的免疫原性和安全性比较:一项随机、平行组研究。
IF 1.5 Q4 ENGINEERING, BIOMEDICAL Pub Date : 2025-10-29 eCollection Date: 2025-01-01 DOI: 10.2147/MDER.S549379
B S Ratta, Ankur Naik, Sarvesh Mutha

Introduction: Pain due to injection and fear of needles is documented barriers to vaccination in children. Needle-Free Injection System ie N-FIS®, is an innovative delivery solution that addresses these limitations. To demonstrate its performance, study was conducted to compare immunogenicity and safety of the MMR vaccine when administered by N-FIS, to that administered using needle-syringe and needle.

Methodology: This was a prospective, randomized, parallel group study (January-September 2024). Eligible participants 9-12 months of age were administered MMR vaccine 0.5 mL, subcutaneously using either N-FIS® or syringe. Immunogenicity and safety were assessed before and 28 days post vaccination.

Results: A total of 60 infants with mean age of 10.5 months were enrolled in the study. At 28 day post dose 1 there was a significant and comparable rise in the MMR GMTs between the two groups. Similarly, at 28 day post vaccination, there was a significant rise in the proportions of seropositive subjects and the rates were comparable between the two groups (Measles: 100% vs 83%, p = 0.046; Mumps: 96.1% vs 95.8%, p = 0.999, Rubella: 92.3% vs 100%, p = 0.491). N-FIS was well tolerated and significantly lower number of subjects experienced pain (12% with N-FIS vs 33% with syringe). No serious or severe adverse events were reported.

Conclusion: Administration of MMR vaccine using N-FIS induced robust and comparable immune responses to that observed with use of Needle-syringe and was well tolerated providing an alternative to paediatrician over conventional delivery system.

由于注射引起的疼痛和对针头的恐惧是记录在案的儿童接种疫苗的障碍。无针注射系统(即N-FIS®)是一种创新的给药解决方案,解决了这些限制。为了证明其性能,进行了一项研究,比较了N-FIS给药与针-注射器和针头给药时MMR疫苗的免疫原性和安全性。方法:这是一项前瞻性、随机、平行组研究(2024年1月至9月)。符合条件的9-12个月大的参与者使用N-FIS®或注射器皮下注射0.5 mL MMR疫苗。接种前和接种后28天分别进行免疫原性和安全性评价。结果:共有60名平均年龄10.5个月的婴儿被纳入研究。在第1次剂量后28天,两组间MMR GMTs有显著且可比的上升。同样,在疫苗接种后28天,血清阳性受试者的比例显著上升,两组之间的比率具有可比性(麻疹:100% vs 83%, p = 0.046;腮腺炎:96.1% vs 95.8%, p = 0.999;风疹:92.3% vs 100%, p = 0.491)。N-FIS耐受性良好,并且明显减少了受试者的疼痛体验(N-FIS组为12%,注射器组为33%)。无严重或严重不良事件报告。结论:使用N-FIS接种MMR疫苗可诱导与使用针头注射器观察到的稳健和可比的免疫反应,并且耐受性良好,提供了一种替代儿科医生的传统给药系统。
{"title":"Comparative Immunogenicity and Safety of Measles-Mumps-Rubella Vaccine When Administered Using Needle Free Injection System to That with Conventional Needle-Syringe in India: A Randomized, Parallel Group Study.","authors":"B S Ratta, Ankur Naik, Sarvesh Mutha","doi":"10.2147/MDER.S549379","DOIUrl":"10.2147/MDER.S549379","url":null,"abstract":"<p><strong>Introduction: </strong>Pain due to injection and fear of needles is documented barriers to vaccination in children. Needle-Free Injection System ie N-FIS<sup>®</sup>, is an innovative delivery solution that addresses these limitations. To demonstrate its performance, study was conducted to compare immunogenicity and safety of the MMR vaccine when administered by N-FIS, to that administered using needle-syringe and needle.</p><p><strong>Methodology: </strong>This was a prospective, randomized, parallel group study (January-September 2024). Eligible participants 9-12 months of age were administered MMR vaccine 0.5 mL, subcutaneously using either N-FIS<sup>®</sup> or syringe. Immunogenicity and safety were assessed before and 28 days post vaccination.</p><p><strong>Results: </strong>A total of 60 infants with mean age of 10.5 months were enrolled in the study. At 28 day post dose 1 there was a significant and comparable rise in the MMR GMTs between the two groups. Similarly, at 28 day post vaccination, there was a significant rise in the proportions of seropositive subjects and the rates were comparable between the two groups (Measles: 100% vs 83%, p = 0.046; Mumps: 96.1% vs 95.8%, p = 0.999, Rubella: 92.3% vs 100%, p = 0.491). N-FIS was well tolerated and significantly lower number of subjects experienced pain (12% with N-FIS vs 33% with syringe). No serious or severe adverse events were reported.</p><p><strong>Conclusion: </strong>Administration of MMR vaccine using N-FIS induced robust and comparable immune responses to that observed with use of Needle-syringe and was well tolerated providing an alternative to paediatrician over conventional delivery system.</p>","PeriodicalId":47140,"journal":{"name":"Medical Devices-Evidence and Research","volume":"18 ","pages":"537-543"},"PeriodicalIF":1.5,"publicationDate":"2025-10-29","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12579826/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145439902","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
A Prospective Clinical Study to Evaluate Safety and Performance of an Intra-Surgical Hydraulic Distractor During Spinal Surgeries. 一项评估脊柱手术中术内液压牵张器安全性和性能的前瞻性临床研究。
IF 1.5 Q4 ENGINEERING, BIOMEDICAL Pub Date : 2025-10-24 eCollection Date: 2025-01-01 DOI: 10.2147/MDER.S537601
Rahel Bornemann, Jens Himstedt, Bronek Boszczyk, Staffan Bowald, Kirsten Sander, Stan Mikulowski, Hesham Mohamed El Saady Farhoud, Robert Pflugmacher

Introduction: Spinal surgeries often require the manipulation of bone structures while at the same time protecting sensitive tissues. Existing surgical systems involve manual force, which has a high risk of damaging the vertebrae and instruments interfere with the visibility between the vertebrae, which complicates work. The OrtoWell® Distractor System, an intra-surgical hydraulic tool for vertebrae separation, offers incremental and stable force control and better visibility during distraction procedures in spinal surgeries. This study aims to fill the gap of still pending clinical trials using the device, by evaluating performance and safety of it across lumbar and thoracic surgeries.

Materials and methods: This study is a prospective, single-arm, multi-center observational study and was conducted between April 2018 and April 2023 at 3 German sites in patients undergoing anterior or lateral spinal surgery requiring vertebral separation. The device's safety and usability were assessed through surgeon-completed questionnaires on the device and its subcomponents, clinical outcomes and cardiovascular assessments. Assessments were conducted at four time points, at baseline, time of the index procedure, discharge and at a 30-day follow-up.

Results: 31 patients (13 males, mean age 62.33) were included. Single and multi-level treatments were performed using different accesses. The device facilitated spinal surgeries without any complications or device-related adverse events. Usability, stability and force control were evaluated as uniformly positive and all procedures proceeded as planned. Mean applied pressure was 147.69±29.48 bar. There were no major delays in surgical time and a mean blood loss of 215.17±113.57 cc. Cardiovascular parameters remained stable, with no significant changes between baseline and follow-up. Hospital stays averaged 16.39±10.25 days. Surgeons unanimously reported that the distractor simplified surgery.

Discussion and conclusions: The investigated device demonstrated high safety and performance, enabling precise force control and simplifying spinal surgeries. These findings suggest that hydraulic distractors could enhance surgical outcomes by improving intraoperative control and reducing procedural challenges.

Trial registration: Registered at ClinicalTrials.gov (NCT03501810, 12 Mar 2018).

脊柱手术通常需要对骨结构进行操作,同时保护敏感组织。现有的手术系统涉及手动操作,这有很高的损伤椎骨的风险,而且器械会干扰椎骨之间的可见性,这使工作变得复杂。OrtoWell®牵开器系统是一种用于脊柱分离的手术内液压工具,在脊柱手术牵开过程中提供增量和稳定的力控制和更好的可视性。本研究旨在通过评估该装置在腰椎和胸椎手术中的性能和安全性来填补尚待临床试验的空白。材料和方法:本研究是一项前瞻性、单臂、多中心观察性研究,于2018年4月至2023年4月在德国3个地点进行了需要椎体分离的前路或侧路脊柱手术患者。该设备的安全性和可用性通过外科医生完成的关于设备及其子部件、临床结果和心血管评估的问卷来评估。在基线、指标程序时间、出院和30天随访四个时间点进行评估。结果:纳入31例患者,其中男性13例,平均年龄62.33岁。采用不同的通路进行单次和多级处理。该设备促进脊柱手术,没有任何并发症或设备相关的不良事件。可用性、稳定性和力控制均被评价为积极的,所有程序都按计划进行。平均施加压力为147.69±29.48 bar。手术时间无重大延误,平均失血量为215.17±113.57 cc,心血管参数保持稳定,基线与随访无明显变化。平均住院时间16.39±10.25天。外科医生一致报告说,牵张器简化了手术。讨论与结论:所研究的装置具有较高的安全性和性能,可实现精确的力控制并简化脊柱手术。这些研究结果表明,液压牵引器可以通过改善术中控制和减少手术挑战来提高手术效果。试验注册:在ClinicalTrials.gov注册(NCT03501810, 2018年3月12日)。
{"title":"A Prospective Clinical Study to Evaluate Safety and Performance of an Intra-Surgical Hydraulic Distractor During Spinal Surgeries.","authors":"Rahel Bornemann, Jens Himstedt, Bronek Boszczyk, Staffan Bowald, Kirsten Sander, Stan Mikulowski, Hesham Mohamed El Saady Farhoud, Robert Pflugmacher","doi":"10.2147/MDER.S537601","DOIUrl":"10.2147/MDER.S537601","url":null,"abstract":"<p><strong>Introduction: </strong>Spinal surgeries often require the manipulation of bone structures while at the same time protecting sensitive tissues. Existing surgical systems involve manual force, which has a high risk of damaging the vertebrae and instruments interfere with the visibility between the vertebrae, which complicates work. The OrtoWell<sup>®</sup> Distractor System, an intra-surgical hydraulic tool for vertebrae separation, offers incremental and stable force control and better visibility during distraction procedures in spinal surgeries. This study aims to fill the gap of still pending clinical trials using the device, by evaluating performance and safety of it across lumbar and thoracic surgeries.</p><p><strong>Materials and methods: </strong>This study is a prospective, single-arm, multi-center observational study and was conducted between April 2018 and April 2023 at 3 German sites in patients undergoing anterior or lateral spinal surgery requiring vertebral separation. The device's safety and usability were assessed through surgeon-completed questionnaires on the device and its subcomponents, clinical outcomes and cardiovascular assessments. Assessments were conducted at four time points, at baseline, time of the index procedure, discharge and at a 30-day follow-up.</p><p><strong>Results: </strong>31 patients (13 males, mean age 62.33) were included. Single and multi-level treatments were performed using different accesses. The device facilitated spinal surgeries without any complications or device-related adverse events. Usability, stability and force control were evaluated as uniformly positive and all procedures proceeded as planned. Mean applied pressure was 147.69±29.48 bar. There were no major delays in surgical time and a mean blood loss of 215.17±113.57 cc. Cardiovascular parameters remained stable, with no significant changes between baseline and follow-up. Hospital stays averaged 16.39±10.25 days. Surgeons unanimously reported that the distractor simplified surgery.</p><p><strong>Discussion and conclusions: </strong>The investigated device demonstrated high safety and performance, enabling precise force control and simplifying spinal surgeries. These findings suggest that hydraulic distractors could enhance surgical outcomes by improving intraoperative control and reducing procedural challenges.</p><p><strong>Trial registration: </strong>Registered at ClinicalTrials.gov (NCT03501810, 12 Mar 2018).</p>","PeriodicalId":47140,"journal":{"name":"Medical Devices-Evidence and Research","volume":"18 ","pages":"525-536"},"PeriodicalIF":1.5,"publicationDate":"2025-10-24","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12560639/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145402295","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Late-Onset Isolated Vitreous Hemorrhage After Micropulse Transscleral Cyclophotocoagulation: A Case Report and Insights for a Rare Complication. 微脉冲经巩膜光凝术后迟发性孤立性玻璃体出血1例及罕见并发症分析。
IF 1.5 Q4 ENGINEERING, BIOMEDICAL Pub Date : 2025-10-11 eCollection Date: 2025-01-01 DOI: 10.2147/MDER.S551399
Wisam Shihadeh, Abdelwahab Aleshawi, Yara Aburamadan, Mohammed Al-Shalakhti, Amr Abu Najm, Ala' Y' Ibrahim

Background: Micropulse transscleral diode cyclophotocoagulation (MP-TSCPC) has been developed as an alternative and safer approach to cyclodestructive procedures. While MP-TSCPC has an improved safety profile, reported complications of this procedure, though uncommon, include anterior chamber inflammation, mydriasis, hypotony, macular edema, corneal edema, phthisis bulbi, vitreous hemorrhage (VH), and conjunctival hemorrhage.

Case presentation: We report a case of a 76-year-old male who underwent uneventful MP-TSCPC for the left eye due to moderate primary-open angle glaucoma. A few days following the procedure, dense VH developed without hyphema or suprachoroidal hemorrhage. The VH persisted despite observation, and pars plana vitrectomy was performed, the patient recovered quickly, and achieved a visual acuity of 6/12 with an intraocular pressure of 10 mmHg.

Conclusion: MP-TSCPC is considered a safe and effective procedure in glaucoma management. However, the treating ophthalmologist should always be aware of rare but potentially serious complications. In our case report, we report a case of dense VH without hyphema or suprachoroidal hemorrhage following MP-TSCPC.

背景:微脉冲经巩膜二极管光凝术(MP-TSCPC)已发展成为一种替代和更安全的环破坏方法。虽然MP-TSCPC具有更好的安全性,但该手术的并发症虽然罕见,但包括前房炎症、瞳孔肿大、低眼压、黄斑水肿、角膜水肿、球疱疮、玻璃体出血(VH)和结膜出血。病例介绍:我们报告一例76岁男性,因中度原发性开角型青光眼,接受了左眼MP-TSCPC手术。手术后几天,致密的VH发展无前房积血或脉络膜上出血。经观察VH持续存在,行玻璃体切割手术,患者恢复迅速,视力6/12,眼压10 mmHg。结论:MP-TSCPC治疗青光眼安全有效。然而,治疗眼科医生应该始终意识到罕见但潜在的严重并发症。在我们的病例报告中,我们报告了一例密集的VH,没有前房积血或脉络膜上出血的MP-TSCPC。
{"title":"Late-Onset Isolated Vitreous Hemorrhage After Micropulse Transscleral Cyclophotocoagulation: A Case Report and Insights for a Rare Complication.","authors":"Wisam Shihadeh, Abdelwahab Aleshawi, Yara Aburamadan, Mohammed Al-Shalakhti, Amr Abu Najm, Ala' Y' Ibrahim","doi":"10.2147/MDER.S551399","DOIUrl":"10.2147/MDER.S551399","url":null,"abstract":"<p><strong>Background: </strong>Micropulse transscleral diode cyclophotocoagulation (MP-TSCPC) has been developed as an alternative and safer approach to cyclodestructive procedures. While MP-TSCPC has an improved safety profile, reported complications of this procedure, though uncommon, include anterior chamber inflammation, mydriasis, hypotony, macular edema, corneal edema, phthisis bulbi, vitreous hemorrhage (VH), and conjunctival hemorrhage.</p><p><strong>Case presentation: </strong>We report a case of a 76-year-old male who underwent uneventful MP-TSCPC for the left eye due to moderate primary-open angle glaucoma. A few days following the procedure, dense VH developed without hyphema or suprachoroidal hemorrhage. The VH persisted despite observation, and pars plana vitrectomy was performed, the patient recovered quickly, and achieved a visual acuity of 6/12 with an intraocular pressure of 10 mmHg.</p><p><strong>Conclusion: </strong>MP-TSCPC is considered a safe and effective procedure in glaucoma management. However, the treating ophthalmologist should always be aware of rare but potentially serious complications. In our case report, we report a case of dense VH without hyphema or suprachoroidal hemorrhage following MP-TSCPC.</p>","PeriodicalId":47140,"journal":{"name":"Medical Devices-Evidence and Research","volume":"18 ","pages":"519-524"},"PeriodicalIF":1.5,"publicationDate":"2025-10-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12523616/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145309612","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Multi-Center Evaluation of Optical Degradation Patterns in Rigid Endoscopes: A Data-Driven Approach. 刚性内窥镜光学退化模式的多中心评估:一种数据驱动的方法。
IF 1.5 Q4 ENGINEERING, BIOMEDICAL Pub Date : 2025-10-09 eCollection Date: 2025-01-01 DOI: 10.2147/MDER.S550187
Herke Jan Noordmans, Menno de Braak, Marco J Brakel, Alexander R Onur, Jeroen De Geus, Joris E N Jaspers

Introduction: The ScopeControl (Dovideq Medical, Hengelo, The Netherlands) measures the optical quality of rigid endoscopes in the sterilization department's cleanroom. Endoscopes with low quality are rejected for use and must be repaired or replaced.

Methods: An analysis of 500,000 measurements from 95 ScopeControls worldwide was conducted to evaluate hospital strategies, such as endoscope replacement timing, repair versus replace decisions, brand selection, and the impact of operating room to central sterile department processes. For the UMC Utrecht, repair data and usage data from surgery were included in the analysis. Custom software was used to filter out inexplicable, deviating measurements and calculate trend parameters such as initial and average quality, average quality decrease, and time to failure.

Results: The ScopeControl measurements are generally of good quality. Data exploration revealed significant variations between endoscopes. Lifespan ranged from 15 to 400 usages, with endoscopes often failing suddenly due to a broken lens, likely caused by mechanical or thermal shock. For specific endoscope types, comparisons between new and repaired endoscopes showed that repaired endoscopes have a median lifespan of 30% fewer usages and median light transmission that is 11% lower than new endoscopes. Entry measurements are highly recommended due to the large variation in the optical quality of both new and repaired endoscopes. Additionally, users appear to accept lower quality for vulnerable, long, thin pediatric endoscopes, possibly due to the high replacement cost, compared to endoscopes used for adult applications.

Conclusion: ScopeControl enables objective assessment of endoscope quality and supports data-driven decisions on repair and replacement. Our analysis suggests that repaired endoscopes generally do not match the optical quality of new ones, and we provide initial estimates of this difference. These findings highlight the need for further data collection and may encourage improvements in repair processes. Sudden failures underline the importance of regular checks, and linking ScopeControl data with repair and usage information remains essential for optimizing maintenance strategies and improving endoscope longevity.

ScopeControl (Dovideq Medical, Hengelo,荷兰)测量灭菌部门洁净室中刚性内窥镜的光学质量。质量差的内窥镜被拒绝使用,必须修理或更换。方法:对来自全球95个scopcontrols的50万项测量数据进行分析,以评估医院的策略,如内窥镜更换时间、修复与更换决策、品牌选择以及手术室对中心无菌部门流程的影响。对于乌得勒支UMC,修复数据和手术使用数据被纳入分析。使用定制软件过滤掉无法解释的、偏离的测量,并计算趋势参数,如初始质量和平均质量、平均质量下降和故障时间。结果:范围控制测量总体质量较好。数据探索揭示了内窥镜之间的显著差异。内窥镜的使用寿命从15到400次不等,通常是由于机械或热冲击造成的镜片破裂而突然失效。对于特定类型的内窥镜,新内窥镜和修复后的内窥镜的比较表明,修复后的内窥镜的使用寿命中位数比新内窥镜少30%,光透射中位数比新内窥镜低11%。由于新内窥镜和修复内窥镜的光学质量差异很大,因此强烈建议进行入门测量。此外,与用于成人应用的内窥镜相比,用户似乎接受易损坏的、长而薄的儿科内窥镜的质量较低,这可能是由于更换成本较高。结论:ScopeControl能够客观评估内窥镜质量,并支持数据驱动的修复和更换决策。我们的分析表明,修复后的内窥镜通常与新内窥镜的光学质量不匹配,我们提供了这种差异的初步估计。这些发现强调了进一步数据收集的必要性,并可能鼓励修复过程的改进。突发故障强调了定期检查的重要性,将ScopeControl数据与维修和使用信息联系起来对于优化维护策略和提高内窥镜使用寿命至关重要。
{"title":"Multi-Center Evaluation of Optical Degradation Patterns in Rigid Endoscopes: A Data-Driven Approach.","authors":"Herke Jan Noordmans, Menno de Braak, Marco J Brakel, Alexander R Onur, Jeroen De Geus, Joris E N Jaspers","doi":"10.2147/MDER.S550187","DOIUrl":"10.2147/MDER.S550187","url":null,"abstract":"<p><strong>Introduction: </strong>The ScopeControl (Dovideq Medical, Hengelo, The Netherlands) measures the optical quality of rigid endoscopes in the sterilization department's cleanroom. Endoscopes with low quality are rejected for use and must be repaired or replaced.</p><p><strong>Methods: </strong>An analysis of 500,000 measurements from 95 ScopeControls worldwide was conducted to evaluate hospital strategies, such as endoscope replacement timing, repair versus replace decisions, brand selection, and the impact of operating room to central sterile department processes. For the UMC Utrecht, repair data and usage data from surgery were included in the analysis. Custom software was used to filter out inexplicable, deviating measurements and calculate trend parameters such as initial and average quality, average quality decrease, and time to failure.</p><p><strong>Results: </strong>The ScopeControl measurements are generally of good quality. Data exploration revealed significant variations between endoscopes. Lifespan ranged from 15 to 400 usages, with endoscopes often failing suddenly due to a broken lens, likely caused by mechanical or thermal shock. For specific endoscope types, comparisons between new and repaired endoscopes showed that repaired endoscopes have a median lifespan of 30% fewer usages and median light transmission that is 11% lower than new endoscopes. Entry measurements are highly recommended due to the large variation in the optical quality of both new and repaired endoscopes. Additionally, users appear to accept lower quality for vulnerable, long, thin pediatric endoscopes, possibly due to the high replacement cost, compared to endoscopes used for adult applications.</p><p><strong>Conclusion: </strong>ScopeControl enables objective assessment of endoscope quality and supports data-driven decisions on repair and replacement. Our analysis suggests that repaired endoscopes generally do not match the optical quality of new ones, and we provide initial estimates of this difference. These findings highlight the need for further data collection and may encourage improvements in repair processes. Sudden failures underline the importance of regular checks, and linking ScopeControl data with repair and usage information remains essential for optimizing maintenance strategies and improving endoscope longevity.</p>","PeriodicalId":47140,"journal":{"name":"Medical Devices-Evidence and Research","volume":"18 ","pages":"507-518"},"PeriodicalIF":1.5,"publicationDate":"2025-10-09","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12517457/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145294056","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Proposed Guidelines for Reporting Parameters and Procedures of High- and Low-Level Laser Therapy in Medical Research Articles. 医学研究文章中报告高、低水平激光治疗的参数和程序的建议指南。
IF 1.5 Q4 ENGINEERING, BIOMEDICAL Pub Date : 2025-10-07 eCollection Date: 2025-01-01 DOI: 10.2147/MDER.S551850
Kawthar Shurrab

Objective: Laser therapy is increasingly used in clinical medicine, ranging from high-level laser therapy (HLLT) to low-level laser therapy (LLLT). This study proposes standardized guidelines for reporting parameters and procedures of HLLT and LLLT in medical research articles, aiming to improve clarity, reproducibility, and scientific rigor in clinical reporting.

Methods: The recommendations are based on insights from peer-reviewed studies that consistently identify gaps in the reporting of laser parameters and procedures. The key parameters include wavelength, power density, energy, energy density, and beam characteristics. Common issues noted are missing data, miscalculated doses, and unverified device specifications. To strengthen these recommendations, selected laser-tissue interactions were modeled using COMSOL Multiphysics. This modeling illustrates how power, spot size, and penetration depth affect treatment outcomes.

Results: The analysis revealed substantial inconsistencies in the way laser parameters are reported, often compromising reproducibility and reliability. The proposed framework introduces a structured checklist to support systematic documentation of treatment protocols, covering laser settings, irradiation, and treatment parameters.

Conclusion: Adopting these proposed guidelines can enhance the accuracy, transparency, and comparability of laser-based studies, filling a critical gap in current reporting practices and ultimately improving treatment validation and advancing the evidence base for HLLT and LLLT.

目的:激光治疗越来越多地应用于临床医学,从高水平激光治疗(HLLT)到低水平激光治疗(LLLT)。本研究提出了医学研究文章中HLLT和LLLT报告参数和程序的标准化指南,旨在提高临床报告的清晰度、可重复性和科学严谨性。方法:这些建议是基于同行评审研究的见解,这些研究一致地确定了激光参数和程序报告中的差距。关键参数包括波长、功率密度、能量、能量密度和光束特性。注意到的常见问题是数据丢失、剂量计算错误和未经验证的设备规格。为了加强这些建议,使用COMSOL Multiphysics对选定的激光与组织的相互作用进行了建模。该模型说明了功率、光斑大小和穿透深度如何影响治疗结果。结果:分析揭示了激光参数报告方式的实质性不一致,经常损害再现性和可靠性。拟议的框架引入了一个结构化的检查表,以支持治疗方案的系统文档,包括激光设置、照射和治疗参数。结论:采用这些拟议的指南可以提高激光研究的准确性、透明度和可比性,填补当前报告实践中的关键空白,最终改善治疗有效性,推进HLLT和LLLT的证据基础。
{"title":"Proposed Guidelines for Reporting Parameters and Procedures of High- and Low-Level Laser Therapy in Medical Research Articles.","authors":"Kawthar Shurrab","doi":"10.2147/MDER.S551850","DOIUrl":"10.2147/MDER.S551850","url":null,"abstract":"<p><strong>Objective: </strong>Laser therapy is increasingly used in clinical medicine, ranging from high-level laser therapy (HLLT) to low-level laser therapy (LLLT). This study proposes standardized guidelines for reporting parameters and procedures of HLLT and LLLT in medical research articles, aiming to improve clarity, reproducibility, and scientific rigor in clinical reporting.</p><p><strong>Methods: </strong>The recommendations are based on insights from peer-reviewed studies that consistently identify gaps in the reporting of laser parameters and procedures. The key parameters include wavelength, power density, energy, energy density, and beam characteristics. Common issues noted are missing data, miscalculated doses, and unverified device specifications. To strengthen these recommendations, selected laser-tissue interactions were modeled using COMSOL Multiphysics. This modeling illustrates how power, spot size, and penetration depth affect treatment outcomes.</p><p><strong>Results: </strong>The analysis revealed substantial inconsistencies in the way laser parameters are reported, often compromising reproducibility and reliability. The proposed framework introduces a structured checklist to support systematic documentation of treatment protocols, covering laser settings, irradiation, and treatment parameters.</p><p><strong>Conclusion: </strong>Adopting these proposed guidelines can enhance the accuracy, transparency, and comparability of laser-based studies, filling a critical gap in current reporting practices and ultimately improving treatment validation and advancing the evidence base for HLLT and LLLT.</p>","PeriodicalId":47140,"journal":{"name":"Medical Devices-Evidence and Research","volume":"18 ","pages":"495-505"},"PeriodicalIF":1.5,"publicationDate":"2025-10-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12515686/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145294010","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Energy Output Variability Among Phacoemulsification Tips Using Calorimetry. 用量热法测定超声乳化尖端的能量输出变异性。
IF 1.5 Q4 ENGINEERING, BIOMEDICAL Pub Date : 2025-10-03 eCollection Date: 2025-01-01 DOI: 10.2147/MDER.S538849
Ivan A Cardenas, Reiker G Ricks, Jenna L Jensen, Tanner K Nelson, William R Barlow Jnr, Randall J Olson, Jeff H Pettey

Purpose: To determine if using different phacoemulsification (phaco) tips will impact thermal energy output utilizing calorimetry.

Methods: The Alcon Centurion was used with the Infiniti OZil handpiece and balanced, hybrid, or MST Dewey A1 mini tip. Settings were 100% power, continuous torsional or longitudinal ultrasound, 0mmHg vacuum limit, and flow rate 12mL/min. Outflow tubing was removed from the handpiece and inflow tubing was manually occluded. The handpiece tip was placed into the calorimeter filled with 50mL of degassed deionized water and ultrasound was initiated for 1 minute. Temperature rise and cumulative dissipated energy (CDE) were recorded; ten trials were performed for each group.

Results: Statistically significant differences in energy output, measured in Joules, were found between all three tips in both ultrasound modes. Balanced tips showed the highest output, Dewey tips had the lowest. Hybrid tips had significantly increased output in longitudinal mode, whereas Dewey tips had significantly decreased output in longitudinal mode. No significant differences were seen in CDE between tips in either ultrasound mode, but significant increases were seen in longitudinal mode within each tip.

Conclusion: This study demonstrated differences in thermal energy production of three different phaco handpiece tips. This provides insight into how thermal injuries occur and what might cause these differences. Our methodology provides a novel way for accurate energy measurement at the tip, though it differs when compared to external energy production measurements. Surgeons should take into consideration the significant variability in thermal energy created by handpiece tips.

目的:利用量热法确定使用不同的超声乳化(phaco)尖端是否会影响热能输出。方法:使用爱尔康百夫长Infiniti OZil手机和平衡,混合,或MST杜威A1迷你尖端。设置为100%功率,连续扭转或纵向超声,0mmHg真空限制,流量12mL/min。从机头上取下流出管,手动堵塞流入管。将手机尖端放入盛有50mL脱气去离子水的量热计中,启动超声1分钟。记录温升和累积耗散能(CDE);每组进行10次试验。结果:在两种超声模式下,以焦耳为单位测量的能量输出在统计上有显著差异。平衡叶片的产量最高,杜威叶片的产量最低。杂交叶尖在纵向模式下显著增加了产量,而杜威叶尖在纵向模式下显著降低了产量。在两种超声模式下,CDE在尖端之间没有显著差异,但在每个尖端的纵向模式下,CDE显著增加。结论:本研究证明了三种不同的phaco手机尖端的热能产生差异。这为热损伤是如何发生以及可能导致这些差异的原因提供了见解。我们的方法为尖端精确的能量测量提供了一种新颖的方法,尽管与外部能量生产测量相比有所不同。外科医生应考虑到手机尖端产生的热能的显著变化。
{"title":"Energy Output Variability Among Phacoemulsification Tips Using Calorimetry.","authors":"Ivan A Cardenas, Reiker G Ricks, Jenna L Jensen, Tanner K Nelson, William R Barlow Jnr, Randall J Olson, Jeff H Pettey","doi":"10.2147/MDER.S538849","DOIUrl":"10.2147/MDER.S538849","url":null,"abstract":"<p><strong>Purpose: </strong>To determine if using different phacoemulsification (phaco) tips will impact thermal energy output utilizing calorimetry.</p><p><strong>Methods: </strong>The Alcon Centurion was used with the Infiniti OZil handpiece and balanced, hybrid, or MST Dewey A1 mini tip. Settings were 100% power, continuous torsional or longitudinal ultrasound, 0mmHg vacuum limit, and flow rate 12mL/min. Outflow tubing was removed from the handpiece and inflow tubing was manually occluded. The handpiece tip was placed into the calorimeter filled with 50mL of degassed deionized water and ultrasound was initiated for 1 minute. Temperature rise and cumulative dissipated energy (CDE) were recorded; ten trials were performed for each group.</p><p><strong>Results: </strong>Statistically significant differences in energy output, measured in Joules, were found between all three tips in both ultrasound modes. Balanced tips showed the highest output, Dewey tips had the lowest. Hybrid tips had significantly increased output in longitudinal mode, whereas Dewey tips had significantly decreased output in longitudinal mode. No significant differences were seen in CDE between tips in either ultrasound mode, but significant increases were seen in longitudinal mode within each tip.</p><p><strong>Conclusion: </strong>This study demonstrated differences in thermal energy production of three different phaco handpiece tips. This provides insight into how thermal injuries occur and what might cause these differences. Our methodology provides a novel way for accurate energy measurement at the tip, though it differs when compared to external energy production measurements. Surgeons should take into consideration the significant variability in thermal energy created by handpiece tips.</p>","PeriodicalId":47140,"journal":{"name":"Medical Devices-Evidence and Research","volume":"18 ","pages":"485-493"},"PeriodicalIF":1.5,"publicationDate":"2025-10-03","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12502956/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145253266","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
A Human Factors Validation of a Smart Autoinjector Accessory Designed to Improve Self-Injection Outcomes and User Confidence. 一种智能自动注射器配件的人为因素验证,旨在提高自我注射效果和用户信心。
IF 1.5 Q4 ENGINEERING, BIOMEDICAL Pub Date : 2025-10-02 eCollection Date: 2025-01-01 DOI: 10.2147/MDER.S545256
Stephan Wegner, Jana Milenkovic, Anthony D Andre, Carson P Whitaker, Tobias Theiler

Purpose: Enhancing user guidance via visual and auditory real-time feedback during subcutaneous injections-combined with automated injection-data logging-has the potential to improve self-injection outcomes and increase user confidence. To meet these objectives, a Smart Autoinjector Accessory (SAA) was developed to help patients, caregivers, and healthcare practitioners (HCPs) manage the injection process more effectively. This human factors validation study assessed whether the SAA, together with its Instructions for Use (IFU), could be operated safely and effectively with an autoinjector by the intended user population.

Methods: Seventy-five untrained participants (patients, n=45; caregivers, n= 15; HCPs, n=15) were evaluated on completion of SAA use tasks and IFU knowledge tasks in a simulated use environment.

Results: All participants (75/75, 100%) successfully delivered the full dose with the associated autoinjector. A total of 1196 out of 1200 use tasks (99.7%) delivering a simulated injection with the SAA and 2772 out of 2775 knowledge tasks (99.9%) concerning content in the IFU were completed successfully. Therefore, the overall results show near perfect success rates. All observed errors were isolated to individual participants, and no patterns of recurring use- or knowledge-task errors were found. Root cause analysis attributed these errors to inattention rather than deficiencies in the IFU clarity or device design.

Conclusion: This study shows that the intended users can operate the SAA safely and effectively. Given the strong usability outcomes, the results warrant further investigation into the SAA's potential to improve treatment adherence and persistence in longer-term studies. In post-launch use, the SAA could play a key role in facilitating the onboarding of inexperienced patients or those undergoing more complex treatment regimens. Beyond individual therapy, the SAA also holds promise in digitalizing real-world evidence collection for post-market insights and clinical drug development, supporting decentralized clinical trial designs, where patients administer injections at home rather than in clinical environments.

目的:在皮下注射过程中,通过视觉和听觉实时反馈加强用户指导,结合自动注射数据记录,有可能改善自我注射效果,增加用户信心。为了实现这些目标,我们开发了智能自动注射器配件(SAA),以帮助患者、护理人员和医疗保健从业者(HCPs)更有效地管理注射过程。这项人为因素验证研究评估了SAA及其使用说明书(IFU)是否可以由预期用户群体安全有效地使用自动进针器操作。方法:对75名未经培训的参与者(患者,n=45;护理人员,n=15;医护人员,n=15)在模拟使用环境中完成SAA使用任务和IFU知识任务进行评估。结果:所有参与者(75/ 75,100%)都成功地通过相关的自动注射器给药。1200个使用任务中有1196个(99.7%)使用SAA进行模拟注射,2775个知识任务中有2772个(99.9%)涉及IFU中的内容,成功完成。因此,总体结果显示出接近完美的成功率。所有观察到的错误都是孤立于个体参与者的,并且没有发现重复使用或知识任务错误的模式。根本原因分析将这些错误归因于注意力不集中,而不是IFU清晰度或设备设计的缺陷。结论:本研究表明预期使用者可以安全有效地操作SAA。鉴于强大的可用性结果,结果值得进一步调查SAA在长期研究中提高治疗依从性和持久性的潜力。在启动后的使用中,SAA可以在促进没有经验的患者或接受更复杂治疗方案的患者入组方面发挥关键作用。除了个人治疗之外,SAA还有望将现实世界的证据收集数字化,用于上市后的见解和临床药物开发,支持分散的临床试验设计,患者可以在家中而不是在临床环境中进行注射。
{"title":"A Human Factors Validation of a Smart Autoinjector Accessory Designed to Improve Self-Injection Outcomes and User Confidence.","authors":"Stephan Wegner, Jana Milenkovic, Anthony D Andre, Carson P Whitaker, Tobias Theiler","doi":"10.2147/MDER.S545256","DOIUrl":"10.2147/MDER.S545256","url":null,"abstract":"<p><strong>Purpose: </strong>Enhancing user guidance via visual and auditory real-time feedback during subcutaneous injections-combined with automated injection-data logging-has the potential to improve self-injection outcomes and increase user confidence. To meet these objectives, a Smart Autoinjector Accessory (SAA) was developed to help patients, caregivers, and healthcare practitioners (HCPs) manage the injection process more effectively. This human factors validation study assessed whether the SAA, together with its Instructions for Use (IFU), could be operated safely and effectively with an autoinjector by the intended user population.</p><p><strong>Methods: </strong>Seventy-five untrained participants (patients, n=45; caregivers, n= 15; HCPs, n=15) were evaluated on completion of SAA use tasks and IFU knowledge tasks in a simulated use environment.</p><p><strong>Results: </strong>All participants (75/75, 100%) successfully delivered the full dose with the associated autoinjector. A total of 1196 out of 1200 use tasks (99.7%) delivering a simulated injection with the SAA and 2772 out of 2775 knowledge tasks (99.9%) concerning content in the IFU were completed successfully. Therefore, the overall results show near perfect success rates. All observed errors were isolated to individual participants, and no patterns of recurring use- or knowledge-task errors were found. Root cause analysis attributed these errors to inattention rather than deficiencies in the IFU clarity or device design.</p><p><strong>Conclusion: </strong>This study shows that the intended users can operate the SAA safely and effectively. Given the strong usability outcomes, the results warrant further investigation into the SAA's potential to improve treatment adherence and persistence in longer-term studies. In post-launch use, the SAA could play a key role in facilitating the onboarding of inexperienced patients or those undergoing more complex treatment regimens. Beyond individual therapy, the SAA also holds promise in digitalizing real-world evidence collection for post-market insights and clinical drug development, supporting decentralized clinical trial designs, where patients administer injections at home rather than in clinical environments.</p>","PeriodicalId":47140,"journal":{"name":"Medical Devices-Evidence and Research","volume":"18 ","pages":"477-483"},"PeriodicalIF":1.5,"publicationDate":"2025-10-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12499574/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145245478","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Modelling and Simulation of the Pediatric Respiratory System with Endotracheal Tube During Pressure-Controlled Ventilation. 压力控制通气过程中气管内插管儿童呼吸系统的建模与仿真。
IF 1.5 Q4 ENGINEERING, BIOMEDICAL Pub Date : 2025-09-30 eCollection Date: 2025-01-01 DOI: 10.2147/MDER.S549778
Mohamed Bourti, Said Younous, Abdelaziz Belaguid

Purpose: To evaluate the impact of pediatric endotracheal tube (ETT) size on airflow and tidal volume (VT) waveforms during pressure-controlled ventilation (PCV) by developing and validating linear and nonlinear respiratory system models that incorporate ETT resistance.

Methods: Five pediatric ETTs (inner diameters: 3.5-5.5 mm) were analyzed using both linear (Poiseuille's law) and nonlinear (Rohrer's equation) mathematical models implemented in MATLAB Simulink. The models simulated PCV conditions and were validated against clinical data collected from nine pediatric ICU patients undergoing mechanical ventilation with ETTs in the same size range. Simulated signals for flow and tidal volume were compared with measured data using percentage differences and Student's t-test.

Results: The nonlinear model closely approximated clinical data, with average percentage differences of 9.85% for tidal volume and 15.68% for peak flow, significantly outperforming the linear model (17.67% and 46.87%, respectively; p < 0.001). Simulation results also demonstrated that smaller-diameter ETTs substantially increased airflow resistance, reducing delivered VT and requiring longer expiratory times to avoid air trapping.

Conclusion: Incorporating nonlinear ETT resistance into respiratory system models improves prediction accuracy under PCV, particularly in pediatric patients with smaller airways. The findings emphasize the importance of ETT size in ventilator parameter adjustment to optimize ventilation efficiency and minimize lung injury. The validated model provides a useful clinical tool for predicting ETT-related effects and guiding safer ventilation strategies in pediatric intensive care.

目的:通过建立和验证包含ETT阻力的线性和非线性呼吸系统模型,评估儿童气管内管(ETT)尺寸对压力控制通气(PCV)期间气流和潮气量(VT)波形的影响。方法:采用MATLAB Simulink实现的线性(Poiseuille’s law)和非线性(Rohrer’s equation)数学模型,对5个内径为3.5 ~ 5.5 mm的小儿外径进行分析。这些模型模拟了PCV条件,并通过收集的9名相同尺寸范围内接受机械通气的儿科ICU患者的临床数据进行了验证。利用百分比差异和学生t检验将流量和潮汐量的模拟信号与测量数据进行比较。结果:非线性模型与临床数据接近,潮汐量和峰值流量的平均百分比差异分别为9.85%和15.68%,显著优于线性模型(分别为17.67%和46.87%,p < 0.001)。模拟结果还表明,较小直径的eta显著增加了气流阻力,降低了输送的VT,并且需要更长的呼气时间来避免空气捕获。结论:将非线性ETT阻力纳入呼吸系统模型可提高PCV下的预测准确性,特别是对气道较小的儿科患者。研究结果强调了ETT大小在调整呼吸机参数以优化通气效率和减少肺损伤中的重要性。经过验证的模型为预测et相关效应和指导儿科重症监护更安全的通气策略提供了有用的临床工具。
{"title":"Modelling and Simulation of the Pediatric Respiratory System with Endotracheal Tube During Pressure-Controlled Ventilation.","authors":"Mohamed Bourti, Said Younous, Abdelaziz Belaguid","doi":"10.2147/MDER.S549778","DOIUrl":"10.2147/MDER.S549778","url":null,"abstract":"<p><strong>Purpose: </strong>To evaluate the impact of pediatric endotracheal tube (ETT) size on airflow and tidal volume (V<sub>T</sub>) waveforms during pressure-controlled ventilation (PCV) by developing and validating linear and nonlinear respiratory system models that incorporate ETT resistance.</p><p><strong>Methods: </strong>Five pediatric ETTs (inner diameters: 3.5-5.5 mm) were analyzed using both linear (Poiseuille's law) and nonlinear (Rohrer's equation) mathematical models implemented in MATLAB Simulink. The models simulated PCV conditions and were validated against clinical data collected from nine pediatric ICU patients undergoing mechanical ventilation with ETTs in the same size range. Simulated signals for flow and tidal volume were compared with measured data using percentage differences and Student's <i>t</i>-test.</p><p><strong>Results: </strong>The nonlinear model closely approximated clinical data, with average percentage differences of 9.85% for tidal volume and 15.68% for peak flow, significantly outperforming the linear model (17.67% and 46.87%, respectively; <i>p</i> < 0.001). Simulation results also demonstrated that smaller-diameter ETTs substantially increased airflow resistance, reducing delivered V<sub>T</sub> and requiring longer expiratory times to avoid air trapping.</p><p><strong>Conclusion: </strong>Incorporating nonlinear ETT resistance into respiratory system models improves prediction accuracy under PCV, particularly in pediatric patients with smaller airways. The findings emphasize the importance of ETT size in ventilator parameter adjustment to optimize ventilation efficiency and minimize lung injury. The validated model provides a useful clinical tool for predicting ETT-related effects and guiding safer ventilation strategies in pediatric intensive care.</p>","PeriodicalId":47140,"journal":{"name":"Medical Devices-Evidence and Research","volume":"18 ","pages":"461-475"},"PeriodicalIF":1.5,"publicationDate":"2025-09-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12495974/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145233783","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
期刊
Medical Devices-Evidence and Research
全部 Acc. Chem. Res. ACS Applied Bio Materials ACS Appl. Electron. Mater. ACS Appl. Energy Mater. ACS Appl. Mater. Interfaces ACS Appl. Nano Mater. ACS Appl. Polym. Mater. ACS BIOMATER-SCI ENG ACS Catal. ACS Cent. Sci. ACS Chem. Biol. ACS Chemical Health & Safety ACS Chem. Neurosci. ACS Comb. Sci. ACS Earth Space Chem. ACS Energy Lett. ACS Infect. Dis. ACS Macro Lett. ACS Mater. Lett. ACS Med. Chem. Lett. ACS Nano ACS Omega ACS Photonics ACS Sens. ACS Sustainable Chem. Eng. ACS Synth. Biol. Anal. Chem. BIOCHEMISTRY-US Bioconjugate Chem. BIOMACROMOLECULES Chem. Res. Toxicol. Chem. Rev. Chem. Mater. CRYST GROWTH DES ENERG FUEL Environ. Sci. Technol. Environ. Sci. Technol. Lett. Eur. J. Inorg. Chem. IND ENG CHEM RES Inorg. Chem. J. Agric. Food. Chem. J. Chem. Eng. Data J. Chem. Educ. J. Chem. Inf. Model. J. Chem. Theory Comput. J. Med. Chem. J. Nat. Prod. J PROTEOME RES J. Am. Chem. Soc. LANGMUIR MACROMOLECULES Mol. Pharmaceutics Nano Lett. Org. Lett. ORG PROCESS RES DEV ORGANOMETALLICS J. Org. Chem. J. Phys. Chem. J. Phys. Chem. A J. Phys. Chem. B J. Phys. Chem. C J. Phys. Chem. Lett. Analyst Anal. Methods Biomater. Sci. Catal. Sci. Technol. Chem. Commun. Chem. Soc. Rev. CHEM EDUC RES PRACT CRYSTENGCOMM Dalton Trans. Energy Environ. Sci. ENVIRON SCI-NANO ENVIRON SCI-PROC IMP ENVIRON SCI-WAT RES Faraday Discuss. Food Funct. Green Chem. Inorg. Chem. Front. Integr. Biol. J. Anal. At. Spectrom. J. Mater. Chem. A J. Mater. Chem. B J. Mater. Chem. C Lab Chip Mater. Chem. Front. Mater. Horiz. MEDCHEMCOMM Metallomics Mol. Biosyst. Mol. Syst. Des. Eng. Nanoscale Nanoscale Horiz. Nat. Prod. Rep. New J. Chem. Org. Biomol. Chem. Org. Chem. Front. PHOTOCH PHOTOBIO SCI PCCP Polym. Chem.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1