首页 > 最新文献

中华放射医学与防护杂志最新文献

英文 中文
Eeffects of tube voltage on image quality and radiation dose of paranasal sinus CT examination 管电压对副鼻窦CT检查图像质量及辐射剂量的影响
Q4 Medicine Pub Date : 2020-03-25 DOI: 10.3760/CMA.J.ISSN.0254-5098.2020.03.013
Jianxing Wu, Y. Niu, Yongxian Zhang, Lili Zhang
Objective To explore the effects of tube voltage on the image quality and radiation dose of paranasal sinus CT on a 16 cm wide-detector CT. Methods The sinus area of head specimen was scanned with sequential scanning mode and tube voltages of 70, 80, 100, 120 and 140 kV, respectively. The parameters were set as follows: NI=9, automatic tube current modulation, Smart mA with the maximum range at the corresponding tube voltage, 120 mm detector width. The images were reconstructed with bone and soft tissue algorithms separately.The slice thickness was 0.625 mm for all reconstructed images. The axial, coronal and sagittal images were reformatted as 2 mm thickness with clinical baseline. The CT value and standard deviation (SD) were measured in the region of interest(ROI) of central axial image and the contrast-to-noise ratio (CNR) was calculated. The volume CT dose index(CTDIvol) and dose length product (DLP) were recorded, and the figure of merit (FOM) of images was computed. The images obtained in the study were subjectively evaluated by two experienced radiologists and one technician using a five-point scoring system. Results The subjective evaluation of images obtained in the study met the diagnostic requirements. With the reconstruction of bone algorithm, the CNR at the tube voltages of 100 and 80 kV were 66.98 and 64.75, respectively, which was significantly higher than those at the tube voltages of 70, 120 and 140 kV (51.61, 61.56 and 57.76, respectively). The maximum and minimum CTDIvol were 34.11 mGy at 140 kV and 17.45 mGy at 70 kV. The FOM value was 152.26 at 100 kV. With the reconstruction of soft tissue algorithm, the maximum CNR was 195.62 at 80 kV, which was markedly greater than those at 70, 100, 120 kV and 140 kV (139.46, 154.49, 148.06 and 155.58 respectively). The maximum FOM value was 1273.56 at 80 kV, which was significantly higher than those at the tube voltages of 70, 100, 120 kV and 140 kV (1114.56, 809.98, 735.63 and 709.62, respectively). Conclusions For the sinus CT of head specimen, the scan protocols with 100 kV combined with bone reconstruction algorithm and 80 kV combined with soft tissue reconstruction algorithm can provide relatively lower radiation dose and good image quality. Key words: CT; Paranasal sinus; Tube voltage; Radiation dose
目的探讨在16 cm宽探测CT上,管电压对鼻窦CT图像质量和辐射剂量的影响。方法采用顺序扫描方式,分别在70、80、100、120、140 kV的管电压下对头部标本窦区进行扫描。参数设置如下:NI=9,自动管电流调制,Smart mA在相应管电压下具有最大量程,检测器宽度120mm。分别用骨和软组织算法重建图像。重建图像的切片厚度均为0.625 mm。轴位、冠状位和矢状位图像重新格式化为2mm厚度,符合临床基线。测量中心轴向图像感兴趣区域(ROI)的CT值和标准差(SD),计算对比噪声比(CNR)。记录体积CT剂量指数(CTDIvol)和剂量长度积(DLP),计算图像的优值(FOM)。研究中获得的图像由两名经验丰富的放射科医生和一名技术人员使用五分制进行主观评估。结果本研究获得的影像主观评价符合诊断要求。通过骨算法重建,100和80 kV管电压下的CNR分别为66.98和64.75,显著高于70、120和140 kV管电压下的CNR(分别为51.61、61.56和57.76)。140 kV时CTDIvol最大值为34.11 mGy, 70 kV时最大值为17.45 mGy。在100 kV时,FOM值为152.26。采用软组织重构算法,80 kV时最大CNR为195.62,显著大于70、100、120、140 kV时的最大CNR(分别为139.46、154.49、148.06、155.58)。80 kV时FOM值最大值为1273.56,显著高于70、100、120、140 kV时的FOM值(分别为1114.56、809.98、735.63、709.62)。结论对于头部标本窦部CT, 100 kV联合骨重建算法和80 kV联合软组织重建算法的扫描方案可以提供相对较低的辐射剂量和较好的图像质量。关键词:CT;副鼻窦;管电压;辐射剂量
{"title":"Eeffects of tube voltage on image quality and radiation dose of paranasal sinus CT examination","authors":"Jianxing Wu, Y. Niu, Yongxian Zhang, Lili Zhang","doi":"10.3760/CMA.J.ISSN.0254-5098.2020.03.013","DOIUrl":"https://doi.org/10.3760/CMA.J.ISSN.0254-5098.2020.03.013","url":null,"abstract":"Objective \u0000To explore the effects of tube voltage on the image quality and radiation dose of paranasal sinus CT on a 16 cm wide-detector CT. \u0000 \u0000 \u0000Methods \u0000The sinus area of head specimen was scanned with sequential scanning mode and tube voltages of 70, 80, 100, 120 and 140 kV, respectively. The parameters were set as follows: NI=9, automatic tube current modulation, Smart mA with the maximum range at the corresponding tube voltage, 120 mm detector width. The images were reconstructed with bone and soft tissue algorithms separately.The slice thickness was 0.625 mm for all reconstructed images. The axial, coronal and sagittal images were reformatted as 2 mm thickness with clinical baseline. The CT value and standard deviation (SD) were measured in the region of interest(ROI) of central axial image and the contrast-to-noise ratio (CNR) was calculated. The volume CT dose index(CTDIvol) and dose length product (DLP) were recorded, and the figure of merit (FOM) of images was computed. The images obtained in the study were subjectively evaluated by two experienced radiologists and one technician using a five-point scoring system. \u0000 \u0000 \u0000Results \u0000The subjective evaluation of images obtained in the study met the diagnostic requirements. With the reconstruction of bone algorithm, the CNR at the tube voltages of 100 and 80 kV were 66.98 and 64.75, respectively, which was significantly higher than those at the tube voltages of 70, 120 and 140 kV (51.61, 61.56 and 57.76, respectively). The maximum and minimum CTDIvol were 34.11 mGy at 140 kV and 17.45 mGy at 70 kV. The FOM value was 152.26 at 100 kV. With the reconstruction of soft tissue algorithm, the maximum CNR was 195.62 at 80 kV, which was markedly greater than those at 70, 100, 120 kV and 140 kV (139.46, 154.49, 148.06 and 155.58 respectively). The maximum FOM value was 1273.56 at 80 kV, which was significantly higher than those at the tube voltages of 70, 100, 120 kV and 140 kV (1114.56, 809.98, 735.63 and 709.62, respectively). \u0000 \u0000 \u0000Conclusions \u0000For the sinus CT of head specimen, the scan protocols with 100 kV combined with bone reconstruction algorithm and 80 kV combined with soft tissue reconstruction algorithm can provide relatively lower radiation dose and good image quality. \u0000 \u0000 \u0000Key words: \u0000CT; Paranasal sinus; Tube voltage; Radiation dose","PeriodicalId":36403,"journal":{"name":"中华放射医学与防护杂志","volume":"403 ","pages":"233-236"},"PeriodicalIF":0.0,"publicationDate":"2020-03-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"72423477","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
The concept of stereotactic ablation brachytherapy and practice 立体定向消融近距离治疗的概念与实践
Q4 Medicine Pub Date : 2020-03-25 DOI: 10.3760/CMA.J.ISSN.0254-5098.2020.03.003
Junjie Wang
The development of tumor brachytherapy and ablation therapy leads to the concept of stereotactic ablation brachytherapy(SABT). It is categorized into H-SABT (high dose rate brachytherapy) and L-SABT (low dose rate brachy therapy). L-SABT is clinically used as radioactive seed implantation therapy. SABT is characterized with the following: hypofraction and short course of treatment, Image-guidance and precision ablation, exact curative effect and preserved organic functions. SABT has gradually become a new disciplinary system and will play an important role in cancer treatments. Key words: Stereotactic; Brachytherapy; Ablation; Interstitial permanent implantation
肿瘤近距离治疗和消融治疗的发展导致了立体定向消融近距离治疗(SABT)的概念。它分为H-SABT(高剂量率近距离治疗)和L-SABT(低剂量率近距离治疗)。L-SABT临床应用于放射性粒子植入治疗。SABT具有体积小、疗程短、影像引导、消融精准、疗效确切、保留器官功能等特点。SABT已逐渐成为一门新的学科体系,并将在肿瘤治疗中发挥重要作用。关键词:立体定向;近距离放射疗法;烧蚀;间质永久性植入
{"title":"The concept of stereotactic ablation brachytherapy and practice","authors":"Junjie Wang","doi":"10.3760/CMA.J.ISSN.0254-5098.2020.03.003","DOIUrl":"https://doi.org/10.3760/CMA.J.ISSN.0254-5098.2020.03.003","url":null,"abstract":"The development of tumor brachytherapy and ablation therapy leads to the concept of stereotactic ablation brachytherapy(SABT). It is categorized into H-SABT (high dose rate brachytherapy) and L-SABT (low dose rate brachy therapy). L-SABT is clinically used as radioactive seed implantation therapy. SABT is characterized with the following: hypofraction and short course of treatment, Image-guidance and precision ablation, exact curative effect and preserved organic functions. SABT has gradually become a new disciplinary system and will play an important role in cancer treatments. \u0000 \u0000 \u0000Key words: \u0000Stereotactic; Brachytherapy; Ablation; Interstitial permanent implantation","PeriodicalId":36403,"journal":{"name":"中华放射医学与防护杂志","volume":"18 1","pages":"173-177"},"PeriodicalIF":0.0,"publicationDate":"2020-03-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"90232927","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 2
Investigation and analysis of CT doses to examined individuals in Henan province 河南省被检查人群CT剂量调查与分析
Q4 Medicine Pub Date : 2020-03-25 DOI: 10.3760/CMA.J.ISSN.0254-5098.2020.03.012
Yanfang Zhao, Caifang Chu, Qinfu Zhang, Xiaojun Cheng, Yongxing Li, Xiaofang Luo, Bingchen Dong
Objective To study the level of doses which children and adults received from CT examinations in Henan province, and to explore the influencing factors, so as to provide basic data for the development of diagnostic reference (guidance) level for CT examination in China and related regions. Methods A total of 1 113 subjects, which have received examinations by 31 CT scanners separately in 7 cities of the province, were selected using stratified quota sampling method to carry out the dose survey, including basic information on subjects, equipment, scanning parameters, and dosimetry, etc. Typical effective dose was estimated by using dose conversion coefficient. Results In the conventional CT scanning of head. The 75% of CTDIw for children aged 0-1, >1-5, >5-10 and >10-15 years were 32.2, 37.2, 43.0 and 46.7 mGy; 75% of DLP were 478.0, 572.0, 715.6 and 743.9 mGy·cm and effective doses were 5.26, 3.83, 2.86 and 2.38 mSv, respectively. In the routine chest scanning, 75% of CTDIw for children aged 0-1, >1-5, >5-10 and >10-15 years were all 9.3 mGy. The 75% of DLP were 141.7, 178.8, 224.0 and 238.7 mGy·cm; and effective doses were 5.53, 4.64 4.03 and 3.10 mSv, respectively. 75% of CTDIw for head, chest and abdomen of the adult subjects from CT examinations were 57.4, 16.2 and 19.4 mGy, respectively. The 75% of DLP were 818.3, 504.7 and 571.1 mGy·cm, respectively, and effective doses were 1.72, 7.07 and 8.57 mSv, respectively. Conclusions As the update of CT equipment, increasing progress in examination technology and rapid increase in CT application frequency, medical institutions should choose the scanning parameters with respect to various types of CT scanner to optimize the image quality and the radiation doses. The diagnostic reference level for CT examination should be improved, especially for children. Key words: CT; CT dose index; Dose length product; Effective dose; Diagnostic reference level
目的了解河南省儿童和成人CT检查的剂量水平,并探讨影响因素,为制定中国及相关地区CT检查诊断参考(指导)水平提供基础数据。方法采用分层定额抽样的方法,在全省7个城市共选取1 113名接受过31台CT扫描仪单独检查的受试者进行剂量调查,调查内容包括受试者基本情况、设备、扫描参数、剂量学等。采用剂量转换系数估算典型有效剂量。结果头颅常规CT扫描。0-1岁、>1-5岁、>5-10岁和>10-15岁儿童CTDIw的75%分别为32.2、37.2、43.0和46.7 mGy;75%的DLP分别为478.0、572.0、715.6和743.9 mGy·cm,有效剂量分别为5.26、3.83、2.86和2.38 mSv。在常规胸部扫描中,0-1岁、>1-5岁、>5-10岁、>10-15岁儿童CTDIw的75%均为9.3 mGy。75%的DLP分别为141.7、178.8、224.0和238.7 mGy·cm;有效剂量分别为5.53、4.64、4.03和3.10毫西弗。成人头部、胸部和腹部CT检查CTDIw的75%分别为57.4、16.2和19.4 mGy。75%的DLP分别为818.3、504.7和571.1 mGy·cm,有效剂量分别为1.72、7.07和8.57 mSv。结论随着CT设备的更新、检查技术的不断进步和CT应用频率的快速增加,医疗机构应针对不同类型的CT扫描仪选择扫描参数,以优化图像质量和辐射剂量。应提高CT检查的诊断参考水平,尤其是儿童。关键词:CT;CT剂量指数;剂量长度积;有效剂量;诊断参考水平
{"title":"Investigation and analysis of CT doses to examined individuals in Henan province","authors":"Yanfang Zhao, Caifang Chu, Qinfu Zhang, Xiaojun Cheng, Yongxing Li, Xiaofang Luo, Bingchen Dong","doi":"10.3760/CMA.J.ISSN.0254-5098.2020.03.012","DOIUrl":"https://doi.org/10.3760/CMA.J.ISSN.0254-5098.2020.03.012","url":null,"abstract":"Objective \u0000To study the level of doses which children and adults received from CT examinations in Henan province, and to explore the influencing factors, so as to provide basic data for the development of diagnostic reference (guidance) level for CT examination in China and related regions. \u0000 \u0000 \u0000Methods \u0000A total of 1 113 subjects, which have received examinations by 31 CT scanners separately in 7 cities of the province, were selected using stratified quota sampling method to carry out the dose survey, including basic information on subjects, equipment, scanning parameters, and dosimetry, etc. Typical effective dose was estimated by using dose conversion coefficient. \u0000 \u0000 \u0000Results \u0000In the conventional CT scanning of head. The 75% of CTDIw for children aged 0-1, >1-5, >5-10 and >10-15 years were 32.2, 37.2, 43.0 and 46.7 mGy; 75% of DLP were 478.0, 572.0, 715.6 and 743.9 mGy·cm and effective doses were 5.26, 3.83, 2.86 and 2.38 mSv, respectively. In the routine chest scanning, 75% of CTDIw for children aged 0-1, >1-5, >5-10 and >10-15 years were all 9.3 mGy. The 75% of DLP were 141.7, 178.8, 224.0 and 238.7 mGy·cm; and effective doses were 5.53, 4.64 4.03 and 3.10 mSv, respectively. 75% of CTDIw for head, chest and abdomen of the adult subjects from CT examinations were 57.4, 16.2 and 19.4 mGy, respectively. The 75% of DLP were 818.3, 504.7 and 571.1 mGy·cm, respectively, and effective doses were 1.72, 7.07 and 8.57 mSv, respectively. \u0000 \u0000 \u0000Conclusions \u0000As the update of CT equipment, increasing progress in examination technology and rapid increase in CT application frequency, medical institutions should choose the scanning parameters with respect to various types of CT scanner to optimize the image quality and the radiation doses. The diagnostic reference level for CT examination should be improved, especially for children. \u0000 \u0000 \u0000Key words: \u0000CT; CT dose index; Dose length product; Effective dose; Diagnostic reference level","PeriodicalId":36403,"journal":{"name":"中华放射医学与防护杂志","volume":"12 1","pages":"227-232"},"PeriodicalIF":0.0,"publicationDate":"2020-03-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"86028266","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Application of a new method for patient positioning based on optical surface monitoring technology to head radiotherapy 基于光学表面监测技术的患者定位新方法在头部放疗中的应用
Q4 Medicine Pub Date : 2020-03-25 DOI: 10.3760/CMA.J.ISSN.0254-5098.2020.03.009
Yixiao Du, H. Yue, Meijiao Wang, Shun Zhou, Songmao Yu
Objective To develop a new method to set up patients using optical surface monitoring system and to compare it with the conventional method in head radiotherapy. Methods A total of 358 set-ups (130 with the conventional method and 228 with the new method ), which were from 99 head tumor patients in Beijing Cancer Hospital treated between May 2018 to April 2019, obtained by using Image Guided Radiotherapy were retrospectively analyzed. The distributions of set-up errors, the number of abnormal positions, and the set-up time were compared to evaluate the potential advantages of the new method . Results The mean (± standard deviation) absolute values of setup errors of the new method were (0.07±0.07) , (0.08±0.06) and (0.06±0.06) cm for the vertical, lateral, and longitudinal, (0.53±0.41)°, (0.59±0.44)° and (0.59±0.46)° for the rotation, pitch and roll, respectively. In the new method , the setup accuracy was improved(t=3.24-6.10, P 0.05). Conclusions The new high-precision method to set up patients using optical surface monitoring system improves the accuracy of patients′ position, decreases the corrections applied by 6DoF couch, reduces the probability of abnormal positions, and suggests the potential benefit in head radiotherapy. Key words: Set-up errors; Image guided radiotherapy; Six degrees of freedom (6DoF) couch; Optical surface monitoring technology
目的建立一种利用光学表面监测系统设置患者的新方法,并将其与传统的头部放疗方法进行比较。方法回顾性分析2018年5月至2019年4月北京肿瘤医院影像引导放疗治疗的99例头部肿瘤患者的358例(常规方法130例,新方法228例)。比较了设置误差分布、异常位置数和设置时间,评价了新方法的潜在优势。结果新方法的设置误差均值(±标准差)绝对值分别为(0.07±0.07)、(0.08±0.06)、(0.06±0.06)cm、(0.53±0.41)°、(0.59±0.44)°、(0.59±0.46)°,分别为纵向、横向和纵向设置误差均值(±标准差)为(0.07±0.07)、(0.08±0.06)cm。新方法提高了设置精度(t=3.24 ~ 6.10, P 0.05)。结论利用光学体表监测系统对患者进行高精度立位,提高了患者的体位精度,减少了6DoF躺椅的校正,降低了体位异常的概率,在头部放疗中具有潜在的应用价值。关键词:设置误差;影像引导放疗;六自由度(6DoF)沙发;光学表面监测技术
{"title":"Application of a new method for patient positioning based on optical surface monitoring technology to head radiotherapy","authors":"Yixiao Du, H. Yue, Meijiao Wang, Shun Zhou, Songmao Yu","doi":"10.3760/CMA.J.ISSN.0254-5098.2020.03.009","DOIUrl":"https://doi.org/10.3760/CMA.J.ISSN.0254-5098.2020.03.009","url":null,"abstract":"Objective \u0000To develop a new method to set up patients using optical surface monitoring system and to compare it with the conventional method in head radiotherapy. \u0000 \u0000 \u0000Methods \u0000A total of 358 set-ups (130 with the conventional method and 228 with the new method ), which were from 99 head tumor patients in Beijing Cancer Hospital treated between May 2018 to April 2019, obtained by using Image Guided Radiotherapy were retrospectively analyzed. The distributions of set-up errors, the number of abnormal positions, and the set-up time were compared to evaluate the potential advantages of the new method . \u0000 \u0000 \u0000Results \u0000The mean (± standard deviation) absolute values of setup errors of the new method were (0.07±0.07) , (0.08±0.06) and (0.06±0.06) cm for the vertical, lateral, and longitudinal, (0.53±0.41)°, (0.59±0.44)° and (0.59±0.46)° for the rotation, pitch and roll, respectively. In the new method , the setup accuracy was improved(t=3.24-6.10, P 0.05). \u0000 \u0000 \u0000Conclusions \u0000The new high-precision method to set up patients using optical surface monitoring system improves the accuracy of patients′ position, decreases the corrections applied by 6DoF couch, reduces the probability of abnormal positions, and suggests the potential benefit in head radiotherapy. \u0000 \u0000 \u0000Key words: \u0000Set-up errors; Image guided radiotherapy; Six degrees of freedom (6DoF) couch; Optical surface monitoring technology","PeriodicalId":36403,"journal":{"name":"中华放射医学与防护杂志","volume":"70 1","pages":"209-215"},"PeriodicalIF":0.0,"publicationDate":"2020-03-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"83491901","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Therapy effect of cimetidine on spleen of mice irradiated with thoracic fractionated irradiation 西咪替丁对胸段放射小鼠脾的治疗作用
Q4 Medicine Pub Date : 2020-03-25 DOI: 10.3760/CMA.J.ISSN.0254-5098.2020.03.004
Yang Bai, Junling Zhang, Dan Yao, Jianghong Zhang
Objective To investigate the protective effect of cimetidine on mice with fractionated X-ray irradiation through right thorax. Methods Thirty male C57BL/6 mice were randomly divided into control group, irradiation group and irradiation + cimetidine group (100 mg/kg), 10 mice in each group. Mice right thorax was locally irradiated with X-rays at a dose rate of 0.883 Gy/min to a cumulative dose of 24 Gy, 8 Gy per day in 3 d. After irradiation, the irradiation + cimetidine group was given cimetidine (100 mg·kg-1·d-1) for 7 d consecutively, and Ctrl and IR groups were given saline. Seven days after irradiation, the mice weight, spleen index and its T lymphocytes, B lymphocytes and NK cells were measured. Spleen tissue was stained with HE for pathological observation. Results The fractionated irradiation lightly suppressed mice growth but obviously induced abscopal damage on spleen so that the spleen index were decreased. The irradiation group (0.369±0.011) had statistically significant difference compared with the control group(0.396±0.022)(t=2.978, P<0.05), the spleen pathological structure became disorder, and the percentages of CD3+ , CD4+ , CD3+ CD4+ , CD3+ CD8+ lymphocytes in spleen were decreased significantly(t=5.754, 3.570, 4.442, 5.281, 4.570, P<0.05). Cimetidine could partially restore the weight of irradiated mice, relieved spleen structure damage and effectively restored lymphocyte subsets, and partly recovered the levels of CD3+ , CD3+ CD4+ , CD3+ CD8+ lymphocytes in the spleen (t=3.523, 2.706, 2.520, P<0.05). Conclusions Cimetidine, as a potential radiation protection drug, could effectively reduce the mouse spleen injury induced by X-ray fractionated irradiation by improving mouse immune function. Key words: Radiotherapy; Abscopal effect; Cimetidine; Spleen lymphocytes
目的探讨西咪替丁对小鼠右胸x线分割照射的保护作用。方法将30只雄性C57BL/6小鼠随机分为对照组、辐照组和辐照+西咪替丁(100 mg/kg)组,每组10只。小鼠右胸局部x射线照射,剂量率为0.883 Gy/min,累计剂量为24 Gy,每天8 Gy,连续照射3 d。照射后,照射+西咪替丁组给予西咪替丁(100 mg·kg-1·d-1),连续7 d,对照组和IR组给予生理盐水。照射后第7天,测定小鼠体重、脾脏指数及其T淋巴细胞、B淋巴细胞和NK细胞。脾组织采用HE染色进行病理观察。结果分级辐照对小鼠生长有轻微抑制作用,但对脾脏有明显的体外损伤,使脾脏指数降低。照射组(0.369±0.011)与对照组(0.396±0.022)(t=2.978, P<0.05)比较差异有统计学意义,脾脏病理结构紊乱,脾脏CD3+、CD4+、CD3+ CD4+、CD3+ CD8+淋巴细胞百分比明显降低(t=5.754、3.570、4.442、5.281、4.570,P<0.05)。西咪替丁能部分恢复照射小鼠体重,减轻脾脏结构损伤,有效恢复淋巴细胞亚群,部分恢复脾脏CD3+、CD3+ CD4+、CD3+ CD8+淋巴细胞水平(t=3.523、2.706、2.520,P<0.05)。结论西咪替丁作为一种潜在的辐射防护药物,可通过改善小鼠免疫功能,有效减轻x线分割照射对小鼠脾脏的损伤。关键词:放疗;Abscopal效果;西咪替丁;脾淋巴细胞
{"title":"Therapy effect of cimetidine on spleen of mice irradiated with thoracic fractionated irradiation","authors":"Yang Bai, Junling Zhang, Dan Yao, Jianghong Zhang","doi":"10.3760/CMA.J.ISSN.0254-5098.2020.03.004","DOIUrl":"https://doi.org/10.3760/CMA.J.ISSN.0254-5098.2020.03.004","url":null,"abstract":"Objective \u0000To investigate the protective effect of cimetidine on mice with fractionated X-ray irradiation through right thorax. \u0000 \u0000 \u0000Methods \u0000Thirty male C57BL/6 mice were randomly divided into control group, irradiation group and irradiation + cimetidine group (100 mg/kg), 10 mice in each group. Mice right thorax was locally irradiated with X-rays at a dose rate of 0.883 Gy/min to a cumulative dose of 24 Gy, 8 Gy per day in 3 d. After irradiation, the irradiation + cimetidine group was given cimetidine (100 mg·kg-1·d-1) for 7 d consecutively, and Ctrl and IR groups were given saline. Seven days after irradiation, the mice weight, spleen index and its T lymphocytes, B lymphocytes and NK cells were measured. Spleen tissue was stained with HE for pathological observation. \u0000 \u0000 \u0000Results \u0000The fractionated irradiation lightly suppressed mice growth but obviously induced abscopal damage on spleen so that the spleen index were decreased. The irradiation group (0.369±0.011) had statistically significant difference compared with the control group(0.396±0.022)(t=2.978, P<0.05), the spleen pathological structure became disorder, and the percentages of CD3+ , CD4+ , CD3+ CD4+ , CD3+ CD8+ lymphocytes in spleen were decreased significantly(t=5.754, 3.570, 4.442, 5.281, 4.570, P<0.05). Cimetidine could partially restore the weight of irradiated mice, relieved spleen structure damage and effectively restored lymphocyte subsets, and partly recovered the levels of CD3+ , CD3+ CD4+ , CD3+ CD8+ lymphocytes in the spleen (t=3.523, 2.706, 2.520, P<0.05). \u0000 \u0000 \u0000Conclusions \u0000Cimetidine, as a potential radiation protection drug, could effectively reduce the mouse spleen injury induced by X-ray fractionated irradiation by improving mouse immune function. \u0000 \u0000 \u0000Key words: \u0000Radiotherapy; Abscopal effect; Cimetidine; Spleen lymphocytes","PeriodicalId":36403,"journal":{"name":"中华放射医学与防护杂志","volume":"15 1","pages":"178-182"},"PeriodicalIF":0.0,"publicationDate":"2020-03-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"75105868","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
南京 192 Ir放射事故患者照后第4年细胞遗传学随访 南京 192 Ir放射事故患者照后第4年细胞遗传学随访
Q4 Medicine Pub Date : 2020-03-25 DOI: 10.3760/CMA.J.ISSN.0254-5098.2020.03.005
陆雪, Xue Lu, 赵骅, Zhao Hua, 王进, Wang Jin, 李爽, Liang Shuang, 田雪蕾, Tian Xuelei, 王福如, Wang Furu, 田梅, Tian Mei, 余宁乐, Yu Ningle, 刘青杰, Liu Qingjie
Objective To screen the indicators of retrospective dose estimation, based on 5 cytogenetic methods to assess the victim followed-up at 4 year after 192Ir radiation accident in Nanjing. Methods The chromosome aberration (dic + r) assay, cytokinesis block micronucleus (MN) and nucleoplasmic bridge (NPB) assay, fluorescence in situ hybridization (FISH)-based and G banding-based translocation analysis were used to retrospective biological dose estimation. Results The estimated doses of FISH-based and G banding -based analysis were 1.45 and 1.21 Gy respectively, which was similar to the biological dose estimated short time after the accident. However, the estimated doses by chromosome aberration, micronucleus and nucleoplasmic bridge method were 0.56, 0.45 and 0.41 Gy respectively, which were lower than the corresponding biodose. Correction factors were used to adjust the biodose. Conclusions In the 4th years after exposure, the estimated biological doses by FISH-based and G banding-based translocation were consistent with the biodose.Therefore, the two methods were suitable for retrospective dose estimation, while correction factors should be considered in chromosome aberration method for retrospective dose estimation. Key words: Chromosome aberration; Fluorescence in situ hybrdization; G banding; Biological dose estimation; Correction coefficient
目的筛选基于5种细胞遗传学方法的回顾性剂量估计指标,对南京地区192Ir辐射事故4年随访患者进行评估。方法采用染色体畸变(dic + r)测定、细胞分裂阻断微核(MN)和核质桥(NPB)测定、基于荧光原位杂交(FISH)和基于G带易位分析进行回顾性生物剂量估计。结果基于fish和G带分析的估计剂量分别为1.45 Gy和1.21 Gy,与事故发生后短时间内估计的生物剂量相近。而染色体畸变法、微核法和核质桥法的估计剂量分别为0.56、0.45和0.41 Gy,均低于相应的生物剂量。校正因子用于调整生物剂量。结论暴露后第4年,基于fish和基于G带易位估计的生物剂量与生物剂量一致。因此,这两种方法适用于回顾性剂量估计,而染色体畸变法用于回顾性剂量估计时应考虑校正因素。关键词:染色体畸变;荧光原位杂交;G显带;生物剂量估计;修正系数
{"title":"南京 192 Ir放射事故患者照后第4年细胞遗传学随访","authors":"陆雪, Xue Lu, 赵骅, Zhao Hua, 王进, Wang Jin, 李爽, Liang Shuang, 田雪蕾, Tian Xuelei, 王福如, Wang Furu, 田梅, Tian Mei, 余宁乐, Yu Ningle, 刘青杰, Liu Qingjie","doi":"10.3760/CMA.J.ISSN.0254-5098.2020.03.005","DOIUrl":"https://doi.org/10.3760/CMA.J.ISSN.0254-5098.2020.03.005","url":null,"abstract":"Objective \u0000To screen the indicators of retrospective dose estimation, based on 5 cytogenetic methods to assess the victim followed-up at 4 year after 192Ir radiation accident in Nanjing. \u0000 \u0000 \u0000Methods \u0000The chromosome aberration (dic + r) assay, cytokinesis block micronucleus (MN) and nucleoplasmic bridge (NPB) assay, fluorescence in situ hybridization (FISH)-based and G banding-based translocation analysis were used to retrospective biological dose estimation. \u0000 \u0000 \u0000Results \u0000The estimated doses of FISH-based and G banding -based analysis were 1.45 and 1.21 Gy respectively, which was similar to the biological dose estimated short time after the accident. However, the estimated doses by chromosome aberration, micronucleus and nucleoplasmic bridge method were 0.56, 0.45 and 0.41 Gy respectively, which were lower than the corresponding biodose. Correction factors were used to adjust the biodose. \u0000 \u0000 \u0000Conclusions \u0000In the 4th years after exposure, the estimated biological doses by FISH-based and G banding-based translocation were consistent with the biodose.Therefore, the two methods were suitable for retrospective dose estimation, while correction factors should be considered in chromosome aberration method for retrospective dose estimation. \u0000 \u0000 \u0000Key words: \u0000Chromosome aberration; Fluorescence in situ hybrdization; G banding; Biological dose estimation; Correction coefficient","PeriodicalId":36403,"journal":{"name":"中华放射医学与防护杂志","volume":"140 1","pages":"183-188"},"PeriodicalIF":0.0,"publicationDate":"2020-03-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"86146439","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 1
Progress on the fundamental biological studies for tumor therapy with heavy particles 重粒子治疗肿瘤的基础生物学研究进展
Q4 Medicine Pub Date : 2020-03-25 DOI: 10.3760/CMA.J.ISSN.0254-5098.2020.03.016
Yingchu Dai, Jiade J. Lu, Guangming Zhou
Due to its unique physical and biological advantages, heavy ion beam has great advantages as tumor radiotherapy ray. Tumor therapy with heavy particles has achieved outstanding clinical outcome and is of the potential to become the most advanced radiotherapy in this century. Fundamental studies on the biological effects are of significance in providing theoretical principle and optimizing beam quality for tumor therapy with heavy particles. Meanwhile, the application of new techniques such as FLASH and GRID in tumor therapy with heavy particles requires fundamental biological studies as well, especially in improving clinical efficacy and security. In this article, we review the progress on the fundamental biological studies for tumor therapy with heavy particles and arouse discussion on the issues necessary for further elucidation. Key words: Heavy ion; Tumor radiotherapy; Relative biological effectiveness; Oxygen enhancement ratio
由于其独特的物理和生物学优势,重离子束作为肿瘤放疗射线具有很大的优势。重粒子治疗肿瘤取得了突出的临床效果,有可能成为本世纪最先进的放疗技术。生物学效应的基础研究对于为重粒子治疗肿瘤提供理论依据和优化束流质量具有重要意义。同时,FLASH、GRID等新技术在重颗粒肿瘤治疗中的应用也需要基础生物学研究,尤其是在提高临床疗效和安全性方面。本文就重粒子治疗肿瘤的基础生物学研究进展进行综述,并对需要进一步阐明的问题进行讨论。关键词:重离子;肿瘤放射治疗;相对生物有效性;氧增强比
{"title":"Progress on the fundamental biological studies for tumor therapy with heavy particles","authors":"Yingchu Dai, Jiade J. Lu, Guangming Zhou","doi":"10.3760/CMA.J.ISSN.0254-5098.2020.03.016","DOIUrl":"https://doi.org/10.3760/CMA.J.ISSN.0254-5098.2020.03.016","url":null,"abstract":"Due to its unique physical and biological advantages, heavy ion beam has great advantages as tumor radiotherapy ray. Tumor therapy with heavy particles has achieved outstanding clinical outcome and is of the potential to become the most advanced radiotherapy in this century. Fundamental studies on the biological effects are of significance in providing theoretical principle and optimizing beam quality for tumor therapy with heavy particles. Meanwhile, the application of new techniques such as FLASH and GRID in tumor therapy with heavy particles requires fundamental biological studies as well, especially in improving clinical efficacy and security. In this article, we review the progress on the fundamental biological studies for tumor therapy with heavy particles and arouse discussion on the issues necessary for further elucidation. \u0000 \u0000 \u0000Key words: \u0000Heavy ion; Tumor radiotherapy; Relative biological effectiveness; Oxygen enhancement ratio","PeriodicalId":36403,"journal":{"name":"中华放射医学与防护杂志","volume":"132 1","pages":"247-252"},"PeriodicalIF":0.0,"publicationDate":"2020-03-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"79659009","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
The application value of low-dose CT scan in pregnant women with corona virus disease 2019 2019年冠状病毒病孕妇低剂量CT扫描的应用价值
Q4 Medicine Pub Date : 2020-03-16 DOI: 10.3760/CMA.J.ISSN.0254-5098.2020.0009
Liang Li, Li Wang, Feifei Zeng, Fang Liu, Zhoufeng Peng, Baojun Xie, Changsheng Liu, Y. Zha
Objective To explore the value of low-dose CT in pregnancy with corona virus disease 2019(COVID-19). Methods A retrospective analysis was performed on the clinical characteristics, laboratory tests, and chest CT findings of 12 patients with pregnancy and COVID-19 diagnosed by nucleic acid testing from Renmin Hospital of Wuhan University from January 20, 2020 to February 16, 2020. Two radiologists blinded to the reconstruction algorithm independently scored subjective image quality on a 5-point Likert scale (grade 1, major noise, heavy artifacts, almost not visible, no diagnosis possible; grade 2, above average noise, poor, substantial, diagnostic confidence substantially reduced; grade 3, average image noise, moderate, but sufficient for diagnosis; grade 4, less than average noise, good, below average; grade 5, minimal noise, excellent). A clinical diagnosis of ≥3 was acceptable. The doctors recorded the CT radiation dose, including CT volume dose index (CTDIvol), dose length product (DLP), and effective radiation dose (E). Then two radiologists observed the distribution, shape, density, and other characteristics of lung lesions, and decided whether hilar, mediastinal lymphadenopathy, and pleural changes. Results Of the 12 pregnancy with COVID-19, 8 had cough, 4 had fever, 2 had chest tightness, and 1 had dyspnea and diarrhea each. The CT image quality score of all patients was (3-4), with an average of 3.46, which fully met the clinical diagnosis requirements. The CTDIvol value was (1.13 - 4.31) mGy, with an average of 3.02 mGy. The DLP value was (34.48 -75.29) mGy • cm, with an average of 55.48 mGy • cm. The E value was (0.48 - 1.05) mSv, with an average of 0.78 mSv. In all cases, chest CT examination showed abnormal manifestations after clinical symptoms, including unilateral lung lesions in 5 cases and bilateral lung lesions in 7 cases. 1 case of ground-glass opacity, 1 case of solidification, 7 cases of ground-glass and consolidation, 1 case of strip opacity, ground-glass, consolidation and strip cable shadow coexisted in 2 cases. Conclusions The application of low-dose CT scan in pregnancy with COVID-19 is completely feasible. CT mainly manifested as bilateral lung patchy, flaky ground-glass opacity with consolidation. Active and effective treatment can completely recover, prognosis better. Key words: Radiation dose; Pregnancy; Novel coronavirus infection; Pneumonia; Tomography, X-ray computed
目的探讨低剂量CT对2019冠状病毒病(COVID-19)妊娠的诊断价值。方法回顾性分析2020年1月20日至2月16日武汉大学人民医院经核酸检测确诊的12例妊娠合并新冠肺炎患者的临床特征、实验室检查及胸部CT表现。两名不知道重建算法的放射科医生独立对5分李克特量表的主观图像质量进行评分(1级,噪声大,伪影重,几乎看不见,无法诊断;2级,噪声高于平均水平,较差,实质,诊断置信度大幅度降低;3级,图像噪声一般,中等,但足以诊断;4级,噪音小于平均水平,良好,低于平均水平;5级,噪音最小,优秀)。临床诊断≥3是可以接受的。医生记录CT辐射剂量,包括CT体积剂量指数(CTDIvol)、剂量长度积(DLP)、有效辐射剂量(E),由两名放射科医师观察肺病变的分布、形态、密度等特征,判断肺门、纵隔淋巴结病变、胸膜有无改变。结果12例妊娠新冠肺炎患者中,咳嗽8例,发热4例,胸闷2例,呼吸困难、腹泻各1例。所有患者的CT图像质量评分为(3-4)分,平均3.46分,完全满足临床诊断要求。CTDIvol值为(1.13 ~ 4.31)mGy,平均为3.02 mGy。DLP值为(34.48 ~ 75.29)mGy•cm,平均55.48 mGy•cm。E值为(0.48 ~ 1.05)mSv,平均为0.78 mSv。所有病例均在出现临床症状后进行胸部CT检查发现异常表现,其中单侧肺病变5例,双侧肺病变7例。毛玻璃混浊1例,凝固1例,毛玻璃合并固结7例,条状混浊1例,毛玻璃、固结、条状电缆影共存2例。结论低剂量CT扫描在妊娠新冠肺炎中的应用是完全可行的。CT主要表现为双侧肺斑片状,片状磨玻璃影伴实变。积极有效治疗可完全康复,预后较好。关键词:辐射剂量;怀孕;新型冠状病毒感染;肺炎;x线计算机断层扫描
{"title":"The application value of low-dose CT scan in pregnant women with corona virus disease 2019","authors":"Liang Li, Li Wang, Feifei Zeng, Fang Liu, Zhoufeng Peng, Baojun Xie, Changsheng Liu, Y. Zha","doi":"10.3760/CMA.J.ISSN.0254-5098.2020.0009","DOIUrl":"https://doi.org/10.3760/CMA.J.ISSN.0254-5098.2020.0009","url":null,"abstract":"Objective \u0000To explore the value of low-dose CT in pregnancy with corona virus disease 2019(COVID-19). \u0000 \u0000 \u0000Methods \u0000A retrospective analysis was performed on the clinical characteristics, laboratory tests, and chest CT findings of 12 patients with pregnancy and COVID-19 diagnosed by nucleic acid testing from Renmin Hospital of Wuhan University from January 20, 2020 to February 16, 2020. Two radiologists blinded to the reconstruction algorithm independently scored subjective image quality on a 5-point Likert scale (grade 1, major noise, heavy artifacts, almost not visible, no diagnosis possible; grade 2, above average noise, poor, substantial, diagnostic confidence substantially reduced; grade 3, average image noise, moderate, but sufficient for diagnosis; grade 4, less than average noise, good, below average; grade 5, minimal noise, excellent). A clinical diagnosis of ≥3 was acceptable. The doctors recorded the CT radiation dose, including CT volume dose index (CTDIvol), dose length product (DLP), and effective radiation dose (E). Then two radiologists observed the distribution, shape, density, and other characteristics of lung lesions, and decided whether hilar, mediastinal lymphadenopathy, and pleural changes. \u0000 \u0000 \u0000Results \u0000Of the 12 pregnancy with COVID-19, 8 had cough, 4 had fever, 2 had chest tightness, and 1 had dyspnea and diarrhea each. The CT image quality score of all patients was (3-4), with an average of 3.46, which fully met the clinical diagnosis requirements. The CTDIvol value was (1.13 - 4.31) mGy, with an average of 3.02 mGy. The DLP value was (34.48 -75.29) mGy • cm, with an average of 55.48 mGy • cm. The E value was (0.48 - 1.05) mSv, with an average of 0.78 mSv. In all cases, chest CT examination showed abnormal manifestations after clinical symptoms, including unilateral lung lesions in 5 cases and bilateral lung lesions in 7 cases. 1 case of ground-glass opacity, 1 case of solidification, 7 cases of ground-glass and consolidation, 1 case of strip opacity, ground-glass, consolidation and strip cable shadow coexisted in 2 cases. \u0000 \u0000 \u0000Conclusions \u0000The application of low-dose CT scan in pregnancy with COVID-19 is completely feasible. CT mainly manifested as bilateral lung patchy, flaky ground-glass opacity with consolidation. Active and effective treatment can completely recover, prognosis better. \u0000 \u0000 \u0000Key words: \u0000Radiation dose; Pregnancy; Novel coronavirus infection; Pneumonia; Tomography, X-ray computed","PeriodicalId":36403,"journal":{"name":"中华放射医学与防护杂志","volume":"32 1","pages":""},"PeriodicalIF":0.0,"publicationDate":"2020-03-16","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"91010881","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Monitoring of CT dose and analysis of radiological protection in CT module applied for COVID-19 prevention and control 新型冠状病毒肺炎防控中CT模块的剂量监测及放射防护分析
Q4 Medicine Pub Date : 2020-03-16 DOI: 10.3760/CMA.J.ISSN.0254-5098.0008
Ansheng Liu, Bolin Huang, Cuiling Li, Kai Yang, H. Du, Guilin Yi
Objective To investigate the safety and protection level of radiological treatment in the CT module in Wuhan during the prevention and treatment of COVID-19. Methods The layout position of the CT module in makeshift hospitals, radiological protection facilities and personal protective equipment were investigated. Based on the national standards, CT dose index and evaluated radiological protection level of the CT module were measured. Results The layout position of the CT module in makeshift hospitals are reasonable, and radiological protection facilities are well-equipped. 87.0% of them had been accepted. In 23 CT module, 20 of which were up to standards. The other three had 1 detection points which exceed the national standards, and then had been modified immediately, so as to reach the national standards. In addition, CT dose index was measured in 7 CT module of them, the variation of body CTDIW was less than 7.5%. Conclusions The CT module in Wuhan meet the requirements of the radiological safety and protection level during the prevention and treatment of COVID-19. Key words: COVID-19; CT module; Radiological protection; CT dose index (CTDI)
目的了解武汉市新型冠状病毒肺炎防治过程中CT模块放射治疗的安全性和防护水平。方法对方舱医院、放射防护设施、个人防护用品的CT模块布置位置进行调查。根据国家标准,测量了CT模块的剂量指数和评估放射防护水平。结果方舱医院CT模块布置位置合理,放射防护设施齐全。录取率为87.0%。在23个CT模块中,20个模块达到标准。另外3家有1个检测点超过国家标准,随即进行了整改,达到国家标准。另外,对其中7个CT模块的CT剂量指数进行了测量,机体CTDIW的变化均小于7.5%。结论武汉市CT模块满足新型冠状病毒肺炎防治期间的放射安全防护水平要求。关键词:COVID-19;CT模块;放射防护;CT剂量指数(CTDI)
{"title":"Monitoring of CT dose and analysis of radiological protection in CT module applied for COVID-19 prevention and control","authors":"Ansheng Liu, Bolin Huang, Cuiling Li, Kai Yang, H. Du, Guilin Yi","doi":"10.3760/CMA.J.ISSN.0254-5098.0008","DOIUrl":"https://doi.org/10.3760/CMA.J.ISSN.0254-5098.0008","url":null,"abstract":"Objective \u0000To investigate the safety and protection level of radiological treatment in the CT module in Wuhan during the prevention and treatment of COVID-19. \u0000 \u0000 \u0000Methods \u0000The layout position of the CT module in makeshift hospitals, radiological protection facilities and personal protective equipment were investigated. Based on the national standards, CT dose index and evaluated radiological protection level of the CT module were measured. \u0000 \u0000 \u0000Results \u0000The layout position of the CT module in makeshift hospitals are reasonable, and radiological protection facilities are well-equipped. 87.0% of them had been accepted. In 23 CT module, 20 of which were up to standards. The other three had 1 detection points which exceed the national standards, and then had been modified immediately, so as to reach the national standards. In addition, CT dose index was measured in 7 CT module of them, the variation of body CTDIW was less than 7.5%. \u0000 \u0000 \u0000Conclusions \u0000The CT module in Wuhan meet the requirements of the radiological safety and protection level during the prevention and treatment of COVID-19. \u0000 \u0000 \u0000Key words: \u0000COVID-19; CT module; Radiological protection; CT dose index (CTDI)","PeriodicalId":36403,"journal":{"name":"中华放射医学与防护杂志","volume":"11 1","pages":""},"PeriodicalIF":0.0,"publicationDate":"2020-03-16","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"73830810","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Investigation and analysis on the infection control and radiation safety of radiodiagnostic workplace for COVID-19 新型冠状病毒肺炎放射诊断工作场所感染控制与辐射安全调查分析
Q4 Medicine Pub Date : 2020-03-09 DOI: 10.3760/CMA.J.ISSN.0254-5098.2020.04.001
L. Tang, Wenshan Zhou, Sheng Liu, Xiangjun Yang, Xixiang Huo, Shuguang Xie, Jing Li, Ansheng Liu, Shi-xiong Tang, Z. Lei, Weimin Chen, Y. Niu
Objectives To survey and supervise the risk of infection control and radiation safety in the radiological diagnostic workplace for novel coronavirus pneumonia(COVID-19), and provide data support for the safety protection of radiographers and related staff. Methods 4 emergency hospitals for COVID-19 including 2 makeshift hospitals, module hospital and brick pattern hospital were performed the testing and evaluation of imaging performance and radiological protection for the 8 new installed CT scanners and places according to the national standards of WS 519-2019 and GBZ 130-2013. The infection control safety factors such as the layout of the equipment room were monitored and investigated. Two COVID-19 designated hospitals including general hospital and infectious disease specialized hospital were selected to carry out field investigation and sampling of environmental biological samples for 4 CT rooms. Then the samples were detected for the nucleic acid of novel coronavirus. The results of radiodiagnostic workplace overall arrangement, infection prevention and the nucleic acid testing were analyzed, and the biological safety reliability and risk point were evaluated. Results The indicators of imaging performance and radiation protection for 8 CT scanners in emergency hospitals could meet the requirements of national standards. Each of 2 makeshift hospitals had 3 CT rooms with the area of 38.8 m2 and shielding protection of 4 mm Pb. The CT rooms in module hospital and brick pattern hospital had the area and shielding protection of 20.0 m2, 4 mm Pb equivalent and 35.8 m2, 3 mm Pb equivalent, respectively. The 8 radiological diagnostic workplaces of the emergency hospitals were designed and constructed based on "three zones with two passageways" . The testing results of the nucleic acid indicated that the multiple sites such as scanning bed, internal of gantry and ground touched by patients in CT scanning room were found positive samples. The areas such as console panel and ground were got polluted risk by the virus infected hands and feet of radiographers. In addition, the areas in scanning room with no touch of patients, such as observation window and air outlet, were found similar positive samples. Conclusions: 8 CT scanners and rooms in 4 emergency hospitals basically meet the requirements of imaging performance and radiation protection. The disinfection of COVID-19 radiodiagnostic workplace should be standardized. Key words: COVID-19; Radiodiagnostic workplace; Radiation safety; Infection control; Module CT
目的调查和监督新型冠状病毒肺炎(COVID-19)放射诊断工作场所感染控制和辐射安全风险,为放射技师及相关工作人员的安全防护提供数据支持。方法根据WS 519-2019和GBZ 130-2013国家标准,对新安装的8台CT扫描仪及场所进行影像学性能和放射防护检测与评价。对机房布置等感染控制安全因素进行监测和调查。选择综合医院和传染病专科医院两家新冠定点医院,对4个CT室进行现场调查和环境生物样本采样。然后对样品进行新型冠状病毒核酸检测。对放射诊断工作场所总体布置、感染预防和核酸检测结果进行分析,并对生物安全可靠性和风险点进行评价。结果急诊医院8台CT扫描仪的成像性能和辐射防护指标均达到国家标准要求。2家方舱医院各设CT室3间,面积38.8 m2,屏蔽防护4 mm Pb。模块式医院CT室和砖式医院CT室的面积和屏蔽防护分别为20.0 m2, 4 mm Pb当量和35.8 m2, 3 mm Pb当量。急诊医院8个放射诊断工作场所按“三区两通道”设计施工。核酸检测结果显示,CT扫描室患者接触的扫描床、龙门架内部、地面等多个部位均有阳性样本。控制台和地面等区域受到辐射技师感染病毒的手脚的污染风险。此外,扫描室观察窗、出风口等无患者接触区域也发现了类似阳性样本。结论:4家急诊医院的8台CT扫描仪和房间基本满足成像性能和辐射防护要求。规范新冠肺炎放射诊断工作场所消毒。关键词:COVID-19;Radiodiagnostic工作场所;辐射安全;感染控制;模块CT
{"title":"Investigation and analysis on the infection control and radiation safety of radiodiagnostic workplace for COVID-19","authors":"L. Tang, Wenshan Zhou, Sheng Liu, Xiangjun Yang, Xixiang Huo, Shuguang Xie, Jing Li, Ansheng Liu, Shi-xiong Tang, Z. Lei, Weimin Chen, Y. Niu","doi":"10.3760/CMA.J.ISSN.0254-5098.2020.04.001","DOIUrl":"https://doi.org/10.3760/CMA.J.ISSN.0254-5098.2020.04.001","url":null,"abstract":"Objectives \u0000To survey and supervise the risk of infection control and radiation safety in the radiological diagnostic workplace for novel coronavirus pneumonia(COVID-19), and provide data support for the safety protection of radiographers and related staff. \u0000 \u0000 \u0000Methods \u00004 emergency hospitals for COVID-19 including 2 makeshift hospitals, module hospital and brick pattern hospital were performed the testing and evaluation of imaging performance and radiological protection for the 8 new installed CT scanners and places according to the national standards of WS 519-2019 and GBZ 130-2013. The infection control safety factors such as the layout of the equipment room were monitored and investigated. Two COVID-19 designated hospitals including general hospital and infectious disease specialized hospital were selected to carry out field investigation and sampling of environmental biological samples for 4 CT rooms. Then the samples were detected for the nucleic acid of novel coronavirus. The results of radiodiagnostic workplace overall arrangement, infection prevention and the nucleic acid testing were analyzed, and the biological safety reliability and risk point were evaluated. \u0000 \u0000 \u0000Results \u0000The indicators of imaging performance and radiation protection for 8 CT scanners in emergency hospitals could meet the requirements of national standards. Each of 2 makeshift hospitals had 3 CT rooms with the area of 38.8 m2 and shielding protection of 4 mm Pb. The CT rooms in module hospital and brick pattern hospital had the area and shielding protection of 20.0 m2, 4 mm Pb equivalent and 35.8 m2, 3 mm Pb equivalent, respectively. The 8 radiological diagnostic workplaces of the emergency hospitals were designed and constructed based on \"three zones with two passageways\" . The testing results of the nucleic acid indicated that the multiple sites such as scanning bed, internal of gantry and ground touched by patients in CT scanning room were found positive samples. The areas such as console panel and ground were got polluted risk by the virus infected hands and feet of radiographers. In addition, the areas in scanning room with no touch of patients, such as observation window and air outlet, were found similar positive samples. \u0000 \u0000 \u0000Conclusions: \u00008 CT scanners and rooms in 4 emergency hospitals basically meet the requirements of imaging performance and radiation protection. The disinfection of COVID-19 radiodiagnostic workplace should be standardized. \u0000 \u0000 \u0000Key words: \u0000COVID-19; Radiodiagnostic workplace; Radiation safety; Infection control; Module CT","PeriodicalId":36403,"journal":{"name":"中华放射医学与防护杂志","volume":"26 1","pages":"253-258"},"PeriodicalIF":0.0,"publicationDate":"2020-03-09","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"83224266","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
期刊
中华放射医学与防护杂志
全部 Acc. Chem. Res. ACS Applied Bio Materials ACS Appl. Electron. Mater. ACS Appl. Energy Mater. ACS Appl. Mater. Interfaces ACS Appl. Nano Mater. ACS Appl. Polym. Mater. ACS BIOMATER-SCI ENG ACS Catal. ACS Cent. Sci. ACS Chem. Biol. ACS Chemical Health & Safety ACS Chem. Neurosci. ACS Comb. Sci. ACS Earth Space Chem. ACS Energy Lett. ACS Infect. Dis. ACS Macro Lett. ACS Mater. Lett. ACS Med. Chem. Lett. ACS Nano ACS Omega ACS Photonics ACS Sens. ACS Sustainable Chem. Eng. ACS Synth. Biol. Anal. Chem. BIOCHEMISTRY-US Bioconjugate Chem. BIOMACROMOLECULES Chem. Res. Toxicol. Chem. Rev. Chem. Mater. CRYST GROWTH DES ENERG FUEL Environ. Sci. Technol. Environ. Sci. Technol. Lett. Eur. J. Inorg. Chem. IND ENG CHEM RES Inorg. Chem. J. Agric. Food. Chem. J. Chem. Eng. Data J. Chem. Educ. J. Chem. Inf. Model. J. Chem. Theory Comput. J. Med. Chem. J. Nat. Prod. J PROTEOME RES J. Am. Chem. Soc. LANGMUIR MACROMOLECULES Mol. Pharmaceutics Nano Lett. Org. Lett. ORG PROCESS RES DEV ORGANOMETALLICS J. Org. Chem. J. Phys. Chem. J. Phys. Chem. A J. Phys. Chem. B J. Phys. Chem. C J. Phys. Chem. Lett. Analyst Anal. Methods Biomater. Sci. Catal. Sci. Technol. Chem. Commun. Chem. Soc. Rev. CHEM EDUC RES PRACT CRYSTENGCOMM Dalton Trans. Energy Environ. Sci. ENVIRON SCI-NANO ENVIRON SCI-PROC IMP ENVIRON SCI-WAT RES Faraday Discuss. Food Funct. Green Chem. Inorg. Chem. Front. Integr. Biol. J. Anal. At. Spectrom. J. Mater. Chem. A J. Mater. Chem. B J. Mater. Chem. C Lab Chip Mater. Chem. Front. Mater. Horiz. MEDCHEMCOMM Metallomics Mol. Biosyst. Mol. Syst. Des. Eng. Nanoscale Nanoscale Horiz. Nat. Prod. Rep. New J. Chem. Org. Biomol. Chem. Org. Chem. Front. PHOTOCH PHOTOBIO SCI PCCP Polym. Chem.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1