首页 > 最新文献

Iranian Journal of Radiation Research最新文献

英文 中文
Mass attenuation coefficients of environmental samples for gamma-ray energies from 46.5 keV to 1408 keV 46.5keV-1408keV伽马射线能量环境样品的质量衰减系数
Q4 Health Professions Pub Date : 2020-04-10 DOI: 10.18869/ACADPUB.IJRR.18.2.201
J. B. Lee, J. Byun, J. Yun
Background: In gamma-ray spectrometry for the radioactivity analysis of bulk samples, self-attenuation correction is necessary when the attenuation rates in test sample differ from the one in the efficiency calibration source. Thus, the mass attenuation coefficient of test samples is an important factor in gamma-ray spectrometry. This study estimates the mass attenuation coefficients for seven kinds of environmental samples. Materials and Methods: An uncollimated transmission system with a high purity germanium detector system was used to measure the gamma-ray transmission rates with and without test samples. The system was calibrated using C (activated charcoal), H2O, MnO2, NaCl, Na2CO3, and (NH4)2SO4 as reference materials. Sea sediment, surface soil, fish, seaweed, Chinese cabbage, milk, and pine needles were selected, and ten identical samples for each sample type were tested using the system. Results: The calibration of the uncollimated transmission system was validated with good agreement within 4% between linear attenuation coefficients by experiment and calculation for K2CrO4 and SiO2. The standard deviation of the mass attenuation coefficients for each kind of sample was estimated as less than 5% above around 100 keV. Conclusion: Mass attenuation coefficient does not significantly depend on the sample type for gamma-ray energy higher than around 100 keV, but mass attenuation coefficient for the lower gamma-ray energy should be considered even with similar kinds of sample. The mass attenuation coefficients tabulated in this paper can be used as a reference or comparable value in gamma-ray spectrometry for environmental samples.
背景:在用于散装样品放射性分析的伽马射线光谱法中,当试样的衰减率与效率校准源的衰减率不同时,需要进行自衰减校正。因此,测试样品的质量衰减系数是伽马射线光谱法中的一个重要因素。本研究估算了七种环境样品的质量衰减系数。材料和方法:使用带有高纯度锗探测器系统的未准直透射系统来测量有测试样品和没有测试样品的伽马射线透射率。使用C(活性炭)、H2O、MnO2、NaCl、Na2CO3和(NH4)2SO4作为参考材料对系统进行校准。选择海洋沉积物、表层土壤、鱼类、海藻、大白菜、牛奶和松针,并使用该系统对每种样本类型的10个相同样本进行测试。结果:通过K2CrO4和SiO2的实验和计算,验证了未准直透射系统的校准,线性衰减系数在4%以内具有良好的一致性。每种样品的质量衰减系数的标准偏差估计为在100keV以上小于5%。结论:对于高于100keV左右的伽马射线能量,质量衰减系数与样品类型无关,但即使使用类似类型的样品,也应考虑较低伽马射线能量的质量衰减系数。本文列出的质量衰减系数可作为环境样品伽马射线光谱法的参考或可比值。
{"title":"Mass attenuation coefficients of environmental samples for gamma-ray energies from 46.5 keV to 1408 keV","authors":"J. B. Lee, J. Byun, J. Yun","doi":"10.18869/ACADPUB.IJRR.18.2.201","DOIUrl":"https://doi.org/10.18869/ACADPUB.IJRR.18.2.201","url":null,"abstract":"Background: In gamma-ray spectrometry for the radioactivity analysis of bulk samples, self-attenuation correction is necessary when the attenuation rates in test sample differ from the one in the efficiency calibration source. Thus, the mass attenuation coefficient of test samples is an important factor in gamma-ray spectrometry. This study estimates the mass attenuation coefficients for seven kinds of environmental samples. Materials and Methods: An uncollimated transmission system with a high purity germanium detector system was used to measure the gamma-ray transmission rates with and without test samples. The system was calibrated using C (activated charcoal), H2O, MnO2, NaCl, Na2CO3, and (NH4)2SO4 as reference materials. Sea sediment, surface soil, fish, seaweed, Chinese cabbage, milk, and pine needles were selected, and ten identical samples for each sample type were tested using the system. Results: The calibration of the uncollimated transmission system was validated with good agreement within 4% between linear attenuation coefficients by experiment and calculation for K2CrO4 and SiO2. The standard deviation of the mass attenuation coefficients for each kind of sample was estimated as less than 5% above around 100 keV. Conclusion: Mass attenuation coefficient does not significantly depend on the sample type for gamma-ray energy higher than around 100 keV, but mass attenuation coefficient for the lower gamma-ray energy should be considered even with similar kinds of sample. The mass attenuation coefficients tabulated in this paper can be used as a reference or comparable value in gamma-ray spectrometry for environmental samples.","PeriodicalId":14498,"journal":{"name":"Iranian Journal of Radiation Research","volume":null,"pages":null},"PeriodicalIF":0.0,"publicationDate":"2020-04-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"48239083","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 2
Optimized production, quality control and biological assessment of 68-Ga-bleomycin as a possible PET imaging agent 68Ga博来霉素作为PET显像剂的优化生产、质量控制和生物学评价
Q4 Health Professions Pub Date : 2020-04-10 DOI: 10.18869/ACADPUB.IJRR.18.2.235
H. R. Tayeri, E. S. Khameneh, S. Zolghadri, S. Kakaei, D. Sardari
Background: Recently, several diagnostic radiopharmaceuticals including bleomycin (BLM) derivatives have been developed. Due to the suitable physical and chemical characteristics of Ga as a radionuclide for PET imaging, in this study, optimized production, quality control and preclinical evaluation of Ga-BLM as a new PET imaging agent is reported for the first time. Materials and Methods: Labeling of BLM with Ga was performed using Ge/Ga generator. Experiments were carried out by changing BLM concentration, temperature and pH of the reaction to determine the optimum parameters while the radiochemical purity was checked by radio thin layer chromatography at different times post labeling. Stability of the radiolabeled complex was studied at room temperature and in human serum at 37 °C. Biodistribution of the complex in BALB/c mice was assessed after intravenous injection and by counting the activity of each organ. Also, images were acquired up to 120 min by dual-head SPECT system. Results: The purity of this complex >96% (ITLC). At the optimized conditions for preparation of Ga-BLM (pH= 3.5-4, temperature = 90 oC, reaction time = 15 minutes and ligand concentration of 1 mg/ml) , the special activity of the labeled BLM reached to around 17.5 GBq/mmol. Biodistribution study showed significant accumulation of radioactivity in lung and bladder that was different pharmacokinetic compared to free Ga cation. Conclusion: Results show that Ga-BLM can be prepared in high radiochemical purity and high special activity only in 15 minutes and totally can be considered as a high potential agent for PET imaging.
背景:近年来,包括博来霉素(BLM)衍生物在内的几种放射性诊断药物已被开发出来。由于Ga作为一种用于PET成像的放射性核素具有合适的物理和化学特性,本研究首次报道了Ga-BLM作为一种新型PET成像剂的优化生产、质量控制和临床前评估。材料和方法:使用Ge/Ga发生器对BLM进行Ga标记。通过改变BLM浓度、反应温度和pH进行实验以确定最佳参数,同时在标记后的不同时间通过放射薄层色谱检查放射化学纯度。放射性标记的复合物在室温下和37°C下在人血清中的稳定性进行了研究。静脉注射后并通过计数每个器官的活性来评估复合物在BALB/c小鼠中的生物分布。此外,通过双头SPECT系统采集了长达120分钟的图像。结果:该配合物的纯度>96%(ITLC)。在优化的Ga-BLM制备条件下(pH=3.5-4,温度=90℃,反应时间=15分钟,配体浓度为1 mg/ml),标记的BLM的比活性达到约17.5 GBq/mmol。生物分布研究表明,与游离镓阳离子相比,肺和膀胱中的放射性积聚具有不同的药代动力学。结论:Ga-BLM在15分钟内即可制备出高放化纯度、高比活性的Ga-BLM,完全可以作为PET成像的一种高潜力试剂。
{"title":"Optimized production, quality control and biological assessment of 68-Ga-bleomycin as a possible PET imaging agent","authors":"H. R. Tayeri, E. S. Khameneh, S. Zolghadri, S. Kakaei, D. Sardari","doi":"10.18869/ACADPUB.IJRR.18.2.235","DOIUrl":"https://doi.org/10.18869/ACADPUB.IJRR.18.2.235","url":null,"abstract":"Background: Recently, several diagnostic radiopharmaceuticals including bleomycin (BLM) derivatives have been developed. Due to the suitable physical and chemical characteristics of Ga as a radionuclide for PET imaging, in this study, optimized production, quality control and preclinical evaluation of Ga-BLM as a new PET imaging agent is reported for the first time. Materials and Methods: Labeling of BLM with Ga was performed using Ge/Ga generator. Experiments were carried out by changing BLM concentration, temperature and pH of the reaction to determine the optimum parameters while the radiochemical purity was checked by radio thin layer chromatography at different times post labeling. Stability of the radiolabeled complex was studied at room temperature and in human serum at 37 °C. Biodistribution of the complex in BALB/c mice was assessed after intravenous injection and by counting the activity of each organ. Also, images were acquired up to 120 min by dual-head SPECT system. Results: The purity of this complex >96% (ITLC). At the optimized conditions for preparation of Ga-BLM (pH= 3.5-4, temperature = 90 oC, reaction time = 15 minutes and ligand concentration of 1 mg/ml) , the special activity of the labeled BLM reached to around 17.5 GBq/mmol. Biodistribution study showed significant accumulation of radioactivity in lung and bladder that was different pharmacokinetic compared to free Ga cation. Conclusion: Results show that Ga-BLM can be prepared in high radiochemical purity and high special activity only in 15 minutes and totally can be considered as a high potential agent for PET imaging.","PeriodicalId":14498,"journal":{"name":"Iranian Journal of Radiation Research","volume":null,"pages":null},"PeriodicalIF":0.0,"publicationDate":"2020-04-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"46444348","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 2
Dosimetric comparison of single and double collimator stereotactic body radiotherapy plans using Cyber Knife for carcinoma prostate 射波刀治疗前列腺癌单、双准直器立体定向放疗方案的剂量学比较
Q4 Health Professions Pub Date : 2020-04-10 DOI: 10.18869/ACADPUB.IJRR.18.2.209
T. Bijina, K. Ganesh, B. Subbulakshmi, A. Pichandi
Purpose: This study was conducted to evaluate single collimator (SC) and double collimator (DC) plans with respect to dosimetric analysis, calculated dose delivery to OAR and treatment time in carcinoma prostate patients treated with cyberknife. Materials and Methods: A retrospective study was conducted among twenty low and intermediate risk carcinoma prostate previously treated with Cyberknife. PTV was created and OARs were delineated. The prescribed dose was set as 37.5 Gy in 5 fractions and a base plan (BP), followed by three reduction plans (time, beam and node) were generated for both single and double collimators with sequential optimization module. The SC and DC plans were compared for the above-said variables. The mean differences were compared using paired t-test. A p-value of <0.05 was taken as statistically significant. Results: The median age of the patients was 63 years. DC plans had tighter isodose lines. The means of minimum doses did not vary significantly across the plans but the mean and maximum doses, PTV D2 and V95 means were significantly higher in single collimator plan. The mean CI and HI values were better in DC plans. The doses to OAR were comparable in both single and double collimator plans in terms of maximum doses. The mean doses received by OAR’s were significantly lesser in DC plans.SC plans resulted in lesser beams, nodes, MU and treatment time. Conclusion: Double collimator plans were better in producing good dosimetric results and reduced OAR doses with lesser estimated treatment efficiency.
目的:评价单准直器(SC)和双准直器(DC)方案在射波刀治疗前列腺癌患者的剂量学分析、OAR的计算剂量传递和治疗时间方面的效果。材料与方法:对20例既往行射波刀治疗的低、中危前列腺癌患者进行回顾性研究。建立了PTV,并划定了桨。将处方剂量设为37.5 Gy,分为5个分量和一个基本计划(BP),然后通过顺序优化模块对单准直器和双准直器分别生成时间、光束和节点3个缩减计划。比较SC和DC计划的上述变量。采用配对t检验比较均数差异。p值<0.05为差异有统计学意义。结果:患者中位年龄为63岁。直流电计有更紧的等剂量线。最小剂量平均值在不同准直器方案中差异不显著,但平均剂量和最大剂量、PTV D2和V95平均值均显著高于单一准直器方案。DC方案的平均CI和HI值较好。在最大剂量方面,单准直器和双准直器计划对OAR的剂量是相当的。在DC计划中,OAR接受的平均剂量明显较少。SC方案导致较少的束、节点、MU和治疗时间。结论:双准直仪方案可获得良好的剂量学结果,降低OAR剂量,但估计治疗效率较低。
{"title":"Dosimetric comparison of single and double collimator stereotactic body radiotherapy plans using Cyber Knife for carcinoma prostate","authors":"T. Bijina, K. Ganesh, B. Subbulakshmi, A. Pichandi","doi":"10.18869/ACADPUB.IJRR.18.2.209","DOIUrl":"https://doi.org/10.18869/ACADPUB.IJRR.18.2.209","url":null,"abstract":"Purpose: This study was conducted to evaluate single collimator (SC) and double collimator (DC) plans with respect to dosimetric analysis, calculated dose delivery to OAR and treatment time in carcinoma prostate patients treated with cyberknife. Materials and Methods: A retrospective study was conducted among twenty low and intermediate risk carcinoma prostate previously treated with Cyberknife. PTV was created and OARs were delineated. The prescribed dose was set as 37.5 Gy in 5 fractions and a base plan (BP), followed by three reduction plans (time, beam and node) were generated for both single and double collimators with sequential optimization module. The SC and DC plans were compared for the above-said variables. The mean differences were compared using paired t-test. A p-value of <0.05 was taken as statistically significant. Results: The median age of the patients was 63 years. DC plans had tighter isodose lines. The means of minimum doses did not vary significantly across the plans but the mean and maximum doses, PTV D2 and V95 means were significantly higher in single collimator plan. The mean CI and HI values were better in DC plans. The doses to OAR were comparable in both single and double collimator plans in terms of maximum doses. The mean doses received by OAR’s were significantly lesser in DC plans.SC plans resulted in lesser beams, nodes, MU and treatment time. Conclusion: Double collimator plans were better in producing good dosimetric results and reduced OAR doses with lesser estimated treatment efficiency.","PeriodicalId":14498,"journal":{"name":"Iranian Journal of Radiation Research","volume":null,"pages":null},"PeriodicalIF":0.0,"publicationDate":"2020-04-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"43994807","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Rational establishment of radon exposure standards for dwellings and workplaces 合理制定住宅和工作场所氡暴露标准
Q4 Health Professions Pub Date : 2020-04-10 DOI: 10.18869/ACADPUB.IJRR.18.2.359
Sang-Tae Kim, J. Yoo
Background: Radon is a normally occurring radioactive material, which is designated as a class 1 human carcinogen. Therefore, it is important to control radon exposure in dwellings and workplaces. Methods: The radon guidelines of 32 countries across Europe, North America, and Asia were examined to determine rational radon exposure standards for minimizing radon risk in dwellings, offices, and workplaces. The exposure standards were classified as standards for people in dwellings and offices, where radon exposure can occur through construction materials, and standards for workers exposed to radon at industrial sites, where they directly handle products containing radioactive matter such as raw materials and by-products from processing. Results: The examination results showed that in South Korea, the advisory reference level (ARL) for dwellings is set to 148 Bq/m. Moreover, ARLs are set for subway stations, libraries, medical institutions, and indoor parking lots, but there are no radon exposure standards to protect workers in manufacturing sites, officers, and other workplaces. In other countries, the ARL or the mandatory reference level (MRL) are usually regulated between 148–400 Bq/m for dwellings and public-use facilities, and between 200–1,000 Bq/m for workplaces. Conclusion: It is recommended to use 148 Bq/m, which is the standard set by the U.S. Environmental Protection Agency, for dwellings and workplaces. For workplaces, it is recommended to set the exposure standard between 400 Bq/m, which is the level adopted in most European Union member countries, and 1,000 Bq/m3, which is the reference level recommended by the International Commission on Radiological Protection.
背景:氡是一种常见的放射性物质,被列为人类1类致癌物质。因此,控制住宅和工作场所的氡暴露非常重要。方法:对欧洲、北美和亚洲32个国家的氡指南进行了审查,以确定合理的氡暴露标准,从而最大限度地降低住宅、办公室和工作场所的氡风险。暴露标准分为住宅和办公室人员的标准,建筑材料可能会导致氡暴露,以及工业场所工人暴露于氡的标准,他们直接处理含有放射性物质的产品,如原材料和加工副产品。结果:检查结果显示,在韩国,住宅的咨询参考水平(ARL)设定为148 Bq/m。此外,地铁站、图书馆、医疗机构和室内停车场都设置了ARL,但没有氡暴露标准来保护制造厂、官员和其他工作场所的工人。在其他国家,住宅和公共设施的ARL或强制性参考水平(MRL)通常在148-400 Bq/m之间,工作场所的ARL通常在200-1000 Bq/m。结论:建议住宅和工作场所使用148 Bq/m,这是美国环境保护局制定的标准。对于工作场所,建议将暴露标准设定在400 Bq/m(大多数欧盟成员国采用的水平)和1000 Bq/m3(国际辐射防护委员会建议的参考水平)之间。
{"title":"Rational establishment of radon exposure standards for dwellings and workplaces","authors":"Sang-Tae Kim, J. Yoo","doi":"10.18869/ACADPUB.IJRR.18.2.359","DOIUrl":"https://doi.org/10.18869/ACADPUB.IJRR.18.2.359","url":null,"abstract":"Background: Radon is a normally occurring radioactive material, which is designated as a class 1 human carcinogen. Therefore, it is important to control radon exposure in dwellings and workplaces. Methods: The radon guidelines of 32 countries across Europe, North America, and Asia were examined to determine rational radon exposure standards for minimizing radon risk in dwellings, offices, and workplaces. The exposure standards were classified as standards for people in dwellings and offices, where radon exposure can occur through construction materials, and standards for workers exposed to radon at industrial sites, where they directly handle products containing radioactive matter such as raw materials and by-products from processing. Results: The examination results showed that in South Korea, the advisory reference level (ARL) for dwellings is set to 148 Bq/m. Moreover, ARLs are set for subway stations, libraries, medical institutions, and indoor parking lots, but there are no radon exposure standards to protect workers in manufacturing sites, officers, and other workplaces. In other countries, the ARL or the mandatory reference level (MRL) are usually regulated between 148–400 Bq/m for dwellings and public-use facilities, and between 200–1,000 Bq/m for workplaces. Conclusion: It is recommended to use 148 Bq/m, which is the standard set by the U.S. Environmental Protection Agency, for dwellings and workplaces. For workplaces, it is recommended to set the exposure standard between 400 Bq/m, which is the level adopted in most European Union member countries, and 1,000 Bq/m3, which is the reference level recommended by the International Commission on Radiological Protection.","PeriodicalId":14498,"journal":{"name":"Iranian Journal of Radiation Research","volume":null,"pages":null},"PeriodicalIF":0.0,"publicationDate":"2020-04-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"45291984","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 3
Radiobiological comparison of 3D conformal and intensity modulated radiation therapy in the treatment of left-sided breast cancer 三维适形和调强放射治疗左癌症的放射生物学比较
Q4 Health Professions Pub Date : 2020-04-10 DOI: 10.18869/ACADPUB.IJRR.18.2.315
A. Shanei, A. Amouheidari, I. Abedi, A. Kazemzadeh, A. Jaafari
Background: The current study aimed to compare the tumor control probability (TCP) and normal tissue complication probability (NTCP) of threedimensional conformal radiation therapy (3D-CRT) and intensity-modulated radiation therapy (IMRT) for left-sided breast cancer using radiobiological models. Methods: This study was conducted on 30 patients with left-sided breast cancer, who were planned for 3D-CRT and 6-9 fields IMRT treatments using the PROWESS treatment planning system. The planning target volume (PTV) dose of 50 Gy was administered for the 3D-CRT and IMRT plans, respectively. The Niemierko’s equivalent uniform dose (EUD) model was utilized for the estimation of tumor control probability (TCP) and normal tissue complication probability (NTCP). Results: According to the results, the mean TCP values for 3D-CRT, 6-fields IMRT, and 9-fields IMRT plans were 99.07 ±0.07, 99.24 ±0.05 and 99.28 ±0.04, respectively, showing no statistically significant difference. The NTCPs of the lung and heart were considerably lower in the IMRT plans, compared to those in the 3D-CRT plans. Conclusions: From the radiobiological point of view, our results indicated that 3D-CRT produces a lower NTCP for ipsilateral lung. In contrast, for TCP calculations, there was a higher gain with IMRT plans compared to 3D-CRT plans.
背景:本研究旨在利用放射生物学模型比较癌症三维适形放射治疗(3D-CRT)和强度调节放射治疗(IMRT)的肿瘤控制概率(TCP)和正常组织并发症概率(NTCP)。方法:采用PROWESS治疗计划系统对30例癌症左侧患者进行3D-CRT和6-9视野IMRT治疗。3D-CRT和IMRT计划分别给予50Gy的计划目标体积(PTV)剂量。Niemierko等效均匀剂量(EUD)模型用于估计肿瘤控制概率(TCP)和正常组织并发症概率(NTCP)。结果:根据结果,3D-CRT、6场IMRT和9场IMRT方案的平均TCP值分别为99.07±0.07、99.24±0.05和99.28±0.04,无统计学差异。与3D-CRT计划相比,IMRT计划中肺和心脏的NTCP要低得多。结论:从放射生物学的角度来看,我们的结果表明3D-CRT对同侧肺产生较低的NTCP。相反,对于TCP计算,与3D-CRT计划相比,IMRT计划有更高的增益。
{"title":"Radiobiological comparison of 3D conformal and intensity modulated radiation therapy in the treatment of left-sided breast cancer","authors":"A. Shanei, A. Amouheidari, I. Abedi, A. Kazemzadeh, A. Jaafari","doi":"10.18869/ACADPUB.IJRR.18.2.315","DOIUrl":"https://doi.org/10.18869/ACADPUB.IJRR.18.2.315","url":null,"abstract":"Background: The current study aimed to compare the tumor control probability (TCP) and normal tissue complication probability (NTCP) of threedimensional conformal radiation therapy (3D-CRT) and intensity-modulated radiation therapy (IMRT) for left-sided breast cancer using radiobiological models. Methods: This study was conducted on 30 patients with left-sided breast cancer, who were planned for 3D-CRT and 6-9 fields IMRT treatments using the PROWESS treatment planning system. The planning target volume (PTV) dose of 50 Gy was administered for the 3D-CRT and IMRT plans, respectively. The Niemierko’s equivalent uniform dose (EUD) model was utilized for the estimation of tumor control probability (TCP) and normal tissue complication probability (NTCP). Results: According to the results, the mean TCP values for 3D-CRT, 6-fields IMRT, and 9-fields IMRT plans were 99.07 ±0.07, 99.24 ±0.05 and 99.28 ±0.04, respectively, showing no statistically significant difference. The NTCPs of the lung and heart were considerably lower in the IMRT plans, compared to those in the 3D-CRT plans. Conclusions: From the radiobiological point of view, our results indicated that 3D-CRT produces a lower NTCP for ipsilateral lung. In contrast, for TCP calculations, there was a higher gain with IMRT plans compared to 3D-CRT plans.","PeriodicalId":14498,"journal":{"name":"Iranian Journal of Radiation Research","volume":null,"pages":null},"PeriodicalIF":0.0,"publicationDate":"2020-04-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"46363240","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 5
Internal mammarial lymph node radiotherapy in obese patients with breast cancer, at what expense? 癌症肥胖患者乳房内淋巴结放疗的费用是多少?
Q4 Health Professions Pub Date : 2020-04-10 DOI: 10.18869/ACADPUB.IJRR.18.2.191
İ. Aral, S. Arslan, A. Yurekli, G. Inan, S. Tekin, H. Bozdemir, S. Sunel, H. Karabuğa, S. Acikgoz
Background and Aim: The incidence of internal lymph node (IMN) involvement was 465% in breast cancer patients. Despite studies indicating the positive effects of IMNRT on the oncological results, most of the clinicians avoided IMNRT because of the toxicity related to the increased dose of organs at risk (OAR). We aimed to compare the dosimetric results of RT plans with and without IMN containing planning target volumes (PTVs) using helical tomotherapy (HT) in obese patients. Materials and Methods: The PMRT data of 23 obese patients were evaluated retrospectively / dosimetrically. Two PTVs with and without IMN were defined and two separate plans were made with HT. Dose received by IMN and OAR were compared. Results: The untargeted IMN V40 were calculated between 0% to 99%. When the plans are evaluated in terms of critical organs, the inclusion of the IMN into the target volume, the most significant adverse effect was observed in heart doses in the left chest wall (CW) irradiation. The significant increases in cardiac V5 (% 62.6 vs %48.6 p=0.007), V10 (%38.2 vs %23.2 p=0.011), V20 (%14.15 vs %9.06 p=0.045) and maximum heart dose (48.04 vs 43.2 p=0.043) were observed in the left-side CW irradiations that involving the IMN. In CW irradiation on the right side with IMN, only a significant increase in mean heart dose (5.44 vs 4.52 p=0.036) was observed. Lung V5 doses were increased by inclusion with IMN in both sides. There was no difference in the contralateral breast doses in both plans for both sides. Conclusions: If the IMN is not targeted, some of the patients are getting appropriate doses in obese patients.
背景与目的:癌症患者的内淋巴结(IMN)受累率为465%。尽管研究表明IMNRT对肿瘤学结果有积极影响,但大多数临床医生避免使用IMNRT,因为其毒性与风险器官(OAR)剂量增加有关。我们的目的是比较在肥胖患者中使用螺旋断层治疗(HT)的带有和不带有IMN的计划靶体积(PTV)的RT计划的剂量测定结果。材料与方法:对23例肥胖患者的PMRT数据进行回顾性/剂量评估。定义了两个有IMN和没有IMN的PTV,并制定了两个单独的HT计划。比较了IMN和OAR接受的剂量。结果:非靶向IMN V40的计算结果在0%至99%之间。当根据关键器官对计划进行评估时,将IMN纳入目标体积,在左胸壁(CW)照射的心脏剂量中观察到最显著的不良影响。在涉及IMN的左侧CW照射中,观察到心脏V5(%62.6 vs%48.6 p=0.007)、V10(%38.2 vs%23.2 p=0.011)、V20(%14.15 vs%9.06 p=0.045)和最大心脏剂量(48.04 vs 43.2 p=0.043)的显著增加。在右侧连续波IMN照射中,仅观察到平均心脏剂量显著增加(5.44 vs 4.52 p=0.036)。肺V5剂量通过在两侧加入IMN而增加。在两种方案中,对侧乳腺的剂量对两侧均无差异。结论:如果IMN不是靶向性的,一些肥胖患者会得到适当的剂量。
{"title":"Internal mammarial lymph node radiotherapy in obese patients with breast cancer, at what expense?","authors":"İ. Aral, S. Arslan, A. Yurekli, G. Inan, S. Tekin, H. Bozdemir, S. Sunel, H. Karabuğa, S. Acikgoz","doi":"10.18869/ACADPUB.IJRR.18.2.191","DOIUrl":"https://doi.org/10.18869/ACADPUB.IJRR.18.2.191","url":null,"abstract":"Background and Aim: The incidence of internal lymph node (IMN) involvement was 465% in breast cancer patients. Despite studies indicating the positive effects of IMNRT on the oncological results, most of the clinicians avoided IMNRT because of the toxicity related to the increased dose of organs at risk (OAR). We aimed to compare the dosimetric results of RT plans with and without IMN containing planning target volumes (PTVs) using helical tomotherapy (HT) in obese patients. Materials and Methods: The PMRT data of 23 obese patients were evaluated retrospectively / dosimetrically. Two PTVs with and without IMN were defined and two separate plans were made with HT. Dose received by IMN and OAR were compared. Results: The untargeted IMN V40 were calculated between 0% to 99%. When the plans are evaluated in terms of critical organs, the inclusion of the IMN into the target volume, the most significant adverse effect was observed in heart doses in the left chest wall (CW) irradiation. The significant increases in cardiac V5 (% 62.6 vs %48.6 p=0.007), V10 (%38.2 vs %23.2 p=0.011), V20 (%14.15 vs %9.06 p=0.045) and maximum heart dose (48.04 vs 43.2 p=0.043) were observed in the left-side CW irradiations that involving the IMN. In CW irradiation on the right side with IMN, only a significant increase in mean heart dose (5.44 vs 4.52 p=0.036) was observed. Lung V5 doses were increased by inclusion with IMN in both sides. There was no difference in the contralateral breast doses in both plans for both sides. Conclusions: If the IMN is not targeted, some of the patients are getting appropriate doses in obese patients.","PeriodicalId":14498,"journal":{"name":"Iranian Journal of Radiation Research","volume":null,"pages":null},"PeriodicalIF":0.0,"publicationDate":"2020-04-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"48805250","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 1
Experimental and theoretical investigation into X-ray shielding properties of thin lead films 铅薄膜X射线屏蔽性能的实验与理论研究
Q4 Health Professions Pub Date : 2020-04-10 DOI: 10.18869/ACADPUB.IJRR.18.2.263
M. Vagheian, D. Sardari, S. Saramad, D. R. Ochbelagh
Background: Among all of the radiations, X-ray has been always the center of attention due to the increasing availability of the X-ray tubes in industry, research institutes and medical centers. In this study, X-ray shielding properties of bulk and nanostructured thin lead films were investigated by means of Monte-Carlo computational and experimental methods, respectively. Materials and Methods: The lead samples were fabricated by the Physical Vapor Deposition technique (PVD) with different thickness of 10, 100 and 1000 nm. To investigate the radiation shielding properties of the nanostructured thin films, all of the prepared samples were subjected to the X-ray ranging from 8 to 14 keV. In order to consider the shielding properties of the bulk-structured thin films, the Monte-Carlo MCNPX code was employed. Results: The results indicated that, for low X-ray energies, the nanostructured thin lead films attenuate more than bulk-structured samples; however, the difference disappears as film thickness increases to 1000 nm or X-ray energy reaches 14 keV. Conclusion: Results imply that the nanostructured thin lead films attenuate more photons than the bulkstructured thin lead films with the same thicknesses.
背景:在所有的辐射中,由于工业、研究机构和医疗中心越来越多地使用X射线管,X射线一直是人们关注的焦点。本研究分别采用蒙特卡罗计算和实验方法研究了体相和纳米结构铅薄膜的X射线屏蔽性能。材料和方法:采用物理气相沉积技术(PVD)制备了厚度分别为10、100和1000nm的铅样品。为了研究纳米结构薄膜的辐射屏蔽性能,对所有制备的样品进行8至14keV的X射线照射。为了考虑体相结构薄膜的屏蔽性能,采用了蒙特卡罗MCNPX程序。结果:结果表明,在低X射线能量下,纳米结构的铅薄膜比大块结构的样品衰减更大;然而,随着膜厚度增加到1000nm或X射线能量达到14keV,差异消失。结论:结果表明,纳米结构的铅薄膜比相同厚度的块状结构铅薄膜衰减更多的光子。
{"title":"Experimental and theoretical investigation into X-ray shielding properties of thin lead films","authors":"M. Vagheian, D. Sardari, S. Saramad, D. R. Ochbelagh","doi":"10.18869/ACADPUB.IJRR.18.2.263","DOIUrl":"https://doi.org/10.18869/ACADPUB.IJRR.18.2.263","url":null,"abstract":"Background: Among all of the radiations, X-ray has been always the center of attention due to the increasing availability of the X-ray tubes in industry, research institutes and medical centers. In this study, X-ray shielding properties of bulk and nanostructured thin lead films were investigated by means of Monte-Carlo computational and experimental methods, respectively. Materials and Methods: The lead samples were fabricated by the Physical Vapor Deposition technique (PVD) with different thickness of 10, 100 and 1000 nm. To investigate the radiation shielding properties of the nanostructured thin films, all of the prepared samples were subjected to the X-ray ranging from 8 to 14 keV. In order to consider the shielding properties of the bulk-structured thin films, the Monte-Carlo MCNPX code was employed. Results: The results indicated that, for low X-ray energies, the nanostructured thin lead films attenuate more than bulk-structured samples; however, the difference disappears as film thickness increases to 1000 nm or X-ray energy reaches 14 keV. Conclusion: Results imply that the nanostructured thin lead films attenuate more photons than the bulkstructured thin lead films with the same thicknesses.","PeriodicalId":14498,"journal":{"name":"Iranian Journal of Radiation Research","volume":null,"pages":null},"PeriodicalIF":0.0,"publicationDate":"2020-04-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"42719281","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 3
Adaptive approach for nasopharyngeal carcinoma patients during Volumetric Modulated Arc Therapy treatment (VMAT) 鼻咽癌患者在容积调制电弧治疗(VMAT)中的适应性方法
Q4 Health Professions Pub Date : 2020-04-10 DOI: 10.18869/ACADPUB.IJRR.18.2.369
B. I. Chibane, F. Benrachi, M. Bali
Background: Patients with head-and-neck cancers receiving radiotherapy show significant anatomical and dosimetric changes, especially during the latter part of treatment. The aim of this study is to evaluate the dosimetric effects of an adaptive Volumetric Modulated Arc Therapy (VMAT) treatment protocol for patients with locally advanced nasopharyngeal cancer (NPC). Material and methods: Ten patients treated with VMAT have benefited from a second computed tomography scan (CT2) after 15 fractions in order to apply a new adapted plan. A hybrid plan has been generated applying the original treatment plan configuration beam to the second CT scan. The dose– volume histograms (DVHs) of hybrid and adapted plans have been compared. Results: At the end of the 3 week of treatment, CT2 shows a considerable shrinkage of GTV N70 volume (45.2%) and a diminution of the left and right parotid glands volume (21.1%, 20.6% respectively). Compared to the initial plans, hybrid plans reveal that the dose delivered to target volume GTV N70 decreased by 15.2%, and the V30 of the left and right parotid glands increased by 47.3% and 25.6% respectively. However, there is no significant difference for the D2 of the brainstem and spinal cord. Conclusion: Our adaptive VMAT protocol improves dosimetric results in terms of GTV N70 coverage and nontoxic doses to parotid glands.
背景:接受放射治疗的头颈癌患者表现出显著的解剖和剂量变化,尤其是在治疗的后期。本研究的目的是评估自适应容量调制电弧疗法(VMAT)治疗方案对局部晚期癌症(NPC)患者的剂量测定效果。材料和方法:10名接受VMAT治疗的患者在15次分割后进行了第二次计算机断层扫描(CT2),以应用新的适应计划。已经生成了将原始治疗计划配置波束应用于第二CT扫描的混合计划。对混合方案和适应方案的剂量-体积直方图(DVH)进行了比较。结果:在治疗3周结束时,CT2显示GTV N70体积显著缩小(45.2%),左右腮腺体积缩小(分别为21.1%和20.6%)。与最初的计划相比,混合计划显示,输送到目标体积GTV N70的剂量减少了15.2%,左右腮腺的V30分别增加了47.3%和25.6%。然而,脑干和脊髓的D2没有显著差异。结论:我们的适应性VMAT方案在GTV N70覆盖率和腮腺无毒剂量方面提高了剂量测定结果。
{"title":"Adaptive approach for nasopharyngeal carcinoma patients during Volumetric Modulated Arc Therapy treatment (VMAT)","authors":"B. I. Chibane, F. Benrachi, M. Bali","doi":"10.18869/ACADPUB.IJRR.18.2.369","DOIUrl":"https://doi.org/10.18869/ACADPUB.IJRR.18.2.369","url":null,"abstract":"Background: Patients with head-and-neck cancers receiving radiotherapy show significant anatomical and dosimetric changes, especially during the latter part of treatment. The aim of this study is to evaluate the dosimetric effects of an adaptive Volumetric Modulated Arc Therapy (VMAT) treatment protocol for patients with locally advanced nasopharyngeal cancer (NPC). Material and methods: Ten patients treated with VMAT have benefited from a second computed tomography scan (CT2) after 15 fractions in order to apply a new adapted plan. A hybrid plan has been generated applying the original treatment plan configuration beam to the second CT scan. The dose– volume histograms (DVHs) of hybrid and adapted plans have been compared. Results: At the end of the 3 week of treatment, CT2 shows a considerable shrinkage of GTV N70 volume (45.2%) and a diminution of the left and right parotid glands volume (21.1%, 20.6% respectively). Compared to the initial plans, hybrid plans reveal that the dose delivered to target volume GTV N70 decreased by 15.2%, and the V30 of the left and right parotid glands increased by 47.3% and 25.6% respectively. However, there is no significant difference for the D2 of the brainstem and spinal cord. Conclusion: Our adaptive VMAT protocol improves dosimetric results in terms of GTV N70 coverage and nontoxic doses to parotid glands.","PeriodicalId":14498,"journal":{"name":"Iranian Journal of Radiation Research","volume":null,"pages":null},"PeriodicalIF":0.0,"publicationDate":"2020-04-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"49515931","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Radiation Shielding Materials: Half-value layer determination for separate and simultaneous photon and neutron emissions by a 252-Cf source 辐射屏蔽材料:252 Cf源单独和同时发射光子和中子的半值层测定
Q4 Health Professions Pub Date : 2020-04-10 DOI: 10.18869/ACADPUB.IJRR.18.2.381
Z. A. Rahvar, M. Ghorbani, M. Khosroabadi, C. Knaup
Background: The present study aimed to determine half-value layers (HVL) of different materials with neutron, photon, and simultaneous neutron and photon emissions by a Cf source. MCNPX code was used for the simulation of the applicator tube (AT) Cf source and the attenuation materials. Materials and methods: The Cf source, a source collimator, a detector, a detector collimator, and HVL material were defined in the present study. HVL was determined for lead, concrete, water, polyethylene, and polyethylene mixtures with various concentrations of boric acid for separate neutron, separate photon, and simultaneous emissions of neutron and photon by the Cf source. Results: HVL values were tabulated for the aforementioned materials for separate neutron, separate photon, and for simultaneous neutron and photon emissions by the Cf source. Among the evaluated materials, polyethylene was the most adequate material for neutron shielding. Conclusion: The obtained HVL results for different materials could be used for shielding design around a Cf source. According to the comparison of lead with concrete, lead was a suitable material for shielding against the photons emission. On the other hand, concrete with a lower HVL value seemed to be superior for neutron shielding. The combination of several HVL layers of polyethylene and lead could result in proper shielding against neutron and photon. In simultaneous neutron and photon emissions, the emitted neutrons from the source generate secondary photons. The issue led to increase HVL, and could be considered in the protection and shielding
背景:本研究旨在确定Cf源同时发射中子、光子和中子的不同材料的半值层(HVL)。MCNPX程序用于模拟敷贴管(AT)Cf源和衰减材料。材料和方法:本研究定义了Cf源、源准直器、探测器、探测器准直器和HVL材料。测定了铅、混凝土、水、聚乙烯和具有不同浓度硼酸的聚乙烯混合物的HVL,以分离中子、分离光子以及Cf源同时发射中子和光子。结果:列出了上述材料的单独中子、单独光子以及Cf源同时发射中子和光子的HVL值。在评估的材料中,聚乙烯是最适合用于中子屏蔽的材料。结论:所获得的不同材料的HVL结果可用于Cf源周围的屏蔽设计。根据铅与混凝土的比较,铅是屏蔽光子发射的合适材料。另一方面,HVL值较低的混凝土似乎更适合中子屏蔽。聚乙烯和铅的几个HVL层的组合可以产生对中子和光子的适当屏蔽。在中子和光子同时发射中,源发射的中子会产生二次光子。该问题导致HVL增加,可在保护和屏蔽中考虑
{"title":"Radiation Shielding Materials: Half-value layer determination for separate and simultaneous photon and neutron emissions by a 252-Cf source","authors":"Z. A. Rahvar, M. Ghorbani, M. Khosroabadi, C. Knaup","doi":"10.18869/ACADPUB.IJRR.18.2.381","DOIUrl":"https://doi.org/10.18869/ACADPUB.IJRR.18.2.381","url":null,"abstract":"Background: The present study aimed to determine half-value layers (HVL) of different materials with neutron, photon, and simultaneous neutron and photon emissions by a Cf source. MCNPX code was used for the simulation of the applicator tube (AT) Cf source and the attenuation materials. Materials and methods: The Cf source, a source collimator, a detector, a detector collimator, and HVL material were defined in the present study. HVL was determined for lead, concrete, water, polyethylene, and polyethylene mixtures with various concentrations of boric acid for separate neutron, separate photon, and simultaneous emissions of neutron and photon by the Cf source. Results: HVL values were tabulated for the aforementioned materials for separate neutron, separate photon, and for simultaneous neutron and photon emissions by the Cf source. Among the evaluated materials, polyethylene was the most adequate material for neutron shielding. Conclusion: The obtained HVL results for different materials could be used for shielding design around a Cf source. According to the comparison of lead with concrete, lead was a suitable material for shielding against the photons emission. On the other hand, concrete with a lower HVL value seemed to be superior for neutron shielding. The combination of several HVL layers of polyethylene and lead could result in proper shielding against neutron and photon. In simultaneous neutron and photon emissions, the emitted neutrons from the source generate secondary photons. The issue led to increase HVL, and could be considered in the protection and shielding","PeriodicalId":14498,"journal":{"name":"Iranian Journal of Radiation Research","volume":null,"pages":null},"PeriodicalIF":0.0,"publicationDate":"2020-04-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"47085588","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Evaluation of health hazards from radionuclides in soil and rocks of North Waziristan, Pakistan 巴基斯坦北瓦济里斯坦土壤和岩石中放射性核素对健康危害的评价
Q4 Health Professions Pub Date : 2020-04-10 DOI: 10.18869/ACADPUB.IJRR.18.2.243
I. Khan, Zhuang Qin, Tianci Xie, Z. Bin, Haopeng Li, Weimin Sun, E. Lewis
Background: The aim & objective of this study is to evaluate the health hazards of the general public from the existence of the radioactive nuclides in soil of North Waziristan and to establish a reference line for future work on radioactivity concentration in this locality. Methods and Materials: A highly background radiation shielded HPGe system is used to measure radioactivity in the collected samples of the geographical study area. Results: The radioactivity concentration is found to be in the range 42.6–106.3 Bq/kg for Ra-226, 4.0–93.8 Bq/kg for Th-232, 49.9–645.5 Bq/kg for K-40, and 2.6–8.33 Bq/kg for Cs-137. The external and internal hazards indices (Hext & Hint), absorbed γdose rate (ADR) and average effective dose (AED) are estimated from the computed activities and determined the average Hext ~ 0.75 and Hint ~ 0.94. The mean absorbed dose rate was found to be 101 nGy/h. The Radium equivalent activity of Datta Khel (422.081 Bq/kg), Khaisur forest (407.938 Bq/kg), Jani Khel (379.213) specific regions and the corresponding outdoor annual effective dose rate was found to be 0.15 mSv/yr, which is higher than the world’s mean value of 0.07 mSv/yr. Conclusion: A strong possitive correlation amongst the radioactive nuclides and, corresponding radiological variables at P ≤ 0.01confirmed that Ra-226 and Th-232 were the significant contributors to γ-radiation emission. Radioactivity concentrations and corresponding radiometric parameters were found to be higher in the specific areas of North Waziristan compared to the world’s mean values poses possible radiological risks for the residents of those areas.
背景:本研究的目的和目的是评价北瓦济里斯坦土壤中放射性核素的存在对公众健康的危害,并为今后该地区放射性浓度的工作建立参考线。方法与材料:采用高本底辐射屏蔽HPGe系统对地理研究区采集的样品进行放射性测量。结果:Ra-226的放射性浓度为42.6 ~ 106.3 Bq/kg, Th-232为4.0 ~ 93.8 Bq/kg, K-40为49.9 ~ 645.5 Bq/kg, Cs-137为2.6 ~ 8.33 Bq/kg。根据计算出的活度,估算出了外、内危害指数(Hext & Hint)、吸收γ剂量率(ADR)和平均有效剂量(AED),确定了平均Hext ~ 0.75, Hint ~ 0.94。平均吸收剂量率为101 nGy/h。达塔克尔(422.081 Bq/kg)、凯瑟尔森林(407.938 Bq/kg)、贾尼克尔(379.213)特定区域的放射性当量活度和相应的室外年有效剂量率为0.15 mSv/yr,高于世界平均值0.07 mSv/yr。结论:放射性核素与相应的放射学变量在P≤0.01时呈强正相关,证实了Ra-226和Th-232是γ辐射的重要贡献者。发现北瓦济里斯坦特定地区的放射性浓度和相应的辐射测量参数高于世界平均值,对这些地区的居民构成可能的辐射风险。
{"title":"Evaluation of health hazards from radionuclides in soil and rocks of North Waziristan, Pakistan","authors":"I. Khan, Zhuang Qin, Tianci Xie, Z. Bin, Haopeng Li, Weimin Sun, E. Lewis","doi":"10.18869/ACADPUB.IJRR.18.2.243","DOIUrl":"https://doi.org/10.18869/ACADPUB.IJRR.18.2.243","url":null,"abstract":"Background: The aim & objective of this study is to evaluate the health hazards of the general public from the existence of the radioactive nuclides in soil of North Waziristan and to establish a reference line for future work on radioactivity concentration in this locality. Methods and Materials: A highly background radiation shielded HPGe system is used to measure radioactivity in the collected samples of the geographical study area. Results: The radioactivity concentration is found to be in the range 42.6–106.3 Bq/kg for Ra-226, 4.0–93.8 Bq/kg for Th-232, 49.9–645.5 Bq/kg for K-40, and 2.6–8.33 Bq/kg for Cs-137. The external and internal hazards indices (Hext & Hint), absorbed γdose rate (ADR) and average effective dose (AED) are estimated from the computed activities and determined the average Hext ~ 0.75 and Hint ~ 0.94. The mean absorbed dose rate was found to be 101 nGy/h. The Radium equivalent activity of Datta Khel (422.081 Bq/kg), Khaisur forest (407.938 Bq/kg), Jani Khel (379.213) specific regions and the corresponding outdoor annual effective dose rate was found to be 0.15 mSv/yr, which is higher than the world’s mean value of 0.07 mSv/yr. Conclusion: A strong possitive correlation amongst the radioactive nuclides and, corresponding radiological variables at P ≤ 0.01confirmed that Ra-226 and Th-232 were the significant contributors to γ-radiation emission. Radioactivity concentrations and corresponding radiometric parameters were found to be higher in the specific areas of North Waziristan compared to the world’s mean values poses possible radiological risks for the residents of those areas.","PeriodicalId":14498,"journal":{"name":"Iranian Journal of Radiation Research","volume":null,"pages":null},"PeriodicalIF":0.0,"publicationDate":"2020-04-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"42858007","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 7
期刊
Iranian Journal of Radiation Research
全部 Acc. Chem. Res. ACS Applied Bio Materials ACS Appl. Electron. Mater. ACS Appl. Energy Mater. ACS Appl. Mater. Interfaces ACS Appl. Nano Mater. ACS Appl. Polym. Mater. ACS BIOMATER-SCI ENG ACS Catal. ACS Cent. Sci. ACS Chem. Biol. ACS Chemical Health & Safety ACS Chem. Neurosci. ACS Comb. Sci. ACS Earth Space Chem. ACS Energy Lett. ACS Infect. Dis. ACS Macro Lett. ACS Mater. Lett. ACS Med. Chem. Lett. ACS Nano ACS Omega ACS Photonics ACS Sens. ACS Sustainable Chem. Eng. ACS Synth. Biol. Anal. Chem. BIOCHEMISTRY-US Bioconjugate Chem. BIOMACROMOLECULES Chem. Res. Toxicol. Chem. Rev. Chem. Mater. CRYST GROWTH DES ENERG FUEL Environ. Sci. Technol. Environ. Sci. Technol. Lett. Eur. J. Inorg. Chem. IND ENG CHEM RES Inorg. Chem. J. Agric. Food. Chem. J. Chem. Eng. Data J. Chem. Educ. J. Chem. Inf. Model. J. Chem. Theory Comput. J. Med. Chem. J. Nat. Prod. J PROTEOME RES J. Am. Chem. Soc. LANGMUIR MACROMOLECULES Mol. Pharmaceutics Nano Lett. Org. Lett. ORG PROCESS RES DEV ORGANOMETALLICS J. Org. Chem. J. Phys. Chem. J. Phys. Chem. A J. Phys. Chem. B J. Phys. Chem. C J. Phys. Chem. Lett. Analyst Anal. Methods Biomater. Sci. Catal. Sci. Technol. Chem. Commun. Chem. Soc. Rev. CHEM EDUC RES PRACT CRYSTENGCOMM Dalton Trans. Energy Environ. Sci. ENVIRON SCI-NANO ENVIRON SCI-PROC IMP ENVIRON SCI-WAT RES Faraday Discuss. Food Funct. Green Chem. Inorg. Chem. Front. Integr. Biol. J. Anal. At. Spectrom. J. Mater. Chem. A J. Mater. Chem. B J. Mater. Chem. C Lab Chip Mater. Chem. Front. Mater. Horiz. MEDCHEMCOMM Metallomics Mol. Biosyst. Mol. Syst. Des. Eng. Nanoscale Nanoscale Horiz. Nat. Prod. Rep. New J. Chem. Org. Biomol. Chem. Org. Chem. Front. PHOTOCH PHOTOBIO SCI PCCP Polym. Chem.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1