首页 > 最新文献

Journal of biomechanics最新文献

英文 中文
Repeatability of two methods for estimating scapular kinematics during dynamic functional tasks 估算动态功能任务中肩胛骨运动学的两种方法的重复性
IF 2.4 3区 医学 Q3 BIOPHYSICS Pub Date : 2024-09-18 DOI: 10.1016/j.jbiomech.2024.112334
Angelica E. Lang

Best practices for scapular motion tracking are still being determined. The repeatability of different scapular kinematic procedures needs to be evaluated. The purpose of this study was to assess the test-retest reliability of two scapular kinematic procedures: double calibration with AMC (D-AMC) and individualized linear modelling (LM). Ten healthy participants had their upper body movement tracked with optical motion capture in two identical sessions. Five scapular calibration poses were performed, and seven dynamic functional tasks were tested. Scapular angles were calculated from both procedures (D-AMC vs LM). The D-AMC approach uses two poses (neutral and maximum elevation) and tracks the scapula with a rigid cluster, while the LM approach predicts scapular positioning from humeral angles based on equations built from the calibration pose data. Angle waveforms and repeatability outcomes were compared. Internal and upward rotation angle waveforms were significantly different (p < 0.05) between kinematic procedures for some tasks, with maximum mean differences up to 17.3° and 23.2°, respectively. Overall, repeatability outcomes were similar between procedures, but the LM approach was slightly better for tilt and the D-AMC approach was notably improved for upward rotation in certain tasks. For example, minimal detectable changes during the Forward Transfer ranged from 6.9° to 11.9° for the D-AMC and 8.9° to 25.3° for the LM. Discrepancies between procedures may be a function of the calibration poses chosen. Additional calibration poses may improve the comparisons between procedures.

肩胛运动跟踪的最佳实践仍在确定之中。需要对不同肩胛运动学程序的重复性进行评估。本研究旨在评估两种肩胛骨运动学程序的重复测试可靠性:AMC 双校准(D-AMC)和个性化线性建模(LM)。十名健康参与者在两次相同的训练中使用光学运动捕捉对他们的上半身运动进行了跟踪。进行了五次肩胛骨校准姿势,并测试了七项动态功能任务。两种方法(D-AMC 与 LM)都计算出了肩胛角。D-AMC 方法使用两个姿势(中立和最大仰角),并用刚性集群跟踪肩胛骨,而 LM 方法则根据校准姿势数据建立的方程,从肱骨角度预测肩胛骨定位。对角度波形和重复性结果进行了比较。在某些任务中,不同运动学程序的内旋和上旋角度波形有显著差异(p < 0.05),最大平均差异分别高达 17.3° 和 23.2°。总体而言,不同程序之间的重复性结果相似,但在某些任务中,LM 方法在倾斜方面略胜一筹,D-AMC 方法在向上旋转方面有明显改善。例如,在 "前向转移 "过程中,D-AMC方法可检测到的最小变化范围为6.9°至11.9°,LM方法为8.9°至25.3°。程序之间的差异可能与所选的校准姿势有关。额外的校准姿势可能会改善程序之间的比较。
{"title":"Repeatability of two methods for estimating scapular kinematics during dynamic functional tasks","authors":"Angelica E. Lang","doi":"10.1016/j.jbiomech.2024.112334","DOIUrl":"10.1016/j.jbiomech.2024.112334","url":null,"abstract":"<div><p>Best practices for scapular motion tracking are still being determined. The repeatability of different scapular kinematic procedures needs to be evaluated. The purpose of this study was to assess the test-retest reliability of two scapular kinematic procedures: double calibration with AMC (D-AMC) and individualized linear modelling (LM). Ten healthy participants had their upper body movement tracked with optical motion capture in two identical sessions. Five scapular calibration poses were performed, and seven dynamic functional tasks were tested. Scapular angles were calculated from both procedures (D-AMC vs LM). The D-AMC approach uses two poses (neutral and maximum elevation) and tracks the scapula with a rigid cluster, while the LM approach predicts scapular positioning from humeral angles based on equations built from the calibration pose data. Angle waveforms and repeatability outcomes were compared. Internal and upward rotation angle waveforms were significantly different (p &lt; 0.05) between kinematic procedures for some tasks, with maximum mean differences up to 17.3° and 23.2°, respectively. Overall, repeatability outcomes were similar between procedures, but the LM approach was slightly better for tilt and the D-AMC approach was notably improved for upward rotation in certain tasks. For example, minimal detectable changes during the Forward Transfer ranged from 6.9° to 11.9° for the D-AMC and 8.9° to 25.3° for the LM. Discrepancies between procedures may be a function of the calibration poses chosen. Additional calibration poses may improve the comparisons between procedures.</p></div>","PeriodicalId":15168,"journal":{"name":"Journal of biomechanics","volume":"176 ","pages":"Article 112334"},"PeriodicalIF":2.4,"publicationDate":"2024-09-18","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S0021929024004123/pdfft?md5=db359fa6f3f2d4ac7ce570c39baf45ce&pid=1-s2.0-S0021929024004123-main.pdf","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"142272729","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
In Silico clinical trial to predict the efficacy of hip protectors for preventing hip fractures 预测髋部保护器预防髋部骨折疗效的硅学临床试验
IF 2.4 3区 医学 Q3 BIOPHYSICS Pub Date : 2024-09-18 DOI: 10.1016/j.jbiomech.2024.112335
Sara Oliviero , Antonino A. La Mattina , Giacomo Savelli , Marco Viceconti

Osteoporosis is characterized by loss of bone mineral density and increased fracture risk. Reduction of hip fracture incidence is of major clinical importance. Hip protectors aim to attenuate the impact force transmitted to the femur upon falling, however different conclusions on their efficacy have been reported; some authors suggest this may be due to differences in compliance. The aim of this study was to apply an In Silico trial methodology to predict the effectiveness of hip protectors and its dependence on compliance.

A cohort of 1044 virtual patients (Finite Element models of proximal femur) were generated. A Markov chain process was implemented to predict fracture incidence with and without hip protectors, by simulating different levels of compliance. At each simulated follow-up year, a Poisson distribution was randomly sampled to determine the number of falls sustained by each patient. Impact direction and force were stochastically sampled from a range of possible scenarios. The effect of wearing a hip protector was simulated by applying attenuation coefficients to the impact force (12.9 %, 19 % and 33.8 %, as reported for available devices). A patient was considered fractured when impact force exceeded the femur strength.

Without hip protector, virtual patients experienced 66 ± 5 fractures in 10 years. Wearing the three devices, fracture incidence was reduced to 43 ± 4, 35 ± 4 and 17 ± 2 respectively, at full compliance. As expected, effectiveness was dependent on compliance.

This In Silico trial technology can be applied in the future to test multiple interventions, optimise intervention strategies, improve clinical trial design and drug development.

骨质疏松症的特点是骨矿物质密度下降和骨折风险增加。降低髋部骨折的发生率具有重要的临床意义。髋部保护器旨在减弱跌倒时传递到股骨的冲击力,但有关其功效的结论却不尽相同;一些作者认为这可能是由于顺应性的差异造成的。本研究的目的是应用 In Silico 试验方法来预测护髋的有效性及其对顺应性的依赖性。通过模拟不同程度的依从性,实施马尔可夫链过程来预测使用和不使用髋部保护器的骨折发生率。在每个模拟随访年,随机抽取泊松分布,以确定每位患者的跌倒次数。撞击方向和撞击力从一系列可能的情况中随机取样。通过对撞击力应用衰减系数(12.9%、19% 和 33.8%,根据现有设备的报告)来模拟佩戴髋部保护器的效果。当冲击力超过股骨强度时,患者即被视为骨折。如果不佩戴髋关节保护器,虚拟患者在 10 年内会发生 66 ± 5 次骨折。佩戴这三种装置后,骨折发生率在完全达标的情况下分别降至 43 ± 4、35 ± 4 和 17 ± 2。正如预期的那样,有效性取决于依从性。这种 In Silico 试验技术今后可用于测试多种干预措施、优化干预策略、改进临床试验设计和药物开发。
{"title":"In Silico clinical trial to predict the efficacy of hip protectors for preventing hip fractures","authors":"Sara Oliviero ,&nbsp;Antonino A. La Mattina ,&nbsp;Giacomo Savelli ,&nbsp;Marco Viceconti","doi":"10.1016/j.jbiomech.2024.112335","DOIUrl":"10.1016/j.jbiomech.2024.112335","url":null,"abstract":"<div><p>Osteoporosis is characterized by loss of bone mineral density and increased fracture risk. Reduction of hip fracture incidence is of major clinical importance. Hip protectors aim to attenuate the impact force transmitted to the femur upon falling, however different conclusions on their efficacy have been reported; some authors suggest this may be due to differences in compliance. The aim of this study was to apply an <em>In Silico</em> trial methodology to predict the effectiveness of hip protectors and its dependence on compliance.</p><p>A cohort of 1044 virtual patients (Finite Element models of proximal femur) were generated. A Markov chain process was implemented to predict fracture incidence with and without hip protectors, by simulating different levels of compliance. At each simulated follow-up year, a Poisson distribution was randomly sampled to determine the number of falls sustained by each patient. Impact direction and force were stochastically sampled from a range of possible scenarios. The effect of wearing a hip protector was simulated by applying attenuation coefficients to the impact force (12.9 %, 19 % and 33.8 %, as reported for available devices). A patient was considered fractured when impact force exceeded the femur strength.</p><p>Without hip protector, virtual patients experienced 66 ± 5 fractures in 10 years. Wearing the three devices, fracture incidence was reduced to 43 ± 4, 35 ± 4 and 17 ± 2 respectively, at full compliance. As expected, effectiveness was dependent on compliance.</p><p>This <em>In Silico</em> trial technology can be applied in the future to test multiple interventions, optimise intervention strategies, improve clinical trial design and drug development.</p></div>","PeriodicalId":15168,"journal":{"name":"Journal of biomechanics","volume":"176 ","pages":"Article 112335"},"PeriodicalIF":2.4,"publicationDate":"2024-09-18","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S0021929024004135/pdfft?md5=3a4fe634f56146bf7ef130db83fd5829&pid=1-s2.0-S0021929024004135-main.pdf","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"142272724","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Simulating vortex generation to investigate the propulsive and braking mechanisms of breaststroke kick using computational fluid dynamics on a breaststroke swimmer 利用蛙泳运动员的计算流体动力学模拟涡流产生,研究蛙泳踢腿的推进和制动机制
IF 2.4 3区 医学 Q3 BIOPHYSICS Pub Date : 2024-09-17 DOI: 10.1016/j.jbiomech.2024.112329
Takahiro Tanaka , Taisei Hayashi , Tadao Isaka

Swimmers primarily increase their forward velocity through lower limb motion in breaststroke, making the breaststroke kick crucial for optimizing race times. Recent studies have highlighted the generation of vortices around the swimmer’s entire body to propel forward during swimming. However, the investigation of vortex generation during breaststroke kicks remains unexplored. This study aimed to reveal the propulsive and braking mechanisms of breaststroke kicks by simulating vortex generation using computational fluid dynamics (CFD). Kinematic data during the breaststroke kick and a three-dimensional digital model were collected to conduct CFD for a male breaststroke swimmer. Vortex generation was determined during one breaststroke kick from the CFD results. Vortices, which potentially induce a decrease in forward velocity, were generated by the swimmer’s lower legs and feet during the recovery phase. The swimmer generated vortices on the dorsal side of the feet and the posterior and lateral sides of the lower legs to increase the forward velocity during the out-sweep phase. The swimmer generated vortices on the lateral sides of the thighs and lower legs and the dorsal and lateral sides of the feet during the in-sweep phase to maintain forward velocity. Moreover, vortices generated from the out-sweep to the in-sweep merged and were shed backward relative to the swimming direction after the in-sweep phase. This study is the first to reveal the propulsive and braking mechanisms of breaststroke kicks by analyzing the vortex generation.

游泳运动员在蛙泳中主要通过下肢运动提高前进速度,因此蛙泳踢腿对于优化比赛时间至关重要。最近的研究强调了游泳时游泳者整个身体周围产生的涡流推动前进。然而,对蛙泳踢腿时涡流产生的研究仍未展开。本研究旨在通过使用计算流体动力学(CFD)模拟涡流的产生,揭示蛙泳踢腿时的推进和制动机制。研究收集了一名男子蛙泳运动员在蛙泳踢腿过程中的运动学数据和三维数字模型,以进行 CFD 模拟。根据 CFD 结果确定了一次蛙泳踢腿过程中产生的涡流。在恢复阶段,有可能导致前进速度降低的涡流是由游泳者的小腿和脚产生的。在外侧扫水阶段,游泳者在脚背、小腿后侧和外侧产生涡流,以提高前进速度。在内侧扫水阶段,游泳者在大腿和小腿外侧以及脚背和外侧产生涡流,以保持前进速度。此外,从出泳到进泳产生的涡流在进泳阶段后合并并相对于游泳方向向后脱落。这项研究首次通过分析涡流的产生揭示了蛙泳踢腿的推进和制动机制。
{"title":"Simulating vortex generation to investigate the propulsive and braking mechanisms of breaststroke kick using computational fluid dynamics on a breaststroke swimmer","authors":"Takahiro Tanaka ,&nbsp;Taisei Hayashi ,&nbsp;Tadao Isaka","doi":"10.1016/j.jbiomech.2024.112329","DOIUrl":"10.1016/j.jbiomech.2024.112329","url":null,"abstract":"<div><p>Swimmers primarily increase their forward velocity through lower limb motion in breaststroke, making the breaststroke kick crucial for optimizing race times. Recent studies have highlighted the generation of vortices around the swimmer’s entire body to propel forward during swimming. However, the investigation of vortex generation during breaststroke kicks remains unexplored. This study aimed to reveal the propulsive and braking mechanisms of breaststroke kicks by simulating vortex generation using computational fluid dynamics (CFD). Kinematic data during the breaststroke kick and a three-dimensional digital model were collected to conduct CFD for a male breaststroke swimmer. Vortex generation was determined during one breaststroke kick from the CFD results. Vortices, which potentially induce a decrease in forward velocity, were generated by the swimmer’s lower legs and feet during the recovery phase. The swimmer generated vortices on the dorsal side of the feet and the posterior and lateral sides of the lower legs to increase the forward velocity during the out-sweep phase. The swimmer generated vortices on the lateral sides of the thighs and lower legs and the dorsal and lateral sides of the feet during the in-sweep phase to maintain forward velocity. Moreover, vortices generated from the out-sweep to the in-sweep merged and were shed backward relative to the swimming direction after the in-sweep phase. This study is the first to reveal the propulsive and braking mechanisms of breaststroke kicks by analyzing the vortex generation.</p></div>","PeriodicalId":15168,"journal":{"name":"Journal of biomechanics","volume":"176 ","pages":"Article 112329"},"PeriodicalIF":2.4,"publicationDate":"2024-09-17","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S002192902400407X/pdfft?md5=0906165a66ed8300e79db0c4f3d8af4b&pid=1-s2.0-S002192902400407X-main.pdf","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"142272725","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Walking in a controlled ankle motion (CAM) boot: In-boot measurement of joint kinematics and kinetics 穿着受控踝关节运动(CAM)靴行走:靴内关节运动学和动力学测量
IF 2.4 3区 医学 Q3 BIOPHYSICS Pub Date : 2024-09-16 DOI: 10.1016/j.jbiomech.2024.112327
Mason L. Stolycia , David E. Lunn , Richard A. Wilkins , Cleveland T. Barnett , Josh Walker

Research investigating ankle function during walking in a controlled ankle motion (CAM) boot has either placed markers on the outside of the boot or made major alterations to the structure of the CAM boot to uncover key landmarks. The aim of this study was to quantify joint kinematics and kinetics using “in-boot” skin markers whilst making only minimal structural alterations. Seventeen healthy participants walked at their preferred walking speed in two conditions: (1) in standard athletic trainers (ASICS patriot 8, ASICS Oceania Pty Ltd, USA), and (2) using a hard-cased CAM boot (Rebound® Air Walker, Össur, Iceland) fitted on the right foot. Kinematic measurements revealed that CAM boots restrict sagittal plane ankle range of motion to less than 5°, and to ∼3° in the frontal plane, which is a reduction of 85% and 73% compared to standard footwear, respectively (p < 0.001). This ankle restriction resulted in a reduction of ankle joint total limb work contribution from 38 ± 5% in normal footwear to 13 ± 4% in the CAM boot (p < 0.001). This study suggests that CAM boots do restrict the ankle joint’s ability to effectively perform work during walking, which leads to compensatory mechanisms at the ipsilateral and contralateral hip and knee joints. Our findings align with previous research that employed “on-boot” kinematic measurements, so we conclude that in-boot approaches do not offer any benefit to the researcher and instead, on-boot measurements are suitable.

对穿着受控踝关节运动(CAM)靴行走时踝关节功能的研究,要么是在靴子外部放置标记,要么是对 CAM 靴子的结构进行重大改动以发现关键地标。本研究旨在使用 "靴内 "皮肤标记对关节运动学和动力学进行量化,同时只对结构进行最小程度的改动。17 名健康参与者在两种条件下以自己喜欢的步行速度行走:(1) 穿着标准运动训练鞋(ASICS patriot 8,ASICS Oceania Pty Ltd,美国);(2) 右脚穿着硬壳 CAM 靴(Rebound® Air Walker,Össur,冰岛)。运动学测量显示,CAM 靴将矢状面的踝关节活动范围限制在 5° 以下,在正面的活动范围限制在 3° 以下,与标准鞋相比分别减少了 85% 和 73%(p < 0.001)。这种踝关节限制导致踝关节的肢体总工作贡献从穿普通鞋时的 38±5% 减少到穿 CAM 靴时的 13±4% (p < 0.001)。这项研究表明,CAM 靴确实限制了踝关节在行走过程中有效做功的能力,从而导致同侧和对侧髋关节和膝关节的代偿机制。我们的研究结果与之前采用 "靴上 "运动学测量方法的研究结果一致,因此我们得出结论,靴上测量方法不会给研究人员带来任何益处,相反,靴上测量方法是合适的。
{"title":"Walking in a controlled ankle motion (CAM) boot: In-boot measurement of joint kinematics and kinetics","authors":"Mason L. Stolycia ,&nbsp;David E. Lunn ,&nbsp;Richard A. Wilkins ,&nbsp;Cleveland T. Barnett ,&nbsp;Josh Walker","doi":"10.1016/j.jbiomech.2024.112327","DOIUrl":"10.1016/j.jbiomech.2024.112327","url":null,"abstract":"<div><p>Research investigating ankle function during walking in a controlled ankle motion (CAM) boot has either placed markers on the outside of the boot or made major alterations to the structure of the CAM boot to uncover key landmarks. The aim of this study was to quantify joint kinematics and kinetics using “in-boot” skin markers whilst making only minimal structural alterations. Seventeen healthy participants walked at their preferred walking speed in two conditions: (1) in standard athletic trainers (ASICS patriot 8, ASICS Oceania Pty Ltd, USA), and (2) using a hard-cased CAM boot (Rebound® Air Walker, Össur, Iceland) fitted on the right foot. Kinematic measurements revealed that CAM boots restrict sagittal plane ankle range of motion to less than 5°, and to ∼3° in the frontal plane, which is a reduction of 85% and 73% compared to standard footwear, respectively (<em>p</em> &lt; 0.001). This ankle restriction resulted in a reduction of ankle joint total limb work contribution from 38 ± 5% in normal footwear to 13 ± 4% in the CAM boot (<em>p</em> &lt; 0.001). This study suggests that CAM boots do restrict the ankle joint’s ability to effectively perform work during walking, which leads to compensatory mechanisms at the ipsilateral and contralateral hip and knee joints. Our findings align with previous research that employed “on-boot” kinematic measurements, so we conclude that in-boot approaches do not offer any benefit to the researcher and instead, on-boot measurements are suitable.</p></div>","PeriodicalId":15168,"journal":{"name":"Journal of biomechanics","volume":"176 ","pages":"Article 112327"},"PeriodicalIF":2.4,"publicationDate":"2024-09-16","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S0021929024004056/pdfft?md5=586cc230f36d8cc9b4ab3e58200ddb76&pid=1-s2.0-S0021929024004056-main.pdf","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"142233811","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Quantitative measurement of the flow depending nasal valve function by elastography with electro-optical distance sensors. A pilot study 利用电子光学距离传感器进行弹性成像,定量测量流量对鼻瓣膜功能的影响。试点研究
IF 2.4 3区 医学 Q3 BIOPHYSICS Pub Date : 2024-09-14 DOI: 10.1016/j.jbiomech.2024.112326
Sonali M. Bauer , Klaus Vogt , Klaus-D. Wernecke , Gerd Rasp , Karen Parker Davidson , Sebastian Roesch

Nasal valve function depends on the intensity of the inspiratory nasal airflow, the geometry of the nasal entrance and the mechanical properties of the lateral nasal wall. It is desirable to obtain objective information on the relation between flow and valve movement. In this study, the deflection of the lateral nasal wall and the inspiratory flow were measured on 30 healthy volunteers, aged 18 to 82 without a history of severe trauma or nasal surgery. Electro-optical distance sensors were housed under a full-face protective mask attached to an analogue inspiratory flowmeter. The mean values for normal breathing were assessed at 675 [cm3/s] for the bilateral flow and −0.57 mm for the total movement. With forced breathing, the mean values for the flow of both nostrils were found to be 1434 cm3/s and for the total movement −1.21 mm. Statistically significant differences between normal and forced breathing were found in all participants and in both sexes, but no significant correlation by age. Electro-optical distance measurement, representing a novel technical way for the ‘elastography’ of the nasal valve should be added to advanced 4-phase-rhinomanometers.

鼻瓣膜的功能取决于吸气时鼻腔气流的强度、鼻腔入口的几何形状和鼻侧壁的机械特性。最好能获得有关气流和瓣膜运动之间关系的客观信息。在这项研究中,对 30 名年龄在 18 至 82 岁之间、无严重外伤或鼻部手术史的健康志愿者进行了鼻侧壁偏转和吸气流量的测量。电子光学距离传感器安装在全脸防护面罩下,与模拟吸气流量计相连。据评估,正常呼吸时,双侧流量的平均值为 675 [cm3/s],总移动量为-0.57 mm。强制呼吸时,双侧鼻孔流量的平均值为 1434 立方厘米/秒,总移动量为-1.21 毫米。所有参与者和男女在正常呼吸和强制呼吸之间都发现了明显的统计学差异,但与年龄没有明显的相关性。电子光学距离测量是鼻腔瓣膜 "弹性成像 "的一种新技术方法,应添加到先进的四相血压计中。
{"title":"Quantitative measurement of the flow depending nasal valve function by elastography with electro-optical distance sensors. A pilot study","authors":"Sonali M. Bauer ,&nbsp;Klaus Vogt ,&nbsp;Klaus-D. Wernecke ,&nbsp;Gerd Rasp ,&nbsp;Karen Parker Davidson ,&nbsp;Sebastian Roesch","doi":"10.1016/j.jbiomech.2024.112326","DOIUrl":"10.1016/j.jbiomech.2024.112326","url":null,"abstract":"<div><p>Nasal valve function depends on the intensity of the inspiratory nasal airflow, the geometry of the nasal entrance and the mechanical properties of the lateral nasal wall. It is desirable to obtain objective information on the relation between flow and valve movement. In this study, the deflection of the lateral nasal wall and the inspiratory flow were measured on 30 healthy volunteers, aged 18 to 82 without a history of severe trauma or nasal surgery. Electro-optical distance sensors were housed under a full-face protective mask attached to an analogue inspiratory flowmeter. The mean values for normal breathing were assessed at 675 [cm<sup>3</sup>/s] for the bilateral flow and −0.57 mm for the total movement. With forced breathing, the mean values for the flow of both nostrils were found to be 1434 cm<sup>3</sup>/s and for the total movement −1.21 mm. Statistically significant differences between normal and forced breathing were found in all participants and in both sexes, but no significant correlation by age. Electro-optical distance measurement, representing a novel technical way for the ‘elastography’ of the nasal valve should be added to advanced 4-phase-rhinomanometers.</p></div>","PeriodicalId":15168,"journal":{"name":"Journal of biomechanics","volume":"176 ","pages":"Article 112326"},"PeriodicalIF":2.4,"publicationDate":"2024-09-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"142272727","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Design and validation of an In Vitro test bench for the investigation of cardiopulmonary resuscitation procedure 设计和验证用于研究心肺复苏程序的体外试验台
IF 2.4 3区 医学 Q3 BIOPHYSICS Pub Date : 2024-09-14 DOI: 10.1016/j.jbiomech.2024.112324
Andréa El-Khoury , Lawrence Leroux , Julien Dupuis Desroches , Giuseppe Di Labbio , Lyes Kadem

Despite recent clinical and technological advancements, the cardiac arrest survival rate remains as low as 10%. To enhance patient outcomes, it is crucial to deepen the understanding of cardiopulmonary resuscitation (CPR) at a fundamental level. Currently, there is a lack of knowledge on the physiological effects of CPR, in particular on the hemodynamics in the heart and the great vessels. The design and validation of a dedicated in vitro heart simulator, capable of replicating the physiological response to CPR, holds the potential to provide valuable insights into the fluid dynamics in the heart during CPR but also to be used as a platform for the development and testing of mechanical CPR machines. The main objective of this study is to design and validate the first in vitro heart simulator that can replicate the physiological response during CPR. For that, a custom-made heart simulator is designed consisting of an elastic model of the complete heart and a controllable linear actuator. The heart model is positioned in an anatomical position, and the linear actuator compresses the model at specific rates and depths. Flow and pressure waveforms are recorded on the newly developed simulator at 60 contractions per minute and results are validated against reported in vivo data in the literature. Finally, the system’s capabilities are evaluated by considering several combinations of compression rates and depths.

尽管近年来临床和技术不断进步,但心脏骤停患者的存活率仍然低至 10%。为了提高患者的治疗效果,从根本上加深对心肺复苏(CPR)的理解至关重要。目前,人们对心肺复苏的生理效应,尤其是对心脏和大血管的血流动力学缺乏了解。专用体外心脏模拟器能够复制心肺复苏术的生理反应,其设计和验证不仅有可能为心肺复苏术期间的心脏流体动力学提供有价值的见解,还可用作开发和测试机械心肺复苏机的平台。本研究的主要目的是设计和验证第一个体外心脏模拟器,该模拟器可以复制心肺复苏过程中的生理反应。为此,我们设计了一个定制的心脏模拟器,由一个完整心脏的弹性模型和一个可控线性致动器组成。将心脏模型置于解剖位置,线性推杆以特定速率和深度压缩模型。在新开发的模拟器上以每分钟 60 次收缩的速度记录流量和压力波形,并根据文献中报告的体内数据对结果进行验证。最后,通过考虑压缩率和深度的几种组合,对系统的能力进行了评估。
{"title":"Design and validation of an In Vitro test bench for the investigation of cardiopulmonary resuscitation procedure","authors":"Andréa El-Khoury ,&nbsp;Lawrence Leroux ,&nbsp;Julien Dupuis Desroches ,&nbsp;Giuseppe Di Labbio ,&nbsp;Lyes Kadem","doi":"10.1016/j.jbiomech.2024.112324","DOIUrl":"10.1016/j.jbiomech.2024.112324","url":null,"abstract":"<div><p>Despite recent clinical and technological advancements, the cardiac arrest survival rate remains as low as 10%. To enhance patient outcomes, it is crucial to deepen the understanding of cardiopulmonary resuscitation (CPR) at a fundamental level. Currently, there is a lack of knowledge on the physiological effects of CPR, in particular on the hemodynamics in the heart and the great vessels. The design and validation of a dedicated in vitro heart simulator, capable of replicating the physiological response to CPR, holds the potential to provide valuable insights into the fluid dynamics in the heart during CPR but also to be used as a platform for the development and testing of mechanical CPR machines. The main objective of this study is to design and validate the first in vitro heart simulator that can replicate the physiological response during CPR. For that, a custom-made heart simulator is designed consisting of an elastic model of the complete heart and a controllable linear actuator. The heart model is positioned in an anatomical position, and the linear actuator compresses the model at specific rates and depths. Flow and pressure waveforms are recorded on the newly developed simulator at 60 contractions per minute and results are validated against reported in vivo data in the literature. Finally, the system’s capabilities are evaluated by considering several combinations of compression rates and depths.</p></div>","PeriodicalId":15168,"journal":{"name":"Journal of biomechanics","volume":"176 ","pages":"Article 112324"},"PeriodicalIF":2.4,"publicationDate":"2024-09-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"142272728","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Exploring the relationship between trunk muscles and lower limb injuries in Australian badminton players 探究澳大利亚羽毛球运动员躯干肌肉与下肢损伤之间的关系
IF 2.4 3区 医学 Q3 BIOPHYSICS Pub Date : 2024-09-14 DOI: 10.1016/j.jbiomech.2024.112325
Julie A. Hides , M. Dilani Mendis , Felix Leung , Brittany Grantham , Brooke Coombes

Due to its dynamic nature, lower limb injuries are common in badminton. Overuse injuries of the knee, including tendon related conditions, are the most common. During jumping and landing, force transference and dissipation through the trunk is required, with the trunk muscles playing a vital role. However, the relationship between knee pain and the ability to voluntarily contract the trunk muscles has not yet been explored in badminton players. A cross-sectional study of Australian badminton players was therefore conducted. Players performed a single leg decline squat to identify those with knee pain. Ultrasound imaging was used to image and measure the size of the multifidus and quadratus lumborum, and the ability to contract the abdominal and multifidus muscles. Voluntary contraction of the trunk muscles was conducted with the subjects lying down. Independent samples T-Tests were performed to test for between group differences. Badminton players with knee pain had larger quadratus lumborum muscles and demonstrated a greater change in muscle thickness from the rested to contracted state. While we cannot comment on causation or direction, over co-contraction of trunk muscles has been shown in other studies to be associated with increased ground reaction forces on landing. Motor control training has been successfully used in other conditions to modify trunk muscle recruitment patterns and may therefore potentially represent a useful approach for badminton players.

由于羽毛球运动的动态性质,下肢受伤在羽毛球运动中很常见。最常见的是膝关节的过度运动损伤,包括与肌腱有关的损伤。在起跳和落地过程中,力量需要通过躯干传递和消散,躯干肌肉起着至关重要的作用。然而,在羽毛球运动员中,膝关节疼痛与躯干肌肉自主收缩能力之间的关系尚未得到探讨。因此,我们对澳大利亚羽毛球运动员进行了一项横断面研究。研究人员对运动员进行了单腿下蹲训练,以找出膝关节疼痛的运动员。研究人员使用超声波成像技术对多裂肌和腰方肌的大小以及腹肌和多裂肌的收缩能力进行了成像和测量。躯干肌肉的自主收缩是在受试者躺下时进行的。对组间差异进行独立样本 T 检验。膝关节疼痛的羽毛球运动员的腰四头肌较大,肌肉厚度从休息状态到收缩状态的变化也较大。虽然我们无法对其因果关系或方向发表评论,但其他研究表明,躯干肌肉的过度共收缩与着地时地面反作用力的增加有关。在其他情况下,运动控制训练已成功用于改变躯干肌肉募集模式,因此可能是羽毛球运动员的一种有用方法。
{"title":"Exploring the relationship between trunk muscles and lower limb injuries in Australian badminton players","authors":"Julie A. Hides ,&nbsp;M. Dilani Mendis ,&nbsp;Felix Leung ,&nbsp;Brittany Grantham ,&nbsp;Brooke Coombes","doi":"10.1016/j.jbiomech.2024.112325","DOIUrl":"10.1016/j.jbiomech.2024.112325","url":null,"abstract":"<div><p>Due to its dynamic nature, lower limb injuries are common in badminton. Overuse injuries of the knee, including tendon related conditions, are the most common. During jumping and landing, force transference and dissipation through the trunk is required, with the trunk muscles playing a vital role. However, the relationship between knee pain and the ability to voluntarily contract the trunk muscles has not yet been explored in badminton players. A cross-sectional study of Australian badminton players was therefore conducted. Players performed a single leg decline squat to identify those with knee pain. Ultrasound imaging was used to image and measure the size of the multifidus and quadratus lumborum, and the ability to contract the abdominal and multifidus muscles. Voluntary contraction of the trunk muscles was conducted with the subjects lying down. Independent samples T-Tests were performed to test for between group differences. Badminton players with knee pain had larger quadratus lumborum muscles and demonstrated a greater change in muscle thickness from the rested to contracted state. While we cannot comment on causation or direction, over co-contraction of trunk muscles has been shown in other studies to be associated with increased ground reaction forces on landing. Motor control training has been successfully used in other conditions to modify trunk muscle recruitment patterns and may therefore potentially represent a useful approach for badminton players.</p></div>","PeriodicalId":15168,"journal":{"name":"Journal of biomechanics","volume":"176 ","pages":"Article 112325"},"PeriodicalIF":2.4,"publicationDate":"2024-09-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S0021929024004032/pdfft?md5=8d1aa0cb7175f251075ed22a759fc63b&pid=1-s2.0-S0021929024004032-main.pdf","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"142241150","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Influence of articular cartilage sample geometry on mechanical response and properties using finite element simulation 利用有限元模拟技术分析关节软骨样本几何形状对机械响应和性能的影响
IF 2.4 3区 医学 Q3 BIOPHYSICS Pub Date : 2024-09-13 DOI: 10.1016/j.jbiomech.2024.112323
Viktor Jönsson , Gustavo A. Orozco , Maria Pierantoni , Hector Dejea , Anna Gustafsson , Lorenzo Grassi , Hanna Isaksson
Mechanical testing of articular cartilage yields highly variable results, posing challenges for tissue characterization. Many factors cause variability, one is sample geometry. Using in-situ phase-contrast enhanced synchrotron micro-tomographs of cartilage samples while tested in unconfined compression (stress relaxation) our group found high variability in the mechanical response. Since all samples originated from a single bovine knee, they were assumed to share mechanical properties. Microscale tomography images showed geometric irregularities in samples that were not accounted for in the often assumed intended cylindrical shape. We aimed to determine the influence of sample shape on mechanical response in unconfined compression and how sample geometry affects identified mechanical properties. Using a parametric FE model incorporating geometric irregularities in a Design of Experiments approach, results were analysed with 2-way ANOVA. Furthermore, a material parameter fitting was done with multiple segmented sample-specific finite element models simultaneously to assess the influence of sample geometry on material parameters. Results revealed that the average inclined sample surface (4°) caused a 15 % decrease in reaction forces compared to the intended cylinder. Fitting multiple sample-specific geometries simultaneously altered material parameters between −70 to +159 % compared to the average model. Strikingly, initial fibril stiffness and permeability increased by 137 % and 159 %, while the root-mean-square error of the fit was reduced by ∼2/3 compared to using parameters from a cylindrical shape model. In conclusion, minor variability in sample geometry affects property characterization and can account for some of the inter-sample variability in the mechanical data for cartilage.
关节软骨的力学测试结果变化很大,给组织表征带来了挑战。造成差异的因素很多,其中之一就是样品的几何形状。我们的研究小组利用原位相位对比增强同步加速器显微层析成像技术,对软骨样本进行无约束压缩(应力松弛)测试,发现机械响应的变异性很高。由于所有样本都来自同一个牛膝,因此假定它们具有相同的机械特性。微尺度断层扫描图像显示样品存在几何不规则性,而通常假定的圆柱形并没有考虑到这些不规则性。我们旨在确定样品形状对无约束压缩机械响应的影响,以及样品几何形状如何影响已识别的机械特性。我们使用一个参数化的 FE 模型,将几何形状的不规则性纳入到实验设计方法中,并通过双向方差分析对结果进行了分析。此外,还同时使用多个分段式特定样品有限元模型进行了材料参数拟合,以评估样品几何形状对材料参数的影响。结果表明,与预定圆柱体相比,平均倾斜样品表面(4°)造成的反作用力降低了 15%。与平均模型相比,同时拟合多个特定样品几何形状会使材料参数发生-70%到+159%的变化。引人注目的是,初始纤维刚度和渗透性分别增加了 137% 和 159%,与使用圆柱形模型参数相比,拟合的均方根误差减少了 ∼2/3。总之,样品几何形状的微小变化会影响性能表征,并能解释软骨力学数据中样品间差异的部分原因。
{"title":"Influence of articular cartilage sample geometry on mechanical response and properties using finite element simulation","authors":"Viktor Jönsson ,&nbsp;Gustavo A. Orozco ,&nbsp;Maria Pierantoni ,&nbsp;Hector Dejea ,&nbsp;Anna Gustafsson ,&nbsp;Lorenzo Grassi ,&nbsp;Hanna Isaksson","doi":"10.1016/j.jbiomech.2024.112323","DOIUrl":"10.1016/j.jbiomech.2024.112323","url":null,"abstract":"<div><div>Mechanical testing of articular cartilage yields highly variable results, posing challenges for tissue characterization. Many factors cause variability, one is sample geometry. Using in-situ phase-contrast enhanced synchrotron micro-tomographs of cartilage samples while tested in unconfined compression (stress relaxation) our group found high variability in the mechanical response. Since all samples originated from a single bovine knee, they were assumed to share mechanical properties. Microscale tomography images showed geometric irregularities in samples that were not accounted for in the often assumed intended cylindrical shape. We aimed to determine the influence of sample shape on mechanical response in unconfined compression and how sample geometry affects identified mechanical properties. Using a parametric FE model incorporating geometric irregularities in a Design of Experiments approach, results were analysed with 2-way ANOVA. Furthermore, a material parameter fitting was done with multiple segmented sample-specific finite element models simultaneously to assess the influence of sample geometry on material parameters. Results revealed that the average inclined sample surface (4°) caused a 15 % decrease in reaction forces compared to the intended cylinder. Fitting multiple sample-specific geometries simultaneously altered material parameters between −70 to +159 % compared to the average model. Strikingly, initial fibril stiffness and permeability increased by 137 % and 159 %, while the root-mean-square error of the fit was reduced by ∼2/3 compared to using parameters from a cylindrical shape model. In conclusion, minor variability in sample geometry affects property characterization and can account for some of the inter-sample variability in the mechanical data for cartilage.</div></div>","PeriodicalId":15168,"journal":{"name":"Journal of biomechanics","volume":"176 ","pages":"Article 112323"},"PeriodicalIF":2.4,"publicationDate":"2024-09-13","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"142322720","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Optimizing distal and proximal splenic artery embolization with patient-specific computational fluid dynamics 利用患者特异性计算流体动力学优化远端和近端脾动脉栓塞术
IF 2.4 3区 医学 Q3 BIOPHYSICS Pub Date : 2024-09-12 DOI: 10.1016/j.jbiomech.2024.112320
Younes Tatari , Tyler Andrew Smith , Jingjie Hu , Amirhossein Arzani

Splenic artery embolization (SAE) has become a favored alternative to splenectomy, offering a less invasive intervention for injured spleens while preserving spleen function. However, our understanding of the role that hemodynamics plays during embolization remains limited. In this study, we utilized patient-specific computational fluid dynamics (CFD) simulations to study distal and proximal embolization strategies commonly used in SAE. Detailed 3D computer models were constructed considering the descending aorta, various major visceral arteries, and the iliac arteries. Subsequently, the blood flow and pressure associated with different coil placement locations in proximal embolization were studied considering the collateral vessels. Coil induced variations in pressure fields were quantified and compared to baseline. The coil induced flow stagnation was also quantified with particle residence time. Distal embolization was modeled with Lagrangian particle tracking and the effect of particle size, release location, and timing on embolization outcome was studied. Our findings highlight the crucial role of collateral vessels in maintaining blood supply to the spleen following proximal embolization. It was demonstrated that coil location can affect distal pressure and that strategic coil placement guided by patient-specific CFD simulations can further reduce this pressure as desired. Additionally, the results point to the critical roles that particle size, release timing, and location play in distal embolization. Our study provides an early attempt to use patient-specific computer modeling for optimizing embolization strategies and ultimately improving patient outcomes during SAE procedures.

脾动脉栓塞术(SAE)已成为脾切除术的首选替代方案,在保留脾脏功能的同时,为损伤的脾脏提供了一种创伤较小的干预方法。然而,我们对栓塞过程中血液动力学所起作用的了解仍然有限。在本研究中,我们利用患者特异性计算流体动力学(CFD)模拟来研究 SAE 中常用的远端和近端栓塞策略。考虑到降主动脉、各种主要内脏动脉和髂动脉,我们构建了详细的三维计算机模型。随后,考虑到侧支血管,研究了近端栓塞中不同线圈放置位置相关的血流和压力。对线圈引起的压力场变化进行了量化,并与基线进行了比较。线圈引起的血流停滞也通过粒子停留时间进行了量化。利用拉格朗日粒子追踪技术建立了远端栓塞模型,并研究了粒子大小、释放位置和时间对栓塞结果的影响。我们的研究结果强调了侧支血管在近端栓塞后维持脾脏血液供应的关键作用。研究结果表明,线圈位置会影响远端压力,而在特定患者的 CFD 模拟指导下进行战略性线圈放置可进一步降低压力。此外,研究结果还指出了颗粒大小、释放时间和位置在远端栓塞中的关键作用。我们的研究为使用患者特异性计算机建模来优化栓塞策略并最终改善 SAE 手术中的患者预后提供了早期尝试。
{"title":"Optimizing distal and proximal splenic artery embolization with patient-specific computational fluid dynamics","authors":"Younes Tatari ,&nbsp;Tyler Andrew Smith ,&nbsp;Jingjie Hu ,&nbsp;Amirhossein Arzani","doi":"10.1016/j.jbiomech.2024.112320","DOIUrl":"10.1016/j.jbiomech.2024.112320","url":null,"abstract":"<div><p>Splenic artery embolization (SAE) has become a favored alternative to splenectomy, offering a less invasive intervention for injured spleens while preserving spleen function. However, our understanding of the role that hemodynamics plays during embolization remains limited. In this study, we utilized patient-specific computational fluid dynamics (CFD) simulations to study distal and proximal embolization strategies commonly used in SAE. Detailed 3D computer models were constructed considering the descending aorta, various major visceral arteries, and the iliac arteries. Subsequently, the blood flow and pressure associated with different coil placement locations in proximal embolization were studied considering the collateral vessels. Coil induced variations in pressure fields were quantified and compared to baseline. The coil induced flow stagnation was also quantified with particle residence time. Distal embolization was modeled with Lagrangian particle tracking and the effect of particle size, release location, and timing on embolization outcome was studied. Our findings highlight the crucial role of collateral vessels in maintaining blood supply to the spleen following proximal embolization. It was demonstrated that coil location can affect distal pressure and that strategic coil placement guided by patient-specific CFD simulations can further reduce this pressure as desired. Additionally, the results point to the critical roles that particle size, release timing, and location play in distal embolization. Our study provides an early attempt to use patient-specific computer modeling for optimizing embolization strategies and ultimately improving patient outcomes during SAE procedures.</p></div>","PeriodicalId":15168,"journal":{"name":"Journal of biomechanics","volume":"176 ","pages":"Article 112320"},"PeriodicalIF":2.4,"publicationDate":"2024-09-12","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"142229566","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Musculoskeletal models determine the effect of a soft active exosuit on muscle activations and forces during lifting and lowering tasks 肌肉骨骼模型确定软质活动外衣对升降任务中肌肉激活和力量的影响
IF 2.4 3区 医学 Q3 BIOPHYSICS Pub Date : 2024-09-11 DOI: 10.1016/j.jbiomech.2024.112322
Chenxi Yan , Jacob J. Banks , Brett T. Allaire , D. Adam Quirk , Jinwon Chung , Conor J. Walsh , Dennis E. Anderson

Exosuits have the potential to mitigate musculoskeletal stress and prevent back injuries during industrial tasks. This study aimed to 1) validate the implementation of a soft active exosuit into a musculoskeletal model of the spine by comparing model predicted muscle activations versus corresponding surface EMG measurements, and 2) evaluate the effect of the exosuit on peak back and hip muscle forces. Fourteen healthy participants performed squat and stoop lift and lower tasks with boxes of 6 and 10 kg, with and without wearing a 2.7 kg soft active exosuit. Participant-specific musculoskeletal models, which included the exosuit, were created in OpenSim. Model validation focused on the back and hip extensors, where temporal agreement between EMG and model estimated muscle activity was generally strong to excellent (average cross-correlation coefficients ranging from 0.84 to 0.98). Root mean square errors of muscle activity (0.05–0.10) were similar with and without the exosuit, and compared well to prior model validation studies without the exosuit (average root mean square errors ranging from 0.05 to 0.19). In terms of performance, the exosuit reduced the estimated peak erector spinae forces during lifting and lowering phases across all lifting tasks but reduced peak hip extensor muscles forces only in a squat lift task of 10 kg. These reductions in total peak muscle forces were approximately 1.7–4.2 times greater than the corresponding exosuit assistance force, which were 146 ± 19 N and 102 ± 14 N at the times of peak erector spinae forces in lifting and lowering, respectively. Overall, the results support the hypothesis that exosuits reduce soft tissue loading, and thereby potentially reduce fatigue and injury risk during manual materials handling tasks. Incorporating exosuits into musculoskeletal models is a valid approach to understand the impact of exosuit assistance on muscle activity and forces.

在工业任务中,防弹衣有可能减轻肌肉骨骼压力并防止背部受伤。本研究的目的是:1)通过比较模型预测的肌肉激活与相应的表面肌电图测量值,验证在脊柱肌肉骨骼模型中实施软质主动式防弹衣的有效性;2)评估防弹衣对背部和臀部肌肉峰值力的影响。14 名健康的参与者分别穿着和未穿着 2.7 千克的软质主动防护服,用 6 千克和 10 千克的箱子完成了下蹲和弯腰抬起和放下任务。在 OpenSim 中创建了包括外穿衣在内的参与者特定肌肉骨骼模型。模型验证主要集中在背部和髋部伸肌,在这两个部位,EMG 和模型估计的肌肉活动之间的时间一致性一般都很强甚至很好(平均交叉相关系数在 0.84 到 0.98 之间)。肌肉活动的均方根误差(0.05-0.10)在穿和不穿外穿衣的情况下相似,与之前不穿外穿衣的模型验证研究(平均均方根误差在 0.05 至 0.19 之间)相比也很好。就性能而言,在所有举重任务中,外固定衣降低了举起和放下阶段的估计峰值竖脊肌力,但只在深蹲举起 10 公斤的任务中降低了峰值髋伸肌力。总峰值肌肉力的减少量大约是相应的外穿衣辅助力的 1.7-4.2 倍,在举起和放下的竖脊肌峰值力分别为 146 ± 19 N 和 102 ± 14 N。总之,研究结果支持了这样的假设,即外穿衣可以减轻软组织负荷,从而降低人工材料搬运任务中的疲劳和受伤风险。在肌肉骨骼模型中加入外防护服是了解外防护服辅助对肌肉活动和力量影响的有效方法。
{"title":"Musculoskeletal models determine the effect of a soft active exosuit on muscle activations and forces during lifting and lowering tasks","authors":"Chenxi Yan ,&nbsp;Jacob J. Banks ,&nbsp;Brett T. Allaire ,&nbsp;D. Adam Quirk ,&nbsp;Jinwon Chung ,&nbsp;Conor J. Walsh ,&nbsp;Dennis E. Anderson","doi":"10.1016/j.jbiomech.2024.112322","DOIUrl":"10.1016/j.jbiomech.2024.112322","url":null,"abstract":"<div><p>Exosuits have the potential to mitigate musculoskeletal stress and prevent back injuries during industrial tasks. This study aimed to 1) validate the implementation of a soft active exosuit into a musculoskeletal model of the spine by comparing model predicted muscle activations versus corresponding surface EMG measurements, and 2) evaluate the effect of the exosuit on peak back and hip muscle forces. Fourteen healthy participants performed squat and stoop lift and lower tasks with boxes of 6 and 10 kg, with and without wearing a 2.7 kg soft active exosuit. Participant-specific musculoskeletal models, which included the exosuit, were created in OpenSim. Model validation focused on the back and hip extensors, where temporal agreement between EMG and model estimated muscle activity was generally strong to excellent (average cross-correlation coefficients ranging from 0.84 to 0.98). Root mean square errors of muscle activity (0.05–0.10) were similar with and without the exosuit, and compared well to prior model validation studies without the exosuit (average root mean square errors ranging from 0.05 to 0.19). In terms of performance, the exosuit reduced the estimated peak erector spinae forces during lifting and lowering phases across all lifting tasks but reduced peak hip extensor muscles forces only in a squat lift task of 10 kg. These reductions in total peak muscle forces were approximately 1.7–4.2 times greater than the corresponding exosuit assistance force, which were 146 ± 19 N and 102 ± 14 N at the times of peak erector spinae forces in lifting and lowering, respectively. Overall, the results support the hypothesis that exosuits reduce soft tissue loading, and thereby potentially reduce fatigue and injury risk during manual materials handling tasks. Incorporating exosuits into musculoskeletal models is a valid approach to understand the impact of exosuit assistance on muscle activity and forces.</p></div>","PeriodicalId":15168,"journal":{"name":"Journal of biomechanics","volume":"176 ","pages":"Article 112322"},"PeriodicalIF":2.4,"publicationDate":"2024-09-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"142272726","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
期刊
Journal of biomechanics
全部 Acc. Chem. Res. ACS Applied Bio Materials ACS Appl. Electron. Mater. ACS Appl. Energy Mater. ACS Appl. Mater. Interfaces ACS Appl. Nano Mater. ACS Appl. Polym. Mater. ACS BIOMATER-SCI ENG ACS Catal. ACS Cent. Sci. ACS Chem. Biol. ACS Chemical Health & Safety ACS Chem. Neurosci. ACS Comb. Sci. ACS Earth Space Chem. ACS Energy Lett. ACS Infect. Dis. ACS Macro Lett. ACS Mater. Lett. ACS Med. Chem. Lett. ACS Nano ACS Omega ACS Photonics ACS Sens. ACS Sustainable Chem. Eng. ACS Synth. Biol. Anal. Chem. BIOCHEMISTRY-US Bioconjugate Chem. BIOMACROMOLECULES Chem. Res. Toxicol. Chem. Rev. Chem. Mater. CRYST GROWTH DES ENERG FUEL Environ. Sci. Technol. Environ. Sci. Technol. Lett. Eur. J. Inorg. Chem. IND ENG CHEM RES Inorg. Chem. J. Agric. Food. Chem. J. Chem. Eng. Data J. Chem. Educ. J. Chem. Inf. Model. J. Chem. Theory Comput. J. Med. Chem. J. Nat. Prod. J PROTEOME RES J. Am. Chem. Soc. LANGMUIR MACROMOLECULES Mol. Pharmaceutics Nano Lett. Org. Lett. ORG PROCESS RES DEV ORGANOMETALLICS J. Org. Chem. J. Phys. Chem. J. Phys. Chem. A J. Phys. Chem. B J. Phys. Chem. C J. Phys. Chem. Lett. Analyst Anal. Methods Biomater. Sci. Catal. Sci. Technol. Chem. Commun. Chem. Soc. Rev. CHEM EDUC RES PRACT CRYSTENGCOMM Dalton Trans. Energy Environ. Sci. ENVIRON SCI-NANO ENVIRON SCI-PROC IMP ENVIRON SCI-WAT RES Faraday Discuss. Food Funct. Green Chem. Inorg. Chem. Front. Integr. Biol. J. Anal. At. Spectrom. J. Mater. Chem. A J. Mater. Chem. B J. Mater. Chem. C Lab Chip Mater. Chem. Front. Mater. Horiz. MEDCHEMCOMM Metallomics Mol. Biosyst. Mol. Syst. Des. Eng. Nanoscale Nanoscale Horiz. Nat. Prod. Rep. New J. Chem. Org. Biomol. Chem. Org. Chem. Front. PHOTOCH PHOTOBIO SCI PCCP Polym. Chem.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1