首页 > 最新文献

Clinical chemistry最新文献

英文 中文
A-302 Evaluation of real-time PCR assays for detection of sexually-transmitted infections A-302 评估用于检测性传播感染的实时 PCR 检测方法
IF 9.3 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2024-10-02 DOI: 10.1093/clinchem/hvae106.299
H Soong
Background Sexually transmitted infections (STIs) have serious negative consequences for reproductive health worldwide. They are associated with infertility, premature birth and neonatal infections. Five STI pathogens, namely Trichomonas vaginalis, Mycoplasma genitalium, Mycoplasma hominis, Ureaplasma parvum, and Ureaplasma urealyticum were selected for the evaluation of three potential real-time PCR assays that can be adopted for diagnostic testing. Methods A total of 36 samples were included, consisting of genital swabs, urine, abscesses and samples provided by quality assurance programmes. DNA extraction was performed using an automated nucleic acid extraction system, abGenixTM. Commercial genomic controls and DNA extracts donated by external laboratories were used as well. The extracted nucleic acids were tested using three different assays, namely Seegene Anyplex II STI-5 Detection, Thermo Fisher Scientific customized STI Panel assay, and TIB MOLBIOL STI LightMix Modular kits, according to the manufacturers’ instructions. Results A total of 67% and 11% of the samples showed equivalent positive and negative results, respectively. However, 22% of the samples had non-concordant results. TIB MOLBIOL STI LightMix Modular kit was unable to differentiate between Ureaplasma parvum and Ureaplasma urealyticum in 9 samples. Genomic controls were tested, and Seegene Anyplex II STI-5 Detection was unable to detect lower concentrations of DNA for Trichomonas vaginalis, Mycoplasma genitalium and Mycoplasma hominis. All assays were able to detect lower concentrations of Ureaplasma parvum DNA, but detectable concentrations were higher for Ureaplasma urealyticum. Conclusions In conclusion, the performance of each assay differed according to pathogen. None of the assays offered equal performance for all pathogens. We have adopted Thermo Fisher Scientific customized STI Panel assay in our diagnostic testing due to its advantage of being user friendly and cost effective.
背景性传播感染(STI)对全世界的生殖健康都有严重的负面影响。它们与不孕、早产和新生儿感染有关。研究人员选择了五种性传播感染病原体,即阴道毛滴虫、生殖器支原体、人型支原体、副脲原体和尿解支原体,对可用于诊断检测的三种潜在实时 PCR 检测方法进行评估。方法 共纳入 36 份样本,包括生殖器拭子、尿液、脓肿和质量保证计划提供的样本。DNA 提取使用自动核酸提取系统 abGenixTM 进行。此外,还使用了商业基因组对照和外部实验室捐赠的 DNA 提取物。提取的核酸按照制造商的说明使用三种不同的检测方法进行检测,即Seegene Anyplex II STI-5检测法、Thermo Fisher Scientific定制的STI Panel检测法和TIB MOLBIOL STI LightMix Modular试剂盒。结果 分别有 67% 和 11% 的样本显示出相同的阳性和阴性结果。但有 22% 的样本结果不一致。在 9 份样本中,TIB MOLBIOL STI LightMix Modular 试剂盒无法区分副脲原体和尿解脲原体。对基因组对照进行了检测,Seegene Anyplex II STI-5 检测试剂盒无法检测到较低浓度的阴道毛滴虫、生殖支原体和人型支原体 DNA。所有检测方法都能检测到较低浓度的副脲原体 DNA,但尿解支原体的可检测浓度较高。结论 总之,病原体不同,每种检测方法的性能也不同。没有一种检测方法能对所有病原体提供相同的性能。由于赛默飞世尔科技公司定制的性传播感染检测板具有用户友好和成本效益高的优点,因此我们在诊断检测中采用了该检测板。
{"title":"A-302 Evaluation of real-time PCR assays for detection of sexually-transmitted infections","authors":"H Soong","doi":"10.1093/clinchem/hvae106.299","DOIUrl":"https://doi.org/10.1093/clinchem/hvae106.299","url":null,"abstract":"Background Sexually transmitted infections (STIs) have serious negative consequences for reproductive health worldwide. They are associated with infertility, premature birth and neonatal infections. Five STI pathogens, namely Trichomonas vaginalis, Mycoplasma genitalium, Mycoplasma hominis, Ureaplasma parvum, and Ureaplasma urealyticum were selected for the evaluation of three potential real-time PCR assays that can be adopted for diagnostic testing. Methods A total of 36 samples were included, consisting of genital swabs, urine, abscesses and samples provided by quality assurance programmes. DNA extraction was performed using an automated nucleic acid extraction system, abGenixTM. Commercial genomic controls and DNA extracts donated by external laboratories were used as well. The extracted nucleic acids were tested using three different assays, namely Seegene Anyplex II STI-5 Detection, Thermo Fisher Scientific customized STI Panel assay, and TIB MOLBIOL STI LightMix Modular kits, according to the manufacturers’ instructions. Results A total of 67% and 11% of the samples showed equivalent positive and negative results, respectively. However, 22% of the samples had non-concordant results. TIB MOLBIOL STI LightMix Modular kit was unable to differentiate between Ureaplasma parvum and Ureaplasma urealyticum in 9 samples. Genomic controls were tested, and Seegene Anyplex II STI-5 Detection was unable to detect lower concentrations of DNA for Trichomonas vaginalis, Mycoplasma genitalium and Mycoplasma hominis. All assays were able to detect lower concentrations of Ureaplasma parvum DNA, but detectable concentrations were higher for Ureaplasma urealyticum. Conclusions In conclusion, the performance of each assay differed according to pathogen. None of the assays offered equal performance for all pathogens. We have adopted Thermo Fisher Scientific customized STI Panel assay in our diagnostic testing due to its advantage of being user friendly and cost effective.","PeriodicalId":10690,"journal":{"name":"Clinical chemistry","volume":"23 1","pages":""},"PeriodicalIF":9.3,"publicationDate":"2024-10-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"142369088","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
B-244 Analytical Performance and Method Comparison Evaluation of a New High Throughput Fully Automated Plasma GFAP Immunoassay B-244 新型高通量全自动血浆 GFAP 免疫测定的分析性能和方法比较评估
IF 9.3 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2024-10-02 DOI: 10.1093/clinchem/hvae106.601
Z Vucetic, M Szabo, M Salvati, B Schlichtmann, J Patzlaff, D Unruh, K Curtis, L Mediger, M Nichkova-Doseva
Background Glial Fibrillary Acidic Protein (GFAP) levels are shown to be an indicator of neurologic injury in conditions like Traumatic Brain Injury (TBI) and stroke and as a marker of disease progression in neuromuscular disorders such as multiple sclerosis. Plasma GFAP is also implicated in detecting AD pathology, correlating to the clinical stage of the disease. The performance characteristics of the plasma GFAP immunoassay currently being developed on Beckman Coulter Access 2 and DxI 9000 analyzers are described. Methods The prototype GFAP assay is a one-step sandwich assay utilizing an anti-GFAP mouse monoclonal (MAb) antibody/alkaline phosphatase conjugate and paramagnetic particles coated with a complementary anti-GFAP mouse MAb. Sample and reactants are incubated and washed, and then a chemiluminescent substrate is added. The light generated is directly proportional to the GFAP concentration in the sample. The assay time to the first result is ∼30 minutes. Analytical performance evaluation and method comparison across different platforms was performed. 43 EDTA plasma samples, consisting of 19 Alzheimer’s Disease (AD) patient samples and 24 age-matched healthy normal samples, were evaluated. Results The prototype GFAP assay has a targeted analytical measuring range of 1.0 pg/ml to approximately 700.0 pg/ml (up to 10,000 pg/ml). All normal samples were quantified with 90% measuring below 20.0 pg/ml and a low dose coefficient of variation <4.0%. A comparison of the prototype GFAP assay on DxI 9000 and Access 2 analyzers shows a Passing-Bablok slope of 0.94 (R=0.992) and an intercept of 0.37 pg/ml. Concordance analysis shows excellent agreement between the Access 2 and DxI 9000 assays. The dose ratio of the median of AD over the median of normal samples for Access 2 and DxI 9000 is 2.31 and 2.41, respectively, with AD patient median dose values of 14.3 pg/ml and 14.6 pg/mL, respectively, and normal sample median dose values of 6.2 pg/ml and 6.0 pg/mL, respectively. Conclusions The prototype GFAP assay provides fast, highly sensitive, and precise results in an automated immunoassay on the Beckman Coulter Access 2 and DxI 9000 platforms. The prototype plasma GFAP assay holds promise for diagnosing and monitoring various neurodegenerative diseases, including AD.
背景 胶质纤维酸性蛋白(GFAP)水平是创伤性脑损伤(TBI)和中风等情况下神经系统损伤的指标,也是多发性硬化症等神经肌肉疾病进展的标志物。血浆 GFAP 还可用于检测注意力缺失症的病理变化,并与该疾病的临床阶段相关联。本文介绍了目前正在贝克曼库尔特 Access 2 和 DxI 9000 分析仪上开发的血浆 GFAP 免疫测定的性能特点。方法 GFAP 检测原型是一种一步式夹心检测法,利用抗 GFAP 小鼠单克隆 (MAb) 抗体/碱性磷酸酶共轭物和涂有互补抗 GFAP 小鼠 MAb 的顺磁颗粒。样品和反应物经过孵育和洗涤,然后加入化学发光底物。产生的光与样品中的 GFAP 浓度成正比。检测时间为 30 分钟。对不同平台进行了分析性能评估和方法比较。对 43 份乙二胺四乙酸(EDTA)血浆样本进行了评估,其中包括 19 份阿尔茨海默病(AD)患者样本和 24 份年龄匹配的健康正常样本。结果 GFAP 检测原型的目标分析测量范围为 1.0 pg/ml 至约 700.0 pg/ml(高达 10,000 pg/ml)。所有正常样本均可定量,其中 90% 的测量值低于 20.0 pg/ml,低剂量变异系数<4.0%。对 DxI 9000 和 Access 2 分析仪上的原型 GFAP 分析进行比较后发现,Passing-Bablok 斜率为 0.94(R=0.992),截距为 0.37 pg/ml。一致性分析表明 Access 2 和 DxI 9000 分析仪之间的一致性非常好。Access 2 和 DxI 9000 的 AD 中位数与正常样本中位数的剂量比分别为 2.31 和 2.41,AD 患者的中位数剂量值分别为 14.3 pg/ml 和 14.6 pg/ml,正常样本的中位数剂量值分别为 6.2 pg/ml 和 6.0 pg/ml。结论 GFAP 检测原型可在 Beckman Coulter Access 2 和 DxI 9000 平台上进行自动免疫测定,结果快速、灵敏度高且精确。血浆 GFAP 检测原型有望用于诊断和监测包括 AD 在内的各种神经退行性疾病。
{"title":"B-244 Analytical Performance and Method Comparison Evaluation of a New High Throughput Fully Automated Plasma GFAP Immunoassay","authors":"Z Vucetic, M Szabo, M Salvati, B Schlichtmann, J Patzlaff, D Unruh, K Curtis, L Mediger, M Nichkova-Doseva","doi":"10.1093/clinchem/hvae106.601","DOIUrl":"https://doi.org/10.1093/clinchem/hvae106.601","url":null,"abstract":"Background Glial Fibrillary Acidic Protein (GFAP) levels are shown to be an indicator of neurologic injury in conditions like Traumatic Brain Injury (TBI) and stroke and as a marker of disease progression in neuromuscular disorders such as multiple sclerosis. Plasma GFAP is also implicated in detecting AD pathology, correlating to the clinical stage of the disease. The performance characteristics of the plasma GFAP immunoassay currently being developed on Beckman Coulter Access 2 and DxI 9000 analyzers are described. Methods The prototype GFAP assay is a one-step sandwich assay utilizing an anti-GFAP mouse monoclonal (MAb) antibody/alkaline phosphatase conjugate and paramagnetic particles coated with a complementary anti-GFAP mouse MAb. Sample and reactants are incubated and washed, and then a chemiluminescent substrate is added. The light generated is directly proportional to the GFAP concentration in the sample. The assay time to the first result is ∼30 minutes. Analytical performance evaluation and method comparison across different platforms was performed. 43 EDTA plasma samples, consisting of 19 Alzheimer’s Disease (AD) patient samples and 24 age-matched healthy normal samples, were evaluated. Results The prototype GFAP assay has a targeted analytical measuring range of 1.0 pg/ml to approximately 700.0 pg/ml (up to 10,000 pg/ml). All normal samples were quantified with 90% measuring below 20.0 pg/ml and a low dose coefficient of variation <4.0%. A comparison of the prototype GFAP assay on DxI 9000 and Access 2 analyzers shows a Passing-Bablok slope of 0.94 (R=0.992) and an intercept of 0.37 pg/ml. Concordance analysis shows excellent agreement between the Access 2 and DxI 9000 assays. The dose ratio of the median of AD over the median of normal samples for Access 2 and DxI 9000 is 2.31 and 2.41, respectively, with AD patient median dose values of 14.3 pg/ml and 14.6 pg/mL, respectively, and normal sample median dose values of 6.2 pg/ml and 6.0 pg/mL, respectively. Conclusions The prototype GFAP assay provides fast, highly sensitive, and precise results in an automated immunoassay on the Beckman Coulter Access 2 and DxI 9000 platforms. The prototype plasma GFAP assay holds promise for diagnosing and monitoring various neurodegenerative diseases, including AD.","PeriodicalId":10690,"journal":{"name":"Clinical chemistry","volume":"23 1","pages":""},"PeriodicalIF":9.3,"publicationDate":"2024-10-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"142369120","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
B-333 Optimizing AFP Cutoffs for Hepatocellular Carcinoma Screening: Insights from the National Cancer Screening Program in Korea B-333 优化肝细胞癌筛查的 AFP 临界值:韩国国家癌症筛查计划的启示
IF 9.3 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2024-10-02 DOI: 10.1093/clinchem/hvae106.690
Y Choi, S Kim, H Kim, H Jeong, H Lee, W Lee, S Chun
Background The incidence rate and mortality rate of liver cancer were 11.6 (ranked 8th among cancers) and 10.7 (4th) per 100,000 individuals respectively globally in 2020, whereas in Korea, they were higher at 29.5 (7th) and 20.6 (2nd) per 100,000 individuals. The National Cancer Screening Program (NCSP) in Korea conducts biannual concurrent liver ultrasound and serum alpha-fetoprotein (AFP) measurement for hepatocellular carcinoma (HCC) surveillance in high-risk groups, including those with chronic hepatitis B and C, and liver cirrhosis. The effectiveness of surveillance is impacted by varying AFP cutoffs; however, within the NCSP, different cutoffs are being used, and it is not clear which cutoffs are being employed. Harmonization of AFP test results by major reagent manufacturers have been achieved, enabling the adoption of a uniform threshold. This study aims to assess the variation of AFP cutoffs among institutions within the NCSP and suggest a unified and optimal AFP threshold. Methods This study examined HCC screening results from the NCSP between 2018 and 2020, investigating unit and cutoff usage for AFP test across institutions each year (number of institutions were 4452 for 2018, 4754 for 2019, and 4847 for 2020). To determine the optimal AFP cutoff for HCC screening, datasets comprised unique patient results from 2018 to 2020 (number of patients = 819,644). Receiver operating characteristic (ROC) curve analyses were conducted for determine best AFP cutoffs. Cancer diagnosis was defined by billing records indicating HCC within one year post-screening. Results More than 96% of institutions used ng/mL as a unit of AFP test. Among institutions using ng/mL, the most frequently used AFP cutoff was 7 ng/mL, with frequency percentages of 75.7%, 72.2%, and 62.2% in 2018, 2019, 2020, respectively. The percentiles (1st, 10th, 90th, 99th) of AFP cutoff usage for the years 2018, 2019, and 2020 were (7, 7, 8.8, 15), (7, 7, 8.9, 15), and (7, 7, 9, 15) respectively. The AUC of ROC was 0.824 and some cut-off points with their sensitivity and specificity values were as follows: 5 ng/mL (0.648, 0.874); 10 (0.457, 0.976); 20 (0.327, 0.993); 40 (0.246, 0.997); 100 (0.169, 0.998); 200 (0.123, 0.999); 400 (0.084, 1). Conclusions Different cutoffs observed across institutions in NCSP. Our study cautiously suggests that the optimal AFP value for screening HCC in NCSP may range from 10 to 20 ng/mL, considering sensitivity and specificity when AFP is combined with ultrasound in regions with a high prevalence of HCC. For a more suitable AFP cutoff recommendation, additional analyses based on ultrasound findings and underlying diseases should be considered.
背景 2020年,全球肝癌的发病率和死亡率分别为每10万人11.6例(在癌症中排名第8位)和10.7例(排名第4位),而韩国的发病率和死亡率较高,分别为每10万人29.5例(排名第7位)和20.6例(排名第2位)。韩国国家癌症筛查计划(NCSP)每半年同时进行一次肝脏超声波和血清甲胎蛋白(AFP)测定,以监测高危人群中的肝细胞癌(HCC),包括慢性乙型肝炎和丙型肝炎患者以及肝硬化患者。不同的 AFP 临界值会影响监测的有效性;然而,在 NCSP 中使用了不同的临界值,目前尚不清楚采用的是哪种临界值。主要试剂制造商已经实现了 AFP 检测结果的统一,从而可以采用统一的临界值。本研究旨在评估 NCSP 内各机构 AFP 临界值的差异,并提出统一的最佳 AFP 临界值。方法 本研究检查了 2018 年至 2020 年 NCSP 的 HCC 筛查结果,调查了每年各机构 AFP 检测的单位和阈值使用情况(2018 年机构数为 4452 家,2019 年为 4754 家,2020 年为 4847 家)。为确定 HCC 筛查的最佳 AFP 临界值,数据集包括 2018 年至 2020 年的唯一患者结果(患者人数 = 819644 人)。为确定最佳 AFP 临界值,进行了接收者操作特征 (ROC) 曲线分析。癌症诊断由筛查后一年内显示 HCC 的账单记录定义。结果 96%以上的机构使用纳克/毫升作为 AFP 检测单位。在使用纳克/毫升的机构中,最常使用的 AFP 临界值是 7 纳克/毫升,2018 年、2019 年和 2020 年的频率百分比分别为 75.7%、72.2% 和 62.2%。2018 年、2019 年和 2020 年 AFP 临界值使用率的百分位数(第 1、10、90、99 位)分别为(7、7、8.8、15)、(7、7、8.9、15)和(7、7、9、15)。ROC 的 AUC 为 0.824,一些临界点及其敏感性和特异性值如下:5毫微克/毫升(0.648,0.874);10毫微克/毫升(0.457,0.976);20毫微克/毫升(0.327,0.993);40毫微克/毫升(0.246,0.997);100毫微克/毫升(0.169,0.998);200毫微克/毫升(0.123,0.999);400毫微克/毫升(0.084,1)。结论 在 NCSP 的不同机构中观察到不同的截断值。我们的研究谨慎地建议,在 HCC 高发地区,考虑到 AFP 与超声波相结合的灵敏度和特异性,NCSP 筛查 HCC 的最佳 AFP 值可能在 10 至 20 纳克/毫升之间。为了推荐更合适的 AFP 临界值,应考虑根据超声检查结果和潜在疾病进行更多分析。
{"title":"B-333 Optimizing AFP Cutoffs for Hepatocellular Carcinoma Screening: Insights from the National Cancer Screening Program in Korea","authors":"Y Choi, S Kim, H Kim, H Jeong, H Lee, W Lee, S Chun","doi":"10.1093/clinchem/hvae106.690","DOIUrl":"https://doi.org/10.1093/clinchem/hvae106.690","url":null,"abstract":"Background The incidence rate and mortality rate of liver cancer were 11.6 (ranked 8th among cancers) and 10.7 (4th) per 100,000 individuals respectively globally in 2020, whereas in Korea, they were higher at 29.5 (7th) and 20.6 (2nd) per 100,000 individuals. The National Cancer Screening Program (NCSP) in Korea conducts biannual concurrent liver ultrasound and serum alpha-fetoprotein (AFP) measurement for hepatocellular carcinoma (HCC) surveillance in high-risk groups, including those with chronic hepatitis B and C, and liver cirrhosis. The effectiveness of surveillance is impacted by varying AFP cutoffs; however, within the NCSP, different cutoffs are being used, and it is not clear which cutoffs are being employed. Harmonization of AFP test results by major reagent manufacturers have been achieved, enabling the adoption of a uniform threshold. This study aims to assess the variation of AFP cutoffs among institutions within the NCSP and suggest a unified and optimal AFP threshold. Methods This study examined HCC screening results from the NCSP between 2018 and 2020, investigating unit and cutoff usage for AFP test across institutions each year (number of institutions were 4452 for 2018, 4754 for 2019, and 4847 for 2020). To determine the optimal AFP cutoff for HCC screening, datasets comprised unique patient results from 2018 to 2020 (number of patients = 819,644). Receiver operating characteristic (ROC) curve analyses were conducted for determine best AFP cutoffs. Cancer diagnosis was defined by billing records indicating HCC within one year post-screening. Results More than 96% of institutions used ng/mL as a unit of AFP test. Among institutions using ng/mL, the most frequently used AFP cutoff was 7 ng/mL, with frequency percentages of 75.7%, 72.2%, and 62.2% in 2018, 2019, 2020, respectively. The percentiles (1st, 10th, 90th, 99th) of AFP cutoff usage for the years 2018, 2019, and 2020 were (7, 7, 8.8, 15), (7, 7, 8.9, 15), and (7, 7, 9, 15) respectively. The AUC of ROC was 0.824 and some cut-off points with their sensitivity and specificity values were as follows: 5 ng/mL (0.648, 0.874); 10 (0.457, 0.976); 20 (0.327, 0.993); 40 (0.246, 0.997); 100 (0.169, 0.998); 200 (0.123, 0.999); 400 (0.084, 1). Conclusions Different cutoffs observed across institutions in NCSP. Our study cautiously suggests that the optimal AFP value for screening HCC in NCSP may range from 10 to 20 ng/mL, considering sensitivity and specificity when AFP is combined with ultrasound in regions with a high prevalence of HCC. For a more suitable AFP cutoff recommendation, additional analyses based on ultrasound findings and underlying diseases should be considered.","PeriodicalId":10690,"journal":{"name":"Clinical chemistry","volume":"191 1","pages":""},"PeriodicalIF":9.3,"publicationDate":"2024-10-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"142368755","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
B-008 An Evaluation of the Analytical Performance of the New Beckman Coulter DxC 500i Clinical Analyzer B-008 对新型贝克曼库尔特 DxC 500i 临床分析仪分析性能的评估
IF 9.3 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2024-10-02 DOI: 10.1093/clinchem/hvae106.372
B Tran, L Mindeman, M Reyes, P Kraght, K Lynch, M Meade, L Frost, T Rice, M Cavalleri, S Frost
Background The DxC 500i Clinical Analyzer* is Beckman Coulter’s latest integrated Clinical Chemistry and Immunoassay analyzer. It provides reliability, scalability, and workflow efficiency via a flexible independent maintenance mode, increased uptime, and a dynamic sample handler to optimize sample management. Standard Beckman Coulter Chemistry and Immunoassay consumables and reagents ensure network standardization and commutability. The purpose of this study was to evaluate the analytical performance of the new DxC 500i analyzer and to compare the performance against the Beckman Coulter stand-alone Clinical Chemistry and Immunoassay analyzers on the market. Methods To assess performance over a range of assay methodologies, several DxC 500i Beckman Coulter Immunoassays and Chemistry assays were evaluated. Precision was assessed using human samples following CLSI EP05-A3 guidelines. Linear range was assessed following CLSI EP06-A guidelines. Method Comparison and Bias Estimation studies compared the DxC 500i Clinical Analyzer against the Access 2 Immunoassay System, the AU480 and the DxC 700 AU Clinical Chemistry analyzers, following CLSI EP09c-ED3 guidelines. To investigate carryover potential as samples are processed across the integrated system, a carryover study was designed. This focused on evaluating potential sample-to-sample carryover caused by the AU sample probe and its effects on immunoassay samples during reflex scenarios. Results DxC 500i Chemistry results: Glucose Serum (OSR6x21) Assessments of repeatability and within laboratory precision at multiple critical analyte concentrations showed total imprecision of <3% and within-run imprecision of <3%. The Glucose serum assay demonstrated acceptable linearity throughout the 10-800 mg/dL analytical measuring range. Method comparison studies compared the DxC 500i analyzer against the AU480 and DxC 700 AU systems. A Weighted Deming regression of serum patient samples (n>100 measured across the analytical range) yielded a 0.986 slope, 0.227 mg/dL intercept, and R=0.9999 correlation coefficient against the DxC 700 AU and a 0.995 slope, 0.356 mg/dL intercept, and R=0.9999 correlation coefficient against the AU480. DxC 500i Immunoassay results: Cortisol (33600) Beckman Coutler’s Cortisol assay exhibited total imprecision of <12%CV at <5 μg/dL or <10CV% at ≥5 μg/dL for serum. The Cortisol assay demonstrated acceptable linearity throughout the 0.4-60 μg/dL analytical measuring range. A method comparison study compared the DxC 500i against the Access 2 system. A Weighted Deming regression of serum patient samples (n>100 measured across the analytical range) yielded a 0.9785 slope, 0.2764 μg/dL intercept, and R=0.9932 correlation coefficient. DxC 500i Carryover Assessment The Total βhCG (5th IS) assay was used to evaluate the potential of chemistry analyzer to immunoassay analyzer sample carryover. The study used High hCG serum samples (>1,000
背景 DxC 500i 临床分析仪*是贝克曼库尔特公司最新推出的集成式临床化学和免疫分析仪。它通过灵活的独立维护模式、更长的正常运行时间以及可优化样本管理的动态样本处理程序,提供可靠性、可扩展性和工作流程效率。标准的贝克曼库尔特化学与免疫分析耗材和试剂确保了网络的标准化和通用性。本研究的目的是评估新型 DxC 500i 分析仪的分析性能,并与市场上的贝克曼库尔特独立临床化学和免疫分析仪进行比较。方法 为了评估各种化验方法的性能,对几种 DxC 500i 贝克曼库尔特免疫化验和化学化验进行了评估。精度按照 CLSI EP05-A3 指南使用人体样本进行评估。线性范围按照 CLSI EP06-A 指南进行评估。方法比较和偏差估计研究根据 CLSI EP09c-ED3 指南,将 DxC 500i 临床分析仪与 Access 2 免疫分析系统、AU480 和 DxC 700 AU 临床化学分析仪进行了比较。为了调查样本在集成系统中处理时的迁移潜力,设计了一项迁移研究。这项研究的重点是评估 AU 样品探针可能造成的样品间迁移及其在反射过程中对免疫测定样品的影响。结果 DxC 500i 化学结果:葡萄糖血清 (OSR6x21) 在多个关键分析物浓度下的重复性和实验室内精密度评估显示,总不精密度为 <3% ,运行内不精密度为 <3% 。在 10-800 mg/dL 的分析测量范围内,葡萄糖血清测定显示出了可接受的线性度。方法比较研究将 DxC 500i 分析仪与 AU480 和 DxC 700 AU 系统进行了比较。对患者血清样本(n>100)的加权戴明回归结果显示,与 DxC 700 AU 相比,斜率为 0.986,截距为 0.227 mg/dL,相关系数为 R=0.9999;与 AU480 相比,斜率为 0.995,截距为 0.356 mg/dL,相关系数为 R=0.9999。DxC 500i 免疫测定结果:皮质醇 (33600) Beckman Coutler 的皮质醇检测法在检测 5 μg/dL 时的总不精确度为 <12%CV 或在检测血清≥5 μg/dL 时的总不精确度为 <10CV% 。在 0.4-60 μg/dL 的分析测量范围内,皮质醇测定法显示出了可接受的线性。方法比较研究将 DxC 500i 与 Access 2 系统进行了比较。对患者血清样本(n>100)进行加权戴明回归,结果显示斜率为 0.9785,截距为 0.2764 μg/dL,相关系数为 R=0.9932。DxC 500i 迁移评估 总 βhCG(第 5 IS)测定用于评估化学分析仪到免疫分析仪样本迁移的可能性。研究使用了高 hCG 血清样本(>1,000,000 mIU/mL)和低 hCG 血清样本(4-6 mIU/mL)。DxC 500i 对血清的累积携带量小于百万分之 2 (PPM)。结论 这些研究表明新型贝克曼库尔特DxC 500i临床分析仪*具有出色的分析性能,并证实其性能可与目前市场上的贝克曼库尔特独立临床化学和免疫分析仪媲美。*产品供应和监管状态取决于各国适用法规的注册情况。
{"title":"B-008 An Evaluation of the Analytical Performance of the New Beckman Coulter DxC 500i Clinical Analyzer","authors":"B Tran, L Mindeman, M Reyes, P Kraght, K Lynch, M Meade, L Frost, T Rice, M Cavalleri, S Frost","doi":"10.1093/clinchem/hvae106.372","DOIUrl":"https://doi.org/10.1093/clinchem/hvae106.372","url":null,"abstract":"Background The DxC 500i Clinical Analyzer* is Beckman Coulter’s latest integrated Clinical Chemistry and Immunoassay analyzer. It provides reliability, scalability, and workflow efficiency via a flexible independent maintenance mode, increased uptime, and a dynamic sample handler to optimize sample management. Standard Beckman Coulter Chemistry and Immunoassay consumables and reagents ensure network standardization and commutability. The purpose of this study was to evaluate the analytical performance of the new DxC 500i analyzer and to compare the performance against the Beckman Coulter stand-alone Clinical Chemistry and Immunoassay analyzers on the market. Methods To assess performance over a range of assay methodologies, several DxC 500i Beckman Coulter Immunoassays and Chemistry assays were evaluated. Precision was assessed using human samples following CLSI EP05-A3 guidelines. Linear range was assessed following CLSI EP06-A guidelines. Method Comparison and Bias Estimation studies compared the DxC 500i Clinical Analyzer against the Access 2 Immunoassay System, the AU480 and the DxC 700 AU Clinical Chemistry analyzers, following CLSI EP09c-ED3 guidelines. To investigate carryover potential as samples are processed across the integrated system, a carryover study was designed. This focused on evaluating potential sample-to-sample carryover caused by the AU sample probe and its effects on immunoassay samples during reflex scenarios. Results DxC 500i Chemistry results: Glucose Serum (OSR6x21) Assessments of repeatability and within laboratory precision at multiple critical analyte concentrations showed total imprecision of <3% and within-run imprecision of <3%. The Glucose serum assay demonstrated acceptable linearity throughout the 10-800 mg/dL analytical measuring range. Method comparison studies compared the DxC 500i analyzer against the AU480 and DxC 700 AU systems. A Weighted Deming regression of serum patient samples (n>100 measured across the analytical range) yielded a 0.986 slope, 0.227 mg/dL intercept, and R=0.9999 correlation coefficient against the DxC 700 AU and a 0.995 slope, 0.356 mg/dL intercept, and R=0.9999 correlation coefficient against the AU480. DxC 500i Immunoassay results: Cortisol (33600) Beckman Coutler’s Cortisol assay exhibited total imprecision of <12%CV at <5 μg/dL or <10CV% at ≥5 μg/dL for serum. The Cortisol assay demonstrated acceptable linearity throughout the 0.4-60 μg/dL analytical measuring range. A method comparison study compared the DxC 500i against the Access 2 system. A Weighted Deming regression of serum patient samples (n>100 measured across the analytical range) yielded a 0.9785 slope, 0.2764 μg/dL intercept, and R=0.9932 correlation coefficient. DxC 500i Carryover Assessment The Total βhCG (5th IS) assay was used to evaluate the potential of chemistry analyzer to immunoassay analyzer sample carryover. The study used High hCG serum samples (>1,000","PeriodicalId":10690,"journal":{"name":"Clinical chemistry","volume":"76 1","pages":""},"PeriodicalIF":9.3,"publicationDate":"2024-10-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"142362984","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
B-169 Combining the EXENT System and LC-MS for the Detection of Monoclonal Immunoglobulins at Low Levels B-169 结合 EXENT 系统和 LC-MS 检测低水平单克隆免疫球蛋白
IF 9.3 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2024-10-02 DOI: 10.1093/clinchem/hvae106.529
D R Barnidge, D Troske, S North, G Wallis, M Perkins, S Harding
Background The EXENT® system is an automated platform designed to quantify monoclonal immunoglobulins in serum. EXENT® combines immunoprecipitation and MALDI-TOF mass spectrometry for the detection of monoclonal immunoglobulins at levels lower than traditional gel based methods. This study was designed to demonstrate if EXENT® prepared samples that were negative could be reflexed to a more sensitive method based on capillary flow Liquid Chromatography- Mass Spectrometry (LC-MS) without any modifications to the EXENT® prepared sample. Methods Patient samples containing a monoclonal immunoglobulin (ranging from 20 to 40 g/L) were diluted into pooled polyclonal serum and the samples were prepared and analyzed by EXENT®. Samples that were negative by EXENT® were reflexed by directly loading EXENT® eluates onto the LC-MS instrument. The abundance of the monoclonal immunoglobulin was defined by the signal-to-noise (S/N) of the light chain peak observed after deconvolution. Results Baseline samples were diluted down to two levels, 0.100 g/L and 0.001 g/L. LC-MS detected the same light chain from the monoclonal immunoglobulin as the EXENT® system in all reflexed samples that were negative by EXENT®. Conclusions LC-MS can detect monoclonal immunoglobulins that can no longer be detected using the EXENT® system by tracking the unique molecular mass of the monoclonal light chain. Reflexing samples to LC-MS did not require additional sample handling resulting in a faster time-to-result than current NGS, NGF, and clonotypic peptide approaches. Furthermore, monitoring the intact monoclonal light chain circumvents enzymatic cleavage to create a unique clonotypic peptide, alleviating a situation where no unique peptide can be generated as has been shown previously by Bergen et al. As a consequence, this reflex-methodology results in a consistent way of measuring the presence of malignant B-cells with high sensitivity.
背景 EXENT® 系统是一种自动平台,用于定量检测血清中的单克隆免疫球蛋白。EXENT® 结合了免疫沉淀和 MALDI-TOF 质谱法,用于检测单克隆免疫球蛋白,其水平低于传统的凝胶法。本研究旨在证明 EXENT® 制备的阴性样本是否可以在不对 EXENT® 制备的样本进行任何修改的情况下,反射到基于毛细管流液相色谱-质谱(LC-MS)的更灵敏的方法中。方法 将含有单克隆免疫球蛋白(从 20 克/升到 40 克/升不等)的患者样本稀释成集合多克隆血清,然后用 EXENT® 制备和分析样本。将 EXENT® 洗脱液直接加载到 LC-MS 仪器上,对 EXENT® 检测阴性的样品进行反射。单克隆免疫球蛋白的丰度根据去卷积后观察到的轻链峰的信噪比(S/N)来确定。结果 基准样品被稀释到 0.100 克/升和 0.001 克/升两个水平。在所有 EXENT® 检测呈阴性的反射样本中,LC-MS 与 EXENT® 系统检测到了相同的单克隆免疫球蛋白轻链。结论 LC-MS 可通过跟踪单克隆轻链的独特分子质量,检测 EXENT® 系统无法检测的单克隆免疫球蛋白。将样本反射到 LC-MS 不需要额外的样本处理,因此比目前的 NGS、NGF 和克隆肽方法更快得出结果。此外,监测完整的单克隆轻链可避免酶裂解产生独特的克隆型肽,从而缓解了 Bergen 等人之前所展示的无法产生独特肽的情况。
{"title":"B-169 Combining the EXENT System and LC-MS for the Detection of Monoclonal Immunoglobulins at Low Levels","authors":"D R Barnidge, D Troske, S North, G Wallis, M Perkins, S Harding","doi":"10.1093/clinchem/hvae106.529","DOIUrl":"https://doi.org/10.1093/clinchem/hvae106.529","url":null,"abstract":"Background The EXENT® system is an automated platform designed to quantify monoclonal immunoglobulins in serum. EXENT® combines immunoprecipitation and MALDI-TOF mass spectrometry for the detection of monoclonal immunoglobulins at levels lower than traditional gel based methods. This study was designed to demonstrate if EXENT® prepared samples that were negative could be reflexed to a more sensitive method based on capillary flow Liquid Chromatography- Mass Spectrometry (LC-MS) without any modifications to the EXENT® prepared sample. Methods Patient samples containing a monoclonal immunoglobulin (ranging from 20 to 40 g/L) were diluted into pooled polyclonal serum and the samples were prepared and analyzed by EXENT®. Samples that were negative by EXENT® were reflexed by directly loading EXENT® eluates onto the LC-MS instrument. The abundance of the monoclonal immunoglobulin was defined by the signal-to-noise (S/N) of the light chain peak observed after deconvolution. Results Baseline samples were diluted down to two levels, 0.100 g/L and 0.001 g/L. LC-MS detected the same light chain from the monoclonal immunoglobulin as the EXENT® system in all reflexed samples that were negative by EXENT®. Conclusions LC-MS can detect monoclonal immunoglobulins that can no longer be detected using the EXENT® system by tracking the unique molecular mass of the monoclonal light chain. Reflexing samples to LC-MS did not require additional sample handling resulting in a faster time-to-result than current NGS, NGF, and clonotypic peptide approaches. Furthermore, monitoring the intact monoclonal light chain circumvents enzymatic cleavage to create a unique clonotypic peptide, alleviating a situation where no unique peptide can be generated as has been shown previously by Bergen et al. As a consequence, this reflex-methodology results in a consistent way of measuring the presence of malignant B-cells with high sensitivity.","PeriodicalId":10690,"journal":{"name":"Clinical chemistry","volume":"14 1","pages":""},"PeriodicalIF":9.3,"publicationDate":"2024-10-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"142369050","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
A-345 Effect of Hemolysis on Routine Blood Gas Analytes A-345 溶血对常规血气分析物的影响
IF 9.3 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2024-10-02 DOI: 10.1093/clinchem/hvae106.339
B Agana, B Overton, K Florendo, C Knezevic
Background Hemolysis is a major pre-analytical concern for most laboratory analytes. Detection of hemolysis and mitigation efforts are especially important for analytes, such as potassium, with high intracellular concentrations. Routine serum and plasma chemistry tests are performed on analyzers with the capacity to detect and measure the degree of hemolysis. However, for whole blood chemistries and blood gas measurements, the instruments utilized lack this capacity. The aim of this study was to evaluate the effect of hemolysis on whole blood and blood gas analytes and to compare visual assessments of hemolysis to measured hemolysis indices (H-index). Methods Remnant whole blood samples were split into two portions and each portion was aspirated and dispensed through a syringe one or more times. For the mock portion, only the syringe was used, while the hemolyzed portion had a needle affixed to the end of the syringe. The needle provided shear stress on the red blood cells to induce hemolysis, while the mock procedure was used to assess the impact of aspirating/dispensing on the sample in the absence of hemolysis. Each portion was then analyzed on a Radiometer ABL800 series instrument, spun down, and the H-index of the plasma portion was measured on a Roche cobas 8000 instrument. Medical technologists recorded their visual assessment of the specimens, with two technologists agreeing to the categorization of the specimen as either slightly, moderately, or severely hemolyzed. Degree of hemolysis was categorized by the delta (hemolyzed - mock) of the measured H-index: slight hemolysis was defined as an H-index delta of <100, moderate as 100-500 and severe as >500. Results The effect of hemolysis, with H-indices ranging from 2 to 3861, on 13 routine blood gas analytes was studied for 85 whole blood specimens. Hemolysis had little effect on metabolites, as percent bias was within ±3% at all levels of hemolysis for glucose, creatinine, and lactate. Similarly, most cooximetry components were minimally affected, with total hemoglobin, oxyhemoglobin, carboxyhemoglobin, and oxygen saturation within ±5% bias at all levels of hemolysis. Methemoglobin had a larger overall negative bias, with slight, moderate, and severe hemolysis levels yielding percent biases of -6.6, -12.3, and -13.3%, respectively. As expected, potassium displayed a significant positive bias with increasing hemolysis, with a generally linear trend. At moderate levels of hemolysis, the average potassium bias was 24.0% and at severe levels, over 100%. Sodium and ionized calcium also displayed overall linear trends but with a negative bias. At slight, moderate, and severe levels of hemolysis, sodium had a -0.56, -1.10, and -3.96% bias, and ionized calcium had a -2.99, -5.65, and -15.5% bias respectively. Conclusions Hemolysis can falsely increase or decrease a range of blood gas analytes and lead to misinterpretation of results and adversely affect clinical decision-making. Ther
背景溶血是大多数实验室分析物分析前的主要问题。对于钾等细胞内浓度较高的分析物而言,溶血的检测和缓解尤为重要。常规血清和血浆化学检测是在具有检测和测量溶血程度功能的分析仪上进行的。然而,在进行全血化学和血气测量时,所使用的仪器却缺乏这种能力。本研究的目的是评估溶血对全血和血气分析物的影响,并将溶血的肉眼评估与溶血指数(H-index)的测量结果进行比较。方法 将残留全血样本分成两份,每份样本都用注射器抽吸并分配一次或多次。模拟部分只使用注射器,而溶血部分则在注射器末端粘贴一根针。针头对红细胞产生剪切应力以诱导溶血,而模拟程序则用于评估在不溶血的情况下抽吸/分配对样本的影响。然后用 Radiometer ABL800 系列仪器对每份样本进行分析、离心,并用 Roche cobas 8000 仪器测量血浆部分的 H 指数。医疗技术人员记录他们对标本的目测评估,由两名技术人员一致同意将标本分为轻微溶血、中度溶血或严重溶血。溶血程度根据测得的 H 指数δ(溶血-模拟)进行分类:H 指数δ为<100 时为轻微溶血,100-500 时为中度溶血,500 时为重度溶血。结果 研究了 85 份全血标本中溶血(H 指数从 2 到 3861 不等)对 13 种常规血气分析物的影响。溶血对代谢物的影响很小,因为在所有溶血水平下,葡萄糖、肌酐和乳酸盐的偏差百分比都在±3%以内。同样,大多数共氧血红蛋白成分受到的影响也很小,总血红蛋白、氧合血红蛋白、碳氧血红蛋白和血氧饱和度在所有溶血程度下的偏差都在±5%以内。高铁血红蛋白的总体负偏差较大,轻度、中度和重度溶血水平的偏差百分比分别为-6.6%、-12.3%和-13.3%。不出所料,随着溶血量的增加,钾的偏倚率呈显著的正值,且总体呈线性趋势。中度溶血时,钾的平均偏差为 24.0%,严重溶血时超过 100%。钠和离子钙也呈总体线性趋势,但偏差为负。在轻度、中度和重度溶血时,钠的偏差分别为-0.56%、-1.10%和-3.96%,离子钙的偏差分别为-2.99%、-5.65%和-15.5%。结论 溶血会使一系列血气分析物发生错误的增减,导致对结果的误读,并对临床决策产生不利影响。因此,在当前的血气分析仪上配备溶血检测功能至关重要,可使实验室减轻这种影响,确保为患者提供准确的结果。
{"title":"A-345 Effect of Hemolysis on Routine Blood Gas Analytes","authors":"B Agana, B Overton, K Florendo, C Knezevic","doi":"10.1093/clinchem/hvae106.339","DOIUrl":"https://doi.org/10.1093/clinchem/hvae106.339","url":null,"abstract":"Background Hemolysis is a major pre-analytical concern for most laboratory analytes. Detection of hemolysis and mitigation efforts are especially important for analytes, such as potassium, with high intracellular concentrations. Routine serum and plasma chemistry tests are performed on analyzers with the capacity to detect and measure the degree of hemolysis. However, for whole blood chemistries and blood gas measurements, the instruments utilized lack this capacity. The aim of this study was to evaluate the effect of hemolysis on whole blood and blood gas analytes and to compare visual assessments of hemolysis to measured hemolysis indices (H-index). Methods Remnant whole blood samples were split into two portions and each portion was aspirated and dispensed through a syringe one or more times. For the mock portion, only the syringe was used, while the hemolyzed portion had a needle affixed to the end of the syringe. The needle provided shear stress on the red blood cells to induce hemolysis, while the mock procedure was used to assess the impact of aspirating/dispensing on the sample in the absence of hemolysis. Each portion was then analyzed on a Radiometer ABL800 series instrument, spun down, and the H-index of the plasma portion was measured on a Roche cobas 8000 instrument. Medical technologists recorded their visual assessment of the specimens, with two technologists agreeing to the categorization of the specimen as either slightly, moderately, or severely hemolyzed. Degree of hemolysis was categorized by the delta (hemolyzed - mock) of the measured H-index: slight hemolysis was defined as an H-index delta of <100, moderate as 100-500 and severe as >500. Results The effect of hemolysis, with H-indices ranging from 2 to 3861, on 13 routine blood gas analytes was studied for 85 whole blood specimens. Hemolysis had little effect on metabolites, as percent bias was within ±3% at all levels of hemolysis for glucose, creatinine, and lactate. Similarly, most cooximetry components were minimally affected, with total hemoglobin, oxyhemoglobin, carboxyhemoglobin, and oxygen saturation within ±5% bias at all levels of hemolysis. Methemoglobin had a larger overall negative bias, with slight, moderate, and severe hemolysis levels yielding percent biases of -6.6, -12.3, and -13.3%, respectively. As expected, potassium displayed a significant positive bias with increasing hemolysis, with a generally linear trend. At moderate levels of hemolysis, the average potassium bias was 24.0% and at severe levels, over 100%. Sodium and ionized calcium also displayed overall linear trends but with a negative bias. At slight, moderate, and severe levels of hemolysis, sodium had a -0.56, -1.10, and -3.96% bias, and ionized calcium had a -2.99, -5.65, and -15.5% bias respectively. Conclusions Hemolysis can falsely increase or decrease a range of blood gas analytes and lead to misinterpretation of results and adversely affect clinical decision-making. Ther","PeriodicalId":10690,"journal":{"name":"Clinical chemistry","volume":"4 1","pages":""},"PeriodicalIF":9.3,"publicationDate":"2024-10-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"142369053","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
A-319 Capillary Blood Collection for Egg-cellent Reproductive Insights A-319 毛细管采血,了解卵子繁殖情况
IF 9.3 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2024-10-02 DOI: 10.1093/clinchem/hvae106.314
W C Brandon, B B Collier, M R Chappell, G Iacovetti, M Peevler, U Y Schaff, G J Sommer, R P Grant
Background In the United States, an estimated 13% of women between the ages of 15 and 49 years, experience impaired reproductive potential (fecundity). Assessment of fecundity typically requires an in-person physician visit and venous blood collection to evaluate hormone levels and identify potential reproductive health conditions. Exploring less invasive, at-home sampling options could offer couples insights into their personal reproductive abilities, providing peace of mind and aiding in family planning. Methods Capillary blood, obtained using the Tasso®+ capillary blood collection device coupled with a BD Microtainer® serum tube, was utilized for measurements of analytes associated with female fecundity (follicle stimulating hormone, luteinizing hormone, estradiol, progesterone, anti-Müllerian hormone, and prolactin). These are all FDA-approved assays measured on Roche cobas® e801, assay parameters remained unchanged. Professional and self-collected capillary serum samples were acquired in parallel with the standard venous serum samples, to compare clinical equivalency of the capillary sampling option. In addition to isothermal stability studies at room temperature (20-25°C) and refrigerated (2-8°C) conditions, shipping excursion studies based on ISTA 7D recommendations were performed using a novel packaging solution to control sample temperature and thus, specimen integrity. Studies to assess matrix equivalency, linearity, and imprecision of this alternate sample type were also performed to demonstrate analytical performance. Results Method comparison of venous and professional collection of capillary samples demonstrated correlations (R) ≥ 0.9720 with slopes ranging from 0.919 to 0.971 and mean biases within ±8.2% for all analytes (n = 19-79). Self-collected capillary serum samples demonstrated correlations (R) ≥ 0.9827 with slopes that ranged from 0.841 to 0.985 and mean biases within ±10.7% (n = 16-26). Isothermal stability was demonstrated for each analyte up to 7 days at room temperature and refrigerated conditions. Simulated shipping excursion results were found to be acceptable for both winter and summer conditions with mean biases for each analyte within ±10.2% and ±19.2%, respectively. Matrix equivalency, linearity, and imprecision results adhered to current CLIA acceptance limits for each respective measurement. Conclusions The analytical and stability results affirm the feasibility of collecting capillary serum in a patient-centric manner, providing women with relevant information about their reproductive health.
背景 在美国,估计有 13% 年龄在 15 岁至 49 岁之间的女性的生殖潜能(生育能力)会受到损害。生育力评估通常需要医生亲自到访并采集静脉血,以评估激素水平并确定潜在的生殖健康状况。探索侵入性较小的居家采样方法,可以让夫妇们深入了解个人的生殖能力,让他们安心并有助于计划生育。方法 使用 Tasso®+ 毛细管采血器和 BD Microtainer® 血清管采集毛细管血液,用于测量与女性生育能力相关的分析物(促卵泡激素、黄体生成素、雌二醇、孕酮、抗穆勒氏管激素和催乳素)。这些都是经美国 FDA 批准的检测方法,使用罗氏 cobas® e801 进行检测,检测参数保持不变。在采集标准静脉血清样本的同时,还采集了专业和自采的毛细管血清样本,以比较毛细管采样方法的临床等效性。除了室温(20-25°C)和冷藏(2-8°C)条件下的等温稳定性研究外,还根据 ISTA 7D 建议使用新型包装解决方案进行了运输偏移研究,以控制样本温度,从而保证样本完整性。此外,还进行了基质等效性、线性度和这种替代样本类型的不精密度评估研究,以证明分析性能。结果 对静脉样本和专业人员采集的毛细管样本进行方法比较,结果显示相关性 (R) ≥ 0.9720,斜率范围为 0.919 至 0.971,所有分析物的平均偏差在 ±8.2% 以内(n = 19-79)。自取毛细管血清样本的相关性 (R) ≥ 0.9827,斜率范围为 0.841 至 0.985,平均偏差在 ±10.7% 以内(n = 16-26)。在室温和冷藏条件下,每种分析物的等温稳定性可达 7 天。在冬季和夏季条件下,模拟运输偏差结果均可接受,每种分析物的平均偏差分别在±10.2%和±19.2%以内。基质等效性、线性度和不精密度结果均符合当前 CLIA 对各项测量的接受范围。结论 分析和稳定性结果证实了以患者为中心收集毛细管血清的可行性,为妇女提供了生殖健康的相关信息。
{"title":"A-319 Capillary Blood Collection for Egg-cellent Reproductive Insights","authors":"W C Brandon, B B Collier, M R Chappell, G Iacovetti, M Peevler, U Y Schaff, G J Sommer, R P Grant","doi":"10.1093/clinchem/hvae106.314","DOIUrl":"https://doi.org/10.1093/clinchem/hvae106.314","url":null,"abstract":"Background In the United States, an estimated 13% of women between the ages of 15 and 49 years, experience impaired reproductive potential (fecundity). Assessment of fecundity typically requires an in-person physician visit and venous blood collection to evaluate hormone levels and identify potential reproductive health conditions. Exploring less invasive, at-home sampling options could offer couples insights into their personal reproductive abilities, providing peace of mind and aiding in family planning. Methods Capillary blood, obtained using the Tasso®+ capillary blood collection device coupled with a BD Microtainer® serum tube, was utilized for measurements of analytes associated with female fecundity (follicle stimulating hormone, luteinizing hormone, estradiol, progesterone, anti-Müllerian hormone, and prolactin). These are all FDA-approved assays measured on Roche cobas® e801, assay parameters remained unchanged. Professional and self-collected capillary serum samples were acquired in parallel with the standard venous serum samples, to compare clinical equivalency of the capillary sampling option. In addition to isothermal stability studies at room temperature (20-25°C) and refrigerated (2-8°C) conditions, shipping excursion studies based on ISTA 7D recommendations were performed using a novel packaging solution to control sample temperature and thus, specimen integrity. Studies to assess matrix equivalency, linearity, and imprecision of this alternate sample type were also performed to demonstrate analytical performance. Results Method comparison of venous and professional collection of capillary samples demonstrated correlations (R) ≥ 0.9720 with slopes ranging from 0.919 to 0.971 and mean biases within ±8.2% for all analytes (n = 19-79). Self-collected capillary serum samples demonstrated correlations (R) ≥ 0.9827 with slopes that ranged from 0.841 to 0.985 and mean biases within ±10.7% (n = 16-26). Isothermal stability was demonstrated for each analyte up to 7 days at room temperature and refrigerated conditions. Simulated shipping excursion results were found to be acceptable for both winter and summer conditions with mean biases for each analyte within ±10.2% and ±19.2%, respectively. Matrix equivalency, linearity, and imprecision results adhered to current CLIA acceptance limits for each respective measurement. Conclusions The analytical and stability results affirm the feasibility of collecting capillary serum in a patient-centric manner, providing women with relevant information about their reproductive health.","PeriodicalId":10690,"journal":{"name":"Clinical chemistry","volume":"23 1","pages":""},"PeriodicalIF":9.3,"publicationDate":"2024-10-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"142369087","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
A-104 Study of the matrix effect for thyroglobulin and calcitonin in fine-needle aspiration of cervical nodules A-104 宫颈结节细针穿刺中甲状腺球蛋白和降钙素的基质效应研究
IF 9.3 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2024-10-02 DOI: 10.1093/clinchem/hvae106.103
D Canali, G Zampieri, C Bonato, G Francisco, C Sabino, V Oliveira, A Lopes
Background Analysis of material from fine needle aspiration of cervical thyroglobulin nodules with Beckman Coulter reagents on the ACCESS equipment and Calcitonin using the Siemens reagent on the Immulite equipment. The matrix effect, linearity and the limit of analytical sensitivity were analyzed. This study aims to analyze the correlation between the serum/plasma results described in the suppliers' package inserts with the results obtained with the puncture matrix. Methods The performance characteristics were tested according to the laboratory's operating procedures, based on the recommendations of ANVISA RDC 166/2017. Such parameters involved doping, which corresponds to the introduction of a calibrator with a known and traceable value into a pool of puncture material that is known to be negative, in order to analyze the behavior of the results obtained, paying attention to the deviations that may eventually occur due to the interference of the characteristics of the puncture material. puncture. After doping the material, dilutions were performed up to the sensitivity limit of the test, also analyzing the functional sensitivity, where ten replicates of the material were dosed, thus obtaining CV%, mean and Standard Deviation values. The same process was carried out and evaluated using the serum material. During the entire process, the materials received and analyzed were centrifuged and always kept in a refrigerated environment. Results The recovery% of the matrix effect study for cervical aspirate puncture showed results within the expected range, ranging from 80% to 120%, as well as the results compared with serum/plasma referenced in the supplier's package insert. The functional sensitivity study for Thyroglobulin showed a CV% of 1.77%, being lower than that recommended by the package insert (2.2%) and for the Calcitonin analyte, the functional sensitivity study showed a CV% of 8.49%, being lower than recommended by the package leaflet (15.7%), in addition to having CV's lower than the 20% recommended by ANVISA. The results obtained with the matrix effect study for serum also showed values between 80% and 120%, both of which agreed. Conclusions The results showed excellent performance for material from fine needle aspiration for both Calcitonin and Thyroglobulin, being consistent with the serum data described by the suppliers' package inserts, thus making the reagent usable for the cervical aspirate puncture matrix.
背景 在 ACCESS 设备上使用贝克曼库尔特试剂分析宫颈甲状腺球蛋白结节的细针穿刺材料,在 Immulite 设备上使用西门子试剂分析降钙素。分析了基质效应、线性度和分析灵敏度极限。本研究旨在分析供应商包装说明书中描述的血清/血浆检测结果与穿刺基质检测结果之间的相关性。方法 根据 ANVISA RDC 166/2017 的建议,按照实验室操作程序对性能特征进行测试。这些参数涉及掺杂,即在已知为阴性的穿刺材料池中引入已知且可溯源的校准物,以分析所获结果的行为,同时注意由于穿刺材料特性的干扰而最终可能出现的偏差。掺入材料后,稀释至测试的灵敏度极限,同时分析功能灵敏度,在此过程中掺入 10 次材料,从而获得 CV%、平均值和标准偏差值。对血清材料也进行了同样的处理和评估。在整个过程中,接收和分析的材料都经过离心,并始终保存在冷藏环境中。结果 宫颈吸液穿刺基质效应研究的回收率在预期范围内,从 80% 到 120% 不等,与供应商包装说明书中提到的血清/血浆相比也是如此。甲状腺球蛋白的功能灵敏度研究显示,CV% 为 1.77%,低于包装说明书中的建议值(2.2%);降钙素分析物的功能灵敏度研究显示,CV% 为 8.49%,低于包装说明书中的建议值(15.7%),此外,CV 值还低于 ANVISA 建议的 20%。血清基质效应研究的结果也显示出 80% 到 120% 之间的数值,两者一致。结论 结果表明,细针抽吸法测得的降钙素和甲状腺球蛋白性能优异,与供应商包装说明书中描述的血清数据一致,因此该试剂可用于宫颈穿刺基质。
{"title":"A-104 Study of the matrix effect for thyroglobulin and calcitonin in fine-needle aspiration of cervical nodules","authors":"D Canali, G Zampieri, C Bonato, G Francisco, C Sabino, V Oliveira, A Lopes","doi":"10.1093/clinchem/hvae106.103","DOIUrl":"https://doi.org/10.1093/clinchem/hvae106.103","url":null,"abstract":"Background Analysis of material from fine needle aspiration of cervical thyroglobulin nodules with Beckman Coulter reagents on the ACCESS equipment and Calcitonin using the Siemens reagent on the Immulite equipment. The matrix effect, linearity and the limit of analytical sensitivity were analyzed. This study aims to analyze the correlation between the serum/plasma results described in the suppliers' package inserts with the results obtained with the puncture matrix. Methods The performance characteristics were tested according to the laboratory's operating procedures, based on the recommendations of ANVISA RDC 166/2017. Such parameters involved doping, which corresponds to the introduction of a calibrator with a known and traceable value into a pool of puncture material that is known to be negative, in order to analyze the behavior of the results obtained, paying attention to the deviations that may eventually occur due to the interference of the characteristics of the puncture material. puncture. After doping the material, dilutions were performed up to the sensitivity limit of the test, also analyzing the functional sensitivity, where ten replicates of the material were dosed, thus obtaining CV%, mean and Standard Deviation values. The same process was carried out and evaluated using the serum material. During the entire process, the materials received and analyzed were centrifuged and always kept in a refrigerated environment. Results The recovery% of the matrix effect study for cervical aspirate puncture showed results within the expected range, ranging from 80% to 120%, as well as the results compared with serum/plasma referenced in the supplier's package insert. The functional sensitivity study for Thyroglobulin showed a CV% of 1.77%, being lower than that recommended by the package insert (2.2%) and for the Calcitonin analyte, the functional sensitivity study showed a CV% of 8.49%, being lower than recommended by the package leaflet (15.7%), in addition to having CV's lower than the 20% recommended by ANVISA. The results obtained with the matrix effect study for serum also showed values between 80% and 120%, both of which agreed. Conclusions The results showed excellent performance for material from fine needle aspiration for both Calcitonin and Thyroglobulin, being consistent with the serum data described by the suppliers' package inserts, thus making the reagent usable for the cervical aspirate puncture matrix.","PeriodicalId":10690,"journal":{"name":"Clinical chemistry","volume":"77 1","pages":""},"PeriodicalIF":9.3,"publicationDate":"2024-10-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"142368756","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
A-306 High Sensitivity Capability of an Anti-müllerian Hormone Assay on the Beckman Coulter DxI 9000 Immunoassay Analyzer*, a Next Step Toward Enabling Ovarian Failure Staging? A-306 Beckman Coulter DxI 9000 免疫测定分析仪*上的高灵敏度抗苗勒管激素测定,实现卵巢功能衰竭分期的下一步?
IF 9.3 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2024-10-02 DOI: 10.1093/clinchem/hvae106.301
S Kuklok, J Thomforde, L Pearson, K Pritchard, T Pfingston, C Bunkelman, L Pearson, T Her, B Bolstad, B Bilyeu, S Santiago, S Thornburgh, M Posnansky, N Malikowski-Hoffarth, K Walt, D Lovett, M Holland, M Szabo, C Carlson
Background Measurement of anti-müllerian hormone (AMH) has proven value in the evaluation of ovarian reserve using assays such as the Beckman Coulter Access AMH assay. If improved sensitivity of AMH measurement can be achieved, enhanced clinical utility during the assessment of reproductive aging and ovarian response to enable staging of ovarian failure is possible (Practice Committee of the American Society of Reproductive Medicine 2020, Cappola et al. 2023). The Beckman Coulter DxI 9000 Immunoassay Analyzer* has been designed with new technology that can be leveraged to improve assay sensitivity. Such technological advancements include the new Lumi-Phos PRO chemiluminescent substrate which yields markedly increased signal-to-noise, a new luminometer with increased dynamic range, and improved low-volume pipetting capabilities. Data herein summarize results from analytical verification studies of the Access AMH assay on the DxI 9000 Immunoassay Analyzer. Results demonstrate an example of DxI 9000 technologies enabling sensitivity claims that are up to 20-fold improved. * The official name is DxI 9000 Access Immunoassay Analyzer. Methods A method comparison study was performed based on CLSI guideline EP09c, 3rd ed. to compare results for the Access AMH assay on DxI 9000 to the predicate device, the Access AMH assay on the Access 2 Immunoassay System. Within-laboratory precision was also evaluated following CLSI EP05-A3. Finally, performance at low analyte levels was evaluated following CLSI EP17-A2, and linearity was evaluated following CLSI EP06-Ed2. Results Method comparison of the Access AMH assay on DxI 9000 compared to the Access AMH assay on the Access 2 yielded excellent agreement with a Passing-Bablok slope of 1.02 for N=126 samples. Further, bias estimates for a number of key medical decision points suggest minimal bias (≤ 2%) across the assay range, and minimal non-linearity was observed. 20-day imprecision studies yielded CV’s between 2 and 5% across a broad range of concentrations evaluated. Sensitivity studies yielded estimates of limit of quantitation (LoQ) between 0.001 and 0.003 ng/mL. This represents an approximately 10- to 20-fold improvement in sensitivity claims compared to the AMH assay on the Access 2 and other commercially available automated immunoassay analyzers. Conclusions The data herein present performance data for the Access AMH assay on the DxI 9000 Immunoassay Analyzer. The assay demonstrates excellent accuracy relative to the predicate device, while showing good linearity and imprecision. Of note, the Access AMH assay exhibits exceptional low-end performance on the DxI 9000 analyzer as demonstrated by the ability to enable sensitivity claims that are up to 20-fold improved in comparison to existing immunoassays for the measurement of AMH. The sensitivity capabilities of the Access AMH assay on DxI 9000 suggest an opportunity for the addition of further intended uses for scenarios where extremely low AMH measurement
背景使用贝克曼库尔特公司的Access AMH测定法等测定抗缪勒氏管激素(AMH)在评估卵巢储备功能方面的价值已得到证实。如果能提高 AMH 测量的灵敏度,就有可能在评估生殖衰老和卵巢反应时提高临床实用性,从而对卵巢功能衰竭进行分期(美国生殖医学会实践委员会 2020 年,Cappola 等 2023 年)。贝克曼库尔特 DxI 9000 免疫分析仪*的设计采用了可用于提高检测灵敏度的新技术。这些技术进步包括:可显著提高信噪比的新型 Lumi-Phos PRO 化学发光底物、可提高动态范围的新型发光仪以及改进的低容量移液能力。本文总结了在 DxI 9000 免疫分析仪上对 Access AMH 检测法进行分析验证研究的结果。结果展示了 DxI 9000 技术使灵敏度提高 20 倍的实例。* 正式名称为 DxI 9000 Access 免疫分析仪。方法 根据 CLSI 指南 EP09c 第 3 版进行了方法比较研究,将 DxI 9000 上的 Access AMH 检测结果与同类设备(Access 2 免疫测定系统上的 Access AMH 检测结果)进行比较。此外,还按照 CLSI EP05-A3 对实验室内的精密度进行了评估。最后,根据 CLSI EP17-A2 评估了低分析物水平下的性能,并根据 CLSI EP06-Ed2 评估了线性度。结果 将 DxI 9000 上的 Access AMH 检测法与 Access 2 上的 Access AMH 检测法进行方法比较,结果显示两者的一致性非常好,N=126 个样本的 Passing-Bablok 斜率为 1.02。此外,对一些关键医疗决策点的偏倚估计表明,整个检测范围内的偏倚极小(≤ 2%),而且观察到的非线性极小。20 天不精密度研究显示,在广泛的评估浓度范围内,CV 值介于 2% 和 5% 之间。灵敏度研究得出的定量限 (LoQ) 估计值介于 0.001 和 0.003 纳克/毫升之间。与 Access 2 和其他市售自动免疫分析仪上的 AMH 检测相比,灵敏度提高了约 10 到 20 倍。结论 本文数据显示了 Access AMH 检测法在 DxI 9000 免疫分析仪上的性能数据。该检测法与同类设备相比具有极佳的准确性,同时还显示出良好的线性和不精密度。值得注意的是,Access AMH 检测试剂盒在 DxI 9000 分析仪上表现出卓越的低端性能,其灵敏度比现有的 AMH 免疫测定试剂盒提高了 20 倍。DxI 9000 上 Access AMH 检测法的灵敏度能力表明,在极低 AMH 测量可提供具有临床意义的额外价值的情况下,有机会增加更多预期用途。要充分了解 DxI 9000 上高灵敏度的 Access AMH 检测法的其他临床用途,包括用于更年期诊断的生殖衰老评估或用于生育治疗的卵巢反应评估,还需要进一步的研究。
{"title":"A-306 High Sensitivity Capability of an Anti-müllerian Hormone Assay on the Beckman Coulter DxI 9000 Immunoassay Analyzer*, a Next Step Toward Enabling Ovarian Failure Staging?","authors":"S Kuklok, J Thomforde, L Pearson, K Pritchard, T Pfingston, C Bunkelman, L Pearson, T Her, B Bolstad, B Bilyeu, S Santiago, S Thornburgh, M Posnansky, N Malikowski-Hoffarth, K Walt, D Lovett, M Holland, M Szabo, C Carlson","doi":"10.1093/clinchem/hvae106.301","DOIUrl":"https://doi.org/10.1093/clinchem/hvae106.301","url":null,"abstract":"Background Measurement of anti-müllerian hormone (AMH) has proven value in the evaluation of ovarian reserve using assays such as the Beckman Coulter Access AMH assay. If improved sensitivity of AMH measurement can be achieved, enhanced clinical utility during the assessment of reproductive aging and ovarian response to enable staging of ovarian failure is possible (Practice Committee of the American Society of Reproductive Medicine 2020, Cappola et al. 2023). The Beckman Coulter DxI 9000 Immunoassay Analyzer* has been designed with new technology that can be leveraged to improve assay sensitivity. Such technological advancements include the new Lumi-Phos PRO chemiluminescent substrate which yields markedly increased signal-to-noise, a new luminometer with increased dynamic range, and improved low-volume pipetting capabilities. Data herein summarize results from analytical verification studies of the Access AMH assay on the DxI 9000 Immunoassay Analyzer. Results demonstrate an example of DxI 9000 technologies enabling sensitivity claims that are up to 20-fold improved. * The official name is DxI 9000 Access Immunoassay Analyzer. Methods A method comparison study was performed based on CLSI guideline EP09c, 3rd ed. to compare results for the Access AMH assay on DxI 9000 to the predicate device, the Access AMH assay on the Access 2 Immunoassay System. Within-laboratory precision was also evaluated following CLSI EP05-A3. Finally, performance at low analyte levels was evaluated following CLSI EP17-A2, and linearity was evaluated following CLSI EP06-Ed2. Results Method comparison of the Access AMH assay on DxI 9000 compared to the Access AMH assay on the Access 2 yielded excellent agreement with a Passing-Bablok slope of 1.02 for N=126 samples. Further, bias estimates for a number of key medical decision points suggest minimal bias (≤ 2%) across the assay range, and minimal non-linearity was observed. 20-day imprecision studies yielded CV’s between 2 and 5% across a broad range of concentrations evaluated. Sensitivity studies yielded estimates of limit of quantitation (LoQ) between 0.001 and 0.003 ng/mL. This represents an approximately 10- to 20-fold improvement in sensitivity claims compared to the AMH assay on the Access 2 and other commercially available automated immunoassay analyzers. Conclusions The data herein present performance data for the Access AMH assay on the DxI 9000 Immunoassay Analyzer. The assay demonstrates excellent accuracy relative to the predicate device, while showing good linearity and imprecision. Of note, the Access AMH assay exhibits exceptional low-end performance on the DxI 9000 analyzer as demonstrated by the ability to enable sensitivity claims that are up to 20-fold improved in comparison to existing immunoassays for the measurement of AMH. The sensitivity capabilities of the Access AMH assay on DxI 9000 suggest an opportunity for the addition of further intended uses for scenarios where extremely low AMH measurement","PeriodicalId":10690,"journal":{"name":"Clinical chemistry","volume":"26 1","pages":""},"PeriodicalIF":9.3,"publicationDate":"2024-10-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"142368791","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
A-353 Lay User testing of Rapid Multiplexed HIV/TP test A-353 艾滋病毒/TP 快速多重检测的非专业用户测试
IF 9.3 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Pub Date : 2024-10-02 DOI: 10.1093/clinchem/hvae106.347
K D Jones, N Bauer
Background Palladium Diagnostics had developed a rapid, multiplexed POC assay for Syphilis and HIV for fingerstick whole blood. To increase potential patient engagement and simplify sample collection it was decided to expand the sample type from fingerstick whole blood to saliva without changing the test time or performance. Additionally, the test was designed to have a lower carbon footprint by using a bioplastic (environmentally friendly foam) for structural components rather than injection molded parts as are commonly used in other assays. The assay was run both by professional users in a lab setting and lay users in the field. Methods A flow through rapid assay was produced and tested both by professional users in a lab setting using library sample and by lay users in a field setting using self collected samples. Results The results are shown in table 1, the assay performance was broadly similar for both whole blood and saliva, however the saliva dataset was much smaller for the professional use setting. The human factor analysis from the lay user field studies showed that they preferred the saliva collection procedure however for a small subset of subjects it proved difficult to collect sufficient sample for testing. For subjects who were able to collect adequate sample the results corresponded well between the device and lab-based ELISA. Conclusions The lay user experience validated test design and function and proved the usability in the field.
背景 Palladium Diagnostics 公司开发了一种快速、多路复用的 POC 化验方法,用于梅毒和艾滋病毒的指血全血检测。为了提高潜在患者的参与度并简化样本采集,公司决定在不改变检测时间或性能的情况下,将样本类型从指模全血扩大到唾液。此外,该检测试剂盒的结构部件采用了生物塑料(环保泡沫),而不是其他检测试剂盒常用的注塑部件,从而降低了碳足迹。实验室的专业用户和实地的非专业用户都可以使用该检测方法。方法 由专业用户在实验室使用库中样本,非专业用户在现场使用自己收集的样本,制作并测试了流 通式快速检测法。结果 如表 1 所示,全血和唾液的检测性能大致相同,但专业用户使用的唾液数据集要小得多。非专业用户实地研究的人为因素分析表明,他们更喜欢唾液采集程序,但对于一小部分受试者来说,很难采集到足够的测试样本。对于能够采集到足够样本的受试者来说,设备和实验室酶联免疫吸附试验的结果非常吻合。结论 非专业用户的体验验证了测试的设计和功能,并证明了其在现场的可用性。
{"title":"A-353 Lay User testing of Rapid Multiplexed HIV/TP test","authors":"K D Jones, N Bauer","doi":"10.1093/clinchem/hvae106.347","DOIUrl":"https://doi.org/10.1093/clinchem/hvae106.347","url":null,"abstract":"Background Palladium Diagnostics had developed a rapid, multiplexed POC assay for Syphilis and HIV for fingerstick whole blood. To increase potential patient engagement and simplify sample collection it was decided to expand the sample type from fingerstick whole blood to saliva without changing the test time or performance. Additionally, the test was designed to have a lower carbon footprint by using a bioplastic (environmentally friendly foam) for structural components rather than injection molded parts as are commonly used in other assays. The assay was run both by professional users in a lab setting and lay users in the field. Methods A flow through rapid assay was produced and tested both by professional users in a lab setting using library sample and by lay users in a field setting using self collected samples. Results The results are shown in table 1, the assay performance was broadly similar for both whole blood and saliva, however the saliva dataset was much smaller for the professional use setting. The human factor analysis from the lay user field studies showed that they preferred the saliva collection procedure however for a small subset of subjects it proved difficult to collect sufficient sample for testing. For subjects who were able to collect adequate sample the results corresponded well between the device and lab-based ELISA. Conclusions The lay user experience validated test design and function and proved the usability in the field.","PeriodicalId":10690,"journal":{"name":"Clinical chemistry","volume":"27 1","pages":""},"PeriodicalIF":9.3,"publicationDate":"2024-10-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"142362989","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
期刊
Clinical chemistry
全部 Acc. Chem. Res. ACS Applied Bio Materials ACS Appl. Electron. Mater. ACS Appl. Energy Mater. ACS Appl. Mater. Interfaces ACS Appl. Nano Mater. ACS Appl. Polym. Mater. ACS BIOMATER-SCI ENG ACS Catal. ACS Cent. Sci. ACS Chem. Biol. ACS Chemical Health & Safety ACS Chem. Neurosci. ACS Comb. Sci. ACS Earth Space Chem. ACS Energy Lett. ACS Infect. Dis. ACS Macro Lett. ACS Mater. Lett. ACS Med. Chem. Lett. ACS Nano ACS Omega ACS Photonics ACS Sens. ACS Sustainable Chem. Eng. ACS Synth. Biol. Anal. Chem. BIOCHEMISTRY-US Bioconjugate Chem. BIOMACROMOLECULES Chem. Res. Toxicol. Chem. Rev. Chem. Mater. CRYST GROWTH DES ENERG FUEL Environ. Sci. Technol. Environ. Sci. Technol. Lett. Eur. J. Inorg. Chem. IND ENG CHEM RES Inorg. Chem. J. Agric. Food. Chem. J. Chem. Eng. Data J. Chem. Educ. J. Chem. Inf. Model. J. Chem. Theory Comput. J. Med. Chem. J. Nat. Prod. J PROTEOME RES J. Am. Chem. Soc. LANGMUIR MACROMOLECULES Mol. Pharmaceutics Nano Lett. Org. Lett. ORG PROCESS RES DEV ORGANOMETALLICS J. Org. Chem. J. Phys. Chem. J. Phys. Chem. A J. Phys. Chem. B J. Phys. Chem. C J. Phys. Chem. Lett. Analyst Anal. Methods Biomater. Sci. Catal. Sci. Technol. Chem. Commun. Chem. Soc. Rev. CHEM EDUC RES PRACT CRYSTENGCOMM Dalton Trans. Energy Environ. Sci. ENVIRON SCI-NANO ENVIRON SCI-PROC IMP ENVIRON SCI-WAT RES Faraday Discuss. Food Funct. Green Chem. Inorg. Chem. Front. Integr. Biol. J. Anal. At. Spectrom. J. Mater. Chem. A J. Mater. Chem. B J. Mater. Chem. C Lab Chip Mater. Chem. Front. Mater. Horiz. MEDCHEMCOMM Metallomics Mol. Biosyst. Mol. Syst. Des. Eng. Nanoscale Nanoscale Horiz. Nat. Prod. Rep. New J. Chem. Org. Biomol. Chem. Org. Chem. Front. PHOTOCH PHOTOBIO SCI PCCP Polym. Chem.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1