首页 > 最新文献

2020 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC)最新文献

英文 中文
Development and Validation of an Accurate Input Function from Carotid Arteries using the uEXPLORER 使用uEXPLORER开发和验证颈动脉的准确输入功能
Pub Date : 2020-10-31 DOI: 10.1109/NSS/MIC42677.2020.9508057
Tao Feng, Hongdi Li, Yizhang Zhao, N. Omidvari, Yang Lv, Elizabeth Li, Debin Hu, Y. Abdelhafez, J. Schmall, R. Badawi, S. Cherry
For a dedicated brain scan, the carotid artery is the best location for acquiring an image-based input function. With improvements in PET spatial resolution, accurate quantitation may be achieved with PET data alone. With the ability to cover both the carotid artery and the thorax at high spatial resolution, the uEXPLORER datasets provide a unique opportunity to develop and validate input functions in multiple regions such as the carotid artery. The regions containing the carotid arteries were first manually identified using reconstructed images consisting of the first 60 seconds of data post-injection. The image-based point spread function (PSF) was measured using off-center phantom scans to approximate the locations of the carotid artery. The same reconstruction approach was used for both the phantom scans and the volunteer scans. The structure of the carotid artery at each slice was generated using a deconvolution approach. An additional constraint of a uniform activity distribution within the carotid artery was added in the deconvolution approach. The acquired carotid artery structure was then applied to the dynamic frames (1-hour data) from volunteer scans for partial volume correction to acquire the input function (CA-IF). The input function from the descending aorta (DA-IF) was also extracted as a gold standard. The area-under-curve (AUC) ratio between the two input functions was used to evaluate the accuracy of the method. Without correction, there was a significant visual difference between CA-IF and the DA-IF, which was reduced dramatically after correction. The quantitation difference was dramatically reduced with the proposed correction method. The AUC ratio between the two input functions was 0.78+-0.04 (original), and was 1.00+-0.03 after correction, suggesting much improved quantitative accuracy. The results demonstrated that with improved image resolution and sensitivity, it is possible to accurately acquire the input function from carotid arteries without reliance on extra anatomical imaging approaches such as MRI.
对于专门的脑部扫描,颈动脉是获取基于图像的输入功能的最佳位置。随着PET空间分辨率的提高,仅用PET数据就可以实现准确的定量。由于能够以高空间分辨率覆盖颈动脉和胸腔,uEXPLORER数据集为开发和验证颈动脉等多个区域的输入功能提供了独特的机会。包含颈动脉的区域首先通过由注射后的前60秒数据组成的重建图像手动识别。基于图像的点扩散函数(PSF)是通过离中心幻像扫描来测量的,以近似于颈动脉的位置。幻像扫描和志愿者扫描采用了相同的重建方法。使用反卷积方法生成每个切片的颈动脉结构。在反褶积方法中增加了颈动脉内均匀活动分布的附加约束。然后将获得的颈动脉结构应用于志愿者扫描的动态帧(1小时数据),进行部分体积校正以获得输入函数(CA-IF)。提取降主动脉输入函数(DA-IF)作为金标准。使用两个输入函数之间的曲线下面积(AUC)比率来评估该方法的准确性。未经矫正,CA-IF与DA-IF之间存在显著的视觉差异,矫正后显著降低。提出的校正方法大大减小了定量差异。两个输入函数之间的AUC比值为0.78+-0.04(原始),校正后为1.00+-0.03,表明定量精度大大提高。结果表明,通过提高图像分辨率和灵敏度,可以准确地获取颈动脉的输入功能,而无需依赖额外的解剖成像方法,如MRI。
{"title":"Development and Validation of an Accurate Input Function from Carotid Arteries using the uEXPLORER","authors":"Tao Feng, Hongdi Li, Yizhang Zhao, N. Omidvari, Yang Lv, Elizabeth Li, Debin Hu, Y. Abdelhafez, J. Schmall, R. Badawi, S. Cherry","doi":"10.1109/NSS/MIC42677.2020.9508057","DOIUrl":"https://doi.org/10.1109/NSS/MIC42677.2020.9508057","url":null,"abstract":"For a dedicated brain scan, the carotid artery is the best location for acquiring an image-based input function. With improvements in PET spatial resolution, accurate quantitation may be achieved with PET data alone. With the ability to cover both the carotid artery and the thorax at high spatial resolution, the uEXPLORER datasets provide a unique opportunity to develop and validate input functions in multiple regions such as the carotid artery. The regions containing the carotid arteries were first manually identified using reconstructed images consisting of the first 60 seconds of data post-injection. The image-based point spread function (PSF) was measured using off-center phantom scans to approximate the locations of the carotid artery. The same reconstruction approach was used for both the phantom scans and the volunteer scans. The structure of the carotid artery at each slice was generated using a deconvolution approach. An additional constraint of a uniform activity distribution within the carotid artery was added in the deconvolution approach. The acquired carotid artery structure was then applied to the dynamic frames (1-hour data) from volunteer scans for partial volume correction to acquire the input function (CA-IF). The input function from the descending aorta (DA-IF) was also extracted as a gold standard. The area-under-curve (AUC) ratio between the two input functions was used to evaluate the accuracy of the method. Without correction, there was a significant visual difference between CA-IF and the DA-IF, which was reduced dramatically after correction. The quantitation difference was dramatically reduced with the proposed correction method. The AUC ratio between the two input functions was 0.78+-0.04 (original), and was 1.00+-0.03 after correction, suggesting much improved quantitative accuracy. The results demonstrated that with improved image resolution and sensitivity, it is possible to accurately acquire the input function from carotid arteries without reliance on extra anatomical imaging approaches such as MRI.","PeriodicalId":6760,"journal":{"name":"2020 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC)","volume":"55 1","pages":"1-4"},"PeriodicalIF":0.0,"publicationDate":"2020-10-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"84875516","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 2
Annihilation Photon Acolinearity with Ultra-fast ToF-PET 超快ToF-PET湮灭光子的非共线性
Pub Date : 2020-10-31 DOI: 10.1109/NSS/MIC42677.2020.9507968
Maxime Toussaint, R. Lecomte, J. Dussault
It was recently demonstrated that using ultra-fast Time-of-Flight (ToF) in positron emission tomography can mitigate the blur induced by the detector size. In another study, it was also observed that resolution beyond the expected Gaussian blur of annihilation photon acolinearity (APA) could be achieved, even though APA was not modeled in the system matrix. This incongruity is further investigated in this work. With perfect ToF resolution, the resulting response function is shown to be the product between the expected Gaussian APA and the 1/r function. A 2D scanner with 1-mm wide detectors and 2-mm ToF resolution is used to further explore the issue. Simulation of point sources shows that the reconstructed profiles were indeed not Gaussian. A Hot Spots phantom was also simulated to evaluate qualitatively the scanner resolving power and, again, resolution beyond the expected 0.0044R, R being the scanner radius, was achieved. These results highlight that the spatial response function induced by neglecting APA in the reconstruction scheme is sharper than expected, approaching an 1/r shape, when ultra-fast ToF is available.
最近的研究表明,在正电子发射断层扫描中使用超高速飞行时间(ToF)可以减轻探测器尺寸引起的模糊。在另一项研究中,也观察到即使APA没有在系统矩阵中建模,也可以实现超过湮灭光子共线性(APA)的预期高斯模糊的分辨率。在这项工作中进一步研究了这种不协调。在完美的ToF分辨率下,得到的响应函数是期望的高斯APA与1/r函数之间的乘积。使用1毫米宽检测器和2毫米ToF分辨率的2D扫描仪来进一步探索这个问题。对点源的仿真表明,重构的轮廓确实不是高斯分布。我们还模拟了一个Hot Spots幻影来定性地评估扫描仪的分辨率,并且再次达到了超出预期的0.0044R (R为扫描仪半径)的分辨率。这些结果表明,当有超高速ToF时,重构方案中忽略APA引起的空间响应函数比预期的更尖锐,接近1/r形状。
{"title":"Annihilation Photon Acolinearity with Ultra-fast ToF-PET","authors":"Maxime Toussaint, R. Lecomte, J. Dussault","doi":"10.1109/NSS/MIC42677.2020.9507968","DOIUrl":"https://doi.org/10.1109/NSS/MIC42677.2020.9507968","url":null,"abstract":"It was recently demonstrated that using ultra-fast Time-of-Flight (ToF) in positron emission tomography can mitigate the blur induced by the detector size. In another study, it was also observed that resolution beyond the expected Gaussian blur of annihilation photon acolinearity (APA) could be achieved, even though APA was not modeled in the system matrix. This incongruity is further investigated in this work. With perfect ToF resolution, the resulting response function is shown to be the product between the expected Gaussian APA and the 1/r function. A 2D scanner with 1-mm wide detectors and 2-mm ToF resolution is used to further explore the issue. Simulation of point sources shows that the reconstructed profiles were indeed not Gaussian. A Hot Spots phantom was also simulated to evaluate qualitatively the scanner resolving power and, again, resolution beyond the expected 0.0044R, R being the scanner radius, was achieved. These results highlight that the spatial response function induced by neglecting APA in the reconstruction scheme is sharper than expected, approaching an 1/r shape, when ultra-fast ToF is available.","PeriodicalId":6760,"journal":{"name":"2020 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC)","volume":"174 1","pages":"1-4"},"PeriodicalIF":0.0,"publicationDate":"2020-10-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"85184850","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 2
Deep Learning Image Transformation under Radon Transform Radon变换下的深度学习图像变换
Pub Date : 2020-10-31 DOI: 10.1109/NSS/MIC42677.2020.9507793
Haoran Chang, Rhodri L. Smith, S. Paisey, R. Boutchko, D. Mitra
Previously, we have shown that an image location, size, or even constant attenuation factor may be estimated by deep learning from the images Radon transformed representation. In this project, we go a step further to estimate a few other mathematical transformation parameters under Radon transformation. The motivation behind the project is that many medical imaging problems are related to estimating similar invariance parameters. Such estimations are typically performed after image reconstruction from detector images that are in the Radon transformed space. The image reconstruction process introduces additional noise of its own. Deep learning provides a framework for direct estimation of required information from the detector images. A specific case we are interested in is dynamic nuclear imaging, where the quantitative estimations of the target tissues are queried. Motion inherent in biological systems, e.g., in vivo imaging with breathing motion, may be modeled as a transformation in the spatial domain. Motion is particularly prevalent in dynamic imaging, while tracer dynamics in the imaged object are a second source of transformation in the time domain. Our neural network model attempts to discern the two types of transformation (motion and intensity variation dynamics), i.e., tries to learn one type of transformation, ignoring the other.
之前,我们已经证明,通过深度学习图像的Radon变换表示,可以估计图像的位置、大小,甚至常数衰减因子。在这个项目中,我们进一步估计了Radon变换下的其他一些数学变换参数。这个项目背后的动机是许多医学成像问题都与估计相似的不变性参数有关。这种估计通常是在Radon变换空间中的探测器图像重建后进行的。图像重建过程本身也引入了额外的噪声。深度学习为从检测器图像中直接估计所需信息提供了一个框架。我们感兴趣的一个具体案例是动态核成像,其中目标组织的定量估计被查询。生物系统中固有的运动,例如,具有呼吸运动的体内成像,可以建模为空间域中的变换。运动在动态成像中特别普遍,而成像对象中的示踪动态是时域变换的第二个来源。我们的神经网络模型试图辨别两种类型的转换(运动和强度变化动力学),即试图学习一种类型的转换,忽略另一种。
{"title":"Deep Learning Image Transformation under Radon Transform","authors":"Haoran Chang, Rhodri L. Smith, S. Paisey, R. Boutchko, D. Mitra","doi":"10.1109/NSS/MIC42677.2020.9507793","DOIUrl":"https://doi.org/10.1109/NSS/MIC42677.2020.9507793","url":null,"abstract":"Previously, we have shown that an image location, size, or even constant attenuation factor may be estimated by deep learning from the images Radon transformed representation. In this project, we go a step further to estimate a few other mathematical transformation parameters under Radon transformation. The motivation behind the project is that many medical imaging problems are related to estimating similar invariance parameters. Such estimations are typically performed after image reconstruction from detector images that are in the Radon transformed space. The image reconstruction process introduces additional noise of its own. Deep learning provides a framework for direct estimation of required information from the detector images. A specific case we are interested in is dynamic nuclear imaging, where the quantitative estimations of the target tissues are queried. Motion inherent in biological systems, e.g., in vivo imaging with breathing motion, may be modeled as a transformation in the spatial domain. Motion is particularly prevalent in dynamic imaging, while tracer dynamics in the imaged object are a second source of transformation in the time domain. Our neural network model attempts to discern the two types of transformation (motion and intensity variation dynamics), i.e., tries to learn one type of transformation, ignoring the other.","PeriodicalId":6760,"journal":{"name":"2020 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC)","volume":"37 1","pages":"1-3"},"PeriodicalIF":0.0,"publicationDate":"2020-10-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"82189389","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 2
High Resolution Combined Neutron/X-ray Imaging Detector 高分辨率中子/ x射线联合成像探测器
Pub Date : 2020-10-31 DOI: 10.1109/NSS/MIC42677.2020.9507745
V. Nagarkar, S. Miller, M. Marshall, C. Brown, C. Sosa, Bipin Singh, L. D'Aries
Federal agencies including DOE (NNSA), CWMD, and DTRA all play a vital role in the U.S. government's efforts to prevent, counter, and respond to a terrorist or other adversary with a nuclear or radiological device. To support its functions and foundational capabilities across nonproliferation, counterterrorism, and emergency response mission areas, these agencies need advanced sensors and instrumentation that may be used to rapidly identify threats and get the information needed to plan a response. At RMD we are developing the tools and techniques that will address some of these mission needs. One ongoing effort involves the development of a portable, high spatial resolution, combined neutron/X-ray radiography detector. The detector is capable of high sensitivity imaging of fast neutrons, thermal neutrons, and high energy X-rays, and will permit data analysis for material identification purposes. The key enabling technology for such a detector is the fabrication of a large area scintillator converter that can simultaneously detect fast neutrons, thermal neutrons, and hard X-rays, with the highest possible efficiency. To preserve spatial resolution the scintillator is microstructured to minimize the traditional trade-off between the detection efficiency and spatial resolution. Additionally, the scintillator should be sufficiently bright for enhanced signal-to-noise-ratio (SNR), and its fabrication method should be amenable for producing large area sensors. We are investigating several approaches to realize such sensors. Here, we report preliminary results on a portable flat panel based large area detector measuring 25 cm × 30 cm in active imaging area, and an intrinsic spatial resolution of 139 µm. The report also includes scintillator design approaches, GEANT4 simulations of scintillator response to various radiation types, and experimental imaging data taken using thermal neutrons, fast neutrons between 1 MeV to 40 MeV, and hard X-rays up to 400 kV. The performance of the detector as a whole has been evaluated at ORNL and NIST neutron beam lines.
联邦机构,包括美国能源部(NNSA)、化学武器防御中心(CWMD)和DTRA,都在美国政府预防、反击和应对恐怖分子或其他拥有核或放射性装置的对手的努力中发挥着至关重要的作用。为了支持其在防扩散、反恐和应急任务领域的职能和基本能力,这些机构需要先进的传感器和仪器,可用于快速识别威胁并获得计划应对所需的信息。在RMD,我们正在开发工具和技术来解决这些任务的一些需求。目前正在进行的一项工作涉及开发便携式、高空间分辨率、中子/ x射线照相组合探测器。该探测器能够对快中子、热中子和高能x射线进行高灵敏度成像,并将允许对材料识别目的进行数据分析。这种探测器的关键技术是制造一个大面积闪烁体转换器,可以同时检测快中子、热中子和硬x射线,并具有最高的效率。为了保持空间分辨率,闪烁体被微结构,以尽量减少传统的检测效率和空间分辨率之间的权衡。此外,闪烁体应足够明亮,以提高信噪比(SNR),其制造方法应适用于生产大面积传感器。我们正在研究几种实现这种传感器的方法。在这里,我们报告了基于便携式平板的大面积探测器的初步结果,该探测器的有效成像区域为25 cm × 30 cm,固有空间分辨率为139 μ m。该报告还包括闪烁体设计方法,GEANT4模拟闪烁体对各种辐射类型的响应,以及使用热中子,1 MeV至40 MeV之间的快中子和高达400 kV的硬x射线获得的实验成像数据。在ORNL和NIST中子束流线上对探测器的整体性能进行了评估。
{"title":"High Resolution Combined Neutron/X-ray Imaging Detector","authors":"V. Nagarkar, S. Miller, M. Marshall, C. Brown, C. Sosa, Bipin Singh, L. D'Aries","doi":"10.1109/NSS/MIC42677.2020.9507745","DOIUrl":"https://doi.org/10.1109/NSS/MIC42677.2020.9507745","url":null,"abstract":"Federal agencies including DOE (NNSA), CWMD, and DTRA all play a vital role in the U.S. government's efforts to prevent, counter, and respond to a terrorist or other adversary with a nuclear or radiological device. To support its functions and foundational capabilities across nonproliferation, counterterrorism, and emergency response mission areas, these agencies need advanced sensors and instrumentation that may be used to rapidly identify threats and get the information needed to plan a response. At RMD we are developing the tools and techniques that will address some of these mission needs. One ongoing effort involves the development of a portable, high spatial resolution, combined neutron/X-ray radiography detector. The detector is capable of high sensitivity imaging of fast neutrons, thermal neutrons, and high energy X-rays, and will permit data analysis for material identification purposes. The key enabling technology for such a detector is the fabrication of a large area scintillator converter that can simultaneously detect fast neutrons, thermal neutrons, and hard X-rays, with the highest possible efficiency. To preserve spatial resolution the scintillator is microstructured to minimize the traditional trade-off between the detection efficiency and spatial resolution. Additionally, the scintillator should be sufficiently bright for enhanced signal-to-noise-ratio (SNR), and its fabrication method should be amenable for producing large area sensors. We are investigating several approaches to realize such sensors. Here, we report preliminary results on a portable flat panel based large area detector measuring 25 cm × 30 cm in active imaging area, and an intrinsic spatial resolution of 139 µm. The report also includes scintillator design approaches, GEANT4 simulations of scintillator response to various radiation types, and experimental imaging data taken using thermal neutrons, fast neutrons between 1 MeV to 40 MeV, and hard X-rays up to 400 kV. The performance of the detector as a whole has been evaluated at ORNL and NIST neutron beam lines.","PeriodicalId":6760,"journal":{"name":"2020 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC)","volume":"3 1","pages":"1-3"},"PeriodicalIF":0.0,"publicationDate":"2020-10-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"79509555","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Performance Benefits of Extending the AFOV of PET Scanners 扩展PET扫描仪的AFOV的性能优势
Pub Date : 2020-10-31 DOI: 10.1109/NSS/MIC42677.2020.9507858
V. Viswanath, M. Daube-Witherspoon, A. Pantel, M. Parma, M. Werner, J. Karp
Long axial field-of-view (AFOV) PET scanners have valuable benefits for both clinical and research applications. Thus far, two of these scanners are currently operational in the US: the 194-cm uExplorer at UC Davis and the PennPET Explorer at the University of Pennsylvania. We had previously reported performance metrics and human imaging studies on the 64-cm PennPET Explorer and have recently completed extending the AFOV of the scanner to a 5-ring, 112-cm system. We extended the NEMA metrics to scanners longer than 65-cm and performed sensitivity, count rate, spatial resolution, and contrast recovery measurements on the 5-ring system. The sensitivity of the system was 104 kcps/MBq; the peak NEC measured with a 20×70 cm count rate phantom was 1.6 Mcps @ 39 kBq/cc; the axial spatial resolution degraded slightly for the 62° acceptance angle; and the contrast recovery did not degrade as a function of increased axial acceptance angle. Therefore, extending the axial length of the PennPET Explorer from 64 cm to 112 cm, while expanding the acceptance angle, resulted in gains in sensitivity and count rate with minimal degradations in spatial resolution and no degradation of contrast recovery.
长轴视场(AFOV) PET扫描仪对临床和研究应用都有宝贵的好处。到目前为止,美国有两台这样的扫描仪:加州大学戴维斯分校的194厘米uExplorer和宾夕法尼亚大学的PennPET Explorer。我们之前报道了64厘米PennPET Explorer的性能指标和人体成像研究,最近完成了将扫描仪的AFOV扩展到5环112厘米系统。我们将NEMA指标扩展到长度超过65厘米的扫描仪上,并对5环系统进行了灵敏度、计数率、空间分辨率和对比度恢复测量。该系统灵敏度为104 kcps/MBq;用20×70 cm计数率模体测量的NEC峰值为1.6 Mcps @ 39 kBq/cc;当接收角为62°时,轴向空间分辨率略有下降;对比度恢复不随轴向接收角的增加而降低。因此,将PennPET Explorer的轴向长度从64 cm延长到112 cm,同时扩大接收角,在最小的空间分辨率下降和对比度恢复没有下降的情况下,获得了灵敏度和计数率的提高。
{"title":"Performance Benefits of Extending the AFOV of PET Scanners","authors":"V. Viswanath, M. Daube-Witherspoon, A. Pantel, M. Parma, M. Werner, J. Karp","doi":"10.1109/NSS/MIC42677.2020.9507858","DOIUrl":"https://doi.org/10.1109/NSS/MIC42677.2020.9507858","url":null,"abstract":"Long axial field-of-view (AFOV) PET scanners have valuable benefits for both clinical and research applications. Thus far, two of these scanners are currently operational in the US: the 194-cm uExplorer at UC Davis and the PennPET Explorer at the University of Pennsylvania. We had previously reported performance metrics and human imaging studies on the 64-cm PennPET Explorer and have recently completed extending the AFOV of the scanner to a 5-ring, 112-cm system. We extended the NEMA metrics to scanners longer than 65-cm and performed sensitivity, count rate, spatial resolution, and contrast recovery measurements on the 5-ring system. The sensitivity of the system was 104 kcps/MBq; the peak NEC measured with a 20×70 cm count rate phantom was 1.6 Mcps @ 39 kBq/cc; the axial spatial resolution degraded slightly for the 62° acceptance angle; and the contrast recovery did not degrade as a function of increased axial acceptance angle. Therefore, extending the axial length of the PennPET Explorer from 64 cm to 112 cm, while expanding the acceptance angle, resulted in gains in sensitivity and count rate with minimal degradations in spatial resolution and no degradation of contrast recovery.","PeriodicalId":6760,"journal":{"name":"2020 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC)","volume":"110 1","pages":"1-7"},"PeriodicalIF":0.0,"publicationDate":"2020-10-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"79642590","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 2
PET Reconstruction with a Spatially Varying Point Spread Function for a Brain Dedicated PET Insert for PET/MR 基于空间变点扩展函数的脑专用PET/MR插入体PET重建
Pub Date : 2020-10-31 DOI: 10.1109/NSS/MIC42677.2020.9507741
Zahra Ashouri, A. Groll, C. Levin
Including accurate modeling of the point spread function (PSF) in positron emission tomography (PET) reconstruction algorithms results in improvements in image spatial resolution and contrast. In this work, we sampled the PSF in our first-generation radio-frequency brain dedicated PET insert for simultaneous PET/MR imaging using a 100 µCi NEMA standard 250 µm diameter Na-22 point source at 13 different positions within a subsection of the system field of view (FoV). The acquired list mode data was converted into the canonical sinogram format from which the spatial positioning of the source and standard deviations were calculated. The subset was then used to extrapolate the PSF for the full system FoV. This model was then fed as an input parameter into a graphical processing unit based ordered subset expectation maximization (OSEM) reconstruction algorithm and used to generate reconstructed images with and without spatially varying PSF modeling for the Na-22 point source and a Hoffman brain phantom. Results indicate that for point source reconstruction, the FWHM of the horizontal profile of the point source is smaller with spatially variant PSF especially closer to the edges. Effect of spatially varying PSF modeling is also presented with Hoffman phantom reconstruction and CNR value has increased with spatially varying PSF.
在正电子发射断层扫描(PET)重建算法中引入点扩展函数(PSF)的精确建模,可以提高图像的空间分辨率和对比度。在这项工作中,我们在第一代射频脑专用PET插入物中取样PSF,使用100µCi NEMA标准250µm直径的Na-22点源在系统视场(FoV)分段内的13个不同位置同时进行PET/MR成像。将获取的列表模式数据转换为标准正弦图格式,计算源的空间定位和标准差。然后使用该子集来推断整个系统FoV的PSF。然后将该模型作为输入参数输入到基于图形处理单元的有序子集期望最大化(OSEM)重建算法中,并用于生成Na-22点源和Hoffman脑幻像的具有和不具有空间变化PSF建模的重建图像。结果表明,对于点源重建,点源水平剖面的波峰宽较小,且波峰宽随空间变化而变化,尤其是靠近边缘的波峰宽。通过Hoffman幻像重建,我们还看到了空间变化的PSF建模效果,并且CNR值随PSF空间变化而增加。
{"title":"PET Reconstruction with a Spatially Varying Point Spread Function for a Brain Dedicated PET Insert for PET/MR","authors":"Zahra Ashouri, A. Groll, C. Levin","doi":"10.1109/NSS/MIC42677.2020.9507741","DOIUrl":"https://doi.org/10.1109/NSS/MIC42677.2020.9507741","url":null,"abstract":"Including accurate modeling of the point spread function (PSF) in positron emission tomography (PET) reconstruction algorithms results in improvements in image spatial resolution and contrast. In this work, we sampled the PSF in our first-generation radio-frequency brain dedicated PET insert for simultaneous PET/MR imaging using a 100 µCi NEMA standard 250 µm diameter Na-22 point source at 13 different positions within a subsection of the system field of view (FoV). The acquired list mode data was converted into the canonical sinogram format from which the spatial positioning of the source and standard deviations were calculated. The subset was then used to extrapolate the PSF for the full system FoV. This model was then fed as an input parameter into a graphical processing unit based ordered subset expectation maximization (OSEM) reconstruction algorithm and used to generate reconstructed images with and without spatially varying PSF modeling for the Na-22 point source and a Hoffman brain phantom. Results indicate that for point source reconstruction, the FWHM of the horizontal profile of the point source is smaller with spatially variant PSF especially closer to the edges. Effect of spatially varying PSF modeling is also presented with Hoffman phantom reconstruction and CNR value has increased with spatially varying PSF.","PeriodicalId":6760,"journal":{"name":"2020 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC)","volume":"37 1","pages":"1-3"},"PeriodicalIF":0.0,"publicationDate":"2020-10-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"84355498","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Neutron and Proton-Induced Radiation Damage in LuAG Scintillating Ceramics LuAG闪烁陶瓷的中子和质子辐射损伤
Pub Date : 2020-10-31 DOI: 10.1109/NSS/MIC42677.2020.9507969
Chen Hu, Jiang Li, B. Jiang, Liyuan Zhang, R. Zhu
Because of its potential low cost, bright and fast LuAG scintillating ceramics have attracted a broad interest in the HEP community. One crucial issue for its application at future high energy colliders is its radiation hardness against hadrons, including both neutrons and protons. LuAG ceramics were irradiated at LANSCE and CERN up to 6.7×1015neq/cm2 and 1.2×1015 p/cm2 respectively, and are found to have a factor of two better radiation hardness than LYSO crystals. Ca2+ codoping in LuAG ceramics improves the fast to total ratio and the radiation hardness against hadrons.
由于其潜在的低成本,明亮和快速的LuAG闪烁陶瓷引起了HEP界的广泛兴趣。在未来高能对撞机中应用的一个关键问题是它对强子(包括中子和质子)的辐射硬度。LuAG陶瓷在LANSCE和CERN的辐照强度分别为6.7×1015neq/cm2和1.2×1015 p/cm2,其辐射硬度比LYSO晶体高2倍。Ca2+共掺杂提高了LuAG陶瓷的快总比和抗强子的辐射硬度。
{"title":"Neutron and Proton-Induced Radiation Damage in LuAG Scintillating Ceramics","authors":"Chen Hu, Jiang Li, B. Jiang, Liyuan Zhang, R. Zhu","doi":"10.1109/NSS/MIC42677.2020.9507969","DOIUrl":"https://doi.org/10.1109/NSS/MIC42677.2020.9507969","url":null,"abstract":"Because of its potential low cost, bright and fast LuAG scintillating ceramics have attracted a broad interest in the HEP community. One crucial issue for its application at future high energy colliders is its radiation hardness against hadrons, including both neutrons and protons. LuAG ceramics were irradiated at LANSCE and CERN up to 6.7×1015neq/cm2 and 1.2×1015 p/cm2 respectively, and are found to have a factor of two better radiation hardness than LYSO crystals. Ca2+ codoping in LuAG ceramics improves the fast to total ratio and the radiation hardness against hadrons.","PeriodicalId":6760,"journal":{"name":"2020 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC)","volume":"81 6 1","pages":"1-2"},"PeriodicalIF":0.0,"publicationDate":"2020-10-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"85310162","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 1
Investigation and Mitigation of Crosstalk in the Prototype ME0 GEM Detector for the Phase-2 Muon System Upgrade of the CMS Experiment: On behalf of the CMS Muon Group CMS实验第二阶段μ子系统升级中ME0 GEM探测器原型串扰的研究与缓解:代表CMS μ子组
Pub Date : 2020-10-31 DOI: 10.1109/NSS/MIC42677.2020.9508058
S. Butalla, M. Hohlmann
The LHe is currently undergoing a high luminosity upgrade, which is set to increase the instantaneous luminosity by at least a factor of five. This luminosity increase will result in a higher muon flux rate in the forward region and overwhelm the current trigger system of the CMS experiment. The MEO, a gas electron multiplier detector, is proposed for the Phase-2 Muon System Upgrade for the CMS experiment to help increase the muon acceptance and to control the Level 1 muon trigger rate. A recent design iteration of this detector features GEM foils that are segmented on both sides, which helps to lower the probability of high voltage discharges. However, during preliminary testing of the chamber, substantial crosstalk between readout sectors was observed. Here, we investigate, characterize, and quantify the crosstalk present in the detector, and also estimate the performance of the chamber as a result of this crosstalk via simulation results of the detector dead time, efficiency loss, and frontend electronics response. The results of crosstalk via signals produced by applying a square voltage pulse directly on the readout strips of the detector with a signal generator are summarized. We also present the efficacy of mitigation strategies including bypass capacitors and increasing the area of the HV segments on the third GEM foil in the detector. We find that the crosstalk is a result of capacitive coupling between the readout strips on the readout board and between the readout strips and the bottom of the third GEM foil. Our results show that the crosstalk generally follows a pattern where the largest magnitude of crosstalk is within the same azimuthal readout segment in the detector, and in the next-nearest horizontal segments in eta. Generally, the bypass capacitors and increased area of the HV segments successfully lower the crosstalk in the sectors where they are located; on average, we observe a maximum decrease of crosstalk in sectors previously experiencing crosstalk from (1.66±0.03)% to (1.11±0.02)% with all HV segments connected in parallel on the bottom of the third GEM foil, with the addition of an HV low-pass filter connected to this electrode, and an HV divider. However, with these mitigation strategies, we also observe slightly increased crosstalk (≨ 0.4%) in readout sectors farther away.
LHe目前正在进行高亮度升级,这将使瞬时亮度增加至少五倍。这种光度的增加将导致前向区的μ子通量率增大,使CMS实验的电流触发系统不堪重负。提出了一种气体电子倍增探测器MEO,用于CMS实验的第二阶段μ子系统升级,以帮助提高μ子接受度并控制1级μ子触发率。该检测器的最新设计迭代具有两面分段的GEM箔,这有助于降低高压放电的概率。然而,在室的初步测试期间,观察到读数扇区之间存在大量串扰。在这里,我们研究,表征和量化检测器中存在的串扰,并通过检测器死区时间,效率损失和前端电子响应的模拟结果估计该串扰导致的腔室性能。总结了用信号发生器在检波器读出带上直接施加方形电压脉冲所产生的信号串扰的结果。我们还介绍了缓解策略的有效性,包括旁路电容器和增加探测器中第三个GEM箔上的HV段的面积。我们发现串扰是读出板上读出带之间以及读出带与第三片GEM箔底部之间电容耦合的结果。我们的结果表明,串扰通常遵循这样一种模式,即最大的串扰在探测器的相同方位读出段内,以及在eta中最近的水平段内。一般来说,旁路电容器和增加高压段的面积成功地降低了它们所在扇区的串扰;平均而言,我们观察到先前经历串扰的扇区的串扰最大减少,从(1.66±0.03)%降至(1.11±0.02)%,所有HV段平行连接在第三个GEM箔的底部,并在该电极上添加一个HV低通滤波器和一个HV分压器。然而,通过这些缓解策略,我们还观察到更远的读出扇区的串扰略有增加(≨0.4%)。
{"title":"Investigation and Mitigation of Crosstalk in the Prototype ME0 GEM Detector for the Phase-2 Muon System Upgrade of the CMS Experiment: On behalf of the CMS Muon Group","authors":"S. Butalla, M. Hohlmann","doi":"10.1109/NSS/MIC42677.2020.9508058","DOIUrl":"https://doi.org/10.1109/NSS/MIC42677.2020.9508058","url":null,"abstract":"The LHe is currently undergoing a high luminosity upgrade, which is set to increase the instantaneous luminosity by at least a factor of five. This luminosity increase will result in a higher muon flux rate in the forward region and overwhelm the current trigger system of the CMS experiment. The MEO, a gas electron multiplier detector, is proposed for the Phase-2 Muon System Upgrade for the CMS experiment to help increase the muon acceptance and to control the Level 1 muon trigger rate. A recent design iteration of this detector features GEM foils that are segmented on both sides, which helps to lower the probability of high voltage discharges. However, during preliminary testing of the chamber, substantial crosstalk between readout sectors was observed. Here, we investigate, characterize, and quantify the crosstalk present in the detector, and also estimate the performance of the chamber as a result of this crosstalk via simulation results of the detector dead time, efficiency loss, and frontend electronics response. The results of crosstalk via signals produced by applying a square voltage pulse directly on the readout strips of the detector with a signal generator are summarized. We also present the efficacy of mitigation strategies including bypass capacitors and increasing the area of the HV segments on the third GEM foil in the detector. We find that the crosstalk is a result of capacitive coupling between the readout strips on the readout board and between the readout strips and the bottom of the third GEM foil. Our results show that the crosstalk generally follows a pattern where the largest magnitude of crosstalk is within the same azimuthal readout segment in the detector, and in the next-nearest horizontal segments in eta. Generally, the bypass capacitors and increased area of the HV segments successfully lower the crosstalk in the sectors where they are located; on average, we observe a maximum decrease of crosstalk in sectors previously experiencing crosstalk from (1.66±0.03)% to (1.11±0.02)% with all HV segments connected in parallel on the bottom of the third GEM foil, with the addition of an HV low-pass filter connected to this electrode, and an HV divider. However, with these mitigation strategies, we also observe slightly increased crosstalk (≨ 0.4%) in readout sectors farther away.","PeriodicalId":6760,"journal":{"name":"2020 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC)","volume":"75 1","pages":"1-6"},"PeriodicalIF":0.0,"publicationDate":"2020-10-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"83803483","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 5
Geometry Optimization of Muon Production Graphite Target by 600 MeV Proton Beam 600 MeV质子束产生μ子石墨靶的几何优化
Pub Date : 2020-10-31 DOI: 10.1109/NSS/MIC42677.2020.9508097
J. Jeong, Jae Chang Kim, J. Son, K. Pak, Yong Kyun Kim, Ju Hahn Lee
The Rare Isotope Science Project was launched in December 2011, and a heavy-ion accelerator complex in Korea, named RAON, has been designed, including a muon facility for muon spin rotation, relaxation, and resonance (μSR). In this study, the graphite target in RAON was designed to have a rotating ring shape and was cooled by radiative heat transfer, which presents advantages in the cool-down process such as a low-temperature gradient in the target and no necessity of a liquid coolant-cooling system. Monte-Carlo simulations and ANSYS calculations were performed to optimize the proton beam size and the dimensions of the target to produce a sufficient number of surface muons in a thermally stable condition. A comparison between the simulation and the experimental data was also included in this paper in order to obtain a reliable result. The expected number of surface muons was 6.942×108 with a 100 kW proton beam and Δp/p~5%. The maximum temperature was 2012 °K and the maximum stress in the target was 8.1598 kPa with the 400 kW proton beam, which guarantees safety during the replacement cycle of the target.
2011年12月启动了“稀有同位素科学事业”,在韩国设计了重离子加速器“RAON”,其中包括μ子自旋、弛豫、共振(μSR)的μ子设施。在本研究中,RAON中的石墨靶被设计成一个旋转的环形,并通过辐射传热进行冷却,这在冷却过程中具有靶内温度梯度低,不需要液冷冷却系统的优点。通过Monte-Carlo模拟和ANSYS计算,优化了质子束的尺寸和靶体的尺寸,以在热稳定的条件下产生足够数量的表面μ子。为了得到可靠的结果,本文还将仿真数据与实验数据进行了比较。在100 kW质子束和Δp/p~5%的条件下,表面介子的期望数为6.942×108。最高温度为2012°K,靶内最大应力为8.1598 kPa,质子束功率为400 kW,保证了靶在更换周期内的安全性。
{"title":"Geometry Optimization of Muon Production Graphite Target by 600 MeV Proton Beam","authors":"J. Jeong, Jae Chang Kim, J. Son, K. Pak, Yong Kyun Kim, Ju Hahn Lee","doi":"10.1109/NSS/MIC42677.2020.9508097","DOIUrl":"https://doi.org/10.1109/NSS/MIC42677.2020.9508097","url":null,"abstract":"The Rare Isotope Science Project was launched in December 2011, and a heavy-ion accelerator complex in Korea, named RAON, has been designed, including a muon facility for muon spin rotation, relaxation, and resonance (μSR). In this study, the graphite target in RAON was designed to have a rotating ring shape and was cooled by radiative heat transfer, which presents advantages in the cool-down process such as a low-temperature gradient in the target and no necessity of a liquid coolant-cooling system. Monte-Carlo simulations and ANSYS calculations were performed to optimize the proton beam size and the dimensions of the target to produce a sufficient number of surface muons in a thermally stable condition. A comparison between the simulation and the experimental data was also included in this paper in order to obtain a reliable result. The expected number of surface muons was 6.942×108 with a 100 kW proton beam and Δp/p~5%. The maximum temperature was 2012 °K and the maximum stress in the target was 8.1598 kPa with the 400 kW proton beam, which guarantees safety during the replacement cycle of the target.","PeriodicalId":6760,"journal":{"name":"2020 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC)","volume":"7 1","pages":"1-2"},"PeriodicalIF":0.0,"publicationDate":"2020-10-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"80543825","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Approaches for Measuring Fast Modulation of Optical Properties Induced by 511keV Photon Interactions 511keV光子相互作用诱导的光学特性快速调制测量方法
Pub Date : 2020-10-31 DOI: 10.1109/NSS/MIC42677.2020.9508093
Diana Jeong, Li Tao, Zander Adams, Yushin Kim, C. Levin
With a goal to achieve less than 10 picosecond (ps) coincidence time resolution for advancing time-of-flight positron emission tomography (TOF-PET), we seek to leverage the effects of transient free carriers generated from 511 keV interactions in a material. Most ultrafast optics methods rely on many excitation particles with a priori information on their arrival times. However, in PET it is critical to find a way to sensitively detect single, randomly emitted 511 keV annihilation photons from the ionization it produces. We investigate two ways to increase the signal: employing probe laser wavelength in the mid-IR and confining the probe beam to match the size of ionization tracks created by the 511 keV photons. We observed a continued decrease of 2% in transmission of the mid-IR probe beam (8μm) in a 10 X 10 X 1 mm3 Cadmium Telluride crystal with 511 keV annihilation irradiation over 300 seconds. The trend was reversed when the radiation source is removed, and reached the previous baseline transmission in 800 seconds. For confining the interaction volume to match the probe beam diameter, we revised a previously investigated interferometry approach to use a focused beam and did not apply a bias voltage across the detector crystal. With a 405nm laser diode as an excitation source, a shift of 43% in transmitted signal voltage was observed, and with Am-241 alpha source irradiation, a 1% shift was measured. With the improved sensitivity, these findings provide insights into possible methods of detecting single 511 keV events.
为了实现小于10皮秒(ps)的符合时间分辨率,以推进飞行时间正电子发射断层扫描(TOF-PET),我们试图利用材料中511 keV相互作用产生的瞬态自由载流子的影响。大多数超快光学方法依赖于许多具有其到达时间先验信息的激发粒子。然而,在PET中,找到一种方法来灵敏地探测它产生的电离中随机发射的511 keV湮灭光子是至关重要的。我们研究了两种增加信号的方法:采用中红外探测激光波长和限制探测光束以匹配511 keV光子产生的电离轨迹的大小。我们观察到在511 keV湮灭辐照300秒后,中红外探测光束(8μm)在10 X 10 X 1 mm3碲化镉晶体中的透射率持续下降2%。当辐射源被移除后,这一趋势发生了逆转,在800秒内达到了之前的基线传输。为了限制相互作用体积以匹配探测光束直径,我们修改了先前研究的干涉测量方法,使用聚焦光束,并且不在探测器晶体上施加偏置电压。以405nm激光二极管作为激发源,观察到传输信号电压偏移43%,在Am-241 α源照射下,测量到传输信号电压偏移1%。随着灵敏度的提高,这些发现为检测单个511 keV事件的可能方法提供了见解。
{"title":"Approaches for Measuring Fast Modulation of Optical Properties Induced by 511keV Photon Interactions","authors":"Diana Jeong, Li Tao, Zander Adams, Yushin Kim, C. Levin","doi":"10.1109/NSS/MIC42677.2020.9508093","DOIUrl":"https://doi.org/10.1109/NSS/MIC42677.2020.9508093","url":null,"abstract":"With a goal to achieve less than 10 picosecond (ps) coincidence time resolution for advancing time-of-flight positron emission tomography (TOF-PET), we seek to leverage the effects of transient free carriers generated from 511 keV interactions in a material. Most ultrafast optics methods rely on many excitation particles with a priori information on their arrival times. However, in PET it is critical to find a way to sensitively detect single, randomly emitted 511 keV annihilation photons from the ionization it produces. We investigate two ways to increase the signal: employing probe laser wavelength in the mid-IR and confining the probe beam to match the size of ionization tracks created by the 511 keV photons. We observed a continued decrease of 2% in transmission of the mid-IR probe beam (8μm) in a 10 X 10 X 1 mm3 Cadmium Telluride crystal with 511 keV annihilation irradiation over 300 seconds. The trend was reversed when the radiation source is removed, and reached the previous baseline transmission in 800 seconds. For confining the interaction volume to match the probe beam diameter, we revised a previously investigated interferometry approach to use a focused beam and did not apply a bias voltage across the detector crystal. With a 405nm laser diode as an excitation source, a shift of 43% in transmitted signal voltage was observed, and with Am-241 alpha source irradiation, a 1% shift was measured. With the improved sensitivity, these findings provide insights into possible methods of detecting single 511 keV events.","PeriodicalId":6760,"journal":{"name":"2020 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC)","volume":"402 1","pages":"1-3"},"PeriodicalIF":0.0,"publicationDate":"2020-10-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"80597500","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
期刊
2020 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC)
全部 Acc. Chem. Res. ACS Applied Bio Materials ACS Appl. Electron. Mater. ACS Appl. Energy Mater. ACS Appl. Mater. Interfaces ACS Appl. Nano Mater. ACS Appl. Polym. Mater. ACS BIOMATER-SCI ENG ACS Catal. ACS Cent. Sci. ACS Chem. Biol. ACS Chemical Health & Safety ACS Chem. Neurosci. ACS Comb. Sci. ACS Earth Space Chem. ACS Energy Lett. ACS Infect. Dis. ACS Macro Lett. ACS Mater. Lett. ACS Med. Chem. Lett. ACS Nano ACS Omega ACS Photonics ACS Sens. ACS Sustainable Chem. Eng. ACS Synth. Biol. Anal. Chem. BIOCHEMISTRY-US Bioconjugate Chem. BIOMACROMOLECULES Chem. Res. Toxicol. Chem. Rev. Chem. Mater. CRYST GROWTH DES ENERG FUEL Environ. Sci. Technol. Environ. Sci. Technol. Lett. Eur. J. Inorg. Chem. IND ENG CHEM RES Inorg. Chem. J. Agric. Food. Chem. J. Chem. Eng. Data J. Chem. Educ. J. Chem. Inf. Model. J. Chem. Theory Comput. J. Med. Chem. J. Nat. Prod. J PROTEOME RES J. Am. Chem. Soc. LANGMUIR MACROMOLECULES Mol. Pharmaceutics Nano Lett. Org. Lett. ORG PROCESS RES DEV ORGANOMETALLICS J. Org. Chem. J. Phys. Chem. J. Phys. Chem. A J. Phys. Chem. B J. Phys. Chem. C J. Phys. Chem. Lett. Analyst Anal. Methods Biomater. Sci. Catal. Sci. Technol. Chem. Commun. Chem. Soc. Rev. CHEM EDUC RES PRACT CRYSTENGCOMM Dalton Trans. Energy Environ. Sci. ENVIRON SCI-NANO ENVIRON SCI-PROC IMP ENVIRON SCI-WAT RES Faraday Discuss. Food Funct. Green Chem. Inorg. Chem. Front. Integr. Biol. J. Anal. At. Spectrom. J. Mater. Chem. A J. Mater. Chem. B J. Mater. Chem. C Lab Chip Mater. Chem. Front. Mater. Horiz. MEDCHEMCOMM Metallomics Mol. Biosyst. Mol. Syst. Des. Eng. Nanoscale Nanoscale Horiz. Nat. Prod. Rep. New J. Chem. Org. Biomol. Chem. Org. Chem. Front. PHOTOCH PHOTOBIO SCI PCCP Polym. Chem.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1