首页 > 最新文献

Clinical chemistry and laboratory medicine最新文献

英文 中文
Interference of therapeutic monoclonal antibodies with electrophoresis and immunofixation of serum proteins: state of knowledge and systematic review. 用血清蛋白电泳和免疫固定干扰治疗性单克隆抗体:知识现状和系统综述。
IF 3.7 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2025-09-01 Print Date: 2025-11-25 DOI: 10.1515/cclm-2025-0678
Sacha Pelletier, Laetitia Florent, Philippe Gillery, Jean-Baptiste Oudart

Introduction: The increasing use of therapeutic monoclonal antibodies (t-mAbs) has improved cancer and autoimmune disorder treatment. These therapeutics can interfere with serum protein electrophoresis (SPEP) and immunofixation (IF), potentially leading to the appearance of monoclonal bands that may be misinterpreted as monoclonal gammopathies. Identifying the migration patterns and detection thresholds of t-mAbs is crucial to avoid misinterpretation in clinical laboratories.

Content: A systematic review following PRISMA guidelines was conducted using Pubmed and ScienceDirect databases, with algorithm-based searches and double-blind article selection. Data on the matrix used, separation methods and type of interference were collected into an extraction table.

Summary: A total of 30 articles were included and 30 t-mAbs were described. 11 t-mAbs migrated at the end of the gamma region, 12 in the mid-gamma region, 5 in the early gamma region, one in the beta-2 globulin region and one in the alpha-2 globulin region. Most t-mAbs were detectable by SPEP and IF at concentrations above 100 mg/L.

Outlook: Caution is needed when a new peak appears on SPEP, as it may be mistaken for a monoclonal spike leading to misdiagnosis. Therefore, understanding the migration profiles of these t-mAbs is essential. Different methods are available to remove t-mAbs interference and could be used in daily practice.

越来越多地使用治疗性单克隆抗体(t- mab)改善了癌症和自身免疫性疾病的治疗。这些疗法可能干扰血清蛋白电泳(SPEP)和免疫固定(IF),潜在地导致单克隆带的出现,可能被误解为单克隆伽玛病。确定t- mab的迁移模式和检测阈值对于避免临床实验室的误解至关重要。内容:使用Pubmed和ScienceDirect数据库,采用基于算法的搜索和双盲文章选择,按照PRISMA指南进行系统评价。将所用基质、分离方法和干扰类型的数据收集到提取表中。总结:共纳入了30篇文章,描述了30个 t- mab。11个 t- mab在γ区末端迁移,12个在γ区中期迁移,5个在γ区早期迁移,1个在β -2球蛋白区迁移,1个在α -2球蛋白区迁移。大多数t- mab在浓度大于100 mg/L时可被SPEP和IF检测到。展望:当SPEP上出现新的峰时需要谨慎,因为它可能被误认为单克隆峰导致误诊。因此,了解这些t- mab的迁移特征是必不可少的。有不同的方法可以消除t- mab的干扰,并可用于日常实践。
{"title":"Interference of therapeutic monoclonal antibodies with electrophoresis and immunofixation of serum proteins: state of knowledge and systematic review.","authors":"Sacha Pelletier, Laetitia Florent, Philippe Gillery, Jean-Baptiste Oudart","doi":"10.1515/cclm-2025-0678","DOIUrl":"10.1515/cclm-2025-0678","url":null,"abstract":"<p><strong>Introduction: </strong>The increasing use of therapeutic monoclonal antibodies (t-mAbs) has improved cancer and autoimmune disorder treatment. These therapeutics can interfere with serum protein electrophoresis (SPEP) and immunofixation (IF), potentially leading to the appearance of monoclonal bands that may be misinterpreted as monoclonal gammopathies. Identifying the migration patterns and detection thresholds of t-mAbs is crucial to avoid misinterpretation in clinical laboratories.</p><p><strong>Content: </strong>A systematic review following PRISMA guidelines was conducted using Pubmed and ScienceDirect databases, with algorithm-based searches and double-blind article selection. Data on the matrix used, separation methods and type of interference were collected into an extraction table.</p><p><strong>Summary: </strong>A total of 30 articles were included and 30 t-mAbs were described. 11 t-mAbs migrated at the end of the gamma region, 12 in the mid-gamma region, 5 in the early gamma region, one in the beta-2 globulin region and one in the alpha-2 globulin region. Most t-mAbs were detectable by SPEP and IF at concentrations above 100 mg/L.</p><p><strong>Outlook: </strong>Caution is needed when a new peak appears on SPEP, as it may be mistaken for a monoclonal spike leading to misdiagnosis. Therefore, understanding the migration profiles of these t-mAbs is essential. Different methods are available to remove t-mAbs interference and could be used in daily practice.</p>","PeriodicalId":10390,"journal":{"name":"Clinical chemistry and laboratory medicine","volume":" ","pages":"2355-2365"},"PeriodicalIF":3.7,"publicationDate":"2025-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"144945339","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
An innovative immunoassay for accurate aldosterone quantification: overcoming low-level inaccuracy and renal dysfunction-associated interference. 一个创新的免疫测定准确醛固酮定量:克服低水平的不准确性和肾功能障碍相关的干扰。
IF 3.7 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2025-08-25 Print Date: 2025-11-25 DOI: 10.1515/cclm-2025-0743
Kaijuan Wang, Hongying Cong, Zhangwei Gao, Xiaojing Gao, Wei Zhang, Xiaocui Shi, Zhou Zhou

Objectives: Accurate quantification of aldosterone is critical for screening and diagnosing primary aldosteronism (PA). Current competitive chemiluminescence immunoassays (cCLIA) overestimate plasma aldosterone concentration (PAC) compared to liquid chromatography-tandem mass spectrometry (LC-MS/MS). However, LC-MS/MS is technically demanding and time-consuming, limiting its widespread clinical utility. Therefore, a novel two-step sandwich chemiluminescence immunoassay (sCLIA) for accurate quantification of PAC was systematically evaluated.

Methods: Precision, trueness, linear range, and maximum dilution factor of the new immunoassay were comprehensively validated. In a multicenter study involving 2,696 samples from seven Chinese centers, PAC measurements were performed in parallel using sCLIA, cCLIA, and LC-MS/MS. The study specifically focused on evaluating the assay's performance at low aldosterone concentrations and in patients with chronic kidney disease (CKD), investigating potential interference from renal impairment by comparing the consistency between immunoassays and LC-MS/MS results across different CKD stages.

Results: The sCLIA exhibited excellent analytical performance for PAC measurement, with intra-assay imprecision <4.64 % and bias <5.71 % against certificated reference materials. The assay exhibited a wide reportable range (30-100,000 ng/L) with a limit of quantification at 30 ng/L and dilution capability ≥50-fold. Compared to cCLIA, sCLIA showed superior agreement with LC-MS/MS, particularly at low PAC concentrations (<110 ng/L) and in subjects with reduced renal function (eGFR<60 mL/min/1.73 m2).

Conclusions: This novel sCLIA method exhibited excellent analytical performance, combining the practical advantages of immunoassays with LC-MS/MS accuracy, thereby offering an ideal solution for large-scale primary aldosteronism screening in clinical practice.

目的:醛固酮的准确定量是筛选和诊断原发性醛固酮增多症(PA)的关键。与液相色谱-串联质谱法(LC-MS/MS)相比,目前的竞争性化学发光免疫测定法(cCLIA)高估了血浆醛固酮浓度(PAC)。然而,LC-MS/MS技术要求高,耗时长,限制了其广泛的临床应用。因此,系统地评估了一种新的两步夹心化学发光免疫分析法(sCLIA),用于准确定量PAC。方法:对新免疫分析法的精密度、准确度、线性范围和最大稀释系数进行综合验证。在一项涉及来自中国7个中心的2,696个样本的多中心研究中,PAC测量采用sCLIA, cCLIA和LC-MS/MS并行进行。该研究特别侧重于评估该方法在低醛固酮浓度和慢性肾脏疾病(CKD)患者中的性能,通过比较不同CKD分期免疫测定和LC-MS/MS结果之间的一致性,研究肾脏损害的潜在干扰。结果:sCLIA在PAC测定中表现出优异的分析性能,但存在分析内不精确(2)。结论:这种新型的sCLIA方法具有优异的分析性能,结合了免疫测定的实用优势和LC-MS/MS的准确性,为临床大规模原发性醛固酮增多症筛查提供了理想的解决方案。
{"title":"An innovative immunoassay for accurate aldosterone quantification: overcoming low-level inaccuracy and renal dysfunction-associated interference.","authors":"Kaijuan Wang, Hongying Cong, Zhangwei Gao, Xiaojing Gao, Wei Zhang, Xiaocui Shi, Zhou Zhou","doi":"10.1515/cclm-2025-0743","DOIUrl":"10.1515/cclm-2025-0743","url":null,"abstract":"<p><strong>Objectives: </strong>Accurate quantification of aldosterone is critical for screening and diagnosing primary aldosteronism (PA). Current competitive chemiluminescence immunoassays (cCLIA) overestimate plasma aldosterone concentration (PAC) compared to liquid chromatography-tandem mass spectrometry (LC-MS/MS). However, LC-MS/MS is technically demanding and time-consuming, limiting its widespread clinical utility. Therefore, a novel two-step sandwich chemiluminescence immunoassay (sCLIA) for accurate quantification of PAC was systematically evaluated.</p><p><strong>Methods: </strong>Precision, trueness, linear range, and maximum dilution factor of the new immunoassay were comprehensively validated. In a multicenter study involving 2,696 samples from seven Chinese centers, PAC measurements were performed in parallel using sCLIA, cCLIA, and LC-MS/MS. The study specifically focused on evaluating the assay's performance at low aldosterone concentrations and in patients with chronic kidney disease (CKD), investigating potential interference from renal impairment by comparing the consistency between immunoassays and LC-MS/MS results across different CKD stages.</p><p><strong>Results: </strong>The sCLIA exhibited excellent analytical performance for PAC measurement, with intra-assay imprecision <4.64 % and bias <5.71 % against certificated reference materials. The assay exhibited a wide reportable range (30-100,000 ng/L) with a limit of quantification at 30 ng/L and dilution capability ≥50-fold. Compared to cCLIA, sCLIA showed superior agreement with LC-MS/MS, particularly at low PAC concentrations (<110 ng/L) and in subjects with reduced renal function (eGFR<60 mL/min/1.73 m<sup>2</sup>).</p><p><strong>Conclusions: </strong>This novel sCLIA method exhibited excellent analytical performance, combining the practical advantages of immunoassays with LC-MS/MS accuracy, thereby offering an ideal solution for large-scale primary aldosteronism screening in clinical practice.</p>","PeriodicalId":10390,"journal":{"name":"Clinical chemistry and laboratory medicine","volume":" ","pages":"2477-2484"},"PeriodicalIF":3.7,"publicationDate":"2025-08-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"144945289","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Validation of a plasma GFAP immunoassay and establishment of age-related reference values: bridging analytical performance and routine implementation. 血浆GFAP免疫测定的验证和年龄相关参考值的建立:连接分析性能和常规实施。
IF 3.7 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2025-08-25 Print Date: 2025-11-25 DOI: 10.1515/cclm-2025-0870
Burak Arslan, Ulf Andreasson, Elzbieta Rembeza, Markus Axelsson, Lenka Novakova, Bjørn-Eivind Kirsebom, Tormod Fladby, Anna Dittrich, Silke Kern, Ingmar Skoog, Kaj Blennow, Henrik Zetterberg, Hlin Kvartsberg

Objectives: Glial fibrillary acidic protein (GFAP) is a well-established biomarker of astrocytic activation associated with neurodegenerative diseases, neuroinflammatory disorders, and traumatic brain injury. With increasing interest in blood-based biomarkers, the need for analytically validated assays and reliable reference intervals is critical for routine clinical implementation. This study aimed to analytically validate the MSD S-Plex® GFAP immunoassay for plasma and to establish age-stratified reference intervals in an apparently healthy population.

Methods: This study was conducted in two phases. First, key analytical validation parameters - including repeatability, intermediate precision, measurement range, interferences, and sample stability - were evaluated following Clinical and Laboratory Standards Institute (CLSI) and published protocol guidelines. Second, reference intervals were derived from 579 apparently healthy individuals aged 17-91 years using a right-sided non-parametric percentile method. Age-specific upper reference limits were calculated for three predefined age groups, and a continuous age-dependent centile model was applied.

Results: MSD S-Plex® GFAP assay demonstrated strong analytical performance, with coefficients of variation for repeatability and intermediate precision below 12 %. After accounting for the 1:2 dilution ratio, the validated measurement range was 0.425-1760 ng/L, with all calibration residuals remaining within ±15 %. GFAP concentrations were unaffected by hemolysis (p=0.85) and remained stable for up to 7 days at 4 °C and under frozen storage conditions. Age-stratified upper reference limits for plasma GFAP were established as 38 pg/mL (18-<50 years), 73 pg/mL (≥50-<70 years), and 156 pg/mL (≥70 years). Additionally, sex-related differences were observed after age 50, with females showing higher absolute GFAP levels than males. A strong positive correlation between age and plasma GFAP levels was observed (Spearman's r=0.832, p<0.0001).

Conclusions: This study demonstrates the robust analytical performance of the MSD S-Plex® GFAP assay and establishes age-related reference values for plasma GFAP. These findings support its suitability for routine clinical use and enhance its applicability in the diagnosis and monitoring of central nervous system (CNS) pathologies, such as neurodegenerative diseases, neuroinflammatory disorders, and acute brain injuries, within biomarker-supported clinical algorithms.

目的:胶质纤维酸性蛋白(GFAP)是一种公认的与神经退行性疾病、神经炎性疾病和创伤性脑损伤相关的星形胶质细胞激活的生物标志物。随着人们对基于血液的生物标志物的兴趣日益增加,对分析验证的分析方法和可靠的参考区间的需求对于常规临床实施至关重要。本研究旨在分析验证MSD S-Plex®血浆GFAP免疫测定,并在表面健康人群中建立年龄分层参考区间。方法:本研究分为两期进行。首先,根据临床和实验室标准协会(CLSI)和公布的协议指南评估关键分析验证参数——包括重复性、中间精度、测量范围、干扰和样品稳定性。其次,采用右侧非参数百分位数法,从579名17-91岁的明显健康个体中获得参考区间。计算三个预定义年龄组的年龄特异性上限,并应用连续年龄相关百分位数模型。结果:MSD S-Plex®GFAP分析具有较强的分析性能,重复性变异系数和中间精密度低于12% %。在考虑1:2稀释比后,验证的测量范围为0.425-1760 ng/L,所有校准残差保持在±15 %范围内。GFAP浓度不受溶血影响(p=0.85),在4 °C和冷冻储存条件下保持稳定长达7天。血浆GFAP的年龄分层参考上限为38 pg/mL(18)。结论:本研究证明了MSD S-Plex®GFAP检测的可靠分析性能,并建立了与年龄相关的血浆GFAP参考值。这些发现支持了其常规临床应用的适用性,并增强了其在生物标志物支持的临床算法中对中枢神经系统(CNS)病理(如神经退行性疾病、神经炎性疾病和急性脑损伤)的诊断和监测的适用性。
{"title":"Validation of a plasma GFAP immunoassay and establishment of age-related reference values: bridging analytical performance and routine implementation.","authors":"Burak Arslan, Ulf Andreasson, Elzbieta Rembeza, Markus Axelsson, Lenka Novakova, Bjørn-Eivind Kirsebom, Tormod Fladby, Anna Dittrich, Silke Kern, Ingmar Skoog, Kaj Blennow, Henrik Zetterberg, Hlin Kvartsberg","doi":"10.1515/cclm-2025-0870","DOIUrl":"10.1515/cclm-2025-0870","url":null,"abstract":"<p><strong>Objectives: </strong>Glial fibrillary acidic protein (GFAP) is a well-established biomarker of astrocytic activation associated with neurodegenerative diseases, neuroinflammatory disorders, and traumatic brain injury. With increasing interest in blood-based biomarkers, the need for analytically validated assays and reliable reference intervals is critical for routine clinical implementation. This study aimed to analytically validate the MSD S-Plex<sup>®</sup> GFAP immunoassay for plasma and to establish age-stratified reference intervals in an apparently healthy population.</p><p><strong>Methods: </strong>This study was conducted in two phases. First, key analytical validation parameters - including repeatability, intermediate precision, measurement range, interferences, and sample stability - were evaluated following Clinical and Laboratory Standards Institute (CLSI) and published protocol guidelines. Second, reference intervals were derived from 579 apparently healthy individuals aged 17-91 years using a right-sided non-parametric percentile method. Age-specific upper reference limits were calculated for three predefined age groups, and a continuous age-dependent centile model was applied.</p><p><strong>Results: </strong>MSD S-Plex<sup>®</sup> GFAP assay demonstrated strong analytical performance, with coefficients of variation for repeatability and intermediate precision below 12 %. After accounting for the 1:2 dilution ratio, the validated measurement range was 0.425-1760 ng/L, with all calibration residuals remaining within ±15 %. GFAP concentrations were unaffected by hemolysis (p=0.85) and remained stable for up to 7 days at 4 °C and under frozen storage conditions. Age-stratified upper reference limits for plasma GFAP were established as 38 pg/mL (18-<50 years), 73 pg/mL (≥50-<70 years), and 156 pg/mL (≥70 years). Additionally, sex-related differences were observed after age 50, with females showing higher absolute GFAP levels than males. A strong positive correlation between age and plasma GFAP levels was observed (Spearman's r=0.832, p<0.0001).</p><p><strong>Conclusions: </strong>This study demonstrates the robust analytical performance of the MSD S-Plex<sup>®</sup> GFAP assay and establishes age-related reference values for plasma GFAP. These findings support its suitability for routine clinical use and enhance its applicability in the diagnosis and monitoring of central nervous system (CNS) pathologies, such as neurodegenerative diseases, neuroinflammatory disorders, and acute brain injuries, within biomarker-supported clinical algorithms.</p>","PeriodicalId":10390,"journal":{"name":"Clinical chemistry and laboratory medicine","volume":" ","pages":"2522-2535"},"PeriodicalIF":3.7,"publicationDate":"2025-08-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"144944950","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Evaluation of measurement uncertainty of 11 serum proteins measured by immunoturbidimetric methods according to ISO/TS 20914: a 1-year laboratory data analysis. 根据ISO/TS 20914免疫比浊法测定11种血清蛋白的测量不确定度评价:1年实验室数据分析
IF 3.7 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2025-08-21 Print Date: 2025-11-25 DOI: 10.1515/cclm-2025-0654
Emine Feyza Yurt, Medine Alpdemir, Mehmet Şeneş

Objectives: Measurement uncertainty (MU) plays an important role in the interpretation of laboratory results, but data on serum proteins analyzed by immunoturbidimetry according to ISO/TS 20914 are limited.

Methods: MU of 11 serum proteins, including CRP, RF, ASO, IgG, IgA, IgM, C3, C4, ceruloplasmin, transferrin, and β2-microglobulin, were estimated using 1-year internal quality control (IQC) data obtained from Roche Cobas analyzers. MU was calculated using uncertainty and calibrator uncertainty according to ISO/TS 20914, assuming negligible deviation from external quality assessment data. Analytical performance specification (APS) models were selected according to the EFLM APS selection criteria, and maximum allowable uncertainty (MAU) values were determined based on sources such as EFLM models and literature.

Results: IgA and RF were the only two analytes that met the required and minimum MAU values, respectively, at both IQC levels. MU values for CRP, ceruloplasmin, transferrin, and β2-microglobulin exceeded targets at both levels. MU for C3, C4, IgG, and IgM exceeded the minimum MAU at IQC1 but remained acceptable at IQC2. MU values for ASO were calculated as 10.01 and 7.22 % but could not be evaluated due to a lack of reference data. Assay precision should be improved for CRP, IgG, IgM, ceruloplasmin, transferrin, and β2-microglobulin. Use of updated calibration materials for CRP may help reduce MU.

Conclusions: Maintaining acceptable precision over a long period remains a challenge for serum proteins analyzed by immunoturbidimetry, highlighting the need for methodological improvements and stricter quality monitoring. In this context, MU assessment is crucial.

目的:测量不确定度(MU)在实验室结果的解释中起着重要作用,但根据ISO/TS 20914免疫比浊法分析血清蛋白的数据是有限的。方法:利用罗氏Cobas分析仪1年的内部质量控制(IQC)数据,对CRP、RF、ASO、IgG、IgA、IgM、C3、C4、铜蓝蛋白、转铁蛋白、β2-微球蛋白等11种血清蛋白的MU进行估算。MU根据ISO/TS 20914使用不确定度和校准器不确定度计算,假设与外部质量评估数据的偏差可以忽略不计。根据EFLM APS选择标准选择分析性能规范(APS)模型,并根据EFLM模型和文献等来源确定最大允许不确定性(MAU)值。结果:IgA和RF是仅有的两种分析物,分别在两个IQC水平上满足要求和最低MAU值。c反应蛋白、铜蓝蛋白、转铁蛋白和β2微球蛋白的MU值在两个水平上都超过了目标值。C3、C4、IgG和IgM的MU在IQC1超过了最小MAU,但在IQC2仍然可以接受。ASO的MU值计算为10.01和7.22 %,但由于缺乏参考资料而无法评估。CRP、IgG、IgM、铜蓝蛋白、转铁蛋白和β2微球蛋白的检测精度有待提高。使用更新的CRP校准材料可能有助于减少MU。结论:免疫比浊法分析血清蛋白长期保持可接受的精度仍然是一个挑战,强调需要改进方法和更严格的质量监测。在这种情况下,MU评估是至关重要的。
{"title":"Evaluation of measurement uncertainty of 11 serum proteins measured by immunoturbidimetric methods according to ISO/TS 20914: a 1-year laboratory data analysis.","authors":"Emine Feyza Yurt, Medine Alpdemir, Mehmet Şeneş","doi":"10.1515/cclm-2025-0654","DOIUrl":"10.1515/cclm-2025-0654","url":null,"abstract":"<p><strong>Objectives: </strong>Measurement uncertainty (MU) plays an important role in the interpretation of laboratory results, but data on serum proteins analyzed by immunoturbidimetry according to ISO/TS 20914 are limited.</p><p><strong>Methods: </strong>MU of 11 serum proteins, including CRP, RF, ASO, IgG, IgA, IgM, C3, C4, ceruloplasmin, transferrin, and β2-microglobulin, were estimated using 1-year internal quality control (IQC) data obtained from Roche Cobas analyzers. MU was calculated using uncertainty and calibrator uncertainty according to ISO/TS 20914, assuming negligible deviation from external quality assessment data. Analytical performance specification (APS) models were selected according to the EFLM APS selection criteria, and maximum allowable uncertainty (MAU) values were determined based on sources such as EFLM models and literature.</p><p><strong>Results: </strong>IgA and RF were the only two analytes that met the required and minimum MAU values, respectively, at both IQC levels. MU values for CRP, ceruloplasmin, transferrin, and β2-microglobulin exceeded targets at both levels. MU for C3, C4, IgG, and IgM exceeded the minimum MAU at IQC1 but remained acceptable at IQC2. MU values for ASO were calculated as 10.01 and 7.22 % but could not be evaluated due to a lack of reference data. Assay precision should be improved for CRP, IgG, IgM, ceruloplasmin, transferrin, and β2-microglobulin. Use of updated calibration materials for CRP may help reduce MU.</p><p><strong>Conclusions: </strong>Maintaining acceptable precision over a long period remains a challenge for serum proteins analyzed by immunoturbidimetry, highlighting the need for methodological improvements and stricter quality monitoring. In this context, MU assessment is crucial.</p>","PeriodicalId":10390,"journal":{"name":"Clinical chemistry and laboratory medicine","volume":" ","pages":"2425-2432"},"PeriodicalIF":3.7,"publicationDate":"2025-08-21","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"144882283","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Reference intervals in value-based laboratory medicine: a shift from single-point measurements to metabolic variation-based models. 基于价值的检验医学的参考区间:从单点测量到基于代谢变化的模型的转变。
IF 3.7 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2025-08-18 Print Date: 2025-10-27 DOI: 10.1515/cclm-2025-0763
Abdurrahman Coskun, Mario Plebani

Laboratory data can be meaningful only when compared with reliable reference data; therefore, the estimation of reliable reference data is just as important as the accurate measurement of measurands in patient samples. Since analyte concentrations in the human body are influenced by both random variations (such as biological fluctuations) and systematic variations (such as physiological rhythms and age-related changes), the conventional model for estimating reference data - based solely on the statistical distribution of single-sample measurements from reference individuals - may not provide sufficiently reliable information for interpreting patient results. Therefore, a paradigm shift from relying solely on single-sample measurement distributions to incorporating metabolic changes observed in the human body when estimating reference intervals may enhance the clinical value of laboratory data for an effective clinical decision making and patient care. This opinion paper aims to summarize how to facilitate this transition and to identify the most suitable model for estimating reference intervals that reflect underlying metabolic dynamics.

只有与可靠的参考数据相比,实验室数据才有意义;因此,估计可靠的参考数据与准确测量患者样本中的测量值同样重要。由于人体内的分析物浓度受到随机变化(如生物波动)和系统变化(如生理节律和与年龄有关的变化)的影响,估计参考数据的传统模型——仅仅基于参考个体的单样本测量的统计分布——可能无法为解释患者结果提供足够可靠的信息。因此,在估计参考区间时,从单纯依赖单样本测量分布到结合在人体中观察到的代谢变化的范式转变,可能会提高实验室数据的临床价值,有助于有效的临床决策和患者护理。本文旨在总结如何促进这种转变,并确定最合适的模型来估计反映潜在代谢动力学的参考区间。
{"title":"Reference intervals in value-based laboratory medicine: a shift from single-point measurements to metabolic variation-based models.","authors":"Abdurrahman Coskun, Mario Plebani","doi":"10.1515/cclm-2025-0763","DOIUrl":"10.1515/cclm-2025-0763","url":null,"abstract":"<p><p>Laboratory data can be meaningful only when compared with reliable reference data; therefore, the estimation of reliable reference data is just as important as the accurate measurement of measurands in patient samples. Since analyte concentrations in the human body are influenced by both random variations (such as biological fluctuations) and systematic variations (such as physiological rhythms and age-related changes), the conventional model for estimating reference data - based solely on the statistical distribution of single-sample measurements from reference individuals - may not provide sufficiently reliable information for interpreting patient results. Therefore, a paradigm shift from relying solely on single-sample measurement distributions to incorporating metabolic changes observed in the human body when estimating reference intervals may enhance the clinical value of laboratory data for an effective clinical decision making and patient care. This opinion paper aims to summarize how to facilitate this transition and to identify the most suitable model for estimating reference intervals that reflect underlying metabolic dynamics.</p>","PeriodicalId":10390,"journal":{"name":"Clinical chemistry and laboratory medicine","volume":" ","pages":"2140-2148"},"PeriodicalIF":3.7,"publicationDate":"2025-08-18","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"144859010","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Clinical validation of a liquid chromatography single quadrupole mass spectrometry (LC-MS) method using Waters Kairos™ Amino Acid Kit reagents. 使用Waters Kairos™氨基酸试剂盒的液相色谱-单四极杆质谱(LC-MS)方法的临床验证。
IF 3.7 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2025-08-08 Print Date: 2025-10-27 DOI: 10.1515/cclm-2025-0424
Rowan Hellier, Luke Griffiths, Chandra Sundas, Annabel Rodham, Stuart J Moat

Objectives: Quantitation of plasma amino acids (AA) is critical for the diagnosis and monitoring of inherited disorders of AA metabolism. AA analysis using ion-exchange chromatography (IEC) with post-column ninhydrin derivatization is time consuming, with run times of ∼2 h, limiting sample throughput. Liquid chromatography mass-spectrometry can potentially address some of the current challenges.

Methods: Performance of components of the Waters Kairos Amino Acid Kit using liquid chromatography single quadrupole mass-spectrometry (LC-MS) following derivatization of samples with 6-aminoquinolyl-N-hydroxysuccinimidyl carbamate (AccQ•Tag™ Ultra Derivatization Reagent) was evaluated. Results were compared with the Biochrom-IEC method using patient specimens (n=115), ClinChek® control and external quality assessment (EQA) material.

Results: The kit reagents and our developed method had a 19-min analysis time, demonstrated acceptable inter-assay imprecision (CV<10 %) and bias vs. IEC-method (overall mean bias <2 %). Excellent correlation (concordance correlation coefficient (CCC) >0.99) with IEC was demonstrated for 10/23 analytes, good correlation (CCC >0.95) for 10/23, with the remaining three amino acids (aspartate, histidine and tryptophan) demonstrating moderate concordance (CCC ≥0.90 but <0.95). 1/23 AAs had a mean bias >10 % using EQA material. The method demonstrated a lower limit of quantitation of ≤2.5 μmol/L for all AA, making this assay suitable for CSF analysis. Calibration stability bias was <5 % over 12-weeks. Derivatized AAs were stable for ≤17 days. The analytical column supplied demonstrated good retention time stability (<0.4 %) and was capable of >2000 injections.

Conclusions: The tested methodology demonstrated good analytical performance and correlation with IEC. This approach confers practical advantages over IEC, including analytical selectivity and workflow time efficiency.

目的:血浆氨基酸(AA)的定量测定对遗传性AA代谢疾病的诊断和监测至关重要。使用柱后茚三酮衍生的离子交换色谱(IEC)进行AA分析非常耗时,运行时间为~ 2 h,限制了样品通量。液相色谱质谱法可以潜在地解决当前的一些挑战。方法:用6-氨基喹啉- n -羟基琥珀酰氨基甲酸酯(AccQ•Tag™超衍生试剂)对样品进行衍生化后,采用液相色谱-单四极杆质谱(LC-MS)对Waters Kairos氨基酸试剂盒的组分进行性能评估。使用患者标本(n=115)、ClinChek®对照和外部质量评价(EQA)材料与Biochrom-IEC方法进行比较。结果:试剂盒试剂和我们开发的方法具有19 min的分析时间,对10/23的分析物具有可接受的分析间不精确(CV0.99),对10/23的分析物具有良好的相关性(CCC >0.95),其余三个氨基酸(天冬氨酸、组氨酸和色氨酸)具有中等一致性(CCC≥0.90,但使用EQA材料时为10 %)。所有AA的定量下限≤2.5 μmol/L,适用于脑脊液分析。校准稳定性偏差为2000注射剂。结论:测试方法具有良好的分析性能和与IEC的相关性。与IEC相比,这种方法具有实际优势,包括分析选择性和工作流程时间效率。
{"title":"Clinical validation of a liquid chromatography single quadrupole mass spectrometry (LC-MS) method using Waters Kairos™ Amino Acid Kit reagents.","authors":"Rowan Hellier, Luke Griffiths, Chandra Sundas, Annabel Rodham, Stuart J Moat","doi":"10.1515/cclm-2025-0424","DOIUrl":"10.1515/cclm-2025-0424","url":null,"abstract":"<p><strong>Objectives: </strong>Quantitation of plasma amino acids (AA) is critical for the diagnosis and monitoring of inherited disorders of AA metabolism. AA analysis using ion-exchange chromatography (IEC) with post-column ninhydrin derivatization is time consuming, with run times of ∼2 h, limiting sample throughput. Liquid chromatography mass-spectrometry can potentially address some of the current challenges.</p><p><strong>Methods: </strong>Performance of components of the Waters Kairos Amino Acid Kit using liquid chromatography single quadrupole mass-spectrometry (LC-MS) following derivatization of samples with 6-aminoquinolyl-N-hydroxysuccinimidyl carbamate (AccQ•Tag™ Ultra Derivatization Reagent) was evaluated. Results were compared with the Biochrom-IEC method using patient specimens (n=115), ClinChek<sup>®</sup> control and external quality assessment (EQA) material.</p><p><strong>Results: </strong>The kit reagents and our developed method had a 19-min analysis time, demonstrated acceptable inter-assay imprecision (CV<10 %) and bias vs. IEC-method (overall mean bias <2 %). Excellent correlation (concordance correlation coefficient (CCC) >0.99) with IEC was demonstrated for 10/23 analytes, good correlation (CCC >0.95) for 10/23, with the remaining three amino acids (aspartate, histidine and tryptophan) demonstrating moderate concordance (CCC ≥0.90 but <0.95). 1/23 AAs had a mean bias >10 % using EQA material. The method demonstrated a lower limit of quantitation of ≤2.5 μmol/L for all AA, making this assay suitable for CSF analysis. Calibration stability bias was <5 % over 12-weeks. Derivatized AAs were stable for ≤17 days. The analytical column supplied demonstrated good retention time stability (<0.4 %) and was capable of >2000 injections.</p><p><strong>Conclusions: </strong>The tested methodology demonstrated good analytical performance and correlation with IEC. This approach confers practical advantages over IEC, including analytical selectivity and workflow time efficiency.</p>","PeriodicalId":10390,"journal":{"name":"Clinical chemistry and laboratory medicine","volume":" ","pages":"2226-2235"},"PeriodicalIF":3.7,"publicationDate":"2025-08-08","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"144798346","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
What Matters Most: an Age-Friendly approach to pathology and laboratory medicine. 最重要的是:对病理学和检验医学采取一种对老年人友好的方法。
IF 3.7 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2025-07-14 Print Date: 2025-10-27 DOI: 10.1515/cclm-2025-0616
Melody Boudreaux Nelson, Candice Coffey

Under the Age-Friendly Health System initiative, "What Matters" is defined as knowing the older adult's specific health outcome goals and care preferences. This involves multiple settings of care (e.g., hospital, skilled nursing facility, outpatient visits) and can include end-of-life care. However, the establishment of testing frequency criterion and the development of wide-scale diagnostic algorithms are often left undefined in older adults with poor prognosis and/or shortened life expectancy. Thus, multidisciplinary development of Geriatric 5M-informed optimization plans at the institution level and quality improvement strategies within the laboratory community may further the successful implementation of age-friendly efforts. While patients, end-users, and systems can attribute to implementation barriers, the development of an evidence-base wherein laboratory expertise is directly associated with patient outcomes is vital. Thus, a concentrated, cooperative age-friendly approach centered on What Matters may present a sustainable strategy for early transformation. Future research centered on piloted interventions on the laboratory's role in older adult care and end of life diagnostic management is needed.

在老年人友好型卫生系统倡议下,“什么重要”被定义为了解老年人的具体健康结果目标和护理偏好。这涉及多种护理环境(例如,医院、专业护理机构、门诊),并可包括临终关怀。然而,在预后不良和/或预期寿命缩短的老年人中,检测频率标准的建立和大规模诊断算法的发展往往不明确。因此,在机构层面多学科发展老年5m信息优化计划和实验室社区的质量改进策略可能会进一步成功实施老年友好工作。虽然患者、最终用户和系统可归因于实施障碍,但建立一个证据基础,其中实验室专业知识与患者预后直接相关,这一点至关重要。因此,以“重要的事情”为中心的集中的、合作的、对老年人友好的方法可能是早期转型的可持续战略。未来的研究需要集中在实验室在老年人护理和临终诊断管理中的作用的试点干预措施上。
{"title":"What Matters Most: an Age-Friendly approach to pathology and laboratory medicine.","authors":"Melody Boudreaux Nelson, Candice Coffey","doi":"10.1515/cclm-2025-0616","DOIUrl":"10.1515/cclm-2025-0616","url":null,"abstract":"<p><p>Under the Age-Friendly Health System initiative, \"What Matters\" is defined as knowing the older adult's specific health outcome goals and care preferences. This involves multiple settings of care (e.g., hospital, skilled nursing facility, outpatient visits) and can include end-of-life care. However, the establishment of testing frequency criterion and the development of wide-scale diagnostic algorithms are often left undefined in older adults with poor prognosis and/or shortened life expectancy. Thus, multidisciplinary development of Geriatric 5M-informed optimization plans at the institution level and quality improvement strategies within the laboratory community may further the successful implementation of age-friendly efforts. While patients, end-users, and systems can attribute to implementation barriers, the development of an evidence-base wherein laboratory expertise is directly associated with patient outcomes is vital. Thus, a concentrated, cooperative age-friendly approach centered on What Matters may present a sustainable strategy for early transformation. Future research centered on piloted interventions on the laboratory's role in older adult care and end of life diagnostic management is needed.</p>","PeriodicalId":10390,"journal":{"name":"Clinical chemistry and laboratory medicine","volume":" ","pages":"2163-2170"},"PeriodicalIF":3.7,"publicationDate":"2025-07-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"144612123","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
No fault or negligence after an adverse analytical finding due to a contaminated supplement: mission impossible. Two examples involving trimetazidine. 没有过错或疏忽后,不利的分析发现,由于污染的补充:不可能的任务。两个涉及曲美他嗪的例子。
IF 3.7 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2025-05-30 Print Date: 2025-10-27 DOI: 10.1515/cclm-2025-0549
Pascal Kintz

Since several years, sports authorities, including national anti-doping organisations and the World Anti-Doping Agency (WADA) have published that the consumption of food supplements can be at risk for athletes due to potential contamination by prohibited substances. Despite these warnings, supplements are largely used by elite athletes and doping violations involving supplements are weekly reported in the media. Anabolic steroids, selective androgen receptor modulators (SARMs), metabolic modulators, stimulants and diuretics are among the most frequently detected substances in contaminated supplements. The author was involved in 2 cases where trimetazidine was identified as the source of the doping violation but the sport authorities sentenced both athletes. Case 1: trimetazidine in urine at 0.1 ng/mL; trimetazidine negative in 3 × 2 cm hair segments (LOQ at 1 ng/g [pg/mg]); trimetazidine at 4 ng/tablet in the supplement. Case 2: trimetazidine in urine at 0.1 and 1.6 ng/mL on 2 occasions; trimetazidine negative in 6 × 1 cm hair segments (LOQ at 1 ng/g [pg/mg]); trimetazidine at 7.2 μg/tablet in the supplement. Despite all pharmacological parameters were consistent with minute amounts of trimetazidine intake during a scenario of proven contamination, the presence of trimetazidine in the urine samples was recognized as an anti-doping rule violation and the athletes were sanctioned with a period of ineligibility of 6 months. From a forensic perspective, contamination is not doping. To avoid these conflicting issues with contaminations, the debate should move to the interest of using hair test results and the need of implementing a threshold for reporting the presence of a drug in urine at very low concentration.

几年来,包括国家反兴奋剂组织和世界反兴奋剂机构(WADA)在内的体育当局已经发布报告称,由于潜在的违禁物质污染,运动员食用食品补充剂可能存在风险。尽管有这些警告,补品在很大程度上被精英运动员使用,媒体每周都会报道涉及补品的兴奋剂违规行为。合成代谢类固醇、选择性雄激素受体调节剂(SARMs)、代谢调节剂、兴奋剂和利尿剂是受污染补充剂中最常检测到的物质。提交人参与了两起案件,其中曲美他嗪被确定为违反兴奋剂规定的来源,但体育当局对两名运动员都判刑。病例1:尿中曲美他嗪0.1 ng/mL;3 × 2 cm毛节曲美他嗪阴性(定量限为1 ng/g [pg/mg]);曲美他嗪4 ng/片。病例2:曲美他嗪2次尿中0.1、1.6 ng/mL;6 × 1 cm毛节曲美他嗪阴性(定量限为1 ng/g [pg/mg]);曲美他嗪7.2 μg/片。尽管所有药理学参数都与在已证实的污染情况下摄入微量曲美他嗪一致,但尿样中存在曲美他嗪被认定为违反了反兴奋剂规则,运动员被处以6个 月的禁赛期。从法医的角度来看,污染不是兴奋剂。为了避免这些与污染相冲突的问题,辩论应该转向使用毛发测试结果的兴趣,以及实施报告极低浓度尿液中药物存在的阈值的必要性。
{"title":"No fault or negligence after an adverse analytical finding due to a contaminated supplement: mission impossible. Two examples involving trimetazidine.","authors":"Pascal Kintz","doi":"10.1515/cclm-2025-0549","DOIUrl":"10.1515/cclm-2025-0549","url":null,"abstract":"<p><p>Since several years, sports authorities, including national anti-doping organisations and the World Anti-Doping Agency (WADA) have published that the consumption of food supplements can be at risk for athletes due to potential contamination by prohibited substances. Despite these warnings, supplements are largely used by elite athletes and doping violations involving supplements are weekly reported in the media. Anabolic steroids, selective androgen receptor modulators (SARMs), metabolic modulators, stimulants and diuretics are among the most frequently detected substances in contaminated supplements. The author was involved in 2 cases where trimetazidine was identified as the source of the doping violation but the sport authorities sentenced both athletes. Case 1: trimetazidine in urine at 0.1 ng/mL; trimetazidine negative in 3 × 2 cm hair segments (LOQ at 1 ng/g [pg/mg]); trimetazidine at 4 ng/tablet in the supplement. Case 2: trimetazidine in urine at 0.1 and 1.6 ng/mL on 2 occasions; trimetazidine negative in 6 × 1 cm hair segments (LOQ at 1 ng/g [pg/mg]); trimetazidine at 7.2 μg/tablet in the supplement. Despite all pharmacological parameters were consistent with minute amounts of trimetazidine intake during a scenario of proven contamination, the presence of trimetazidine in the urine samples was recognized as an anti-doping rule violation and the athletes were sanctioned with a period of ineligibility of 6 months. From a forensic perspective, contamination is not doping. To avoid these conflicting issues with contaminations, the debate should move to the interest of using hair test results and the need of implementing a threshold for reporting the presence of a drug in urine at very low concentration.</p>","PeriodicalId":10390,"journal":{"name":"Clinical chemistry and laboratory medicine","volume":" ","pages":"2171-2176"},"PeriodicalIF":3.7,"publicationDate":"2025-05-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"144207845","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
The accuracy of presepsin in diagnosing neonatal late-onset sepsis in critically ill neonates: a prospective study. presepsin在诊断危重新生儿迟发性脓毒症中的准确性:一项前瞻性研究。
IF 3.7 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2025-04-21 Print Date: 2025-08-26 DOI: 10.1515/cclm-2025-0128
Cinzia Auriti, Domenico Umberto De Rose, Chiara Maddaloni, Lucilla Ravà, Ludovica Martini, Eleonora Di Tommaso, Paola Bernaschi, Emanuel Paionni, Ottavia Porzio, Fiammetta Piersigilli, Marco Iannetta, Andrea Dotta, Maria Paola Ronchetti

Objectives: The diagnostic accuracy of presepsin (P-SEP) in the newborn is still under evaluation.

Methods: In a multicenter study, we studied the accuracy of P-SEP as a diagnostic marker of late-onset sepsis (LOS) in critical newborns with underlying disorders, to define the most accurate cut-off to distinguish infected from uninfected patients.

Results: Sixty-nine/351 newborns without infections at admission developed LOS. The median P-SEP value at T0 (admission) was 518.0 ng/L (IQR 313.0-789.0), without significant differences related to underlying diseases (p=0.52). In neonates who developed LOS, P-SEP increased at the onset of infection (T1) (median: 816.0 ng/L) and after 24-48 h (median: 901.0 ng/L) compared with their value at admission (median: 560.0 ng/L) (p<0.01 and p=0.03, respectively). The area under the ROC curve at T1 was 0.71 (95 % CI 0.65-0.78) when all cases of sepsis were included in the analysis and increased to 0.74 (95 % CI 0.66-0.81) considering only confirmed sepsis. Approximately two-thirds of patients were correctly classified, setting the cut-off at 713 ng/L, with a negative predictive value of 89.0 %.

Conclusions: At a cut-off of 713 ng/L, P-SEP has good accuracy in diagnosing LOS in critically ill newborns. In uninfected newborns, the median value of P-SEP is not influenced by any underlying pathology.

目的:产前压血素(P-SEP)在新生儿诊断中的准确性仍有待评估。方法:在一项多中心研究中,我们研究了P-SEP作为有潜在疾病的危重新生儿迟发性脓毒症(LOS)诊断标志物的准确性,以确定区分感染和未感染患者的最准确截止值。结果:入院时无感染的新生儿有69 /351发生LOS。入院时p - sep值中位数为518.0 ng/L (IQR为313.0-789.0),与基础疾病无显著性差异(p=0.52)。在发生LOS的新生儿中,P-SEP在感染开始时(T1)(中位数:816.0 ng/L)和24-48 h(中位数:901.0 ng/L)较入院时(中位数:560.0 ng/L)升高(结论:P-SEP在临界值为713 ng/L时,对诊断危重新生儿LOS具有良好的准确性。在未感染的新生儿中,P-SEP的中位数不受任何潜在病理的影响。
{"title":"The accuracy of presepsin in diagnosing neonatal late-onset sepsis in critically ill neonates: a prospective study.","authors":"Cinzia Auriti, Domenico Umberto De Rose, Chiara Maddaloni, Lucilla Ravà, Ludovica Martini, Eleonora Di Tommaso, Paola Bernaschi, Emanuel Paionni, Ottavia Porzio, Fiammetta Piersigilli, Marco Iannetta, Andrea Dotta, Maria Paola Ronchetti","doi":"10.1515/cclm-2025-0128","DOIUrl":"10.1515/cclm-2025-0128","url":null,"abstract":"<p><strong>Objectives: </strong>The diagnostic accuracy of presepsin (P-SEP) in the newborn is still under evaluation.</p><p><strong>Methods: </strong>In a multicenter study, we studied the accuracy of P-SEP as a diagnostic marker of late-onset sepsis (LOS) in critical newborns with underlying disorders, to define the most accurate cut-off to distinguish infected from uninfected patients.</p><p><strong>Results: </strong>Sixty-nine/351 newborns without infections at admission developed LOS. The median P-SEP value at T0 (admission) was 518.0 ng/L (IQR 313.0-789.0), without significant differences related to underlying diseases (p=0.52). In neonates who developed LOS, P-SEP increased at the onset of infection (T1) (median: 816.0 ng/L) and after 24-48 h (median: 901.0 ng/L) compared with their value at admission (median: 560.0 ng/L) (p<0.01 and p=0.03, respectively). The area under the ROC curve at T1 was 0.71 (95 % CI 0.65-0.78) when all cases of sepsis were included in the analysis and increased to 0.74 (95 % CI 0.66-0.81) considering only confirmed sepsis. Approximately two-thirds of patients were correctly classified, setting the cut-off at 713 ng/L, with a negative predictive value of 89.0 %.</p><p><strong>Conclusions: </strong>At a cut-off of 713 ng/L, P-SEP has good accuracy in diagnosing LOS in critically ill newborns. In uninfected newborns, the median value of P-SEP is not influenced by any underlying pathology.</p>","PeriodicalId":10390,"journal":{"name":"Clinical chemistry and laboratory medicine","volume":" ","pages":"1876-1887"},"PeriodicalIF":3.7,"publicationDate":"2025-04-21","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"143962217","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Assessment of serum free light chain measurements in a large Chinese chronic kidney disease cohort: a multicenter real-world study. 中国慢性肾脏疾病队列中血清游离轻链测量的评估:一项多中心真实世界研究
IF 3.7 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2025-04-14 Print Date: 2025-08-26 DOI: 10.1515/cclm-2024-1226
Xia Luo, Xiaomeng Zhang, Xu Yuan, Pan Zhao, Wenqian Zhang, Lingyan Deng, Liming Cheng

Objectives: Diagnosing monoclonal gammopathy in chronic kidney disease (CKD) patients is challenging due to the complex interpretation of serum free light chain (FLC) levels. This study aimed to assess the FLC levels in a large cohort of Chinese CKD patients.

Methods: We enrolled 5,287 patients with all-cause nondialysis CKD from three medical centers between February 2018 and April 2023. Central 95 % reference ranges were established using data from one center and validated in an independent subpopulation from the other two centers. The crude prevalence of light chain monoclonal gammopathy of undetermined significance (LC-MGUS) was assessed against established norms for the entire cohort and subgroups based on estimated glomerular filtration rate (eGFR).

Results: A notable proportion of patients exceeded the standard reference limits for kappa (89.0 %), lambda (72.1 %), and FLC ratio (10.5 %), whereas non-eGFR-based renal reference interval identified only 0.3 % with abnormal FLC ratios. The iStopMM reference ranges showed higher out-of-range rates for absolute FLC levels in patients with preserved kidney function and lower rates in those with impaired kidney function, while the FLC ratio references remained robust across all groups. The crude prevalence of LC-MGUS was 10.4 % using standard ranges, predominantly kappa LC-MGUS (99.8 %). This prevalence decreased to 0.3 % with non-eGFR-based renal reference interval and 0.5 % with the iStopMM ranges. Using the newly established reference ranges, the crude prevalence was found to be 0.9 %.

Conclusions: Our findings suggest that current FLC reference ranges are inadequate for the Chinese population, underscoring the need for eGFR-based reference ranges tailored to this demographic.

目的:由于血清游离轻链(FLC)水平的复杂解释,慢性肾脏疾病(CKD)患者单克隆伽玛病的诊断具有挑战性。本研究旨在评估中国CKD患者的FLC水平。方法:我们在2018年2月至2023年4月期间从三个医疗中心招募了5287例全因非透析性CKD患者。中心95 %参考范围使用来自一个中心的数据建立,并在来自另外两个中心的独立亚人群中进行验证。根据估计的肾小球滤过率(eGFR),对整个队列和亚组的未确定意义的轻链单克隆伽玛病(LC-MGUS)的粗患病率进行评估。结果:kappa(89.0 %)、lambda(72.1 %)和FLC比率(10.5 %)超过标准参考限度的患者比例显著,而非基于egfr的肾脏参考区间仅鉴定出0.3 %的FLC比率异常。iStopMM参考范围显示,保留肾功能的患者绝对FLC水平的超范围率较高,而肾功能受损的患者的超范围率较低,而FLC比率参考在所有组中都保持稳健。按标准范围计算,LC-MGUS的粗患病率为10.4% %,主要是kappa LC-MGUS(99.8% %)。在非egfr为基础的肾参考区间,患病率降至0.3 %,在iStopMM范围内,患病率降至0.5 %。使用新建立的参考范围,发现粗患病率为0.9 %。结论:我们的研究结果表明,目前的FLC参考范围不适合中国人群,强调需要针对这一人群量身定制基于egfr的参考范围。
{"title":"Assessment of serum free light chain measurements in a large Chinese chronic kidney disease cohort: a multicenter real-world study.","authors":"Xia Luo, Xiaomeng Zhang, Xu Yuan, Pan Zhao, Wenqian Zhang, Lingyan Deng, Liming Cheng","doi":"10.1515/cclm-2024-1226","DOIUrl":"10.1515/cclm-2024-1226","url":null,"abstract":"<p><strong>Objectives: </strong>Diagnosing monoclonal gammopathy in chronic kidney disease (CKD) patients is challenging due to the complex interpretation of serum free light chain (FLC) levels. This study aimed to assess the FLC levels in a large cohort of Chinese CKD patients.</p><p><strong>Methods: </strong>We enrolled 5,287 patients with all-cause nondialysis CKD from three medical centers between February 2018 and April 2023. Central 95 % reference ranges were established using data from one center and validated in an independent subpopulation from the other two centers. The crude prevalence of light chain monoclonal gammopathy of undetermined significance (LC-MGUS) was assessed against established norms for the entire cohort and subgroups based on estimated glomerular filtration rate (eGFR).</p><p><strong>Results: </strong>A notable proportion of patients exceeded the standard reference limits for kappa (89.0 %), lambda (72.1 %), and FLC ratio (10.5 %), whereas non-eGFR-based renal reference interval identified only 0.3 % with abnormal FLC ratios. The iStopMM reference ranges showed higher out-of-range rates for absolute FLC levels in patients with preserved kidney function and lower rates in those with impaired kidney function, while the FLC ratio references remained robust across all groups. The crude prevalence of LC-MGUS was 10.4 % using standard ranges, predominantly kappa LC-MGUS (99.8 %). This prevalence decreased to 0.3 % with non-eGFR-based renal reference interval and 0.5 % with the iStopMM ranges. Using the newly established reference ranges, the crude prevalence was found to be 0.9 %.</p><p><strong>Conclusions: </strong>Our findings suggest that current FLC reference ranges are inadequate for the Chinese population, underscoring the need for eGFR-based reference ranges tailored to this demographic.</p>","PeriodicalId":10390,"journal":{"name":"Clinical chemistry and laboratory medicine","volume":" ","pages":"1757-1765"},"PeriodicalIF":3.7,"publicationDate":"2025-04-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"143976879","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
期刊
Clinical chemistry and laboratory medicine
全部 Acc. Chem. Res. ACS Applied Bio Materials ACS Appl. Electron. Mater. ACS Appl. Energy Mater. ACS Appl. Mater. Interfaces ACS Appl. Nano Mater. ACS Appl. Polym. Mater. ACS BIOMATER-SCI ENG ACS Catal. ACS Cent. Sci. ACS Chem. Biol. ACS Chemical Health & Safety ACS Chem. Neurosci. ACS Comb. Sci. ACS Earth Space Chem. ACS Energy Lett. ACS Infect. Dis. ACS Macro Lett. ACS Mater. Lett. ACS Med. Chem. Lett. ACS Nano ACS Omega ACS Photonics ACS Sens. ACS Sustainable Chem. Eng. ACS Synth. Biol. Anal. Chem. BIOCHEMISTRY-US Bioconjugate Chem. BIOMACROMOLECULES Chem. Res. Toxicol. Chem. Rev. Chem. Mater. CRYST GROWTH DES ENERG FUEL Environ. Sci. Technol. Environ. Sci. Technol. Lett. Eur. J. Inorg. Chem. IND ENG CHEM RES Inorg. Chem. J. Agric. Food. Chem. J. Chem. Eng. Data J. Chem. Educ. J. Chem. Inf. Model. J. Chem. Theory Comput. J. Med. Chem. J. Nat. Prod. J PROTEOME RES J. Am. Chem. Soc. LANGMUIR MACROMOLECULES Mol. Pharmaceutics Nano Lett. Org. Lett. ORG PROCESS RES DEV ORGANOMETALLICS J. Org. Chem. J. Phys. Chem. J. Phys. Chem. A J. Phys. Chem. B J. Phys. Chem. C J. Phys. Chem. Lett. Analyst Anal. Methods Biomater. Sci. Catal. Sci. Technol. Chem. Commun. Chem. Soc. Rev. CHEM EDUC RES PRACT CRYSTENGCOMM Dalton Trans. Energy Environ. Sci. ENVIRON SCI-NANO ENVIRON SCI-PROC IMP ENVIRON SCI-WAT RES Faraday Discuss. Food Funct. Green Chem. Inorg. Chem. Front. Integr. Biol. J. Anal. At. Spectrom. J. Mater. Chem. A J. Mater. Chem. B J. Mater. Chem. C Lab Chip Mater. Chem. Front. Mater. Horiz. MEDCHEMCOMM Metallomics Mol. Biosyst. Mol. Syst. Des. Eng. Nanoscale Nanoscale Horiz. Nat. Prod. Rep. New J. Chem. Org. Biomol. Chem. Org. Chem. Front. PHOTOCH PHOTOBIO SCI PCCP Polym. Chem.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1