Rebekah Goldblatt, Florencia Monge, Cory Hansen, Melissa M Budelier
Background: Tacrolimus therapeutic drug monitoring is a routine and critical assay for managing transplant patients. Tacrolimus can be measured from whole blood samples via liquid chromatography-tandem mass spectrometry (LC-MS/MS) or immunoassay (IA). LC-MS/MS provides accurate parent drug quantification with minimal metabolite interference but requires expensive analyzers, highly trained staff, and is often a very manual process. Immunoassays offer faster, simpler workflows, but historically have been prone to significant cross-reaction with the drug's metabolites or other therapeutic drugs, leading to a falsely elevated result. Newer immunoassays show promising improvements in specificity.
Methods: Using 73 residual patient samples, we compared our in-house LC-MS/MS tacrolimus assay to the Roche Elecsys® tacrolimus electrochemiluminescence immunoassay on the cobas® e801 analyzer. Comparative analysis was done with Deming regression and Bland-Altman comparison. Linearity, within-individual imprecision, and total imprecision post immunoassay implementation were also evaluated, along with changes and improvements to workflow and turnaround times (TAT).
Results: The Roche Elecsys tacrolimus assay correlated well with our LC-MS/MS method (Deming regression equation: y = 1.004x + 0.4848). The bias was 0.52 ng/mL (an average of 8.1%) as determined by Bland-Altman comparison. When switching to the immunoassay we observed a 42%-51% improvement in TAT across the median, 75th percentile, and 90th percentile.
Conclusions: We observed excellent correlation between the Roche Elecsys tacrolimus assay on an e801 analyzer vs our in-house LC-MS/MS assay, with good precision and linearity across reportable ranges. Changing from mass spectrometry to tacrolimus immunoassay substantially improved our TAT and workflow.
{"title":"Tacrolimus Assay Comparison: Analytical and Operational Considerations for Immunoassay vs Mass Spectrometry Testing in a High Volume Clinical Laboratory.","authors":"Rebekah Goldblatt, Florencia Monge, Cory Hansen, Melissa M Budelier","doi":"10.1093/jalm/jfaf199","DOIUrl":"https://doi.org/10.1093/jalm/jfaf199","url":null,"abstract":"<p><strong>Background: </strong>Tacrolimus therapeutic drug monitoring is a routine and critical assay for managing transplant patients. Tacrolimus can be measured from whole blood samples via liquid chromatography-tandem mass spectrometry (LC-MS/MS) or immunoassay (IA). LC-MS/MS provides accurate parent drug quantification with minimal metabolite interference but requires expensive analyzers, highly trained staff, and is often a very manual process. Immunoassays offer faster, simpler workflows, but historically have been prone to significant cross-reaction with the drug's metabolites or other therapeutic drugs, leading to a falsely elevated result. Newer immunoassays show promising improvements in specificity.</p><p><strong>Methods: </strong>Using 73 residual patient samples, we compared our in-house LC-MS/MS tacrolimus assay to the Roche Elecsys® tacrolimus electrochemiluminescence immunoassay on the cobas® e801 analyzer. Comparative analysis was done with Deming regression and Bland-Altman comparison. Linearity, within-individual imprecision, and total imprecision post immunoassay implementation were also evaluated, along with changes and improvements to workflow and turnaround times (TAT).</p><p><strong>Results: </strong>The Roche Elecsys tacrolimus assay correlated well with our LC-MS/MS method (Deming regression equation: y = 1.004x + 0.4848). The bias was 0.52 ng/mL (an average of 8.1%) as determined by Bland-Altman comparison. When switching to the immunoassay we observed a 42%-51% improvement in TAT across the median, 75th percentile, and 90th percentile.</p><p><strong>Conclusions: </strong>We observed excellent correlation between the Roche Elecsys tacrolimus assay on an e801 analyzer vs our in-house LC-MS/MS assay, with good precision and linearity across reportable ranges. Changing from mass spectrometry to tacrolimus immunoassay substantially improved our TAT and workflow.</p>","PeriodicalId":46361,"journal":{"name":"Journal of Applied Laboratory Medicine","volume":" ","pages":""},"PeriodicalIF":1.9,"publicationDate":"2026-01-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145918924","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
Maurizio Pasi, Ilaria Giardini, Eleonora Buscaglia, Azzurra Schicchi, Carlo A Locatelli, Andrea Giampreti, Mariapina Gallo, Riccardo Albertini, Pietro Papa, Antonella Valli
Background: The rise of synthetic cathinones (SC), and in particular 3,4-MDPHP (3,4-methylenedioxy-α-pyrrolidinohexanophenone), is a concerning health threat. This paper represents the most extensive case series on 3,4-MDPHP suspected intoxications to date, offering a detailed analysis of its impact.
Methods: This study presents the analytical findings for 108 hospitalized patients who tested positive for 3,4-MDPHP, out of a total of 465 subjects admitted for suspected new psychoactive substances (NPS) intoxication. Analyses were performed by GC-MS and LC-MS/MS. The potential cross-reactivity of 3,4-MDPHP with the Syva EMIT II Plus urine ecstasy immunoassay is also investigated.
Results: 3,4-MDPHP was detected in 56% of NPS positive cases. A wide range of 3,4-MDPHP concentrations was observed in both blood (1-257 ng/mL) and urine (2-32 250 ng/mL). 3,4-MDPHP was the sole detected substance in 73.1% of cases, while 26.9% involved co-consumption with other drugs of abuse, primarily cocaine. Investigations about 3,4-MDPHP cross-reactivity with Syva EMIT II Plus urine ecstasy demonstrated that unmodified 3,4-MDPHP does not trigger a positive result, but its metabolites seem to be involved in a positive cross-reaction.
Conclusion: This study confirms the high prevalence (108/465) of 3,4-MDPHP in NPS related intoxication cases in our cohort. It is essential for clinical laboratories and emergency departments to be aware of the potential cross-reactivity in ecstasy immunoassay, as this may lead to significant diagnostic errors and misinterpretation of results.
背景:合成卡西酮(SC)的兴起,特别是3,4- mdphp(3,4-亚甲二氧基-α-吡咯烷二己烯酮),是一个令人担忧的健康威胁。本文代表了迄今为止最广泛的3,4- mdphp疑似中毒病例系列,并提供了其影响的详细分析。方法:在465例疑似新型精神活性物质(NPS)中毒住院患者中,108例3,4- mdphp检测呈阳性。采用GC-MS和LC-MS/MS进行分析。我们还研究了3,4- mdphp与Syva EMIT II Plus尿狂喜免疫分析法的潜在交叉反应性。结果:56%的NPS阳性病例检出3,4- mdphp。在血液(1-257 ng/mL)和尿液(2-32 250 ng/mL)中均观察到3,4- mdphp浓度的大范围变化。在73.1%的病例中,3,4- mdphp是唯一检测到的物质,而26.9%的病例涉及与其他滥用药物(主要是可卡因)共同使用。对3,4- mdphp与Syva EMIT II Plus尿液摇头丸的交叉反应性的研究表明,未经修饰的3,4- mdphp不会引发阳性结果,但其代谢物似乎参与了阳性交叉反应。结论:本研究证实了我们队列中NPS相关中毒病例中3,4- mdphp的高患病率(108/465)。临床实验室和急诊科必须意识到摇头丸免疫测定中潜在的交叉反应性,因为这可能导致重大的诊断错误和对结果的误解。
{"title":"Prevalence of 3,4-Methylenedioxy-α-Pyrrolidinohexanophenone (3,4-MDPHP) in 465 Suspected Intoxication Cases and Cross-Reactivity Evaluation in an Ecstasy Immunoassay.","authors":"Maurizio Pasi, Ilaria Giardini, Eleonora Buscaglia, Azzurra Schicchi, Carlo A Locatelli, Andrea Giampreti, Mariapina Gallo, Riccardo Albertini, Pietro Papa, Antonella Valli","doi":"10.1093/jalm/jfaf165","DOIUrl":"10.1093/jalm/jfaf165","url":null,"abstract":"<p><strong>Background: </strong>The rise of synthetic cathinones (SC), and in particular 3,4-MDPHP (3,4-methylenedioxy-α-pyrrolidinohexanophenone), is a concerning health threat. This paper represents the most extensive case series on 3,4-MDPHP suspected intoxications to date, offering a detailed analysis of its impact.</p><p><strong>Methods: </strong>This study presents the analytical findings for 108 hospitalized patients who tested positive for 3,4-MDPHP, out of a total of 465 subjects admitted for suspected new psychoactive substances (NPS) intoxication. Analyses were performed by GC-MS and LC-MS/MS. The potential cross-reactivity of 3,4-MDPHP with the Syva EMIT II Plus urine ecstasy immunoassay is also investigated.</p><p><strong>Results: </strong>3,4-MDPHP was detected in 56% of NPS positive cases. A wide range of 3,4-MDPHP concentrations was observed in both blood (1-257 ng/mL) and urine (2-32 250 ng/mL). 3,4-MDPHP was the sole detected substance in 73.1% of cases, while 26.9% involved co-consumption with other drugs of abuse, primarily cocaine. Investigations about 3,4-MDPHP cross-reactivity with Syva EMIT II Plus urine ecstasy demonstrated that unmodified 3,4-MDPHP does not trigger a positive result, but its metabolites seem to be involved in a positive cross-reaction.</p><p><strong>Conclusion: </strong>This study confirms the high prevalence (108/465) of 3,4-MDPHP in NPS related intoxication cases in our cohort. It is essential for clinical laboratories and emergency departments to be aware of the potential cross-reactivity in ecstasy immunoassay, as this may lead to significant diagnostic errors and misinterpretation of results.</p>","PeriodicalId":46361,"journal":{"name":"Journal of Applied Laboratory Medicine","volume":" ","pages":"61-72"},"PeriodicalIF":1.9,"publicationDate":"2026-01-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145507639","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
{"title":"The Past, the Present, and the Future for The Journal of Applied Laboratory Medicine.","authors":"Ian S Young, Allison B Chambliss","doi":"10.1093/jalm/jfaf176","DOIUrl":"10.1093/jalm/jfaf176","url":null,"abstract":"","PeriodicalId":46361,"journal":{"name":"Journal of Applied Laboratory Medicine","volume":" ","pages":"1-3"},"PeriodicalIF":1.9,"publicationDate":"2026-01-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145543103","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
Jacob B Nielsen, Rongrong Huang, Roger L Bertholf, Xin Yi
{"title":"Incidental Detection of a Hemoglobin Variant Reveals the Cause of Chronic Dyspnea.","authors":"Jacob B Nielsen, Rongrong Huang, Roger L Bertholf, Xin Yi","doi":"10.1093/jalm/jfaf136","DOIUrl":"10.1093/jalm/jfaf136","url":null,"abstract":"","PeriodicalId":46361,"journal":{"name":"Journal of Applied Laboratory Medicine","volume":" ","pages":"188-192"},"PeriodicalIF":1.9,"publicationDate":"2026-01-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145356420","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
Christine L H Snozek, Matthew D Krasowski, Jennifer M Colby, Kamisha L Johnson-Davis, Rebecca E Bruccoleri, Stacy E Melanson
Toxicology testing for drugs associated with scenarios such as recreational use or substance use disorder can be performed in support of the emergency department (ED) for specific patient populations such as pediatrics and trauma. These compounds were historically referred to as drugs of abuse (DOA); although the word "abuse" is recognized as potentially stigmatizing, no replacement terminology for DOA has emerged in current guidelines. This document refers to these compounds as drugs or substances of misuse, and acknowledges the need for less-stigmatizing language that more fully encompasses the range of uses for these drugs. This literature-driven, consensus guidance document provides recommendations primarily targeted to US hospital-based laboratories performing urine drug testing (UDT) in support of the ED. Indications for ordering UDT and related testing in both pediatric and adult populations are summarized. Further, recommendations are made for testing that should be available at all facilities with rapid turnaround, and how to perform and report testing. The advantages and disadvantages of immunoassays and mass spectrometry, as well as common challenges, are reviewed. Indications for mass-spectrometry assays and more extensive testing (e.g., novel psychoactive substances) are also provided. Future directions for improvements in laboratory technology to improve the utility of this testing are outlined. All laboratories should collaborate with ED leadership, medical toxicologists and poison control centers to optimize and update test menus to reflect local drug use patterns, ensure test methodologies and results meet clinical needs, and educate clinical staff regarding assay limitations and accurate test interpretation.
{"title":"ADLM Guidance Document on Laboratory Testing for Drugs of Misuse to Support the Emergency Department.","authors":"Christine L H Snozek, Matthew D Krasowski, Jennifer M Colby, Kamisha L Johnson-Davis, Rebecca E Bruccoleri, Stacy E Melanson","doi":"10.1093/jalm/jfaf172","DOIUrl":"https://doi.org/10.1093/jalm/jfaf172","url":null,"abstract":"<p><p>Toxicology testing for drugs associated with scenarios such as recreational use or substance use disorder can be performed in support of the emergency department (ED) for specific patient populations such as pediatrics and trauma. These compounds were historically referred to as drugs of abuse (DOA); although the word \"abuse\" is recognized as potentially stigmatizing, no replacement terminology for DOA has emerged in current guidelines. This document refers to these compounds as drugs or substances of misuse, and acknowledges the need for less-stigmatizing language that more fully encompasses the range of uses for these drugs. This literature-driven, consensus guidance document provides recommendations primarily targeted to US hospital-based laboratories performing urine drug testing (UDT) in support of the ED. Indications for ordering UDT and related testing in both pediatric and adult populations are summarized. Further, recommendations are made for testing that should be available at all facilities with rapid turnaround, and how to perform and report testing. The advantages and disadvantages of immunoassays and mass spectrometry, as well as common challenges, are reviewed. Indications for mass-spectrometry assays and more extensive testing (e.g., novel psychoactive substances) are also provided. Future directions for improvements in laboratory technology to improve the utility of this testing are outlined. All laboratories should collaborate with ED leadership, medical toxicologists and poison control centers to optimize and update test menus to reflect local drug use patterns, ensure test methodologies and results meet clinical needs, and educate clinical staff regarding assay limitations and accurate test interpretation.</p>","PeriodicalId":46361,"journal":{"name":"Journal of Applied Laboratory Medicine","volume":"11 1","pages":"155-180"},"PeriodicalIF":1.9,"publicationDate":"2026-01-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145906801","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
Sara Martínez-Rodríguez, Alberto Izquierdo-Martínez, Juan Miguel Guerrero Montávez, Ricardo Rubio-Sánchez
Background: Analytical interferences are very common in clinical laboratories, so professionals must develop strategies for their detection, avoiding incorrect results that can lead to inappropriate diagnoses and treatments.
Methods: An isolated 1040 error (absorbance-related) in the Alanine Aminotransferase2 (ALT2) assay performed on the Abbott Alinity c that occurred in 158 samples over 7 months was investigated. Highly lipemic or hemolyzed samples were excluded, and an error due to an increased concentration of total proteins was ruled out, all of which are documented analytical interferences.
Results: We isolated immunoglobulin (Ig) with an increased concentration (monoclonal components: 149 IgM, 7 IgG, and 2 IgA) from all analyzed samples, so the presence of this error solely in the alanine aminotransferase (ALT) assay had a 100% positive predictive value for monoclonal gammopathy. Serum viscosity was elevated in all cases, which is the reason for the detected interference. Treatment of IgM samples with dithiothreitol confirmed that dissociation of the pentamers eliminates the error in ALT determination.
Conclusions: The detection of this interference in samples from patients without recent immunoglobulin determinations indicates the presence of a significant and isolated increase in the concentration of one of them, potentially leading to the diagnosis of a previously unknown monoclonal gammopathy.
{"title":"Analytical Interference in Alanine Aminotransferase Determination in Patients with Monoclonal Gammopathy.","authors":"Sara Martínez-Rodríguez, Alberto Izquierdo-Martínez, Juan Miguel Guerrero Montávez, Ricardo Rubio-Sánchez","doi":"10.1093/jalm/jfaf115","DOIUrl":"10.1093/jalm/jfaf115","url":null,"abstract":"<p><strong>Background: </strong>Analytical interferences are very common in clinical laboratories, so professionals must develop strategies for their detection, avoiding incorrect results that can lead to inappropriate diagnoses and treatments.</p><p><strong>Methods: </strong>An isolated 1040 error (absorbance-related) in the Alanine Aminotransferase2 (ALT2) assay performed on the Abbott Alinity c that occurred in 158 samples over 7 months was investigated. Highly lipemic or hemolyzed samples were excluded, and an error due to an increased concentration of total proteins was ruled out, all of which are documented analytical interferences.</p><p><strong>Results: </strong>We isolated immunoglobulin (Ig) with an increased concentration (monoclonal components: 149 IgM, 7 IgG, and 2 IgA) from all analyzed samples, so the presence of this error solely in the alanine aminotransferase (ALT) assay had a 100% positive predictive value for monoclonal gammopathy. Serum viscosity was elevated in all cases, which is the reason for the detected interference. Treatment of IgM samples with dithiothreitol confirmed that dissociation of the pentamers eliminates the error in ALT determination.</p><p><strong>Conclusions: </strong>The detection of this interference in samples from patients without recent immunoglobulin determinations indicates the presence of a significant and isolated increase in the concentration of one of them, potentially leading to the diagnosis of a previously unknown monoclonal gammopathy.</p>","PeriodicalId":46361,"journal":{"name":"Journal of Applied Laboratory Medicine","volume":" ","pages":"120-124"},"PeriodicalIF":1.9,"publicationDate":"2026-01-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"144973823","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
{"title":"Commentary on Incidental Detection of a Hemoglobin Variant Reveals the Cause of Chronic Dyspnea.","authors":"Catherine L Omosule","doi":"10.1093/jalm/jfaf162","DOIUrl":"10.1093/jalm/jfaf162","url":null,"abstract":"","PeriodicalId":46361,"journal":{"name":"Journal of Applied Laboratory Medicine","volume":" ","pages":"193-194"},"PeriodicalIF":1.9,"publicationDate":"2026-01-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145356422","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
{"title":"Commentary on A Pediatric Case of Recurrent Dehydration and Shock Revealing an Endocrine Disorder.","authors":"Kyle P McNerney","doi":"10.1093/jalm/jfaf163","DOIUrl":"10.1093/jalm/jfaf163","url":null,"abstract":"","PeriodicalId":46361,"journal":{"name":"Journal of Applied Laboratory Medicine","volume":" ","pages":"186-187"},"PeriodicalIF":1.9,"publicationDate":"2026-01-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145402363","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
Background: In vitro diagnostic (IVD) devices are essential for disease detection and clinical decision-making. India's IVD market has grown rapidly, but regulatory oversight has lagged. Although the Indian Medical Device Rules introduced a risk-based framework in 2017, persistent gaps in enforcement, quality standards, and post-market surveillance undermine diagnostic reliability.
Content: This review examines India's IVD regulatory framework, focusing on device classification, approval pathways, quality management requirements, and post-market mechanisms. Particular attention is given to challenges such as enforcement issues, inconsistent compliance among small and medium enterprises, adoption barriers for national/international standards, weak post-market surveillance systems, and the absence of transparent public databases. Comparative analysis highlights India's shortfalls relative to the European and US systems, especially in mandatory clinical evidence, structured post-market surveillance, and stakeholder engagement.
Summary: Despite progress, India's IVD ecosystem remains constrained by fragmented oversight and weak harmonization with global standards. Strengthening regulatory capacity, mandating evidence-based evaluation, and fostering digital health readiness are critical to ensure patient safety and global competitiveness.
{"title":"Regulatory Framework for In Vitro Diagnostic Devices in India: Current Landscape, Challenges, and Future Perspectives.","authors":"Mahima Ramesh, Harsh Sah, Amrutha C","doi":"10.1093/jalm/jfaf164","DOIUrl":"10.1093/jalm/jfaf164","url":null,"abstract":"<p><strong>Background: </strong>In vitro diagnostic (IVD) devices are essential for disease detection and clinical decision-making. India's IVD market has grown rapidly, but regulatory oversight has lagged. Although the Indian Medical Device Rules introduced a risk-based framework in 2017, persistent gaps in enforcement, quality standards, and post-market surveillance undermine diagnostic reliability.</p><p><strong>Content: </strong>This review examines India's IVD regulatory framework, focusing on device classification, approval pathways, quality management requirements, and post-market mechanisms. Particular attention is given to challenges such as enforcement issues, inconsistent compliance among small and medium enterprises, adoption barriers for national/international standards, weak post-market surveillance systems, and the absence of transparent public databases. Comparative analysis highlights India's shortfalls relative to the European and US systems, especially in mandatory clinical evidence, structured post-market surveillance, and stakeholder engagement.</p><p><strong>Summary: </strong>Despite progress, India's IVD ecosystem remains constrained by fragmented oversight and weak harmonization with global standards. Strengthening regulatory capacity, mandating evidence-based evaluation, and fostering digital health readiness are critical to ensure patient safety and global competitiveness.</p>","PeriodicalId":46361,"journal":{"name":"Journal of Applied Laboratory Medicine","volume":" ","pages":"125-142"},"PeriodicalIF":1.9,"publicationDate":"2026-01-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145490594","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
Nicholas E Larkey, Derek C Waggoner, Erica M Fatica
{"title":"Changing Chemistry Instruments? Don't Forget to Ask These Questions.","authors":"Nicholas E Larkey, Derek C Waggoner, Erica M Fatica","doi":"10.1093/jalm/jfaf158","DOIUrl":"https://doi.org/10.1093/jalm/jfaf158","url":null,"abstract":"","PeriodicalId":46361,"journal":{"name":"Journal of Applied Laboratory Medicine","volume":"11 1","pages":"197-202"},"PeriodicalIF":1.9,"publicationDate":"2026-01-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"145906807","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}