首页 > 最新文献

Physica Medica-European Journal of Medical Physics最新文献

英文 中文
Optimization of microcalcification cluster detection in wide-angle flying focal spot digital breast tomosynthesis and synthetic mammography: A virtual imaging study 广角飞焦点数字乳房断层合成和合成乳房x线摄影中微钙化簇检测的优化:一项虚拟成像研究。
IF 2.7 3区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2025-11-17 DOI: 10.1016/j.ejmp.2025.105676
Katrien Houbrechts , Astrid Van Camp , Lesley Cockmartin , Liesbeth Vancoillie , Nicholas Marshall , Hilde Bosmans

Purpose

To investigate the impact of angular range and dose distribution on microcalcification cluster detection in digital breast tomosynthesis (DBT) and synthetic mammography (SM) of a clinical wide-angle DBT system with flying focal spot (FFS), through virtual imaging techniques.

Approach

DBT projection sets were acquired from ten patients at twice the automatic exposure controlled (AEC) dose. Noise was added to each projection to create four projection sets: (1) 25 projections at AEC dose (“AEC”), (2) 25 projections with a convex dose distribution (“convex”), (3) 25 projections with increased dose in the central three projections (“peak 3”), and (4) 19 projections covering a 40° angular range instead of the standard 50° (“40 degrees”). Total scan dose and angular spacing were maintained constant across all setups. Microcalcification clusters were simulated within these projection sets, followed by a human observer detection study with DBT and SM patches. Performance was analysed using jackknife-alternative free-response receiver operating characteristic (JAFROC) analysis.

Results

For DBT, the area under the curve (AUC) was 0.89 ± 0.03 (AEC), 0.91 ± 0.03 (convex), 0.88 ± 0.03 (peak 3), and 0.91 ± 0.04 (40 degrees). For SM, AUC values were lower: 0.73 ± 0.04, 0.74 ± 0.02, 0.73 ± 0.04, and 0.74 ± 0.02, respectively. No significant improvements were observed compared to the AEC setup for either modality.

Conclusions

Increasing the dose to the central projections or reducing the angular range from 50° to 40° did not significantly affect calcification detection in DBT or SM compared to the standard AEC setup for an FFS system.
目的:通过虚拟成像技术,探讨带飞行焦斑(FFS)的临床广角DBT系统数字乳房断层合成(DBT)和合成乳房x线摄影(SM)中角度范围和剂量分布对微钙化簇检测的影响。方法:从10例患者获得两倍自动暴露控制(AEC)剂量的DBT投影装置。将噪声添加到每个投影中以创建四个投影集:(1)25个AEC剂量投影(“AEC”),(2)25个凸剂量分布投影(“凸”),(3)25个中心三个投影剂量增加的投影(“峰3”),以及(4)19个覆盖40°角范围而不是标准50°角范围的投影(“40度”)。在所有设置中,总扫描剂量和角间距保持恒定。在这些投影集内模拟微钙化簇,然后使用DBT和SM补丁进行人类观察者检测研究。性能分析采用jackknife-alternative自由响应接收机工作特性(JAFROC)分析。结果:DBT的曲线下面积(AUC)分别为0.89±0.03 (AEC)、0.91±0.03(凸)、0.88±0.03(峰3)和0.91±0.04(40度)。SM的AUC值较低,分别为0.73±0.04、0.74±0.02、0.73±0.04和0.74±0.02。与AEC设置相比,两种方式均未观察到显著改善。结论:与FFS系统的标准AEC设置相比,增加中心投影剂量或将角度范围从50°减小到40°对DBT或SM的钙化检测没有显着影响。
{"title":"Optimization of microcalcification cluster detection in wide-angle flying focal spot digital breast tomosynthesis and synthetic mammography: A virtual imaging study","authors":"Katrien Houbrechts ,&nbsp;Astrid Van Camp ,&nbsp;Lesley Cockmartin ,&nbsp;Liesbeth Vancoillie ,&nbsp;Nicholas Marshall ,&nbsp;Hilde Bosmans","doi":"10.1016/j.ejmp.2025.105676","DOIUrl":"10.1016/j.ejmp.2025.105676","url":null,"abstract":"<div><h3>Purpose</h3><div>To investigate the impact of angular range and dose distribution on microcalcification cluster detection in digital breast tomosynthesis (DBT) and synthetic mammography (SM) of a clinical wide-angle DBT system with flying focal spot (FFS), through virtual imaging techniques.</div></div><div><h3>Approach</h3><div>DBT projection sets were acquired from ten patients at twice the automatic exposure controlled (AEC) dose. Noise was added to each projection to create four projection sets: (1) 25 projections at AEC dose (“AEC”), (2) 25 projections with a convex dose distribution (“convex”), (3) 25 projections with increased dose in the central three projections (“peak 3”), and (4) 19 projections covering a 40° angular range instead of the standard 50° (“40 degrees”). Total scan dose and angular spacing were maintained constant across all setups. Microcalcification clusters were simulated within these projection sets, followed by a human observer detection study with DBT and SM patches. Performance was analysed using jackknife-alternative free-response receiver operating characteristic (JAFROC) analysis.</div></div><div><h3>Results</h3><div>For DBT, the area under the curve (AUC) was 0.89 ± 0.03 (AEC), 0.91 ± 0.03 (convex), 0.88 ± 0.03 (peak 3), and 0.91 ± 0.04 (40 degrees). For SM, AUC values were lower: 0.73 ± 0.04, 0.74 ± 0.02, 0.73 ± 0.04, and 0.74 ± 0.02, respectively. No significant improvements were observed compared to the AEC setup for either modality.</div></div><div><h3>Conclusions</h3><div>Increasing the dose to the central projections or reducing the angular range from 50° to 40° did not significantly affect calcification detection in DBT or SM compared to the standard AEC setup for an FFS system.</div></div>","PeriodicalId":56092,"journal":{"name":"Physica Medica-European Journal of Medical Physics","volume":"140 ","pages":"Article 105676"},"PeriodicalIF":2.7,"publicationDate":"2025-11-17","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145551938","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Validation of DICOM extraction and Structuration Toolkit (DEST) for automated extraction and structuration of radiotherapy data for large-scale analysis 验证DICOM提取和结构化工具包(DEST)用于大规模分析放射治疗数据的自动提取和结构化
IF 2.7 3区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2025-11-15 DOI: 10.1016/j.ejmp.2025.105215
Ibrahima Diallo , Mamadou Hady Balde , Duyen Do , Véronique Letort , Thomas Ménard , Sarah Lemler , Guillaume Auzac , Anne Beaudré , Eric Deutsch , Nicolas Meillan , Thomas Sarrade , Anne-Laure Martin , Catherine Gaudin , Dominique Delmas , Antonio Di Meglio , Ines Vaz-Luis , Charlotte Robert , Florent de Vathaire , Sofia Rivera , Rodrigue S. Allodji

Purpose

Converting radiotherapy (RT) data from DICOM-RT into datasets suitable for statistical modelling remains challenging. We developed the DICOM Extraction and Structuration Toolkit (DEST), an automated solution that streamlines data extraction and ensures compatibility with statistical software. The reliability of DEST was also assessed by comparing its outputs with those from treatment planning systems (TPS) in study of cardiopulmonary dose-volume histograms (DVH) after radiotherapy (RT) for localised breast cancer.

Methods

DEST comprises two main modules: a data extraction module and a viewer/analysis module. It processes DICOM-RT objects, including RT structure sets, RT plans, and RT dose files. Extractions are performed per treatment for predefined patient lists, after which structured data is consolidated. A 3D visualisation module verifies dose distributions for selected regions of interest, ensuring consistency and accuracy.

Results

DEST was successfully applied to 404 patients from the “CANcer TOxicities – Radiation Therapy” (CANTO-RT) cohort. In this initial implementation, DEST showed strong overall agreement with regard to TPS for heart and lung dose metrics, including mean doses and dose-volume measurements. Specifically, near-minimum dose, median dose, near-maximum dose and percentages of volume receiving at least 10 Gy (V10Gy), 20 Gy (V20Gy), 30 Gy (V30Gy) and 40 Gy (V40Gy) showed high consistency between DEST and TPS.

Conclusions

DEST enhances accessibility to dose-volume metrics and will facilitate advanced modelling of medical outcomes (efficiency and risk) at the voxel level. By providing streamlined access to voxel spatial coordinates and local dose information, DEST enables more sophisticated analyses, such as clustering and localized region selection, supporting deeper insights into dose–response relationships.
目的将放射治疗(RT)数据从DICOM-RT转换为适合统计建模的数据集仍然具有挑战性。我们开发了DICOM提取和结构化工具包(DEST),这是一个自动化的解决方案,可以简化数据提取并确保与统计软件的兼容性。通过比较治疗计划系统(TPS)在局部乳腺癌放疗(RT)后心肺剂量-体积直方图(DVH)研究中的输出结果,也评估了DEST的可靠性。方法dest包括两个主要模块:数据提取模块和查看/分析模块。它处理DICOM-RT对象,包括RT结构集、RT计划和RT剂量文件。为预定义的患者列表执行每次治疗的提取,然后对结构化数据进行整合。3D可视化模块验证选定感兴趣区域的剂量分布,确保一致性和准确性。结果dest成功应用于404例“癌症毒性-放射治疗”(CANTO-RT)队列患者。在这项初步实施中,DEST在心肺剂量指标(包括平均剂量和剂量-体积测量)的TPS方面显示出强烈的总体一致性。其中,近最小剂量、中位剂量、近最大剂量和接受至少10Gy (V10Gy)、20Gy (V20Gy)、30Gy (V30Gy)和40Gy (V40Gy)的体积百分比在DEST和TPS之间表现出高度一致性。dest增强了剂量-体积指标的可及性,并将促进在体素水平上对医疗结果(效率和风险)进行高级建模。通过提供对体素空间坐标和局部剂量信息的简化访问,DEST可以进行更复杂的分析,例如聚类和局部区域选择,从而支持对剂量-反应关系的更深入了解。
{"title":"Validation of DICOM extraction and Structuration Toolkit (DEST) for automated extraction and structuration of radiotherapy data for large-scale analysis","authors":"Ibrahima Diallo ,&nbsp;Mamadou Hady Balde ,&nbsp;Duyen Do ,&nbsp;Véronique Letort ,&nbsp;Thomas Ménard ,&nbsp;Sarah Lemler ,&nbsp;Guillaume Auzac ,&nbsp;Anne Beaudré ,&nbsp;Eric Deutsch ,&nbsp;Nicolas Meillan ,&nbsp;Thomas Sarrade ,&nbsp;Anne-Laure Martin ,&nbsp;Catherine Gaudin ,&nbsp;Dominique Delmas ,&nbsp;Antonio Di Meglio ,&nbsp;Ines Vaz-Luis ,&nbsp;Charlotte Robert ,&nbsp;Florent de Vathaire ,&nbsp;Sofia Rivera ,&nbsp;Rodrigue S. Allodji","doi":"10.1016/j.ejmp.2025.105215","DOIUrl":"10.1016/j.ejmp.2025.105215","url":null,"abstract":"<div><h3>Purpose</h3><div>Converting radiotherapy (RT) data from DICOM-RT into datasets suitable for statistical modelling remains challenging. We developed the DICOM Extraction and Structuration Toolkit (DEST), an automated solution that streamlines data extraction and ensures compatibility with statistical software. The reliability of DEST was also assessed by comparing its outputs with those from treatment planning systems (TPS) in study of cardiopulmonary dose-volume histograms (DVH) after radiotherapy (RT) for localised breast cancer.</div></div><div><h3>Methods</h3><div>DEST comprises two main modules: a data extraction module and a viewer/analysis module. It processes DICOM-RT objects, including RT structure sets, RT plans, and RT dose files. Extractions are performed per treatment for predefined patient lists, after which structured data is consolidated. A 3D visualisation module verifies dose distributions for selected regions of interest, ensuring consistency and accuracy.</div></div><div><h3>Results</h3><div>DEST was successfully applied to 404 patients from the <em>“CANcer TOxicities – Radiation Therapy”</em> (CANTO-RT) cohort. In this initial implementation, DEST showed strong overall agreement with regard to TPS for heart and lung dose metrics, including mean doses and dose-volume measurements. Specifically, near-minimum dose, median dose, near-maximum dose and percentages of volume receiving at least 10 Gy (V<em><sub>10Gy</sub></em>), 20 Gy (V<em><sub>20Gy</sub></em>), 30 Gy (V<em><sub>30Gy</sub></em>) and 40 Gy (V<em><sub>40Gy</sub></em>) showed high consistency between DEST and TPS.</div></div><div><h3>Conclusions</h3><div>DEST enhances accessibility to dose-volume metrics and will facilitate advanced modelling of medical outcomes (efficiency and risk) at the voxel level. By providing streamlined access to voxel spatial coordinates and local dose information, DEST enables more sophisticated analyses, such as clustering and localized region selection, supporting deeper insights into dose–response relationships.</div></div>","PeriodicalId":56092,"journal":{"name":"Physica Medica-European Journal of Medical Physics","volume":"140 ","pages":"Article 105215"},"PeriodicalIF":2.7,"publicationDate":"2025-11-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145529204","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Evaluation of domain shift sources and generalisability in AI-based prostate MRI autocontouring for radiotherapy 基于人工智能的前列腺MRI放射治疗自动轮廓的域移源评价和通用性
IF 2.7 3区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2025-11-14 DOI: 10.1016/j.ejmp.2025.105188
Maram Alqarni , Emma-Louise Jones , Vinod Mullassery , Stephen Morris , Jorge Mariscal Harana , Esther Puyol Antón , Sian Cooper , Hema Verma , Teresa Guerrero Urbano , Andrew P. King

Aim

Deep learning (DL) models have been widely proposed to automate MRI-based delineation, but their use is hindered by differences in image characteristics between training and evaluation datasets (i.e. domain shift). This paper aims to (i) analyse the impacts of different sources of domain shift and (ii) externally evaluate a model trained using heterogeneous public data and compare it with an in-house model.

Methods

The nnU-Net DL framework was trained for prostate autocontouring using axial T2-weighted (T2W) prostate MRIs from five public datasets. By controlling training set size, three sources of domain shift were evaluated: dataset, scanner vendor/field strength, and image acquisition/annotation protocol. 66 prostate MRIs were used for external evaluation and training/evaluation of an in-house model. The Dice Similarity Coefficient (DSC) and 95 % Hausdorff distance (HD) evaluated the model-produced contours.

Results

The performance gap (Δ) between intra/inter-domain evaluation showed that domain shift from scanner vendor/field strength (ΔDSC = 0.33, Δ95 % HD = 246.86 mm) and image acquisition/annotation protocols (ΔDSC = 0.20, Δ95 % HD = 14.70 mm) had greater impact than that from dataset (ΔDSC = 0.06, Δ95 % HD = 3.69 mm), although all were significant (p < 0.05). External evaluation showed that the mixed-domain trained model performed well but was less robust than the in-house model (median (IQR) DSC/95 % HD = 0.87 (0.06)/3.75 (4.47)mm, 0.90 (0.03)/1.03 (1.29) mm, p < 0.05, respectively).

Conclusions

We highlight for the first time the domain shift effect of image acquisition/annotation protocol, even with images acquired using the same scanner vendor/field strength. Understanding the effect of multiple sources of domain shift has enabled us to train a robust model that can be safely clinically deployed.
深度学习(DL)模型已被广泛提出用于自动化基于mri的描绘,但它们的使用受到训练和评估数据集之间图像特征差异(即域移位)的阻碍。本文旨在(i)分析不同来源的领域转移的影响,(ii)外部评估使用异构公共数据训练的模型,并将其与内部模型进行比较。方法使用来自5个公共数据集的轴向t2加权(T2W)前列腺mri,训练nnU-Net DL框架进行前列腺自动轮廓。通过控制训练集的大小,评估了三个领域转移的来源:数据集、扫描仪供应商/场强和图像采集/注释协议。66台前列腺核磁共振成像用于外部评估和内部模型的培训/评估。骰子相似系数(DSC)和95%豪斯多夫距离(HD)评估模型产生的轮廓。结果域内/域间评价的性能差距(Δ)表明,扫描仪供应商/场强(ΔDSC = 0.33, Δ95 % HD = 246.86 mm)和图像采集/注释协议(ΔDSC = 0.20, Δ95 % HD = 14.70 mm)的域偏移比数据集(ΔDSC = 0.06, Δ95 % HD = 3.69 mm)的域偏移影响更大,但均显著(p < 0.05)。外部评价表明,混合域训练模型表现良好,但鲁棒性不如内部模型(中位数(IQR) DSC/ 95% HD = 0.87 (0.06)/3.75 (4.47)mm, 0.90 (0.03)/1.03 (1.29) mm, p < 0.05)。结论我们首次强调了图像采集/注释协议的域移位效应,即使使用相同的扫描仪供应商/场强获取的图像。了解多个域移位源的影响,使我们能够训练一个可以安全临床部署的鲁棒模型。
{"title":"Evaluation of domain shift sources and generalisability in AI-based prostate MRI autocontouring for radiotherapy","authors":"Maram Alqarni ,&nbsp;Emma-Louise Jones ,&nbsp;Vinod Mullassery ,&nbsp;Stephen Morris ,&nbsp;Jorge Mariscal Harana ,&nbsp;Esther Puyol Antón ,&nbsp;Sian Cooper ,&nbsp;Hema Verma ,&nbsp;Teresa Guerrero Urbano ,&nbsp;Andrew P. King","doi":"10.1016/j.ejmp.2025.105188","DOIUrl":"10.1016/j.ejmp.2025.105188","url":null,"abstract":"<div><h3>Aim</h3><div>Deep learning (DL) models have been widely proposed to automate MRI-based delineation, but their use is hindered by differences in image characteristics between training and evaluation datasets (i.e. domain shift). This paper aims to (i) analyse the impacts of different sources of domain shift and (ii) externally evaluate a model trained using heterogeneous public data and compare it with an in-house model.</div></div><div><h3>Methods</h3><div>The nnU-Net DL framework was trained for prostate autocontouring using axial T2-weighted (T2W) prostate MRIs from five public datasets. By controlling training set size, three sources of domain shift were evaluated: dataset, scanner vendor/field strength, and image acquisition/annotation protocol. 66 prostate MRIs were used for external evaluation and training/evaluation of an in-house model. The Dice Similarity Coefficient (DSC) and 95 % Hausdorff distance (HD) evaluated the model-produced contours.</div></div><div><h3>Results</h3><div>The performance gap (Δ) between intra/inter-domain evaluation showed that domain shift from scanner vendor/field strength (ΔDSC = 0.33, Δ95 % HD = 246.86 mm) and image acquisition/annotation protocols (ΔDSC = 0.20, Δ95 % HD = 14.70 mm) had greater impact than that from dataset (ΔDSC = 0.06, Δ95 % HD = 3.69 mm), although all were significant (p &lt; 0.05). External evaluation showed that the mixed-domain trained model performed well but was less robust than the in-house model (median (IQR) DSC/95 % HD = 0.87 (0.06)/3.75 (4.47)mm, 0.90 (0.03)/1.03 (1.29) mm, p &lt; 0.05, respectively).</div></div><div><h3>Conclusions</h3><div>We highlight for the first time the domain shift effect of image acquisition/annotation protocol, even with images acquired using the same scanner vendor/field strength. Understanding the effect of multiple sources of domain shift has enabled us to train a robust model that can be safely clinically deployed.</div></div>","PeriodicalId":56092,"journal":{"name":"Physica Medica-European Journal of Medical Physics","volume":"140 ","pages":"Article 105188"},"PeriodicalIF":2.7,"publicationDate":"2025-11-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145529249","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Toward the use of Silicon Carbide based detector for protontherapy microdosimetry 碳化硅基探测器在质子治疗微剂量测定中的应用。
IF 2.7 3区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2025-11-10 DOI: 10.1016/j.ejmp.2025.105675
G. Petringa , C. Verona , R. Catalano , M. Guarrera , A. Kurmanova , L. Brighel , A. Sciuto , S. Tudisco , G.A.P. Cirrone

Purpose:

This study explores the potential of a Silicon Carbide (SiC)-based microdosimeter for use in proton therapy. Thanks to its high radiation hardness, minimal leakage current, and stable charge collection efficiency, SiC is proposed as an innovative solid-state material for measuring microdosimetric quantities and enhancing radiobiological modeling in hadrontherapy.

Methods:

A custom SiC p–n junction diode with a 10 μm epitaxial layer was fabricated and electrically characterized. Microdosimetric spectra were acquired under proton beams at 30, 70, and 150 MeV at various depths in water. The SiC detector’s performance was compared with benchmark microdosimeters, including a silicon-based MicroPlus probe and a single-crystal diamond device. Monte Carlo simulations were used to determine the water-equivalent thickness and to validate the measured spectra. Key microdosimetric quantities, such as frequency-mean and dose-mean lineal energy, were extracted and used to estimate the relative biological effectiveness (RBE) via the Microdosimetric Kinetic Model (MKM).

Results:

The SiC detector demonstrated full charge collection efficiency and an energy resolution of approximately 2% under alpha irradiation. The measured spectra showed depth-dependent trends consistent with LET variations. The microdosimetric parameters derived from SiC data were in good agreement with both reference detectors and simulations. RBE values estimated from SiC measurements accurately reflected experimental data for U87 glioblastoma cells.

Conclusion:

SiC-based detectors prove to be promising tools for microdosimetry in proton therapy. The consistency with reference systems and simulations supports their use in RBE estimation and biologically optimized treatment planning. Future work will target further miniaturization and application to heavier ions.
目的:本研究探讨了基于碳化硅(SiC)的微剂量计用于质子治疗的潜力。由于其高辐射硬度,最小的泄漏电流和稳定的电荷收集效率,SiC被提出作为一种创新的固态材料,用于测量微剂量量和增强强子治疗中的放射生物学建模。方法:制作具有10 μm外延层的SiC p-n结二极管,并对其进行电学表征。在30、70和150 MeV的质子束下,在不同深度的水中获得了微剂量学光谱。SiC探测器的性能与基准微剂量计(包括硅基MicroPlus探针和单晶金刚石器件)进行了比较。蒙特卡罗模拟用于确定水当量厚度和验证测量光谱。提取关键的微剂量学量,如频率-平均和剂量-平均线性能量,并通过微剂量动力学模型(MKM)估计相对生物有效性(RBE)。结果:在α辐照下,SiC探测器具有完全的电荷收集效率和约2%的能量分辨率。实测光谱显示出与LET变化相一致的随深度变化趋势。从SiC数据中得到的微剂量学参数与参考探测器和模拟结果一致。从SiC测量中估计的RBE值准确地反映了U87胶质母细胞瘤细胞的实验数据。结论:硅基探测器被证明是质子治疗中有前途的微剂量测定工具。与参考系统和模拟的一致性支持它们在RBE估计和生物优化处理计划中的使用。未来的工作将进一步小型化和应用于更重的离子。
{"title":"Toward the use of Silicon Carbide based detector for protontherapy microdosimetry","authors":"G. Petringa ,&nbsp;C. Verona ,&nbsp;R. Catalano ,&nbsp;M. Guarrera ,&nbsp;A. Kurmanova ,&nbsp;L. Brighel ,&nbsp;A. Sciuto ,&nbsp;S. Tudisco ,&nbsp;G.A.P. Cirrone","doi":"10.1016/j.ejmp.2025.105675","DOIUrl":"10.1016/j.ejmp.2025.105675","url":null,"abstract":"<div><h3>Purpose:</h3><div>This study explores the potential of a Silicon Carbide (SiC)-based microdosimeter for use in proton therapy. Thanks to its high radiation hardness, minimal leakage current, and stable charge collection efficiency, SiC is proposed as an innovative solid-state material for measuring microdosimetric quantities and enhancing radiobiological modeling in hadrontherapy.</div></div><div><h3>Methods:</h3><div>A custom SiC p–n junction diode with a 10 <span><math><mi>μ</mi></math></span>m epitaxial layer was fabricated and electrically characterized. Microdosimetric spectra were acquired under proton beams at 30, 70, and 150 MeV at various depths in water. The SiC detector’s performance was compared with benchmark microdosimeters, including a silicon-based MicroPlus probe and a single-crystal diamond device. Monte Carlo simulations were used to determine the water-equivalent thickness and to validate the measured spectra. Key microdosimetric quantities, such as frequency-mean and dose-mean lineal energy, were extracted and used to estimate the relative biological effectiveness (RBE) via the Microdosimetric Kinetic Model (MKM).</div></div><div><h3>Results:</h3><div>The SiC detector demonstrated full charge collection efficiency and an energy resolution of approximately 2% under alpha irradiation. The measured spectra showed depth-dependent trends consistent with LET variations. The microdosimetric parameters derived from SiC data were in good agreement with both reference detectors and simulations. RBE values estimated from SiC measurements accurately reflected experimental data for U87 glioblastoma cells.</div></div><div><h3>Conclusion:</h3><div>SiC-based detectors prove to be promising tools for microdosimetry in proton therapy. The consistency with reference systems and simulations supports their use in RBE estimation and biologically optimized treatment planning. Future work will target further miniaturization and application to heavier ions.</div></div>","PeriodicalId":56092,"journal":{"name":"Physica Medica-European Journal of Medical Physics","volume":"140 ","pages":"Article 105675"},"PeriodicalIF":2.7,"publicationDate":"2025-11-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145497564","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Patterns of practice of image guided particle therapy for brain tumours: A site specific multi-institutional survey of the European particle therapy network 脑肿瘤图像引导粒子治疗的实践模式:欧洲粒子治疗网络的特定地点多机构调查。
IF 2.7 3区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2025-11-09 DOI: 10.1016/j.ejmp.2025.105211
Iuliana Toma-Dasu , Alexandru Dasu , Dante Amelio , Lorenzo Placidi , Juliette Thariat , Markus Stock , Petra Trnková , Daniëlle Eekers , Anneleen Goedgebeur , Morten Høyer , Aswin Hoffmann , Alessandra Bolsi

Purpose

To investigate the current practice patterns in image-guided proton therapy (IGPT) for brain tumours.

Methods

A multi-institutional survey was distributed to European particle therapy centres to analyse the current practice of IGPT for neuro-oncology. The survey was subsequently used for driving a DELPHI consensus analysis aiming at defining the minimum requirements and the optimal workflow.

Results

Seven centres participated in the survey on proton therapy for brain tumours. All reported access to pencil beam scanning and rotating gantries; one also used passive scattering. Supine positioning with standard immobilisation tools was common, while prone and paediatric-specific methods were rare. Multimodal imaging with CT and MRI was standard; PET use was limited and SPECT absent. Rigid registration between imaging modalities was widely used, though MR imaging in treatment position was uncommon. Verification practices varied. Six centres joined the DELPHI consensus, reaching agreement on minimum requirements for immobilisation, imaging for treatment planning, image registration and pre-treatment setup. Disagreement remained on robustness criteria, imaging frequency, and dose tracking, highlighting the need for unified clinical guidelines and workflow optimisation.

Conclusion

There is generally agreement across European proton centres, but variability remained in key components of treatment planning, verification and workflow optimisation, including the frequency and modality of control imaging, plan robustness criteria, and treatment position imaging protocols. These differences reflect both local resource availability and the absence of harmonised guidelines. The minimal requirements for image guidance in brain proton therapy achieved good consensus level and will be very useful for new centres.
目的:探讨图像引导质子治疗(IGPT)治疗脑肿瘤的实践模式。方法:对欧洲粒子治疗中心进行多机构调查,分析目前IGPT治疗神经肿瘤学的实践。该调查随后被用于驱动DELPHI共识分析,旨在定义最低需求和最佳工作流程。结果:有7个中心参与了脑肿瘤质子治疗的调查。所有报告都使用了铅笔束扫描和旋转龙门;一种方法是被动散射。使用标准固定工具的仰卧位很常见,而俯卧位和儿科专用方法则很少见。标准的CT和MRI多模态成像;PET使用受限,SPECT缺失。成像模式之间的刚性配准被广泛使用,尽管MR成像在治疗位置并不常见。验证实践各不相同。六个中心加入了DELPHI共识,就固定、治疗计划成像、图像配准和治疗前设置的最低要求达成一致。在稳健性标准、成像频率和剂量跟踪方面仍然存在分歧,强调需要统一的临床指南和工作流程优化。结论:欧洲质子中心的观点普遍一致,但在治疗计划、验证和工作流程优化的关键组成部分仍然存在差异,包括控制成像的频率和方式、计划稳健性标准和治疗位置成像协议。这些差异既反映了当地资源的可用性,也反映了缺乏统一的指导方针。脑质子治疗中影像引导的最低要求达到了良好的共识水平,对新中心非常有用。
{"title":"Patterns of practice of image guided particle therapy for brain tumours: A site specific multi-institutional survey of the European particle therapy network","authors":"Iuliana Toma-Dasu ,&nbsp;Alexandru Dasu ,&nbsp;Dante Amelio ,&nbsp;Lorenzo Placidi ,&nbsp;Juliette Thariat ,&nbsp;Markus Stock ,&nbsp;Petra Trnková ,&nbsp;Daniëlle Eekers ,&nbsp;Anneleen Goedgebeur ,&nbsp;Morten Høyer ,&nbsp;Aswin Hoffmann ,&nbsp;Alessandra Bolsi","doi":"10.1016/j.ejmp.2025.105211","DOIUrl":"10.1016/j.ejmp.2025.105211","url":null,"abstract":"<div><h3>Purpose</h3><div>To investigate the current practice patterns in image-guided proton therapy (IGPT) for brain tumours.</div></div><div><h3>Methods</h3><div>A multi-institutional survey was distributed to European particle therapy centres to analyse the current practice of IGPT for neuro-oncology. The survey was subsequently used for driving a DELPHI consensus analysis aiming at defining the minimum requirements and the optimal workflow.</div></div><div><h3>Results</h3><div>Seven centres participated in the survey on proton therapy for brain tumours. All reported access to pencil beam scanning and rotating gantries; one also used passive scattering. Supine positioning with standard immobilisation tools was common, while prone and paediatric-specific methods were rare. Multimodal imaging with CT and MRI was standard; PET use was limited and SPECT absent. Rigid registration between imaging modalities was widely used, though MR imaging in treatment position was uncommon. Verification practices varied. Six centres joined the DELPHI consensus, reaching agreement on minimum requirements for immobilisation, imaging for treatment planning, image registration and pre-treatment setup. Disagreement remained on robustness criteria, imaging frequency, and dose tracking, highlighting the need for unified clinical guidelines and workflow optimisation.</div></div><div><h3>Conclusion</h3><div>There is generally agreement across European proton centres, but variability remained in key components of treatment planning, verification and workflow optimisation, including the frequency and modality of control imaging, plan robustness criteria, and treatment position imaging protocols. These differences reflect both local resource availability and the absence of harmonised guidelines. The minimal requirements for image guidance in brain proton therapy achieved good consensus level and will be very useful for new centres.</div></div>","PeriodicalId":56092,"journal":{"name":"Physica Medica-European Journal of Medical Physics","volume":"140 ","pages":"Article 105211"},"PeriodicalIF":2.7,"publicationDate":"2025-11-09","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145491015","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Increasing the use of prospective risk assessment through training 通过培训增加前瞻性风险评估的使用
IF 2.7 3区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2025-11-06 DOI: 10.1016/j.ejmp.2025.105213
Elizabeth Claridge Mackonis , Anna Ralston , Johnson Yuen , Nicholas Hardcastle , Annette Haworth

Purpose

Whilst a formal prospective risk assessment is an important part of risk management in the implementation of new techniques and technologies, a lack of time and practical knowledge can prevent its use. Here we present an innovative training program aiming to increase the knowledge and practical skills of radiation oncology professionals in using prospective risk assessment.

Methods

The training program explained the need for prospective risk assessment and provided practical experience in the use of an easy-to-use prospective risk assessment and project management tool. Participants were surveyed before the training and 12-months post-training to determine current risk assessment processes and whether the training affected participants views and behaviours in this area.

Results

A time-efficient training program in prospective risk assessment was created and successful delivered to multi-disciplinary groups from 12 radiation therapy providers. In the 12-months post-training, the number of centres completing prospective risk assessments increased from 8% to 30%. The training removed many of the perceived barriers, with insufficient time remaining the largest barrier.

Conclusions

The training was found to be useful and applicable by all respondents. Future work needs to focus on how to expand the reach of the training program while maintaining its success, and how to ensure prospective risk assessments are seen as an essential part of implementation of new techniques and technologies in radiation therapy centres.
虽然正式的前瞻性风险评估是实施新技术和新技术风险管理的重要组成部分,但缺乏时间和实用知识可能会阻碍其使用。在这里,我们提出了一个创新的培训计划,旨在提高放射肿瘤学专业人员在使用前瞻性风险评估方面的知识和实践技能。方法通过培训项目,说明前瞻性风险评估的必要性,并提供使用易于使用的前瞻性风险评估和项目管理工具的实践经验。在培训前和培训后12个月对参与者进行了调查,以确定当前的风险评估过程,以及培训是否影响了参与者在这方面的观点和行为。结果建立了一套高效的前瞻性风险评估培训方案,并成功地对来自12家放射治疗提供者的多学科小组进行了培训。在培训后的12个月里,完成前瞻性风险评估的中心数量从8%增加到30%。训练消除了许多感知到的障碍,时间不足仍然是最大的障碍。结论所有被调查者都认为培训是有用的和适用的。今后的工作需要侧重于如何在保持培训计划成功的同时扩大培训计划的范围,以及如何确保将前瞻性风险评估视为在放射治疗中心实施新技术和技术的重要组成部分。
{"title":"Increasing the use of prospective risk assessment through training","authors":"Elizabeth Claridge Mackonis ,&nbsp;Anna Ralston ,&nbsp;Johnson Yuen ,&nbsp;Nicholas Hardcastle ,&nbsp;Annette Haworth","doi":"10.1016/j.ejmp.2025.105213","DOIUrl":"10.1016/j.ejmp.2025.105213","url":null,"abstract":"<div><h3>Purpose</h3><div>Whilst a formal prospective risk assessment is an important part of risk management in the implementation of new techniques and technologies, a lack of time and practical knowledge can prevent its use. Here we present an innovative training program aiming to increase the knowledge and practical skills of radiation oncology professionals in using prospective risk assessment.</div></div><div><h3>Methods</h3><div>The training program explained the need for prospective risk assessment and provided practical experience in the use of an easy-to-use prospective risk assessment and project management tool. Participants were surveyed before the training and 12-months post-training to determine current risk assessment processes and whether the training affected participants views and behaviours in this area.</div></div><div><h3>Results</h3><div>A time-efficient training program in prospective risk assessment was created and successful delivered to multi-disciplinary groups from 12 radiation therapy providers. In the 12-months post-training, the number of centres completing prospective risk assessments increased from 8% to 30%. The training removed many of the perceived barriers, with insufficient time remaining the largest barrier.</div></div><div><h3>Conclusions</h3><div>The training was found to be useful and applicable by all respondents. Future work needs to focus on how to expand the reach of the training program while maintaining its success, and how to ensure prospective risk assessments are seen as an essential part of implementation of new techniques and technologies in radiation therapy centres.</div></div>","PeriodicalId":56092,"journal":{"name":"Physica Medica-European Journal of Medical Physics","volume":"140 ","pages":"Article 105213"},"PeriodicalIF":2.7,"publicationDate":"2025-11-06","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145442562","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Corrigendum to “Ultra-high dose rate dosimetry for pre-clinical experiments with mm-small proton fields” [Phys. Med. 104 (2022) 101–111] “极小质子场临床前实验的超高剂量率剂量学”的勘误表[物理学。中华医学杂志,2010(5):391 - 391。
IF 2.7 3区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2025-11-01 DOI: 10.1016/j.ejmp.2025.105100
M. Togno , K.P. Nesteruk , R. Schäfer , S. Psoroulas , D. Meer , M. Grossmann , J.B. Christensen , E.G. Yukihara , A.J. Lomax , S. Safai
{"title":"Corrigendum to “Ultra-high dose rate dosimetry for pre-clinical experiments with mm-small proton fields” [Phys. Med. 104 (2022) 101–111]","authors":"M. Togno ,&nbsp;K.P. Nesteruk ,&nbsp;R. Schäfer ,&nbsp;S. Psoroulas ,&nbsp;D. Meer ,&nbsp;M. Grossmann ,&nbsp;J.B. Christensen ,&nbsp;E.G. Yukihara ,&nbsp;A.J. Lomax ,&nbsp;S. Safai","doi":"10.1016/j.ejmp.2025.105100","DOIUrl":"10.1016/j.ejmp.2025.105100","url":null,"abstract":"","PeriodicalId":56092,"journal":{"name":"Physica Medica-European Journal of Medical Physics","volume":"139 ","pages":"Article 105100"},"PeriodicalIF":2.7,"publicationDate":"2025-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145031336","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Evaluating the accuracy of SPECT/CT LSF estimations in SIRT therapies using Monte Carlo simulations with virtual 4D anthropomorphic phantoms 利用蒙特卡罗模拟与虚拟四维拟人化幻象评估SIRT治疗中SPECT/CT LSF估计的准确性
IF 2.7 3区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2025-11-01 DOI: 10.1016/j.ejmp.2025.105196
Niamh McArdle , Jackie McCavana , Darragh McCague , Seán Cournane , Luis León Vintró

Purpose

Recent publications advocate the use of SPECT/CT for LSF estimation for more accurate lung dose estimations in Selective Internal Radiation Therapy (SIRT) treatments. These studies have been limited to patient studies where the ground truth is not known or to a small number of phantom studies of a single body size. This research aimed to investigate the accuracy of LSF estimation derived using a Monte Carlo SPECT/CT model incorporating 4D virtual phantoms of varying BMIs.

Methods

Six 4D virtual XCAT phantoms were used to model patients of varying BMIs. SPECT acquisitions were simulated and assessed using the SIMIND Monte Carlo modelling framework. LSF estimates derived from SPECT imaging were evaluated and subsequently compared to corrected planar LSF estimates.

Results

SPECT LSF estimates which incorporated scatter correction and with a CT based attenuation correction were shown to be the most accurate. Moreover, the SPECT LSF estimates were shown to be comparable with planar LSF estimation with incorporated scatter correction.

Conclusion

Given that the LSF estimate is a critical input in calculating the absorbed dose to the target volume in SIRT treatments, errors in its estimation can propagate through the dosimetric workflow. This work showed that both the SPECT/CT and the planar LSF estimation with scatter correction methods can significantly improve the LSF accuracy, over the current geometric mean method, and are easily implementable into the clinical workflow.
最近的出版物提倡使用SPECT/CT进行LSF估计,以便在选择性内放射治疗(SIRT)治疗中更准确地估计肺剂量。这些研究局限于不知道基本真相的病人研究或少数单一体型的虚幻研究。本研究旨在探讨利用蒙特卡罗SPECT/CT模型结合不同bmi的四维虚拟幻影得出的LSF估计的准确性。方法采用6个4D虚拟XCAT模型对不同bmi的患者进行建模。使用SIMIND蒙特卡罗建模框架对SPECT采集进行模拟和评估。评估从SPECT成像得到的LSF估计值,并随后与校正后的平面LSF估计值进行比较。结果结合散射校正和基于CT的衰减校正的spect LSF估计是最准确的。此外,SPECT LSF估计与纳入散点校正的平面LSF估计具有可比性。鉴于LSF估计是计算SIRT治疗中靶体积吸收剂量的关键输入,其估计误差可能会在剂量学工作流程中传播。研究表明,与现有的几何平均方法相比,SPECT/CT和平面LSF估计的散点校正方法都能显著提高LSF的精度,并且易于在临床工作流程中实现。
{"title":"Evaluating the accuracy of SPECT/CT LSF estimations in SIRT therapies using Monte Carlo simulations with virtual 4D anthropomorphic phantoms","authors":"Niamh McArdle ,&nbsp;Jackie McCavana ,&nbsp;Darragh McCague ,&nbsp;Seán Cournane ,&nbsp;Luis León Vintró","doi":"10.1016/j.ejmp.2025.105196","DOIUrl":"10.1016/j.ejmp.2025.105196","url":null,"abstract":"<div><h3>Purpose</h3><div>Recent publications advocate the use of SPECT/CT for LSF estimation for more accurate lung dose estimations in Selective Internal Radiation Therapy (SIRT) treatments. These studies have been limited to patient studies where the ground truth is not known or to a small number of phantom studies of a single body size. This research aimed to investigate the accuracy of LSF estimation derived using a Monte Carlo SPECT/CT model incorporating 4D virtual phantoms of varying BMIs.</div></div><div><h3>Methods</h3><div>Six 4D virtual XCAT phantoms were used to model patients of varying BMIs. SPECT acquisitions were simulated and assessed using the SIMIND Monte Carlo modelling framework. LSF estimates derived from SPECT imaging were evaluated and subsequently compared to corrected planar LSF estimates.</div></div><div><h3>Results</h3><div>SPECT LSF estimates which incorporated scatter correction and with a CT based attenuation correction were shown to be the most accurate. Moreover, the SPECT LSF estimates were shown to be comparable with planar LSF estimation with incorporated scatter correction.</div></div><div><h3>Conclusion</h3><div>Given that the LSF estimate is a critical input in calculating the absorbed dose to the target volume in SIRT treatments, errors in its estimation can propagate through the dosimetric workflow. This work showed that both the SPECT/CT and the planar LSF estimation with scatter correction methods can significantly improve the LSF accuracy, over the current geometric mean method, and are easily implementable into the clinical workflow.</div></div>","PeriodicalId":56092,"journal":{"name":"Physica Medica-European Journal of Medical Physics","volume":"139 ","pages":"Article 105196"},"PeriodicalIF":2.7,"publicationDate":"2025-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145417696","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Corrigendum to “Evaluation of cumulative radiation burden and associated risk in smouldering multiple myeloma patients monitored using recurrent whole-body low-dose CT imaging”. [Phys. Med. 138 (2025) 105173] “评估反复全身低剂量CT成像监测的阴燃型多发性骨髓瘤患者的累积辐射负担和相关风险”的勘误表。(理论物理。医学杂志。138(2025):105173]。
IF 2.7 3区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2025-11-01 DOI: 10.1016/j.ejmp.2025.105212
Pascaline Wangechi Abissi , Georgia Lymperopoulou , Vasileios-Petros Agrokostas , Vassilis Koutoulidis , Panagiotis Grigoropoulos , Maria Gavriatopoulou , Evangelos Terpos , Lia Angela Moulopoulos , Martin Fiebich , Ioannis Seimenis
{"title":"Corrigendum to “Evaluation of cumulative radiation burden and associated risk in smouldering multiple myeloma patients monitored using recurrent whole-body low-dose CT imaging”. [Phys. Med. 138 (2025) 105173]","authors":"Pascaline Wangechi Abissi ,&nbsp;Georgia Lymperopoulou ,&nbsp;Vasileios-Petros Agrokostas ,&nbsp;Vassilis Koutoulidis ,&nbsp;Panagiotis Grigoropoulos ,&nbsp;Maria Gavriatopoulou ,&nbsp;Evangelos Terpos ,&nbsp;Lia Angela Moulopoulos ,&nbsp;Martin Fiebich ,&nbsp;Ioannis Seimenis","doi":"10.1016/j.ejmp.2025.105212","DOIUrl":"10.1016/j.ejmp.2025.105212","url":null,"abstract":"","PeriodicalId":56092,"journal":{"name":"Physica Medica-European Journal of Medical Physics","volume":"139 ","pages":"Article 105212"},"PeriodicalIF":2.7,"publicationDate":"2025-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145402898","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Insights, robustness and practical considerations of global noise level measurement in chest CT 胸部CT全局噪声级测量的见解、鲁棒性和实际考虑
IF 2.7 3区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2025-11-01 DOI: 10.1016/j.ejmp.2025.105207
Kwinten Torfs , Dimitar Petrov , Louise D’hondt , Tim Busselot , Janne Vignero , Walter De Wever , Annemiek Snoeckx , Klaus Bacher , Hilde Bosmans

Background

Accuracy of global noise level (GNL) has been benchmarked predominantly for soft tissue. This work extends the use of GNL to lung tissue and investigates the effect of patient, protocol and method parameters on the robustness of the algorithm.

Procedures

Optimal GNL algorithm parameters for lung tissue were established after comparing manual measurements in one hundred patient images. Automatic determination of patient-specific GNL was performed from all chest CT slices with outlier verifications and lung tissue area estimation. Accuracy of GNL mean and variation were assessed with increasing number of included slices to convert GNLs into a stack-overarching parameter. Finally, GNL in soft and lung tissue were compared in identical patient images.

Main findings

Optimal GNL parameters for lung tissue were: 4 mm x 4 mm sliding window, Hounsfield units below −590 HU, and 1 HU histogram bin width. In chest CT slices with a lung tissue area above 40 cm2, 99.8 % of GNL calculations were clinically relevant. A 95 % accurate mean lung GNL required minimally 17 equidistant slices. Finally, lung GNL in thicker slices was higher than soft tissue GNL, but the reverse was observed in thin slices.

Conclusion

This study provides insight into important practical considerations on the calculation and subsequent use of GNL for monitoring and image quality evaluation.
全球噪声水平(GNL)的精度主要针对软组织进行基准测试。这项工作将GNL的使用扩展到肺组织,并研究了患者、方案和方法参数对算法鲁棒性的影响。通过比较100张患者图像的人工测量值,建立肺组织的最佳GNL算法参数。通过离群值验证和肺组织面积估计,对所有胸部CT切片自动确定患者特异性GNL。随着切片数量的增加,对GNL均值和变异的精度进行了评估,将GNL转化为叠加参数。最后,在相同的患者图像中比较软组织和肺组织的GNL。肺组织的最佳GNL参数为:4 mm × 4 mm滑动窗,Hounsfield单位低于- 590 HU,直方图bin宽度为1 HU。在肺组织面积大于40 cm2的胸部CT切片中,99.8%的GNL计算具有临床相关性。95%准确率的平均肺GNL至少需要17片等距切片。最后,厚片肺GNL高于软组织GNL,薄片肺GNL则相反。本研究为GNL在监测和图像质量评估中的计算和后续使用提供了重要的实际考虑。
{"title":"Insights, robustness and practical considerations of global noise level measurement in chest CT","authors":"Kwinten Torfs ,&nbsp;Dimitar Petrov ,&nbsp;Louise D’hondt ,&nbsp;Tim Busselot ,&nbsp;Janne Vignero ,&nbsp;Walter De Wever ,&nbsp;Annemiek Snoeckx ,&nbsp;Klaus Bacher ,&nbsp;Hilde Bosmans","doi":"10.1016/j.ejmp.2025.105207","DOIUrl":"10.1016/j.ejmp.2025.105207","url":null,"abstract":"<div><h3>Background</h3><div>Accuracy of global noise level (GNL) has been benchmarked predominantly for soft tissue. This work extends the use of GNL to lung tissue and investigates the effect of patient, protocol and method parameters on the robustness of the algorithm.</div></div><div><h3>Procedures</h3><div>Optimal GNL algorithm parameters for lung tissue were established after comparing manual measurements in one hundred patient images. Automatic determination of patient-specific GNL was performed from all chest CT slices with outlier verifications and lung tissue area estimation. Accuracy of GNL mean and variation were assessed with increasing number of included slices to convert GNLs into a stack-overarching parameter. Finally, GNL in soft and lung tissue were compared in identical patient images.</div></div><div><h3>Main findings</h3><div>Optimal GNL parameters for lung tissue were: 4 mm x 4 mm sliding window, Hounsfield units below −590 HU, and 1 HU histogram bin width. In chest CT slices with a lung tissue area above 40 cm<sup>2</sup>, 99.8 % of GNL calculations were clinically relevant. A 95 % accurate mean lung GNL required minimally 17 equidistant slices. Finally, lung GNL in thicker slices was higher than soft tissue GNL, but the reverse was observed in thin slices.</div></div><div><h3>Conclusion</h3><div>This study provides insight into important practical considerations on the calculation and subsequent use of GNL for monitoring and image quality evaluation.</div></div>","PeriodicalId":56092,"journal":{"name":"Physica Medica-European Journal of Medical Physics","volume":"139 ","pages":"Article 105207"},"PeriodicalIF":2.7,"publicationDate":"2025-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145417697","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
期刊
Physica Medica-European Journal of Medical Physics
全部 Acc. Chem. Res. ACS Applied Bio Materials ACS Appl. Electron. Mater. ACS Appl. Energy Mater. ACS Appl. Mater. Interfaces ACS Appl. Nano Mater. ACS Appl. Polym. Mater. ACS BIOMATER-SCI ENG ACS Catal. ACS Cent. Sci. ACS Chem. Biol. ACS Chemical Health & Safety ACS Chem. Neurosci. ACS Comb. Sci. ACS Earth Space Chem. ACS Energy Lett. ACS Infect. Dis. ACS Macro Lett. ACS Mater. Lett. ACS Med. Chem. Lett. ACS Nano ACS Omega ACS Photonics ACS Sens. ACS Sustainable Chem. Eng. ACS Synth. Biol. Anal. Chem. BIOCHEMISTRY-US Bioconjugate Chem. BIOMACROMOLECULES Chem. Res. Toxicol. Chem. Rev. Chem. Mater. CRYST GROWTH DES ENERG FUEL Environ. Sci. Technol. Environ. Sci. Technol. Lett. Eur. J. Inorg. Chem. IND ENG CHEM RES Inorg. Chem. J. Agric. Food. Chem. J. Chem. Eng. Data J. Chem. Educ. J. Chem. Inf. Model. J. Chem. Theory Comput. J. Med. Chem. J. Nat. Prod. J PROTEOME RES J. Am. Chem. Soc. LANGMUIR MACROMOLECULES Mol. Pharmaceutics Nano Lett. Org. Lett. ORG PROCESS RES DEV ORGANOMETALLICS J. Org. Chem. J. Phys. Chem. J. Phys. Chem. A J. Phys. Chem. B J. Phys. Chem. C J. Phys. Chem. Lett. Analyst Anal. Methods Biomater. Sci. Catal. Sci. Technol. Chem. Commun. Chem. Soc. Rev. CHEM EDUC RES PRACT CRYSTENGCOMM Dalton Trans. Energy Environ. Sci. ENVIRON SCI-NANO ENVIRON SCI-PROC IMP ENVIRON SCI-WAT RES Faraday Discuss. Food Funct. Green Chem. Inorg. Chem. Front. Integr. Biol. J. Anal. At. Spectrom. J. Mater. Chem. A J. Mater. Chem. B J. Mater. Chem. C Lab Chip Mater. Chem. Front. Mater. Horiz. MEDCHEMCOMM Metallomics Mol. Biosyst. Mol. Syst. Des. Eng. Nanoscale Nanoscale Horiz. Nat. Prod. Rep. New J. Chem. Org. Biomol. Chem. Org. Chem. Front. PHOTOCH PHOTOBIO SCI PCCP Polym. Chem.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1