首页 > 最新文献

Investigative Radiology最新文献

英文 中文
Countering Calcium Blooming With Personalized Contrast Media Injection Protocols: The 1-2-3 Rule for Photon-Counting Detector CCTA. 用个性化的造影剂注射方案对抗钙盛行:光子计数探测器 CCTA 的 1-2-3 规则。
IF 7 1区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2024-10-01 Epub Date: 2024-05-15 DOI: 10.1097/RLI.0000000000001078
Michael C McDermott, Thomas Sartoretti, Lion Stammen, Bibi Martens, Gregor Jost, Hubertus Pietsch, Ralf Gutjahr, Bernhard Schmidt, Thomas G Flohr, Hatem Alkadhi, Joachim E Wildberger

Objective: Photon-counting detector computed tomography (PCD-CT) enables spectral data acquisition of CT angiographies allowing for reconstruction of virtual monoenergetic images (VMIs) in routine practice. Specifically, it has potential to reduce the blooming artifacts associated with densely calcified plaques. However, calcium blooming and iodine attenuation are inversely affected by energy level (keV) of the VMIs, creating a challenge for contrast media (CM) injection protocol optimization. A pragmatic and simple rule for calcium-dependent CM injection protocols is investigated and proposed for VMI-based coronary CT angiography with PCD-CT.

Materials and methods: A physiological circulation phantom with coronary vessels including calcified lesions (maximum CT value >700 HU) with a 50% diameter stenosis was injected into at iodine delivery rates (IDRs) of 0.3, 0.5, 0.7, 1.0, 1.5, 2.0, 2.5, and 3.0 g I/s. Images were acquired using a first-generation dual-source PCD-CT and reconstructed at various VMI levels (between 45 and 190 keV). Iodine attenuation in the coronaries was measured at each IDR for each keV, and blooming artifacts from the calcified lesions were assessed including stenosis grading error (as % overestimation vs true lumen). The IDR to achieve 300 HU at each VMI level was then calculated and compared with stenosis grading accuracy to establish a general rule for CM injection protocols.

Results: Plaque blooming artifacts and intraluminal iodine attenuation decreased with increasing keV. Fixed windowing (representing absolute worst case) resulted in stenosis overestimation from 77% ± 4% at 45 keV to 5% ± 2% at 190 keV, whereas optimized windowing resulted in overestimation from 29% ± 3% at 45 keV to 4% ± 1% at 190 keV. The required IDR to achieve 300 HU showed a strong linear correlation to VMI energy ( R2 = 0.98). Comparison of this linear plot versus stenosis grading error and blooming artifact demonstrated that multipliers of 1, 2, and 3 times the reference IDR for theoretical clinical regimes of no, moderate, and severe calcification density, respectively, can be proposed as a general rule.

Conclusions: This study provides a proof-of-concept in an anthropomorphic phantom for a simple pragmatic adaptation of CM injection protocols in coronary CT angiography with PCD-CT. The 1-2-3 rule demonstrates the potential for reducing the effects of calcium blooming artifacts on overall image quality.

目的:光子计数探测器计算机断层扫描(PCD-CT光子计数探测器计算机断层扫描(PCD-CT)可获取 CT 血管造影的光谱数据,从而在常规实践中重建虚拟单能图像(VMI)。特别是,它有可能减少与密集钙化斑块相关的钙化伪影。然而,钙华和碘衰减与 VMI 的能级(keV)成反比,这给造影剂(CM)注射方案的优化带来了挑战。针对基于 VMI 的 PCD-CT 冠状动脉 CT 血管造影,研究并提出了一个实用、简单的钙依赖性 CM 注射方案规则:生理循环模型中的冠状动脉血管包括直径狭窄 50%的钙化病变(最大 CT 值大于 700 HU),碘注射率(IDR)分别为 0.3、0.5、0.7、1.0、1.5、2.0、2.5 和 3.0 g I/s。使用第一代双源 PCD-CT 采集图像,并在不同的 VMI 水平(45 至 190 千伏之间)下进行重建。在每个IDR、每个KeV下测量冠状动脉的碘衰减,并评估钙化病变产生的开花伪影,包括狭窄分级误差(高估与真实管腔的百分比)。然后计算每个 VMI 水平达到 300 HU 的 IDR,并将其与狭窄分级的准确性进行比较,以制定 CM 注射方案的一般规则:结果:斑块开花伪影和腔内碘衰减随 keV 的增加而减少。固定开窗(代表绝对最差情况)导致血管狭窄高估,从 45 keV 时的 77% ± 4% 降至 190 keV 时的 5% ± 2%,而优化开窗导致血管狭窄高估,从 45 keV 时的 29% ± 3% 降至 190 keV 时的 4% ± 1%。达到 300 HU 所需的 IDR 与 VMI 能量呈很强的线性相关(R2 = 0.98)。该线性图与血管狭窄分级误差和出血伪影的比较表明,对于无钙化密度、中度钙化密度和重度钙化密度的临床理论机制,可将参考 IDR 的 1 倍、2 倍和 3 倍作为一般规则:本研究在一个拟人化模型中对 PCD-CT 冠状动脉 CT 血管造影中的 CM 注射方案进行了简单实用的调整,提供了一个概念验证。1-2-3 规则证明了减少钙花伪影对整体图像质量影响的潜力。
{"title":"Countering Calcium Blooming With Personalized Contrast Media Injection Protocols: The 1-2-3 Rule for Photon-Counting Detector CCTA.","authors":"Michael C McDermott, Thomas Sartoretti, Lion Stammen, Bibi Martens, Gregor Jost, Hubertus Pietsch, Ralf Gutjahr, Bernhard Schmidt, Thomas G Flohr, Hatem Alkadhi, Joachim E Wildberger","doi":"10.1097/RLI.0000000000001078","DOIUrl":"10.1097/RLI.0000000000001078","url":null,"abstract":"<p><strong>Objective: </strong>Photon-counting detector computed tomography (PCD-CT) enables spectral data acquisition of CT angiographies allowing for reconstruction of virtual monoenergetic images (VMIs) in routine practice. Specifically, it has potential to reduce the blooming artifacts associated with densely calcified plaques. However, calcium blooming and iodine attenuation are inversely affected by energy level (keV) of the VMIs, creating a challenge for contrast media (CM) injection protocol optimization. A pragmatic and simple rule for calcium-dependent CM injection protocols is investigated and proposed for VMI-based coronary CT angiography with PCD-CT.</p><p><strong>Materials and methods: </strong>A physiological circulation phantom with coronary vessels including calcified lesions (maximum CT value >700 HU) with a 50% diameter stenosis was injected into at iodine delivery rates (IDRs) of 0.3, 0.5, 0.7, 1.0, 1.5, 2.0, 2.5, and 3.0 g I/s. Images were acquired using a first-generation dual-source PCD-CT and reconstructed at various VMI levels (between 45 and 190 keV). Iodine attenuation in the coronaries was measured at each IDR for each keV, and blooming artifacts from the calcified lesions were assessed including stenosis grading error (as % overestimation vs true lumen). The IDR to achieve 300 HU at each VMI level was then calculated and compared with stenosis grading accuracy to establish a general rule for CM injection protocols.</p><p><strong>Results: </strong>Plaque blooming artifacts and intraluminal iodine attenuation decreased with increasing keV. Fixed windowing (representing absolute worst case) resulted in stenosis overestimation from 77% ± 4% at 45 keV to 5% ± 2% at 190 keV, whereas optimized windowing resulted in overestimation from 29% ± 3% at 45 keV to 4% ± 1% at 190 keV. The required IDR to achieve 300 HU showed a strong linear correlation to VMI energy ( R2 = 0.98). Comparison of this linear plot versus stenosis grading error and blooming artifact demonstrated that multipliers of 1, 2, and 3 times the reference IDR for theoretical clinical regimes of no, moderate, and severe calcification density, respectively, can be proposed as a general rule.</p><p><strong>Conclusions: </strong>This study provides a proof-of-concept in an anthropomorphic phantom for a simple pragmatic adaptation of CM injection protocols in coronary CT angiography with PCD-CT. The 1-2-3 rule demonstrates the potential for reducing the effects of calcium blooming artifacts on overall image quality.</p>","PeriodicalId":14486,"journal":{"name":"Investigative Radiology","volume":" ","pages":"684-690"},"PeriodicalIF":7.0,"publicationDate":"2024-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11460796/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"140921877","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
A Novel Reconstruction Technique to Reduce Stair-Step Artifacts in Sequential Mode Coronary CT Angiography. 减少顺序模式冠状动脉 CT 血管造影中台阶伪影的新型重建技术
IF 7 1区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2024-09-01 Epub Date: 2024-01-30 DOI: 10.1097/RLI.0000000000001066
Lukas Jakob Moser, Victor Mergen, Thomas Allmendinger, Robert Manka, Matthias Eberhard, Hatem Alkadhi

Purpose: Prospective electrocardiography-triggering is one of the most commonly used cardiac computed tomography (CT) scan modes but can be susceptible to stair-step artifacts in the transition areas of an acquisition over multiple cardiac cycles. We evaluated a novel reconstruction algorithm to reduce the occurrence and severity of such artifacts in sequential coronary CT angiography.

Materials and methods: In this institutional review board-approved, retrospective study, 50 consecutive patients (16 females; mean age, 58.9 ± 15.2) were included who underwent coronary CT angiography on a dual-source photon-counting detector CT in the sequential ultra-high-resolution mode with a detector collimation of 120 × 0.2 mm. Each scan was reconstructed without (hereafter called standard reconstruction) and with the novel ZeeFree reconstruction algorithm, which aims to minimize stair-step artifacts. The presence and extent of stair-step artifacts were rated by 2 independent, blinded readers on a 4-point discrete visual scale. The relationship between the occurrences of artifacts was correlated with the average and variability of heart rate and with patient characteristics.

Results: A total of 504 coronary segments were included into the analyses. In standard reconstructions, reader 1 reported stair-step artifacts in 40/504 (7.9%) segments, from which 12/504 led to nondiagnostic image quality (2.4% of all segments). Reader 2 reported 56/504 (11.1%) stair-step artifacts, from which 11/504 lead to nondiagnostic image quality (2.2% of all segments). With the ZeeFree algorithm, 9/12 (75%) and 8/11 (73%) of the nondiagnostic segments improved to a diagnostic quality for readers 1 and 2, respectively. The ZeeFree reconstruction algorithm significantly reduced the frequency and extent of stair-step artifacts compared with standard reconstructions for both readers ( P < 0.001, each). Heart rate variability and body mass index were significantly related to the occurrence of stair-step artifacts ( P < 0.05).

Conclusions: Our study demonstrates the feasibility and effectiveness of a novel reconstruction algorithm leading to a significant reduction of stair-step artifacts and, hence, a reduction of coronary segments with a nondiagnostic image quality in sequential ultra-high-resolution coronary photon-counting detector CT angiography.

目的:前瞻性心电图触发是最常用的心脏计算机断层扫描(CT)扫描模式之一,但在多个心动周期的采集过渡区域容易出现阶梯状伪影。我们评估了一种新型重建算法,以减少顺序冠状动脉 CT 血管造影中此类伪影的出现和严重程度:在这项机构审查委员会批准的回顾性研究中,50 名连续患者(16 名女性;平均年龄为 58.9 ± 15.2)在双源光子计数探测器 CT 上进行了冠状动脉 CT 血管造影,CT 采用顺序超高分辨率模式,探测器准直度为 120 × 0.2 毫米。每次扫描均采用无重建(以下称为标准重建)和新型 ZeeFree 重建算法进行重建,该算法旨在最大限度地减少阶梯伪影。阶梯伪影的存在和程度由两名独立的盲人读者根据 4 点离散视觉量表进行评分。伪影的出现与心率的平均值和变异性以及患者特征之间的关系也有关联:共有 504 个冠状动脉节段被纳入分析。在标准重建中,读者 1 报告 40/504 个节段(7.9%)出现阶梯状伪影,其中 12/504 个节段的图像质量无法诊断(占所有节段的 2.4%)。阅读器 2 报告了 56/504 个(11.1%)阶梯状伪影,其中 11/504 个导致图像质量无法诊断(占所有节段的 2.2%)。使用 ZeeFree 算法后,读者 1 和读者 2 中分别有 9/12 个(75%)和 8/11 个(73%)非诊断节段的图像质量提高到了诊断质量。与标准重建相比,ZeeFree 重建算法大大降低了两位读者的阶梯伪影频率和程度(P < 0.001,各不相同)。心率变异性和体重指数与阶梯伪影的发生有明显关系(P < 0.05):我们的研究证明了一种新型重建算法的可行性和有效性,该算法可显著减少阶梯伪影,从而减少连续超高分辨率冠状动脉光子计数探测器 CT 血管造影中图像质量不达标的冠状动脉节段。
{"title":"A Novel Reconstruction Technique to Reduce Stair-Step Artifacts in Sequential Mode Coronary CT Angiography.","authors":"Lukas Jakob Moser, Victor Mergen, Thomas Allmendinger, Robert Manka, Matthias Eberhard, Hatem Alkadhi","doi":"10.1097/RLI.0000000000001066","DOIUrl":"10.1097/RLI.0000000000001066","url":null,"abstract":"<p><strong>Purpose: </strong>Prospective electrocardiography-triggering is one of the most commonly used cardiac computed tomography (CT) scan modes but can be susceptible to stair-step artifacts in the transition areas of an acquisition over multiple cardiac cycles. We evaluated a novel reconstruction algorithm to reduce the occurrence and severity of such artifacts in sequential coronary CT angiography.</p><p><strong>Materials and methods: </strong>In this institutional review board-approved, retrospective study, 50 consecutive patients (16 females; mean age, 58.9 ± 15.2) were included who underwent coronary CT angiography on a dual-source photon-counting detector CT in the sequential ultra-high-resolution mode with a detector collimation of 120 × 0.2 mm. Each scan was reconstructed without (hereafter called standard reconstruction) and with the novel ZeeFree reconstruction algorithm, which aims to minimize stair-step artifacts. The presence and extent of stair-step artifacts were rated by 2 independent, blinded readers on a 4-point discrete visual scale. The relationship between the occurrences of artifacts was correlated with the average and variability of heart rate and with patient characteristics.</p><p><strong>Results: </strong>A total of 504 coronary segments were included into the analyses. In standard reconstructions, reader 1 reported stair-step artifacts in 40/504 (7.9%) segments, from which 12/504 led to nondiagnostic image quality (2.4% of all segments). Reader 2 reported 56/504 (11.1%) stair-step artifacts, from which 11/504 lead to nondiagnostic image quality (2.2% of all segments). With the ZeeFree algorithm, 9/12 (75%) and 8/11 (73%) of the nondiagnostic segments improved to a diagnostic quality for readers 1 and 2, respectively. The ZeeFree reconstruction algorithm significantly reduced the frequency and extent of stair-step artifacts compared with standard reconstructions for both readers ( P < 0.001, each). Heart rate variability and body mass index were significantly related to the occurrence of stair-step artifacts ( P < 0.05).</p><p><strong>Conclusions: </strong>Our study demonstrates the feasibility and effectiveness of a novel reconstruction algorithm leading to a significant reduction of stair-step artifacts and, hence, a reduction of coronary segments with a nondiagnostic image quality in sequential ultra-high-resolution coronary photon-counting detector CT angiography.</p>","PeriodicalId":14486,"journal":{"name":"Investigative Radiology","volume":" ","pages":"622-628"},"PeriodicalIF":7.0,"publicationDate":"2024-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11827681/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"139570330","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Clinical Safety of Gadobutrol: Review of Over 25 Years of Use Exceeding 100 Million Administrations. 钆布醇的临床安全性:超过 1 亿次用药的 25 年使用回顾。
IF 7 1区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2024-09-01 Epub Date: 2024-03-01 DOI: 10.1097/RLI.0000000000001072
Jan Endrikat, Matthias Gutberlet, Karl-Titus Hoffmann, Laura Schöckel, Aasia Bhatti, Cornelia Harz, Jörg Barkhausen

Background: The macrocyclic gadolinium-based contrast agent gadobutrol was introduced to the market in February 1998. Over the last 25 years, gadobutrol has been administered more than 100 million times worldwide providing a wealth of data related to safety.

Objective: The aim of this study was to perform a thorough review and status update on gadobutrol's safety.

Materials and methods: Safety data from the clinical phase II-IV program and postmarketing surveillance were descriptively analyzed from February 1998 until December 31, 2022. Literature on special at-risk populations and specific safety aspects was critically summarized.

Results: Forty-five clinical phase II-IV studies recruited 7856 patients receiving gadobutrol. Drug-related adverse events (AEs) were reported in 3.4% and serious AEs in <0.1% of patients. Nausea (0.7%) and dysgeusia (0.4%) were the most reported AEs. All other drug-related AEs occurred ≤0.3%. After more than 100 million gadobutrol administrations, overall adverse drug reactions (ADRs) from postmarketing surveillance (including clinical trials) were rare with an overall reporting rate of 0.0356%, hypersensitivity reactions (0.0147%), nausea (0.0032%), vomiting (0.0025%), and dyspnea (0.0010%). All other ADRs were <0.001%. No trend for higher rates of AEs was found in patients with reduced renal or liver function. Seven clinical studies reported safety findings in 7292 children ≤18 years, thereof 112 newborns/toddlers younger than 2 years. Overall, 61 ADRs (0.84%) were reported, including 3 serious ones. Adverse events in patients ≥65 years of age ("elderly") were significantly less frequent than in younger patients. A total of 4 reports diagnostic of or consistent with nephrogenic systemic fibrosis have been received. No causal relationship has been established between clinical signs and symptoms and the presence of small amounts of gadolinium in the body in patients with normal renal function after use of gadobutrol.

Conclusions: More than 100 million administrations worldwide have shown gadobutrol's well-established benefit-risk profile in any approved indication and populations.

背景:大环钆基造影剂钆布醇于 1998 年 2 月投放市场。在过去的 25 年中,钆布醇在全球的使用次数已超过 1 亿次,提供了大量与安全性相关的数据:本研究的目的是对钆布醇的安全性进行全面回顾和现状更新:对 1998 年 2 月至 2022 年 12 月 31 日期间临床 II-IV 期项目和上市后监测的安全性数据进行了描述性分析。对特殊高危人群和特定安全性方面的文献进行了批判性总结:45项临床II-IV期研究招募了7856名接受钆布醇治疗的患者。据报告,3.4%的患者发生了与药物相关的不良事件(AEs),结论为严重不良事件:全球超过 1 亿次的用药表明,钆布醇在任何已获批准的适应症和人群中都具有公认的效益-风险特征。
{"title":"Clinical Safety of Gadobutrol: Review of Over 25 Years of Use Exceeding 100 Million Administrations.","authors":"Jan Endrikat, Matthias Gutberlet, Karl-Titus Hoffmann, Laura Schöckel, Aasia Bhatti, Cornelia Harz, Jörg Barkhausen","doi":"10.1097/RLI.0000000000001072","DOIUrl":"10.1097/RLI.0000000000001072","url":null,"abstract":"<p><strong>Background: </strong>The macrocyclic gadolinium-based contrast agent gadobutrol was introduced to the market in February 1998. Over the last 25 years, gadobutrol has been administered more than 100 million times worldwide providing a wealth of data related to safety.</p><p><strong>Objective: </strong>The aim of this study was to perform a thorough review and status update on gadobutrol's safety.</p><p><strong>Materials and methods: </strong>Safety data from the clinical phase II-IV program and postmarketing surveillance were descriptively analyzed from February 1998 until December 31, 2022. Literature on special at-risk populations and specific safety aspects was critically summarized.</p><p><strong>Results: </strong>Forty-five clinical phase II-IV studies recruited 7856 patients receiving gadobutrol. Drug-related adverse events (AEs) were reported in 3.4% and serious AEs in <0.1% of patients. Nausea (0.7%) and dysgeusia (0.4%) were the most reported AEs. All other drug-related AEs occurred ≤0.3%. After more than 100 million gadobutrol administrations, overall adverse drug reactions (ADRs) from postmarketing surveillance (including clinical trials) were rare with an overall reporting rate of 0.0356%, hypersensitivity reactions (0.0147%), nausea (0.0032%), vomiting (0.0025%), and dyspnea (0.0010%). All other ADRs were <0.001%. No trend for higher rates of AEs was found in patients with reduced renal or liver function. Seven clinical studies reported safety findings in 7292 children ≤18 years, thereof 112 newborns/toddlers younger than 2 years. Overall, 61 ADRs (0.84%) were reported, including 3 serious ones. Adverse events in patients ≥65 years of age (\"elderly\") were significantly less frequent than in younger patients. A total of 4 reports diagnostic of or consistent with nephrogenic systemic fibrosis have been received. No causal relationship has been established between clinical signs and symptoms and the presence of small amounts of gadolinium in the body in patients with normal renal function after use of gadobutrol.</p><p><strong>Conclusions: </strong>More than 100 million administrations worldwide have shown gadobutrol's well-established benefit-risk profile in any approved indication and populations.</p>","PeriodicalId":14486,"journal":{"name":"Investigative Radiology","volume":" ","pages":"605-613"},"PeriodicalIF":7.0,"publicationDate":"2024-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"139996237","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
The Small Pixel Effect in Ultra-High-Resolution Photon-Counting CT of the Lumbar Spine. 腰椎超高分辨率光子计数 CT 中的小像素效应
IF 7 1区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2024-09-01 Epub Date: 2024-02-08 DOI: 10.1097/RLI.0000000000001069
Henner Huflage, Robin Hendel, Piotr Woznicki, Nora Conrads, Philipp Feldle, Theresa Sophie Patzer, Süleyman Ergün, Thorsten Alexander Bley, Andreas Steven Kunz, Jan-Peter Grunz

Objectives: Image acquisition in ultra-high-resolution (UHR) scan mode does not impose a dose penalty in photon-counting CT (PCCT). This study aims to investigate the dose saving potential of using UHR instead of standard-resolution PCCT for lumbar spine imaging.

Materials and methods: Eight cadaveric specimens were examined with 7 dose levels (5-35 mGy) each in UHR (120 × 0.2 mm) and standard-resolution acquisition mode (144 × 0.4 mm) on a first-generation PCCT scanner. The UHR images were reconstructed with 3 dedicated bone kernels (Br68 [spatial frequency at 10% of the modulation transfer function 14.5 line pairs/cm], Br76 [21.0], and Br84 [27.9]), standard-resolution images with Br68 and Br76. Using automatic segmentation, contrast-to-noise ratios (CNRs) were established for lumbar vertebrae and psoas muscle tissue. In addition, image quality was assessed subjectively by 19 independent readers (15 radiologists, 4 surgeons) using a browser-based forced choice comparison tool totaling 16,974 performed pairwise tests. Pearson's correlation coefficient ( r ) was used to analyze the relationship between CNR and subjective image quality rankings, and Kendall W was calculated to assess interrater agreement.

Results: Irrespective of radiation exposure level, CNR was higher in UHR datasets than in standard-resolution images postprocessed with the same reconstruction parameters. The use of sharper convolution kernels entailed lower CNR but higher subjective image quality depending on radiation dose. Subjective assessment revealed high interrater agreement ( W = 0.86; P < 0.001) with UHR images being preferred by readers in the majority of comparisons on each dose level. Substantial correlation was ascertained between CNR and the subjective image quality ranking (all r 's ≥ 0.95; P < 0.001).

Conclusions: In PCCT of the lumbar spine, UHR mode's smaller pixel size facilitates a considerable CNR increase over standard-resolution imaging, which can either be used for dose reduction or higher spatial resolution depending on the selected convolution kernel.

目的:在超高分辨率(UHR)扫描模式下采集图像不会对光子计数 CT(PCCT)造成剂量损失。本研究旨在探讨在腰椎成像中使用超高分辨率而非标准分辨率 PCCT 可节省剂量的潜力:在第一代 PCCT 扫描仪上以 UHR(120 × 0.2 毫米)和标准分辨率采集模式(144 × 0.4 毫米)对 8 具尸体标本进行了 7 个剂量水平(5-35 mGy)的检查。UHR 图像用 3 个专用骨核(Br68 [空间频率为调制传递函数 14.5 线对/厘米的 10%]、Br76 [21.0] 和 Br84 [27.9])重建,标准分辨率图像用 Br68 和 Br76 重建。通过自动分割,确定了腰椎和腰肌组织的对比度-噪声比(CNR)。此外,19 位独立读者(15 位放射科医生和 4 位外科医生)使用基于浏览器的强制选择比较工具对图像质量进行了主观评估,共进行了 16974 次配对测试。皮尔逊相关系数(r)用于分析CNR与主观图像质量排名之间的关系,Kendall W用于评估判读者之间的一致性:结果:无论辐照水平如何,UHR 数据集的 CNR 均高于使用相同重建参数后处理的标准分辨率图像。根据辐射剂量的不同,使用更清晰的卷积核会导致更低的 CNR,但主观图像质量却更高。主观评估结果显示,在每个剂量水平的大多数比较中,UHR 图像受到读者的青睐,这显示了较高的校准间一致性(W = 0.86;P < 0.001)。CNR与主观图像质量排名之间存在显著相关性(所有r均≥0.95;P < 0.001):结论:在腰椎的 PCCT 中,UHR 模式的像素尺寸较小,与标准分辨率成像相比,CNR 大幅提高,根据所选卷积核的不同,可用于降低剂量或提高空间分辨率。
{"title":"The Small Pixel Effect in Ultra-High-Resolution Photon-Counting CT of the Lumbar Spine.","authors":"Henner Huflage, Robin Hendel, Piotr Woznicki, Nora Conrads, Philipp Feldle, Theresa Sophie Patzer, Süleyman Ergün, Thorsten Alexander Bley, Andreas Steven Kunz, Jan-Peter Grunz","doi":"10.1097/RLI.0000000000001069","DOIUrl":"10.1097/RLI.0000000000001069","url":null,"abstract":"<p><strong>Objectives: </strong>Image acquisition in ultra-high-resolution (UHR) scan mode does not impose a dose penalty in photon-counting CT (PCCT). This study aims to investigate the dose saving potential of using UHR instead of standard-resolution PCCT for lumbar spine imaging.</p><p><strong>Materials and methods: </strong>Eight cadaveric specimens were examined with 7 dose levels (5-35 mGy) each in UHR (120 × 0.2 mm) and standard-resolution acquisition mode (144 × 0.4 mm) on a first-generation PCCT scanner. The UHR images were reconstructed with 3 dedicated bone kernels (Br68 [spatial frequency at 10% of the modulation transfer function 14.5 line pairs/cm], Br76 [21.0], and Br84 [27.9]), standard-resolution images with Br68 and Br76. Using automatic segmentation, contrast-to-noise ratios (CNRs) were established for lumbar vertebrae and psoas muscle tissue. In addition, image quality was assessed subjectively by 19 independent readers (15 radiologists, 4 surgeons) using a browser-based forced choice comparison tool totaling 16,974 performed pairwise tests. Pearson's correlation coefficient ( r ) was used to analyze the relationship between CNR and subjective image quality rankings, and Kendall W was calculated to assess interrater agreement.</p><p><strong>Results: </strong>Irrespective of radiation exposure level, CNR was higher in UHR datasets than in standard-resolution images postprocessed with the same reconstruction parameters. The use of sharper convolution kernels entailed lower CNR but higher subjective image quality depending on radiation dose. Subjective assessment revealed high interrater agreement ( W = 0.86; P < 0.001) with UHR images being preferred by readers in the majority of comparisons on each dose level. Substantial correlation was ascertained between CNR and the subjective image quality ranking (all r 's ≥ 0.95; P < 0.001).</p><p><strong>Conclusions: </strong>In PCCT of the lumbar spine, UHR mode's smaller pixel size facilitates a considerable CNR increase over standard-resolution imaging, which can either be used for dose reduction or higher spatial resolution depending on the selected convolution kernel.</p>","PeriodicalId":14486,"journal":{"name":"Investigative Radiology","volume":" ","pages":"629-634"},"PeriodicalIF":7.0,"publicationDate":"2024-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"139706719","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Morphological and Quantitative Parametric MRI Follow-up of Cartilage Changes Before and After Intra-articular Injection Therapy in Patients With Mild to Moderate Knee Osteoarthritis: A Randomized, Placebo-Controlled Trial. 轻度至中度膝关节骨性关节炎患者关节内注射治疗前后软骨变化的形态学和定量参数磁共振成像随访:随机安慰剂对照试验。
IF 7 1区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2024-09-01 Epub Date: 2024-02-28 DOI: 10.1097/RLI.0000000000001067
Marcel Tschopp, Christian W A Pfirrmann, Florian Brunner, Sandro F Fucentese, Julien Galley, Christoph Stern, Reto Sutter, Sabrina Catanzaro, Nathalie Kühne, Andrea B Rosskopf

Background: Intra-articular injections are routinely used for conservative treatment of knee osteoarthritis (OA). The detailed comparative therapeutic effects of these injections on cartilage tissue are still unclear.

Objective: The aim of this study was to detect and compare knee cartilage changes after intra-articular injection of glucocorticoid, hyaluronic acid, or platelet-rich plasma (PRP) to placebo using quantitative (T2 and T2* mapping) and morphological magnetic resonance imaging parameters in patients with mild or moderate osteoarthritis.

Materials and methods: In a double-blinded, placebo-controlled, single-center trial, knees with mild or moderate osteoarthritis (Kellgren-Lawrence grade 1-3) were randomly assigned to an intra-articular injection with 1 of these substances: glucocorticoid, hyaluronic acid, PRP, or placebo. Cartilage degeneration on baseline and follow-up magnetic resonance imaging scans (after 3 and 12 months) was assessed by 2 readers using quantitative T2 and T2* times (milliseconds) and morphological parameters (modified Outerbridge grading, subchondral bone marrow edema, subchondral cysts, osteophytes).

Results: One hundred twenty knees (30 knees per treatment group) were analyzed with a median patient age of 60 years (interquartile range, 54.0-68.0 years). Interreader reliability was good for T2 (ICC, 0.76; IQR, 0.68-0.83) and T2* (ICC, 0.83; IQR, 0.76-0.88) measurements. Morphological parameters showed no significant changes between all groups after 3 and 12 months. T2 mapping after 12 months showed the following significant ( P = 0.001-0.03) changes between groups in 6 of 14 compartments: values after PRP injection decreased compared with glucocorticoid in 4 compartments (complete medial femoral condyle and central part of lateral condyle) and compared with placebo in 2 compartments (anterior and central part of medial tibial plateau); values after glucocorticoid injection decreased compared with placebo in 1 compartment (central part of medial tibial plateau). No significant changes were seen for T2 and T2* times after 3 months and T2* times after 12 months. No correlation was found between T2/T2* times and Kellgren-Lawrence grade, age, body mass index, or pain (Spearman ρ, -0.23 to 0.18).

Conclusions: Platelet-rich plasma injection has a positive long-term effect on cartilage quality in the medial femoral compartment compared to glucocorticoid, resulting in significantly improved T2 values after 12 months. For morphological cartilage parameters, injections with glucocorticoid, PRP, or hyaluronic acid showed no better effect in the short or long term compared with placebo.

背景:关节内注射是膝关节骨性关节炎(OA)保守治疗的常规方法。这些注射剂对软骨组织治疗效果的详细比较仍不清楚:本研究旨在使用定量(T2 和 T2* 映射)和形态学磁共振成像参数检测和比较轻度或中度骨关节炎患者在关节内注射糖皮质激素、透明质酸或富血小板血浆(PRP)和安慰剂后膝关节软骨的变化:在一项双盲、安慰剂对照、单中心试验中,患有轻度或中度骨关节炎(Kellgren-Lawrence 1-3 级)的膝关节被随机分配到关节内注射糖皮质激素、透明质酸、PRP 或安慰剂中的一种。由两名阅读者使用定量 T2 和 T2* 时间(毫秒)以及形态学参数(改良的 Outerbridge 分级、软骨下骨髓水肿、软骨下囊肿、骨质增生)对基线和随访磁共振成像扫描(3 个月和 12 个月后)上的软骨退变进行评估:对 120 个膝关节(每个治疗组 30 个膝关节)进行了分析,患者的中位年龄为 60 岁(四分位间范围为 54.0-68.0 岁)。T2(ICC,0.76;IQR,0.68-0.83)和T2*(ICC,0.83;IQR,0.76-0.88)测量的读数间可靠性良好。3 个月和 12 个月后,形态学参数在各组之间均无明显变化。12 个月后的 T2 图谱显示,14 个区段中有 6 个区段的组间变化显著(P = 0.001-0.03):与糖皮质激素相比,注射 PRP 后 4 个区段(股骨内侧髁和外侧髁中部)的数值下降,与安慰剂相比 2 个区段(胫骨内侧平台的前部和中部)的数值下降;与安慰剂相比,注射糖皮质激素后 1 个区段(胫骨内侧平台的中部)的数值下降。3 个月后的 T2 和 T2* 时间以及 12 个月后的 T2* 时间均无明显变化。T2/T2* 时间与 Kellgren-Lawrence 分级、年龄、体重指数或疼痛之间没有相关性(Spearman ρ,-0.23 至 0.18):结论:与糖皮质激素相比,注射富血小板血浆对股骨内侧软骨质量有积极的长期影响,12个月后T2值明显改善。就形态学软骨参数而言,与安慰剂相比,注射糖皮质激素、PRP或透明质酸在短期或长期都没有更好的效果。
{"title":"Morphological and Quantitative Parametric MRI Follow-up of Cartilage Changes Before and After Intra-articular Injection Therapy in Patients With Mild to Moderate Knee Osteoarthritis: A Randomized, Placebo-Controlled Trial.","authors":"Marcel Tschopp, Christian W A Pfirrmann, Florian Brunner, Sandro F Fucentese, Julien Galley, Christoph Stern, Reto Sutter, Sabrina Catanzaro, Nathalie Kühne, Andrea B Rosskopf","doi":"10.1097/RLI.0000000000001067","DOIUrl":"10.1097/RLI.0000000000001067","url":null,"abstract":"<p><strong>Background: </strong>Intra-articular injections are routinely used for conservative treatment of knee osteoarthritis (OA). The detailed comparative therapeutic effects of these injections on cartilage tissue are still unclear.</p><p><strong>Objective: </strong>The aim of this study was to detect and compare knee cartilage changes after intra-articular injection of glucocorticoid, hyaluronic acid, or platelet-rich plasma (PRP) to placebo using quantitative (T2 and T2* mapping) and morphological magnetic resonance imaging parameters in patients with mild or moderate osteoarthritis.</p><p><strong>Materials and methods: </strong>In a double-blinded, placebo-controlled, single-center trial, knees with mild or moderate osteoarthritis (Kellgren-Lawrence grade 1-3) were randomly assigned to an intra-articular injection with 1 of these substances: glucocorticoid, hyaluronic acid, PRP, or placebo. Cartilage degeneration on baseline and follow-up magnetic resonance imaging scans (after 3 and 12 months) was assessed by 2 readers using quantitative T2 and T2* times (milliseconds) and morphological parameters (modified Outerbridge grading, subchondral bone marrow edema, subchondral cysts, osteophytes).</p><p><strong>Results: </strong>One hundred twenty knees (30 knees per treatment group) were analyzed with a median patient age of 60 years (interquartile range, 54.0-68.0 years). Interreader reliability was good for T2 (ICC, 0.76; IQR, 0.68-0.83) and T2* (ICC, 0.83; IQR, 0.76-0.88) measurements. Morphological parameters showed no significant changes between all groups after 3 and 12 months. T2 mapping after 12 months showed the following significant ( P = 0.001-0.03) changes between groups in 6 of 14 compartments: values after PRP injection decreased compared with glucocorticoid in 4 compartments (complete medial femoral condyle and central part of lateral condyle) and compared with placebo in 2 compartments (anterior and central part of medial tibial plateau); values after glucocorticoid injection decreased compared with placebo in 1 compartment (central part of medial tibial plateau). No significant changes were seen for T2 and T2* times after 3 months and T2* times after 12 months. No correlation was found between T2/T2* times and Kellgren-Lawrence grade, age, body mass index, or pain (Spearman ρ, -0.23 to 0.18).</p><p><strong>Conclusions: </strong>Platelet-rich plasma injection has a positive long-term effect on cartilage quality in the medial femoral compartment compared to glucocorticoid, resulting in significantly improved T2 values after 12 months. For morphological cartilage parameters, injections with glucocorticoid, PRP, or hyaluronic acid showed no better effect in the short or long term compared with placebo.</p>","PeriodicalId":14486,"journal":{"name":"Investigative Radiology","volume":" ","pages":"646-655"},"PeriodicalIF":7.0,"publicationDate":"2024-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"139990023","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
AI-Based Measurement of Lumbar Spinal Stenosis on MRI: External Evaluation of a Fully Automated Model. 基于人工智能的磁共振成像腰椎管狭窄测量:全自动模型的外部评估
IF 7 1区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2024-09-01 Epub Date: 2024-03-01 DOI: 10.1097/RLI.0000000000001070
Sanja Bogdanovic, Matthias Staib, Marco Schleiniger, Livio Steiner, Leonardo Schwarz, Christoph Germann, Reto Sutter, Benjamin Fritz

Objectives: The aim of this study was to clinically validate a fully automated AI model for magnetic resonance imaging (MRI)-based quantifications of lumbar spinal canal stenosis.

Materials and methods: This retrospective study included lumbar spine MRI of 100 consecutive clinical patients (56 ± 17 years; 43 females, 57 males) performed on clinical 1.5 (51 examinations) and 3 T MRI scanners (49 examinations) with heterogeneous clinical imaging protocols. The AI model performed segmentations of the thecal sac on axial T2-weighted sequences. Based on these segmentations, the anteroposterior (AP) and mediolateral (ML) distance, and the area of the thecal sac were measured in a fully automated manner. For comparison, 2 fellowship-trained musculoskeletal radiologists performed the same segmentations and measurements independently. Statistics included 1-sample t tests, the intraclass correlation coefficient (ICC), Bland-Altman plots, and Dice coefficients. A P value of <0.05 was considered statistically significant.

Results: The average measurements of the AI model, reader 1, and reader 2 were 194 ± 72 mm 2 , 181 ± 71 mm 2 , and 179 ± 70 mm 2 for thecal sac area, 13 ± 3.3 mm, 12.6 ± 3.3 mm, and 12.6 ± 3.2 mm for AP distance, and 19.5 ± 3.9 mm, 20 ± 4.3 mm, and 19.4 ± 4 mm for ML distance, respectively. Significant differences existed for all pairwise comparisons, besides reader 1 versus AI model for the ML distance and reader 1 versus reader 2 for the AP distance ( P = 0.1 and P = 0.21, respectively). The pairwise mean absolute errors among reader 1, reader 2, and the AI model ranged from 0.59 mm and 0.75 mm for the AP distance, from 1.16 mm to 1.37 mm for the ML distance, and from 7.9 mm 2 to 15.54 mm 2 for the thecal sac area. Pairwise ICCs among reader 1, reader 2, and the AI model ranged from 0.91 and 0.94 for the AP distance and from 0.86 to 0.9 for the ML distance without significant differences. For the thecal sac area, the pairwise ICC between both readers and the AI model of 0.97 each was slightly, but significantly lower than the ICC between reader 1 and reader 2 of 0.99. Similarly, the Dice coefficient and Hausdorff distance between both readers and the AI model were significantly lower than the values between reader 1 and reader 2, overall ranging from 0.93 to 0.95 for the Dice coefficients and 1.1 to 1.44 for the Hausdorff distances.

Conclusions: The investigated AI model is reliable for assessing the AP and the ML thecal sac diameters with human level accuracies. The small differences for measurement and segmentation of the thecal sac area between the AI model and the radiologists are likely within a clinically acceptable range.

研究目的本研究旨在对基于磁共振成像(MRI)量化腰椎管狭窄的全自动人工智能模型进行临床验证:这项回顾性研究包括在临床 1.5 T(51 次检查)和 3 T MRI 扫描仪(49 次检查)上对 100 名连续临床患者(56 ± 17 岁;43 名女性,57 名男性)进行的腰椎 MRI 检查,临床成像方案各不相同。人工智能模型在轴向 T2 加权序列上对椎管囊进行了分割。在这些分割的基础上,以全自动的方式测量了椎管内囊的前后(AP)距离、内外侧(ML)距离和面积。为了进行比较,两名受过研究培训的肌肉骨骼放射科医生独立完成了相同的分割和测量。统计数据包括单样本 t 检验、类内相关系数 (ICC)、Bland-Altman 图和 Dice 系数。结果的 P 值:AI 模型、阅读器 1 和阅读器 2 的平均测量结果分别为:膀胱囊面积为 194 ± 72 平方毫米、181 ± 71 平方毫米和 179 ± 70 平方毫米;AP 距离为 13 ± 3.3 毫米、12.6 ± 3.3 毫米和 12.6 ± 3.2 毫米;ML 距离为 19.5 ± 3.9 毫米、20 ± 4.3 毫米和 19.4 ± 4 毫米。除了阅读器 1 与人工智能模型的 ML 距离比较和阅读器 1 与阅读器 2 的 AP 距离比较存在显著差异外(P = 0.1 和 P = 0.21),所有成对比较均存在显著差异。阅读器 1、阅读器 2 和 AI 模型之间的成对平均绝对误差范围为:AP 距离为 0.59 毫米至 0.75 毫米,ML 距离为 1.16 毫米至 1.37 毫米,睾丸囊面积为 7.9 平方毫米至 15.54 平方毫米。阅读器 1、阅读器 2 和 AI 模型之间的配对 ICC 在 AP 距离上介于 0.91 和 0.94 之间,在 ML 距离上介于 0.86 和 0.9 之间,无显著差异。在钙囊面积方面,两个读者和人工智能模型之间的成对 ICC 均为 0.97,略低于读者 1 和读者 2 之间的 ICC 0.99,但差异显著。同样,两名读者与人工智能模型之间的狄斯系数和豪斯多夫距离也明显低于读者 1 和读者 2 之间的值,狄斯系数的总体范围为 0.93 至 0.95,豪斯多夫距离的总体范围为 1.1 至 1.44:所研究的人工智能模型可以可靠地评估 AP 和 ML 膀胱囊直径,其精确度达到了人类水平。人工智能模型与放射科医生在测量和分割椎囊面积方面的微小差异可能在临床可接受的范围内。
{"title":"AI-Based Measurement of Lumbar Spinal Stenosis on MRI: External Evaluation of a Fully Automated Model.","authors":"Sanja Bogdanovic, Matthias Staib, Marco Schleiniger, Livio Steiner, Leonardo Schwarz, Christoph Germann, Reto Sutter, Benjamin Fritz","doi":"10.1097/RLI.0000000000001070","DOIUrl":"10.1097/RLI.0000000000001070","url":null,"abstract":"<p><strong>Objectives: </strong>The aim of this study was to clinically validate a fully automated AI model for magnetic resonance imaging (MRI)-based quantifications of lumbar spinal canal stenosis.</p><p><strong>Materials and methods: </strong>This retrospective study included lumbar spine MRI of 100 consecutive clinical patients (56 ± 17 years; 43 females, 57 males) performed on clinical 1.5 (51 examinations) and 3 T MRI scanners (49 examinations) with heterogeneous clinical imaging protocols. The AI model performed segmentations of the thecal sac on axial T2-weighted sequences. Based on these segmentations, the anteroposterior (AP) and mediolateral (ML) distance, and the area of the thecal sac were measured in a fully automated manner. For comparison, 2 fellowship-trained musculoskeletal radiologists performed the same segmentations and measurements independently. Statistics included 1-sample t tests, the intraclass correlation coefficient (ICC), Bland-Altman plots, and Dice coefficients. A P value of <0.05 was considered statistically significant.</p><p><strong>Results: </strong>The average measurements of the AI model, reader 1, and reader 2 were 194 ± 72 mm 2 , 181 ± 71 mm 2 , and 179 ± 70 mm 2 for thecal sac area, 13 ± 3.3 mm, 12.6 ± 3.3 mm, and 12.6 ± 3.2 mm for AP distance, and 19.5 ± 3.9 mm, 20 ± 4.3 mm, and 19.4 ± 4 mm for ML distance, respectively. Significant differences existed for all pairwise comparisons, besides reader 1 versus AI model for the ML distance and reader 1 versus reader 2 for the AP distance ( P = 0.1 and P = 0.21, respectively). The pairwise mean absolute errors among reader 1, reader 2, and the AI model ranged from 0.59 mm and 0.75 mm for the AP distance, from 1.16 mm to 1.37 mm for the ML distance, and from 7.9 mm 2 to 15.54 mm 2 for the thecal sac area. Pairwise ICCs among reader 1, reader 2, and the AI model ranged from 0.91 and 0.94 for the AP distance and from 0.86 to 0.9 for the ML distance without significant differences. For the thecal sac area, the pairwise ICC between both readers and the AI model of 0.97 each was slightly, but significantly lower than the ICC between reader 1 and reader 2 of 0.99. Similarly, the Dice coefficient and Hausdorff distance between both readers and the AI model were significantly lower than the values between reader 1 and reader 2, overall ranging from 0.93 to 0.95 for the Dice coefficients and 1.1 to 1.44 for the Hausdorff distances.</p><p><strong>Conclusions: </strong>The investigated AI model is reliable for assessing the AP and the ML thecal sac diameters with human level accuracies. The small differences for measurement and segmentation of the thecal sac area between the AI model and the radiologists are likely within a clinically acceptable range.</p>","PeriodicalId":14486,"journal":{"name":"Investigative Radiology","volume":" ","pages":"656-666"},"PeriodicalIF":7.0,"publicationDate":"2024-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"139996236","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Evaluation of the Contrast Enhancement Performance of Gadopiclenol for Magnetic Resonance Angiography in Healthy Rabbits and Pigs. 评估钆喷酸醇在健康兔子和猪磁共振血管造影中的对比增强性能
IF 7 1区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2024-09-01 Epub Date: 2024-05-07 DOI: 10.1097/RLI.0000000000001083
Gaëlle Hugon, Hans Adriaensen, Mélody Wintrebert, Laurent Arnould, Jean-Michel Serfaty, Philippe Robert

Objectives: Unexpected accumulations of gadolinium in various organs were reported after the administration of gadolinium-based contrast agents, making desirable to reduce the dose while maintaining equivalent diagnostic performance. The aim of this study was to evaluate the contrast enhancement performance of high relaxivity gadopiclenol compared with gadoterate meglumine in abdominal contrast-enhanced magnetic resonance angiography (CE-MRA).

Materials and methods: In a first study in healthy rabbits, axial 3D gradient echo sequences were applied at 4.7 T to study arterial enhancement as a function of gadopiclenol dose (0.025, 0.05, 0.075, and 0.1 mmol Gd/kg) or gadoterate meglumine at 0.1 mmol Gd/kg (n = 5-6/group). The increase in signal-to-noise ratio (ΔSNR) in the aorta at the first pass was measured and compared. In a second, crossover study in 6 healthy pigs, abdominal CE-MRA sequences were acquired at 3 T with gadopiclenol at 0.05 mmol Gd/kg or gadoterate meglumine at 0.1 mmol Gd/kg at a 1-week interval. Quantitatively on the maximum intensity projection (MIP) images, the mean MIP SNR within the aorta of both groups was compared. Qualitatively, a blinded comparison of the angiograms was performed by an experienced radiologist to determine the preferred contrast agent.

Results: In the rabbit, ∆SNR is linearly correlated with the gadopiclenol dose ( P = 0.0010). Compared with gadoterate meglumine 0.1 mmol Gd/kg, an increase in the ∆SNR is observed after 0.05, 0.075, and 0.1 mmol Gd/kg of gadopiclenol (+63% P = 0.0731, +78% P = 0.0081, and +72% P = 0.0773, respectively), whereas at 0.025 mmol Gd/kg, ∆SNR is in the same range as with gadoterate meglumine 0.1 mmol Gd/kg (+15% P > 0.9999). In pigs, contrast enhancement after gadopiclenol at 0.05 mmol/kg is +22% superior to MIP SNR after gadoterate meglumine at 0.1 mmol Gd/kg ( P = 0.3095). Qualitatively, a preference was shown for gadopiclenol images (3/6) over the gadoterate meglumine examinations (1/6), with no preference being shown for the remainder (2/6).

Conclusions: First-pass CE-MRA is feasible with gadopiclenol at 0.05 mmol Gd/kg with at least the same arterial signal enhancement and image quality as gadoterate meglumine at 0.1 mmol Gd/kg.

目的:据报道,使用钆基造影剂后,钆会在不同器官中意外蓄积,因此,在保持同等诊断性能的同时减少剂量是可取的。本研究旨在评估在腹部对比增强磁共振血管造影(CE-MRA)中,高弛豫度钆喷酸醇与钆喷酸葡胺的对比增强性能:在对健康兔子进行的首次研究中,在 4.7 T 下应用轴向三维梯度回波序列研究了钆喷酸诺剂量(0.025、0.05、0.075 和 0.1 mmol Gd/kg)或 0.1 mmol Gd/kg 的钆喷酸葡胺(n = 5-6/组)对动脉增强的影响。测量并比较了主动脉首次通过时信噪比(ΔSNR)的增加情况。在对 6 头健康猪进行的第二项交叉研究中,腹部 CE-MRA 序列是在 3 T 下用 0.05 mmol Gd/kg 的钆喷酸仑或 0.1 mmol Gd/kg 的钆喷酸葡胺采集的,间隔时间为 1 周。在最大强度投影(MIP)图像上,对两组主动脉内的平均 MIP SNR 进行定量比较。定性方面,由一名经验丰富的放射科医生对血管造影进行盲比对,以确定首选造影剂:在兔子身上,∆SNR 与钆喷酸酯剂量呈线性相关(P = 0.0010)。与钆特酸葡胺 0.1 毫摩尔 Gd/kg 相比,在使用 0.05、0.075 和 0.1 毫摩尔 Gd/kg 的钆吡特仑醇后,观察到 ∆SNR 增加(+63% P = 0.0731、+78% P = 0.0081 和 +72% P = 0.0773),而在 0.025 mmol Gd/kg 时,ΔSNR 与钆喷酸葡胺 0.1 mmol Gd/kg 时的范围相同(+15% P > 0.9999)。在猪身上,0.05 mmol/kg 的钆喷酸葡胺造影剂的对比增强效果比 0.1 mmol Gd/kg 的钆喷酸葡胺造影剂的 MIP SNR 高 22%(P = 0.3095)。定性分析显示,钆喷酸葡胺图像(3/6)优于钆喷酸葡胺检查(1/6),其余图像(2/6)无优劣之分:结论:使用 0.05 毫摩尔 Gd/kg 的钆喷酸葡胺进行第一道 CE-MRA 是可行的,其动脉信号增强和图像质量至少与 0.1 毫摩尔 Gd/kg 的钆喷酸葡胺相同。
{"title":"Evaluation of the Contrast Enhancement Performance of Gadopiclenol for Magnetic Resonance Angiography in Healthy Rabbits and Pigs.","authors":"Gaëlle Hugon, Hans Adriaensen, Mélody Wintrebert, Laurent Arnould, Jean-Michel Serfaty, Philippe Robert","doi":"10.1097/RLI.0000000000001083","DOIUrl":"10.1097/RLI.0000000000001083","url":null,"abstract":"<p><strong>Objectives: </strong>Unexpected accumulations of gadolinium in various organs were reported after the administration of gadolinium-based contrast agents, making desirable to reduce the dose while maintaining equivalent diagnostic performance. The aim of this study was to evaluate the contrast enhancement performance of high relaxivity gadopiclenol compared with gadoterate meglumine in abdominal contrast-enhanced magnetic resonance angiography (CE-MRA).</p><p><strong>Materials and methods: </strong>In a first study in healthy rabbits, axial 3D gradient echo sequences were applied at 4.7 T to study arterial enhancement as a function of gadopiclenol dose (0.025, 0.05, 0.075, and 0.1 mmol Gd/kg) or gadoterate meglumine at 0.1 mmol Gd/kg (n = 5-6/group). The increase in signal-to-noise ratio (ΔSNR) in the aorta at the first pass was measured and compared. In a second, crossover study in 6 healthy pigs, abdominal CE-MRA sequences were acquired at 3 T with gadopiclenol at 0.05 mmol Gd/kg or gadoterate meglumine at 0.1 mmol Gd/kg at a 1-week interval. Quantitatively on the maximum intensity projection (MIP) images, the mean MIP SNR within the aorta of both groups was compared. Qualitatively, a blinded comparison of the angiograms was performed by an experienced radiologist to determine the preferred contrast agent.</p><p><strong>Results: </strong>In the rabbit, ∆SNR is linearly correlated with the gadopiclenol dose ( P = 0.0010). Compared with gadoterate meglumine 0.1 mmol Gd/kg, an increase in the ∆SNR is observed after 0.05, 0.075, and 0.1 mmol Gd/kg of gadopiclenol (+63% P = 0.0731, +78% P = 0.0081, and +72% P = 0.0773, respectively), whereas at 0.025 mmol Gd/kg, ∆SNR is in the same range as with gadoterate meglumine 0.1 mmol Gd/kg (+15% P > 0.9999). In pigs, contrast enhancement after gadopiclenol at 0.05 mmol/kg is +22% superior to MIP SNR after gadoterate meglumine at 0.1 mmol Gd/kg ( P = 0.3095). Qualitatively, a preference was shown for gadopiclenol images (3/6) over the gadoterate meglumine examinations (1/6), with no preference being shown for the remainder (2/6).</p><p><strong>Conclusions: </strong>First-pass CE-MRA is feasible with gadopiclenol at 0.05 mmol Gd/kg with at least the same arterial signal enhancement and image quality as gadoterate meglumine at 0.1 mmol Gd/kg.</p>","PeriodicalId":14486,"journal":{"name":"Investigative Radiology","volume":" ","pages":"614-621"},"PeriodicalIF":7.0,"publicationDate":"2024-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"140857049","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Addressing the Contrast Media Recognition Challenge: A Fully Automated Machine Learning Approach for Predicting Contrast Phases in CT Imaging. 应对对比介质识别挑战:预测 CT 成像中对比相位的全自动机器学习方法。
IF 7 1区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2024-09-01 Epub Date: 2024-03-04 DOI: 10.1097/RLI.0000000000001071
Giulia Baldini, René Hosch, Cynthia S Schmidt, Katarzyna Borys, Lennard Kroll, Sven Koitka, Patrizia Haubold, Obioma Pelka, Felix Nensa, Johannes Haubold

Objectives: Accurately acquiring and assigning different contrast-enhanced phases in computed tomography (CT) is relevant for clinicians and for artificial intelligence orchestration to select the most appropriate series for analysis. However, this information is commonly extracted from the CT metadata, which is often wrong. This study aimed at developing an automatic pipeline for classifying intravenous (IV) contrast phases and additionally for identifying contrast media in the gastrointestinal tract (GIT).

Materials and methods: This retrospective study used 1200 CT scans collected at the investigating institution between January 4, 2016 and September 12, 2022, and 240 CT scans from multiple centers from The Cancer Imaging Archive for external validation. The open-source segmentation algorithm TotalSegmentator was used to identify regions of interest (pulmonary artery, aorta, stomach, portal/splenic vein, liver, portal vein/hepatic veins, inferior vena cava, duodenum, small bowel, colon, left/right kidney, urinary bladder), and machine learning classifiers were trained with 5-fold cross-validation to classify IV contrast phases (noncontrast, pulmonary arterial, arterial, venous, and urographic) and GIT contrast enhancement. The performance of the ensembles was evaluated using the receiver operating characteristic area under the curve (AUC) and 95% confidence intervals (CIs).

Results: For the IV phase classification task, the following AUC scores were obtained for the internal test set: 99.59% [95% CI, 99.58-99.63] for the noncontrast phase, 99.50% [95% CI, 99.49-99.52] for the pulmonary-arterial phase, 99.13% [95% CI, 99.10-99.15] for the arterial phase, 99.8% [95% CI, 99.79-99.81] for the venous phase, and 99.7% [95% CI, 99.68-99.7] for the urographic phase. For the external dataset, a mean AUC of 97.33% [95% CI, 97.27-97.35] and 97.38% [95% CI, 97.34-97.41] was achieved for all contrast phases for the first and second annotators, respectively. Contrast media in the GIT could be identified with an AUC of 99.90% [95% CI, 99.89-99.9] in the internal dataset, whereas in the external dataset, an AUC of 99.73% [95% CI, 99.71-99.73] and 99.31% [95% CI, 99.27-99.33] was achieved with the first and second annotator, respectively.

Conclusions: The integration of open-source segmentation networks and classifiers effectively classified contrast phases and identified GIT contrast enhancement using anatomical landmarks.

目的:在计算机断层扫描(CT)中准确获取和分配不同的对比度增强相位,对临床医生和人工智能协调选择最合适的序列进行分析都很重要。然而,这些信息通常是从 CT 元数据中提取的,而元数据往往是错误的。本研究旨在开发一种自动管道,用于对静脉注射(IV)造影剂阶段进行分类,以及识别胃肠道(GIT)中的造影剂:这项回顾性研究使用了研究机构在2016年1月4日至2022年9月12日期间收集的1200张CT扫描照片,以及来自癌症影像档案馆的多个中心的240张CT扫描照片进行外部验证。使用开源分割算法 TotalSegmentator 识别感兴趣区域(肺动脉、主动脉、胃、门静脉/脾静脉、肝脏、门静脉/肝静脉、下腔静脉、十二指肠、小肠、结肠、左/右肾、膀胱)、通过 5 次交叉验证对机器学习分类器进行训练,以对 IV 造影剂阶段(非造影剂、肺动脉、动脉、静脉和尿路造影剂)和 GIT 造影剂增强进行分类。使用接收器操作特征曲线下面积(AUC)和 95% 置信区间(CIs)对组合的性能进行了评估:在 IV 期分类任务中,内部测试集的 AUC 得分如下非对比相为 99.59% [95% CI,99.58-99.63],肺动脉相为 99.50% [95% CI,99.49-99.52],动脉相为 99.13% [95% CI,99.10-99.15],静脉相为 99.8% [95% CI,99.79-99.81],尿路相为 99.7% [95% CI,99.68-99.7]。对于外部数据集,第一位和第二位标注者在所有对比阶段的平均 AUC 分别为 97.33% [95% CI,97.27-97.35] 和 97.38% [95% CI,97.34-97.41]。在内部数据集中,GIT 中对比介质的识别率为 99.90% [95% CI, 99.89-99.9],而在外部数据集中,第一和第二注释者的识别率分别为 99.73% [95% CI, 99.71-99.73] 和 99.31% [95% CI, 99.27-99.33]:开源分割网络和分类器的整合有效地对对比度阶段进行了分类,并利用解剖地标识别了 GIT 对比度增强。
{"title":"Addressing the Contrast Media Recognition Challenge: A Fully Automated Machine Learning Approach for Predicting Contrast Phases in CT Imaging.","authors":"Giulia Baldini, René Hosch, Cynthia S Schmidt, Katarzyna Borys, Lennard Kroll, Sven Koitka, Patrizia Haubold, Obioma Pelka, Felix Nensa, Johannes Haubold","doi":"10.1097/RLI.0000000000001071","DOIUrl":"10.1097/RLI.0000000000001071","url":null,"abstract":"<p><strong>Objectives: </strong>Accurately acquiring and assigning different contrast-enhanced phases in computed tomography (CT) is relevant for clinicians and for artificial intelligence orchestration to select the most appropriate series for analysis. However, this information is commonly extracted from the CT metadata, which is often wrong. This study aimed at developing an automatic pipeline for classifying intravenous (IV) contrast phases and additionally for identifying contrast media in the gastrointestinal tract (GIT).</p><p><strong>Materials and methods: </strong>This retrospective study used 1200 CT scans collected at the investigating institution between January 4, 2016 and September 12, 2022, and 240 CT scans from multiple centers from The Cancer Imaging Archive for external validation. The open-source segmentation algorithm TotalSegmentator was used to identify regions of interest (pulmonary artery, aorta, stomach, portal/splenic vein, liver, portal vein/hepatic veins, inferior vena cava, duodenum, small bowel, colon, left/right kidney, urinary bladder), and machine learning classifiers were trained with 5-fold cross-validation to classify IV contrast phases (noncontrast, pulmonary arterial, arterial, venous, and urographic) and GIT contrast enhancement. The performance of the ensembles was evaluated using the receiver operating characteristic area under the curve (AUC) and 95% confidence intervals (CIs).</p><p><strong>Results: </strong>For the IV phase classification task, the following AUC scores were obtained for the internal test set: 99.59% [95% CI, 99.58-99.63] for the noncontrast phase, 99.50% [95% CI, 99.49-99.52] for the pulmonary-arterial phase, 99.13% [95% CI, 99.10-99.15] for the arterial phase, 99.8% [95% CI, 99.79-99.81] for the venous phase, and 99.7% [95% CI, 99.68-99.7] for the urographic phase. For the external dataset, a mean AUC of 97.33% [95% CI, 97.27-97.35] and 97.38% [95% CI, 97.34-97.41] was achieved for all contrast phases for the first and second annotators, respectively. Contrast media in the GIT could be identified with an AUC of 99.90% [95% CI, 99.89-99.9] in the internal dataset, whereas in the external dataset, an AUC of 99.73% [95% CI, 99.71-99.73] and 99.31% [95% CI, 99.27-99.33] was achieved with the first and second annotator, respectively.</p><p><strong>Conclusions: </strong>The integration of open-source segmentation networks and classifiers effectively classified contrast phases and identified GIT contrast enhancement using anatomical landmarks.</p>","PeriodicalId":14486,"journal":{"name":"Investigative Radiology","volume":" ","pages":"635-645"},"PeriodicalIF":7.0,"publicationDate":"2024-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"140021740","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Optimal Spectral Performance on Pediatric Photon-Counting CT: Investigating Phantom-Based Size-Dependent kV Selection for Spectral Body Imaging. 小儿光子计数 CT 的最佳光谱性能:研究光谱人体成像中基于模型大小的 kV 选择。
IF 7 1区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2024-08-20 DOI: 10.1097/RLI.0000000000001119
Wei Zhou, Afrouz Ataei, Donglai Huo, Liqiang Ren, Lorna P Browne, Xin Zhou, Jason P Weinman

Purpose: The comprehensive evaluation of kV selection on photon-counting computed tomography (PCCT) has yet to be performed. The aim of the study is to evaluate and determine the optimal kV options for variable pediatric body sizes on the PCCT unit.

Materials and methods: In this study, 4 phantoms of variable sizes were utilized to represent abdomens of newborn, 5-year-old, 10-year-old, and adult-sized pediatric patients. One solid water and 4 solid iodine inserts with known concentrations (2, 5, 10, and 15 mg I/mL) were inserted into phantoms. Each phantom setting was scanned on a PCCT system (Siemens Alpha) with 4 kV options (70 and 90 kV under Quantum Mode, 120 and 140 kV under QuantumPlus Mode) and clinical dual-source (3.0 pitch) protocol. For each phantom setting, radiation dose (CTDIvol) was determined by clinical dose settings and matched for all kV acquisitions. Sixty percent clinical dose images were also acquired. Reconstruction was matched across all acquisitions using Qr40 kernel and QIR level 3. Virtual monoenergetic images (VMIs) between 40 and 80 keV with 10 keV interval were generated on the scanner. Low-energy and high-energy images were reconstructed from each scan and subsequently used to generate an iodine map (IM) using an image-based 2-material decomposition method. Image noise of VMIs from each kV acquisition was calculated and compared between kV options. Absolute percent error (APE) of iodine CT number accuracy in VMIs was calculated and compared. Root mean square error (RMSE) and bias of iodine quantification from IMs were compared across kV options.

Results: At the newborn size and 50 keV VMI, noise is lower at low kV acquisitions (70 kV: 10.5 HU, 90 kV: 10.4 HU), compared with high kV acquisitions (120 kV: 13.8 HU, 140 kV: 13.9 HU). At the newborn size and 70 keV VMI, the image noise from different kV options is comparable (9.4 HU for 70 kV, 8.9 HU for 90 kV, 9.7 HU for 120 kV, 10.2 HU for 140 kV). For APE of VMI, high kV (120 or 140 kV) performed overall better than low kV (70 or 90 kV). At the 5-year-old size, APE of 90 kV (median: 3.6%) is significantly higher (P < 0.001, Kruskal-Wallis rank sum test with Bonferroni correction) than 140 kV (median: 1.6%). At adult size, APE of 70 kV (median: 18.0%) is significantly higher (P < 0.0001, Kruskal-Wallis rank sum test with Bonferroni correction) than 120 kV (median: 1.4%) or 140 kV (median: 0.8%). The high kV also demonstrated lower RMSE and bias than the low kV across all controlled conditions. At 10-year-old size, RMSE and bias of 120 kV are 1.4 and 0.2 mg I/mL, whereas those from 70 kV are 1.9 and 0.8 mg I/mL.

Conclusions: The high kV options (120 or 140 kV) on the PCCT unit demonstrated overall better performance than the low kV options (70 or 90 kV), in terms of image quality of VMIs and IMs. Our results recommend the use of high kV for general body imaging on the PCCT.

目的:目前尚未对光子计数计算机断层扫描(PCCT)的 kV 选择进行全面评估。本研究旨在评估和确定 PCCT 设备上不同儿童体型的最佳 kV 选择:在这项研究中,使用了 4 个不同大小的模型,分别代表新生儿、5 岁、10 岁和成年儿科病人的腹部。将已知浓度(2、5、10 和 15 毫克碘/毫升)的 1 个固体水和 4 个固体碘插入模型中。每个模型均在 PCCT 系统(西门子 Alpha)上扫描,该系统有 4 个千伏选择(量子模式下为 70 和 90 千伏,量子增强模式下为 120 和 140 千伏)和临床双源(3.0 间距)协议。对于每种模型设置,辐射剂量(CTDIvol)都是根据临床剂量设置确定的,并与所有 kV 采集相匹配。还采集了 60% 的临床剂量图像。所有采集的重建均使用 Qr40 内核和 QIR 3 级进行匹配。扫描仪上生成的虚拟单能量图像(VMI)介于 40 和 80 keV 之间,间隔为 10 keV。从每次扫描中重建低能和高能图像,然后使用基于图像的双材料分解法生成碘图(IM)。计算每次千伏采集的 VMI 图像噪声,并在不同千伏选项之间进行比较。计算并比较了 VMI 中碘 CT 编号准确性的绝对百分比误差 (APE)。比较了不同 kV 选项中 IM 的均方根误差 (RMSE) 和碘定量偏差:在新生儿大小和 50 keV VMI 条件下,低 kV 采集(70 kV:10.5 HU,90 kV:10.4 HU)与高 kV 采集(120 kV:13.8 HU,140 kV:13.9 HU)相比,噪声更低。在新生儿大小和 70 千伏 VMI 的情况下,不同千伏选项的图像噪声相当(70 千伏:9.4 HU;90 千伏:8.9 HU;120 千伏:9.7 HU;140 千伏:10.2 HU)。就 VMI 的 APE 而言,高千伏(120 或 140 千伏)的总体表现优于低千伏(70 或 90 千伏)。在 5 岁儿童中,90 千伏的 APE(中位数:3.6%)明显高于 140 千伏(中位数:1.6%)(P < 0.001,Kruskal-Wallis 秩和检验,Bonferroni 校正)。成年体型时,70 千伏(中位数:18.0%)的 APE 明显高于 120 千伏(中位数:1.4%)或 140 千伏(中位数:0.8%)(P < 0.0001,Kruskal-Wallis 秩和检验,Bonferroni 校正)。在所有控制条件下,高千伏值的均方根误差和偏差也低于低千伏值。在 10 岁时,120 千伏的有效值和偏差分别为 1.4 和 0.2 毫克 I/毫升,而 70 千伏的有效值和偏差分别为 1.9 和 0.8 毫克 I/毫升:结论:就 VMI 和 IM 的图像质量而言,PCCT 设备上的高千伏选项(120 或 140 千伏)比低千伏选项(70 或 90 千伏)表现出更好的整体性能。我们的结果建议在 PCCT 上使用高千伏进行全身成像。
{"title":"Optimal Spectral Performance on Pediatric Photon-Counting CT: Investigating Phantom-Based Size-Dependent kV Selection for Spectral Body Imaging.","authors":"Wei Zhou, Afrouz Ataei, Donglai Huo, Liqiang Ren, Lorna P Browne, Xin Zhou, Jason P Weinman","doi":"10.1097/RLI.0000000000001119","DOIUrl":"https://doi.org/10.1097/RLI.0000000000001119","url":null,"abstract":"<p><strong>Purpose: </strong>The comprehensive evaluation of kV selection on photon-counting computed tomography (PCCT) has yet to be performed. The aim of the study is to evaluate and determine the optimal kV options for variable pediatric body sizes on the PCCT unit.</p><p><strong>Materials and methods: </strong>In this study, 4 phantoms of variable sizes were utilized to represent abdomens of newborn, 5-year-old, 10-year-old, and adult-sized pediatric patients. One solid water and 4 solid iodine inserts with known concentrations (2, 5, 10, and 15 mg I/mL) were inserted into phantoms. Each phantom setting was scanned on a PCCT system (Siemens Alpha) with 4 kV options (70 and 90 kV under Quantum Mode, 120 and 140 kV under QuantumPlus Mode) and clinical dual-source (3.0 pitch) protocol. For each phantom setting, radiation dose (CTDIvol) was determined by clinical dose settings and matched for all kV acquisitions. Sixty percent clinical dose images were also acquired. Reconstruction was matched across all acquisitions using Qr40 kernel and QIR level 3. Virtual monoenergetic images (VMIs) between 40 and 80 keV with 10 keV interval were generated on the scanner. Low-energy and high-energy images were reconstructed from each scan and subsequently used to generate an iodine map (IM) using an image-based 2-material decomposition method. Image noise of VMIs from each kV acquisition was calculated and compared between kV options. Absolute percent error (APE) of iodine CT number accuracy in VMIs was calculated and compared. Root mean square error (RMSE) and bias of iodine quantification from IMs were compared across kV options.</p><p><strong>Results: </strong>At the newborn size and 50 keV VMI, noise is lower at low kV acquisitions (70 kV: 10.5 HU, 90 kV: 10.4 HU), compared with high kV acquisitions (120 kV: 13.8 HU, 140 kV: 13.9 HU). At the newborn size and 70 keV VMI, the image noise from different kV options is comparable (9.4 HU for 70 kV, 8.9 HU for 90 kV, 9.7 HU for 120 kV, 10.2 HU for 140 kV). For APE of VMI, high kV (120 or 140 kV) performed overall better than low kV (70 or 90 kV). At the 5-year-old size, APE of 90 kV (median: 3.6%) is significantly higher (P < 0.001, Kruskal-Wallis rank sum test with Bonferroni correction) than 140 kV (median: 1.6%). At adult size, APE of 70 kV (median: 18.0%) is significantly higher (P < 0.0001, Kruskal-Wallis rank sum test with Bonferroni correction) than 120 kV (median: 1.4%) or 140 kV (median: 0.8%). The high kV also demonstrated lower RMSE and bias than the low kV across all controlled conditions. At 10-year-old size, RMSE and bias of 120 kV are 1.4 and 0.2 mg I/mL, whereas those from 70 kV are 1.9 and 0.8 mg I/mL.</p><p><strong>Conclusions: </strong>The high kV options (120 or 140 kV) on the PCCT unit demonstrated overall better performance than the low kV options (70 or 90 kV), in terms of image quality of VMIs and IMs. Our results recommend the use of high kV for general body imaging on the PCCT.</p>","PeriodicalId":14486,"journal":{"name":"Investigative Radiology","volume":" ","pages":""},"PeriodicalIF":7.0,"publicationDate":"2024-08-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"142004230","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Visualization of Renal Glomeruli in Human Native Kidneys With Sensing Ultrasound Localization Microscopy. 利用传感超声定位显微镜观察人体原生肾脏的肾小球
IF 7 1区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2024-08-01 Epub Date: 2024-01-12 DOI: 10.1097/RLI.0000000000001061
Sylvain Bodard, Louise Denis, Georges Chabouh, Jacques Battaglia, Dany Anglicheau, Olivier Hélénon, Jean-Michel Correas, Olivier Couture

Objectives: Kidney diseases significantly impact individuals' quality of life and strongly reduce life expectancy. Glomeruli play a crucial role in kidney function. Current imaging techniques cannot visualize them due to their small size. Sensing ultrasound localization microscopy (sULM) has shown promising results for visualizing in vivo the glomeruli of human kidney grafts. This study aimed to evaluate the ability of sULM to visualize glomeruli in vivo in native human kidneys despite their depth and a shorter duration of ultrasound acquisition limited by the period of the patient's apnea. Sensing ultrasound localization microscopy parameters in native kidneys and kidney grafts and their consequence regarding glomeruli detection were also compared.

Materials and methods: Exploration by sULM was conducted in 15 patients with native kidneys and 5 with kidney allografts. Glomeruli were counted using a normalized distance metric projected onto sULM density maps. The difference in the acquisition time, the kidney depth, and the frame rate between native kidneys and kidney grafts and their consequence regarding glomeruli detection were assessed.

Results: Glomerular visualization was achieved in 12 of 15 patients with native kidneys. It failed due to impossible breath-holding for 2 patients and a too-deep kidney for 1 patient. Sensing ultrasound localization microscopy found 16 glomeruli per square centimeter in the native kidneys (6-31) and 33 glomeruli per square centimeter in kidney transplant patients (18-55).

Conclusions: This study demonstrated that sULM can visualize glomeruli in native human kidneys in vivo. The proposed method may have many hypothetical applications, including biomarker development, assisting biopsy, or potentially avoiding it. It establishes a framework for improving the detection of local microstructural pathology, influencing the evaluation of allografts, and facilitating disease monitoring in the native kidney.

目的:肾脏疾病严重影响个人的生活质量,并大大缩短预期寿命。肾小球在肾功能中起着至关重要的作用。由于肾小球体积小,目前的成像技术无法观察到它们。传感超声定位显微镜(sULM)在活体观察人体肾移植肾小球方面显示出良好的效果。本研究旨在评估传感超声定位显微镜(sULM)在体内观察原生人类肾脏肾小球的能力,尽管其深度较深,而且超声采集时间较短,受患者呼吸暂停时间的限制。此外,还比较了原生肾脏和移植物肾脏的传感超声定位显微镜参数及其对肾小球检测的影响:对 15 名原生肾脏患者和 5 名异体肾脏患者进行了超声定位显微镜检查。使用投射到 sULM 密度图上的归一化距离指标对肾小球进行计数。评估了原生肾和移植肾在采集时间、肾脏深度和帧频上的差异,以及它们对肾小球检测的影响:结果:15 名原生肾脏患者中有 12 人实现了肾小球可视化。2名患者因无法憋气而失败,1名患者因肾脏太深而失败。传感超声定位显微镜在原生肾脏中发现每平方厘米 16 个肾小球(6-31 个),在肾移植患者中发现每平方厘米 33 个肾小球(18-55 个):本研究表明,sULM 可以观察到人体原生肾脏中的肾小球。所提出的方法可能会有许多假定的应用,包括生物标记物的开发、辅助活组织检查或有可能避免活组织检查。它为改善局部微观结构病理学的检测、影响异体移植的评估以及促进原生肾脏的疾病监测建立了一个框架。
{"title":"Visualization of Renal Glomeruli in Human Native Kidneys With Sensing Ultrasound Localization Microscopy.","authors":"Sylvain Bodard, Louise Denis, Georges Chabouh, Jacques Battaglia, Dany Anglicheau, Olivier Hélénon, Jean-Michel Correas, Olivier Couture","doi":"10.1097/RLI.0000000000001061","DOIUrl":"10.1097/RLI.0000000000001061","url":null,"abstract":"<p><strong>Objectives: </strong>Kidney diseases significantly impact individuals' quality of life and strongly reduce life expectancy. Glomeruli play a crucial role in kidney function. Current imaging techniques cannot visualize them due to their small size. Sensing ultrasound localization microscopy (sULM) has shown promising results for visualizing in vivo the glomeruli of human kidney grafts. This study aimed to evaluate the ability of sULM to visualize glomeruli in vivo in native human kidneys despite their depth and a shorter duration of ultrasound acquisition limited by the period of the patient's apnea. Sensing ultrasound localization microscopy parameters in native kidneys and kidney grafts and their consequence regarding glomeruli detection were also compared.</p><p><strong>Materials and methods: </strong>Exploration by sULM was conducted in 15 patients with native kidneys and 5 with kidney allografts. Glomeruli were counted using a normalized distance metric projected onto sULM density maps. The difference in the acquisition time, the kidney depth, and the frame rate between native kidneys and kidney grafts and their consequence regarding glomeruli detection were assessed.</p><p><strong>Results: </strong>Glomerular visualization was achieved in 12 of 15 patients with native kidneys. It failed due to impossible breath-holding for 2 patients and a too-deep kidney for 1 patient. Sensing ultrasound localization microscopy found 16 glomeruli per square centimeter in the native kidneys (6-31) and 33 glomeruli per square centimeter in kidney transplant patients (18-55).</p><p><strong>Conclusions: </strong>This study demonstrated that sULM can visualize glomeruli in native human kidneys in vivo. The proposed method may have many hypothetical applications, including biomarker development, assisting biopsy, or potentially avoiding it. It establishes a framework for improving the detection of local microstructural pathology, influencing the evaluation of allografts, and facilitating disease monitoring in the native kidney.</p>","PeriodicalId":14486,"journal":{"name":"Investigative Radiology","volume":" ","pages":"561-568"},"PeriodicalIF":7.0,"publicationDate":"2024-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"139424709","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
期刊
Investigative Radiology
全部 Acc. Chem. Res. ACS Applied Bio Materials ACS Appl. Electron. Mater. ACS Appl. Energy Mater. ACS Appl. Mater. Interfaces ACS Appl. Nano Mater. ACS Appl. Polym. Mater. ACS BIOMATER-SCI ENG ACS Catal. ACS Cent. Sci. ACS Chem. Biol. ACS Chemical Health & Safety ACS Chem. Neurosci. ACS Comb. Sci. ACS Earth Space Chem. ACS Energy Lett. ACS Infect. Dis. ACS Macro Lett. ACS Mater. Lett. ACS Med. Chem. Lett. ACS Nano ACS Omega ACS Photonics ACS Sens. ACS Sustainable Chem. Eng. ACS Synth. Biol. Anal. Chem. BIOCHEMISTRY-US Bioconjugate Chem. BIOMACROMOLECULES Chem. Res. Toxicol. Chem. Rev. Chem. Mater. CRYST GROWTH DES ENERG FUEL Environ. Sci. Technol. Environ. Sci. Technol. Lett. Eur. J. Inorg. Chem. IND ENG CHEM RES Inorg. Chem. J. Agric. Food. Chem. J. Chem. Eng. Data J. Chem. Educ. J. Chem. Inf. Model. J. Chem. Theory Comput. J. Med. Chem. J. Nat. Prod. J PROTEOME RES J. Am. Chem. Soc. LANGMUIR MACROMOLECULES Mol. Pharmaceutics Nano Lett. Org. Lett. ORG PROCESS RES DEV ORGANOMETALLICS J. Org. Chem. J. Phys. Chem. J. Phys. Chem. A J. Phys. Chem. B J. Phys. Chem. C J. Phys. Chem. Lett. Analyst Anal. Methods Biomater. Sci. Catal. Sci. Technol. Chem. Commun. Chem. Soc. Rev. CHEM EDUC RES PRACT CRYSTENGCOMM Dalton Trans. Energy Environ. Sci. ENVIRON SCI-NANO ENVIRON SCI-PROC IMP ENVIRON SCI-WAT RES Faraday Discuss. Food Funct. Green Chem. Inorg. Chem. Front. Integr. Biol. J. Anal. At. Spectrom. J. Mater. Chem. A J. Mater. Chem. B J. Mater. Chem. C Lab Chip Mater. Chem. Front. Mater. Horiz. MEDCHEMCOMM Metallomics Mol. Biosyst. Mol. Syst. Des. Eng. Nanoscale Nanoscale Horiz. Nat. Prod. Rep. New J. Chem. Org. Biomol. Chem. Org. Chem. Front. PHOTOCH PHOTOBIO SCI PCCP Polym. Chem.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1