首页 > 最新文献

Biomedical Physics & Engineering Express最新文献

英文 中文
Impact of geometric correction on echo-planar imaging-based apparent diffusion coefficient maps for abdominal radiotherapy. 几何校正对基于回声平面成像的腹部放射治疗表观扩散系数图的影响。
IF 1.3 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2024-08-30 DOI: 10.1088/2057-1976/ad7597
Signe Winther Hasler, Uffe Bernchou, Claus P Behrens, Ivan Richter Vogelius, Anne L H Bisgaard, Minea Jokivuolle, Anders Bertelsen, Tine Schytte, Carsten Brink, Faisal Mahmood

Objective: The apparent diffusion coefficient (ADC) extracted from diffusion-weighted magnetic resonance imaging (DWI) is a potential biomarker in radiotherapy (RT). DWI is often implemented with an Echo-planar imaging (EPI) read-out due to speed, but unfortunately low geometric accuracy follows. This study aimed to investigate the influence of geometric distortions on the ADCs extracted from the gross tumor volume (GTV) and on the shape of the GTV in abdominal EPI-DWI. Approach: Twenty-one patients had EPI-DWI scans on a 1.5 T MRI sim before treatment and on a 1.5 T MRI-Linac at one of the first treatment fractions. Off-resonance correction with and without eddy current correction were applied to ADC maps. The clinical GTVs were deformed based on the same (but inverted) corrections to assess the local-regional geometric influence of distortions. Mean surface distance (MSD), Hausdorff distance (HD), and Dice similarity coefficient (DSC) were calculated to compare the original and distorted GTVs, and ADC values were calculated based on a mono-exponential model. Phantom measurements were performed to validate the applied correction method. Main results: The median (range) ADC change within the GTV after full distortion correction was 1.3% (0.02-6.9%) for MRI-Sim and 1.5% (0.1-6.4%) for MRI-Linac. The additional effect of the eddy current correction was small in both systems. The median (range) MSD, HD, and DSC comparing the original and off-resonance distorted GTVs for all patients were 0.43 mm (0.11-0.94 mm), 4.00 mm (1.00-7.81 mm) and 0.93 (0.82-0.99), respectively. Significance: Overall effect of distortion correction was small in terms of derived ADC values, indicating that distortion correction is unimportant for prediction of outcomes based on ADC. However, large local geometric changes occurred after off-resonance distortion correction for some patients, suggesting that if the spatial information from ADC maps is to be used for dose painting strategies, corrections should be applied. .

目的:从弥散加权磁共振成像(DWI)中提取的表观弥散系数(ADC)是放射治疗(RT)的潜在生物标志物。由于速度快,DWI 通常使用回声平面成像(EPI)读出,但不幸的是,几何精度较低。本研究旨在探讨几何失真对腹部 EPI-DWI 中从肿瘤总体积(GTV)提取的 ADCs 以及 GTV 形状的影响:21 名患者在治疗前在 1.5 T MRI sim 上进行了 EPI-DWI 扫描,并在第一次治疗分段时在 1.5 T MRI-Linac 上进行了扫描。对 ADC 图进行了带或不带涡流校正的非共振校正。临床 GTV 根据相同(但倒置)的校正进行变形,以评估变形对局部区域几何的影响。计算平均表面距离(MSD)、豪斯多夫距离(HD)和戴斯相似系数(DSC),以比较原始和变形的 GTV,并根据单指数模型计算 ADC 值。为了验证所应用的校正方法,还进行了模型测量:完全畸变校正后,MRI-Sim 和 MRI-Linac GTV 内 ADC 变化的中位数(范围)分别为 1.3% (0.02-6.9%)和 1.5% (0.1-6.4%)。涡流校正对这两种系统的额外影响都很小。所有患者的原始和非共振畸变 GTV 的 MSD、HD 和 DSC 比较中位数(范围)分别为 0.43 毫米(0.11-0.94 毫米)、4.00 毫米(1.00-7.81 毫米)和 0.93(0.82-0.99):就得出的 ADC 值而言,畸变校正的总体影响较小,这表明畸变校正对于根据 ADC 预测结果并不重要。然而,一些患者在进行非共振失真校正后出现了较大的局部几何变化,这表明如果要将 ADC 图的空间信息用于剂量绘制策略,则应进行校正。
{"title":"Impact of geometric correction on echo-planar imaging-based apparent diffusion coefficient maps for abdominal radiotherapy.","authors":"Signe Winther Hasler, Uffe Bernchou, Claus P Behrens, Ivan Richter Vogelius, Anne L H Bisgaard, Minea Jokivuolle, Anders Bertelsen, Tine Schytte, Carsten Brink, Faisal Mahmood","doi":"10.1088/2057-1976/ad7597","DOIUrl":"https://doi.org/10.1088/2057-1976/ad7597","url":null,"abstract":"<p><strong>Objective: </strong>The apparent diffusion coefficient (ADC) extracted from diffusion-weighted magnetic resonance imaging (DWI) is a potential biomarker in radiotherapy (RT). DWI is often implemented with an Echo-planar imaging (EPI) read-out due to speed, but unfortunately low geometric accuracy follows. This study aimed to investigate the influence of geometric distortions on the ADCs extracted from the gross tumor volume (GTV) and on the shape of the GTV in abdominal EPI-DWI. &#xD;Approach: Twenty-one patients had EPI-DWI scans on a 1.5 T MRI sim before treatment and on a 1.5 T MRI-Linac at one of the first treatment fractions. Off-resonance correction with and without eddy current correction were applied to ADC maps. The clinical GTVs were deformed based on the same (but inverted) corrections to assess the local-regional geometric influence of distortions. Mean surface distance (MSD), Hausdorff distance (HD), and Dice similarity coefficient (DSC) were calculated to compare the original and distorted GTVs, and ADC values were calculated based on a mono-exponential model. Phantom measurements were performed to validate the applied correction method. &#xD;Main results: The median (range) ADC change within the GTV after full distortion correction was 1.3% (0.02-6.9%) for MRI-Sim and 1.5% (0.1-6.4%) for MRI-Linac. The additional effect of the eddy current correction was small in both systems. The median (range) MSD, HD, and DSC comparing the original and off-resonance distorted GTVs for all patients were 0.43 mm (0.11-0.94 mm), 4.00 mm (1.00-7.81 mm) and 0.93 (0.82-0.99), respectively.&#xD;Significance: Overall effect of distortion correction was small in terms of derived ADC values, indicating that distortion correction is unimportant for prediction of outcomes based on ADC. However, large local geometric changes occurred after off-resonance distortion correction for some patients, suggesting that if the spatial information from ADC maps is to be used for dose painting strategies, corrections should be applied.&#xD.</p>","PeriodicalId":8896,"journal":{"name":"Biomedical Physics & Engineering Express","volume":null,"pages":null},"PeriodicalIF":1.3,"publicationDate":"2024-08-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"142103980","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Multimodal radiomics-based methods using deep learning for prediction of brain metastasis in non-small cell lung cancer with 18F-FDG PET/CT images. 基于多模态放射组学的深度学习方法,利用 18F-FDG PET/CT 图像预测非小细胞肺癌的脑转移。
IF 1.3 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2024-08-30 DOI: 10.1088/2057-1976/ad7595
Yuan Zhu, Shan Cong, Qiyang Zhang, Zhenxing Huang, Xiaohui Yao, You Cheng, Dong Liang, Zhanli Hu, Shao Dan

Objective: Approximately 57% of non-small cell lung cancer (NSCLC) patients face a 20% risk of brain metastases (BMs). The delivery of drugs to the central nervous system is challenging because of the blood-brain barrier, leading to a relatively poor prognosis for patients with BMs. Therefore, early detection and treatment of BMs are highly important for improving patient prognosis. This study aimed to investigate the feasibility of a multimodal radiomics-based method using 3D neural networks trained on 18F-FDG PET/CT images to predict BMs in NSCLC patients.

Approach: We included 226 NSCLC patients who underwent 18F-FDG PET/CT scans of areas, including the lung and brain, prior to EGFR-TKI therapy. Moreover, clinical data (age, sex, stage, etc.) were collected and analyzed. Shallow lung features and deep lung-brain features were extracted using PyRadiomics and 3D neural networks, respectively. A support vector machine (SVM) was used to predict BMs. The receiver operating characteristic (ROC) curve and F1 score were used to assess BM prediction performance.

Main result: The combination of shallow lung and shallow-deep lung-brain features demonstrated superior predictive performance (AUC=0.96±0.01). Shallow-deep lung-brain features exhibited strong significance (P<0.001) and potential predictive performance (coefficient>0.8). Moreover, BM prediction by age was significant (P<0.05).

Significance: Our approach enables the quantitative assessment of medical images and a deeper understanding of both superficial and deep tumor characteristics. This noninvasive method has the potential to identify BM-related features with statistical significance, thereby aiding in the development of targeted treatment plans for NSCLC patients.

目的:大约 57% 的非小细胞肺癌(NSCLC)患者面临 20% 的脑转移风险。由于血脑屏障的存在,将药物输送到中枢神经系统具有挑战性,导致脑转移瘤患者的预后相对较差。因此,早期发现和治疗脑转移瘤对改善患者预后非常重要。本研究旨在探讨一种基于多模态放射组学的方法的可行性,该方法使用在 18F-FDG PET/CT 图像上训练的三维神经网络来预测 NSCLC 患者的 BMs:我们纳入了 226 例 NSCLC 患者,这些患者在接受 EGFR-TKI 治疗前接受了包括肺部和脑部在内的 18F-FDG PET/CT 扫描。此外,我们还收集并分析了临床数据(年龄、性别、分期等)。使用 PyRadiomics 和三维神经网络分别提取了肺部浅层特征和肺部-大脑深层特征。支持向量机(SVM)用于预测BMs。主要结果:主要结果:浅肺和浅深肺-脑特征的组合显示出更优越的预测性能(AUC=0.96±0.01)。浅深肺-脑特征具有很强的显著性(P0.8)。此外,年龄对 BM 预测也有显著性(PS:Psignificance:我们的方法可对医学影像进行定量评估,加深对肿瘤表层和深层特征的理解。这种无创方法有可能识别出具有统计学意义的BM相关特征,从而帮助NSCLC患者制定有针对性的治疗方案。
{"title":"Multimodal radiomics-based methods using deep learning for prediction of brain metastasis in non-small cell lung cancer with 18F-FDG PET/CT images.","authors":"Yuan Zhu, Shan Cong, Qiyang Zhang, Zhenxing Huang, Xiaohui Yao, You Cheng, Dong Liang, Zhanli Hu, Shao Dan","doi":"10.1088/2057-1976/ad7595","DOIUrl":"https://doi.org/10.1088/2057-1976/ad7595","url":null,"abstract":"<p><strong>Objective: </strong>Approximately 57% of non-small cell lung cancer (NSCLC) patients face a 20% risk of brain metastases (BMs). The delivery of drugs to the central nervous system is challenging because of the blood-brain barrier, leading to a relatively poor prognosis for patients with BMs. Therefore, early detection and treatment of BMs are highly important for improving patient prognosis. This study aimed to investigate the feasibility of a multimodal radiomics-based method using 3D neural networks trained on 18F-FDG PET/CT images to predict BMs in NSCLC patients.</p><p><strong>Approach: </strong>We included 226 NSCLC patients who underwent 18F-FDG PET/CT scans of areas, including the lung and brain, prior to EGFR-TKI therapy. Moreover, clinical data (age, sex, stage, etc.) were collected and analyzed. Shallow lung features and deep lung-brain features were extracted using PyRadiomics and 3D neural networks, respectively. A support vector machine (SVM) was used to predict BMs. The receiver operating characteristic (ROC) curve and F1 score were used to assess BM prediction performance.</p><p><strong>Main result: </strong>The combination of shallow lung and shallow-deep lung-brain features demonstrated superior predictive performance (AUC=0.96±0.01). Shallow-deep lung-brain features exhibited strong significance (P<0.001) and potential predictive performance (coefficient>0.8). Moreover, BM prediction by age was significant (P<0.05).</p><p><strong>Significance: </strong>Our approach enables the quantitative assessment of medical images and a deeper understanding of both superficial and deep tumor characteristics. This noninvasive method has the potential to identify BM-related features with statistical significance, thereby aiding in the development of targeted treatment plans for NSCLC patients.</p>","PeriodicalId":8896,"journal":{"name":"Biomedical Physics & Engineering Express","volume":null,"pages":null},"PeriodicalIF":1.3,"publicationDate":"2024-08-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"142103982","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Thickness-extensible Higher Order Plate Theory with Enforced C1 Continuity for the Analysis of PEEK Medical Implants. 用于分析聚醚醚酮(PEEK)医疗植入物的具有强制 C1 连续性的增厚高阶板理论。
IF 1.3 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2024-08-30 DOI: 10.1088/2057-1976/ad7591
Mohamad Hasan Bin Tasneem, Farooq Al-Jahwari, Mahmood Al-Kindi, Imad Al-Lawati, Abdulmonem Al-Lawati

Plate-like structures had been thoroughly studied in literature over years to reduce the computational space from 3D to 2D. Many of these theories suffer either from satisfying the free traction condition or thickness extensibility in addition to the consistency of transverse shear strain energy. This work presents a higher order shear deformation thickness-extensible plate theory (eHSDT) for the analysis of plates. The proposed eHSDT satisfies the condition of free traction as other theories do but it also satisfies the condition of consistency of transverse shear strain energy which is neglected by many theories in the area of plates and shells. The implementation of the proposed theory in displacement-based finite element procedure requires continuity of derivatives across elements. This necessary condition was achieved using the penalty enforcement method for derivative-based nodal degrees of freedom across the standard 9-nodes Lagrange element. The theory was tested for elastic bending deformation of Polyether-ether-ketone (PEEK) which is one of the basic materials for medical implants. The theory showed good accuracy compared to experimental data of the three-points bending test. The present eHSDT was also tested for different conditions with a wide range of aspects ratios (thin to thick plates) and different boundary conditions. The accuracy of the proposed eHSDT was verified against exact solutions for these conditions which showed the advantage over other approaches and commercial finite element packages.

多年来,为了将计算空间从三维缩小到二维,文献中对板状结构进行了深入研究。其中许多理论除了横向剪切应变能的一致性外,还存在满足自由牵引条件或厚度可延伸性的问题。本研究提出了一种用于板分析的高阶剪切变形厚度可扩展板理论(eHSDT)。所提出的 eHSDT 与其他理论一样满足自由牵引的条件,但它也满足横向剪切应变能一致性的条件,而在板和壳领域,许多理论都忽略了这一点。在基于位移的有限元程序中实施所提出的理论需要跨元素导数的连续性。这一必要条件是通过对标准 9 节点拉格朗日元素中基于导数的节点自由度采用惩罚执行法来实现的。该理论针对聚醚醚酮(PEEK)的弹性弯曲变形进行了测试,聚醚醚酮是医疗植入物的基本材料之一。与三点弯曲试验的实验数据相比,该理论显示出良好的准确性。本 eHSDT 还在不同条件下进行了测试,包括宽范围的纵横比(薄板与厚板)和不同的边界条件。针对这些条件的精确解验证了所提出的 eHSDT 的准确性,这表明它比其他方法和商用有限元软件包更具优势。
{"title":"Thickness-extensible Higher Order Plate Theory with Enforced C1 Continuity for the Analysis of PEEK Medical Implants.","authors":"Mohamad Hasan Bin Tasneem, Farooq Al-Jahwari, Mahmood Al-Kindi, Imad Al-Lawati, Abdulmonem Al-Lawati","doi":"10.1088/2057-1976/ad7591","DOIUrl":"https://doi.org/10.1088/2057-1976/ad7591","url":null,"abstract":"<p><p>Plate-like structures had been thoroughly studied in literature over years to reduce the computational space from 3D to 2D. Many of these theories suffer either from satisfying the free traction condition or thickness extensibility in addition to the consistency of transverse shear strain energy. This work presents a higher order shear deformation thickness-extensible plate theory (eHSDT) for the analysis of plates. The proposed eHSDT satisfies the condition of free traction as other theories do but it also satisfies the condition of consistency of transverse shear strain energy which is neglected by many theories in the area of plates and shells. The implementation of the proposed theory in displacement-based finite element procedure requires continuity of derivatives across elements. This necessary condition was achieved using the penalty enforcement method for derivative-based nodal degrees of freedom across the standard 9-nodes Lagrange element. The theory was tested for elastic bending deformation of Polyether-ether-ketone (PEEK) which is one of the basic materials for medical implants. The theory showed good accuracy compared to experimental data of the three-points bending test. The present eHSDT was also tested for different conditions with a wide range of aspects ratios (thin to thick plates) and different boundary conditions. The accuracy of the proposed eHSDT was verified against exact solutions for these conditions which showed the advantage over other approaches and commercial finite element packages.</p>","PeriodicalId":8896,"journal":{"name":"Biomedical Physics & Engineering Express","volume":null,"pages":null},"PeriodicalIF":1.3,"publicationDate":"2024-08-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"142103984","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Investigation of a novel 2.5 MV sintered diamond target beam for intracranial linac-based stereotactic treatments. 新型 2.5 MV 烧结金刚石靶束在颅内亚麻加速器立体定向治疗中的应用研究。
IF 1.3 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2024-08-29 DOI: 10.1088/2057-1976/ad7031
Jennifer M Borsavage, Amanda J Cherpak, James L Robar

Purpose. This work investigates the small-field dosimetric characteristics of a 2.5 MV sintered diamond target beam and its feasibility for use in linac-based intracranial stereotactic treatments. Due to the increased proportion of low energy photons in the low-Z beam, it was hypothesized that this novel beam would provide sharper dose fall-off compared to the 6 MV beam owing to the reduced energy, and therefore range, of secondary electrons.Methods. Stereotactic treatments of ocular melanoma and trigeminal neuralgia were simulated for 2.5 MV low-Z and 6 MV beams using Monte Carlo to calculate dose in a voxelized anatomical phantom. Two collimation methods were investigated, including a 5 × 3 mm2HDMLC field and a 4 mm cone to demonstrate isolated and combined effects of geometric and radiological contributions to the penumbral width.Results. The measured 2.5 MV low-Z dosimetric profiles demonstrated reduced penumbra by 0.5 mm in both the inline and crossline directions across all depths for both collimation methods, compared to 6 MV. In both treatment cases, the 2.5 MV low-Z beam collimated with the 4 mm cone produced the sharpest dose fall off in profiles captured through isocenter. This improved fall-off resulted in a 59% decrease to the maximum brainstem dose in the trigeminal neuralgia case for the 2.5 MV low-Z MLC collimated beam compared to 6 MV. Reductions to the maximum and mean doses to ipsilateral and contralateral OARs in the ocular melanoma case were observed for the 2.5 MV low-Z beam compared to 6 MV with both collimation methods.Conclusions. While the low dose rate of this novel beam prohibits immediate clinical translation, the results of this study support the further development of this prototype beam to decrease toxicity in intracranial SRS treatments.

目的:这项研究调查了 2.5 MV 烧结钻石靶光束的小场剂量测定特性及其用于基于线性加速器的颅内立体定向治疗的可行性。由于低Z光束中低能量光子的比例增加,我们假设这种新型光束会比6 MV光束提供更锐利的剂量衰减,原因是次级电子的能量和范围减少了:使用蒙特卡洛模拟 2.5 MV 低 Z 和 6 MV 射束对眼部黑色素瘤和三叉神经痛的立体定向治疗,计算体素化解剖模型中的剂量。研究了两种准直方法,包括 5x3 mm2 HDMLC 磁场和 4 mm 锥体,以展示几何和放射对半影宽度的单独和综合影响:测得的 2.5 MV 低 Z 剂量曲线显示,与 6 MV 相比,两种准直方法在所有深度的内线和横线方向上的半影都减少了 0.5 毫米。在这两种治疗方法中,用 4 毫米锥体准直的 2.5 MV 低 Z 射束在通过等中心捕获的剖面中产生的剂量衰减最明显。与 6 MV 相比,2.5 MV 低 Z MLC 准直光束在三叉神经痛病例中的最大脑干剂量下降了 59%。在眼部黑色素瘤病例中,采用两种准直方法,2.5 MV lowZ 射束的同侧和对侧 OAR 最大剂量和平均剂量均比 6 MV 减少:虽然这种新型射束的剂量率较低,无法立即应用于临床,但本研究结果支持进一步开发这种原型射束,以降低颅内 SRS 治疗的毒性。
{"title":"Investigation of a novel 2.5 MV sintered diamond target beam for intracranial linac-based stereotactic treatments.","authors":"Jennifer M Borsavage, Amanda J Cherpak, James L Robar","doi":"10.1088/2057-1976/ad7031","DOIUrl":"10.1088/2057-1976/ad7031","url":null,"abstract":"<p><p><i>Purpose</i>. This work investigates the small-field dosimetric characteristics of a 2.5 MV sintered diamond target beam and its feasibility for use in linac-based intracranial stereotactic treatments. Due to the increased proportion of low energy photons in the low-Z beam, it was hypothesized that this novel beam would provide sharper dose fall-off compared to the 6 MV beam owing to the reduced energy, and therefore range, of secondary electrons.<i>Methods</i>. Stereotactic treatments of ocular melanoma and trigeminal neuralgia were simulated for 2.5 MV low-Z and 6 MV beams using Monte Carlo to calculate dose in a voxelized anatomical phantom. Two collimation methods were investigated, including a 5 × 3 mm<sup>2</sup>HDMLC field and a 4 mm cone to demonstrate isolated and combined effects of geometric and radiological contributions to the penumbral width.<i>Results</i>. The measured 2.5 MV low-Z dosimetric profiles demonstrated reduced penumbra by 0.5 mm in both the inline and crossline directions across all depths for both collimation methods, compared to 6 MV. In both treatment cases, the 2.5 MV low-Z beam collimated with the 4 mm cone produced the sharpest dose fall off in profiles captured through isocenter. This improved fall-off resulted in a 59% decrease to the maximum brainstem dose in the trigeminal neuralgia case for the 2.5 MV low-Z MLC collimated beam compared to 6 MV. Reductions to the maximum and mean doses to ipsilateral and contralateral OARs in the ocular melanoma case were observed for the 2.5 MV low-Z beam compared to 6 MV with both collimation methods.<i>Conclusions</i>. While the low dose rate of this novel beam prohibits immediate clinical translation, the results of this study support the further development of this prototype beam to decrease toxicity in intracranial SRS treatments.</p>","PeriodicalId":8896,"journal":{"name":"Biomedical Physics & Engineering Express","volume":null,"pages":null},"PeriodicalIF":1.3,"publicationDate":"2024-08-29","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"141995223","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Numerical assessment of using various outlet boundary conditions on the hemodynamics of an idealized left coronary artery model. 使用各种出口边界条件对理想化左冠状动脉模型血液动力学的数值评估。
IF 1.3 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2024-08-29 DOI: 10.1088/2057-1976/ad7030
Asif Equbal, Paragmoni Kalita

Vascular diseases are greatly influenced by the hemodynamic parameters and the accuracy of determining these parameters depends on the use of correct boundary conditions. The present work carries out a two-way fluid-structure interaction (FSI) simulation to investigate the effects of outlet pressure boundary conditions on the hemodynamics through the left coronary artery bifurcation with moderate stenosis (50%) in the left anterior descending (LAD) branch. The Carreau viscosity model is employed to characterise the shear-thinning behaviour of blood. The results of the study reveal that the employment of zero pressure at the outlet boundaries significantly overestimates the values of hemodynamic variables like wall shear stress (WSS), and time-averaged wall shear stress (TAWSS) compared with human healthy and pulsatile pressure outlet conditions. However, the difference between these variables is marginally low for human healthy and pulsatile pressure outlets. The oscillatory shear index (OSI) remains the same across all scenarios, indicating independence from the outlet boundary condition. Furthermore, the magnitude of negative axial velocity and pressure drop across the plaque are found to be higher at the zero pressure outlet boundary condition.

血管疾病在很大程度上受血液动力学参数的影响,而确定这些参数的准确性取决于使用正确的边界条件。本研究进行了双向流固耦合(FSI)模拟,以研究出口压力边界条件对通过左冠状动脉分叉处左前降支(LAD)中度狭窄(50%)的血液动力学的影响。采用 Carreau 粘度模型来描述血液的剪切稀化行为。研究结果表明,与人体健康和搏动压力出口条件相比,出口边界采用零压力会明显高估血流动力学变量的值,如壁剪切应力(WSS)和时间平均壁剪切应力(TAWSS)。不过,这些变量之间的差异在人体健康和搏动压力出口条件下略低。振荡剪切指数(OSI)在所有情况下都保持不变,这表明与出口边界条件无关。此外,在零压力出口边界条件下,负轴向速度的大小和穿过斑块的压降都较高。
{"title":"Numerical assessment of using various outlet boundary conditions on the hemodynamics of an idealized left coronary artery model.","authors":"Asif Equbal, Paragmoni Kalita","doi":"10.1088/2057-1976/ad7030","DOIUrl":"10.1088/2057-1976/ad7030","url":null,"abstract":"<p><p>Vascular diseases are greatly influenced by the hemodynamic parameters and the accuracy of determining these parameters depends on the use of correct boundary conditions. The present work carries out a two-way fluid-structure interaction (FSI) simulation to investigate the effects of outlet pressure boundary conditions on the hemodynamics through the left coronary artery bifurcation with moderate stenosis (50%) in the left anterior descending (LAD) branch. The Carreau viscosity model is employed to characterise the shear-thinning behaviour of blood. The results of the study reveal that the employment of zero pressure at the outlet boundaries significantly overestimates the values of hemodynamic variables like wall shear stress (WSS), and time-averaged wall shear stress (TAWSS) compared with human healthy and pulsatile pressure outlet conditions. However, the difference between these variables is marginally low for human healthy and pulsatile pressure outlets. The oscillatory shear index (OSI) remains the same across all scenarios, indicating independence from the outlet boundary condition. Furthermore, the magnitude of negative axial velocity and pressure drop across the plaque are found to be higher at the zero pressure outlet boundary condition.</p>","PeriodicalId":8896,"journal":{"name":"Biomedical Physics & Engineering Express","volume":null,"pages":null},"PeriodicalIF":1.3,"publicationDate":"2024-08-29","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"141995225","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
3D printed heterogeneous paediatric head and adult thorax phantoms for linear accelerator radiotherapy quality assurance: from fabrication to treatment delivery. 用于直线加速器放射治疗质量保证的 3D 打印异质儿童头部和成人胸部模型:从制造到治疗实施。
IF 1.3 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2024-08-29 DOI: 10.1088/2057-1976/ad6f13
John Paul O Bustillo, Julia Rebecca D Posadas, Jacob L Mata, Elrick T Inocencio, Anatoly B Rosenfeld, Michael L F Lerch

Objective. This study aims to design and fabricate a 3D printed heterogeneous paediatric head phantom and to customize a thorax phantom for radiotherapy dosimetry.Approach. This study designed, fabricated, and tested 3D printed radiotherapy phantoms that can simulate soft tissue, lung, brain, and bone. Various polymers were considered in designing the phantoms. Polylactic acid+, nylon, and plaster were used in simulating different tissue equivalence. Dimensional accuracy, and CT number were investigated. The phantoms were subjected to a complete radiotherapy clinical workflow. Several treatment plans were delivered in both the head and the thorax phantom from a simple single 6 MV beam, parallel opposed beams, and five-field intensity modulated radiotherapy (IMRT) beams. Dose measurements using an ionization chamber and radiochromic films were compared with the calculated doses of the Varian Eclipse treatment planning system (TPS).Main results. The fabricated heterogeneous phantoms represent paediatric human head and adult thorax based on its radiation attenuation and anatomy. The measured CT number ranges are within -786.23 ± 10.55, 0.98 ± 3.86, 129.51 ± 12.83, and 651.14 ± 47.76 HU for lung, water/brain, soft tissue, and bone, respectively. It has a good radiological imaging visual similarity relative to a real human head and thorax depicting soft tissue, lung, bone, and brain. The accumulated dose readings for both conformal radiotherapy and IMRT match with the TPS calculated dose within ±2% and ±4% for head and thorax phantom, respectively. The mean pass rate for all the plans delivered are above 90% for gamma analysis criterion of 3%/3 mm.Significance and conclusion. The fabricated heterogeneous paediatric head and thorax phantoms are useful in Linac end-to-end radiotherapy quality assurance based on its CT image and measured radiation dose. The manufacturing and dosimetry workflow of this study can be utilized by other institutions for dosimetry and trainings.

目的: 本研究旨在设计和制造一个 3D 打印的异质儿科头部模型,并定制一个胸部模型用于放射治疗剂量测定:本研究旨在设计和制造一个 3D 打印的异质儿科头部模型,并定制一个用于放射治疗剂量测定的胸部模型:本研究设计、制造并测试了可模拟软组织、肺部、大脑和骨骼的三维打印放疗模型。在设计模型时考虑了多种聚合物。聚乳酸+、尼龙和石膏被用于模拟不同的组织等值。对尺寸精度和 CT 数量进行了研究。对模型进行了完整的放射治疗临床工作流程测试。在头部和胸部模型中采用了多种治疗方案,包括简单的单个 6 MV 射束、平行对置射束和五场调强放射治疗(IMRT)射束。使用电离室和放射性变色胶片测量的剂量与瓦里安 Eclipse 治疗计划系统(TPS)计算的剂量进行了比较:根据辐射衰减和解剖结构,制作的异质模型代表了儿童头部和成人胸部。测得的肺部、水/脑部、软组织和骨骼的 CT 数值范围分别为 -786.23±10.55、0.98±3.86、129.51±12.83 和 651.14±47.76HU。相对于真实的人体头部和胸部,它在软组织、肺、骨和脑的描绘上具有良好的放射成像视觉相似性。适形放射治疗和 IMRT 的累积剂量读数与 TPS 计算的头部和胸部模型剂量的吻合度分别在 ±2% 和 ±4% 以内。在伽马分析标准为 3% / 3 mm 时,所有计划的平均合格率均超过 90%:制作的异质儿科头部和胸部模型可根据其 CT 图像和测量的辐射剂量,用于 Linac 端到端放疗质量保证。本研究的制造和剂量测定工作流程可供其他机构用于剂量测定和培训。
{"title":"3D printed heterogeneous paediatric head and adult thorax phantoms for linear accelerator radiotherapy quality assurance: from fabrication to treatment delivery.","authors":"John Paul O Bustillo, Julia Rebecca D Posadas, Jacob L Mata, Elrick T Inocencio, Anatoly B Rosenfeld, Michael L F Lerch","doi":"10.1088/2057-1976/ad6f13","DOIUrl":"10.1088/2057-1976/ad6f13","url":null,"abstract":"<p><p><i>Objective</i>. This study aims to design and fabricate a 3D printed heterogeneous paediatric head phantom and to customize a thorax phantom for radiotherapy dosimetry.<i>Approach</i>. This study designed, fabricated, and tested 3D printed radiotherapy phantoms that can simulate soft tissue, lung, brain, and bone. Various polymers were considered in designing the phantoms. Polylactic acid+, nylon, and plaster were used in simulating different tissue equivalence. Dimensional accuracy, and CT number were investigated. The phantoms were subjected to a complete radiotherapy clinical workflow. Several treatment plans were delivered in both the head and the thorax phantom from a simple single 6 MV beam, parallel opposed beams, and five-field intensity modulated radiotherapy (IMRT) beams. Dose measurements using an ionization chamber and radiochromic films were compared with the calculated doses of the Varian Eclipse treatment planning system (TPS).<i>Main results</i>. The fabricated heterogeneous phantoms represent paediatric human head and adult thorax based on its radiation attenuation and anatomy. The measured CT number ranges are within -786.23 ± 10.55, 0.98 ± 3.86, 129.51 ± 12.83, and 651.14 ± 47.76 HU for lung, water/brain, soft tissue, and bone, respectively. It has a good radiological imaging visual similarity relative to a real human head and thorax depicting soft tissue, lung, bone, and brain. The accumulated dose readings for both conformal radiotherapy and IMRT match with the TPS calculated dose within ±2% and ±4% for head and thorax phantom, respectively. The mean pass rate for all the plans delivered are above 90% for gamma analysis criterion of 3%/3 mm.<i>Significance and conclusion</i>. The fabricated heterogeneous paediatric head and thorax phantoms are useful in Linac end-to-end radiotherapy quality assurance based on its CT image and measured radiation dose. The manufacturing and dosimetry workflow of this study can be utilized by other institutions for dosimetry and trainings.</p>","PeriodicalId":8896,"journal":{"name":"Biomedical Physics & Engineering Express","volume":null,"pages":null},"PeriodicalIF":1.3,"publicationDate":"2024-08-29","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"141981572","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
A cross-shaped terahertz metamaterial absorber for brain cancer detection. 用于检测脑癌的十字形太赫兹超材料吸收器。
IF 1.3 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2024-08-28 DOI: 10.1088/2057-1976/ad6f16
W Bhowmik, B Appasani, A Gorai, A K V Jha, B P De, P K Samanta

The article presents, for the first time, a terahertz metamaterial absorber (TMA) designed in the shape of a cross consisting of four orthogonally positioned horn-shaped patches in succession, to detect brain cancer cells. The design exhibits the property of mu-negative material, indicating magnetic resonance. The proposed TMA has achieved an impressive absorption rate of 99.43% at 2.334 THz and a high Q-factor of 47.15. The sensing capability has been investigated by altering the refractive index of the surrounding medium in the range of 1.3 to 1.48, resulting in a sensitivity of 0.502 THz/RIU. The proposed TMA exhibits complete polarization insensitivity, highlighting this as one of its advantageous features. The adequate sensing capability of the proposed TMA in differentiating normal and cancerous brain cells makes it a viable candidate for an early and efficient brain cancer detector. This research can be the foundation for future research on using THz radiation for brain cancer detection.

文章首次介绍了一种太赫兹超材料吸收器(TMA),该吸收器设计成十字形,由四个连续正交定位的喇叭形贴片组成,用于检测脑癌细胞。该设计显示了μ负材料的特性,表明存在磁共振。拟议的 TMA 在 2.334 THz 频率下的吸收率高达 99.43%,Q 系数高达 47.15。通过改变周围介质在 1.3 至 1.48 范围内的折射率,对传感能力进行了研究,结果灵敏度达到 0.502 THz/RIU。拟议的 TMA 具有完全的极化不敏感性,这是它的优势特点之一。所提出的 TMA 在区分正常和癌变脑细胞方面具有充分的感应能力,使其成为早期和高效脑癌检测器的可行候选者。这项研究可为今后利用太赫兹辐射检测脑癌的研究奠定基础。
{"title":"A cross-shaped terahertz metamaterial absorber for brain cancer detection.","authors":"W Bhowmik, B Appasani, A Gorai, A K V Jha, B P De, P K Samanta","doi":"10.1088/2057-1976/ad6f16","DOIUrl":"10.1088/2057-1976/ad6f16","url":null,"abstract":"<p><p>The article presents, for the first time, a terahertz metamaterial absorber (TMA) designed in the shape of a cross consisting of four orthogonally positioned horn-shaped patches in succession, to detect brain cancer cells. The design exhibits the property of mu-negative material, indicating magnetic resonance. The proposed TMA has achieved an impressive absorption rate of 99.43% at 2.334 THz and a high Q-factor of 47.15. The sensing capability has been investigated by altering the refractive index of the surrounding medium in the range of 1.3 to 1.48, resulting in a sensitivity of 0.502 THz/RIU. The proposed TMA exhibits complete polarization insensitivity, highlighting this as one of its advantageous features. The adequate sensing capability of the proposed TMA in differentiating normal and cancerous brain cells makes it a viable candidate for an early and efficient brain cancer detector. This research can be the foundation for future research on using THz radiation for brain cancer detection.</p>","PeriodicalId":8896,"journal":{"name":"Biomedical Physics & Engineering Express","volume":null,"pages":null},"PeriodicalIF":1.3,"publicationDate":"2024-08-28","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"141981573","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Evaluation of TOPAS MC tool performance in optical photon transport and radioluminescence-based dosimetry. 评估 TOPAS MC 工具在光学光子传输和基于放射线的剂量测定中的性能。
IF 1.3 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2024-08-28 DOI: 10.1088/2057-1976/ad6f14
A Khodaei, F Moradi, A Oresegun, H T Zubair, D A Bradley, S A Ibrahim, H A Abdul-Rashid

Radiation therapy plays a pivotal role in modern cancer treatment, demanding precise and accurate dose delivery to tumor sites while minimizing harm to surrounding healthy tissues. Monte Carlo simulations have emerged as indispensable tools for achieving this precision, offering detailed insights into radiation transport and interaction at the subatomic level. As the use of scintillation and luminescence dosimetry becomes increasingly prevalent in radiation therapy, there arises a need for validated Monte Carlo tools tailored to optical photon transport applications. In this paper, an evaluation process of the TOPAS (TOol for PArticle Simulation) Monte Carlo tool for Cerenkov light generation, optical photon transport and radioluminescence based dosimetry is presented. Three distinct sources of validation data are utilized: one from a published set of experimental results and two others from simulations performed with the Geant4 code. The methodology employed for evaluation includes the selection of benchmark experiments, making use of opt3 and opt4 Geant4 physics models and simulation setup, with observed slight discrepancies within the calculation uncertainties. Additionally, the complexities and challenges associated with modeling optical photons generation through luminescence or Cerenkov radiation and their transport are discussed. The results of our evaluation suggests that TOPAS can be used to reliably predict Cerenkov generation, luminescence phenomenon and the behavior of optical photons in common dosimetry scenarios.

放射治疗在现代癌症治疗中起着举足轻重的作用,它要求精确地将剂量传递到肿瘤部位,同时尽量减少对周围健康组织的伤害。蒙特卡洛模拟是实现这种精确性不可或缺的工具,它提供了亚原子水平辐射传输和相互作用的详细见解。随着闪烁和发光剂量测定在放射治疗中的应用越来越普遍,人们需要为光学光子传输应用量身定制经过验证的蒙特卡罗工具。本文介绍了 TOPAS(粒子模拟工具)蒙特卡洛工具的评估过程,该工具适用于塞伦科夫光生成、光学光子传输和基于辐射的剂量测定。利用了三个不同来源的验证数据:一个来自已公布的实验结果集,另两个来自使用 Geant4 代码进行的模拟。评估采用的方法包括选择基准实验,利用 opt3 和 opt4 Geant4 物理模型和模拟设置,在计算不确定性范围内观察到轻微差异。此外,还讨论了通过发光或切伦科夫辐射产生的光学光子及其传输建模的复杂性和挑战。我们的评估结果表明,TOPAS 可用于可靠地预测常见剂量测定方案中的切伦科夫辐射生成、发光现象和光学光子行为。
{"title":"Evaluation of TOPAS MC tool performance in optical photon transport and radioluminescence-based dosimetry.","authors":"A Khodaei, F Moradi, A Oresegun, H T Zubair, D A Bradley, S A Ibrahim, H A Abdul-Rashid","doi":"10.1088/2057-1976/ad6f14","DOIUrl":"10.1088/2057-1976/ad6f14","url":null,"abstract":"<p><p>Radiation therapy plays a pivotal role in modern cancer treatment, demanding precise and accurate dose delivery to tumor sites while minimizing harm to surrounding healthy tissues. Monte Carlo simulations have emerged as indispensable tools for achieving this precision, offering detailed insights into radiation transport and interaction at the subatomic level. As the use of scintillation and luminescence dosimetry becomes increasingly prevalent in radiation therapy, there arises a need for validated Monte Carlo tools tailored to optical photon transport applications. In this paper, an evaluation process of the TOPAS (TOol for PArticle Simulation) Monte Carlo tool for Cerenkov light generation, optical photon transport and radioluminescence based dosimetry is presented. Three distinct sources of validation data are utilized: one from a published set of experimental results and two others from simulations performed with the Geant4 code. The methodology employed for evaluation includes the selection of benchmark experiments, making use of opt3 and opt4 Geant4 physics models and simulation setup, with observed slight discrepancies within the calculation uncertainties. Additionally, the complexities and challenges associated with modeling optical photons generation through luminescence or Cerenkov radiation and their transport are discussed. The results of our evaluation suggests that TOPAS can be used to reliably predict Cerenkov generation, luminescence phenomenon and the behavior of optical photons in common dosimetry scenarios.</p>","PeriodicalId":8896,"journal":{"name":"Biomedical Physics & Engineering Express","volume":null,"pages":null},"PeriodicalIF":1.3,"publicationDate":"2024-08-28","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"141981576","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Microfluidic systems for modeling digestive cancer: a review of recent progress. 用于消化系统癌症建模的微流控系统:最新进展回顾。
IF 1.3 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2024-08-28 DOI: 10.1088/2057-1976/ad6f15
ZahraSadat Razavi, Madjid Soltani, Hamidreza Pazoki-Toroudi, Mahsa Dabagh

Purpose. This review aims to highlight current improvements in microfluidic devices designed for digestive cancer simulation. The review emphasizes the use of multicellular 3D tissue engineering models to understand the complicated biology of the tumor microenvironment (TME) and cancer progression. The purpose is to develop oncology research and improve digestive cancer patients' lives.Methods. This review analyzes recent research on microfluidic devices for mimicking digestive cancer. It uses tissue-engineered microfluidic devices, notably organs on a chip (OOC), to simulate human organ function in the lab. Cell cultivation on modern three-dimensional hydrogel platforms allows precise geometry, biological components, and physiological qualities. The review analyzes novel methodologies, key findings, and technical progress to explain this field's advances.Results. This study discusses current advances in microfluidic devices for mimicking digestive cancer. Micro physiological systems with multicellular 3D tissue engineering models are emphasized. These systems capture complex biochemical gradients, niche variables, and dynamic cell-cell interactions in the tumor microenvironment (TME). These models reveal stomach cancer biology and progression by duplicating the TME. Recent discoveries and technology advances have improved our understanding of gut cancer biology, as shown in the review.Conclusion. Microfluidic systems play a crucial role in modeling digestive cancer and furthering oncology research. These platforms could transform drug development and treatment by revealing the complex biology of the tumor microenvironment and cancer progression. The review provides a complete summary of recent advances and suggests future research for field professionals. The review's major goal is to further medical research and improve digestive cancer patients' lives.

目的: 本综述旨在重点介绍目前用于消化系统癌症模拟的微流体设备的改进情况。综述强调利用多细胞三维组织工程模型来了解肿瘤微环境(TME)和癌症进展的复杂生物学过程。方法: 本综述分析了最近关于模拟消化道癌症的微流控装置的研究。它利用组织工程微流控装置,特别是芯片上的器官(OOC),在实验室中模拟人体器官的功能。在现代三维水凝胶平台上进行细胞培养,可以获得精确的几何形状、生物成分和生理特性。本综述分析了新方法、主要发现和技术进展,以解释这一领域的进展。重点介绍了具有多细胞三维组织工程模型的微生理系统。这些系统捕捉了肿瘤微环境(TME)中复杂的生化梯度、生态位变量和动态的细胞-细胞相互作用。这些模型通过复制肿瘤微环境来揭示胃癌生物学和进展。正如综述所示,最近的发现和技术进步提高了我们对肠癌生物学的认识。这些平台可以揭示肿瘤微环境和癌症进展的复杂生物学过程,从而改变药物开发和治疗方法。这篇综述全面总结了近期的研究进展,并为该领域的专业人士提出了未来的研究建议。该综述的主要目标是促进医学研究,改善消化系统癌症患者的生活。
{"title":"Microfluidic systems for modeling digestive cancer: a review of recent progress.","authors":"ZahraSadat Razavi, Madjid Soltani, Hamidreza Pazoki-Toroudi, Mahsa Dabagh","doi":"10.1088/2057-1976/ad6f15","DOIUrl":"10.1088/2057-1976/ad6f15","url":null,"abstract":"<p><p><i>Purpose</i>. This review aims to highlight current improvements in microfluidic devices designed for digestive cancer simulation. The review emphasizes the use of multicellular 3D tissue engineering models to understand the complicated biology of the tumor microenvironment (TME) and cancer progression. The purpose is to develop oncology research and improve digestive cancer patients' lives.<i>Methods</i>. This review analyzes recent research on microfluidic devices for mimicking digestive cancer. It uses tissue-engineered microfluidic devices, notably organs on a chip (OOC), to simulate human organ function in the lab. Cell cultivation on modern three-dimensional hydrogel platforms allows precise geometry, biological components, and physiological qualities. The review analyzes novel methodologies, key findings, and technical progress to explain this field's advances.<i>Results</i>. This study discusses current advances in microfluidic devices for mimicking digestive cancer. Micro physiological systems with multicellular 3D tissue engineering models are emphasized. These systems capture complex biochemical gradients, niche variables, and dynamic cell-cell interactions in the tumor microenvironment (TME). These models reveal stomach cancer biology and progression by duplicating the TME. Recent discoveries and technology advances have improved our understanding of gut cancer biology, as shown in the review.<i>Conclusion</i>. Microfluidic systems play a crucial role in modeling digestive cancer and furthering oncology research. These platforms could transform drug development and treatment by revealing the complex biology of the tumor microenvironment and cancer progression. The review provides a complete summary of recent advances and suggests future research for field professionals. The review's major goal is to further medical research and improve digestive cancer patients' lives.</p>","PeriodicalId":8896,"journal":{"name":"Biomedical Physics & Engineering Express","volume":null,"pages":null},"PeriodicalIF":1.3,"publicationDate":"2024-08-28","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"141981578","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Commissioning of MRI-guided gynaecological brachytherapy using an MR-linac. 使用核磁共振近距离治疗仪进行核磁共振引导的妇科近距离治疗。
IF 1.3 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2024-08-27 DOI: 10.1088/2057-1976/ad6c54
José Alejandro Rojas-López, Alexis Cabrera-Santiago, Celeste Adragna, Brenda Elisabeth Ibarra-Ortega, José Eleazar López-Luna, Jonathan Agustín Contreras-Rodríguez, Efraín Martínez-Ortiz

Purpose. To evaluate the feasibility of use of an 1.5 T magnetic resonance (MR)-linear accelerator MR-linac for imaging in gynaecologic high-dose-rate (HDR) brachytherapy.Method. Commissioning measurements for MR images quality control, geometric distortion, dwell position accuracy, applicator reconstruction and end-to-end test for a tandem-and-ring applicator were performed following the recommendations of American Brachytherapy Society, International Commission on Radiation Units and Measurements and Report of the Brachytherapy Working Group of the Spanish Society of Medical Physics. The values for MR-based IGABT were compared to the corresponding values with computed tomography (CT).Results. Measured distorsions for the MR images were less than 0.50 mm compared to the CT images. The differences between 3D displacements for all dwell positions were 0.66 mm and 0.62 mm for the tandem and ring, respectively. The maximum difference is 0.64 mm for the distances from the applicator tip obtained using the films. The CT and MR dose differences for the right and left 'A' points were 0.9% and -0.7%, respectively. Similar results were observed in terms of dose distribution for CT and Mr The gamma passing rate was 99.3% and 99.5%, respectively.Conclusion. The use of MR images from an MR-linac used in a radiotherapy service for gynaecological brachytherapy was proved to be feasible, safe and precise as the geometrical differences were less than 1 mm, and the dosimetric differences were less than 1% when comparing to the use of CT images for the same purpose.

目的:评估在妇科高剂量率(HDR)近距离放射治疗中使用 1.5 T 磁共振(MR)直线加速器 MR-linac 进行成像的可行性:按照美国近距离放射治疗学会、国际辐射单位和测量委员会以及西班牙医学物理学会近距离放射治疗工作组的报告的建议,对串联环形涂药器的 MR 图像质量控制、几何失真、停留位置精度、涂药器重建和端对端测试进行了调试测量。基于 MR 的 IGABT 值与计算机断层扫描 (CT) 的相应值进行了比较:与 CT 图像相比,MR 图像的测量偏差小于 0.50 毫米。所有停留位置的三维位移差值分别为 0.66 毫米(串联)和 0.62 毫米(环形)。使用胶片获得的涂抹器尖端距离的最大差异为 0.64 毫米。左右 "A "点的 CT 和 MR 剂量差异分别为 0.9% 和 -0.7%。CT 和 MR 的剂量分布结果相似。伽马通过率分别为 99.3% 和 99.5%:事实证明,在妇科近距离放射治疗服务中使用磁共振成像仪的磁共振图像是可行、安全和精确的,因为与使用 CT 图像相比,两者的几何差异小于 1 毫米,剂量差异小于 1%。
{"title":"Commissioning of MRI-guided gynaecological brachytherapy using an MR-linac.","authors":"José Alejandro Rojas-López, Alexis Cabrera-Santiago, Celeste Adragna, Brenda Elisabeth Ibarra-Ortega, José Eleazar López-Luna, Jonathan Agustín Contreras-Rodríguez, Efraín Martínez-Ortiz","doi":"10.1088/2057-1976/ad6c54","DOIUrl":"10.1088/2057-1976/ad6c54","url":null,"abstract":"<p><p><i>Purpose</i>. To evaluate the feasibility of use of an 1.5 T magnetic resonance (MR)-linear accelerator MR-linac for imaging in gynaecologic high-dose-rate (HDR) brachytherapy.<i>Method</i>. Commissioning measurements for MR images quality control, geometric distortion, dwell position accuracy, applicator reconstruction and end-to-end test for a tandem-and-ring applicator were performed following the recommendations of American Brachytherapy Society, International Commission on Radiation Units and Measurements and Report of the Brachytherapy Working Group of the Spanish Society of Medical Physics. The values for MR-based IGABT were compared to the corresponding values with computed tomography (CT).<i>Results</i>. Measured distorsions for the MR images were less than 0.50 mm compared to the CT images. The differences between 3D displacements for all dwell positions were 0.66 mm and 0.62 mm for the tandem and ring, respectively. The maximum difference is 0.64 mm for the distances from the applicator tip obtained using the films. The CT and MR dose differences for the right and left 'A' points were 0.9% and -0.7%, respectively. Similar results were observed in terms of dose distribution for CT and Mr The gamma passing rate was 99.3% and 99.5%, respectively.<i>Conclusion</i>. The use of MR images from an MR-linac used in a radiotherapy service for gynaecological brachytherapy was proved to be feasible, safe and precise as the geometrical differences were less than 1 mm, and the dosimetric differences were less than 1% when comparing to the use of CT images for the same purpose.</p>","PeriodicalId":8896,"journal":{"name":"Biomedical Physics & Engineering Express","volume":null,"pages":null},"PeriodicalIF":1.3,"publicationDate":"2024-08-27","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"141900855","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":0,"RegionCategory":"","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
期刊
Biomedical Physics & Engineering Express
全部 Acc. Chem. Res. ACS Applied Bio Materials ACS Appl. Electron. Mater. ACS Appl. Energy Mater. ACS Appl. Mater. Interfaces ACS Appl. Nano Mater. ACS Appl. Polym. Mater. ACS BIOMATER-SCI ENG ACS Catal. ACS Cent. Sci. ACS Chem. Biol. ACS Chemical Health & Safety ACS Chem. Neurosci. ACS Comb. Sci. ACS Earth Space Chem. ACS Energy Lett. ACS Infect. Dis. ACS Macro Lett. ACS Mater. Lett. ACS Med. Chem. Lett. ACS Nano ACS Omega ACS Photonics ACS Sens. ACS Sustainable Chem. Eng. ACS Synth. Biol. Anal. Chem. BIOCHEMISTRY-US Bioconjugate Chem. BIOMACROMOLECULES Chem. Res. Toxicol. Chem. Rev. Chem. Mater. CRYST GROWTH DES ENERG FUEL Environ. Sci. Technol. Environ. Sci. Technol. Lett. Eur. J. Inorg. Chem. IND ENG CHEM RES Inorg. Chem. J. Agric. Food. Chem. J. Chem. Eng. Data J. Chem. Educ. J. Chem. Inf. Model. J. Chem. Theory Comput. J. Med. Chem. J. Nat. Prod. J PROTEOME RES J. Am. Chem. Soc. LANGMUIR MACROMOLECULES Mol. Pharmaceutics Nano Lett. Org. Lett. ORG PROCESS RES DEV ORGANOMETALLICS J. Org. Chem. J. Phys. Chem. J. Phys. Chem. A J. Phys. Chem. B J. Phys. Chem. C J. Phys. Chem. Lett. Analyst Anal. Methods Biomater. Sci. Catal. Sci. Technol. Chem. Commun. Chem. Soc. Rev. CHEM EDUC RES PRACT CRYSTENGCOMM Dalton Trans. Energy Environ. Sci. ENVIRON SCI-NANO ENVIRON SCI-PROC IMP ENVIRON SCI-WAT RES Faraday Discuss. Food Funct. Green Chem. Inorg. Chem. Front. Integr. Biol. J. Anal. At. Spectrom. J. Mater. Chem. A J. Mater. Chem. B J. Mater. Chem. C Lab Chip Mater. Chem. Front. Mater. Horiz. MEDCHEMCOMM Metallomics Mol. Biosyst. Mol. Syst. Des. Eng. Nanoscale Nanoscale Horiz. Nat. Prod. Rep. New J. Chem. Org. Biomol. Chem. Org. Chem. Front. PHOTOCH PHOTOBIO SCI PCCP Polym. Chem.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1